Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600 1800 m/z 0 20 40 60 80 100 Relative Abundance

Upload: others

Post on 22-Mar-2020

13 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Block Diagram of a Mass Spectrometer

IonSource

Mass Analyzer Detector

“High” Vacuum

SampleIn

Computer

400 600 800 1000 1200 1400 1600 1800m/z

0

20

40

60

80

100

Re

lative

Ab

un

da

nce

Page 2: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Mechanism of Electrospray Ionization (ESI)

+ -++

+

++ +

+

+ +

+-

--

+ +

+ +

-+

+

++ +

+ -+ +

+

-

++

+

+ +

+ ++ -

+ ++

-

++

++ +

++

+

-++

+-

++

+++

+ ++

-

++

+

++ -

++

++

-

+

+

++ +

++ +

++ +

++ ++ + +

Charged

Capillary

Needle

Heated

Capillary

+3-8 kV

Taylor cone Rayleigh

limit hit:

Coulombic

explosion

Page 3: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

ESI-MS Spectrum of Horse Apomyoglobin

ApomyoglobinMW=16,951.2

+18

+20

+16 +14

+12

+10+22

m/z

Page 4: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

ESI-MS Spectrum of Bovine Serum Albumin

AlbuminMW=66,621

+41

+52

+34

m/z

Page 5: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Block Diagram of a Mass Spectrometer

IonSource

MassAnalyzer

Detector

“High” Vacuum

SampleIn

Computer

400 600 800 1000 1200 1400 1600 1800m/z

0

20

40

60

80

100

Re

lative

Ab

un

da

nce

MassAnalyzer

Activation

orReaction

Tandem Mass Spectrometry or MS/MS

Page 6: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Block Diagram of a Mass Spectrometer

IonSource

Detector

“High” Vacuum

SampleIn

Computer

400 600 800 1000 1200 1400 1600 1800m/z

0

20

40

60

80

100

Re

lative

Ab

un

da

nce

Tandem Mass Spectrometry or MS/MS

Page 7: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Block Diagram of a Mass Spectrometer

IonSource

Detector

“High” Vacuum

SampleIn

Computer

Tandem Mass Spectrometry or MS/MS

m/z

Page 8: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

S Q H L F A P I S E K

200 400 600 800 1000 12000

20

40

60

80

100

Re

lative

Abu

dan

ce

m/z

Page 9: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

S Q

H L

F A

P I S

E K

147

276

363

477

574

645

792

905

1042

1170 216

353

467

614

685

782

895

982

1111

200

40

06

00

800

100

01

20

00

20

40

60

80

10

0

Relative Abudance

m/z

881111

1042

905

982

792895

685644

M-NH3

574

1042 / 2

467

353

Page 10: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

200 400 600 800 1000 1200 14000

10

20

30

40

50

Rela

tive A

bundance

m/z

Identification of Peptide Sequences Using Tandem Mass Spectra and SEQUEST

200 400 600 800 1000 1200 14000

20

40

60

80

100

Rela

tive A

bundance

m/z

Tandem Mass Spectrum

Preprocessing

200 400 600 800 1000 1200 14000

20

40

60

80

100

Re

lative A

bunda

nce

m/z

VHNFLEAGNLGLVK

EVDANGNVTLNFFA

VHNFKFKLNHTK

SKNGKKKKNFNMS

RANRIKASVRIVQ Th

eo

reti

cal S

pectr

um

Cro

ss

-Co

rre

lati

on

Page 11: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Data Dependant Acquisition

Is Intensity >

Threshold

Full Scan

380-2000 m/z

Isolate m/z and

Acquire MS/MS

Add Mass to

Mass List

Is Ion

Present In

Mass List

No

No

Yes

Yes

Yates et al. Anal Chem, 67, 1426 (1995)

Stahl et al. JASMS, 7, 532 (1996)

Page 12: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Proteomics: Traditional

Technology

Characterize peptides by mass spectrometry

N

C

CN

2D Gel 1D Gel

Page 13: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

µLC

Ab

un

da

nc

e

Time

Ab

un

da

nc

e

Time

Shotgun Proteomics

Protein

Mixture

proteolysis

Peptide

MixtureA

bu

nd

an

ce

m/z

Ab

un

da

nc

e

m/z

MS

MS/MS1

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

1

2

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

2

3

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

3

Page 14: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Shotgun Characterization of Protein Mixtures

