ESA612 Electrical Safety Analyzer
VT MOBILE Gas Flow Analyzer
ESA620 Electrical Safety Analyzer
Impulse 7000DP Defibrillator/ Transcutaneous Pacemaker Analyzer
Biomedical Test Product Catalog
2009/2010
Fluke Biomedical.Better products. More choices. One company.
2
2009/2010
Biomedical TestProduct Catalog
Providing solutions, not just productsToday, biomeds, physicists, RSO’s, other medical personnel must meet increasing regulatory pressures, higher quality standards, and rapid technological growth, while performing their work faster and more efficiently than ever. Fluke Biomedical provides a diverse range of software and hardware tools to meet today’s challenges.
ServiceFluke Biomedical is dedicated to providing the best service within the healthcare industry. Equipped with the best-credentialed facilities, onsite experts, and full asset-management capabilities, Fluke Biomedical’s service team is always on call to take care of its customers. Fluke Biomedical’s world-class staff leads the industry in post- and pre-sale support, including helping customers choose the best products and accessories for their needs, technical support, product calibration, and repairs.
Regulatory complianceFluke Biomedical’s benchmark quality operates to the most rigorous standards in the industry, including compliance with ISO 9001:2000, ISO 13485:2003, FDA/QSR as applicable, and NRC/Part 50, Appendix B/Part 21 and adheres to ISO 17025:2005, ANSI Z540, Mammography MQSA and CNSC. Many of the Fluke Biomedical products are CE-marked and CSA-certified. In addition, the Global Calibration Laboratory holds its NVLAP Lab Code 200566-0 certification and is traceable to both the NIST & PTB.
LegacyYou may be familiar with some of our legacy brand names, including:
• Victoreen® • Metron• Nuclear Associates • DNI Nevada• Keithley • Bio-Tek Instruments
Fluke Biomedical has taken the best elements and products of these former brands and has incorporated them into the Fluke Biomedical culture and product line available today.
Our newest catalogOur Biomedical Test catalog emphasizes the complete line of biomedical test and simulation products for Biomedical/Clinical Engineers and Technicians. The catalog contains information about Fluke Biomedical’s test and simulation products, including standalone electrical safety testers, patient simulators, and per-formance analyzers, as well as fully integrated and automated performance-testing and documentation systems.
If you are interested in receiving catalogs or information about any of Fluke Biomedical’s other product-lines, please visit www.flukebiomedical.com/catalogs.
Catalogs are also available for the following product lines:
• Radiation Safety • Diagnostic Imaging QA • Radiation Oncology QA • Service
About Fluke BiomedicalFluke Biomedical leads the world in the manufacture of biomedical test and simu-lation products, including standalone electrical safety testers to fully integrated and automated performance testing and documentation systems. Fluke Biomedical also provides some of the most trusted and accurate radiation safety, medical imaging, and oncology quality-assurance solutions for regulatory compliance.
About Fluke CorporationFluke Biomedical is a division of Fluke Corporation. Fluke Corporation is the world leader in the manufacture, distribution, and service of electronic test tools and software and is a wholly owned subsidiary of Danaher Corporation (NYSE:DHR).
Fluke Biomedical6045 Cochran Road Cleveland, OH 44139-3303 U.S.A.
Fluke Biomedical EuropeScience Park Eindhoven 5110 5692EC Son, The Netherlands
For more information, contact usIn the U.S.A. (800) 850-4608 or Fax (440) 349-2307In Europe/M-East/Africa +31 40 267 5435 orFax +31 40 267 5436From other countries +1 (440) 248-9300 orFax +1 (440) 349-2307Email: [email protected] access: www.flukebiomedical.com
Service Center/Repair/ Calibration USFluke Biomedical6045 Cochran RoadCleveland OH 44139-3303Tel: 440-498-2560Toll free: 800-850-4608 ext 2564Email: [email protected]
Service Center/Repair/ Calibration EuropeFluke Biomedical EuropeScience Park Eindhoven 51105692EC Son, The NetherlandsTel: +31 (40) 267 5435Fax: +31 (40) 267 5436Email: [email protected]
Biomedical TestProduct Catalog
Contents
3
Defibrillator/Transcutaneous Paceer AnalyzersImpulse 6000D/7000DP Defibrillator/Transcutaneous Pacemaker Analyzer ..............4 QED 6 Defibrillator Analyzer ....................................................................................................8
Pacemaker AnalyzerSigmaPace 1000 External Pacemaker Analyzer ...............................................................10
Electrosurgery AnalyzersQA-ES Series II Electrosurgery Analyzer ............................................................................12RF303RS Electrosurgery Analyzer .........................................................................................14
Infusion Devise AnalyzerIDA 4 Plus Infusion Devise Analyzer ...................................................................................16
Electrical Safety AnalyzersESA620 Electrical Safety Analyzer .......................................................................................18ESA612 Electrical Safety Analyzer ...................................................................................... 20ESA601 Electrical Safety Analyzer ...................................................................................... 22601 Pro SeriesXL Electrical Safety Analyzer ...................................................................... 24QA-90 MKII Electrical Safety Analyzer ................................................................................26180 Electrical Safety Analyzer ............................................................................................. 28175 Electrical Safety Analyzer .............................................................................................. 29 LT544DLITE Digital Safety Tester ........................................................................................ 30 LT544DPLUS Digital Safety Tester ........................................................................................31
Ultrasound Transducer Leakage Tester ULT800 Ultrasound Transducer Leakage Tester ................................................................32
Non-Invasive Blood Pressure SimulatorBP Pump 2 Non-Invasive Blood Pressure Simulator ........................................................ 33CuffLink Non-Invasive Blood Pressure Simulator............................................................. 35
Patient SimulatorsMPS450 Patient Simulator ......................................................................................................37medSim 300B Patient Simulator .......................................................................................... 39PS420 Patient Simulator .........................................................................................................41PS415 Patient Simulator ......................................................................................................... 44PS410 Patient Simulator ......................................................................................................... 45PS400 Patient Simulator ........................................................................................................ 46 DataSim 6100 Patient Simulator ...........................................................................................47
ControllerHHC3 Hand Held Controller ................................................................................................... 48
Fetal Simulator PS320 Fetal Simulator ............................................................................................................ 49
Pulse Oximeter SimulatorIndex 2 Patient Oximeter Simulator ......................................................................................51
Gas Flow AnalyzersACCU LUNG Portable Precision Test Lung ...........................................................................52VT MOBILE Portable Gas Flow Analyzer ............................................................................ 53VT PLUS HF Gas Flow Analyzer ........................................................................................... 55
Pressure MetersDPM4 Parameter Tester ...........................................................................................................57DPM1B Pneumatic Transducer Tester .................................................................................. 59DPM2Plus Pressure Tester ..................................................................................................... 60
Automation SolutionsAnsur Test Automation Software ..........................................................................................61medTester 5000C Automated Biomedical Equipment Test System ............................. 63
Incubator AnalyzerINCU Incubator Analyzer ........................................................................................................ 65
Oxygen Analyzermax02 PLUS AE Oxygen Analyzer ....................................................................................... 66
Service and CalibrationService and Calibration Info............................................................................................... 67
Impulse 6000D/7000DP
Defibrillator/Transcutaneous Pacemaker Analyzer
Defibrillator/Transcutaneous Pacer Analyzers4
The Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer Test Systems are rugged, portable precision test instruments that ensure proper operation and ultimate perfor-mance of critical life-support cardiac-resuscitation equip-ment. The Impulse 6000D and Impulse 7000DP test capabili-ties encompass the spectrum
of worldwide-established pulse shapes, showcase breakthrough AED technology compatibility, and outperform in accuracy and standards. Additionally, the Impulse 7000DP incorporates the tests and the extensive range of test loads and measurement algorithms needed to test external transcutaneous pacemakers.
In conjuntion with an Impulse 7000DP, the Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω for defibrillator performance testing. A standard USB interface enables computer control and data transfer, and optional Ansur PC-based automation software increases productivity by outfitting users with an easy-to-use method to standardize testing procedures and cap-ture, print and document data.
N10140
Model QED 6 Impulse 6000D Impulse 7000DP
Monophasic and dc biphasic energy capability
Yes Yes Yes
Pulsed biphasic engery capability
No Yes Yes
Defibrillator tests Output energy Output energy Output energy
Cardioversion Cardioversion Cardioversion
Peak measurements Max energy/charge-time overshoot
Max energy/charge-time overshoot
— Peak and average current Peak and average current
— Voltage measurement Voltage measurement
Normal ECG/performance waves
No Yes Yes
Transcutaneous pacer tests No No Yes
Product comparison chart
Key features• Impulse 7010 Defibrillator
Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional)
• Lown, Edmark, trapezoidal, biphasic and pulsed biphasic defibrillation technology compatibility
• AED technology compatibility• First-class measurement
accuracy: ± 1 % of reading 0.1 J
• Intuitive user interface and backlight, easy-to-ready display
• Portable, rugged, easy to carry• Long-lasting, rechargeable
battery• Internal pacer brand
selections• Pacer input protected against
defibrillator output (7000DP only)
• 10 isolated ECG electrodes that provide 12 combina-tions for standardized clinical signals
• Flexible heart-rate settings (1 BPM step) facilitate rate meter accuracy and alarm testing
• DSP-based measurements enable future firmware and waveforms upgrade
• Unique integrated posts for secure connections
• Optional Ansur test automa-tion software to standardize testing procedures, capture waveforms, and print and document test result
4
Impulse 6000D/7000DP
Defibrillator/Transcutaneous Pacemaker Analyzer
Defibrillator/Transcutaneous Pacer Analyzers 5
Specifications
Defibrillator Analyzer
Energy output measurement Compatible defibrillator waveshapes: Lown, Edmark, trapezoidal, dc bi-phasic, and ac pulsed bi-phasic
Autoranged measurement 0.1 J to 600 J
Accuracy 0.1 J to 360 J: ± 1 % of reading +0.1 J360 J to 600 J: ± 1 % of reading +0.1 J, typicalNote: For pulsed bi-phasic defibrillator, specified accuracy is ± (1.5 % of reading + 0.3 J) on both ranges
Load resistance Resistence: 50 ΩAccuracy 1 %, non-inductive (< 2 μH)
Charge time measurement Range: 0.1 s to 100 s
Accuracy: ± 0.05 s, typical
Synchronization test (cardioversion)
Delay time measurement
• Timing window: ECG R-wave peak to the defib pulse peak• Range: -120 ms to 380 ms; measures timing from 120 ms prior
to the R-wave peak to up to 380 ms following the R-wave peak
Automated defibrillator test ECG waves
• Normal sinus: 10 BPM to 300 BPM in 1 BPM steps• Ventricular fibrilation: Coarse and fine• Monomorphic ventricular tachycardia: 120 BPM to 300 BPM
in 1 BPM steps• Polymorphic ventricular tachycardia: 5 types• Asystole
ECG waves
ECG general Lead configuration: 12-lead simulation; RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance 1000 ΩRate accuracy ± 1 % nominal
ECG amplitudes Reference lead: Lead II (default) or Lead I
Settings: 0.05 mV to 0.45 mV by 0.05 mV and 0.5 mV to 5 mV by 0.05 mV
Accuracy: ± 2 % of setting (Lead II), ± 5 % for all other leads and defib paddles
ECG normal sinus Rates: 10 BPM to 360 BPM in 1 BPM steps
ECG on defibrillator input load
Same as the Lead II amplitude but limited to ± 4 mV
ECG performance waves Square wave: 2 Hz and 0.125 Hz
Triangular wave: 2 Hz and 2.5 Hz
Sine waves: 0.05 Hz, 0.5 Hz, 5 Hz, 10 Hz, 40 Hz, 50 Hz, 60 Hz, 100 Hz, 150 Hz, and 200 Hz
Pulse: 30 BPM and 60 BPM, 60 ms pulse width
R-wave detection Waveform: Haver-triangle
Rate: 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
Widths: 8 ms, 10 ms, 12 ms, and 20 ms to 200 ms in 10 ms steps
Accuracy: ± 1 % setting 0.2 mV
Noise immunity Wave sine
Line frequency: 50 Hz or 60 Hz (± 0.5 Hz)
Amplitude: 0 mV to 10 mV (by 0.5 mV ± 5 %)
Arrhythmia selections Pacer interactive (Impulse 7000DP only)Supraventricular Premature Ventricular Conduction Transveneous paced with selectable pacer spike amplitudesand widths
5
Impulse 6000D/7000DP
Defibrillator/Transcutaneous Pacemaker Analyzer
Defibrillator/Transcutaneous Pacer Analyzers
Specifications
Transcutaneous Pacemaker Analyzer (Impulse 7000DP only)
Defibrillator input Fixed load: 50 ΩAccuracy: ± 1 %, non-inductive (< 2 μH)
Pacemaker input Variable load: 50 Ω to 1500 Ω by 50 ΩAccuracy: ± 1 %, non-inductive (< 2 μH)
Manufacturer specific algorithms
• Medtronic/Physio Control LIFEPAK• Philips/Agilent/HP• ZOLL Medical• GE Responder (1500 and 1700)• MRL/Welch Allyn• Schiller Medical• MDE300 (Medical Data Electronics), plus a general purpose
default algorithm selection
Current Range: 4 mA to 250 mAAccuracy: ± 1 % of reading +0.02 mA
Pulse rate Range: 5 PPM to 800 PPMAccuracy: ± 0.5 % of reading +0.1 PPM
Pulse width Range: 1 ms to 100 msAccuracy: ± 0.5 % of reading +0.01 ms
Demand and asynchronous mode test
Underdrive rate: 10 BPM minimumOverdrive rate: 300 BPM maximum
Sensitivity test Automatic interactive threshold detection
Compatible pacer rates: 30 PPM to 120 PPM
ECG R wave
Waveforms: Square, triangle, sine
Widths: 1 ms to 19 ms (by 1 ms), 20 ms to 95 ms (by 5 ms), 100 ms to 300 ms (by 25 ms)
Accuracy: ± 5 % of setting
Amplitude: 0.05 mV to 0.95 mV (by 0.05 mV), 1 mV to 5 mV (by 0.5 mV)
Accuracy: ± 5 % of setting
Refractory period tests Paced refractory period 20 ms to 500 msSensed refractory period 15 ms to 500 msAccuracy: ± 1 ms
General information
Dimensions (LxWxH) 32 cm x 24 cm x 13 cm (13 in x 9.5 in x 5 in)
Weight 3.02 kg (6.6 lb)
Standards
Safety standards CE: IEC/EN61010-1 2nd Edition; Pollution degree 2; CSA: CAN/CSA-C22.2 NO,61010-1, UL61010-1; C-Tick: Australian EMC
Optional accessories3091370 Ansur Impulse 6000D/7000DP Plug-In3065489 MedtronicERS/Physio-Control (FAST PATCH) (set of two): 4 mm defibrillator adapters3065450 Kimberly Clark/R2 Darox MRL/MDE/NK: 4 mm defibrillator adapters3065438 Internal discharge paddle contacts (set of two)3065477 Medtronic ERS/Physio-Control (QUIK PACE) (set of two): 4 mm pacer adapters3065527 Zoll Medical NTP/PD1400: 4 mm pacer adapters3065461 Medtronic ERS/Physio-Control (QUIK COMBO): 4 mm defib/pacer adapters3065492 Philips/Agilent/HP (CODEMASTER Series-Round): 4 mm defib/pacer adapters3065509 Philips/Agilent HEARTSTART FR2/MRX: 4 mm defib/pacer adapters3065511 Zoll PD-2200 Multi-Function PD-Series, M-Series, M-Series CCT, AED PRO® and AED Plus™ defib/pacer adapters3065423 GE Marquette (RESPONDER 1500/1700 Series) (set of two): 4 mm defib/pacer adapters3158544 Impulse 7010 Defibrillator Selectable Load Accessory
6
Defibrillator/Transcutaneous Pacemaker Analyzer
Impulse 6000D/7000DP
Specifications Included accessories3028681 Users Manual CD3028662 Getting Started GuideXXXXXXX Battery Eliminator (country specific)2814980 Carrying Case2795773 Defibrillator Paddle Contact Plates1626219 USB Computer Communication Cable
Ordering informationImpulse 6000D Defibrillator Analyzer2811928 United States, 120 V3077031 Schuko3077046 United Kingdom3077054 Japan3085270 Australia3085281 IndiaImpulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer2811919 United States, 120 V3077005 Schuko3077010 United Kingdom3077022 Japan3085296 Australia3085308 IndiaImpulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer with test automation3326874 United States, 120 V3326888 Schuko3326895 United Kingdom3326901 Japan3326912 Australia3326920 India
Defibrillator/Transcutaneous Pacer Analyzers
Impulse 7010 Defibrillator Selectable Load Accessory
Maximum voltage 5000 V
Maximum continuous power 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Inductance < 2 μH, @25 Ω < 3 μH, @50 Ω < 4 μH, @75 Ω and 100 Ω < 5 μH, @125 Ω < 6 μH, @150 Ω < 7 μH, @175 Ω < 8 μH, @200 Ω
Temperature Operating: 10 °C to 40 °C (50 °F to 104 °F) Storage: -20 °C to 60 °C (-4 °F to 140 °F)
Humidity 10 % to 90 % non-condensing
Dimensions (WxDxH) 154 mm x 272 mm x 138.7 mm (6.07 in x 10.71 in x 5.46 in)
Weight (net) 1.54 kg (3 lb 6.2 oz)
Safety class Complies with EN61010-1 2nd Edition, Class II product
Safety standards CE: IEC/EN61010-1 2nd Edition; Pollution degree 2; CSA: CAN/CSA-C22.2 NO,61010-1, UL61010-1; C-Tick: Australian EMC
Warranty Two-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center)
Calibration interval One-year
Electrical specifications (for load accessory and analyzer together)
Load settings 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω ± 1 %
Accuracy Energy (all except pulsed biphasic): 2 % of reading + 0.1 J with 25 Ω, 75 Ω though 200 Ω loads, 1 % of reading + 0.1 J with 50 Ω load
Energy (pulsed biphasic): 2.5 % of reading + 0.3 J with 25 Ω, 75 Ω though 200 Ω loads, 1.5 % of reading + 0.3 J with 50 Ω load
Voltage: 1 % of reading + 2 V with 25 Ω and 50 Ω loads, 2 % of reading + 2 V with 75 Ω through 200 Ω loads
Current: 2 % of reading + 0.1 A with 25 Ω load, 1 % of reading + 0.1 A with 50 Ω through 200 Ω loads
N10140
Discharge curve at 25 Ohms using Ansur and the 7010 load box. Discharge curve at 175 Ohms using Ansur and the 7010 load box.Note the differences in the shape, the peak currents and the time of the discharges.
7
QED 6
Defibrillator Analyzer
Defibrillator/Transcutaneous Pacer Analyzers
Specifications
Key features• Defibrillator analyzer• Monophasic and biphasic dc
energy measurement• Energy and cardioversion
measurement• Peak voltage, peak current,
and overshoot measurement• 2-line x 24-character display• Bidirectional RS-232 port for
computer control• Storage and playback of
output waveform so results can be viewed in greater detail
• Optional Ansur test automa-tion software to standardize testing procedures, capture waveforms and test results, and print and document test results
The QED 6 provides a scal-able solution to accurately test defibrillators. Lightweight and portable, the QED 6 measures a wide range of defibrillator energy output parameters.
An RS-232 serial port sup-ports remote computer control and test documentation.
Output energy test
Load 50 ± 1 %, with inductance < 70 μ h
Resolution High-range: 1 J Low-range: 0.1 J
Low-range 0 J to 100 J
High-range 0 J to 1000 J
Pulse width 1 ms to 50 ms
Maximum current Low: 35 A High: 110 A
Maximum voltage Low: 1750 V High: 5500 V
Minimum voltage Low: 20 V High: 66 V
Accuracy 1000 J range: ± 2 % of reading 100 J to 1000 J: ± 2 J 100 J Range: ± 2 % of reading, ± 0.1 J
Waveform storage Discharge viewable via ECG output, paddles, and scope output
Time expansion lead II amplitude High = 3000 V / mV Low = 900 V / mV
Peak/overshoot
Voltage accuracy 1000 J range: ± 10 V 100 J Range: ± 25 V
Current accuracy: ± 1 A
Cardioversion synchronization test Measurement from peak or base of simulated R-wave: 0 ms to 199.9 ms
Accuracy: 1 % of full scale or ± 2 ms, whichever is greater
88
Included accessories2204510 Operator’s Manual2204198 Internal Paddle Adapters
Ordering information2251457 QED 6 Defibrillator Analyzer
Optional accessories2817226 Ansur QED 6 Plug-in 2204282 Carrying Case2204472 RS-232/Printer Cable2248899 Serial/Parallel Printer2720054 9 V dc Adapter
Defibrillator/Transcutaneous Pacer Analyzers
QED 6
Defibrillator Analyzer
Defib waveform playback
Time base expansion 100:1 @ 25 mm/s paper speed, each division equals 40 ms
Amplitude scaling Lead II (RA-LL) 1000 J range:1 mV = 3000 V 100 J range:1 mV = 900 V
ECG output 1000 J range: 0.5 V = 3000 V 100 J range: 0.5 V = 900 V
Scope outputs
ECG hi-level Fixed at 1 V
Accuracy ± 2 %
Defib output Real time
Waveform output 5 ECG lead adapters, front-panel paddles, and high-level scope output
Calibration screen
Load 50 ± 1% (Apex-Sternum)
Amplitude scaling Apex (+) to sternum (-)
Zero voltage input 0 ± 2 counts
RS-232 output/computer control Computer control allows the user to operate the QED 6 remotely via a serial RS-232 interface. It requires an RS-232 cable and a bidirectional D-9 connector.
Selectable communications parameters
Baud rate 300, 600, 1200, 2400, and 9600
Parity None, even, odd
Stop bits 1 or 2
Data bits 7 or 8
Environmental requirements
Storage temperature -25 ˚C to 50 ˚C (-13 ˚F to 122 ˚F)
Operating temperature 0 ˚C to 40 ˚C (32 ˚F to 104 ˚F)
Maximum humidity 90 % relative humidity
General information
Display 2-line x 24-character super twist LCD
Power One 9 V alkaline battery or 9 V battery elimina-tor; 12 hours continuous operation; low-battery indication; 120/240 V battery eliminator input
Dimensions (WxDxH) 24.13 cm x 26.67 cm x 10.16 cm (9.5 in x 10.5 in x 4 in)
Weight 2.06 kg (4.54 lb)
9
Pacemaker Analyzer
Fluke Biomedical’s premier SigmaPace 1000 analyzes both transvenous and transcutaneous external pacemakers and comes loaded with features to save time and money. This powerful hand-held tool conducts the full suite of tests specified by major pace-maker manufacturers in less than half the time it would take using originally prescribed testing methods.
Output data is displayed on three selectable screens for easy viewing, including an AV delay time readout providing a performance snapshot for both pacer channels.
With capability for long-term trend testing, the SigmaPace
1000 can interrogate a pacer for up to 11 days, capturing data pulse by pulse to detect intermittent and hard-to-find problems.
For maximum efficiency, the SigmaPace 1000 doubles as a training tool. Interactive ECG simulation lets users test patient monitoring equipment as well as teach nurses how to operate the pacemaker.
