br comtec 04.07 gbbpharma-basic.net/com.tec.pdfbmsc bone marrow stem cell processing p1y + bmsc...
TRANSCRIPT
T R A N S F U S I O N T E C H N O L O G Y
COM.TEC®THE MULTI-PROCEDURAL APHERESIS PLATFORM
Br_ComTec_04.07_GB.indd 1Br_ComTec_04.07_GB.indd 1 25.04.2007 11:57:53 Uhr25.04.2007 11:57:53 Uhr
COM.TEC®
The multi-procedural apheresis platform
Donation as well as therapeutic procedures
Continuous flow technology
Component collection outside of the centrifuge
Optional single-needle mode for selected protocols
Ultimate donor/patient safety
Apheresis sets with pre-donation capability
Easy (drawer like) installation of the separation chamber
Automatic or manual process management for all protocols
Equipped as standard with printer
Apheresismaster – the optional data management
Br_ComTec_04.07_GB.indd 2Br_ComTec_04.07_GB.indd 2 25.04.2007 11:57:58 Uhr25.04.2007 11:57:58 Uhr
The “Apheresismaster” provides a bi-directional
radio transmission capability of cell separators with
a server PC. As an option, the server PC can be
linked to the lab and the central data base.
The total monitoring of the production process with
central documentation provides completely new
perspectives. Apheresismaster documents all proce-
dures available on COM.TEC®.
CENTRALIZED MONITORING DOCUMENTATION
• For supervising the production process.
• All relevant separation data are immediately
available when alarms occur.
• Create your individual separation records match-
ing your requirements. The complete records can
be transmitted and printed on any PC.
ApheresismasterThe intelligent system
COM.TEC® 1 COM.TEC® 3
COM.TEC® 2 COM.TEC® 4
CentralDatabase
ServerPC
Br_ComTec_04.07_GB.indd 5Br_ComTec_04.07_GB.indd 5 25.04.2007 11:57:29 Uhr25.04.2007 11:57:29 Uhr
PLATELET COLLECTION
• Leuko-depleted platelet units without filtration
• Platelet collection in dual- or single-needle mode available
• Plasma recirculation for optimal flow conditions
• Donor-tailored separation parameters, for example “best flow”
• Protocol also recommended for therapeutic platelet depletion
THERAPEUTIC PROTOCOLS
By the plasma recircula-
tion the hematocrit in
the separation chamber
is adjusted to 37% and
almost constant flow con-
ditions are maintained.
This also prevents mod-
ulations of the interface,
and high level platelet
quality is achieved.
The RBC barrier ensures
the leuko-depleted flow
of platelets into the sec-
ond separation step.
Therapeutic plasma exchange
• Plasma exchange efficacy 87% on average
• Single-needle mode available
RBC exchange
• Implemented RBC replacement calculator
• Accurate achievement of desired post Hct
• Very efficient HbS depletion
RBC depletion
• Accurate achievement of desired post Hct
• Short procedure times
Plasma treatment on adsorber columns
• Flexible use of different adsorber columns
Therapeutic depletion of platelets
Platelet collection bag Whole blood inlet
Recircu-lation pump
Whole blood pump
Plasma pump
Drip chamber
Return fl ow
Cell pump
Seperation chamber
Br_ComTec_04.07_GB.indd 3Br_ComTec_04.07_GB.indd 3 25.04.2007 11:58:04 Uhr25.04.2007 11:58:04 Uhr
STEM CELL COLLECTION
• Variable protocol selection due to different donor/patient conditions
• Protocol selection dependent on targets
• Lymphocyte collection for consecutive photochemotherapy and donor lymphocyte infusions (DLI)
• CD34 + prediction
AutoMNC procedure for automated collection:
The spillover sensor detects the passing cell fraction and triggers
the spillover clamp from “return” to “collection”.
Correlation of predicted vs. collected CD34+ yield
[x106 Cells].
10000
1000
100
10
CD
34+
(in
teri
m y
ield
) [1
0e6]
CD34+ (predicted) [10e6]
10 100 1000 10000
Speaman’s correlation coefficient: r1 = 0.936 (p < 0.0001): r2 = 0.913 (p < 0.0001)
� MNC� RV-PBSC
Spillover Clamp
Drip Chamber
Spillover Detector
Collection Bag
Br_ComTec_04.07_GB.indd 4Br_ComTec_04.07_GB.indd 4 25.04.2007 11:58:23 Uhr25.04.2007 11:58:23 Uhr
Fresenius HemoCare GmbH A company of the Fresenius Kabi group
Pfingstweide 5361169 Friedberg · Germany Phone: +49-61 72-6 08 8499Fax: +49-61 72-6 08 8498
www.fresenius-kabi.com
Your local sales organization:
B 201 GB 04/2007 (Pure:Comm) © Copyright 2007, Fresenius HemoCare GmbH
Programs and disposables COM.TEC®
Programs Platelet Disposables
PLT 5d Platelet apheresis, 5-day storage C5L(simultaneous plasma harvest possible)
PLT 5d Platelet apheresis, Triple-concentrates, 5 day storage C5LT(simultaneous plasma harvest possible)
PLT 5d-SN Single needle platelet apheresis, 5 day storage S5L(simultaneous plasma harvest possible)
Programs Stem cell Disposables
PBSC-Lym Peripheral blood stem cell collection, C4Ylymphocyte collection or depletion
RV-PBSC Peripheral blood stem cell collection, RVYreduced collection volume
MNC Peripheral blood stem cell collection, P1Ymononuclear cell collection or depletion
autoMNC Automatic collection of peripheral blood stem cells, P1YAstem cells mononuclear cell collection
autoMNC Automated collection of lymphocytes in dual needle P1YAlymphocytes mode for consecutive photochemotherapy or DLI*
Granulo Granulocyte collection P1Y
BMSC Bone marrow stem cell processing P1Y + BMSC
Programs Therapy Disposables
Deplet Depletion of platelets or white blood cells C4L
RBC Therapeutic red cell exchange or depletion PL1; US: RBC
TPE Therapeutic plasma exchange PL1; US: TPE
Adsorb Therapeutic plasma treatment e. g. immunoadsorption P1R
*DLI (donor lymphocyte infusion)
Br_ComTec_04.07_GB.indd 6Br_ComTec_04.07_GB.indd 6 25.04.2007 11:57:40 Uhr25.04.2007 11:57:40 Uhr