C. elegans

Extract Protein Mixture

Denature

Digest into Small Peptides

Page 15: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Measurement of Peptides by Microcapillary Liquid Chromatography Tandem Mass Spectrometry

Traditional Chromatography Column

Microcapillary Chromatography Column

Page 16: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Measurement of Peptides by Microcapillary Liquid Chromatography Tandem Mass Spectrometry

Traditional Chromatography Column

Microcapillary Chromatography Column

Peptide Mixture

Time

Abund

ance

Peptides are loaded onto the column

and separated by liquid chromatography

Page 17: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Acquiring Mass Spectrometry Data of Peptides

• Peptides from the column are ionized and transmitted into the vacuum system using electrospray and trapped in the 2-dimensional ion trap

Microcapillary

Column

2 kV 30 V

Page 18: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

m/z

Abund

ance

• Peptides from the column are ionized and transmitted into the vacuum system using electrospray and trapped in the 2-dimensional ion trap

• Ions are selectively ejected radially by their m/z to produce a mass spectrum

• The instrument data-system automatically selects a m/z for further analysis

Selected m/z

Acquiring Mass Spectrometry Data of Peptides

Page 19: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Acquiring a Fragmentation Spectrum

• The trap is then refilled

Page 20: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

• The trap is then refilled

• All ions are ejected from the trap except the predetermined m/z

• The energy of the isolated ion is increased resulting in collisions with the helium gas in the trap

Acquiring a Fragmentation Spectrum

Page 21: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

m/z

Abund

ance

• The collisions with the helium molecules result in structure specific fragment ions

• The fragment ions are then selectively ejected to produce a tandem mass spectrum

Acquiring a Fragmentation Spectrum

Page 22: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Ab

un

da

nc

e

m /z

Ab

un

da

nce

m/z

Ab

un

da

nc

e

m/z

Mass Spectrum

Ab

un

da

nce

m /z

Review

1. Acquire mass spectrum

2. Select precursors for tandem mass spectrometry

3. Selectively isolate and fragment precursor ions.

4. Repeat for 3 different precursor ions

1 2 3 Tandem Mass Spectra of Peptides

1 to 2 Seconds per cycle

Page 23: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Spectrum Identification

Database: fasta file

…Beowulf Computer Cluster

SEQUESTPeptide ID list

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

>MEKK1 (kinase)MDRILARMKKSTRRGGDKNITPVRRLERR…

>ATMKK5 (kinase kinase)MKPIQSPSGVASPMKNRLRKRPDLSPPLPHRDVALAVLP…

MS/MS spectra

ID proteins

from peptides…

Scan1 0.7 EGSSDEEVP…

Scan1 0.3 TFAEILNPI…

Scan1 0.2 ARFDLNNHD…

-------------------

Scan2 0.5 EDEESIRAV…

Scan2 0.2 WLGDDCFMV…

Scan2 0.1 IDRAAWKAV…

-------------------

Scan3 0.2 EITTRDMGN…

Scan3 0.1 GRNMCTAKL…

Page 24: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

>MEKK1 (kinase)MDRILARMKKSTRRGGDKNITPVRRLERR…

>ATMKK5 (kinase kinase)MKPIQTFAEILNPITWRNRLRKRPDLSPPLPHRDVALAVP…

Spectrum Identification

Database: fasta file

SEQUESTPeptide ID list

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

MS/MS spectra

TFAEILNPITWR

score = 0.3

Scan1 0.7 EGSSDEEVP…

Scan1 0.3 TFAEILNPI…

Scan1 0.2 ARFDLNNHD…

-------------------

Scan2 0.5 EDEESIRAV…

Scan2 0.2 WLGDDCFMV…

Scan2 0.1 IDRAAWKAV…

-------------------

Scan3 0.2 EITTRDMGN…

Scan3 0.1 GRNMCTAKL…

z = 2, m/z = 720.6

mass = 1441.2

Page 25: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 26: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 27: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 28: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 29: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 30: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 31: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 32: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

MudPIT

• 100 µm i.d. fuse silica

Link et al, Nat Biotech (1999)

Washburn, Wolters, Yates, Nat Biotech (2001)

Page 33: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

• 100 µm i.d. fuse silica

• Tip is pulled to ~5 µm I.D.

MudPIT

Page 34: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

• 100 µm i.d. fuse silica

• Tip is pulled to ~5 µm I.D.

• Pack with C18 material 1st

MudPIT

Page 35: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

• 100 µm i.d. fuse silica

• Tip is pulled to ~5 µm I.D.

