BIOMETRY & PEDIATRICS IOL CALCULATION
BY DR RAKESH JAISWAL
![Page 2: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/2.jpg)
BIOMETRY
Biometry is the process of measuring thepower of the cornea (keratometry) and theAxial length of the eye, by using this data todetermine the ideal intraocular lens power
Measurement of axial length(AL)• A-scan • IOL master
![Page 3: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/3.jpg)
A-scan Biometry• A scan: Amplitude Scan;
utilizes ultrasound waves of 10 - 12 MHz frequency.
• 2 Principles: Piezoelectric Phenomenon
Acoustic Impedence.
• Components: Transducer Amplifier
Display Monitor • Ultrasound biometry
machines use the formula Distance = Velocity x Time
![Page 4: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/4.jpg)
• In A-scan, thin, parallel sound beam is emitted from the probe tip, with an echo bouncing back into the probe tip as the sound beam strikes each interface.
• An interface is the junction between any two media of different densities and velocities.
anterior corneal surface aqueous/anterior lens surface posterior lens capsule/anterior vitreous posterior vitreous/retinal surface choroid/anterior scleral surface.
![Page 5: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/5.jpg)
• The echoes received back into the probe from these interfaces are converted by the biometer to spikes arising from baseline.
• The greater the difference in the two media at each interface, the stronger the echo and the higher the spike.
![Page 6: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/6.jpg)
• Average Axial Length of Normal Eye 23.06 mm
•Majority 22.0 to 24.5 mm
•Error of 0.4mm in the measurement of axial length may result in a one diopter change in calculated IOL power.
•Difference in AL measurement Between both eyes + 0.3 mm
![Page 7: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/7.jpg)
• Contact - Applanation Method - Hand-Held Method
• Immersion
![Page 8: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/8.jpg)
Applanation A-scan Biometry
• A-scan biometry by applanation requires that the ultrasound probe be placed directly on the corneal surface. This can either be done at the slit lamp
![Page 9: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/9.jpg)
Applanation A-scan Biometry.
• a: Initial spike (probe tip and cornea)b: Anterior lens capsulec: Posterior lens capsuled: Retinae: Scleraf: Orbital fat
![Page 10: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/10.jpg)
Applanation A-scan Biometry
• When echoes b through d are high and steeply rising, the ultrasound beam is most likely on visual axis.
• If no scleral or orbital fat echoes visible, then ultrasound beam is most likely aligned with optic nerve.
![Page 11: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/11.jpg)
The five basic limitations of applanation A-scan biometry are:
1. Variable corneal compression. 2. Broad sound beam without precise localization3. Limited resolution. 4. Incorrect assumptions regarding sound velocity
5. Potential for incorrect measurement distance.
![Page 12: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/12.jpg)
PROCEDURE: Hand-Held Method A probe is placed on the patient’s cornea.The probe is attached to a device that delivers adjustable sound waves.The measurements are displayed as spikes on the screen of an oscilloscope (Visual monitor).The appearance of the spikes and the distance between them can be correlated to structures within the eye and the distance between them.
![Page 13: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/13.jpg)
Probe positioning: The probe lightly touches the cornea and is positioned, such that the barrel of the probe is aligned with the optical axis or visual axis of the eye.The operator aims the probe towards the macula of the eye.Alignment with the optical axis will be indicated by high lens spikes and a high retina spike on the scan graph.
![Page 14: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/14.jpg)
• Spike height is affected by the difference in density & by the angle of incidence, which is determined by the probe orientation to the visual axis.
• If the probe is held nonparallel, part of the echo is diverted at an angle away from the probe tip, and is not received by the machine.
• A perfect high, steeply rising retinal spike may be impossible when macular pathology is present (eg, macular edema, macular degeneration, epiretinal membranes, posterior staphylomas).