Key features• Transcutaneous and
transvenous external pacemaker tests
• Pulse-output tests (rate, current, volts, energy, pulse width, and AV interval)
• Amplitude sensitivity and refractory tests
• Demand and asynchronous-mode tests
• DC load current test• Output-leakage tests• Line-frequency
noise-rejection tests• Wide range of test loads,
from 50 Ω to 1500 Ω, specified by manufacturer for transcutaneous pacers
• Full range of IEC specified test loads for transvenous pacers 200 Ω, 500 Ω, and 1000 Ω
• Pacer output displayed on three different screens
• AV readout showing both pacer channels on one screen
• Long-term trend test to detect intermittent errors and hard-to-find problems
• Interactive ECG pacer simulation with 5-lead output for patient monitor evaluation and pacer operation training
• 8-line x 21-character display
Specifications
ECG disposable snap electrode adapters
3.2 mm and 4 mm
Modes of operation Manual, remote
Transcutaneous pacer tests
Pulse output test Current: 4 mA to 250 mA Rate: 5 PPM to 300 PPMWidth: 1 ms to 99.9 msEnergy: 1 μJ to 1.99J
Demand model test
Async mode test
Amplitude sensitivity test
Noise immunity test
Paced refractory period test
Sensed refractory period test
Internal test loads: 31 selections: 50 Ω to 1550 Ω
SigmaPace™ 1000
External Pacemaker Analyzer
SigmaPace 1000 standard accessories
10
SigmaPace™ 1000
External Pacemaker Analyzer
Pacemaker Analyzer
Specifications
Transvenous pacer tests Pulse output testDisplay output test: (3) single (A or V) and dual (A+V)Current: 0.05 mA to 30 mARate: 10 PPM to 999 PPMWidth: 0.02 ms to 9.99 msVoltage: 0.05 V peak to 30 V peakEnergy: 1 nJ to 999 μJ
Demand model test
Async mode test
Amplitude sensitivity test
Noise immunity test
Paced refractory period test
Sensed refractory period test
AV delay time test
DC leakage current test
Measurement of dc offset on the pacemaker output
Test types: Static/continuous and dynamic/sync’d with output
Inputs/test loads: (3) 500 ΩDC current range: 0.1 mA to 99.9 mA
Battery load current testLoad current drawn by the pacemaker: 99.9 mA max
Transvenous measurement algorithm
Default: Derived from device manufacturer(s)
Available internal test loads
Ventricular and atrial channels: 200 Ω, 500 Ω, and 1000 ΩDefault selection: 500 Ω (both A and V channels)
Long-term trend test Tests the fundamental stability of the pacer outputTotal pulse count: 999,999 (max)Elapsed time: 999:59:59 (max)Maximum error count: 200 pulsesTest limits: Selectable rate and output percentages
Interactive pacer ECG simulation
Simulates dynamic patient ECG activity in response to pacer output: Interactive NSR heart rate: 0 BPM to 25 BPMNSR PR interval: 0.05 s to 0.399 sTransvenous threshold: 1 mA to 25 mATranscutaneous threshold: 10 mA to 250 mA
Serial port Type: RS-232Baud rate: 2400, 9600, and 192000
Power Internal lithium ion battery pack (rechargeable)Battery operation: 20 hours (minute)Universal/external ac-to-dc power supply
Dimensions (WxDxH) 10.1 cm 20.3 cm x 5 cm (4 in x 8 in x 2 in)
Weight 0.9 kg (2 lb)
Included accessories2243306 Users Manual2392906 Soft-sided Vinyl Carrying Case2201166 Transvenous Pacer Test Leads (2 sets, red)2201153 Transvenous Pacer Test Leads (2 sets, black)2392272 SigmaPace 9 V dc Load Test Cable2392260 Serial PC Interface Cable2184298 Universal-input Battery Charger2198724 Power cord set USA 120 V ac
Ordering informationSigmaPace 1000 External Pacemaker Analyzer2247700 United States, 120 V 2394548 Japan, 100 V 2394553 Schuko, 250 V 2394566 United Kingdom, 250 V
Optional accessories2245006 Electrode adapters (including the brand/model-specific interface connector and a pair of 4 mm “safety-typebanana plugs)2201109 Aligent (HP) CodeMaster Series2201111 GE Marquette Medical2201127 Medical Data Electronics (MDE); Medical Research Laboratories (MRL)2201095 Medtronic Physio-Control Quick Combo2201088 Medtronic Physio-Control Quick Pace2201323 Philips/Agilent Codemaster Series2201130 Zoll Medical NTP Series2201148 Zoll Medical PD Series and M Series2200102 Interface Cable (RS-232; female DB9 to female DB25; medTester to SigmaPace™ 1000/PC/Index 2XL/IDA 4 Plus; Impulse 4000 to PC)2201419 Detachable cord set—Japan (IEC 320 C6 type 3-pin inlet)2201437 Detachable cord set— Schuko—Euro (IEC 320 C6 type 3-pin inlet)2201428 Detachable cord set—UKI (IEC 320 C6 type 3-pin inlet)2201455 Detachable cord set—USA (IEC 320 C6 type 3-pin inlet)2201443 Detachable cord set— Australia (IEC 320 C6 type 3-pin inlet)
11
Electrosurgery Analyzers
Electrosurgery Analyzer
QA-ES Series II
QA-ES Series II analyzes electrosurgical units quickly and accurately.
A wide load-resistance range provides 128 user-selectable loads, including very low loads for testing many of today’s ESUs.
An accuracy of ± 2 % of reading down to 20 mA guarantees reliable high-frequency leakage results. With capability to run an automatic-power-distribution test in as little as 1 minute, the QA-ES works fast so technicians save time.
An Ansur QA-ES software plug-in allows users to create and automatically run tests, capture data, and produce easy-to-read reports with a PC.
Key features• Automatic power
distribution measurement, including power, current, peak-to-peak voltage (closed load only), and crest factor
• Oscilloscope output • High-frequency leakage
measurements with accuracy of ± 2 % of reading
• 128 internal user-selectable test loads from 10 Ω to 5200 Ω
• Foot-switch output for triggering the ESU under test
• Ansur QA-ES software plug-in for automated test protocols
• Large display• RS-232 and Centronic-Printer
interface
Specifications
Modes of operation
Continuous operation Continuous measurement of power, current, peak-to-peak voltage (closed load only), and crest factor
Single operation Single measurement after the set delay time of the ESU output of power, current, peak-to-peak voltage (closed load only), and crest factor
Power distribution Automatic measurement of power, current, peak-to-peak voltage (closed load only), and crest factor through a user-selectable load range
RF leakage current Provides connections and load configurations to measure HF leakage from both grounded and isolated equipment
RECQM Test the “return electrode control quality monitoringusing the QA-ES internal loads
Manual/remote via Ansur test automation software
12
Electrosurgery Analyzers
QA-ES Series II
Electrosurgery Analyzer
Specifications Optional accessories2461794 Carrying Case2461802 Ansur Test Software, QA-ES plug-in license2461993 Data Transfer Cable, RS-2322716059 QA-ES II Calibration Manual2523266 Clamp, crocodile style, grip C, black2523275 Clamp, crocodile style, grip C, red
Included accessories2716044 QA-ES Series II Users Manual (electronic, CD)2716032 QA-ES Series II Users Manual (printed)2772171 ESU-Dispersive Safety Lead2772180 ESU-CQM Safety Lead2772209 ESU-Jumper Safety Lead2826194 Test Lead with stackable plugs1903307 Test Lead Set with retractable sheaths1610159 Sure-Grip Large Alligator Clip SetXXXXXXX Power Cord (country specific)
Ordering informationQA-ES Series II Electrosurgery Analyzer2649769 United States, 115 V2651725 Schuko, 230 V2770445 United Kingdom, 230 V2770450 Australia, 230 V3096390 Japan, 100 V3319736 QA-ES, US 115V w/Test Automation3319749 QA-ES, SHK 230V w/Test Automation3319751 QA-ES, UK 230V w/Test Automation3319760 QA-ES, AUS 230V w/Test Automation3319772 QA-ES, JPN 100V w/Test Automation
Generator output
Load resistance (128 loads)
10 Ω to 2500 Ω in step of 25 Ω2500 Ω to 5200 Ω in step of 100 Ω
Measurement True-rms value of applied waveform
RMS bandwidth 30 Hz to 10 MHz (-3 dB) for instrumentation only30 Hz to 2.5 MHz (-3 dB) with loads
Low frequency filter 100 Hz filter to avoid low-frequency disturbance or interference
Current 20 mA to 2200 mA
Current accuracy ± 2 % of reading
Additional fixed load 200 Ω 400 W for 30 s; max 15 % duty cycle
Crest factor The higher of the two peak voltage measurements is used for computationRange: 1.4 to 16 (V peak/V rms).
Foot-switch output The foot switch output can be used to trigger the electrosurgical unit.
Peak-to-peak voltage 0 kV to 10 kV (closed load only) accuracy: ± 10 %Note: Measurement is taken between the active and dispersive electrodes with closed load only.
Oscilloscope output 5 V/A uncalibrated, 100 mA RF current minimum input
Ansur QA-ES plug-In remote control
All functions and tests in QA-ES may be performed from the PC
User-programmable test sequences
Allows unlimited numbers of test sequences with user-programmable templates and test limits. These tests include power distribution test, output test, HF leakage, and RECQM verification.
Storage and recall Protocol formats and data may be stored, recalled, printed out, or transferred.
General information
Display LCD graphic displayAlphanumeric format8 lines x 40 charactersGraphic mode240 x 64 pixel matrix
Display control Five f-keys, enter, cancel, control knob, and an encoder
Data input/outputs Parallel printer port and bidirectional RS-232
Power 115/230 V ac; 48 Hz to 66 Hz, 35 VA
Housing Metal case
Dimensions (LxWxH) 39.5 cm x 34.2 cm x 13.2 cm (15.6 in x 13.5 in x 5.2 in)
Weight 9.8 kg (21.6 lb)
Example of a power distribution curve created in 30 seconds with the Ansur QA-ES plug-in.
13
Electrosurgery Analyzers
Electrosurgery Analyzer
RF303RS
Key features• Easy to use simple configuration• Oscilloscope output,
high-frequency leakage, and return electrode contact quality monitor tests
• Instantaneous and signal-averaging measurement mode
• Ability to connect with Fluke Biomedical’s medTester 5000C for automated solution
• RS-232 port for computer control
• Battery powered• 4-digit numeric LCD with
backlight and power-save mode
RF303RS Electrosurgery Analyzer provides enough user-selectable test loads to do routine maintenance checks on most electrosurgery units on the market today. Compact and portable, the device is so simple to use that technicians can become proficient with the RF303RS within minutes.
The unit measures ESU output and high-frequency leakage, allows for verification tests on the return electrode contact quality monitors, and has an
oscilloscope output for waveform viewing. Instantaneous output or selectable sample times provide extra versatility. The instantaneous mode is sufficient for most units, but if output readings are variable and require stabilizing, the signal- averaging mode allows users to manually select two additional, slower sampling times to produce an accurate average reading.
Specifications
Modes of operation Line powered, battery powered, offline (battery maintenance charge)
Test parameters Power (W), HF current (mA), test load (Ω)
Tests performed generator output
HF leakage (performs to IEC 601 2-2, 1289-2, ANSI/AAMI standards)
Type BF test 1: Earth-referenced monopolar output
Type BF test 2: Earth-referenced monopolar output
Type CF/bipolar: Isolated monopolar or bipolar output
Current measurement (leakage)
Range: 30 mA to 2500 mA rmsResolution: 1 mAAccuracy: ± 2.5 % of reading or ± 15 mA (whichever is greater)
Power measurement (output)
Range: 1 W to 400 WResolution: 0.1 WAccuracy: ± 5 % of reading or ± 3 W (whichever is greater)
Bandwidth of rms converter circuit (1 % accuracy)
Flat response: 10 kHz to 10 MHz-3 dB points: 1 kHz to 20 MHz
Frequency response System response: -3 dB points, 1 kHz to 10 MHz at 300 ΩRECQM test 50 Ω to 750 Ω, 50 Ω steps
Test load section Number of selections: 15
Range: 50 Ω to 750 Ω
Step size: 50 ΩAccuracy (dc to 500 KHz): ± 4 % of selected value measured at calibration to ± 1 % (across the entire operating temperature range)
Duty cycle: 50 % @ 400 W (max 30 sec ON during any 1 minute period)
Resonance impedance variation: ± 0.5 dB max (< 10 MHz)
Auxiliary leakage test load
Fixed: 200 ΩAccuracy: ± 4 %Power rating: 225 W
Input capacitance(nominal)
Active to dispersive: 30 pFActive or dispersive to earth ground: 40 pF
Oscilloscope output Transformer coupled output, uncalibratedConnector type: BNC
14
Electrosurgery Analyzers
Electrosurgery Analyzer
RF303RS
Included accessories2202027 Users Manual2202009 Accessory Kit
Accessory kit includes the following:2200904 Active Safety Lead2200872 ESU Dispersive Safety Lead2200860 ESU CQM Safety Lead2200885 ESU Case Safety Lead2200897 ESU Jumper Safety Leads (2)2196071 Active Safety Clip, yellow2196080 Case Safety Clip, green2183792 Fuses (2) 5X20 F3.15A 250V CE2242165 Ground Pin AdapterXXXXXXX Detachable Power Cord (country specific)
Ordering informationRF-303RS Electrosurgery Analyzer2251504 United States, 120 V2394461 Australia, 250 V2394477 Denmark, 250 V2394489 Schuko, 250 V2394492 Israel, 250 V2394509 Italy, 250 V2394511 India, 250 V2394527 Switzerland, 250 V2394530 United Kingdom, 250 V
Specifications Optional accessories2248587 Multipurpose Hard-Sided, Watertight Carrying Case2204472 Serial Cable for D9F-D9F2238659 Interface Cable, medTester to RF303RS (RS-232; male DB9 to female DB9; adapter required, p/n 2391789)2391789 Adapter for Interface Cable, medTester to RF303RS (male DB9 to female DB25; used with interface cable, p/n 2238659)
Battery Type: Sealed lead-acid
Time between recharge: Two hours (continuous use)
Time to full charge: Eight hours
Number of cycles: 200
Capacity: 2.2 A H
Field serviceable: No
Recharging: Internal, automatic charger; power cord required
Front-panel controls/push buttons
Measurement select (1)
Load select: Increment test load (+) one step; decrement test load (-) one step
Top-panel input connections
Designations:• Generator output-active (1)• Generator output-dispersive (2)• Signal earth/ground reference (2)• Auxiliary HF leakage load (2)
Connector type: 4 mm (0.16 in) diameter safety sockets
Input voltage limit: 10,000 V peak
Input current limit: 3 A rms
Installation category: II
Side input connection Designation: Signal reference
Power requirements Universal input switching supply (12 V dc output)
Operating voltages:• Specified: 115 V ac/230 V ac• Max Range: 83 V ac to 264 V ac
Operating frequencies:• Specified: 50 Hz/60 Hz• Max range: 47 Hz to 63 Hz
Ventilation Internal fan with variable speed control; over-temperature detector; magnetic tachometer sensor to detect blocked fan rotor
Display LCD, 7-segment, 4 full digits, 2 in x 0.75 in
Case construction High-impact plastic, UL94-VO
Dimensions (WxDxH) 33.7 cm x 29.2 cm x 15.2 cm (13.25 in x 11.5 in x 6 in)
Weight 5.6 kg (14.2 lb)
Model RF303RS QA-ES Series II
Test loads 50 Ω to 750 Ω in step of 50 Ω 15 loads
10 Ω to 2500 Ω in step of 25 Ω, 2500 Ω to 5200 Ω in step of 100 Ω
128 loads
Displayed-result parameters
W, mA, Ω W, mA, Vpp, CF, Ω
High-frequency current 30 mA to 2500 mA RMS 20 mA to 2200 mA ± 2 % of reading
Accuracy: ± 2.5 % or reading or ± 15 mA, whichever is greater
20 mA to 2200 mA± 2 % of reading
Automation capabilities medTester 5000C Ansur
Additional benefits Battery operated Foot-switch output for triggering the ESU under test
Product comparison chart
15
Multi-Channel Infusion Device Analyzer
Multi-Channel Infusion Device Analyzer
IDA 4 Plus
Specifications
IDA 4 Plus Multi-Channel Infusion Device Analyzer maximizes produc-tivity with multiple, independent channels for testing upto four infusion pumps at once.
The device measures instantaneous flow, average flow, occlusion pressure, and analyzes patient-control analgesia (PCA) units. An optional PCA trigger box provides automated PCA pump control, allowing technicians to set up tests and walk away.
An autostart feature simplifies syringe pump testing or other tests that have long startup times.
With built-in memory, the IDA 4 Plus records test results internally and provides easy-to-read test-result graphs right on the analyzer’s screen. The display is so large numbers can
be read from across the roomAdditionally, the IDA 4 Plus comes with Hydrograph PC soft-
ware for creating full-color graphs and reports. For automated testing, the IDA 4 Plus is compatible with Fluke Biomedical’s medTester 5000C (US only).
Key features• Tests up to four infusion
pumps simultaneously• Compatible with virtually any
type of infusion device• Instantaneous and average
flow measurement• Occlusion pressure
measurements to 45 psi• Single- and dual-flow
(piggyback) testing• Full PCA testing (bolus volume,
lockout time, and automated external triggering)
• Autostart mode enables unit to begin testing only when fluid is detected
• On-board graphing of pressure and flow
• Built-in memory to save test results for printing or downloading to computer
• Hydrograph graphical software to control unit and display results via PC
• Automated testing through Fluke Biomedical medTester 5000C (US only)
• RS-232 ports• Optional keyboard, printer,
and alarm/PCA
Flow-rate measurement
Technique Calculated by measuring a volume over time
Range 0.5 ml/hr to 1000 ml/hr
Accuracy 1 % of reading ± 1 LSD for flows of 16 ml/hr to 200 ml/hr for volumes over 20 ml; otherwise, 2 % of reading ± 1 LSD after delivery of 10 ml
Volume measurement
Technique Volume measured directly by the transducer in minimum sample sizes of 60 μl
Range 0.06 ml to 9999 ml
Accuracy 1 % of reading ± 1 LSD for flows of 16 ml/hr to 200 ml/hr for volumes over 20 ml; otherwise, 2 % of reading ± 1 LSD after delivery of 10 ml
PCA bolus measurement
Technique Volume is measured directly by the transducer in minimum bolus volumes of 0.5 ml. The measurement is made with a continuous rate between 0 ml/hr and 30 ml/hr. The bolus flow rate should be at least four times the basal flow rate for reliable detection of boluses
Minimum bolus volume 0.5 ml
Accuracy See volume measurement
Pressure measurement
Technique Direct occlusion of the infusion line and measurement of pressure prior to the glass transducer
Range 0 psi to 45 psi and equivalents in mmHg and kPa
Accuracy 1 % of full scale ± 1 LSD
Back pressure -100 mmHg to 300 mmHg
Optional PCA Trigger Box
N10140
16
Multi-Channel Infusion Device Analyzer
Multi-Channel Infusion Device Analyzer
IDA 4 Plus
Specifications
Electrical specifications
Supply voltage Autoswitching 90 V ac to 260 V ac
Supply frequency 50 Hz to 60 Hz
Supply power < 30 VA
Fuse 20 mm 250 V, 1 A (T) (slow blow)
Earth leakage current < 1 mA in single fault condition
Environmental conditions
Operating temperature 15 °C to 30 °C (59 °F to 86 °F)
Storage temperature 0 °C to 40 °C (32 °F to 104 °F) at 85 % RH or less for storage (Do not leave for more than 48 hours at -20 °C/-4 °F)
General information
Dimensions (LxWxH) 19.05 cm x 18.11 cm x 30.18 cm (7.5 in x 7.2 in x 11.9 in)
Weight 5 kg (11 lb)
Optional accessories2245061 External mini-keyboard, 83-key with PS/2 connector and AT adapter2238072 Parallel Printer Cable (D25M-36M) 2209703 PCA Trigger/Nurse Call Box2248899 Printer, Seiko DPU-414-30B (120 V power supply) (additional purchase required: parallel printer cable, p/n 2238072)2399531 Printer, Seiko DPU-414-30B (220 V power supply) (additional pur-chase required: parallel printer cable, p/n 2238072)2235375 Printer (120 V power supply)2235382 Printer (220 V power supply)2200102 Interface Cable, medTester to IDA 4 Plus (without wedge adapter) (RS-232; female DB25 to female DB9)2201042 Interface Cable, medTester to IDA 4 Plus (with or without wedge adapter) (RS-232; female DB9 to female DB25)2245092 Barcode Scanner (with long-reach coil cable with Y connector for keyboard attachment)2238626 Null Modem Cable (female DB9 to female DB9)
Included accessories2213506 Electronic Users Manual and HydroGraph software2217231 20 ml Priming Syringe2391750 Luerlock-3 way (one for each channel)2238909 5-foot Plastic Drain Line2238626 Null Modem Cable (female DB9 to female DB9)XXXXXXX Detachable Power Cord (country specific)
Ordering informationIDA 4 Plus One-Channel Infusion Device Analyzer2250063 United States, 120 V2394575 Australia, 250 V2394582 Denmark, 250 V2394594 Shuko, 250 V2394608 Israel, 250 V2394613 Italy, 250 V2394624 India, 250 V2394636 Switzerland, 250 V2394649 United Kingdom, 250 V
IDA 4 Plus Two-Channel Infusion Device Analyzer Full testing for up to two infusion pumps simultaneously2250088 United States, 120 V2394651 Australia, 250 V2394660 Denmark, 250 V2394672 Shuko, 250 V2394685 Israel, 250 V2394697 Italy, 250 V2394703 India, 250 V2394715 Switzerland, 250 V2394726 United Kingdom, 250 V
IDA 4 Plus Three-Channel Infusion Device Analyzer Full testing capability for up to three infusion pumps simultaneously2250109 United States, 120 V2394732 Australia, 250 V2394744 Denmark, 250 V2394759 Shuko, 250 V2394767 Israel, 250 V2394771 Italy, 250 V2394780 India, 250 V2394798 Switzerland, 250 V2394800 United Kingdom, 250 V
IDA 4 Plus Four-Channel Infusion Device AnalyzerFull testing capability for up to four infusion pumps simultaneously2250127 United States, 120 V2394817 Australia, 250 V2394821 Denmark, 250 V2394839 Shuko, 250 V2394842 Israel, 250 V2394856 Italy, 250 V2394863 India, 250 V2394874 Switzerland, 250 V2394888 United Kingdom, 250 V
HydroGraph™ Graphics SoftwareUse the moving color visual advantage of HydroGraph to troubleshoot up to four infusion pumps at once. Data is taken directly off the transducer and transmitted to HygroGraph. The flowing graphs provide an electronic means to display, store, and recall flow patterns for comparison at a later date. Each test window can display instantaneous and average flow rates, cumulative, and bolus volumes; and occlusion pressure.
17
Electrical Safety Analyzers
ESA620
Electrical Safety Analyzer
The ESA620 Electrical Safety Analyzer represents the next generation in manual, portable electrical safety testers. With selections of three test loads, two protective earth test currents, and two insulation test voltages this versatile product can be used worldwide to enhance productivity and test to standards of choice.
New DSP technology offers better accuracy of leakage measurements throughout the ranges specified in the standards.
Equipped with ten safety-enhanced ECG posts, the ESA620 offers simulation of ECG and performance waveforms so both electrical safety and basic tests on patient monitors can be performed with a single connection. When used with optional Ansur computer-based software plug-in, the ESA620 becomes automated. This allows for standardization of test procedures, capturing and storage of results, comparison to standard limits, and printing of reports thus enabling the sophisticated performance of the high-end electrical safety analyzers.
N10140
Specifications
Voltage
Range (mains voltage) 90 V ac to 132 V ac rms, 180 V ac to 264 V ac rms
Range (accessible voltage) 0 V ac to 300 V ac rms
Accuracy ± (2 % of reading +2 LSD)
Earth resistance
Modes Two terminal or four terminal
Test current > 200 mA ac or 10 A ac to 25 A ac
Ranges 0 Ω to 2 ΩAccuracy ± (2 % of reading 0.015 Ω)
Equipment current
Mode AC rms
Range 0 A to 20 A
Accuracy ± 5 % of reading ± (2 counts or 0.2 A, whichever is greater)
Leakage current
Patient load selection (input impedance)
AAMI ES1-1993 Fig 1 IEC 60601: Fig 15 IEC 61010: Fig A-1
Crest factor ≤ 3
Ranges 0 μA to 199.9 μA200 μA to 1999 μA2.0 μA to 10.0 mA
Frequency response DC to 1 kHz 1 kHz to 100 kHz100 kHz to 1 MHz
Accuracy ± (1 % of reading + 1 μA)± (2 % of reading + 1 μA)± (5 % of reading + 1 μA)
Key features• Superior compliance with
multiple standards: IEC60601, IEC62353, VDE 751, ANSI/AAMI ES1:1993, NFPA-99, AN/NZS 3551, IEC61010
• Three test loads• Expanded leakage ranges
through 10,000 μA• Dual-lead resistance, leakage,
and voltage tests• AC only, dc only and true-rms
leakage readings• 100 % and 110 % mains
voltage for mains on applied part (lead isolation) test
• 200 mA and 25 A AC PE test current
• DSP filter technology for improved accuracy in leakage measurements
• 20 A equipment current• More applied parts
selections • ECG and performance
waveforms• Intuitive user interface• Easy-to-use applied parts
(ECG) connections• Insulation posts on applied
parts connections• Five different insulation tests• Varying insulation test voltage
500 V dc and 250 V dc• 2- or (optional) 4-wire ground
wire resistance• Large display with adjustable
contrast• Ergonomic design• Optional Ansur plug-in
software• USB connection• CE, C-TICK and CSA for USA
and Canada• RoHS compliance• Optional Ansur test automa-
tion software to standardize testing procedures, capture waveforms, and print and document test result
18
Electrical Safety Analyzers
ESA620
Electrical Safety Analyzer
Specifications
Leakage current (continued)
Mains on applied part test voltage
110 % of mains @ 230 V for IEC 60601100 % of mains @ 115 V per AAMI100 % of mains @ 230 V per 62353
Differential leakage
Ranges 10 μA to 199 μA200 μA to 1999 μA2 mA to 20 mA
Accuracy ± 10 % of reading ± (2 counts or .2 μA, whichever is greater)
Insulation resistance
Ranges 0.5 MΩ to 20 MΩ20 MΩ to 100 MΩ
Accuracy ± (2 % of reading + 2 counts)± (7.5 % of reading + 2 counts)
Source test voltage 500 V dc250 V dc
ECG performance waveforms
Accuracy ± 2 %± 5 % for amplitude of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
RatesECG complex (BPM): 30, 60, 120, 180, and 240
Ventricular fibrillation Square wave (50 % duty cycle) (Hz): 0.125 and 2Sine wave (Hz): 10, 40, 50, 60, and 100Triangle wave (Hz): 2Pulse (63 ms pulse width): 30 and 60
Power ratings
Mains voltage outlet 120 V ac230 V ac
Mains voltage inlet power range
90 V ac to 132 V ac rms180 V ac to 264 V ac rms
Maximum current 20 A @ 120 V ac16 A @ 230 V ac
Hz 50 or 60
Physical case
Dimensions (LxWxH) 31 cm x 23 cm x 10 cm (12.2 in x 9 in x 2.9 in)
Weight 4.7 kg (10.25 lb)
Certifications
Certifications CE: IEC/EN61010-1 2nd Edition; Pollution degree 2CSA: CAN/CSA-C22.2 No 61010-1; UL61010-1C-Tick: Australian EMC
Environmental
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity 10 % to 90 % non-condensing
Altitude To 2,000 meters @ 230 V ac (main supply voltage)
To 5,000 meters @ 120 V ac (mains supply voltage)
Included accessories2814967 Operator’s Manual CD2814971 Multilingual Getting Started GuideESA620 Accessory Kit (country specific)2195732 15 A to 20 A adapter (US only)2814980 Carry case1626219 Data Transfer cable3326842 Null Post assemblyXXXXXXX Detachable Power Cord (country specific) Test Lead Set
Ordering informationESA620 Electrical Safety Analyzer2785725 United States, 115 V, 20 A3051408 Europe, 230 V3051390 France, 230 V3051413 Israel, 230 V3051436 Australia, 230 V3051449 United Kingdom, 230 V3051451 Switzerland, 230 V3326935 ESA620, US 115V 20A w/Test Automation3326947 ESA620, EUR 230V w/Test Automation3326958 ESA620, FR 230V w/Test Automation3326964 ESA620, ISR 230V w/Test Automation3326986 ESA620, AUS 230V w/Test Automation3326999 ESA620, SWI 230V w/Test Automation3327002 ESA620, UK 230V w/Test Automation
Optional accessories3116463 Ansur ESA620 Plug-In1903307 Retractable Test Leads2242165 Ground Pin Adapter2067864 Kelvin Cable Set for 4-wire Measurement
19
Electrical Safety Analyzers
The ESA612 Electrical Safety Analyzer rep-resents the next generation in testers for biomedical professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.
