blood component fractionation: manual versus automatic ... · blood component fractionation: manual...
TRANSCRIPT
www.elsevier.com/locate/transci
Transfusion and Apheresis Science 30 (2004) 23–28
Blood component fractionation: manual versusautomatic procedures
Daniela Pasqualetti a,*, Alessandro Ghirardini b, Maria Cristina Arista a,Stefania Vaglio a, Azis Fakeri a, Alan A. Waldman c, Gabriella Girelli a
a Department of Cell Biotechnology and Hematology, University ‘‘La Sapienza’’, Blood Bank, Via Chieti 7, Rome 00161, Italyb Istituto Superiore di Sanit�a, Laboratory of Epidemiology and Biostatistics, Rome, Italy
c Waldman Biomedical Consultancy, Oceaside, New York, USA
Abstract
Over the last few years, quality system requirements have been introduced for blood components. The necessary
compliance with standard productions will have a considerable impact on Blood Banks. The introduction of automated
methods is the most satisfactory means to meet these requirements for blood component preparation.
A new device has been developed to automate the fractionation of blood into components. We evaluated the efficacy
of this instrument as compared to manual methods. A total of 218 units of blood have been collected, into several
different commercial blood bag systems (77 into standard quadruple bag systems, 141 into bag systems with integrated
in line filters), and used to evaluate the universality of the instrument. Whole blood units were processed using the Top/
Top system and the Compomat G4 (Fresenius HemoCare). A separate program protocol was developed for each kind
of bag.
Use of the Compomat G4 resulted in a statistically significant (p < 0:001) increase of the hemoglobin in filtered red
cell concentrates (RCC) in comparison with the manual procedure, and a similar trend, even not statistically significant,
has been observed for filtered RCC. Regardless of bag systems, we were able to observe a statistically significant in-
crease of platelets in the platelet concentrates (PCs), when comparing automatic versus manual procedure.
The automated procedure has been shown to be fast, and easy for the operators. This device reliably produces
acceptable blood components, and has been shown adaptable to use with different blood bag systems.
� 2003 Elsevier Ltd. All rights reserved.
1. Introduction
The implementation of a quality system for
blood component production has been much dis-
* Corresponding author. Tel.: +39-06-85795-524; fax: +39-
06-85795-501.
E-mail address: [email protected] (D. Pasqua-
letti).
1473-0502/$ - see front matter � 2003 Elsevier Ltd. All rights reserve
doi:10.1016/j.transci.2003.07.002
cussed over the last few years, and bringing stan-
dard production into compliance will have a
considerable impact on Blood Banks. In actual
practice, it often has been necessary to change
working methods in order to guarantee repro-
ducible production of the blood components,
introducing careful controls of the processing
procedures involved and performing quality testsof the blood products on a routine basis. While in
the past the procedures for preparing fractions of
d.
24 D. Pasqualetti et al. / Transfusion and Apheresis Science 30 (2004) 23–28
blood components were based completely on
manual operations, automated methods have re-
cently been introduced as the most efficient and
effective way to fulfil the standards for bloodcomponents production and conform with the
European guidelines [1].
Using collection bags which allow for removal
of materials only through a port on the top (Top/
Top system), standard blood fractionation in the
United States begins with the separation, by low-
speed centrifugation, of platelet-rich plasma
(PRP), red cell concentrates (RCC) and buffy-coat(BC) from whole blood (WB). In a subsequent
high-speed centrifugation, platelet concentrates
(PCs) and platelet poor plasma (PPP) are derived
from the PRP [2]. In Europe, in the 1990�s, a dif-
ferent approach was introduced, one designed to
decrease development of the platelet storage lesion
and in vitro activation due to the second centri-
fugation [3,4]. Using collection bags which allowfor removal of materials through ports on both the
top and the bottom (Top/Bottom system), it is
possible, following an initial high-speed centrifu-
gation of the WB, to produce PPP on the top, with
the platelets concentrating into a BC on a cushion
of RCC at the bottom [5]. After collection of this
BC, the platelets can be separated by a gentle low-
speed centrifugation, a process that presumablyproduces less damage than high-speed packing,
which causes formation of tight pellets of plate-
lets [6].
In actual practice, however, examination of
platelets stored for up to five days indicated no
significant differences in platelet survival and
function with the two preparative techniques [7,8],
and it is still being debated whether or not plateletsfrom PRP or from BC are better [8].
While there have been several studies performed
to validate the performance of an automatic
instrument with the Top & Bottom system [9–11],
the results of using an automatic extractor for the
Top/Top system for blood component preparation
have not been reported.
