bk-250 proving instructions - honeywell gas depot · bk-250 diaphragm meters can be tested on snap...

2
1. Compatibility a. SNAP I, II, III or SNAP MMX BK-250 diaphragm meters can be tested on SNAP I, II, III or MMX proving systems with the use of an optical sensor fixture when testing flow rates at normal 0.5 in.WC differential pressure drop across the meter inlet and outlet. The maximum flow rate should not exceed 250 CFH. With SNAP MMX proving systems, program modifications can be made to shorten test time. 2. SNAP Setup a. Sensor An available OPTIC SENSOR ASSEMBLY is recommended for the most accurate testing method of BK-250 diaphragm meters. When using the OPTIC SENSOR ASSEMBLY, it must read off of the primary hand (1FT, 2FT or 0.05M). Consult your SNAP Operator’s Manual for complete instructions. b. Configuration New meter test parameters can be added to any SNAP proving system meter files. Consult your SNAP Operator’s Manual for complete instructions. c. Test Parameters Recommended Open Rate: 200 – 250 CFH air (80 – 100%) Recommended Check Rate: 50 – 75 CFH air (20 – 30%) Meter Rev/ Unit Vol: 1FT³ Index Hand: 1.000 2FT³ Index Hand: 0.500 0.05M³ Index Hand: 0.5663 Figure 1. 1FT³ Index Meter File Setup This document provides the instructions for proving an Elster American Meter BK-250 diaphragm meter on a Sonic Nozzle Automatic Prover (SNAP). Contact the Elster Meter Services Metrology department before making any physical changes to a SNAP proving system to ensure proper operation for any new application. BK-250 Diaphragm Meter SNAP Proving Instructions Recommended Pulses/Revs 1FT³ Index 2FT³ and 0.05M³ Indexes Open Flow Rate 2 1 Check Flow Rate 2 1 Exercise 2 1

Upload: others

Post on 13-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: BK-250 Proving Instructions - Honeywell Gas Depot · BK-250 diaphragm meters can be tested on SNAP I, II, III or MMX proving systems with the use of an optical sensor fixture when

1. Compatibility a. SNAP I, II, III or SNAP MMX BK-250 diaphragm meters can be tested on SNAP I, II, III or

MMX proving systems with the use of an optical sensor fixture when testing flow rates at normal 0.5 in.WC differential pressure drop across the meter inlet and outlet. The maximum flow rate should not exceed 250 CFH. With SNAP MMX proving systems, program modifications can be made to shorten test time.

2. SNAP Setup a. Sensor An available OPTIC SENSOR ASSEMBLY is recommended for the

most accurate testing method of BK-250 diaphragm meters. When using the OPTIC SENSOR ASSEMBLY, it must read off of the primary hand (1FT, 2FT or 0.05M). Consult your SNAP Operator’s Manual for complete instructions.

b. Configuration New meter test parameters can be added to any SNAP

proving system meter files. Consult your SNAP Operator’s Manual for complete instructions.

c. Test Parameters Recommended Open Rate: 200 – 250 CFH air (80 – 100%) Recommended Check Rate: 50 – 75 CFH air (20 – 30%) Meter Rev/ Unit Vol: 1FT³ Index Hand: 1.000

2FT³ Index Hand: 0.500 0.05M³ Index Hand: 0.5663

Figure 1. 1FT³ Index Meter File Setup

This document provides the instructions for proving an Elster American Meter BK-250 diaphragm meter on a Sonic Nozzle Automatic Prover (SNAP). Contact the Elster Meter Services Metrology department before making any physical changes to a SNAP proving system to ensure proper operation for any new application.

BK-250 Diaphragm Meter

SNAP Proving Instructions

Recommended Pulses/Revs 1FT³ Index2FT³ and 0.05M³

Indexes

Open Flow Rate 2 1

Check Flow Rate 2 1

Exercise 2 1

Page 2: BK-250 Proving Instructions - Honeywell Gas Depot · BK-250 diaphragm meters can be tested on SNAP I, II, III or MMX proving systems with the use of an optical sensor fixture when

2221 Industrial Road | Nebraska City, NE 68410-6889 USA | T +1 402 873 8200 | F +1 402 873 7616 | www.elster.com EAM-II3013 11/15

© 2015 by Elster. All rights reserved. Information contained herein is subject to change without notice. Contact your Elster representative for the most current product information.Elster and Elster Logo are trademarks of Elster American Meter Company LLC.

Figure 2. 2FT³ Index Meter File Setup

Figure 3. 0.05M³ Index Meter File Setup