With selection of two test loads, this ver-satile product can be used worldwide to test to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751), and AN/NZS 3551.
The versatility of the multifaceted ESA612 is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.
Key features• Portable, ergonomic, light-
weight and easy to use• Large, easy-to-read display
with adjustable contrast• Human-factors-designed user
interface• Tilt stand design for stand-up
testing in field environments• Five applied parts jacks and
easy ECG snap connection with optional expander box
• ECG waveform tests and dual-lead measurements combine the functionality of a simula-tor, multimeter and electrical-safety analyzer in a single test tool
• Replaceable mains fuses keep the device in the field and out of the repair shop
• Internal memory for 100 test records
• 20 A at 120 V current capability
• USB connection for use with Ansur and Data Viewer soft-ware (for memory download to PC)
• Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center)
• Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
• Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval
Electrical Safety Analyzer
ESA612
Voltage
Range (mains voltage) 90 V ac to 132 V ac rms, 180 V ac to 264 V ac rms
Range (accessible voltage) 0 V ac to 300 V ac rms
Accuracy ± (2 % of reading + 0.2 V)
Voltage tests Mains and point-to-point
Earth resistance
Mode Two terminal
Test current > 200 mA ac
Range 0 Ω to 2 ΩAccuracy ± (2 % of reading + 0.015 Ω)
Resistance tests Earth resistance and point-to-point
Equipment current
Mode AC rms
Range 0 A to 20 A
Accuracy ± 5 % of reading + (2 counts or 0.2 A, whichever is greater)
Duty cycle 15 A to 20 A, 5 min on/5 min off, 10 A to 15 A, 7 min on/3 min off, 0 A to 10 A continuous
Leakage current
Modes* AC + DC (true-rms)
AC only
DC only*Modes are available in all leakage tests with the exception of MAP leakages that are available only in true-rms.
Patient load selection (input impedance)
AAMI ES1-1993 Fig. 1, IEC 60601: Fig 15
Crest factor ≤ 3
Specifications
Optional accessories1903307 Retractable Test Leads2242165 Ground Pin Adapter (US receptacle testing ground lug)3392119 1210 Adapter Box Assembly3454829 Ansur ESA612 Plug-In License Key
20
Ranges 0 μA to 199.9 μA, 200 μA to 1999 μA, 2 mA to 10 mA
Frequency response/ accuracy
DC to 1 kHz ± (1 % of reading + (1 μA or 1 LSD, whichever is greater))
1 kHz to 100 kHz ± (2 % of reading + (1 μA or 1 LSD, whichever is greater))
1 kHz to 5 kHz (current > 1.6 mA)
± (4 % of reading + (1 μA or 1 LSD, whichever is greater))
100 kHz to 1 MHz ± (5 % of reading + (1 μA or 1 LSD, whichever is greater))
Note: Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are + (2.5 μA or 1 LSD, whichever is greater)
Leakage tests Ground wire (earth), Chassis (enclosure), Lead to ground (patient), Lead to lead (patient auxiliary), Lead isolation (mains on applied part), Direct equipment , Direct applied part, Alternative equipment, Alternative applied part, Point to point
Mains on applied part test voltage
100 % of mains
Differential leakage
Ranges 10 μA to 199 μA, 200 μA to 2000 μA, 2 mA to 20 mA
Accuracy ± 10 % of reading + (2 counts or 20 μA, whichever is greater)
Insulation resistance
Ranges 0.5 MΩ to 20 MΩ, 20 MΩ to 100 MΩAccuracy ± (2 % of reading + 0.2 MΩ), ± (7.5 % of reading + 0.2 MΩ)
Source test voltage 500 V dc , 250 V dc
Insulation resistance tests Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part)
ECG performance waveforms
Accuracy ± 2 %± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV Lead II configuration
Waveforms: rates Ventricular fibrillation
ECG complex (BPM): 30, 60, 120, 180, and 240
Square wave (50 % duty cycle) (Hz): 0.125 and 2
Sine wave (Hz): 10, 40, 50, 60, and 100
Triangle wave (Hz): 2
Pulse (63 ms pulse width): 30 BPM and 60 BPM
Power ratings
Mains voltage outlet 120 V ac or 230 V ac
Mains voltage inlet power range
90 to 132 V ac rms 180 to 264 V ac rms
Maximum current 20 A 16 A
Hz 50 or 60 50 or 60
Physical case
Dimensions (L x W x H) 17.63 cm x 8.38 cm x 28.45 cm (6.94 in x 3.30 in x 11.20 in)
Weight 1.6 kg (3.5 lb)
Environmental specifications
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity 10 % to 90 % non-condensing
Altitude 120 V ac mains supply voltage up to 5,000 m,230 V ac mains supply voltage up to 2,000 m
General
Warranty Two-year extended warranty (no-cost, available after first-year cali-bration at the Fluke Biomedical Cleveland Service Center, otherwise standard one-year warranty applies)
Electrical Safety Analyzers
ESA612
Electrical Safety Analyzer
Included accessories3334509 Operator’s Manual (mul-tilingual CD)3334511 Getting-Started Guide (hard copy, multilingual)2795488 Ansur ESA612 Plug-In, CD with demo version1626219 Data Transfer CableESA612 Accessory Kit (country specific)2195732 15 A to 20 A Adapter (US only)3326842 Null Post Adapter3359538 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter2248650 Carry CaseXXXXXXX Detachable Power Cord (country specific) Test Lead Set
Ordering information3367232 ESA612 Electrical Safety Analyzer (US), 115 V 20 A3367259 ESA612 Electrical Safety Analyzer (Europe), 230 V3367244 ESA612 Electrical Safety Analyzer (France), 230 V3367267 ESA612 Electrical Safety Analyzer (Israel), 230 V3367271 ESA612 Electrical Safety Analyzer (Australia), 230 V3367280 ESA612 Electrical Safety Analyzer (UK), 230 V3367298 ESA612 Electrical Safety Analyzer (Switzerland), 230 V3454793 ESA612 Electrical Safety Analyzer (Thailand), 230 V3461965 ESA612 Electrical Safety Analyzer (Japan), 100 V3460932 ESA612 Electrical Safety Analyzer (US), 115 V 20 A w/Test Automation3460959 ESA612 Electrical Safety Analyzer (Europe), 230 V w/Test Automation3460944 ESA612 Electrical Safety Analyzer (France), 230 V w/Test Automation3460967 ESA612 Electrical Safety Analyzer (Israel), 230 V w/Test Automation3460971 ESA612 Electrical Safety Analyzer (Australia), 230 V w/Test Automation3460980 ESA612 Electrical Safety Analyzer (UK), 230 V w/Test Automation3460998 ESA612 Electrical Safety Analyzer (Switzerland), 230 V w/Test Automation3461001 ESA612 Electrical Safety Analyzer (Thailand), 230 V w/Test Automation3462285 ESA612 Electrical Safety Analyzer (Japan), 100 V w/Test Automation
Specifications
21
Electrical Safety Analyzer
ESA601
Electrical Safety Analyzers
ESA601 Electrical Safety Analyzer tests laboratory and hospital equipment to both US and international standards. Users simply flip a switch to change between AAMI or IEC electrical safety testing load. The US version includes overlays in AAMI or IEC nomen-clature so technicians use the terms that are most familiar to them. Multiple language over-lays, outlets, and power cords are available for convenient use in many countries.
Ten applied part connections allow for lead-to-ground (patient), lead-to-lead (patient auxil-iary), and lead isolation (mains on applied part) leakage testing of equipment with multiple applied parts.
Designed for on-the-go testing, the portable analyzer is lightweight and compact and comes with a sturdy handle for easy carrying.
For an automated solution, Fluke Biomedical’s Ansur software plug-in for the ESA601 allows technicians to use a PC to run autosequences, document results, and print reports.
Key features• Selectable AAMI or IEC test
loads• Ten applied parts-lead
connectors• 90 V (min) to 264 V (max)
autoswitching power supply• Dual-lead leakage and
dual-lead voltage tests• Easy-to-read display• Compact and portable• Ansur plug-in software
available to automate testing and document results
• Multiple outlets and power cords for compatibility in multiple countries
• Overlays in English, German, French, and Italian
• Overlay for USA version with either IEC or AAMI nomenclature
• RS-232 serial port for PC control and printing
• Optional Ansur test automa-tion software to standardize testing procedures, capture waveforms, and print and document test result
Specifications
Power
Mains power operating range
90 min to 264 max V ac (47 Hz to 63 Hz), autoswitching
Power ratings 16 A at 264 V max
Mains inlet
Mains inlet The instrument uses a standard IEC 60320-1/C20 mains inlet rated at 16 A and 250 V for class-1 equipment in cold conditions.
Model 175 180 232D ESA620
Test modes
Manual Manual Manual Manual
— — — Automated with Ansur
— — — —
Test loads AAMI AAMI AAMI AAMI
— — — IEC60601-1
— — — IEC61010
PE test current 10 mA dc 10 mA dc 10 mA dc 200 mA ac25 mA ac
20A device testing No Yes No Yes
ECG waveforms no No Yes Yes
Product comparison chart
N10140
Optional accessories2556755 Ansur ESA601 Plug-in 2004175 Alligator Clamp2391669 Banana/ECG Adapter2248899 Printer, Seiko DPU-414-30B (with choice of 120 V power supply, p/n 2235375; or 220 V power supply p/n 2235382)2185787 North American 220 V Adapter Kit2238659 Interface Cable (RS-232; male DB9 to female DB9)
22
Electrical Safety Analyzer
ESA601
Electrical Safety Analyzers
Included accessories2388919 Users Manual and Ansur ESA601 Plug-in Demo Software on CD2243822 Getting Started Guide2248650 Soft-sided Carrying Case2391738 Red Lead2391723 Black Lead2391714 Alligator/Banana Adapters (five each)
Ordering informationESA601 Electrical Safety Analyzer2249883 Australian outlet, Australian line cord, and English 1 overlay2249909 Schuko outlet, Schuko line cord, and German overlay2249927 Schuko outlet, Schuko line cord, and French overlay2249948 Schuko outlet, Schuko line cord, and Italian overlay2434154 Israeli outlet, Israeli line cord, and English 1 overlay2404834 Schuko outlet, Schuko line cord, and English 1 overlay2249966 UK outlet, UK line cord, and English 1 overlay2249982 US outlet, US line cord, and English 1 overlay utilizing IEC nomenclature2404852 US outlet, US line cord, and English 2 overlay utilizing AAMI (NFPA99) nomenclature3334732 ESA601, 230V AUS w/Test Automation3334744 ESA601, DEU 230V SHK w/Test Automation3334759 ESA601, FRA 230V SHK w/Test Automation3334767 ESA601, ITA 230V SHK w/Test Automation3334771 ESA601, ISR 230V w/Test Automation3334780 ESA601, 230V SHK w/Test Automation3334798 ESA601, 230V UK w/Test Automation3334800 ESA601, 115V IEC w/Test Automation3334817 ESA601, 115 AAMI w/Test Automation
Specifications
Voltage measurements
Range (mains voltage) 90 V to 264 V true-rms
Range (accessible voltage) 0 V to 264 V true-rms
Accuracy ± 2 % of reading ± 2 V dc
Earth-resistance measurements
Range 0 Ω to 1.999 ΩAccuracy ± 2 % of reading ± 5 MΩCurrent source amplitude 1 A dc (± 10 %)
Insulation measurements
Range 0.5 M to 100 MΩAccuracy 0.5 MΩ to 20 MΩ, ± 2 % of reading, ± 200 kΩ;
above 20 MΩ, ± 5 % of reading, ± 200 kΩ
Voltage source amplitude 500 V dc (± 10 %)
VDE differential current
Range 10 μA to 10000 μA
Accuracy ± 2 % of full scale
Leakage-current measurements
Measurement range 0 μA to 8000 μA true-rms
Accuracy ± 1 % of reading ( ± 2 μA) @ dc and 25 Hz to 1000 Hz*;± 2.5 % of reading (± 2 μA) 1 kHz to 200 kHz* Full scale inputAccuracy of mains-on-applied-part leakage currents: ± 2 % of reading ± 6 μA
Frequency response: DC to 1 MHz (-3 dB)
Crest factor < 3
Input impedance Per figure 15 of IEC601-1 or AAMI ES 1
Dual lead leakage 500 μA to 8000 μA
Environmental specs
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -25 °C to 50 °C (-13 °F to 122 °F)
Maximum humidity 80 % relative humidity up to 31 °C (88 °F), decreasing linearly to 50 % relative humidity at 40 °C (104 °F)
Dimensions (LxWxH) 24.0 cm x 21.1 cm x 7.6 cm (9.8 in x 8.2 in x 3.0 in)
Weight 2.4 kg (5.2 lb)
Model ESA601 ESA601 Pro SeriesXL medTester 5000C
Test modes Manual Automatic
Automatic
Automated with Ansur Stepwise Manual
— Manual —
Test loads AAMI AAMI AAMI
IEC60601 IEC60601 —
— IEC61010 —
PE test current 1 A dc 1 A ac 10 A ac25 A ac
100 mA dc
20A device testing No No Yes
ECG waveforms No Yes Yes
23
Electrical Safety Analyzers
Electrical Safety Analyzer
601 Pro SeriesXL
Key features• IEC60601-1, IEC61010-1, and
ANSI/AAMI ES1 test loads, user selectable
• Multiple patient-applied-part types
• Power ON/OFF delays• DC-only current for patient-
and auxiliary-leakage tests• User-programmable test
sequences• Manual, auto, step, and
computer-control modes• 1 A, 10 A, or 25 A ac
protective-earth-resistance test circuit
• Memory for up to 1000 device-information records
• Integrated printer
The 601 Pro SeriesXL is the most advanced Electrical Safety Analyzer on the market. The One-Touch-Testing user interface is an industry exclusive that allows the user to perform rapid tests on various medical devices without having to maneuver around cumbersome menus. This full-featured safety analyzer combines the IEC60601-1, IEC61010-1, and ANSI/AAMI ES1 standard test loads into one device, so you can do all your testing at once.
Templates feature reduces your data entry, making your electrical-safety testing faster and easier!
Save the protective-earth test current at the default value you prefer, whether it is 1 A, 10 A, or 25 A. You are no longer limited to the 1 A default.
Specifications
Voltage (single and dual lead)
Insulation resistance
Protective earth resistance
Current consumption
Range 0 V to 300 V 0.5 MΩ to 400 MΩ 0 to 2.999 0 A to 15 A
Accuracy DC to 100 Hz ± 1.5 % of reading ± 1 LSD
± 5 % of reading ± 2 LSD
± 5 % of reading ± 4 mΩ (1 A, 10 A, and 25 A test currents)*
± 5 % of reading ± 2 LSD
*For additional specifications qualifying the varying effects on accuracy, please contact us.
N10140
24
Electrical Safety Analyzers
Electrical Safety Analyzer
601 Pro SeriesXL
Specifications
Included accessories2234222 Operator’s Manual2391723 Black Test-Lead Set 2391738 Red Test-Lead Set
Ordering informationESA601 Pro SeriesXL
2250323 Schuko receptacle, english overlay, with internal printer 2250314 Schuko receptacle, english overlay, without internal printer2250361 UK receptacle, english overlay, with internal printer2250350 UK receptacle, english overlay, without internal printer 2250306 Australian receptacle, english overlay, with internal printer 2250298 Australian receptacle, english overlay, without internal printer 2250389 601 Pro SeriesXL—USP: US receptacle, english overlay, with internal printer 2250377 US receptacle, english overlay, without internal printer 2250345 Schuko receptacle, german overlay, with internal printer 2250338 Schuko receptacle, english overlay, without internal printer
Optional accessories2248899 External Parallel Printer Port2245061 Compact Keyboard 2234065 Carrying Case2238659 RS-232 Cable2245092 Barcode Scanner (optical) 2238072 Parallel Printer Cable, D25M-C36M 2235375 120 V ac Adapter 2235382 220 V ac Adapter
IEC60601-1 and AAMI leakage current
Range 0 μA to 8000 μA true-rms or dc only
Accuracy (per IEC60601-1 or AAMI) DC to 1 kHz ± 1 % of reading ± 1 μA1 kHz to 100 kHz ± 2 % of reading ± 1 μA100 kHz to 1 MHz ± 5 % of reading ± 1 μA
Mains on applied part , equivalent device, and equivalent patient l eakage current tests the following exceptions apply
Applied voltage ≥ 110 % of mains voltage
Accuracy ± 2 % of reading ± 6 μA
IEC61010-1 leakage current
Range 0 μA to 16000 μA true-rms or dc only
Accuracy (per IEC 61010-1 Fig. A.1 filter)DC to 1 kHz ± 2 % of reading ± 1 μA1 kHz to 100 kHz ± 4 % of reading ± 2 μA100 kHz to 1 MHz ± 10 % of reading ± 10 μA
Waveform simulation
Normal sinus rhythm 30 BPM, 60 BPM, 120 BPM, 180 BPM, and 240 BPM
Performance pulse 30 BPM and 60 BPM
Sine 10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz
Square 0.125 Hz2 Hz (50 % duty cycle)
Triangle 2 mV2 Hz
Arrhythmia A-FibA-FlutterA-TACIdioventricularPVC1R-on-TRunV-FibV-Tach
Performance ± 2 % of reading for rate of ± 5 % of reading for amplitude, fixed at 1 mV peak on a Lead II ECG connection (except for triangle wave, which is 2 mV peak to peak)
General
Power Autoswitching, 90 V ac to 265 V ac
Dimensions (LxWxH) 42.2 cm x 30 cm x 14.1 cm (16.62 in x 11.75 in x 5.56 in)
Weight 7.7 kg (17 lb)
25
Electrical Safety Analyzers
QA-90 MKII
Electrical Safety Analyzer
Key features• Small size—easy to carry• Can operate completely stand-
alone or by remote control with the ansur software
• Tests instruments containing modules with different protec-tion classes in one test run i.e. CF and BF (defibrillator)
• Tests module separation• Residual current measure-
ments (optional)• 11 patient inputs (laboratory
style)• Test according to: IEC
60601.1, IEC 60601.1.1, IEC 60601.2.4, IEC 61010, VDE 751-1:1990, VDE 751-1: 2001, UL 2601.1, AS 3200.1, DB 9801, HEI 95, AAMI
• Testing against standard or user defined limits
• Automatic, stepwise and manual test modes
• Programmable power up delay time before measurements
• Integrated keypad• Internal memory for uploading
test sequences and storing typically up to 200 equipment test results
• Built-in standard A4 protocols for on site documentation (localized)
• RS-232, centronics and bar code interface
• Optional Ansur test automa-tion software to standardize testing procedures, capture waveforms, and print and document test result
The QA-90 represents a new generation of safety testing equipment. QA-90 is the only automatic safety tester that can test Instruments containing modules with different classes of protec-tion in one test run (i.e. CF and BF defibrillators). It is simple to use. All you need
to do is select the type and class of the equipment to be tested. When you press START, the QA-90 will execute the tests prescribed in the selected standard.
The test result can either be printed out immediately or stored internally in the unit for later use. The QA-90 can be fully remotely controlled via ansur software. With ansur, you can make your own test protocols, store the informa-tion on disk and export formatted data to equipment management databases.
Specifications
Test sequence
Individual test sequences may be compiled to satisfy national and international standards: IEC 60601.1, VDE 0750 T1/12-91, BS 5724, UL 2601.1, CAN/CSA-C22.2 No 601.1-M90, AS 3200.1, NZS 6150:1990, VDE 0751:1990, VDE 0751:2001, IEC 60601.1.1, IEC 60601.2.4, UL 544, HEI 95, DB9801 and more
Voltage measurement
The voltage measurement may be executed in the following ways
– Between lead 1 and 2 – Between lead 1 and Earth – Between lead 2 and Earth – Between input/output E+ and E- (floating inputs/outputs)
Range 0 V to 400 V true-rms
Resolution 0.1 V
Accuracy DC to 100 Hz 1 % of full scale ± 1 LSD 100 Hz to1 kHz 2 % of scale ± 1 LSD
Number of tests Four or multiple
Current consumption
The current measurement may be executed in lead no. 1 (live)
Range 1 0 mA to 1000 mA true-rms (@ < 250 V ac)
Resolution : 1 mA
Accuracy : ± 2 % of full scale ± 1 LSD
Number of tests: 1 or multiple
Range 2 1 A to 16 A true-rms (@ < 250 V ac)
Resolution: 1 mA
Accuracy: ± 1 % of full scale ± 1 LSD
Number of tests: 1 or multiple
Protective earth
The test current is 25 A or 1 A, delivered from a transformer with a maximum idle voltage of 6 V. The measurement can be performed on ground leads or between E+ and E- (floating inputs/ outputs)
Range 0 MΩ to 2000 MΩResolution 1 MΩAccuracy ± 2 % of full scale or 5 % of reading
Number of tests 1, 2, or multiple
Insulating resistance
The measurement of the insulating resistance may be executed between casing and power unit, or between patient module and power unit
Test voltage 500 V dc through a 130 kΩ limiting resistor
Number of tests 1, 2, or multiple
26
Electrical Safety Analyzers
Optional accessoriesData transfer cable, RS-232
2462909 Carrying Case 2463016 Ansur test software, QA-90 plug-in license
Electrical Safety Analyzer
QA-90 MKII
Specifications
Range 1 1 MΩ to 50 MΩResolution: 1 MΩAccuracy: ± 2 % of full scale ± 1 LSD
Range 2 5 MΩ to 200 MΩResolution: 1 MΩAccuracy: ± 2 % of full scale ± 1 LSD
Leakage currents
All measurements can be performed with a IEC 601.1 filter (patient equivalent), or without (flat frequency response). The filter can be exchanged with filters covering other standards. All measurements can be performed as true-rms measurements, or ac/dc measurements.
Leakage currents are measured
Number of tests: Earth leakage current – 4 Enclosure leakage current – 6 or multiple Residual current – 2 Substitute equipment current – 1 Direct current – 2 Current Fig. 9 – 1
Leakage currents are measured for each module In one test run a maximum of 11 modules with differ-ent protection classes may be tested
Number of tests: Patient leakage current – 6 Mains on applied part leakage current – 2 Patient Auxiliary current – 6 Floating dual lead measurement of leakage currents – Multiple Substitute patient current – 1
Accuracy
Range 1 0 to 99.9 μAResolution: 0.1 μAAccuracy: ± 2 % of full scale ± 1 LSD
Range 2 100 μ to 1000 μAResolution: 1 μAAccuracy: ± 2 % of full scale ± 1 LSD
Range 3 1.0 mA to 10.0 mAResolution: 1 μAAccuracy: ± 1 % of full scale ± 1 LSD
Frequency response
Dc to 1 MHz (-3 dB) with a crest factor: > 2. The test voltage for the mains on applied part measurement is 110 % of the line voltage, delivered through a limiting resistor of 47 kΩ.Ansur software
Remote control All functions and tests in QA-90 may be performed from the PC
User-definition of test standards
Select predefined standards or create your own local/new standard with test limits
User-programmable test sequences
Allows user-defined test sequences with a selection of tests from the selected test standard
Customized protocols Create your own protocol format including a header, checklist, job instructions, a command field and a test sequence
Storage and recall Protocol formats and data may be stored, recalled, printed out, or transferred to D-base systems
General information
Temperature requirements + 15 °C to +35 °C while operating; 0 °C to +50 °C for storage
Display Type: LCD Alphanumeric format: Four lines by 40 characters Display control: 7 F-keys and a keypad
Data input/outputs (2) Parallel printer port (1); bi-directional RS-232 (1) for computer control Bar code interface
Power From 100 V ac to 240 V ac, 47 Hz to 63 Hz
Housing Metal case
Dimensions (LxWxH) 305 mm x 342 mm x 132 mm
Weight 5.8 kg
Recommended printers HP Desk Jet, Canon Bubble Jet or compatible
Included accessories2462072 Banana Adapter, 10 pk2462975 E-Input Measuring Cable, 2 m, black2523266 Clamp, Crocodile Style, blackQA-90-1CD Users Manual on CDXXXXXXX Detachable Power Cord (country specific)
Ordering informationQA-90 MKII Electrical Safety Analyzer2462865 United States, 115 V2462876 Europe, 230 V2556858 United Kingdom, 230 V2557155 Australia, 230 V3277167 Denmark, 230 V
27
Electrical Safety Analyzers
Key features• Handheld • Self-switching 120 V and
240 V operation• 15 A and 20 A capabilities• Five patient applied parts
jacks• Dual-lead testing
Included accessories2185754 Users Manual2392409 Test Lead 8 foot, black
Ordering information2249852 180 Electrical Safety Analyzer, US, 90 V to 265 V
Optional accessories2248864 Soft-sided Vinyl Carrying Case2392409 Chassis Cable, coil cord, 8-foot extended2392411 Chassis Cable, coil cord, 16 foot extended2392448 External Leakage Cable, coil cord, 8 foot extended2231563 External Leakage Cable, coil cord, 16 foot extended2185787 North American 220 V adapter kit
Electrical Safety Analyzer
180
Specifications
The handheld 180 Electrical Safety Analyzer is a lightweight, portable device for testing electrical systems, medical devices, and physiological instrumentation. The device includes the AAMI test load and has five jacks for patient applied parts testing.