![Page 15: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/15.jpg)
![Page 16: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/16.jpg)
Corneal Compression • If pressure is applied on the cornea, the axial
length measurment may be falsely too short.• It can be monitored by observing the anterior
chamber depth, read out by an instrument.• Most eyes will have an ACD readings between
2.5 to 4.0mm.• The corneal compression error factor can be
avoided by using the immersion technique
![Page 17: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/17.jpg)
Error caused by 1 mm Corneal Compression
Average eye 2.5 DLong eye 1.75 DShort eye 3.75 D
![Page 18: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/18.jpg)
Immersion A-scan Biometry
• The immersion technique is accomplished by placing a small scleral shell between the patient's lids, filling it with saline, and immersing the probe into the fluid, being careful to avoid contact with the cornea.
• More accurate than contact method because corneal
compression is avoided.
• Eyes measured with the immersion method are, on average, 0.1-0.3 mm longer.
![Page 19: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/19.jpg)
Immersion A-scan Biometry
• The immersion technique requires the use of a Prager Scleral Shell .
![Page 20: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/20.jpg)
Immersion A-scan Biometry • . • a: Probe tip. Echo from tip of
probe, now moved away from the cornea and has become visible.
• b: Cornea. Double-peaked echo will show both the anterior and posterior surfaces.
• c: Anterior lens capsule.• d: Posterior lens capsule.• e: Retina. This echo needs to have
sharp 90 degree take-off from the baseline.
• f: Sclera.• g: Orbital fat.
![Page 21: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/21.jpg)
Immersion A-scan Biometry
• When the ultrasound beam is properly aligned with the center of the macula, all five spikes will be steeply rising and of maximum height.
• Both the peaks of corneal spike should be equal in
height ideally.
• Other advantage: Easier, better repeatability.
![Page 22: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/22.jpg)
• The gain setting on A-scan is measured in decibels and affects amplification and resolution of spikes.
• When on highest gain, spike height and sensitivity of display screen are maximized, enabling visualization of weaker signals, but resolution is affected adversely.
• When gain is lowered, the spike amplitude and sensitivity are decreased, which eliminates the weaker signals but improves resolution.
• Error can occur when the gain is set too high or too low .
• Very high gain short reading• Very low gain long reading
![Page 23: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/23.jpg)
• Resolution: ability to display two interfaces that lie in close proximity, one directly behind the other, as separate echoes or spikes.
• The more dense the cataract, the higher the necessary gain.
• Gain setting may vary not only from patient to patient but from one eye to the next in the same patient, depending on cataract density.
![Page 24: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/24.jpg)
• Gates are electronic calipers on the display screen that measure distance between two points.
• Proper gate placement is on the ascending edge of each appropriate spike.
• If the biometer does not allow for movement of gates, scans must be repeated until they automatically align properly.
![Page 25: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/25.jpg)
![Page 26: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/26.jpg)
The scan of an aphakic eye• It will either have no lens spikes,
(or) it will have one lens spike (A) that represents an intact posterior lens capsule. ( C )
• The velocity of sound will be different because the beam is not passing through the lens.
• A velocity of sound of 1532m/s is typically used for aphakic measurements
• .• Its required for scondary IOL
• Immersion techique is choice
![Page 27: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/27.jpg)
• Biometry of pseudophakic eye performed:- To compare to the fellow phakic eye for accuracy- IOL exchange- Checking an unwanted postoperative refractive error.
• A scan of pseudophakic eye → multiple reverberation echoes in the vitreous cavity that tend to decrease in amplitude from left to right.
• Decreasing the gain in pseudophakic eye is helpful.
![Page 28: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/28.jpg)
Factors to Consider for IOL material The type of material is important because the
velocity of sound is a function of the material that the sound is passing through.
A-Scan does not actually measure length, they measure how long it takes a sound beam to bounce off an object ( Anterior lens, Posterior lens,and Retina) & return to the probe.
The instrument is pre programmed with the velocity of sound factors for the aqueous, the lens material, and the vitreous.
![Page 29: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/29.jpg)
PMMA protocol : (Polymethylmethacrylate)
•Sound travels faster through PMMA than it does through the natural lens.
•If a pseudophakic mode is not available,you can measure the eye in the aphakic mode and add a standard compensating factor of 0.4mm to the resultant axial length.