As part of our efforts to improve productquality and to reduce manufacturing problems, we
have tested the Compomat G4 (Fresenius Hemo-
Care) equipment developed to automate Top/Top
fractionation, and compared the quality of the
components produced with those obtained by the
standard manual method. As part of this study, we
evaluated Top/Top blood bag systems from dif-
ferent manufacturers.
2. Materials and methods
To check on the universality of the automatic
instrument Compomat G4 (Fresenius HemoCare),
we collected from blood donors 218 blood units,
each of 450 ml ± 10%, into 63 ml of CPD antico-
agulant solution, using blood bag collection sys-
tems from different manufacturers.Standard quadruple bag systems were used for
77 of the blood collections, while blood bag
collection systems with integrated in-line filters
were used for 141 of the blood collections. In the
first group of bag systems, manual procedures
were used to manufacture components from 13
units collected into Maco Pharma bags, 12 units
collected into Terumo bags, and 10 units col-lected into Fresenius bags, while the automatic
device was used to manufacture components
from 8 units collected into Maco Pharma bags,
20 units collected into Terumo bags, and 14
units collected into Fresenius bags. In the second
group of bags system with in-line filters, manual
procedures were used to manufacture compo-
nents from 80 units collected into Fresenius bagsand from 18 units collected into bags manufac-
tured by Baxter, while the automated device was
used to manufacture components from 31 units
collected into Fresenius bags and 12 units col-
lected into Baxter bags.
The quadruple bag systems were processed
immediately after blood collection, while all the in-
line bag systems were processed after 2 h sinceblood collection and after cooling at a conditioned
temperature between 18 and 22 �C. All the WB
units were centrifuged at 550g for 8 min at 20 �C,with an acceleration rate of 2 and deceleration 1 in
a Beckman J-6M/E centrifuge, to produce RCC.
PC were prepared from PRP by a second centri-
fugation at 2200g for 10 min.
The automated fractionation was performedusing Compomat G4 (Fresenius-HemoCare).
D. Pasqualetti et al. / Transfusion and Apheresis Science 30 (2004) 23–28 25
Following each of the centrifugations, the blood
packs are taken out of the bucket, the primary
bag hung on the pins on the front of the machine,
and the satellite bags placed on the top of theCompomat G4, with the integrated tubing being
placed in built-in automated clamper/sealers. After
the first centrifugation of the WB, as directed
by the software, three presses combined with
the integrated detectors express out PRP and,
if desired, BC, into satellite bags, leaving the
RCC. After the second centrifugation, the pre-
sses and integrated detectors express out the PPPinto a satellite bag, leaving the PCs. The detectors,
the sensitivity of which were set to meet our stan-
dards, allow control over the blood component
separation, and standardization for different bags.
The separation of the 141 blood units with in-
line filters was performed with the same procedure
as described above, placing the filter upside down
on the right hand side panel of the Compomat. Inaddition, this equipment has been used on Frese-
nius filters with the partial squeezing of Sag-M
through the filter by means of the top press. The
filtration of the 141 RCC with in-line filters was
carried out by gravity.
By using Compomaster, a software running on
Windows NT that manages the Compomat, the
Compomat G4 is able to process the blood unitswith a programmed and personalized protocol for
each kind of bag. The Compomaster/Compomat
programs every procedure, from the start to end of
separation, and registers all the data. At our site,
in order to obtain the best recovery results for the
RCC, defined as lowest hemoglobin loss into the
small BC volume and a low level of WBC con-
tamination, we developed specific parameters forthe processing programs for the different bag sys-
tems, as follows:
(a) Fresenius and Maco Pharma blood bag sys-
tems:
The upper press is pushed out to 54.0 mm and
the red cell level reaches first the DET A2 by
means of both presses, then the DET A6+B usingonly the lower press with a speed 20% to reduce
the squeezing flow. The combination of these
parameters mainly defines the features of the BC,
in terms of volume and hematocrit.
(b) Terumo and Baxter blood bag systems:
Upper press out to 53.0 mm, both presses out
DET A1 and lower press out DET A8+B with a
speed 25%.(c) Fresenius and Baxter blood bag systems with
in-line filters:
Upper press out 50.0 mm, lower press out DET
A6 with speed 15% and then lower press out DET
A8+B with speed 20%.
(d) PRP derived from all systems:
Both presses to 45.0 mm, upper press to 56.0
mm and lower press to 58.0 mm.