Small enough to fit in a briefcase, the analyzer works well as a bench-top instrument in the laboratory or a portable testing device in the mobile engineer’s toolbox. With its uncomplicated design, the 180 is simple to use.
A single master function switch, directly labeled with the test to be performed, leads the user through a complete measurement procedure.
The analyzer uses simple, yet sophisticated, electronics for true-rms measurement of current and voltage. The 180 also performs dual-lead leakage and resistance tests.
Operating mains voltage range
90 V ac to 240 V ac
Current capacity for DUT Line 90 V to 140 V: 20 A for 5 min, 15 A for 30 min; line over 140 V: 10 A
Line-voltage measurement
Range 90 V to 240 V
Accuracy ± 3 % of reading ± 1 LSD
Load-Current measurement
Range 1 A to 19.99 A
Accuracy ± 5 % of reading ± 1 LSD
Leakage-current measurement
Range 0 μA to 1999 μA
Accuracy DC and 25 Hz to 1 kHz: ± 1 % of reading ± 1 μA; 1 kHz to 100 kHz: ± 2.5 % of reading ± 1 μA; 100 kHz to 1 MHz: ± 5 % of reading ± 1 μA
Measurement type True-rms; input impedance per AAMI ES1-1993
Isolation test
Isolation source voltage 110 % of mains, ± 5 % of reading
Current limit 1 mA @ 120 V ac
Resistance measurement
Range 0.01 Ω to 19.99 ΩAccuracy ± 1 % of reading ± 1 LSD
Resolution 0.01 ΩCurrent source 10 mA dc
Environmental requirements
Operating temperature 15 °C to 40 °C (59 °F to 104 °F)
Storage temperature -20 °C to 65 °C (-4 °F to 149 °F)
Relative humidity 90 % max
Mains voltage range 90 V to 240 V
General information
Display LCD 3.5 digit
Dimensions (LxWxH) 13.3 cm x 18.4 cm x 5.4 cm (5.25 in x 7.25 in x 2.1 in)
Weight Max weight 1 kg (2.25 lb)
N10140
28
Electrical Safety Analyzers
Electrical Safety Analyzer
175
Included accessories2185754 Operator’s Manual2392409 Test Lead 8 foot, black
Ordering information2249841 175 Electrical Safety Analyzer, US, 90 V to 265 V
Optional accessories2248864 Soft-sided Vinyl Carrying Case2392409 Chassis Cable, coil cord, 8-foot extended2392411 Chassis Cable, coil cord, 16 foot extended2392448 External Leakage Cable, coil cord, 8 foot extended2231563 External Leakage Cable, coil cord, 16 foot extended2185787 North American 220 V adapter kit
Key features• Portable • IEC601-1 and AAMI test loads• Self-switching 120 V and
240 V operation• Dual-lead testing
Specifications
The 175 Electrical Safety Analyzer is ideal for performing quick elec-trical safety checks on electrical systems, medical devices, and physiological instrumentation.
Small enough to fit in a briefcase, the analyzer works well as a bench-top instrument in the laboratory or a portable testing device in the mobile engineer’s toolbox.
With its uncomplicated design, the 175 is simple to use. A single master function switch, directly
labeled with the test to be performed, leads the user through a complete measurement procedure. The unit features both IEC601-1 and AAMI test loads. Technicians simply flip a switch to perform leakage measurements to a particular standard.
The analyzer uses simple, yet sophisticated, electronics for true-rms measurement of current and voltage. The 175 also performs dual-lead leakage tests.
Operating mains voltage range
90 V ac to 240 V ac
Current-capacity for DUT 15 A for 10 min
Voltage measurement
Range 90 V to 240 V
Accuracy ± 3 % of reading ± 1 LSD
Load-current measurement
Range 1 A to 19.99 A
Accuracy ± 5 % of reading ± 1 LSD
Resistance measurement
Range 0.01 Ω to 19.99 ΩAccuracy ± 1 % of reading ± 1 LSD
Resolution 0.01 ΩCurrent source 10 mA dc
Leakage-current measurement
Range 0 μA to 1999 μA
Accuracy DC and 25 Hz to 1 KHz: ± 1 % of reading ± 1 μA; 1 KHz to 100 KHz: ± 2.5 % of reading ± 1 μA; 100 KHz to 1 MHz: ± 5 % of reading ± 1 μA
Measurement type True-rms
Input impedance 1000 Ω per AAMI ES1-1993, IEC601-1
Environmental requirements
Operating temperature 15 °C to 40 °C (59 °F to 104 °F)
Storage temperature -20 °C to 50 °C (-4 °F to 122 °F)
Relative humidity Max 80 % up to 31 °C (88 °F), decreasing linearly to 50 % at 40 °C (104 °F)
General information
Display LCD 3.5 digit
Dimensions (LxWxH) 13.3 cm x 18.4 cm x 5.4 cm (5.25 in x 7.25 in x 2.1 in)
Weight Max weight: 1 kg (2.25 lb)
N10140
29
Electrical Safety Analyzers
Included accessories600/100 Chassis Cable, coil cord, 8 ft 600/101 Chassis Cable, coil cord, 16 ft 600/102 Chassis Ground Probe, 8 ft 600/215 Conductivity Cable – PCI
600/600 Soft Carrying Case (6 in x 10 in x 4 in)
Ordering informationLT544DLITE Digital Safety Tester2250751 United States, 85 V ac to 265 V ac, 50/60 Hz2250805 Europe, 85 V ac to 265 V ac, 50/60 Hz2250797 Australia, 85 V ac to 265 V ac, 50/60 Hz2250810 Israel, 85 V ac to 265 V ac, 50/60 Hz2250822 Italy, 85 V ac to 265 V ac, 50/60 Hz2250831 Switzerland, 85 V ac to 265 V ac, 50/60 Hz2250846 United Kingdom, 85 V ac to 265 V ac, 50/60 Hz
Key features• Small, lightweight, self-
contained portable instrument• Universal power supply –
input voltage from 85 V ac to 265 V ac
• Chassis resistance measure-ments are made with a four-wire Kelvin bridge to eliminate errors due to cable length
• Test method complies with AAMI load per ANSI/AAMI ES1-1993
• Leakage current can be measured in Open/Closed Neutral, Normal/Reversed Polarity and Open/Closed Ground
• All Leakage current measure-ments are displayed directly in μA
Digital Safety Tester
LT544DLITE
Specifications
This device is designed to perform simple electrical safety tests on any type of medical equipment, whenever patient lead testing is not required.
The rugged, hand-held instrument is built for quick and easy use. A simple selector knob controls the functions: Leakage Current and Chassis Resistance. Accurate test measurements are shown on the large display.
Heavy-duty switches change polarity, open and close the
neutral and select Chassis or Earth Leakage Current measurements. A four-wire Kelvin bridge eliminates reading errors in resistance due to cable length and contact resistance. True-rms measurements are provided for all current readings. The AAMI test load is utilized.
Ground resistance
Range 0 Ω to 19.99 ΩAccuracy ± 1% R, ± 1 LSD
Resolution 10 mΩLeakage current
Range 0 μA to 1, 999 μA
Accuracy DC and 25 Hz to 1kHz: ± 1 % of reading, ± 3 LSD 1.0 KHz to 100 KHz: ± 2.5 % of reading, ± 3 LSD 100 KHz to 1 MHz: ± 5 % of reading, ± 3 LSD
Current capacity
LT544DLite- 1515 15 A at 120 V ac
LT544DLite- 230 V 10 A at 230 V ac
LT544DLite- 1520/2020 20 A, at 230 V ac, 20 % duty cycle (2 minutes on, 8 minutes off)
Controls 1. Function switch – 4 position rotary 2. Neutral switch – 2 position rocker (Open/Closed) 3. Polarity switch – 3 position rocker (Normal/Off/Reversed) 4. Leakage switch – 2 position rocker (Chassis/Earth)
DUT power 85 V ac to 265 V ac, 16 A @ 120 V, 10 A @ 230 V
Physical size 90 mm x 180 mm x 38 mm (3.5 in x 6.2 in x 1.6 in)
30
Digital Safety Tester
LT544DPLUS
Electrical Safety Analyzers
Included accessories600/100 Chassis Cable, coil cord, 8 ft 600/101 Chassis Cable, coil cord, 16 ft 600/102 Chassis Ground Probe, 8 ft 600/215 Conductivity Cable – PCI
600/600 Soft Carrying Case (6 in x 10 in x 4 in)
Ordering informationLT544DPLUS Digital Safety Tester2250854 United States, 85 V ac to 265 V ac, 50/60 Hz2250893 Europe, 85 V ac to 265 V ac, 50/60 Hz2250887 Australia, 85 V ac to 265 V ac, 50/60 Hz2250902 Israel, 85 V ac to 265 V ac, 50/60 Hz2250916 Italy, 85 V ac to 265 V ac, 50/60 Hz2250925 Switzerland, 85 V ac to 265 V ac, 50/60 Hz2250933 United Kingdom, 85 V ac to 265 V ac, 50/60 Hz
Key features• Small, lightweight, self-
contained portable instrument• Universal power supply –
input voltage from 85 V ac to 265 V ac
• Chassis resistance measure-ments are made with a four-wire Kelvin bridge to eliminate errors due to cable length
• Test method complies with AAMI load per ANSI/AAMI ES1-1993
• Leakage current can be measured in Open/Closed Neutral, Normal/Reversed Polarity and Open/Closed Ground
• All Leakage current measure-ments are displayed directly in μA
Specifications
This device is designed to perform simple electrical safety tests on any type of medical equipment, whenever patient lead testing is not required.
The rugged, hand-held instrument is built for quick and easy use. A simple selector knob controls the functions: Line Voltage, Instrument Current, Leakage Current and Chassis Resistance. Accurate test measurements are shown on the large display.
Heavy-duty switches change polarity, open and close the neutral and select Chassis or Earth Leakage Current measurements. A four-wire Kelvin bridge eliminates reading errors in resistance due to cable length and contact resistance. True-rms measurements are provided for all current readings. The AAMI test load is utilized.
Line voltage
Range 85 V ac 265 V ac
Resolution 1 Volt
Accuracy 2 % R ± 1 LSD
Instrument current
Range 0.0 A to 19.99 A
Resolution 10 mA
Accuracy 5 % R ± 1 LSD
Ground resistance
Range 0 W 19.99 W
Accuracy ± 1 % R, ± 1 LSD
Resolution 10 mW
Load current
Range 0 A to 14.99 A (LT544Dplus – 1515) 0 A to 19.99 A (LT544Dplus – 2020)0 A to 9.99 A (LT544Dplus – 230 V)
Accuracy ± 4 % R, ± 1 LSD
Leakage current
Range 0 μA to 1, 999 μA
Accuracy DC and 25 Hz to 1 kHz: ± 1 % g, ± 3 LSD1.0 KHz to 100 KHz: ± 2.5 % of reading, ± 3 LSD100 KHz to 1 MHz: ± 5 % of reading, ± 3 LSD
Current capacity LT544Dplus-1515: 15 A at 120 V acLT544Dplus-230V: 10 A at 230 V ac LT544Dplus-2020: 20 A, at 120 V ac, 20 % duty cycle (2 minutes on, 8 minutes off)
Controls 1. Function Switch - 4 position rotary 2. Neutral Switch - 2 position rocker (Open/Closed) 3. Polarity Switch - 3 position rocker(Normal/Off/Reversed) 4. Leakage Switch - 2 position rocker (Chassis/Earth)
DUT power 85 V ac to 265 V ac, 16 A @ 120 V, 10 A @ 230 V
Power connectors LT544Dplus – Standard US LT544Dplus-EUR – Schuko
Physical size 90 mm x 180 mm x 38 mm (3.5 in x 6.2 in x 1.6 in)
31
Ultrasound Transducer Leakage Tester
Ultrasound Transducer Leakage Tester
Key features• Handheld• Stand-alone operation• Direct measurement of
leakage current• Battery-operated• Independent of 120 V or
240 V systems• Built-in self-test circuit• Auto shut-off to conserve
battery
ULT800
ULT800 tests the electrical safety of ultrasound transducers independent of their ultrasound machines. A variety of adapters allow for testing of many different makes and models, including transesophageal echocardiography (TEE) transducers.
With the ULT800, transducer testing easily fits into routine disinfecting procedures. Technicians conduct the tests in a cleaning basin at the beginning of the day and between patients. Simple Pass/Fail indicators make it simple to use even non-technical medical personnel, such as sonographers and central sterile-supply technicians, can perform the tests.
The ULT800 is available by itself or as a kit, which includes two transducer adapters, a dual-conductivity electrode, and carrying case.
Specifications
Power 9 V alkaline battery, approximately 1000 uses per battery
Conductivity Limit to pass: > 133 μA +1 %
Leakage Limit to pass: < 185 μA +1 %; > 20 μA +1 %
Dimensions (LxWxH) 17 cm x 10 cm x 4 cm (6.5 in x 3.7 in x 1.5 in)
Weight 0.34 kg (0.75 lb)
Operating temperature 15 °C to 40 °C (59 °F to 104 °F)
Storage temperature 15 °C to 65 °C (59 °F to 149 °F)
Humidity 90 % max
Ordering information2434187 ULT800 Ultrasound Transducer Leakage Tester Kits2491569 Cust1 Ultrasound Transducer Leakage Tester KitStandard accessories: ULT800 (2491569), dual-conductivity electrode (2392502), and hard-sided carrying case (2392525)
2491578 Cust2 Ultrasound Transducer Leakage Tester Kit Standard accessories: ULT800 (2491569), 800-Cleaning Tub, Cleaning/Testing Basin for ULT800 Testing System (2743055, includes foam, Velcro strap, and shipping container), 600/220PLUS, Dual conduc-tivity electrode for use with 800-Cleaning Tub, Cleaning/Testing Basin for ULT800 Testing System (2801776)
2392502 Dual Conductivity Electrode2392525 Hard-sided Carrying Case2392569 Dual Conductivity Electrode—for use with Cidex 2032 Tray2801776 600/220PLUS, Dual Conductivity Electrode for use with 800-Cleaning Tub, 2392533 Conductivity Adapter Cable for PCI GUS cleaning2231616 Universal Ultrasound Probe for DALE601 and DALE601E2392427 Chassis Ground Probe for DALE601/DALE601E2392430 Acuson/Siemens ultrasound transducer adapter for Acuson 156 and V510B Transducers2743055 800-Cleaning Tub, Cleaning/Testing Basin for ULT800 Testing System2392578 Acuson/Siemens Ultrasound Transducer Adapter for Acuson 260 Transducers2392591 Acuson/Siemens Ultrasound Transducer Adapter for Acuson MP and 3V2c Transducers
2392516 Acuson/Toshiba Ultrasound Transducer Adapter for Acuson XP, Acuson Aspen, Acuson Capasee, Acuson 3-Needle Guide C3, ATL 3.5 DFT, Toshiba PSF-37HT, and Toshiba F Series Transducers2392482 HP/Agilent Ultrasound Transducer Adapter (600/210)2392494 HP/Agilent Ultrasound Transducer Adapter (600/211)2392475 GE LogiQ Ultrasound Transducer Adapter (600/204)2231602 GE LogiQ Ultrasound Transducer Adapter (600/205)2392453 GE YMS/RT Ultrasound Transducer Adapter for GE YMS/RT Transducers2392466 GE CGR Ultrasound Transducer Adapter for GE CGR radius and GE SONO Transducers2392557 Philips Ultrasound Transducer Adapter (ATL/600/218)800/PHILIPS-04 Ultrasound Transducer Adapter (for use with Philips iE33 and iU22 diagnostic ultrasound TEE transducers)2540999 SonoSite TEE Ultrasound Transducer Adapter (for use with all SonoSite Transducers-including TEE)
Optional accessories
ULT800 Ultrasound Electrical Leakage testing system (TEE transducer not included)
32
Key features• Dynamic BP simulators for
arm- and wrist-cuff monitors• ECG and arrhythmia simulation
synchronizedwith BP (optional)• Internal pump for high- and
low-pressure release verifica-tion, leak testing, and pressure sourcing
• Internal adult/neonatal cuffs elimates need for external cuffs
• Four respiratory artifacts, including spontaneous breath-ing and controlled ventilation
• Multiple arrhythmia simula-tions, including premature atrial contractions #1 and #2, atrial fibrillation, and PVCs
• Optional Ansur test automation software to standardize testing procedures, capture wave-forms, and print and document test result
Pressure generation/ measurement
Static-pressure range: 0 mmHg to 400 mmHg (53 kPa)
Difference between target pressure and actual pressure: -5 mmHg
Internal leak rate: < 2 mmHg per minute with minimum volume of 300 cc
Four respiratory artifacts
3 spontaneous breathing; controlled ventilation
Three adult wrist-cuff simulations
Normal, Hyper, Hypo
Pressure source Specified pressure generated from 50 mmHg to 400 mmHg in selectable increments of 1 mmHg
Pressure gauge Static pressure measured from 0 mmHg to 400 mmHg at the pressure port
Pressure relief test Test for the NIBPM pressure relief valve (0 mmHg to 400 mmHg) with display of peak pressure
Neonate internal cuff simulations
Internal neonate cuff; four standard neonate pressures
Neonate simulations Cuff #1: Blood pressure: 35/15 Heart rate: 120 BPM Pulse volume: 0.3Cuff #2: Blood pressure: 60/30 Heart rate: 120 BPM Pulse volume: 0.3Cuff #3: Blood pressure: 80/50 Heart rate: 120 BPM Pulse volume: 0.3Cuff #4: Blood pressure: 100/70 Heart rate: 120 BPM Pulse volume: 0.3
Normal sinus rhythm BP and ECG: Healthy heart, weak pulse, mild exercise strenuous exercise, obese subject, geriatric subject, tachycardia, bradycardia
Irregular pulse BP and ECG: Premature atrial contractions # 1, premature atrial contractions # 2, premature ventricular contractions, atrial fibrillation and PVCs
User-definable simulations
User-definable systolic and diastolic values, along with heart rate and pulse volume Ranges:Systolic pressure range Diastolic pressure range Heart rate Pulse volume 20 mmHG to 250 mmHG 10 mmHG to 200 mmHG 30 BPM to 250 BPM 0.1 cc to 2.4 cc in increments of 0.1 cc
Simulation parameters performance
Max pulse volume: 2.4 cc
Max heart rate: 200 BPM at 2.4 cc pulse volume; 250 BPM at 1.2 cc pulse volume
Internal neonatal cuff volume: 20 cc
Internal adult cuff volume (including NN volume): 310 cc
Heart rate setting accuracy: ± 1 BPM
Simulation units: kPa and mmHg (user selectable)
Pressure leak test The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the pressure loss over time. Peak pressure and present pressure are displayed at all times; leak rate is displayed when it is available.
Autosequences Nine autosequences are provided for four tests and up to five simulations
Electrical ECG (optional)
Signals: RA, LA, RL, LL, V
Waveform: Lead II
Amplitude: 1 mV peak (± 10 %) NIBP peripheral pulse synchronized with ECG signal
Connections: Optional external ECG adapter, physiological synchronization with NIBP
Non-Invasive Blood Pressure Simulators
Non-Invasive Blood Pressure Simulator
BP Pump 2
Specifications
The BP Pump 2 is a second- generation non-invasive blood pres-sure (NIBP) monitor analyzer that efficiently verifies oscillometric adult and neonatal NIBP. The BP Pump 2’s unique feature set includes tests to accurately interrogate wrist-cuff monitors, internal cuff volumes, and optional 5-lead synchronized ECG simulations for spot checks on the
monitor. The simulated peripheral pulse is synchronized with this electrical ECG signal for testing NIBP monitors utilizing gated measurement for noise/artifact rejection.
33
Included accessories2391882 Accessory Kit (tubings and fittings)Users ManualXXXXXXX Power Cord (country specific)
Ordering informationBP Pump 2L (standard pressure transducer)2249036 United States, 120 V2394895 Australia, 250 V2394901 Denmark, 250 V2394912 Schuko, 250 V2394920 Israel, 250 V2394935 Italy, 250 V2394947 India, 250 V2394958 Switzerland, 250 V2394964 United Kingdom, 250 V3334821 US 120V w/Test Automation3334839 AUS w/Test Automation3334842 DEN w/Test Automation3334856 SHK w/Test Automation3334863 ISR w/Test Automation3334874 ITAL w/Test Automation3334888 IND w/Test Automation3334895 SWZ w/Test Automation3334901 UK w/Test Automation
Non-Invasive Blood Pressure Simulators
Non-Invasive Blood Pressure Simulator
BP Pump 2
Optional accessories2755836 Ansur BP Pump 2 Plug-in2222822 Soft-sided Vinyl Carrying Case2391894 ECG Adapter Block (allows simulation of 5-lead ECG waveforms)2248899 Printer, Seiko DPU-414-30 B, 120 V power supply2399531 Printer, Seiko DPU-414-30B, 200 V power supply2238659 Serial Cable, D9M-D9F2392328 Neonatal/external cuff mandrel (truncated plastic cylinder diameters: 7.6, 10, and 14 cm)2391875 Wrist cuff mandrel (adult)
Serial port Bidirectional RS-232 port; baud rate of 9600 with no parity, one stop bit, and eight data bits
Pressure measurement Pressure-measurement units: kPa, mmHg, cmH20, cmH2O and psi (user selectable)Range: 0 mmHg to 400 mmHg
Accuracy Basic model (BP Pump 2L): 0 mmHg to 300 mmHg: + 0.5 % of reading + 1 mmHg; 301 mmHg to 400 mmHg: + 2 % of readingHigh-accuracy version (BP Pump 2M): < 0.8 mmHg (0.1 kPa) throughout range
Parallel port 25-pin female connector, with D-subminiature style and pinouts conforming to IBM PC printer port (unidirectional), HP and ASCII printers
Sample adult arm-cuff simulation (standard parameters)
Standard set of blood pressures: BP #1: Blood pressure: 120/80 (93) Heart rate: 80 Pulse volume: 0.68 ccBP #2: Blood pressure: 150/100 (116) Heart rate: 80 Pulse volume: 0.65 ccBP #3: Blood pressure: 200/150 (166) Heart rate: 80 Pulse volume: 0.6 ccBP #4: Blood pressure: 255/195 (215) Heart rate: 80 Pulse volume: 0.55 ccBP #5: Blood pressure: 60/30 (40) Heart rate: 80 Pulse volume: 0.75 ccBP #6: Blood pressure: 80/50 (60) Heart rate: 80 Pulse volume: 0.7 ccBP #7: Blood Pressure: 100/65 (76) Heart rate: 80 Pulse volume: 0.69 cc
Patient condition simulations
Healthy heart, weak pulse, mild exercise #1, strenuous exercise #2, obese subject, geriatric subject, tachycardia, bradycardia
Arrhythmia simulations Premature atrial cont. #1, premature atrial cont. #2, premature ventricular cont., atrial fib and PVCs
Wrist simulations Simulation #1: Blood pressure 120/80 (93) Heart rate: 80 BPM Pulse volume: 0.5 ccSimulation #2: Blood pressure 160/100 (120) Heart rate: 80 BPM Pulse volume: 0.5 ccSimulation #3: Blood pressure: 80/55 (63) Heart rate: 80 BPM Pulse volume: 0.5 cc
Temperature Operating: 15 °C to 40 °C (59 °F to 104 °F) Storage: -20 °C to 65 °C (-4 °F to 149 °F)
Display Bright, large 4-line x 40-character alphanumeric display with backlighting
Dimensions (WxDxH) 25.4 cm x 25.4 cm x 12.7 cm (10 in x 10 in x 5 in)
Weight 3.4 kg (7.5 lb)
BP Pump 2M (high-accuracy pressure transducer)2249049 United States, 120 V2394973 Australia, 250 V2394986 Denmark, 250 V2394999 Schuko, 250 V2395003 Israel, 250 V2395015 Italy, 250 V2395026 India, 250 V2395032 Switzerland, 250 V2395044 United Kingdom, 250 V3334912 US w/Test Automation3334920 AUS w/Test Automation3334935 DEN w/Test Automation3334947 SHK w/Test Automation3334958 ISR w/Test Automation3334964 ITAL w/Test Automation3334973 IND w/Test Automation3334986 SWZ w/Test Automation3334999 UK w/Test Automation
34
Key features• Dynamic oscillometric
non-invasive blood-pressure simulation
• Automated static-pressure measurements, leakage testing, and relief-valve testing
• Five automated NIBP testing autosequences
• Five arrhythmia selections• Adult and neonatal NIBP
selections• Adjustable heart rate values• Direct interface with
medTester 5000C
The CuffLink Non-Invasive Blood Pressure (NIBP) Analyzer offers a quick, reliable, and consistent way to evaluate the operation and performance of oscillometric NIBP signals.