Silicon protocol:
•Silicon IOL’s are foldable. Sound travels much slower through a silicon lens than it does through the natural lens.
•If not taking this in to account could result in a –3.0D post - op refractive error.
•If your biometer does not have a pseudophakic or silicon mode, use the aphakic mode and subtract a compensation factor of 0.8mm.
![Page 30: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/30.jpg)
Acrylic protocol:
•Sound travels faster through acrylic than it does through the natural lens.
•If biometer does not have an acrylic mode,use the aphakic mode and add a compensation factor of 0.2mm.
![Page 31: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/31.jpg)
1 Inadequate Patient Fixation Low Vision Nystagmus Blepharospasm Strabismus2. Posterior StaphylomaPosterior staphylomas often causeresulting in an inability to display adistinct , high retinal spike , leadingto a significant error in A-scan measurement . 3. High Hyperopia Immersion technique is preferable4. Macular Lesions 5 Vitreous Lesions6 Dense Cataract
![Page 32: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/32.jpg)
IOL MASTER
WORKING PRINCIPAL• IOL Master (Zeiss Humphrey system) is a
combined biometric instrument that measures quickly and preciesely parameters of human eye needed for IOL power calculation by non-contact technique.
• It also incorporates the softwere to calculate IOL power from various formulae
![Page 33: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/33.jpg)
WORKING RINCIPALIt is a non contact optical device that measures the
various parameters based on the following principal1. AL measurement is based on patented interference
optical method known as Partial Coherence Interferometry (PCI).This technique relies on a laser Doppler technique to measure the echo delay and intensity of infrared light reflected back from the tissue interface-cornea and retinal pigment epithelium .
2. Corneal curvature(K) is determined by measuring the distance between reflected light images as in conventional keratometry
![Page 34: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/34.jpg)
Anterior chamber depth (ACD) is determined as the distance between the optical section of the crystalline lens and cornea produced by lateral slit illumination
White-to-White is determine from the image of the iris
Calculation of IOL power by softwere incorporating internationally accepted calculation formulae
![Page 35: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/35.jpg)
Advantages of the IOL master• Patient comfort as the technique involves non
contact measurements• User friendly • Single instrument for measuring AL,corneal
curvature and ACD• Cross-infection risk is not there,as technique is
non-contact• More accurate AL measurement as compared
to A-scan (five times) with AL ranging b/w 14 to 40mm
![Page 36: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/36.jpg)
• Specially useful in certain ocular conditions like small corneal scar,ant cortical spokes,extream myopia, post staphyloma
• Incorporates IOL power
• This method cannot be used in significant media opacity (eg. dense cataracts or corneal or vitreal opacity) due to absorption of light or inability of the patient to fixate on target.
![Page 37: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/37.jpg)
Accuracy of axial length by different machine
Applanation A -scan
Immersion A-scan
IOL Master
+/- 0.24mm +/- 0.12mm +/- .01mm
![Page 38: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/38.jpg)
Immersion ultrasound
IOL master
Posterior staphylomaSilicone oilPseudophakia4++brunescent lensCentral PSC plaqueVitreous hemorrhageCentral corneal scar
DifficultDifficultVariable•Yes•Yes•Yes•Yes
•Yes•Yes•YesNoNoNoNo
![Page 39: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/39.jpg)
![Page 40: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/40.jpg)
Introduction• Keratometry .1. Keratometry is the measurement of a patients corneal
curvature .2. It provides on objective,quantitative measurement of
corneal astigmatism,measuring the curvature in each meridian as well as the axis.
3. Keratometry is also helpful in determining the appropriate fit of contact lens .
4. It is use for measurement the corneal dioptric power.5. The measurement of the curvature of the anterior
corneal surface by using the first Purkinje image.
![Page 41: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/41.jpg)
Keratometer:• Determines corneal curvature by measuring the size
of a reflected “mire”. n - 1
• Surface power formula: D = ......... R
• D = the dioptric power of the cornea n = the refractive index of the cornea used (1.3375) R = the radius of curvature of the cornea in meters
• Keratometer measures only the central 3mm of the corneal diameter.
