c105 – general checklist: a2la policy on ... · web viewaccredited clinical laboratories are...

28
American Association for Laboratory Accreditation C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing Document Issued: March 27, 2014 Page 1 of 28 Accredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory Testing and P903 - Policy on Estimating Measurement Uncertainty for ISO 15189 Testing Laboratories A2LA Assessor Instructions : Please note that for all N/A indications, you must document the reason why this requirement is N/A in the comments section. IMPORTANT NOTE: An asterisk (*) in the comments section indicates that the assessor must document the specific traceable objective evidence reviewed in association with that requirement. Objective evidence information is mandatory for those clauses. To the best of my knowledge, all laboratory document references below as well as actual laboratory practice have been assessed for compliance with the relevant clauses of P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory Testing and P903 - Policy on Estimating Measurement Uncertainty for ISO 15189 Testing Laboratories . I hereby attest that all ‘Yes’ marked compliance clauses, whether initialed or not, meet the aforementioned requirements. Any areas of noncompliance have been fully described in the Assessor Deficiency Report. Lab Name: Address: Contact: Phone: Email: Master Code: Assessment ID: Certificate: Conformity Standard: Assessment Dates: Assessment Type: L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Upload: hoangdan

Post on 23-Jun-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 1 of 16

Accredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory Testing and P903 - Policy on Estimating Measurement Uncertainty for ISO 15189 Testing Laboratories

A2LA Assessor Instructions: Please note that for all N/A indications, you must document the reason why this requirement is N/A in the comments section.IMPORTANT NOTE: An asterisk (*) in the comments section indicates that the assessor must document the specific traceable objective evidence reviewed in association with that requirement. Objective evidence information is mandatory for those clauses.

To the best of my knowledge, all laboratory document references below as well as actual laboratory practice have been assessed for compliance with the relevant clauses of P905 - A2LA Metrological Traceability Policy for ISO 15189 Laboratory Testing and P903 - Policy on Estimating Measurement Uncertainty for ISO 15189 Testing Laboratories. I hereby attest that all ‘Yes’ marked compliance clauses, whether initialed or not, meet the aforementioned requirements. Any areas of noncompliance have been fully described in the Assessor Deficiency Report.

Lab Name:Address:

Contact:Phone: Email:

Master Code: Assessment ID:Certificate: Conformity Standard:Assessment Dates: Assessment Type:Assessor(s): Assessor Signature(s):

CAcO:

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 2: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 2 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NA

(M1) A2LA accredited medical laboratories are required to ensure that:

(A) All calibrations and verifications of measuring and test equipment and reference standards be conducted by:

A calibration laboratory accredited to ISO/IEC 17025 by a mutually recognized Accreditation Body; or,

A recognized National Metrology Institute (NMI) including designated institutes. Recognition of the NMI is based on the Institute or designated institute being a signatory to the CIPM (Comité International des PoidsetMesures) MRA (Mutual Recognition Arrangement) and supporting the measurement comparison activities of the CIPM. A listing of these recognized Institutes can be found at http://www.bipm.org/en/cipm-mra/participation/signatories.html; or,

A laboratory accredited by A2LA to ISO 15189 and found to meet the M4 requirements of this document for their in-house calibrations; or

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 3: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 3 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NA A State Weights and Measures facility with a current

certificate of measurement traceability. Please see http://www.nist.gov/pml/wmd/labmetrology/lab-contacts-ac.cfmfor a copy of current certificates.

(B) When possible, and/or not in conflict with the manufacturer’s instructions for use, all reference materials shall be obtained from:

A reference material producer accredited to ISO Guide 34 in combination with ISO/IEC 17025 by a recognized Asia Pacific Laboratory Accreditation Cooperation (APLAC) signatory recognized for accrediting reference material producers; or,

A recognized National Metrology Institute (NMI) or designated institute.

Note: This requirement does not apply to routine quality control materials.