With user-programmable selections, the CuffLink will sim-ulate the full range or normal, hypertensive, and hypotensive dynamic NIBP waveforms repre-senting typical adult, infant, and neonatal patients. The analyzer can also generate normal, brady-cardia, and tachycardia rhythm selections with a wide range
of weak, normal, and strong peripheral pulses. A variety of parameters allow creation and storage of five custom autosequences within the unit. In addition to the programmable blood-pressure target values, these autosequences can include static pressure, leak, and over-pressure relief valve pop-off tests.
Power 120/250 V ac, 50 Ω average, 100 Ω peak, 50/60 Hz
Analog outputs Cuff pressure: 0 mmHg to 499.95 mmHg FS ± 1 % of reading, 10 mV/mmHg
Pulse pressure: 0 mmHg to 5 mmHg FS ± 1 of reading, 1 V/mmHg
Digital manometer Pressure: -499.75 mmHg to 499.75 mmHg
Measurement parameters: instantaneous and peak
Displayed graphics Dynamic real-time NIBP cuff-pressure waveform programmed peripheral pulse and envelope waveforms
Display Alphanumeric graphic display (LCD)
Alphanumeric mode: 8 lines x 40 characters
Graphics mode: 64 vertical x 240 horizontal dot matrix, backlight with viewing angle adjustment
Digital interfaces RS-232/serial: bidirectional: downloads cuff measurement data and con-trols test features with a compatible computer or via the medTester 5000C with the medCheck option.
Parallel printer: Centronics compatible
Pulse sync: 0 V dc to 5 V dc (TTL)
Cuff mandrel Interlocking plastic blocks: four cuff circumferences, including 39.5 cm (large adult), 33 cm (adult), 26.6 cm (small adult) and 20 cm (child); maxi-mum cuff: 15.25 cm
Truncated plastic cylinders: three neonatal cuff circumferences, including 14 cm, 10 cm, and 7.6 cm; maximum cuff width of 7.6 cm
Pop-off valve testing Automatic test for operation of the monitor’s relief valve
Measurement parameters: instantaneous and peak pressure
Maximum pressure: 499.75 mmHg
System leak testing Start pressure: 499.75 mmHg max
Elapsed time: 60 s (fixed)
Leak-rate range: 0.25 mmHg/min to 499.75 mmHg/min
Pump: 2 liters/minute minimum (free flow)
Accuracy Dynamic NIBP Response Repeatability (Systolic/Diastalic/Mean): 1 % of target value
Cuff pressure: ± 1 % of reading ± 1 mmHg
Input overpressure limit: ± 1500 mmHg
Specifications
CuffLink standard accessories
Non-Invasive Blood Pressure Simulator
CuffLink
Non-Invasive Blood Pressure Simulator 35
Non-Invasive Blood Pressure Simulator
Included accessories2242915 Operator’s Manual2392381 Adult Cuff Mandrel Spacer Blocks2230305 Adult Cuff Mandrel End Blocks (qty 2)2392328 External Cuff Mandrel Neonatal (truncated plastic cylinder diameters: 7, 6, 10, and 14 cm)2245300 CuffLink Adapter Kit2198846 Hospital Grade Power Cord Set2248408 Soft Vinyl Accessory Pouch
Ordering informationCuffLink Non-Invasive Blood Pressure Simulator2246985 United States, 120 V2399820 Australia, 250 V2399835 Denmark, 250 V2399873 India, 250 V2399858 Israel, 250 V2399864 Italy, 250 V2399847 Schuko, 250 V2399886 Switzerland, 250 V2399899 United Kingdom, 250 V
Specifications
Autosequences Up to five user-programmable sequences to test NIBP monitors with a specific series of cuffLink performance tests, including static pressure test, leak test, and pop-off test
Up to eight adult-neonatal-arrhythmia dynamic NIBP selections, each of which can be cycled up to 99 times during the sequence
Printable test report
Displayed real-time parameters
Instantaneous cuff pressure: 0 mmHg to 300 mmHgPeak cuff pressure: 0 mmHg to 500 mmHg peakInflate/deflate time: 0.1 s to 999.9 sInflate/deflate rate: 0.1 mmHg/s to 999.9 mmHg/sTotal measurement time: 0 s to 999.9 s max
Selected heart rate:Selected systolic/diastolic (mean) target valuesUser-programmed vertical and horizontal shifts
Dynamic non-invasive blood pressure
Simulation of a range of normal, hypertensive and hypotensive dynamic noninvasive blood pressures for typical adult, infant, and neonatal patients. Generation of normal, bradycardia, and tachycardia rhythm selections with a wide range of user-programmable peripheral pulse amplitudes (weak, normal and strong). Compatible with oscillometric NIBP devices.
Preprogrammed target value selections: Adult systolic/diastolic (MAP) (mmHg): 60/30 (40), 80/50 (62), 100/65 (75), 120/80 (90), 150/100 (115), 200/150 (165) and 255/195 (215)
Neonatal/pediatric systolic/diastolic: above selections, excluding 255/195 and 200/150
Repeatability: ± 1 % of selected target value
Adult arrhythmia selections:• Baseline NIBP target value: (120/80) (NSR)• Atrial fibrillation (A-Fib)• Premature atrial contraction (PAC)• Premature ventricular contraction (PVC)• Missed beat (MB)• Aberrant sinus conduction (AS)
Preprogrammed peripheral pulse waveforms:• Pulse amplitude at MAP: 2 mmHg (typical adult value)• Pulse volume range: 0 ml to 5.1 ml• Pulse rise time: 80 ms (min)• Heart rates (adult and neonate): 30 BPM, 40 BPM, 60 BPM, 80 BPM,
120 BPM, 160 BPM, 200 BPM, and 240 BPM• Heart-rate accuracy: ± 1 % of selected rate
Pre-programmable target value shifts: Horizontal axis: preprogrammed target value selections shifted in 1.0 mmHg steps over a maximum range of ± 300 mmHg to increase or decrease dynamic pressure valuesVertical axis: relative amplitude shifted in 1 % increments over a maxi-mum range from 0 % to 200 % to simulate weak, normal, and strong peripheral pulses
Environmental Operating temperature: 15 °C to 35 °C (59 °F to 95 °F)
Storage temperature: 0 °C to 50 °C (32 °F to 122 °F)
Dimensions (LxWxH) 38.1 cm x 31.75 cm x 12.7 cm (15 in x 12.5 in x 5 in)
Weight 6.82 kg (15 lb)
Model BP Pump 2 CuffLink
Wrist cuff simulations Yes No
ECG syncronized Option No
Leak test Yes Yes
Over pressure test Yes Yes
Manometer Yes Yes
medTester compatible No Yes
Product comparison chart
CuffLink
Non-Invasive Blood Pressure Simulator
36
Patient Simulator
MPS450
The MPS450 is Fluke Biomedical’s next-generation, portable, multiparameter patient simulator for your comprehen-sive testing and training needs. Whether it’s a quick check on a bedside monitor, arrhythmia recognition training, or pe forming a complete PM on the latest patient-monitoring systems, this simulator is a clear choice with its broad range of physiological wave-forms, easy-to-use interface, and compact, portable size.
ECG normal sinus rhythm
12-lead configuration with independent outputs
Amplitude: 0.05 mV to 5.5 mV
Rates: 30 BPM to 300 BPM
ECG waveform selections: Adult or pediatric
Superimposed artifact: 50 Hz and 60 Hz, muscle, baseline wander, respi-ration
ECG performance Amplitude: 0.05 mV to 5.5 mVSquare wave: 2 Hz , 0.125 HzPulse: 30 BPM, 60 BPM, 60 ms pulse widthSine waves: 0.5 Hz to 100 HzTriangle wave: 2 Hz, 2.5 Hz
ST segments Elevated/depressed: -0.8 mV to 0.8 mV in 0.1 mV steps; plus -0.05 mV and 0.05 mV steps
Accuracy All amplitudes ± 2 % of setting Lead IIAll rates ± 1 %All widths ± 1 %
Arrhythmia selections (43 Total)
Premature rhythmsSupraventricular rhythmsVentricular rhythmsConduction defectsPacemaker
Respiration Baseline impedance: 500 Ω to 2000 Ω, leads I, II, IIIImpedance variations: 3 Ω, 1 Ω, 0.5 Ω, 0.2 ΩRates: 15 BrPM to 120 BrPM and APNEAApnea periods: 12 seconds, 22 seconds, 32 seconds, and continuous
Model MPS450 medSim 300B PS420 PS415
Arrhythmia selections 43 32 35 14
Respiration Yes Yes Yes Yes
BP channels Yes, four Yes, four Yes, two Yes, two
Swan-Ganz procedure Yes Yes Yes Yes
Temperature channels Yes, one Yes, two Yes, one Yes, one
User-programmable auto-sequences Yes (with HHC3) Yes, internal feature plus HHC3 capability
No Yes
Cardiac output Optional Optional Standard No
Product comparison chart
Key features• 12-lead ECG simulation with
independent outputs• 43 arrhythmia selections• Four invasive BP channels,
including Swan-Ganz simulation
• Respiration and temperature simulations
• ECG performance testing, including R-Wave detection
• Large, bright 4-line x 20-character display
• RS-232 serial port• Expansion port• Compact and lightweight
Patient Simulators
Specifications
MPS450 optional accessories
37
Blood pressure channels
Channels 4; synchronized with normal sinus rhythm rates; tracks arrhythmia activity
TransducerExciter voltages: ac and dc compatibleSensitivity: 5 μV/V/mmHg and 40 μV/V/mmHgCalibrated Rate: 80 BPM
Available selections Static pressure
Dynamic pressure: Art (120/80), Radial Art (120/80), LV (120/0), RA/CVP (15/10), RV (25/0), PA (25/10), PAW (10/2), and LA (14/4)
Swan-Ganz procedure: automated and manual control
Temperatures 0 °C, 24 °C, 37 °C, and 40 °C
Cardiac output (optional)
Faulty-injectate curveLeft-to-right shunt curveC.O. for 0°: 2.5 l/min, 5 l/min, and 10 l/minC.O. for 24°: 2.5 l/min, 5 l/min, and 10 l/minCal Pulse: 1.5° for 1 second
Fetal/Maternal ECG and IUP simulations (optional)
Fixed fetal heart rates: 60 BPM to 240 BPM
Dynamic fetal heart activity: Uniform deceleration, uniform acceleration, early deceleration, late deceleration
Maternal heart rate: 80 BPM
Dynamic intrauterine pressure (IUP)
Waveform: positive bell-shaped pressure curve
Peak pressure: 90 mmHg, ± 4 mmHg (max)
Contraction interval: 2 minutes, 3 minutes, and 5 minutes (manual)
Duration: 90 seconds
Dimensions (WxDxH) 18.4 cm x 19 cm x 5 cm (7.3 in x 7.5 in x 2 in)
Weight 0.6 kg (1.4 lb)
Included accessories2720054 AC Battery Eliminator2243350 Users Manual
Ordering information2251364 MPS450 (ECG 12-lead simulation; invasive BP; respiration; temperature; BP in sync with ECG; large, bright 4-line x 20-character display; R-wave-detection test; RS-232 port for computer control; soft-key navigation; universal ECG connectors; and flash memory for easy program upgrade)
2251373 MPS450-CO (base model plus cardiac-output simulation)
2251399 MPS450-FET (base model plus direct fetal/maternal ECG simulations with maternal heart rate, selectable fetal heart rate, and dynamic intrauterine pressure waveform [IUP])
2251386 MPS450-CO/FET (base model plus cardiac-output simulation and direct fetal/maternal ECG simulations with maternal heart rate, selectable fetal heart rate, and dynamic intrauterine pressure waveform [IUP])
Optional accessories2248623 Soft-Sided Vinyl Carrying Case2238659 Serial Cable D9M-D9F2226608 Cardiac-Output Adapter Box2645641 HHC3 Handheld Controller
Cardiac output adapters2392285 GE Medical/Marquette Cardiac Output Cable (interface cable for GE Medical/Marquette plus monitors, including in-line switch box to select injec-tate temperature)2227016 Gould/Spectramed 1445 Injectate Temperature Adapter (4 pin)
2227025 Gould/Spectramed 1465 Injectate Temperature Adapter (phone jack)2226973 HP Injectate Temperature Adapter (1/4 in phone plug)2391990 Universal Injectate Temperature Adapter Pigtail (unterminated)2392158 General Purpose Connector
Model PS410 PS400 DataSim 6100
Arrhythmia selections 35 12 34
Respiration No No Yes
BP channels No No Yes, three
Swan-Ganz procedure No No Yes
Temperature channels No No No
User-programmable auto-sequences No No Yes
Cardiac output No No Optional
For a complete list of temperature and blood-pressure cables, contact us.
Patient Simulator
MPS450
Patient Simulators
Specifications
38
If your physiological-monitor testing requirements call for a full-featured, comprehensive instrument, the medSim 300B is your answer. Used by thousands of biomedical departments and many medical-device manufacturers around the world, this simulator delivers an entire array of parameters with ultimate flexibility.
Specifications
Arrhythmia selections Supraventricular, premature, conduction, ventricular, and transvenous pacemaker
Programmable sequences 15 preprogrammed selections or user programmable
ECG parameters Normal sinus rhythmPerformanceST segmentsArrhythmia selections
medSim 300B option 1 Fetal/maternal ECG and IUPIntra-aortic balloon assist pumpCardiac catheterizationCardiac Output OptionFlow rates: 3 lpm to 7 lpmCurves: Normal, interrupt, slow, L/R shuntTrends: Avg = 5 l/min
Controller option Direct waveform access
Premature beat selections Manual: Push button insertions of one PVC, PAC, or PNCAutomated: User-programmable PVC activityPVC types: 4 LV and RV fociPVC timing: Standard, early, and R-on-TPVC rate: 0 to 25/min or automated PVC-rate variance
Pacemaker Amplitude: -700 mV to 700 mVWidth: 0.1 ms to 2 ms
Defibrillator training Three scenarios available
Key features• 12-lead ECG simulation with
independent output• Four invasive BP channels
including Swan-Ganz simulation
• Respiration and temperature• 9 V dc battery and lead test• ECG-performance testing
including R-Wave detection• Interactive defibrillation
training• 2-line x 20-character display• RS-232 serial port
Patient Simulator
medSim 300B
Patient Simulators 39
Blood pressure channels ChannelsFour synchronized with normal sinus rhythm rates; tracks arrhythmia activity
TransducerExciter voltages: ac and dc compatibleSensitivity: 5 μV/V/mmHg and 40 μV/V/mmHgCalibrated rate: 80 BPM
Available selections Static pressure
Dynamic pressure: Art (120/80), LV (120/0), RA/CVP (15/10), RV (25/0), PA (25/10), PAW (10/2)
Swan-Ganz procedure (manual)
Triangle wave
Respiration Normal physiological simulationBaseline impedance: 500 Ω to 2000 ΩImpedance variations: 0 Ω to 3 ΩRates: 15 BrPM to 120 BrPMApnea: Off, momentary, continuous, timed (12 sec and 32 sec)
Auxiliary features I/E Ratio: 5/1, 4/1, 3/1, 2/1, 1/1
Baseline shift: Delta impedance is reduced to 1/6 and shifted either positive or negative. The rate is changed to 120 BrPM for 12 sec/min.
Ventilator simulation: 40 BrPM at fixed ratio
Artifact ECG: 50 Hz, 60 Hz, muscle, and baseline wander. All or single leads.BP/respiration: Two pressure values
Temperature Channel 1: Fixed at 37 °C (98.6 °F)Channel 2: 34 °C (86 °F), 37 °C (98.6 °F), 40 °C (104 °F)Hypothermia, hyperthermia, and spike
Power Two 9 V dc alkaline batteries, battery eliminator
Dimensions (WxDxH) 17.8 cm x 25.4 cm x 7.6 cm (7 in x 10 in x 3 in)
Weight 1.6 kg (3.5 lb)
Optional accessories2248554 Multipurpose Hard-sided Watertight Carrying Case 2199070 PC remote control interface cable (right-angle DIN to female DB25) 2199225 Patient Simulator to medTester interface cable (right-angle DIN to female DB25) 2199747 DS-A Datascope System 9 (DS-1 BP cable required) intra-aortic balloon assist sync cable2199786 KT-A Kontron K200, KAAT (KT-1 BP cable required) intra-aortic balloon assist sync cable 2645641 HHC3 Handheld Controller
For a complete list of cardiac adapters, temperature and blood-pressure cables, contact us.
Patient Simulators
Patient Simulator
medSim 300B
Specifications
40
Key features• Compact, lightweight,
pocket size• Labeled hot keys for common
settings• 12-lead ECG• Respiration and temperature
selection• Two-channel blood pressure
simulation• Optional cardiac output• Adult and pediatric normal
sinus rhythms• 35 arrhythmia selections• ECG performance waveforms• ST segment levels• ECG artifact• Pacemaker simulation• RS-232 serial port for
computer control• Battery operated
Specifications
ECG
Normal rate 80 BPM
Selectable rates 30 BPM, 40 BPM, 60 BPM, 80 BPM, 100 BPM, 120 BPM, 140 BPM, 160 BPM, 180 BPM, 200 BPM, 220 BPM, 240 BPM, 260 BPM, 280 BPM, and 300 BPM
Accuracy ± 1 %
Output impedance 500 Ω, 1000 Ω, 1500 Ω, and 2000 Ω for leads I, II, and III
ECG amplitudes 0.5 mV, 1 mV, 1.5 mV, and 2 mV
Amplitude accuracy ± 2 % lead II
Adult or pediatric ECG waveform performance waveform
Lead II square wave 2 Hz, 0.125 Hz
Pulse 30 BPM and 60 BPM, 60 ms pulse width
Sine wave 0.5 Hz, 4 Hz, 10 Hz, 40 Hz, 50 Hz, and 60 Hz (1 mV amplitude, lead II)
Triangle wave 2 Hz
ST segment analysis
Elevated or depressed -0.8 mV to +0.8 mV in 0.1 mV steps
Pacemaker
Pacer spike Amplitude: 2 mV, 4 mV, 8 mV, and 10 mV in lead IIAccuracy: ± 5 %, Lead II
Pacer spike Duration: 0.1 ms, 0.5 ms, 1 ms, 1.5 ms, and 2 msAccuracy: ± 5 %
Functions Asynchronous pacemakerPacer non-functionPacer non-captureDemand occasional sinusDemand frequent sinusAV sequential
The PS420 is a handheld, high-performance simulator for testing patient monitors. Small enough to fit in a pocket, the handy PS420 features a wide variety of simulation capability, including a full range of ECG, respiration, blood pres-sure, temperature and cardiac output conditions. The tool includes 12-lead ECG, two-channel blood pressure simulation, 35 arrhythmia selections, pacemaker simulation as well as adult and pediatric normal sinus rhythms. For convenient use, labeled hot keys on the keypad guide users to the most common settings.
Patient Simulator
PS420
Patient Simulators 41
Blood pressure
Input/output impedance 350 ΩExciter input limit ± 10 V
Exciter input frequency range
DC to 4000 Hz
Transducer sensitivity 5 μV/V/mmHg or 40 μV/V/mmHg
Level accuracy ± 1 %, ± 1 mmHg
Static levels BP1 -10 mmHg, 0 mmHg, 50 mmHg, 100 mmHg, 150 mmHg, 200 mmHg, and 250 mmHg
Static levels BP2 -10 mmHg, 0 mmHg, 80 mmHg, 160 mmHg, 240 mmHg, 320 mmHg, and 400 mmHg
Channel selections: Arterial 120/80, channel 1 and 2Radial artery 120/80, channel 1 and 2Left ventricle 120/00, channel 1 and 2Right ventricle 25/00, channel 1 and 2Central venous 15/10, channel 2Pulmonary artery 25/10, channel 2Pulmonary wedge 10/2, channel 2Left atrium 14/4; automatic Swan/Ganz (every 20 sec)Manual swan/ganz (changes when entry is selected), channel 2Synchronized with all normal sinus rates.Physiologically track all arrhythmia selection
Cardiac output (must have optional cardiac output adapter box p/n 2462200)
Catheter type Baxter edwards, 10 cc
Blood temperature 37 °C (98.6 °F)CO for 2 °C (35.6 °F): 3, 5, 7 l/minCO for 20 °C (68 °F): 3, 5, 7 l/min
Cal pulse Of 1 °C for 1 sec; of Delta 402 Ω for 4 sec.
Computational constant For 2 °C (35.6 °F) is 0.561; for 20 °C (68 °F) is 0.608
Left to right shunt 2 °C and 20 °C (35.6 °F and 68 °F)
Faulty injectate 2 °C and 20 °C (35.6 °F and 68 °F)
Accuracy ± 5 %Calibrated or uncalibrated cardiac output waves for 4 different CO values
Respiration
Baseline impedance 500 Ω, 1000 Ω, 1500 Ω, and 2000 Ω, leads I, II, and III
Lead selections LL or LA
Impedance variations 3 Ω, 1 Ω, 0.5 Ω, and 0.2 ΩAccuracy ± 5 %
Rates 15 BPM, 20 BPM, 30 BPM, 40 BPM, 60 BPM, 80 BPM, 100 BPM, 120 BPM, and 0 BPM for apnea
Accuracy ± 2 %
Apnea 12 seconds, 22 seconds, 32 seconds, and continuous
Temperature
Compatibility YSI 400/700 series
Temperature 30 °C, 35 °C, 37 °C, 40 °C, and 42 °C (86 °F, 95 °F, 98.6 °F, 104 °F, and 107.6 °F)
Temperature simulation accuracy
± 0.25 °C
Specifications
Patient Simulators
Patient Simulator
PS420
42
Key features• Range: 10-1 to 107 mR/h• Energy response: 10 % from
55 keV to 3 MeV
Included accessories2631808 PS420 Users Manual (printed)2631721 PS420 Users Manual (electronic, CD)2647372 Battery Eliminator 100 V ac to 240 V ac9 V Battery
Ordering information2631290 PS420 Patient Simulator
Optional accessories2462072 Universal Banana Adapter (17024)2462189 Carrying Case, single pocket2462177 Carrying Case, double pocket2651740 Cardiac Output Adapter Box PS420 (17290)2462295 BP Cable, unterminated PS4202462312 Temperature Cable, unterminated PS4202462217 RS-232 Cable
Arrhythmias Base rate of 80 BPMSinus arrhythmiaAtrial (PAC)*Missed beat*Atrial tachycardiaAtrial flutterNodal (PNC)*Nodal rhythmSupraventricular tachycardiaPVC1 left ventricular focus*PVC 1 early, LV focus *PVC1 R on T, LV focus*PVC2 right ventricular focus*PVC2 early, RV focus*PVC2 R on T, RV focus*Multifocal PVCs*Atrial fibrillation coarse/finePVCs 6/minutePVCs 12/minutePVCs 24/minuteFrequent multifocal PVCsBigeminyTrigeminyPair PVCs*Run 5 PVCs *Run 11 PVCs*Ventricular tachycardiaVentricular fibrillation coarse/fineAsystoleConduction defectsFirst degreeSecond degreeThird degreeRight bundle branch blockLeft bundle branch block*Will go to NSR ECG @ 80 BPM after completion
Artifacts 50/60 HzMuscleBaselineRespiration
General
Dimensions (LxWxH) 15.6 cm x 9.4 cm x 3.4 xm (6.1 in x 3.7 in x 1.3 in)
Weight 0.4 kg (0.9 lb)
Specifications
Patient Simulators
Patient Simulator
PS420
43
The PS415 is a high-performance multiparameter patient simulator for testing full-function monitors. Compact and lightweight, the handheld PS415 provides the versatile features and convenience to meet the needs of mobile field service technicians.
The device features 12-lead ECG output, two invasive blood-pressure channels, a wide range of arrhythmia selections, and respiration and temperature simulation. With its broad capability, the PS415 provides one of the best cost-benefit ratios among simulators on the market today.
Technicians control the PS415 with manual commands or onboard autosequences. A convenient pullout card provides quick menu access for all the manual simulations, and the front panel features an easy-to-use soft-key interface.
Included accessories2572338 Operating Manual2183983 Battery Eliminator2392729 Lead-test Adapter2248505 Soft-sided Carrying Case
Ordering information2558944 PS415 Patient Simulator
Optional accessoriesFor a complete list of cardiac adapters, temperature and blood-pressure cables, contact us.