![Page 42: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/42.jpg)
Principle of Keratometry • Cornea is a convex refracting surface
• In order to find the refracting power of the cornea, we need to reflect an object of a known size at a known distance off the corneal surface.
• Then determine the size of the reflecting image with measuring telescope and calculate the refractive power of the cornea based on the refractive index of n= 1.3375
Doubling Principle Helps us to measure the size of something that moves
around since the small movements of the eye makes the measurement impossible
![Page 43: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/43.jpg)
Keratometry 43
Bausch & Lomb Keratometer
• One position keratometer: it measures two meridians at the same time
• It contains two prisms
![Page 44: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/44.jpg)
Principle of Keratometry
![Page 45: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/45.jpg)
• The object used is an illuminated circle with plus and minus rings .
• The two prisms inside the instrument give two additional one displaced horizontally and another displaced vertically. Three images are seen.
• While taking the reading the pluses and minuses coincide. This is achieved by moving the keratometer with the object forward or backward in front of the eye.
• When coincidence takes place the size of the images of fixed value. The distance of the object is different for different curvatures. The instrument is calibrated. As the drum rotate the distance varies.
![Page 46: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/46.jpg)
Parts of keratometer
![Page 47: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/47.jpg)
Parts of keratometer1. Telescope2. Eye piece at one end (near to examiner)3. Objective (near to patient)4. Knobs for adjustment of vertical and horizontal
curvature on the telescope.5. Knobs adjusting height of telescope.6. Chinrest-adjusting knobs.7. Bulb for illumination.8. Knob for focusing of mires.9. Model cornea with occluder.
![Page 48: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/48.jpg)
Performing Keratometry1. Looking the through the eye piece of the
keratometer,use the eye piece to focus the reticule (cross hair) in the same way as for the lens meter.
2. The patient can comfortably put the chin and forehead on the appropriate rests.
3. Use the occluder attached to the keratometer to cover the eye not being measured.
4. Then use the height adjustment knob of the keratometer to position the light reflections at the level of the cornea.
![Page 49: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/49.jpg)
1. To obtain proper focus, rotate the focus knob until the bottom-right circles converge to form a fused image.(fig)
1. To locate the proper axis, rotate the keratometer until the pluses between the two bottom circles are in the same plane.(fig)
![Page 50: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/50.jpg)
Range of keratometer
• 36-52D• Its lower limit can be extented up to 30D by
interposing -1.0D in front of objective of telescope
• Its upper limit up to 61D by interposing a lens of +1.25D in front of objective of telescope
![Page 51: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/51.jpg)
Source of keratometry errors
• Unfocused eye piece• Failure to calibrate unit• Poor patient fixation• Dry eye• Drooping eye lids• Irregular cornea
![Page 52: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/52.jpg)
Repeat Keratometery If
• Corneal curvature more than 47D or less than 40D.
• The difference in corneal cylinder is more than one diopter between eyes.
• The average keratometry (K) → 43.0-44.0D,
![Page 53: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/53.jpg)
Difficult Situations
• Post Refractive Surgery• Corneal Transplantation• Corneal Scar• Keratoconus etc.
![Page 54: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/54.jpg)
Formula forIOL Power Calculation
![Page 55: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/55.jpg)
Depending upon the basis of their deviation ,the various formulae for calculating IOL power have been grouped into
•Theoretical formulae –these is based on mathematical principal revolving around the schematic eye
•Regression formulae-these were arrived at by looking at postoperative outcomes retrospectively
![Page 56: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/56.jpg)
IOL Formulae
• First generation
• Second generation
• Third generation
• Fourth generation
![Page 57: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/57.jpg)
First generation
Theoretical
1. Binkhorst 2. Colenbrander-
Hoffer3. Gilll’s
4. Clayman 5. Fyodorov
Regression
SRK - I
![Page 58: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/58.jpg)
IOL FORMULA Ist generation• Most are based on regression formula developed by
Sander ,Retzlaff & Kraff• Known as SRK formula.