After standardization of the method for stain-
ing of leukocytes with a commercial kit (Leuco-
Count, Becton Dickinson, San Jose, Costa
Mesa-CA, USA), this kit and calibrated flow cyto-
metry (FACScan, Becton Dickinson, San Jose,
Costa Mesa-CA, USA) were used to count the
absolute number of the residual [12]. Otherhematological parameters were measured with an
automated analyser (AcT.5DiffCP-Coulter, Instru-
mentation Laboratory, Milan, Italy).
Statistical analysis was performed by comparing
mean value of the relevant variables, by using the
Epi Info software (version 6) produced by the
Centers forDisease Control andPrevention (CDC).
3. Results
The average gain in hemoglobin recovery in
RCC was equal to or more than 3 g/unit for both
filtered and not filtered RCC. As shown in Tables
1 and 2, the automatic procedure is associated with
a statistically (p < 0:001) significant increase in
hemoglobin recovery in filtered RCC, while for
RCC the difference does not reach the significant
level (p ¼ 0:12). The apparent greater increase inred cell recovery for the Fresenius bags (Table 3) is
felt, in part, to be due to the large standard devi-
ation in the volumes of the individual collections.
With regard to the higher values for hemoglo-
bin in the filtered RCC (Table 2), this is apparently
due to a decreased loss in the filters, a loss that, for
the Baxter bags, falls to 11.17% (Table 4).
It should be noted that the Fresenius filterwas hard-housing and the Baxter filter was
Table 1
Comparison of results between manual and automatic proce-
dure in standard bag systems from all different manufactures
Overall summary
of standard bag
systems
Manual procedure
(N ¼ 35)
Compomat G4
(N ¼ 42)
Mean (SD) Mean (SD)
RCC
Volume (ml) 335.7 (22.6) 351.6 (24.6)
Hb (g/unit) 61.0 (7.7) 64.6 (5.7)
Ht (%) 52.9 (5.3) 53.2 (2.2)
PCs
Plts (·10e11) 0.62 (0.17) 0.81 (0.2)
Recovery (%) 73.8 (14.1) 92.6 (14.5)
Table 2
Comparison of results between manual and automatic proce-
dure in in-line filter systems from all different manufactures
Overall Summary of
in-line filter systemsManual procedure
(N ¼ 98)
Compomat G4
(N ¼ 43)
Mean (SD) Mean (SD)
Filtered RCC
Hb (g/unit) 53.3 (7.8) 59.2 (7.8)
Hb loss (%) 13.0 (7.6) 10.4 (4.4)
WBC Log depletion 4.36 (0.6) 4.14 (0.3)
PCs
Plts (·10e11) 0.61 (0.2) 0.71 (0.3)
Recovery (%) 73.5 (15.1) 82.0 (24)
Table 3
Comparison of the results produced by Compomat G4 of red
cell concentrates using bag systems from three different manu-
facturers
Results with
different bag
system
manufacturers
Terumo
(N ¼ 20)
Fresenius
(N ¼ 14)
Maco Pharma
(N ¼ 8)
Mean (SD) Mean (SD) Mean (SD)
RCC
Volume (ml) 340.0 (12.4) 370.4 (25.1) 365.4 (18)
Hb (g/unit) 62.6 (3.2) 68.4 (6.5) 57.2 (3.9)
Ht (%) 52.8 (2.1) 53.8 (2.6) 52.8 (2.2)
PCs
Plts (·10e11) 0.82 (0.16) 0.85 (0.3) 0.60 (0.33)
Recovery (%) 92.9 (11.8) 95.0 (14.4) 67.1 (10.2)
Table 4
Comparison of the results produced by Compomat G4 of fil-
tered red cell concentrates with bag systems with in-line filters
from the two different manufacturers used
Results with the two
bag systems with in-line
filters manufacturers
Fresenius
(N ¼ 31)
Baxter
(N ¼ 12)
Mean (SD) Mean (SD)
Filtered RCC
Hb (g/unit) 58.1 (7.8) 62.09 (5.24)
Hb loss (%) 10.4 (4.4) 11.17 (8.6)
WBC Log depletion 4.14 (0.30) 4.08 (0.27)
PCs
Plts (·10e11) 0.70 (0.3) 0.68 (0.7)
Recovery (%) 84 (22) 76 (30)
26 D. Pasqualetti et al. / Transfusion and Apheresis Science 30 (2004) 23–28
soft-housing. The priming of the first, before thepassive filtration, was performed automatically by
use of the top press, while for the second both
priming and filtration was carried out by gravity.There was no statistically significant difference
(p ¼ 0:52) in residual WBC (Table 4) for both
types of WBC filtered components.
