bios 532 laboratory module on uv-visible light absorption spectroscopy
DESCRIPTION
Bios 532 Laboratory Module on UV-Visible Light Absorption Spectroscopy. Absorption Spectroscopy. The Beer-Lambert Law A= lc Where A is absorbance or optical density (no units, since A = log [I 0 /I]); is the molar absorptivity with units of (L mol -1 cm -1 ); - PowerPoint PPT PresentationTRANSCRIPT
Bios 532
Laboratory Module on UV-VisibleLight Absorption Spectroscopy
Absorption Spectroscopy
The Beer-Lambert Law
A=lc
Where A is absorbance or optical density (no units, since A = log [I0/I]);
is the molar absorptivity with units of (L mol-1 cm-1);l is the path length of the sample - that is,the interior dimension (cm) of the cuvette
in which the sample is contained, parallel to the light path; c is the concentration of the compound in solution,
expressed in (mol L-1).
is a measure of the amount of light absorbedper unit concentration.
Molar absorptivity is a constant for a particular substance, so if the concentration of the solution is halved, so is the
absorbance at dilute concentrations.
Molar AbsorptivityA = lc
A
concentration
this relationship fails at extremely high concentrations
A = bc
A
concentration
Limitations
• deviations in absorptivity coefficients at high concentrations (>0.01M) due to electrostatic interactions between molecules in close proximity• changes in refractive index at high analyte concentration• shifts in chemical equilibria as a function of concentration
Absorbance (A) versus the measured transmission (%T):
A sample which allows 100% of the selected light to be transmitted clearly has zero absorbance.
A sample which allows 0% transmittance exhibitsinfinite absorbing power: ∞ = A .
These limits are properly accounted for through the following logarithmic relationship between absorbance and percent transmittance:
A = - log (%T /100)
Limitations
Errors in %T readings which are too low (e.g., below ~12%) or too high (e.g., above ~70%) are significantly magnified in their corresponding A values.
Design experiments so that only samples with %T values above 12% and below 70% are prepared and measured.
This corresponds to solutions with absorbance values A between ~ 0.16 and 0.92.
A = - log (%T /100)Limitations
Review of these principles can be found on-line in the US Naval Academy Dept. of ChemistryGeneral Chemistry Laboratory Manual:
http://www.chemistry.usna.edu/manual/ApdxI.pdf.
• monochromatic light - light of a single, constant wavelength.
• coherent light - wavelengths are in phase in space and time.
• white light is a mixture of all the colors of the visible spectrum, which includes light with wavelengths of 400 - 700 nm.
• ordinary light (emitted from the everyday light bulb, the sun, a candle) will consist of wavetrains of unrelated phases, frequency and polarization in all directions.
• in contrast, laser light is monochromatic, directional and coherent.
Spectroscopic light sources - terms
Light Source in the Shimadzu Specs
1. The halogen visible light source - typical spectral emission range for a halogen lamp is ~ 300 -1100 nm.
2. The deuterium UV light source - typical spectral emission range is ~ 180 - 500 nm.
3. Switching between the two light sources is performed automatically by the spec, according to a pre-selected switching wavelength that has been chosen at installation. The switching wavelength is usually somewhere between 280 - 390 nm.
Difference Spectroscopy
• Two solutions are compared - usually start out identical.
• All common spectral features cancel out.
• Reference solution is unperturbed.
• The sample solution is varied by additives.
• Double-beam operation - the spec does the subtraction (the Shimadzu specs have a beam chopper).
Difference Spectroscopy
The chopper resembles a fan that allows light to pass for a characteristic period of time, then blocks the light for the same time period. The unblocked signal is reference + sample, the blocked signal is reference only. The difference between these two signals (blocked and unblocked) must be the desired analytical signal.
• beam chopper - an optical chopper is a mechanical or electromagnetic device for passing and then interrupting a beam of light for a known brief interval. Examples include tuning forks, rotating shutters and the more sophisticated Kerr cells.
• monochromator - an instrument for isolating narrow portions of the spectrum The spectrum of any light source is formed by a prism or grating, and an exit slit placed in the spectrum selects a narrow band of wavelengths for emission. By moving the spectrum of a source internally past the slit, the color of the emitted light can be varied at will. As most monochromators emit several percentages of unwanted light -- either white or of the wrong wavelength -- along with the desired wavelength, two monochromators often are used in tandem, both being set to transmit the same wavelength. In this way the percentage of unwanted light can be reduced drastically.
monochromators
1. Grating monochromator - uses a diffraction grating as the dispersive element that is a ray of fine, parallel, equally spaced reflecting or transmitting lines that mutually enhance the effects of diffraction to concentrate the diffracted light in specific directions determined by the spacing of the lines and by the wavelength of the light.
2. Crystal monochromator - the crystal lattice serves as a 3-dimensional diffraction grating that separates light by wavelength (think prism).
Factors other than sample path length and sample concentration that can affect absorbance
reflections at the cuvette interface scratches, fingerprints on the cuvette absorptive materials in buffer (non-sample) fluorescence (increases apparent intensity) high sample concentrations change in the chemical properties of the sample
Hemoglobin
1 2
1
2
A B
D C
Fe
MethemoglobinFe(II)-heme is oxidized to Fe(III)-heme to form metHb.
MetHb does not bind O2.At low pH, H2O occupies the space between the Fe and the
distal histidine - aquomethemoglobin.
At high pH, Fe binds OH- - hydroxymethemoglobin.
oxy-Hb met-Hb
When a ligandis bound to Hb,the heme ironis 6-coordinated.
Absolute Spectrum of Methemoglobin
metHb/CNlow spin
aquometHbhigh spin
ß
ß
490-510 600-630
500
630
deoxyHbhigh spin
oxyHblow spin
High Spin State vs. Low Spin State
Elemental Fe [Ar]3d64s2
Fe2+ (ferrous) [Ar]3d6
Fe3+ (ferric) [Ar]3d5
metHb- low pH- high pH-If there +H2O +OH -
were no ligands
1/21/2
1/21/21/2
5/2
1/2 1/2 - 1/21/2 - 1/2
1/2high spin low spin
Absolute Spectrum of Methemoglobin
low spin
high spin
ß
ß
490-510 600-630
500
630
Inositol hexaphosphoric acid (IP6 or phytic acid) binding to met Hb:
Physiological relevance - the binding of organic phosphates to ferrohemoglobinresults in a decrease in oxygen affinity, linking erythrocyte metabolism to gas
transport.
IP6 binding causes a change in spin character ofß chain hemes.
p-hydroxymercuribenzoate (pMB) binding to met Hb
ß - Cys 93
93145
94
146
92
subunit in red
pMB binding changes the spin character of met Hb
Remember
KCN, NaN3, PMB, NaF are all toxic. (NaF eats glass when wet.)
Use a reference cell or a baseline reading for subtracting buffer.
Save all files in the directory UVPC:Data:Bios532.
All buffers, chemicals are pre-prepared, use only what you need.
Store buffers, chemicals, protein solutions in 4C between labs.
Clean cuvettes with ethanol, air dry.
Rinse syringes with millipore H2O.