• P = A - 2.5(L) - 0.9(K)
P=lens implant power for emetropiaL= Axial length (mm)K=average keratometric reading (diaopters)A= lens constant
![Page 59: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/59.jpg)
Second generation
Theoretical
Modified Binkhorst
Regression
SRK - II
![Page 60: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/60.jpg)
IOL FORMULA 2nd generation• SRK formula – works well for average eyes. less accurate for long, short eyes.• SRK II formula Modification of SRK A-constant is modified on the basis of AL P = A1 – 2.5L – 0.9K
A1 = A + 3 AL < 20mmA1 = A + 2 AL 20-21A1 = A + 1 AL 21-22A1 = A AL 22-24.5A1 = A – 0.5 AL >24.5
![Page 61: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/61.jpg)
Third generation
• Holladay I• SRK/T• Hoffer Q
![Page 62: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/62.jpg)
IOL FORMULA 3rd generation
• Third generation formulas-
• SRK/T -very long eyes >26mm• Holladay I -long eyes 24-26 mm• HofferQ -Short eyes<22mm
![Page 63: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/63.jpg)
Fourth generation
• Holladay - II
![Page 64: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/64.jpg)
IOL FORMULA 4th generation
• Holladay II• Haigis formula- d = a0 + (a1 * ACD) + (a2 * AL)ACD is the measured anterior chamber depthAL is the axial length of the eye
The a0, a1 and a2 constants are set by optimizinga set of surgeon- and IOL-specific outcomes for a widerange of ALs and ACDs.
![Page 65: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/65.jpg)
• SRK/T formula — uses "A-constant“
• Holladay 1 formula — uses "Surgeon Factor“
• Holladay 2 formula — uses "Anterior Chamber Depth“
• Hoffer Q formula — uses "Anterior Chamber Depth"
![Page 66: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/66.jpg)
AL < 19 mm (<0.1%) Holladay 2AL 19-22 mm (8%) Hoffer-Q
AL 22-24.5 mm (72%) SRK II
AL 24.5-26 mm (15%) Holladay 1 , AL > 26 mm ( 15%) SRK/T
![Page 67: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/67.jpg)
When capsular tear does not allow bag placement of the lens → change IOL power for sulcus placement
• >=28.5 D Decrease by 1.5 D• +17 To 28 D Decrease by 1.0 D • +9 To 17 D Decrease by 0.5 D• <+ 9 D No change
![Page 68: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/68.jpg)
Pediatric biometry• Pediatric eye is not a miniaturized adult eye. • It has shorter axial length, steeper cornea with
higher keratometry value and smaller anterior chamber depth.
• Errors in axial length measurement affect IOL power calculation the most, it increases to 3.75 D per mm in children.
![Page 69: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/69.jpg)
• no formula has been proven to have an advantage over others.
• newer theoretic formulas such as Holladay II that take anterior segment measurements into account, may be recommended for pediatric IOL power calculation or Hoffer Q
• 20% undercorrection if the child is less than age 2 years 10% undercorrection for age 2–8 years .
![Page 70: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/70.jpg)
![Page 71: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/71.jpg)
![Page 72: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/72.jpg)
A-SCAN:A-SCAN We measure AL with an immersion technique, often under sedation. KERATOMETRY For the K measurement, we use a handheld keratometer. Alcon autokeratometer produces accurate measurements of curvature of the cornea in pediatric eyes.
• Measurements should be taken without the use of an eyelid speculum• To avoid the problems associated with corneal dryness, measurements should be taken as soon as possible (following IOP measurement) after induction of anesthesia. •Balanced salt solution should be instilled as necessary to maintain a smooth corneal surface
![Page 73: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/73.jpg)
Using hand –held instruments , reliable keratometry & axial length measurement. For each individual case the IOL power should be customized based on many characteristics including the age , one eye or both , amblyopia status , likely compliance with glasses & family history of myopia.
![Page 74: Biometry & Iol calculations](https://reader037.vdocuments.net/reader037/viewer/2022102518/58a233081a28ab527c8b69fb/html5/thumbnails/74.jpg)