bioimaging &optics platformbiop.epfl.ch/pdf/quantitativeimaging.pdf · dr. arne seitz pt-biop...

25
Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform Quantitative imaging

Upload: others

Post on 12-Jul-2020

9 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

Quantitative imaging

Page 2: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

2

What is a quantitative imaging• Getting quantitative information about the specimen from the image• It is necessary to avoid subjective bias and to present the overall pattern of the data• Relative quantification – measure relative intensity• Absolute quantification – measure the number of photons• Thick specimens are difficult to quantify due to scattering and absorption• Important for the most of light microscopy methods which are intensity based (e.g. Colocalisation, FRET)

Page 3: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

Concentration of fluorophoresAbsorption: Calculation of absolute concentration of moleculesis easily possible

I0 IT

)lg()(0

TdcEIIT T

−=⋅⋅=

=

λε

Page 4: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

Concentration of fluorophores

Fluorescence: Calculation of concentration of molecules is difficult

I0 IF

),( 0IcfIF =

Absolute measurements of fluorophore concentration is almost impossible but relative measurements are possible. Comparable and reproducable imaging conditions are needed.

Page 5: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

5

Overview

• System calibration •Noise issues• Image processing

Page 6: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

6

System calibration

• AOTF calibration• Detector linearity• Gain calibration

Page 7: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

7

AOTF calibration

488 nm, PE RS• AOTF transmission is not linear • In single spot confocal it is normally more linear

Page 8: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

8

Signal linearity

• Normally no fluorophore saturation for spinning disc• More pronounced in single beam confocal• Depends on fluorophore coefficient of exstinction

PE RS, 1 µm beads , 488 nm excitation

Page 9: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

9

CCD linearity

1 µm beads , 488 nm excitation

• (EM)CCD is very linear

Page 10: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

10

Analog (amplifier) gain

• Normally linear in whole gain range• Sometimes used in confocals as well

Hamamatsu Orca ERG CDD camera

Page 11: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

11

PMT gain

Hamamatsu C9100-50 EMCCD camera600 700 800 900 1000

0

200

400

600

800

1000

1200

1400

1600

1800

2000

pixe

l inte

nsity

/ a.

u.

PMT gain / a.u.

Page 12: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

12

Electron multiplier gain

• For PMT and EMCCD gain is exponential• Typical max gain values: EMCCD ~103, PMT ~ 106

Hamamatsu C9100-50 EMCCD camera

Page 13: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

Confocal microscopeDwell time: 50 µs Dwell time: 6 µs Dwell time: 1.6 µs

Pixel intensity is not proportional to the collected photons !!!

Page 14: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

14

System calibration: summary

• System calibration is necessary for comparison between different images• Image with the same microscope settings if possible• Direct monitoring of excitation power is necessary for comparison of different systems• Check for absence of fluorophore saturation• Measuring system PSF is a good way to monitor the system performance • Calibration should be repeated regularly to monitor stability of the system

Page 15: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

15

Noise issues

• Role of noise• Noise and resolution• Main sources of noise•How to control noise

Page 16: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

16

Role of noise in quantification

• Detectable difference in intensity is approximately two times bigger than noise level • Example: SNR 5 (noise is 20% of signal). Only 40% difference in signal can be reliably detected.

1 µm beads , 488 nm excitation

SNR 10 SNR 5 SNR 3

Page 17: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

17

Noise decreases resolution

OTF for PlanApo 63x/1.4 objective, 500 nm emission

• SNR 5 results in about 30% decrease of resolution

Page 18: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

18

How to control noise

• More optimal filters• Better objectives• Accurate specimen preparation• Better detectors• Brighter fluorophores• Longer exposure/dwell time• Larger pixel size • Higher excitation power

Image quality depends not only on noise, but also on signal and background

Page 19: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

19

Noise issues: summary

• Noise effects both contrast and resolution• Average over several frames (e.g. time laps) or several pixels (e.g. roi in FRAP)• Get imaging system optimised• For noisy stacks use average instead of maximum intensity projection• Confocal images more noisy than brightfield• Average or accumulate noisy signal. This also helps to avoid saturation

Page 20: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

20

Image processing

• Background subtraction• Flat field correction• Photobleaching correction• Correct sampling• Filtering /deconvolution

Page 21: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

21

Background subtraction

Leica AF600063 x/1.3 Imm objective

original background background corrected

• Useful especially in transmission mode• Removes of dust particles, spots from the image

Page 22: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

22

Flat field correction

• Useful especially in fluorescence mode• Flat field can be measured with calibration slide• Subtract dark image before flat field correction

PerkinElmer ERS100 x/1.3 Oil objective

Page 23: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

23

Photobleaching correction

• Max bleaching: fixed sample - 50%, live sample - 20%• For strong bleaching SNR is different at the beginning and at the end of the series• Fluctuations of laser or lamp power are corrected similarly

Page 24: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

24

Image processing: summary

• Flatfield correction and background subtraction are already necessary for analysis within single image• For quantitation use images with at least 12 bit grey values• Use full dynamic range of detector with offset and gain settings preventing clipping or saturation• Use correct image sampling with pixel size 2 to 3 times smaller than resolution• To remove noise use Gaussian or mean filter with kernel size close to the resolution of the system

Page 25: BioImaging &Optics Platformbiop.epfl.ch/pdf/QuantitativeImaging.pdf · Dr. Arne Seitz PT-BIOP Course, Confocal Microscopy 2010, EPFL BioImaging &Optics Platform 24 Image processing:

Dr. Arne SeitzPT-BIOP Course, Confocal Microscopy 2010, EPFL

BioImaging &Optics Platform

25

Quantitative imaging: summary

• Quantitative imaging is a powerful tool to get unbiased/reproducable results•Quantitative imaging requires proper calibration of

•PMT•AOTF

• Image processing can be used to correct for:•Uneven illumination (flatfield correction)•Photobleaching•Background artefacts