• Pack with C18 material 1st

• Pack with SCX material 2nd

MudPIT

Page 36: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

• 100 µm i.d. fuse silica

• Tip is pulled to ~5 µm I.D.

• Pack with C18 material 1st

• Pack with SCX material 2nd

• Peptide digest is loaded off-

line

MudPIT

Page 37: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

• 100 µm i.d. fuse silica

• Tip is pulled to ~5 µm I.D.

• Pack with C18 material 1st

• Pack with SCX material 2nd

• Peptide digest is loaded off-

line

MudPIT

Page 38: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Multidimensional Protein Identification Technology (MudPIT)

MudPIT Column

Page 39: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

2 kV RP-C18SCX

HPLC

H2O MeCN NH4OAc

waste

~200 nl/min

MudPIT

Page 40: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

2 kV RP-C18SCX

HPLC

H2O MeCN NH4OAc

waste

• Step 1 – Reverse Phase Gradient

MudPIT

Page 41: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

2 kV RP-C18SCX

HPLC

H2O MeCN NH4OAc

waste

• Step 1 – Reverse Phase Gradient

• Step 2 – Salt Pulse

MudPIT

Page 42: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

2 kV RP-C18SCX

HPLC

H2O MeCN NH4OAc

waste

• Step 1 – Reverse Phase Gradient

• Step 2 – Salt Pulse

• Step 3 – Reverse Phase Gradient

MudPIT

Page 43: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

2 kV RP-C18SCX

HPLC

H2O MeCN NH4OAc

waste

• Step 1 – Reverse Phase Gradient

• Step 2 – Salt Pulse

• Step 3 – Reverse Phase Gradient

• Step 4 – Increase Salt Pulse

• Repeat

MudPIT

Page 44: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

µLC/µLC/MS/MS of C. elegans Peptides

1 2 3 4

5 6 7 8

9 10 11 12

MS/MS Spectra: 280290

Peptides Identified: 2753

Proteins IdentifiedWith >95% Confidence: 941

RT: 0.00 - 120.00

0 10 20 30 40 50 60 70 80 90 100 110

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rela

tive

Abu

nda

nce

29.49 34.93

45.53

10.6110.87

11.41

27.34 57.6227.05

35.40

38.8222.90

0.15

40.90

64.17

12.66

48.81 57.088.65

3.65 50.73

114.8864.50

71.3165.67

71.71 114.5877.60 81.56 111.5686.77 100.89

90.15 101.37

NL: 4.86E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061504-worm-trizol-urea-15n-J1-01

RT: 0.00 - 120.03

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

37.18

37.40

48.33

48.73

21.24

37.8065.92

21.6220.86

66.11

44.58

71.1738.0757.06

75.9263.93

39.8530.48 60.44 80.17

29.14

24.33 52.83 84.50

109.8384.9720.07 111.475.5286.15 108.71

18.715.824.78 88.49 104.8814.237.32

NL: 3.00E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061504-worm-trizol-urea-15n-J1-02