The results for PCs from the standard bags are
presented in Table 1. A statistically significant
difference (p < 0:003) has been observed between
the automatic procedure compared to manual
procedure, with a remarkable increment of theplatelet count, roughly 20%, in the PCs. Table 3
presents the results obtained for each of several
different manufacturers. The automatic prepara-
tion method was found to yield excellent and
comparable platelet counts with both the Terumo
and Fresenius blood bag systems, with a recovery
greater than 92.9% and 95.0%, respectively. This is
a statistically (p < 0:001) significant difference, ascompared to the results with the Maco Pharma
bags, which had a recovery of 62.1%.
Similarly, the automated method proved supe-
rior to the manual method for the preparation
of PCs from blood bag systems with filters, with
a mean recovery per unit of 0.71 · 1011 versus
0.61 · 1011 (Table 2). Table 4 presents the results
obtained for the two different bag manufacturers.No statistically significant differences were found
in the results obtained with the Fresenius and
Baxter blood bag systems with in-line filters.
The average processing time using the Compo-
mat was 2 min and 15 s for each bag. The total
time for the procedure using automatic equipment
was 4 min and 50 s, while the total time for the
manual operation was 6 min and 36 s.
D. Pasqualetti et al. / Transfusion and Apheresis Science 30 (2004) 23–28 27
There were no problems with the equipment,
and no breakdowns during the study. The only
unexpected events during the study were two
leaking seals, failures traced to incorrect posi-tioning of the tubing in the welding clamp.
4. Discussion
It is well known that, for transfusion therapy
to be effective, there must be quality and consis-
tent blood components. This study provides data
on the performance of an available automated
instrument, Compomat G4 (Fresenius Hemo-
Care), for separating blood components. Perfor-mance, the accuracy of the separation process, and
the reliability of the device using different bag
systems, were evaluated and compared with the
manual method previously used in our Blood
Center.
Performance, accuracy and reproducibility of
the automatic method were assessed in two ways,
both on ability to meet the standards in theEuropean guidelines, and on the extent of varia-
tion in the results. The results obtained for the
components produced were all well above the
standards of the European guidelines, and the
standard deviation of the parameters measured
were either the same or lower for the automatic
procedure than for the manual procedure.
Our results demonstrate that when this devicewas used to prepare red cell concentrates (RCC)
following a low-speed centrifugation to prepared
PRP, the level of hemoglobin in the RCC was
higher than for the manual method. When this
device was used in conjunction with in-line filters,
there was less loss of hemoglobin during prepara-
tion of the filtered RCC than for the manual
method. In Europe, it is required [1] that RCC andfiltered RCC units contain more than 43 and 40 g
hemoglobin/unit, respectively. This study suggests
that, in the light of the ongoing standardization of
these methods, higher values for the expected lev-
els of hemoglobin could be set.
Similarly, our results demonstrate that when
this device is used to separate PPP to yield PC,
there is a marked improvement in the plateletcontent of the component. Indeed, the level of
recovery was high enough to allow us to reduce the
number of units used for the preparation of pooled
PCs. This reduces the risk of exposure to infectious
agents and to allogeneic stimuli as well as reducingthe costs of this product. In our facility, where the
increased request for PCs does not allow storage of
the PCs for more than one or at most two days, we
have established that it is best to prepare platelets
by this two-step method, using the automated
device Compomat G4.
Usage of pools of PC made by the two-step
centrifugation method has, in the past, beenquestioned. Since the beginning of the 1990�s in the
United States, in order to reduce risks of alloim-
munization, there has been an increase in the use
of apheresis-derived platelet concentrates (APCs),
rather than single unit PCs, for treatment of
thrombocytopenic patients. In 1994, more than
half of all transfused platelets were performed with
APC�s [13], with markedly increased costs. How-ever, the TRAP Study group [14] demonstrated
that, when blood products are leucodepleted, there
is no advantage of APCs over PCs with regard to
alloimmunization rates. It also was suggested, by
several studies in the past, that the quality of BC-
PCs was better than PRP-PCs [15,16]. However,
recent studies demonstrate that there are no dif-
ferences in low morphological score and recoverybetween filtered pooled BC-PCs and PRP-PCs [17]
and that, in vitro, the differences of activation,
fibrinogen and glycoproteins were not significant
after one day of storage [7].
Our Center processes approximately 20,000
blood units/year, and for each product uses a
specific kind of bag, with the aim of improving the
quality process, so we evaluated the universalapplicability of the instrument, using the same
separation process with bag systems from several
different manufacturers. The results confirm that
this device can be used successfully with bags from
numerous sources.