Normal sinus rhythm:
Selectable rates 30 BPM, 40 BPM, 60 BPM, 70 BPM, 80 BPM, 90 BPM, 100 BPM, 120 BPM, 140 BPM, 150 BPM, 160 BPM, 180 BPM, 200 BPM, 210 BPM, 220 BPM, 240 BPM, 270 BPM, 300 BPM, and 350 BPM
Accuracy ± 1 %
Auto-step rate 30 BPM, 60 BPM, 80 BPM, 120 BPM, 180 BPM, and 240 BPM at 30-second intervals
Amplitude lead II 0.5 mV, 1 mV, 1.5 mV, and 2 mV
Amplitude accuracy +5 % 2 Hz square wave at 1 mV p-p (Lead II)
Limb-lead impedance selections
500 Ω or 1000 Ω
V-lead impedance 1000 ΩAccuracy ± 5 %
General information
Power 9 V battery for 50-hour life or line-operated via battery eliminator
Output 7.7 V ac, 100 mA, unregulated
Display 2-digit LCD
Lead test If the resistance is less than 1 kΩ, the display flashes at the rate of 4 Hz
Data interface RS-232 compatible, bidirectional port
5-pin DIN female connector
Baud rate: 2400
Connector 2.5 mm, center (+)
Dimensions (LxWxH) 16 cm x 10.4 cm x 3.8 cm (6.3 in x 4 in x 1.5 in)
Weight 0.3 kg (0.8 lb)
Temperature Operating: 15 ˚C to 35 ˚C (59 ˚F to 95 ˚F)
Storage: 0 ˚C to 55 ˚C (32 ˚F to 131 ˚F)
Patient Simulators
Patient Simulator
PS415
Key features• Handheld• 12-lead ECG simulations• Wide range of arrhythmia
selections• Automatic or manual
operation• RS-232 port for computer
control• Two invasive blood pressure
channels• Respiration simulation• Temperature simulation• Battery operated• Pullout card for quick menu
access
44
Key features• Handheld• 12-lead ECG simulation• 12 arrhythmia selections• Universal ECG jacks• Auto sequencing of
performance waveforms• Battery operated
The PS410 ECG/Arrythmia Simulator is a compact, high-performance simulator for patient monitor testing.
This handheld device simulates a full range of cardiac rhythms and a wide variety of ECG con-ditions. It includes pacemaker simulation, 35 arrhythmia selec-tions, and adult and pediatric normal-sinus rhythms.
Small enough to fit in a pocket, the handy PS410 weighs less than a pound and is easy to operate. Technicians simply connect the simulator to the device under test and use the PS410 keypad to enter the code presets. The simu-lator then transmits the selected preset simulations to the device.
Normal sinus rhythm: 12-lead with independent outputs referenced to RL
Normal rate 80 BPM
Selectable rates 30 BPM, 40 BPM, 60 BPM, 80 BPM, 100 BPM, 120 BPM, 140 BPM, 160 BPM, 180 BPM, 200 BPM, 220 BPM, 240 BPM, 260 BPM, 280 BPM, and 300 BPM
Accuracy ± 1 %
Output impedance 940 Ω between leads
ECG amplitudes 0.5 mV, 1 mV, and 2 mV
Amplitude accuracy ± 2 % Lead II
High-level output 1000x Lead II
Waveforms Adult or pediatric ECG waveform
ECG performance: Lead II
Square wave 2 Hz and 0.125 Hz
Pulse 30 BPM, 60 BPM, and 120 BPM; 60 ms pulse width
Sine waves 0.5 Hz, 5 Hz, 10 Hz, 40 Hz, 50, and 60 Hz (1 mV amplitude only, lead II)
Triangle wave 2 Hz
ST Segment analysis Elevated or depressed: -0.6 mV to 6 mV in 0.2 mV steps
Pacemaker Pacemaker rhythmPacer non-capturePacer non-functionDemand pacer with occasional sinusDemand pacer with frequent sinusA-V sequential
Artifact selection 50 Hz artifact60 Hz artifactMuscle artifactBaseline artifactRespiration artifact
Specifications
Included accessories2631742 PS410 Users Manual (electronic, CD)2631795 PS410 Users Manual (printed)2647372 Battery Eliminator 100 V ac to 240 V ac9 V battery
Ordering information2631276 PS410 Patient Simulator
Optional accessories2647372 Battery Eliminator 100 V ac to 240 V ac2462189 Carrying Case, single pocket2462177 Carrying Case, double pocket2462072 10 Universal Banana Adapters
Patient Simulator
PS410
Patient Simulators 45
Included accessories2572345 Operating Manual2248424 Soft-sided Carrying Case2183983 Battery Eliminator9 V Battery
Ordering information2558932 PS400 ECG/Arrhythmia Simulator
Key features• Handheld• 12-lead ECG simulation• 12 arrhythmia selections• Universal ECG jacks• Auto sequencing of
performance waveforms• Battery operated
The PS400 is a basic patient simulator for quality-assurance testing of ECG machines, monitors and telemetry units.
The lightweight, handheld device features 12-lead ECG output and a single high-level output. Technicians can use manual or automatic sequence modes to check low- and high-frequency response, linearity, gain, damping, and paper speed. The PS400 also includes 12 arrhythmia selections to check arrhythmia detection systems and teach health care workers how to recognize normal and abnormal heart rhythms.
Using the PS400 is easy. The user interface features a menu selector switch and a simple-to-use rotary knob for selecting waveforms.
Specifications
Waveforms
ECG 30 BPM, 60 BPM, 120 BPM, 180 BPM, and 240 BPM
Square wave 2 Hz
Sine waves 10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz
Arrhythmias Atrial fibrillationSecond-degree A-V block, type 1Right-bundle-branch block Premature atrial contractionPremature ventricular contraction, earlyPremature ventricular contraction, R on TMultifocal PVCsBigeminy Run of 5 PVCsVentricular tachycardiaVentricular fibrillationPaced
Automated sequence
Pulse 4 seconds
Sine waves 10 Hz, 40 Hz, 60 Hz, and 100 Hz
Triangle wave 2 Hz
Rate accuracy ± 0.5 %
Amplitudes
Lead I 0.5 mV, 1 mV, 1.5 mV, and 2 mV
High-level output 0.25 V, 0.5 V, 0.75 V, and 1 V
Accuracy ± 2 %
Hz 2 Hz square wave at 2 mV (all leads)
General information
Power 9 V alkaline battery for 200-hour life or line-operated via battery eliminator
115 V ac or 230 V ac or 7.7 V dc, 100 mA, unregulated
Connector, 2.5 mm center (+)
Dimensions (LxWxH) 13.2 cm x 9.9 cm x 3.6 cm (5.2 in x 3.9 in x 1.4 in)
Weight 0.4 kg (0.9 lb)
Operating temperature 15 °C to 35 °C (59 °F to 95 °F)
Storage temperature 0 °C to 55 °C (32 °F to 131 °F)
Optional accessories2200116 High-level ECG Cable
Patient Simulators
Patient Simulator
PS400
46
Key features• Interface with
Resusci-Anne™, Arrhythmia Anne™, and Chris Clean™
• Synchronized hemodynamic waveforms
• ECG/Respiration• Manual PAC and PVC
insertions• Swan-Ganz procedure• Expansion modules• Training capabilities
The DataSim 6100 offers six channels for generating ECG arrhythmias, blood pressure, and respiration. Features include a standard defibrillator training capa-bility, along with an internal battery. All control keys and display prompts can be found on DataSim’s con-venient, handheld keypad, which is connected to the unit with a flexible, 6.1 m (20 ft) telephone-style cord for exceptional range of motion.
Critical Care Nurses, ACLS program instructors, and other clinical educators can
use DataSim 6100 to generate an extensive range of simulations, from a simple normal sinus rhythm to a complex Swan-Ganz catheter insertion.
Specifications
Optional personality modules
• Pediatric ECG• Intracranial pressures• Advanced pacemaker• MCL1 atrials• MCL1 blocks• MCL1 ectopy/aberrancy• MCL1 set• Normal/diseased left heart• 12-lead set• ST segments• Intra-aortic balloon assist (interactive)• Cardiac output
Dimensions (WxDxH) 25.4 cm x 33 cm x 10.2 cm) (10 in x 13 in x 4.7 in )
Weight 3.2 kg (7.1 lb)
Included accessories2242959 Operator’s manual2392337 LCD pendant controller2426360 Universal Battery ChargerXXXXXXX Power Cord (country specific)
Ordering informationDataSim 6100 Patient Simulator2247742 United States, 120 V2395218 Australia, 250 V2395229 Denmark, 250 V2395234 Schuko, 250 V2395241 Israel, 250 V2395252 Italy, 250 V2395265 India, 250 V2395276 Switzerland, 250 V2395283 United Kingdom, 250 V
Optional accessoriesFor a complete list of blood-pressure cables and personality modules, contact us.
Patient Simulator
Patient Simulator
DataSim 6100
47
Hand Held Controller
HHC3
Included accessories2671068 Users Manual2712829 medSim 300B Serial Interface Cable2702279 MPS450 and Marq III Serial Interface Cable2702287 Serial Interface Cable HHC3 to Computer2671031 Utility Software2671046 medSim 300B Instruction Card2671022 MPS450 and Marq III Instruction Card2242666 4-way Stop Cock AdapterAA alkaline batteries (4)
Ordering information2645641 HHC3 Hand Held Controller
Optional accessories2720054 Battery Eliminator
Key features• Full-functionality control
of the simulator, up to 6.1 m (20 ft) away
• Small and light weight• Factory-defined sequences
provided• Easy programming for
user-defined sequences• Ability to run defined
sequences repeatedly• PC interface for simple
configuration (utility)
The HHC3 Hand Held Controller is used to remotely operate medSim 300B, MPS450 and Marq III simulators in an easy and efficient manner. The HHC3 has all the output controls for these simulators and enables current simulator users, including hos-pital biomedical technicians and manufacturers, to simplify and standardize their testing, training, and preventive maintenance protocols. In addition, the HHC3 is an excellent device for biomedical training and demonstration use.
The HHC3 facilitates the direct selection of parameters for the Fluke Biomedical medSim 300B,
MPS450, and Marq III simulators. The HHC3 uses flexible coiled cable to connect to a simulator. The HHC3 provides single-key commands, dual-key commands, factory-defined sequences and easy programming of user-defined sequences. Customers can use the HHC-Utility software to upload user-defined sequences from a PC and download the sequences to multiple controllers.
Power requirements
medSim 300B RS-232 cable supplies power to the HHC3
MPS450 Four alkaline AA batteries or battery eliminator
Marq III Four alkaline AA batteries or battery eliminator
Battery power supply
Four alkaline AA cells, non-rechargeable voltage
1.5 V dc x 4 V dc
Battery life (continuous use) 60 hours
Battery eliminator supply Output voltage: 9 V dc
Output current: 50 mA
General information
Display 2 x 16 LCD, adjustable viewing angle
Controls 20 control keys and ON/OFF power switch
Embossed keys in 4 x 5 matrix
Interface RS-232 bidirectional interface
Auto connect to simulator parameters
Altitude Up to 2000 m
Dimensions (LxWxH) 3.6 cm x 8.1 cm x 16 cm (1.4 in x 3.2 in x 6.3 in )
Weight (with batteries) 0.36 kg (0.8 lb)
Specifications
Controller48
Key features• Mechanical heart for ultra-
sound simulation• TOCO simulation (External or
IUP)• Ultrasound simulation
(including twins)• Maternal ECG simulation• Fetal ECG (tracks ultrasound #1)• Internal (DECG) and external
fetal ECG• Uterine-activity selections• Fetal beat-to-beat variability• Periodic and non-periodic
fetal ECG changes• Arrhythmia selections• Compact, lightweight,
pocket-size plastic housing• Battery operated with status
indications• Special kits available with
all required accessories and cables to test fetal monitors for specified manufacturers
The PS320 simulates fetal and maternal ECG as well as uterine activity to test and troubleshoot fetal electronic monitors and to train clinical staff.
The unit is battery operated and small enough to fit in a pocket so mobile technicians and clinical instructors can take it anywhere.
The PS320 simulates several fetal parameters, including twins, as well as a wide range of clinical scenarios for training
labor-and-delivery staff in how to recognize normal and abnormal responses. An optional mechanical heart creates fetal heart sounds for testing fetal monitor ultrasound cables and transducers.
PS320 offers an easy user interface, providing a 2 x 16-character LCD display with adjust-able contrast. The unit operates on a 9 V battery with low-battery monitoring or functions with the supplied battery eliminator.
Specifications
Fetal ECG
Static rates 30 BPM, 60 BPM, 90 BPM, 120 BPM, 150 BPM, 180 BPM, 210 BPM, and 240 BPM
ECG sensitivity 50 μV, 100 μV, 200 μV, 0.5 mV, 1 mV, and 2 mV
US-1 tracks primary fetal ECG rates
US-2 tracks secondary fetal activity for either independent “normalor “twinssimulation, US-2 rate fixed at 140 BPM
Fetal patternsNote: US-1 and fetal ECG track these selections. US-2 is in normal pattern, except during TREND #1 selection.
Trend #1: Twin fetal patterns
Normal: Normal patternTachycardia: Tachycardia patternBradycardia: Bradycardia patternArrhythmias: Arrhythmia patternLate deceler.: Late deceler.Early deceler.: Early deceler.Moderate deceler.: Moderate variable deceler.Acceler.: #1: Acceler. wave #1Acceler.: #2: Acceler. wave #2Sinusoidal (high): Sinusoidal pattern, large changeSinusoidal (low): Sinusoidal pattern, small changeSevere var. deceler.: #1: Severe deceler. wave #2Severe var. deceler.: #2: Severe variable decelerProlonged deceler.: Prolonged decelerBiphasic deceler.: Biphasic decelerExaggerated deceler.: Exaggerated decelerNon-uniform deceler.: Non-uniform decelerVar. deceler. (u): Variable deceler, “U shapedVar. deceler. Tach: Variable deceler. with high-rate BPMVar. deceler. (v): Variable deceler. “V shapedVar. deceler. (post): Variable deceler. post exaggeratedVar. deceler.: Variable deceler.Deceler. (position): Variable deceler. with position changesLong deceler.: Long deceler.Compensatory acceler.: Compensatory acceler
Optional accessories2647372 Battery Eliminator 100 V ac to 240 V ac2462177 Carrying Case, Double Pocket2462478 Philips 50 Series—Ultrasound Cable2462491 Agilent 50 Series TOCO—External Cable 2462528 Agilent 50 and 8040 Series TOCO—IUP Cable2462469 Corometrics TOCO—External Cable2462484 Corometrics—Ultrasound Cable2462519 Corometrics TOCO—IUP Cable2462528 HP/AG/PHILIPS IUP TOCO Simulation Cable2462537 HP (8040 Series) Ultrasound Simulation Cable2462543 HP (8040 Series) Ext TOCO Simulation Cable2462555 2462562 Oxford Ultrasound Simulation Cable 2.0 MHz (blue)2462570 Oxford IUP Simulation Cable2462217 RS-232 Cable2651757 Mechanical Fetal Heart Probe2462123 Mechanical Fetal Heart Cable
Fetal Simulator
PS320
Fetal Simulator 49
Ordering information2583030 PS320 Fetal SimulatorKit #1: GE Corometrics2794057 PS320 Fetal Monitoring Kit, GE Corometrics, includes:2583030 PS320 Fetal Simulator (includes Operator Manual (2631693), Battery Eliminator (2647372), 9 V Battery)2651757 MFH-1 Mechanical Fetal Heart Probe [includes Fetal Heart Cable (2462123)]2397628 Soft-Sided Carrying Case for kit2462217 17291 RS-232 Cable, PS320/4202462484 Corometrics Ultrasound Simulation Cable2462469 Corometrics EXT. TOCO Simulation Cable2462519 Corometrics IUP TOCO Simulation CableKit #2: Hewlett-Packard, Agilent, Philips Medical2794069 PS320 Fetal Monitoring Kit, Series 50/8040 Philips Medical, Hewlett-Packard, Agilent, includes:2583030 PS320 Fetal Simulator (includes Operator Manual (2631693), Battery Eliminator (2647372), 9 V Battery)2651757 MFH-1 Mechanical Fetal Heart Probe [includes: Fetal Heart Cable (2462123)]2397628 Soft-Sided Carrying Case for kit2462217 RS-232 Cable, PS320/4202462478 HP/AG/PHILIPS (50 Series) Ultrasound Simulation Cable PS3202462491 HP/AG/PHILIPS (50 Series) EXT TOCO Simulation Cable PS3202462528 HP/AG/PHILIPS (50 & 8040 Series) IUP TOCO Simulation Cable PS3202462537 HP (8040 Series) Ultrasound Simulation Cable PS3202462543 HP (8040 Series) EXT TOCO Simulation Cable PS320Kit #3: Oxford Medical2794078 PS320 Fetal Monitoring Kit, Oxford Medical, includes:2583030 PS320 Fetal Simulator (includes Operator Manual (2631693), Battery Eliminator (2647372), 9 V Battery)2651757 MFH-1 Mechanical Fetal Heart Probe [includes: Fetal Heart Cable (2462123)]2397628 Soft-Sided Carrying Case for kit2462217 RS-232 Cable, PS320/4202462570 Oxford TOCO IUP Simulation Cable PS3202462562 Oxford Ultrasound Simulation Cable 2 MHz (blue) PS3202462555 Oxford Ultrasound Simulation Cable 1.5 MHz (yellow) PS320Kit #4: Spacelabs Medical2794040 PS320 Fetal Monitoring Kit, Spacelabs Medical, includes:2583030 PS320 Fetal Simulator (includes Operator Manual 2631693), Battery Eliminator (2647372), 9 V Battery)2651757 MFH-1 Mechanical Fetal Heart Probe [includes Fetal Heart Cable (2462123)]2397628 Soft-Sided Carrying Case for kit2462217 17291 RS-232 Cable, PS320/4202462581 Spacelabs Ultrasound Simulation Cable PS3202462596 Spacelabs TOCO Simulation Cable PS320
Fetal ECG (continued)
Variability selections (added to fetal ECG)
Absent variability, low variability, mild variability, high variability severe variability, long-term variabilityNote: These patterns repeat and toco channel will perform toco wave selected.
Optional mechanical heart
Provides a mechanical interface to the ultrasound trans-ducer; can be connected to either ultrasound channels. This option, due to its power consumption, requires an ac adapter to be connected.
Maternal ECG ECG static rates: 60 BPM, 80 BPM, 100 BPM, 120 BPM, 140 BPM, and 160 BPMECG sensitivity: 0.5 mV, 1 mV, and 2 mVPattern selected during Trend #1 selection
Uterine activity Note: Toco waveform selection not available during Trend #1.
Execute waveform: Start toco waveformUterine wave Off: Stop toco waveformAnalog 0 V TO 1 V: Analog range 0 V to 1 V (1 V = 100 mmHg)Uterine wave 0 to 25: Range of toco waveformUterine wave 0 to 50: Range of toco waveformUterine wave 0 to 100: Range of toco waveformShort duration: Toco waveform of short durationNormal duration: Normal duration of toco waveformIncreased duration: Long duration of toco waveformUterine level = Zero: Zero toco channel (automatic on power up)Uterine static +20: Increase toco static level by 20 mmHg (0 mmHg to 100 mmHg)Incr. resting tone: Resting tone increasesCouping: 2 close toco wavesTripling: 3 close toco wavesUterine pressure sensitivity: 5 μV or 40 μV on power up
Important notes • US-1 tracks the fetal ECG rates• US-2 is the second ultrasound channel with a normal
fetal ECG pattern• On the fetal and maternal ECG, the fetal ECG is 1/4
the size of the maternal ECG
The PS320 turns on in the following state:
• Fetal ECG static rate @ 150 BPM• US-1 tracks @ 150 BPM• US-2 normal pattern• Pressure sensitivity @ 5 μV/mmHg• Pressure/Toco set to zero• Maternal ECG rate @ 80 BPM• ECG sensitivity @ 1 mV• Toco wave is normal duration @ 0 to 50 divisions
(i.e. 0 mmHg to 50 mmHg)
Temperature
Operating 15 °C to 35 °C (59 °F to 95 °F)
Storage 0 °C to 50 °C (32 °F to 122 °F)
General information
Display 2-line x 16-character LCD with keypad
RS-232 Bidirectional interface, 9600 baud
Power 9 V battery/battery eliminator; low battery indication set at 6 V
Housing Plastic case
Dimensions 15.6 cm x 9.4 cm x 3.4 cm (6.1 in x 3.7 in x 1.3 in)
Weight 0.4 kg (0.9 lb)
Specifications
Fetal Simulator
Fetal Simulator
PS320
50
Patient Oximeter Simulator
Pulse Oximeter Simulator
Index 2
Key features• Portable• 10 preloaded manufacturers’
R-curves• User-definable “makesfor most other manufacturers• New R-curves for
Masimo, Nonin and Philips Medical Systems (formerly Agilent/H-P) oximeters
• Six downloadable R-curve spaces available
• Simultaneous simulation of motion and arterial-oxygen levels
• Arterial wave-amplitude scale, calibrated in units of perfusion
• Tap/shiver motion simulations to explore the impact of motion
• RS-232 port for computer control
• Physiological finger for complete SpO2 tests
• Electrical simulations with probe testing (optional)
The Index 2 is the most versatile optical simulator for oximeters on the market today. This lightweight, portable tool includes preloaded manufac-turers’ R-curves and the ability to define other “makesfor most pulse oximeters.
Motion presets, player mode, transmission level control (TLC), and computer commands boost testing ability. The Index 2 can also be configured to include an optional electrical simula-
tion feature with probe test. Optical and electrical simulations allow technicians to isolate problems quickly. The probe test identifies defective probes with quantitative test results.
Included accessories2226196 Operator’s Manual2521465 Battery ChargerNellcor and Ohmeda Electrical Simulation and Probe Test Cable (for Index2XLFE only)
Ordering informationIndex 2XLF Pulse Oximeter Simulator – Optical Finger Simulation2250232 United States, 120 V2395290 Shuko, 250 V2399900 Australia, 250 V2399917 United Kingdom, 250 V2447476 Japan, 100 VIndex 2XLFE Pulse Oximeter Simulator – Optical Finger and Electrical Simulation with Probe Test2250244 United States, 120 V2395309 Shuko, 250 V2399921 Australia, 250 V2399939 United Kingdom, 250 V2447465 Japan, 100 V
O2 Range: 35 % to 100 %
Resolution: 1 %
Accuracy: 100 % to 75 %: ± 1 % ± accuracy of the pulse oximeter under test; 74 % to 50 %: ± 3 % ± the accuracy of the pulse oximeter under test; < 50 % unspecified
Repeatability: ± 1 standard deviation
Rate Range: 30 BPM to 250 BPM
Resolution: 1 BPM
Accuracy: 1 % ± 1 BPM
Pulse amplitude Range: 0 % to 100 % of nominal pleth amplitude
Resolution: 1 %
Pulse amplitude is 20 % of maximum pass-through brightness
Probe test Continuity/resistance test matrix: Measures all combinations of possible interconnections in an XX point matrix
Range: 250 Ω to 150 kΩAccuracy: ± 5 % of reading
LED/detector voltage test
Test format: Measures the voltage drop across Red LED, infrared LED, and the photo detector when the internally generated test signal is applied
Test signal: Constant current source @ 1 mA
Open circuit: 2.5 V max
Measurement/display range: 0 V to 4 V
Accuracy: ± 5 % of reading, 0.4 V to 4 V
Dynamic test Test format: Photodetector/diode response to both the red and infrared light generated by the probe when pulsed by an internal test signal
Test signal: Pulsed between the two LEDs; constant current level @ 1 mA
Test results: Nominal range of 0 to 2000
Checksum Sum of all locations in the program chip; for service use only
General information
Display 2-line x 24-character super twist LCD
Battery life At least four hours of continuous use
Dimensions (LxWxH) 45.7 cm x 40.6 cm x 22.9 cm (18 in x 16 in x 9 in)
Weight 4.5 kg (10 lb)
Optional accessories2204282 Soft Vinyl Carrying Case2200102 Interface Cable, medTester to Index 2 (RS-232; female DB25 to female DB9)For a complete list of electrical simulation and probe test cables, contact us.
Specifications
51
Specifications
Included accessoriesOperator’s/Service Manual2397628 Soft-sided Carrying Case
Ordering information2387318 ACCU LUNG Precision Test Lung
Key features• Portable (light weight,
small footprint) • User-selectable compliance
and resistance settings (three selections each)
• Calibrated accuracy for both resistance and compliance
• Complies with IEC standard for breathing-circuit connection
• Certified test lung for test system traceability to standards
The ACCU LUNG Precision Test Lung is a lung simulator that presents a specific load comprised of a user-selectable compliance and resistance for the purpose of evaluating ventilator performance according to clinical expectation and manufacturers’ specification. It is a portable unit that can be hung from a cart, the ventilator itself, or can be hand-held, thus presenting a “zerofootprint.
Environmental parameters
Operating temperature 10 °C to 40 °C
Storage temperature 0 °C to 50 °C
Performance characteristics
Static compliance C50 0.5 l/kPa ± 10 % at 500 ml tidal volume
C20 0.2 l/kPa ± 10 % at 500 ml tidal volume
C10 0.1 l/kPa ± 10 % at 300 ml tidal volume
Resistance Parabolic (orifice) resistor pressure drops selected from ASTM F1100 (K values), for inspiratory flows at 2, 1, and 0.5 l/s, respectively.
Rp5 K 2.70 ± 20 % (equivalent orifice size = 8.48 mm) pressure drop 10.80 cmH2O at 2 l/s
Rp20 K 17.61 ± 20 % (equivalent orifice size = 5.31 mm) pressure drop 17.61 cmH2O at 1 l/s
Rp50 K 108.70 ± 20 % (equivalent orifice size = 3.37 mm) pressure drop 27.20 cmH2O at 0.5 l/s
Physical characteristics
Ventilator circuit connection
ISO 22 mm female
Warranty specifications
15 month extended warranty on all parts and labor with the following limitations:a) All rubber parts (including bellows made from Hypalon®)
are warranted to be free from defects at the time of deliveryb) Springs are considered limited lifecycle parts and
are expected to survive 1 x 106 cycles
Dimensions (LxWxH) 27.9 cm x 21.6 cm x 10.2 cm (11 in x 8.5 in x 4 in)
Weight 1.8 kg (4 lb)
Gas Flow Analyzers
Portable Precision Test Lung
ACCU LUNG
52
Key features• Bidirectional flow (high- and
low-flow ranges), volume, vacuum, pressure and oxygen concentration measurements
• 16 ventilator parameter measurements
• Trending and statistical analysis of all measured values
• Onboard graphical display• Portable and compact• RS-232 for computer control• Memory for storing results• VT for Windows PC software• Optional sensor assembly for
temperature and humidity measurements
Specifications
The VT MOBILE is a compact and portable general purpose gas-flow analyzer designed to meet the needs of the traveling technician or engineer. This versatile tool evaluates performance of a wide variety of medical gas-flow/pressure devices and measures 16 ventilator parameters.