RT: 0.00 - 120.02

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

42.9723.31

39.00

36.26

61.04

35.84

61.28

27.2560.66

50.54

50.7030.3549.91

20.24

57.01 66.73

56.64 96.28

13.31 76.4167.3177.0677.30

87.4478.7567.65 87.616.45 14.31

97.84102.416.13 109.82

4.32

NL: 3.22E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-03

RT: 0.00 - 120.04

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

30.7933.70

34.49

46.8046.31

37.27

47.31

29.7544.78

44.64 49.4129.29

51.1513.96

71.99

19.61 39.2424.01

72.1656.7697.16

96.6065.30 75.799.7298.028.06 64.83

78.22 98.2061.7983.826.34

93.8385.68

98.7999.263.80

103.52 110.00

NL: 3.90E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-04

RT: 0.00 - 120.02

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

42.83

42.54

43.19

33.70

28.87

37.98

57.48

57.88

61.2944.0519.65 26.08 65.47 73.3725.909.47 56.2445.17 73.88

54.734.06 18.81 23.89 93.3574.1367.81 90.55112.8494.0486.6578.16 100.45

NL: 4.86E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-05

RT: 0.00 - 120.04

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

34.47

34.71

33.94

35.07

40.17

32.53

40.54

30.97 88.67

30.68 88.4911.2461.0741.29 56.73

89.0614.73 88.1746.97

15.03 54.9723.148.13 73.837.92 76.1268.6318.36 89.2448.25 68.27 96.654.08 84.68

62.67 93.26 111.27100.0377.12112.18101.87

113.49

NL: 3.59E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-06

RT: 0.00 - 120.02

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

38.51

41.72

41.99

34.3811.36

86.8650.93 69.6220.35 49.28

52.92

69.9754.4472.76

56.25

56.74

68.9218.23 73.2160.5960.83

73.3932.28 61.38

23.49 24.04 42.89 73.5967.11

79.9231.29

88.1092.9280.25

4.18 81.2693.48

99.03112.53100.76

NL: 1.44E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-07

RT: 0.00 - 120.01

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

89.83

73.2348.03

73.70

89.47

90.2488.22

36.26

74.16

33.3338.43

69.9338.63

40.3432.7711.70

24.1917.71 40.82

19.59

59.25

80.6658.7941.3626.71

58.53 69.3680.99

49.99

50.3159.67

91.174.77 93.13

61.06 84.52 94.729.18 98.66 108.64108.02 108.95

100.80 110.93

NL: 1.39E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-08

RT: 0.00 - 120.00

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

30.16

30.32

90.94

67.78

91.62

33.5567.63

74.33

92.0921.85

20.8740.24

40.6688.63

35.6288.4825.97

48.2911.83 59.53 108.9392.60

108.3411.6548.77 109.6743.6110.38 12.12 78.31 108.11

64.6858.64 109.9378.6912.68 92.9997.1579.20

107.623.6181.48 111.0298.37

4.82 113.79

NL: 1.55E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-09

RT: 0.00 - 120.00

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

21.42

67.19

66.96

67.47

109.24

108.4435.21

26.4267.84

109.6029.97 87.91

89.9068.1276.36

51.2266.0951.0244.85

55.11 110.3240.66 63.75 75.64 76.93

63.3811.31 107.55 110.86

55.5511.80 77.5369.9419.023.70 94.42 103.99 112.57

4.30

NL: 2.25E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-10

RT: 0.00 - 120.00

0 10 20 30 40 50 60 70 80 90 100 110

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

20.95

27.33

106.92106.57

70.27

106.3770.5231.61 107.70

43.55

43.09108.04

40.5143.86 108.62

44.5085.9837.92 46.26 105.6052.64 108.86

72.4476.3446.4576.64 109.68

57.01 86.9369.60 110.1577.1015.889.36 57.56 64.09 78.54 111.374.07 91.98

92.61112.11

NL: 2.45E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061704-worm-trizol-urea-15n-H-11

RT: 0.00 - 120.01

0 10 20 30 40 50 60 70 80 90 100 110 120

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Abu

nd

ance

26.72 28.65

42.56

20.27

42.82

14.938.67

25.21

62.3811.68

65.97 72.52

31.37 39.67111.39

59.61111.0256.2039.09 77.0844.82

77.688.1950.19

112.206.41

87.1492.32

79.56116.2993.36 99.44

5.60

NL: 3.47E7

Base Peak F: ITMS + c ESI Full ms [ 400.00-1400.00] MS 061504-worm-trizol-urea-15n-J1-12

Page 45: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Qualitative Analysis of Tandem Mass SpectraA

bu

nd

an

ce

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

Ab

un

da

nc

e

m/z

100,000’s of

MS/MS Spectra

Peptide IDsSpectrum 1: ERTYILDFGH

Spectrum 2: WEQTMNVCSW

Spectrum 3: ERTYEWIPWQ

Spectrum 4: …

…Beowulf Computer Cluster

Database Search

Using SEQUEST

Output Sorting and

Filtering

Protein

Identifications

Assembly Using

DTASelect

Page 46: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 47: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 48: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 49: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 50: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 51: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 52: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 53: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 54: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 55: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 56: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 57: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 58: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 59: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 60: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 61: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 62: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 63: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 64: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

Writing a Manuscript Review

• Overview (~1-2 paragraphs) – summarizing the

general point of the manuscript and your broad

overall enthusiasm

• Major comments – bullet points

– Items that must be fixed before publication

• Minor comments – bullet points

– These should be stylistic recommendations

– Typos

– Etc…

Page 65: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 66: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600
Page 67: Block Diagram of a Mass Spectrometer · Block Diagram of a Mass Spectrometer Ion Source Mass Analyzer Detector “High” Vacuum Sample In Computer 400 600 800 1000 1200 1400 1600

benign prostate hyperplasia

Prostate Cancer