In conclusion, given the improvement in the
production and content of the blood components
produced, the compliance with quality require-ments and the easy, rapid and reproducible tech-
nology, we would recommended the use of the
Compomat/Compomaster system to process blood
into its components.
28 D. Pasqualetti et al. / Transfusion and Apheresis Science 30 (2004) 23–28
Acknowledgements
The authors express their appreciation to Mrs
M.L. Salvagnini and Mrs G. Alimenti for thetechnical work and for help with the preparation
of the samples tested in this study. The authors are
also grateful to Dr. Chiara Franciosi for helpful
discussions.
References
[1] Guide to the preparation, use and quality assurance of
blood components. 6th ed. Strasbourg: Council of Europe
Publishing; 2000.
[2] The American Association of Blood Banks Annual Report
1990. Arlington, Va, American Association of Blood
Banks, 1990.
[3] Prins HK, de Bruijn JCGH, Henrichs HPJ, Loos JA.
Prevention of microaggregate formation by removal of
‘‘buffycoats’’. Vox Sang 1980;39:48–51.
[4] Fijnheer R, Veldman HA, van der Eertwegh AJ, Gouw-
erok CW, Homburg CH, Boomgaard MN, et al. In vitro
evaluation of buffy-coat-derived platelet concentrates
stored in a synthetic medium. Vox Sang 1991;60:16–22.
[5] Hogman CF, Eriksson L, Hedlund K, Wallvik J. The
bottom and top system: a new technique for blood
component preparation and storage. Vox Sang 1988;55:
211–7.
[6] Fijnheer R, Pieters RNI, de Korte D, Gouwerok CWN,
Dekker WJA, Reesink HW, et al. Platelet activation during
preparation of platelet concentrates: a comparison of the
platelet-rich plasma and the buffy coat methods. Transfu-
sion 1990;30:634–8.
[7] Lozano M, Estebanell E, Cid J, Diaz-Ricart M, Mazzara
R, Ordinas A, et al. Platelet concentrates prepared and
stored under currently optimal conditions: minor impact
on platelet adhesive and cohesive functions after storage.
Transfusion 1999;39:951–9.
[8] Anderson KC, Ness PM. Scientific basis of Transfusion
medicine. 2nd ed. Implications for clinical basis. Saunders
Company; 2000.
[9] Kretschmer V, Tesche U, Richter P. Improvement of the
separation of blood by modification of separators Opti-
press and Biotrans Separator. Beitr Infus 1992;30:112–21.
[10] Hurtado C, Bonanad S, Soler MA, Mirabet V, Blasco I,
Planelles MD, et al. Quality analysis of blood components
obtained by automated buffy-coat layer removal with a
Top & Bottom system (Optipress II). Haematologica
2000;85:390–5.
[11] van Rhenen DJ, Vermeij J, de Voogt J, Bernes JC, Payrat
JM. Quality and standardization in blood component
preparation with an automated blood processing tech-
nique. Transfus Med 1998;8:319–24.
[12] Dzik S, Moroff G, Dumont L, for the BEST Working
Party of the ISBT. A multicenter study evaluating three
methods for counting residual WBCs in WCB-reduced
blood components: nageotte hemocytometry, flow cytom-
etry, and microfluorometry. Transfusion 2000;40:513–20.
[13] Wallace EL, Churchill WH, Surgenor DM, Cho GS,
McGurk S. Collection and transfusion of blood and blood
components in the United States, 1994. Transfusion
1998;38:625–36.
[14] The Trial to reduce Alloimmunization to Platelet Study
Group. Leukocyte reduction and ultraviolet B irradiation
of platelets to prevent alloimmunization and refratoriness
to platelet transfusions. N Engl J Med 1997;25:1861–9.
[15] KretschmerV,BierannE, LohH. Separation of platelet con-
centrates (PC) from buffy-coats (BC) using the bottom and
top (BAT) drainage system. Transfus Sci 1990;11:363–6.
[16] Rebulla P, Bertolini F, Porretti L, Marangoni F, Smacchia
C, Marconi M, et al. Platelet concentrates from buffy coats:
improved conditions for preparation and evaluation in
routine clinical use. Transfus Sci 1993;14:41–6.
[17] Kostelijk EH, Klomp AJA, Engbers GHM, Gouwerok
CWN, Verhoeven AJ, van Aken WG, et al. Improved
platelet compatibility of water vapour glow discharge
treated non woven poly(ethylene terephthalate)leukocyte-
reduction filters for different types of platelet concentrates.
Transfus Med 2001;11:199–205.