EC.6.20 now requires completion of 100 % of life-support device preventive maintenance every year. VT MOBILE can help you meet those requirements.
The base unit measures high- and low-flow ranges, volume, pressure, and oxygen concentration. Additionally, the temperature and relative humidity option can be ordered separately to ensure the most accurate gas-flow measurements.
Display 64 pixels x 128 pixels, reflective LCD, blue on yellow
Gas types Air, N2, N2O, CO2, O2, N2O bal O2, N2 bal O2
Battery power supply
Input voltage range: 9 V dc
Battery life: > 7 hours
Dimensions (LxWxH) 20 cm x 10 cm x 3.8 cm (8 in x 4 in x 1.5 in)
Weight 0.45 kg (1 lb)
Low-pressure port High-pressure port Airway pressure
Maximum applied pressure
5 psi 125 psi 5 psi
Operating ressure -20 cmH20 to 120 cm H2O
-2 psi to 100 psi -20 cmH2O to 120 cmH2O
Span accuracy +2 % of reading or 1.5 mmHg
+2 % of reading or +0.2 psig
+2 % of reading or +0.5 cmH2O
Model VT MOBILE VT PLUS HF
Flow range 25 to 200 l/min (high flow sensor) 0 to 25 l/min (low flow sensor)
25 to 300 l/min (high flow channel) 0 to 25 l/min (low flow channel)
Features and benefits
Portable, battery-powered, all ranges of pressure, flow, temperature, and RH measurement, easy-to-use
Bench-top or portable, line-powered, all ranges of pressure, flow, easy-to-use
16 ventilator parameters on three screens 21 ventilator parameters on one screen
On-screen pressure, flow and volume waveforms On-screen pressure, flow and volume waveforms
Ventilator and non-ventilator flow measurements Ventilator and non-ventilator flow measurements
On-board memory for temporary test result storage On-screen pressure, flow and volume waveforms
— Special modes for High Frequency ventilators and RT-200 emulation
Compatible with VT for Windows® PC software (standard accessory)
Compatible with VT for Windows® PC software (standard accessory)
Product comparison chart
Gas Flow Analyzers
Portable Gas-Flow Analyzer
VT MOBILE
VT MOBILE Tilt Stand in low-tilt position
53
Included accessories2548405 Accessory Kit2544903 CD, includes: quick-reference card, operators manual, getting-started manual, other matter2544892 Getting Started Manual2544630 Quick Reference Card2548431 High-flow Sensor2548422 Low-flow Sensor2548315 High-pressure Adapter, Male to Female2454175 Low-pressure Adapter2448051 Oxygen-sensor Cable, 6 ft2457028 T Adapter for Oxygen Sensor2248801 Oxygen Sensor614487 9 V dc Battery (alkaline)2075257 Serial Communications Cable (RS-232), 6 ft2558269 VT for Windows® PC Software2551236 Soft Carrying Case
Ordering informationVT MOBILE Portable Gas-Flow Analyzer2427911 United States, English overlay2553550 France, French overlay2542531 Denmark, German overlay2542546 Italy, Italian overlay2542554 Spanish, Spanish overlay2553610 Japan, Japanese overlay2553605 Chinese, Chinese overlay
High-flow port Low-flow port
Operating flow range ± 200 lpm +25 lpm
Accuracy ± 3 % of reading or ± 2 % of range +3 % of reading or +1 % of range
Floor for absolute accuracy 25 lpm 3 lpm
Low-flow dropout 2.5 lpm 0.24 lpm
Volume range > ± 60 l +60 l
Tidal volume accuracy
± 3% of reading or ± 20 ml, whichever is greater
+3 % of reading or +2 ml
Oxygen measurement Barometric pressure measurement
Range 0 % to 100 % 8 psia to 18 psia (400 mmHg to 900 mmHg)
Accuracy +2 % full-scale output +2 % of reading
Sensor technology Galvanic fuel cell —
Calibration Allows user calibration using air and 100 % O2
Not required; however, device allows user calibration of offset
Notes:• Automatic partial pressure compensation for barometric and airway pressure changes.
• Recommended interval for changing oxygen sensor is one year. However, sensor may last longer. During user calibration of the sensor, the VT MOBILE can detect if the sensor needs to be replaced.
Secondary parameter-accuracy specifications
Resolution Range Accuracy
Inspiratory and expiratory tidal volume
0.1 ml > 10 l ± 3 % expiratory minute volume
0.001 lpm 0 l to 60 l ± 3 %
Breath rate 0.1 BPM 2 BPM to 150 BPM ± 1 % inspiratory-to-expiratory time ratio (I:E ratio)
0.01 Range: 0.25 to 9.99
— ± 2 % or 0.1 s
Peak inspiratory pressure 0.1 cmH2O ± 120 cmH2O +3 % or 1 cmH2O
Inspiratory pause pressure 0.1 cmH2O ± 120 cmH2O +3 % or 1 cmH2O
Mean airway pressure 0.1 cmH2O +80 cmH2O +3 % or 0.5 cmH2O
Positive-end expiratory pressure (PEEP)
0.1 cmH2O -5 cmH2O to 40 cmH2O
+3 % or 0.5 cmH2O
Peak expiratory flow 0.01 lpm 0 lpm to 150 lpm ± 3 % or 2 % of range
Temperature 0.1 °C 0 °C to 50 °C ± 1 °C
Units: °C, °F, °K
Humidity 0.1 % 0 % to 100 % ± 5
RS-232 serial communications
4-pin modular connector located on upper-left side of panel. RS-232 compatible with the VT Plus for Windows software application (version 2.01.00 or higher.)
Environmental specifications
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -25 °C to 50 °C (-13 °F to 122 °F)
Operating humidity 0 % to 80 % non-condensing at temperatures to 31 °C, decreasing linearly to 50 % relative humidity at 40 °C (104 °F)
Storage humidity 0 % to 95 % non-condensing
Operating barometric 7 psia to 18 psia
Storage barometric 787.9 mmHg to 522.7 mmHg (-1000 ft to 10000 ft)
Specifications Optional accessories2548405 Accessory Kit2548431 High-flow Sensor2548422 Low-flow Sensor2548303 High-pressure Adapter, male to male2548315 High-pressure Adapter, male to female2454175 Low-pressure Adapter2541622 Temperature and RH Sensor, Cable and T Adapter, 6 ft2457028 Oxygen-Sensor Cable, 6 ft2448051 T Adapter for Oxygen Sensor2558269 VT for Windows® PC Software2075257 Serial Communications Cable (RS-232), 6 ft2547455 Power adapter, universal (USA and international)2551236 Soft Carrying Case2387318 ACCU Lung Portable Precision Test Lung
Gas Flow Analyzers
Portable Gas-Flow Analyzer
VT MOBILE
54
Key features• Bidirectional flow,
pressure, volume, and oxygen concentration, and pressure measurements
• Low- and high-pressure, and flow measurement capability
• Special HF mode—up to 900 BPM (15 Hz)
• RS-232 and printer ports• Included Windows-compatible
graphics software• All 21 ventilator parameters
displayed at once on one screen
• Operation by user-friendly VT PLUS HF command mode or special RT-200 command mode
• Minimum, maximum, average, absolute, and graph for all parameters
• Multiple special-function tests for efficient troubleshooting
Optional features• Operation with a variety
of precision test lungs available from Fluke Biomedical to complete a fully NIST- traceable ventilator testing system
The VT PLUS HF is Fluke Biomedical’s premier general-purpose gas flow analyzer. In addition, special display modes and bidirectional flow make it perfect for fully and efficiently testing both conventional mechanical ventilators and high-frequency ventilators. EC.6.20 now requires 100 % completion of scheduled life-support device preventive maintenance every year, and VT PLUS HF can help meet
those requirements. Multiple special-function tests make troubleshooting quick and efficient.
Learning to use the VT PLUS HF is simple. Technicians control the unit using the VT PLUS HF user-friendly command system, or, if they’re familiar with the RT-200, they can switch to a special control mode that uses RT-200-style commands.
Low-pressure High-pressure Airway-pressure
Range ± 500 mmHg (10 psi) ± 100 psi ± 120 cmH2O
Accuracy ± 0.5 % of reading or ± 1.5 mmHg, which-ever is greater
± 1 % of reading or ± 0.1 psig, whichever is greater
± 0.75 % of read-ing or ± 0.5 cmH2O, whichever is greater
Note Fluid pressure may be applied to the positive port; however, fluids should be kept from entering the pressure port by using a suitable length of connec-tion tubing.
Airway pressure is internally tapped off the proximal-flow sensor port, which is the port closest to the exhaust port on the VT PLUS HF
Low-flow High-flow
Flow range -25 lpm to 25 lpm -300 lpm to 300 lpm
Accuracy ± 2 % of reading or ± 1 % of range, whichever is greater
± 2 % of reading or ± 2 % of range, whichever is greater
Low-flow dropout 0.01 lpm —
High-flow dropout — 25 lpm
Volume range > ± 60 l > ± 60 l
Notes • Tidal-volume accuracy: ± 3 % of reading or ± 2 ml, whichever is greater
• Volume accuracy tested to 1 liter
• Flow accuracy is specified for dry air or oxygen
• Below 3.0 lpm, measurement accuracy is obtained by allowing the VT PLUS HF to fully warm up or manually zeroing before reading or documenting measurement.
• Tidal-volume accuracy: ± 3 % of reading or ± 10 ml, whichever is greater
• Volume accuracy tested to 7 liters
• Flow accuracy is specified for dry air or oxygen
General
Dimensions (LxWxH) 25.4 cm x 25.4 cm x 12.7 cm (10 in x 10 in x 5 in )
Weight 4.53 kg (10 lb)
Specifications
Gas Flow Analyzer
VT PLUS HF
Gas Flow Analyzers
VT PLUS HF standard accessories
VT for Windows PC Software (PC not included)
55
Ventilator parameter Resolution Range Accuracy
Inspiratory and expiratory tidal volume 0.1 ml As specified in high-flow/low-flow specification
Expiratory minute volume 0.001 lpm 0 L to 60 L ± 3 %
Breath rate 0.1 BPM 0.5 BPM to 150 BPM ± 1 %
Inspiratory-to expiratory time ratio (I:E ratio) 0.01 1:200 to 200:1 ± 2 % or ± 0.1 s
Inspiratory time 0.01 s 0 s to 60 s ± 1 % or ± 0.02 s
Expiratory time 0.01 s 0 s to 90 s ± 1 % or ± 0.01 s
Peak inspiratory pressure 0.1 cmH20 ± 120 cmH2O ± 3 % or ± 1 cmH2O
Inspiratory pause pressure 0.1 cmH20 ± 120 cmH2O ± 3 % or ± 1 cmH2O
Mean airway pressure 0.1 cmH20 ± 80 cmH2O ± 3 % or ± 0.5 cmH2O
Positive end-expiratory pressure (PEEP) 0.1 cmH20 -5 cmH2O to 40 cmH2O ± 3 % or ± 0.5 cmH2O
Inspiratory hold time 0.01 s 0 s to 60 s ± 1 % or ± 0.1 s
Expiratory hold time 0.01 s 0 s to 90 s ± 1 % or ± 0.1 s
Peak expiratory flow 0.01 lpm 0 lpm to 300 lpm ± 3 % or ± 2 lpm
Peak inspiratory flow 0.01 lpm 0 lpm to 300 lpm ± 3 % or ± 2 lpm
Lung compliance 0.1 ml/cmH20 0 ml/cmH2O to 150 ml/cmH2O ± 5 % or ± 5 ml/cmH2O
Inspiratory pause time: > 0.5 s
Flow bias 0.01 lpm 0 lpm to 30 lpm ± 2 % or ± 0.5 lpm
Expiratory pause time: > 0.5 s
Specifications
Optional accessories2222822 Soft Vinyl Carrying Case for VT PLUS HF2248587 Hard-sided Protective Carrying Case for VT PLUS HF (limited to stock on hand)2397628 Soft-sided Carrying Case for ACCU LUNG
Test Lungs2387318 ACCU LUNG Portable Precision Test Lung (with Soft-sided carrying case for ACCU LUNG, model 2397628)2251049 Michigan Instruments Non-instrumented Single-adult Test Lung2251008 Michigan Instruments Non-instrumented Dual-adult Test Lung2251013 Michigan Instruments Non-instrumented Adult/Infant Test Lung2213774 Siemens 190 Test Lung
Parabolic Airway Resistors (for use with Michigan Instruments test lungs)2212830 Parabolic Airway Resistor: RP52212934 Parabolic Airway Resistor: RP102212848 Parabolic Airway Resistor: RP202212853 Parabolic Airway Resistor: RP502212918 Parabolic Airway Resistor: RP2002213140 Parabolic Airway Resistor: RP500
Printers2248762 Printer 110 V, Citizen IDP 31102719653 Printer 220 V, Citizen IDP 31102238072 Parallel Printer Cable, D25M-C36M
Included accessories2137275 Operator’s Manual2392054 VT for Windows® PC Software2238659 Serial Cable2133387 Tilt StandPower cord (country specific)2131367 Accessory Kit (includes 16 accessories)
Ordering informationVT PLUS HF Gas Flow Analyzer2128272 United States, 120 V2399376 Australia, 250 V2399383 Schuko, 250 V2399390 United Kingdom, 250 V
Premium Precision Ventilator Test Kit(VT PLUS HF Gas Flow Analyzer; and ACCU LUNG portable precision test lung)2387329 United States, 120 V2425682 Australia, 250 V2425694 Schuko, 250 V2425701 United Kingdom, 250 V
VT-Plus Upgrades(adds HF capability and RT-200 mode)2240945 VT PLUS HF hardware and firmware factory service upgrade (for units lower than hardware v1.01.01; additional flat-rate charge required for factory service/calibration)
Accessory Kit Parts2133712 Filter, External (bacterial), 1 each2391777 Adapter, DISS 02 nut and nipple with 1/4 in I.D. hose barb, 1 each2133310 Tubing Adapter, Directional (15 mm OD x 15 mm OD), 2 each2133305 Tubing Adapter (22 mm OD x 22 mm ID), 2 each2133291 Tubing Adapter (22 mm OD x 22 mm OD), 2 each2133269 Tubing Adapter (15 mm OD x 22 mm OD), 2 each2133278 Tubing Adapter (15 mm OD x 15 mm OD), 2 each2133284 Tubing Adapter (15 mm ID x 15 mm OD), 2 each2133322 Tubing Adapter, Narrow Bore, 2 each2213679 Barb (luer lock – male to 1/89 in ID tubing), 2 each2133240 Tubing Adapter (1/4 in NPT male to 1/8 in ID tubing barb fitting), 2 each2133202 Tubing Adapter (luer lock 1/16 in to bulkhead connection), 2 each
Gas Flow Analyzer
VT PLUS HF
Gas Flow Analyzers56
Key features
All models• Palm size• High accuracy• Differential pressure,
vacuum, and temperature measurements
• Multiple user-selectable units of measurement
• Simultaneous display of multiple parameter measurements
• Leak-detection/leak-rate calculation
• Peak test function to capture peak pressure
• RS-232 for computer control
Model 1G• Pressure measurements in
-700 mmHg to 5000 mmHg range
Model 1H• Pressure measurements in
-350 mmHg to 350 mmHg range
Model 2G• Barometric pressure, gas flow,
and humidity measurements• Pressure measurements in
-700 mmHg to 5000 mmHg range
Model 2H• Barometric pressure, gas flow,
and humidity measurements• Pressure measurements in
-350 mmHg to 350 mmHg range
The versatile DPM4 tests and calibrates flow and pressure generators used in many medical devices. With several measurements combined in a single, handheld device, the DPM4 provides a cost-effective solution, eliminating the need for multiple test meters.
The DPM4 features a menu-driven interface for simple operation and an easy-to-read screen that displays multiple parameter measurements simultaneously.
Model 1H or 2H
Pressure measurement
Operating range -350 mmHg to 350 mmHg
Accuracy ± 0.3 % of range
Units of measure mmHg, mBar, cmH2O, psi, inHg, inH2O, kg/cm2, and kPa
Model 1G or 2G
Pressure measurement
Operating range -700 mmHg to +5000 mmHg
Accuracy ± 0.3 % of range for temperatures from 21 °C to 25 °C and relative humidity from 30 % to 70 % ± 0.3 % of range;
± 0.02 % of range per degree °C for temperatures < 21 °C or > 25 °C with relative humidity from 30 % to 70 %
Units of measure mmHg, mBar, cmH2O, psi, inHg, inH2O, kg/cm2, and kPa
Temperature measurement (with optional temperature probe)
Operating range -40 °C to 200 °C (-40 °F to 392 °F)
Accuracy ± (2 % of reading, + 0.5 °C)
Temperature units °C, °F
Specifications
N10140
Pressure Meters
Parameter Tester
DPM4
57
Included accessories2572323 Users Manual2647372 Battery EliminatorXXXXXXX Power Cord (country specific)XXXXXXX One 9 Volt Alkaline Battery
Ordering informationDPM4 Parameter Tester2583121 Model 1H (± 350 mmHg)2631330 Model 1G (-700 to 5000 mmHg)2637760 Model 2H (± 350 mmHg, Press, Temp, Flow, RH)2637772 Model 2G (-700 to 5000 mmHg, Press, Temp, Flow, RH)
Optional accessories2462177 Soft-Sided Carrying Case2461910 PT-100 Temperature Probe2461922 PT-1000 Temperature Probe2461905 Expansion Chamber2461946 Tubing Kit with Inflation Bulb2462335 RS-232 Cable
Temperature Probe PT-100 and PT-1000
PT-100 Operating range -200 °C to 750 °C (-328 °F to 1382 °F)
Accuracy ± 0.13 °C @ 100 °C (0.23 °F at 212 °F)± 0.1 °C @ 0 °C (0.18 °F @ 32 °F)± 0.2 °C @ 100 °C (0.36 °F @ 212 °F)
PT-1000 operating range -200 °C to 750 °C (-328 °F to 1382 °F)
Accuracy 0.3 °C (0.5 °F)
Model 2G or 2H Note: It is possible to compensate for the sea level and calibrate for offsets.
Pressure measurement
Operating range 380 mmHg to 825 mmHg
Accuracy ± 2 % of reading
Units of measure mmHg, mBar, cmH2O, psi, inHg, inH2O, kg/cm2, and kPa
Relative humidity Note: An integrated sensor in the instrument determines relative humidity measurements.
Operating range 12 % RH to 95 % RH
Accuracy ± 3.5 % of reading ± 2 % @ 25 °C (77 °F)
Gas flow Note: Gas flow measures with an embedded sensor with 11 calibration points to compensate non-linearity: calibration constants are stored in firmware.
Gas compatibility Air, N2, O2, CO, NO, CO2, N2O, NO2
Operating range -750 ml/min to 750 ml/min
Accuracy ± 1 % of range or ± 5 % of reading
Gas flow units ml/min or sccm (Standard Cubic Centimeter per Minute)
Peak flow test Peak flow is captured in the unit selected for flow. A reset key allows to restart the test.
Model 1G, 1H, 2G and 2H
Leak test and peak test
Leak test Leak rate is computed in the unit selected for pressure over 15, 30, 45 or 60 seconds
Peak test Peak pressure is captured in the unit selected for pressure. A reset key allows to restart the test.
Temperature
Operating 15 °C to 35 °C (59 °F to 95 °F)
Storage 0 °C to 50 °C (32 °F to 122 °F)
General information
Power 9 V alkaline battery RG9 or battery eliminator
Battery life > 7 hours
Dimensions (LxWxH) 156 mm x 94 mm x 34 mm (6.1 in x 3.7 in x 1.3 in)
Weight 0.4 kg (0.9 lb) with battery
Specifications
Pressure Meters
Parameter Tester
DPM4
58
Included accessories2572314 Users Manual2242666 3-way Stop Cock Adapter 6144879 9 V alkaline battery
Ordering information2249779 DPM 1B Pneumatic Transducer Tester
Key features• Battery operated• Generates and measures
positive or negative pressures • Operates with gas and liquid• Troubleshooting with 1 %
accuracy
The DPM1B pneumatic transducer tester is designed to measure the positive and negative pressures of medical devices in either liquid or gaseous form, and to generate pressure within the ± 300 mmHg range to assist in repair and quality control.
Specifications
Pressure measurement
Operating range ± 300 mmHg
Pressure generation range
± 300 mmHg
Accuracy ± 1 % of reading or ± 1 mmHg
Resolution 0.1 mmHg
Units of measure mmHg
Environmental requirements
Operating temperature 10 °C to +40 °C (+50 °F to +104 °F)
General information
Display/control 0.5 in LCD with LO BATT indication
Power 9 V alkaline battery
Dimensions (WxDxH) 15.9 cm x 14.6 cm x 3.8 cm (3.6 in x 5.8 in x 1.5 in)
Weight 260 g (10 oz)
Model DPM1B DPM2Plus DPM4-1G
Pressure measurement range
-300 mmHg to +300 mmHg
-698 mmHg to +802 mmHg-949 cmH2O to +1090 cmH2O
-374 inH2O to +429 inH2O-13.50 psi to +15.50 psi-13.50 psi to +100.00 psi
-700 mmHg to +5000 mmHg-950 cmH2O to +6797 cmH2O-374 inH2O to +2678 inH2O
-13.5 psi to +96.7 psi-93.4 kPa to +666 kPa
Pressure units 1 unit: mmHg 4 units: mmHg, cmH2O, inH2O, psi 8 units: mmHg, cmH2O, inHg, inH2O, psi, mBar, kg/cm2, and kPa
Pressure Generation Internal With optional inflation bulb With optional inflation bulb
Gas/Liquid operation Both Both Gas only
Temperature measurement — — With optional temperature probe in °C or °F
Barometric pressure — — —
Relative Humidity measurement — — —
Gas flow measurement — — —
Product comparison chart
Pressure Meters
Pneumatic Transducer Tester
DPM1B
59
Included accessories2572323 Operator’s Manual2242653 Catheter Adapter2249768 One 9 V Alkaline Battery
Ordering information2249768 DPM2Plus Pressure Meter
Optional accessories2242653 Catheter Adapter2249177 Parabolic Flow Adapter2200116 Phone to BNC Cable
Key features• Five selectable pressure
ranges• Voltage output to drive a
recorder for assessing electronics of pressure- measurement circuit
• Capability to test ophthalmology equipment, lasers, dialysis machines, automatic tourniquets, drainage devices, IV pumps, pressure gauges, ventilators, diagnostic, surgical suction devices, and more
• Air or liquid measurement
The DPM2Plus Pressure Meter is designed to measure the positive and negative pressures of med-ical devices in either liquid or gaseous form to assist in repair and quality control.
When coupled with the optional Parabolic Flow Adapter accessory, the displayed pressure can be interpreted, using the look up table supplied with the parabolic flow adapter to deter-mine flow from medical devices.
Operating range -698 mmHg to 802 mmHg-949 cmH2O to 1090 cmH2O-374 inH2O to 429 inH2O-13.50 PSI to 15.50 PSI-13.5 PSI to 100 PSI
Accuracy ± 1 % of range
Units of measure mmHg, cmH2O, PSI, inH2O
Temperature operating requirements
0 °C to 30 °C (32 °F to 86 °F)
Display/control 0.5 in LCD with LO BATT and negative polarity indication
Data outputs V/psi (all ranges except 100 psi) 0.01 V/psi for 100 psi range
Power 9 V alkaline battery
Dimensions (LxWxH) 14.61 cm x 9.14 cm x 4.83 cm (5.75 in x 3.6 in x 1.9 in)
Weight 0.4 kg (1 lb)
Specifications
Model DPM4-1H DPM4-2G DPM4-2H
Pressure measurement range -350 mmHg to +350 mmHg-475 cmH2O to +476 cmH2O-187 inH2O to +187 inH2O
-6.8 psi to +6.8 psi-46.7 kPa to +46.7 kPa
-700 mmHg to +5000 mmHg-950 cmH2O to +6797 cmH2O-374 inH2O to +2678 inH2O
-13.5 psi to +96.7 psi-93.4 kPa to +666 kPa
-350 mmHg to +350 mmHg-475 cmH2O to +476 cmH2O-187 inH2O to +187 inH2O
-6.8 psi to +6.8 psi-46.7 kPa to +46.7 kPa
Pressure units 8 units: mmHg, cmH2O, inHg, inH2O, psi, mBar, kg/cm2, and kPa
8 units: mmHg, cmH2O, inHg, inH2O, psi, mBar, kg/cm2, and kPa
8 units: mmHg, cmH2O, inHg, inH2O, psi, mBar, kg/cm2, and kPa
Pressure Generation With optional inflation bulb With optional inflation bulb With optional inflation bulb
Gas/Liquid operation Gas only Gas only Gas only
Temperature measurement With optional temperature probe in °C or °F
With optional temperature probe in °C or °F
With optional temperature probe in °C or °F
Barometric pressure — Yes, 4 units: mmHg, inHg, mBar, and hPa
Yes, 4 units: mmHg, inHg, mBar, and hPa
Relative Humidity measurement — Yes Yes
Gas flow measurement —
-750 ml/min to +750 ml/minCompatible with Air, N2, O2, CO,
NO, CO2, N2O, NO2
-750 ml/min to +750 ml/minCompatible with Air, N2, O2, CO,
NO, CO2, N2O, NO2
Product comparison chart
Pressure Meters
Pressure Meter
DPM2Plus
60
How well do your PM Inspection and post-repair performance-testing processes eliminate sources of human error? Wish that all technicians woulddocument results the same way?Do you have enough time to complete all PM Inspection and repair work on your shelf?
Ansur offers a solution:Repeatability—Creates standard work since everyone tests the same way every timeQuality—Can automatically configure and col-lect data from the compatible test devices to minimize human error and save time
Key features• General framework software
for performing all types of tests and inspections
• Remote control of Fluke Biomedical testers, and acquisition of test results via RS-232
• Manual/visual tests, performance tests, and electrical-safety tests all executed in one procedure
• Test-procedure and test-result files stored in industry-standard XML format
• Interface capability with some equipment management systems and computerized maintenance management systems
• Ready-to-use or customized test templates make creating standard work easy
• Compatible with a variety of test instruments by Fluke Biomedical for easy standardization
Specifications
Test Automation Software
Ansur
Automation Solutions
Productivity—Ensure that the amountof time required to perform testing is uniform and therefore predictableAnsur test automation system collects all the observe-and-record manual entries as well as automated measurements from compatible sim-ulators and performance analyzers from Fluke Biomedical.
Automate with AnsurLook for this logo in the Fluke Biomedical product catalog to see where test automation can benefit you.
PC requirements 64 MB RAM
50 MB unused hard drive space for software
IBM PC/XT compatable Pentium 266 MHz or faster processor
Hard drive space for result and template files
32-bit Microsoft Windows® operating system (2000/XP)
RS-232 ports or USB-RS-232 adapter
Other requirements License key for each Fluke Biomedical or Metron simulator/analyzer plug-in (accesses full functionality of Ansur and its plug-ins)
One or more Ansur-compatible Fluke Biomedical or Metron simulators/ analyzers (ensures best results for minimizing human error and opportunity for best productivity)
61
Automation Solutions
Test Automation Software
Ansur
Ordering information2462982 Ansur Test Executive
Plug-ins available:Purchase the modules you need, and then add modules as you acquire new Fluke Biomedical analyzers and simulators.2755836 BP Pump 22556755 ESA6013454829 ESA6123116463 ESA6202817641 Impulse 40003091370 Impulse 6000D/7000DP2461775 QA-40M/452463016 QA-902462644 QA-12902461802 QA-ES2461979 QA-IDS/lagu2463002 QA-ST2462024 QA-VTM2817226 QED 63337356 TNT 12000
Test Automation Bundles available: Purchase modules along with the Fluke Biomedical instrument of choice3334821 BP Pump 2L NIBP, US 120V w/Test Automation3334839 BP Pump 2l NIBP, AUS 250V w/Test Automation3334842 BP Pump 2L NIBP, DEN 250V w/Test Automation3334856 BP Pump 2L NIBP, SHK 250V w/Test Automation3334863 BP Pump 2L NIBP, ISR 250V w/Test Automation3334874 BP Pump 2L NIBP, ITAL 250V w/Test Automation3334888 BP Pump 2L NIBP, IND 250V w/Test Automation3334895 BP Pump 2L NIBP, SWZ 250V w/Test Automation3334901 BP Pump 2L NIBP, UK 250V w/Test Automation3334912 BP Pump 2M NIBP, US 120V w/Test Automation
3461001 ESA612, THAI 230V w/Test Automation3462285 ESA612, JPN 230V w/Test Automation3326935 ESA620, US 115V 20A w/Test Automation3326947 ESA620, EUR 230V w/Test Automation3326958 ESA620, FR 230V w/Test Automation3326964 ESA620, ISR 230V w/Test Automation3326986 ESA620, AUS 230V w/Test Automation3326999 ESA620, SWI 230V w/Test Automation3327002 ESA620, UK 230V w/Test Automation3326874 Impulse 7KDP, US 120V w/Test Automation3326888 Impulse 7KDP, SHK w/Test Automation3326895 Impulse 7KDP, UK w/Test Automation3326901 Impulse 7KDP, JPN w/Test Automation3326912 Impulse 7KDP, AUS w/Test Automation3326920 Impulse 7KDP, IN w/Test Automation3319736 QA-ES, US 115V w/Test Automation3319749 QA-ES, SHK 230V w/Test Automation3319751 QA-ES, UK 230V w/Test Automation3319760 QA-ES, AUS 230V w/Test Automation3319772 QA-ES, JPN 100V w/Test Automation3327016 Kit, TA-LAGU, 1 Channel Lagu, with Test Automation3327025 Kit, TA-LAGU, 2 Channel Lagu with Test Automation3335538 TNT 12000, w/Test Automation3340639 TNT 12000WD, w/Test Automation
62
3334920 BP Pump 2M NIBP, AUS 250V w/Test Automation3334935 BP Pump 2M NIBP, DEN 250V w/Test Automation3334947 BP Pump 2M NIBP, SHK 250V w/Test Automation3334958 BP Pump 2M NIBP, ISR 250V w/Test Automation3334964 BP Pump 2M NIBP, ITAL 250V w/Test Automation3334973 BP Pump 2M NIBP, IND 250V w/Test Automation3334986 BP Pump 2M NIBP, SWZ 250V w/Test Automation3334999 BP Pump 2M NIBP, UK 250V w/Test Automation3334732 ESA601, 230V AUS w/Test Automation3334744 ESA601, DEU 230V SHK w/Test Automation3334759 ESA601, FRA 230V SHK w/Test Automation3334767 ESA601, ITA 230V SHK w/Test Automation3334771 ESA601, ISR 230V w/Test Automation3334780 ESA601, 230V SHK w/Test Automation3334798 ESA601, 230V UK w/Test Automation3334800 ESA601, 115V IEC w/Test Automation3334817 ESA601, 115 AAMI w/Test Automation3460932 ESA612, US 115 V w/Test Automation3460959 ESA612, EUR 230 V w/Test Automation3460944 ESA612, FR 230 V w/Test Automation3460967 ESA612, ISR 230 V w/Test Automation3460971 ESA612, AUS 230V w/Test Automation3460980 ESA612, UK 230 V w/Test Automation3460998 ESA612, SWI 230V w/Test Automation
“We used Ansur to create test sequences that match service manual procedures so every inspection is done the same way every time. We improved quality and uniformity by creating standard work.”
-Robert Dorrian, TBS U.K. Telematic & Biomedical Services Ltd. Hope Hospital
Key features•Easy verification of biomedical
equipment to manufacturer’s specifications
•Ten preprogrammed and five user-programmable electrical-safety-testing sequences
•Convenient transfer of equipment inventory and testing procedures from CMMS
•Module options to automate testing of most Fluke Biomedical testing devices
•Automatic storage of detailed test results for printing or transfer to CMMS
•Compliant with ANSI/AAMI (1993) and NFPA-99 (2005) standards
•20 A device testing with GFCI protection
•Wedge hardware option for extended serial port use, optional PC-style keyboard and barcode scan gun
The medTester 5000C is an automated system designed for electrical safety testing and performance verifica-tion. It is compatible with most Fluke Biomedical testing devices and a majority of the popular Computerized Maintenance Management Systems (CMMS) in the US. The medTester 5000C pro-vides a completely integrated solution for standardized and streamlined testing and record keeping.
Modes of operation Fully equipped, with four operational modes: manual, autosequence, medCheck, and remote control
Input power supply Line voltage/frequency input: 115 V ac ± 10 %/60 Hz
Test-receptacle type USA, 20 A
System/line voltage
Range (full scale) 200 V
Accuracy ± 5 % of range ± 1 LSD
Resolution 0.1 V
Equipment current
Range (full scale) 0 A to 20 A
Accuracy ± 5 % of range
Resolution 0.01 A
Ground resistance
Range (full scale) 0 Ω to 2 ΩAccuracy ± 1 % of range
Resolution 0.001 Ω (1 mΩ)
Current source 100 mA dc
Measurement type True four-terminal technique
Test leads Kelvin (2) insulated clip
Leakage-current/voltage gradient
Ranges (full scale) 200 µA and 2000 µA or mV
Accuracy DC and 48 Hz to 1 kHz, ± 1 % of reading; 1 kHz to 100 kHz, ± 2.5 % of reading; 100 kHz to 1 MHz, ± 5 % of reading
Resolution 0.1 µA or 0.1 mV
Measurement type True-rms (autoranging) (ac + dc or dc only response)
Test-load selection ANSI/AAMI ES1 1993
Test-load impedence 1000 Ω ± 1 % at dc
Isolation test
Test selection (full scale) Patient leads to ground
Lead combinations All leads; or individual leads RL, RA, LA, LL, and V1/V6 (V1 through V6 tested as a single lead)
Available current Limited by internal 120 kΩ resistor
Resolution 0.1 µA
Ranges (full scale) 200 µA and 2000 µA
Specifications
Optional accessories2245136 Performance Enhancement Module 2: RS-232/printer2245149 Performance Enhancement Module 3: 100 records2245151 Performance Enhancement Module 4: Expanded memory2245160 Performance Enhancement Module 5: Waves/extended test2245172 Performance Enhancement Module 6: Data transfer2245185 Performance Enhancement Module 7: medCheck2245197 Performance Enhancement Module 8: Defibrillator2245201 Performance Enhancement Module 9: IV pump2245212 Performance Enhancement Module 10: CMMS interface2245220 Performance Enhancement Module 11: ESU2245235 Performance Enhancement Module 12: Sp02
2245247 Performance Enhancement Module 13: Pacer2245258 Performance Enhancement Module 14: NIBP
Automated Biomedical Equipment Test System
medTester 5000C
Automation Solutions 63
Isolation test (continued)
ECG binding posts 10 posts, American Hospital Association color-coded RL, RA, LA, LL, V1-V6
Compatibility Compatible with both 3.2 mm and 4 mm pins and disposable snap electrodes
Performance waveforms
ECG performance test waves (lead I, Vp-p)
Square wave: 2 Hz, 1 mV
DC pulse 4 s, 1 mV
Sine wave 0.5 Hz, 10 Hz, 40 Hz, 60 Hz, and 100 Hz, 1 mV
Square wave 1 kHz, 1 mV
Triangle 2 Hz, 1 mV
CMRR 60 Hz sine wave with 1 kΩ imbalance in LA
Normal sinus 30 BPM, 60 BPM, 120 BPM, and 240 BPM
Arrhythmias Atrial fibrillationSecond-degree A-V block, type 1Premature atrial contractionsMissed beat at 80 BPM and 120 BPMPVC 1 leftPVC 2 rightMultifocal PVCsPVC 1, R on TA pair of PVCsRun of 5 PVCsRun of 11 PVCs, MFRight bundle branch block;Ventricular tachycardiaVentricular fibrillationAsystole
Environmental requirements
Operating temperature 15 °C to 55 °C (59 °F to 95 °F)
Storage temperature 0 °C to 50 °C (32 °F to 122 °F)
General information
Clock/date functions Time and date formats: 24 hour (hh:mm:ss) and mm/dd/yy
Safety certification Canadian Standards Association CSA C22.2 No 231-M89 (1989)
Display characteristics Type: 80 character, alphanumeric liquid crystal display (LCD)
Size: 2 lines x 40 characters
Backlight LED with adjustable brightness control
Dimensions (LxWxH) 25.4 cm x 35.0 cm x 10.2 cm (10 in x 13.8 in x 4 in)
Weight 5 kg (11 lb)
Specifications Optional accessories (continued)2245264 Wedge Adapter (eight 25 in serial ports, as well as AT or PS/2 keyboard port)2245061 Mini PC-style External Keyboard (83 keys, AT or PS/2, wedge adapter required 2245092 Laser Barcode Gun (wedge adapter required)2245515 5000C-PRINTER, Brady TLS Test Label Printer Kit medTester 5000C V 5.10 or greater and 115 V ac only2248606 Multipurpose Hard-sided Carrying Case for medTest-er 5000C with wedge adapter2248587 Multi-purpose Hard-sided Carrying Case for medTester 5000C without wedge adapterInterface cables: Call for specific test-device connection cables
Included accessoriesmedTester 5000C2243153 Users Manual2392871 Soft Vinyl Accessory Pouch2392617 Two Kelvin Cables2392639 Two Ground-pin Adapters
medTester 5000C/B, CMMS Connectivity Bundle2245136 RS-232/Printer Module2245149 100 Record Storage Module2245151 Expanded Record Storage Module2245172 Data-Transfer Module Med2245185 Check Module2245212 CMMS Interface Module
Ordering information2247382 medTester 5000C (20 A, 115 V ac)2585098 medTester 5000C/B, CMMS Connectivity Bundle
medTester 5000C is compat-ible with the following test tools:• Impulse4000DefibAnalyzer• IDA4plusIVPumpAnalyzer• RF303RSESUAnalyzer• Index2XLSpO2Analyzer• SigmaPace1000Pacemaker
Analyzer• CufflinkNIBPSimulator
medTester also interfaces with the following legacy test tools:• Impulse3000DefibAnalyzer• Infustest2000SeriesD
IVPUMP Analyzer• IPT-1IVPUMPAnalyzer• IPT-MCIVIVPUMPAnalyzer• 402AESUAnalyzer• 454AESUAnalyzer• OxitestPlus/Plus7SpO2
Analyzer• CardioSat100SpO2 Analyzer
Automation Solutions
Automated Biomedical Equipment Test System
medTester 5000C
64
Incubator Analyzer
INCU™
Incubator Analyzers
Key features• Simultaneous measurement of
humidity, airflow, sound, and 4 independent temperatures
• 24-hour continuous testing (battery); 35-hour continuous testing (main power)
• Battery operated • Adjustable measurement
intervals• Compatible with closed,
forced-convection incubators and open infant warmers
• Stand-alone measurement or automated testing with PC
• Windows® compatible INCU software for easy data collection, analysis, and documentation
• Numerical and full-color graphical reports
Faulty incubator con-trols lengthen hospital stays, and increase healthcare costs, making thorough incu-bator testing essential.
Designed around AAMI and IEC standards that specify incubator and radiant warmer sound levels and thermal characteristics, the INCU simultaneously measures airflow, rela-tive humidity, sound, and four independent
temperatures. Adjustable measurement intervals allow technicians to configure the unit to meet their facilities’ testing requirements. Technicians are free to do other work while the device collects and records data.
INCU software lets technicians upload setup parameters and download the test results to a PC file, or print the data in reports with full-color charts and graphs.
Power supply
Input voltage range 100 V ac to 240 V ac
Input frequency range 47 Hz to 63 Hz
Battery Rechargeable sealed lead-acid type NP7-6 YUASA, 6 V, 7 Ah; operates for 24 hours continuously; low-battery alert
Sound level
Measuring range 30 dbA to 80 dbA
Resolution 0.1 dbA
Accuracy +5 dbA @ 30 dbA to 80 dbA
Relative humidity (without condensation)
Measuring range 0 % to 100 % RH, non-condensing
Resolution 0.1 % RH
Accuracy ± 5 % RH for 0 % to 90 % RH at 77 °F to 104 °F (25 °C to 40 °C) or ± 5.3 % RH for 0 % to 100 % RH at 77 °F to 104 °F (25 °C to 40 °C)
Temperature measurement
Measuring range 5 °C to 70 °C (41 °F to 158 °F)
Resolution 0.1 °C (32 °F)
Accuracy +0.5 °C (+0.9 °F) +1 LSB of range from 25 °C to 40 °C (77 °F to 104 °F)
Airflow
Measurement range 0.1 m/s to 0.7 m/s
Resolution 0.01 m/s
Accuracy from 0.1 m/s to 0.5 m/s reading ± 0.1 m/s at temperature 25 °C to 40 °C (77 °F to 104 °F) and humidity 50 % RH ± 15 % RH
General information
Measurement interval Via PC: Adjustable from 1 minute to 10 minutes
Storage temperature -20 °C to 50 °C (-4 °F to 122 °F)
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Dimensions (LxWxH) 27 cm x 20 cm x 14 cm (10.6 in x 7.8 in x 5.5 in)
Weight 3 kg (6.6 lb)
Specifications
Included accessories2206965 Users Manual2248900 Soft-Sided Carrying Case2239025 Airflow Sensor2391761 Universal AC Battery Charger with Worldwide Mains Adapter Set2391866 Serial Cable DB9F to DB9F2391789 Serial Adapter DB-9M to DB-25F2213928 Outside Temperature probe holder2239002 Adapter for Radiant Infant Warmer Assembly2213919 INCU PC Software (one CD)
Ordering information2250148 Incubator Analyzer
65
Included accessoriesUsers ManualMAX-250 External Oxygen SensorOxygen Sensor Cable2448051 Breathing Circuit tee
Ordering information2511920 MAXO2+AE Oxygen Analyzer
Oxygen Analyzer
Oxygen Analyzer
Key features• One-touch calibration, with
reminder• Long battery life
(approximately 5,000 hours)• Impact resistant and drip
proof• External MAX-250E Oxygen
Sensor
MAXO2+AE
The MAXO2+AE is an oxygen analyzer that measures the oxygen concentration in a flow of gas from a medical gas source or through a medical gas-flow device such as a ventilator or anesthesia system, or within an infant incubator. It is handheld and rugged to suit the needs of portable use. The MAXO2+ AE comes equipped with a two-year warranty on both analyzer and sensor.
Specifications
Measurement range 0 % to 100 %
Resolution 0.1 %
Accuracy and linearity 1 % of full scale at constant temperature, RH and pressure when calibrated at full scale
Total accuracy ± 3 % actual oxygen level over full operating range
Response time 90 % of final value in approx. 15 sec at 23 °C
Warm-up time None required
Power supply
Battery life Approx. 5000 hours with continuous use
Low battery indication “BATicon displayed on LCD
Sensor type Maxtec® MAX-250E for AE model
Expected sensor life > 900,000 O2 % hours minimum, 2 years in typical medical applications
Power requirements 2, AA alkaline batteries
Environmental requirements
Operating temperature 15 °C to 40 °C (59 °F to 104 °F)
Storage temperature -15 °C to 50 °C (5 °F to 122 °F)
Atmospheric pressure -800 mBar to 1013 mBar
Relative humidity Operating range: 0 % to 95 % (non-condensing)
General information
Dimensions (LxWxH) 38 mm x 76 mm x 914 mm (1.5 in x 3.0 in x 36.0 in)
Weight 285 g (0.6 lb)
66
Fluke Biomedical’s Global Calibration Laboratory is equipped to calibrate and repair the following types of instruments:
• Area Monitors• Barometers• Blood Pressure Simulators• Defibrillators/External Pace
Maker Analyzers• Densitometers• Diode Detectors• Dosimeters• Electrical Safety Analyzers• Incubator Analyzers• Ion Chambers • IV Pump Analyzers • kVp Meters• mAs Meters• Electrical Multimeters• Oscilloscopes• Patient Simulators • Pressure Meters/Parameter
Testers• Radiation Multimeters• Sensitometers• SpO2 Simulators/Analyzers• Thermometers• Test Lungs• Ultra Sound Analyzers• Velometers• Ventilators/Gas flow
Analyzers
World-class facility. World-class service.
Service and Calibration
Service and Calibration
Calibration Beam Specifications
Fluke Biomedical’s Global Calibration Lab is NVLAP Lab Code 200566-0 accredited, adheres to ISO 17025:2005, ANSI Z540, Mammography MQSA, and CNSC, and is traceable to national and interna-tional standards.
Fluke Biomedical offers one-stop, bulk contracts for managing larger instrument pools, including various asset-management alternatives for pools larger than 150 units. Fluke Biomedical’s asset-management program takes over your grueling task of instrument tracking and allows you to use your time more productively.
If you have a large number of instruments that require service, you
can greatly benefit from this quality service. Proper protocols are strictly fol-lowed, eliminating the problems with inspectors and audits that can result when other less-qualified labs perform the calibrations. Instrumentation includes Fluke Biomedical as well as other industry models.
Tungsten AnodeNIST-Traceable Techniques
Equivalent Beam Code
Potential(kV)
Filtration HVLmm Al
mm Al
mm Cu
mm Sn
mm Pb
L20 20 0.07
L100 100 1.98 2.75
M30 30 0.50 0.33
M50 50 1.00 0.98
M60 60 1.50 1.68
M80 80 2.6 2.98
M100 100 5.0 5.1
M150 150 5.0 0.25 10.2
M200 200 4.1 1.12 14.9
M250 250 5.0 3.2 18.5
H50 50 4.0 0.12 4.4
H60 60 4.0 0.6 6
H100 100 4.0 5 13.5
H150 150 4.0 4 1.5 16.8
H200 200 4 0.6 4 0.7 19.5
H250 250 4 0.6 1 2.7 21.5
PTB-Traceable Techinques
Equivalent Beam Code
Potential(kV)
Filtration HVLmm Al
mm Al
mm Cu
mm Sn
mm Pb
DV30 30 2.5 0.98
DV40 40 2.5 1.44
DV50 50 2.5 1.81
DV60 60 2.5 2.13
DV70 70 2.5 2.45
DV80 80 2.5 2.78
DV90 90 2.5 3.1
DV100 100 2.5 3.48
DV120 120 2.5 4.15
DV150 150 2.5 5.36
DH40 40 4 2.2
DH50 50 10 3.75
DH60 60 16 5.35
DH70 70 21 6.77
DH80 80 26.0 8.12
DH90 90 30.0 9.26
DH100 100 34.0 10.15
DH120 120 40.0 11.67
DH150 150 45.0 13.36
67
World-class facility. World-class service.
Service and Calibration
Service and Calibration
Calibration Beam Specifications
Molybdenum/Rhodium AnodeNIST-Traceable Techniques
Equivalent Beam Code
Potential(kV)
Filtration HVLmm Almm Mo mm Rh mm Al
Mo/Mo 28 28 0.032 0.33
Mo/Mo 35 35 0.032 0.39
Mo/Rh 28 28 0.029 0.41
Rh/Rh 25 25 0.029 0.35
Rh/Rh 40 40 0.029 0.56
Mo/Mo28x 28 0.030 2 0.63
Rh/Rh/35x 35 0.029 2 0.898
PTB-Traceable Techinques
Equivalent Beam Code
Potential(kV)
Filtration HVLmm Almm Mo mm Rh mm Al
MV20 20 0.030 0.223
MV25 25 0.030 0.282
MV30 30 0.030 0.337
MV35 35 0.030 0.374
MV40 40 0.030 0.402
MV50 50 0.030 0.440
MV20 20 0.030 2 0.450
MV25 25 0.030 2 0.580
MV30 30 0.030 2 0.670
MV35 35 0.030 2 0.749
MV40 40 0.030 2 0.825
MV50 50 0.030 2 0.968
Service Center/Repair/Calibration USFluke Biomedical6045 Cochran RoadCleveland OH 44139-3303Tel: 440-498-2560Toll free: 800-850-4608 ext 2564Email: [email protected]
Service Center/Repair/Calibration EuropeFluke Biomedical EuropeScience Park Eindhoven 51105692EC Son, The NetherlandsTel: +31 (40) 267 5435Fax: +31 (40) 267 5436Email: [email protected]
www.flukebiomedical.com/service
68
PublicationsThe following Fluke Biomedical
catalogs are also available
Your authorized Fluke Biomedical distributor
Fluke Biomedical Radiation Oncology QAThe Fluke The Radiation Oncology QA catalog provides a full range of QA solutions for the Radiation Oncology Physicist, Therapist, and Dosimetrist. The catalog contains information about the linear accelerator QA instruments, radiation oncol-ogy chambers, phantoms, and accessories needed to manage radiation oncology QA and maintain a safe, regulatory-compliant facility.
For more information, contact [email protected]
Fluke Biomedical Radiation SafetyThe Radiation Safety catalog highlights devices used to measure radiation levels, manage regulatory QA requirements, and aide in radiation emergencies. These devices are intended for Radiation Safety Officers (RSOs), Health Physicists, Emergency Responders and other radiation-minded professionals. The catalog contains information about a variety of survey meters and probes, area monitors, and other radiation-monitoring accessories.
For more information, contact [email protected]
Fluke Biomedical Diagnostic Imaging QAThe Diagnostic Imaging QA catalog is a comprehensive source book of solutions for the Imaging QA Technologist, Physicist, Biomedical/Clinical Engineer, or Service Engineer. The catalog contains information about the test devices, phantoms, and accessories needed to manage diagnostic imaging QA and maintain regulatory-compliance.
For more information, contact [email protected]
Fluke Biomedical. Better products. More choices. One company.
Fluke Biomedical6045 Cochran Road Cleveland, OH 44139-3303 U.S.A.
Fluke Biomedical EuropeScience Park Eindhoven 5110 5692EC Son, The Netherlands
For more information, contact us:In the U.S.A. (800) 850-4608 orFax (440) 349-2307In Europe/M-East/Africa +31 40 267 5435 orFax +31 40 267 5436From other countries +1 (440) 248-9300 orFax +1 (440) 349-2307Email: [email protected] access: www.flukebiomedical.com
©2007-2009 Fluke Biomedical. Specifications subject to change without notice. Printed in U.S.A. 6/2009 3082069B C-EN-N