(M2) A2LA requires that:

(A) For those external calibrations and verifications performed by an A2LA Accredited calibration laboratory or a calibration laboratory accredited by an MRA partner, these must be recorded in a calibration certificate or report and must include:

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 4: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 4 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NA1. An endorsement by the recognized Accreditation Body’s

symbol (or otherwise makes reference to accredited status by a specific, recognized accreditation body);

2. An indication of the type of entity that is accredited (e.g., inclusion of “calibration laboratory” with the symbol, etc.).  A2LA-accredited organizations are also required to include their A2LA certificate number along with an indication of the type of entity with every use of the A2LA Accredited symbol (or narrative reference) (see P101, Section B2.2). and,

3. The measurement uncertainty.(B) For those external calibrations and verifications performed by an NMI, these must be recorded in a calibration certificate or report and must include:

1. An endorsement by the National Metrology Institute (NMI); and,2. The measurement uncertainty.

(C) For internal calibrations and verifications performed by the medical laboratory, those requirements outlined in requirement M4 of P605 apply.

(D)When possible for reference materials (see M1, B.), these must be recorded in a certificate meeting the requirements of ISO Guide 31 and must also include:

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 5: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 5 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NA1. an endorsement by the recognized Accreditation Body’s

symbol (or otherwise makes reference to accredited status by a specific, recognized accreditation body); and

2. An indication of the type of entity that is accredited (e.g., “reference material producer”) or endorsed by the recognized NMI.  A2LA-accredited reference material producers are also required to include their A2LA certificate number (see P101, Section B2.2).

3. It is common practice for a reference material producer to package their reference materials under a different organization’s name.  In these instances, it is possible for the reference materials to meet the A2LA Traceability Policy if the accompanying certificate includes reference to the specific, recognized accreditation body, an indication of the type of entity that is accredited and the accreditation certificate number.

(M3) All A2LA-Accredited medical laboratories must define their policy for achieving measurement traceability and also for achieving traceability for reference materials if applicable. The policy shall ensure compliance with P605.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 6: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 6 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NAa. Achievement of traceability means the measurements made by

medical devices or ancillary equipment are traceable to the SI through calibration performed by a national metrology institute (NMI) such as the National Institute for Standards and Technology (NIST), or by a calibration laboratory that is accredited by an ILAC signatory accreditation body.

This policy applies to calibrations traceable to the SI (e.g. balance calibration, temperature calibrations) and also addresses traceability of reference materials and medical devices (see Table 1 of P605).

b. If a medical laboratory uses a commercially available or other outside source of calibration for any of the Reference Standards or Working Standards within Table 1 of P605, the calibration laboratory, or reference material producer where possible, must be accredited by an ILAC signatory accreditation body and the type and range of calibration performed must be within the scope of accreditation for that accredited calibration laboratory or reference material producer.

(M4) If a medical laboratory chooses to calibrate its own Reference Standards or Working Standards within Table 1 of P605, the medical laboratory must meet the following requirements:

a) The in-house medical laboratory must maintain documented procedures for the in-house calibrations.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 7: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 7 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NAb) The in-house calibrations must be evidenced by a calibration

report, certificate, or sticker, or other suitable method; *

c) Calibration records must be retained for an appropriate, prescribed time;

*

d) The in-house medical laboratory must maintain training records for calibration personnel and these records must demonstrate the technical competence of the personnel performing the calibrations: evidence of competence includes, for example, documented training and the results of measurement audits;

*

e) The in-house medical laboratory shall be able to demonstrate traceability to national or international standards of measurement by procuring calibration services from appropriately accredited calibration labs or an NMI for the reference standards and/or purchasing reference materials from appropriately accredited reference material producers or an NMI;

f) The in-house medical laboratory must have and apply procedures for evaluating measurement uncertainty. Measurement uncertainty shall be calculated in accordance with the Guide to the Expression of Uncertainty in Measurement (GUM) for each type of calibration and records of these calculations shall be maintained. Measurement uncertainty must be taken into account when statements of compliance with specifications are made.

*

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 8: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 8 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NAG) Reference standards must be recalibrated at appropriate intervals to

ensure that the reference value is reliable. Policy and procedures for establishing and changing calibration intervals must be based on the historical behavior of the reference standard.

WHAT IS REQUIRED ON ACCREDITED CALIBRATION CERTIFICATES?

(M5) Traceability statement, statement of compliance, accreditation symbol endorsement:

A. For the purpose of demonstrating measurement traceability, calibration certificates shall, wherever applicable, indicate the traceability to national or international standards of measurement and should provide the measurement result and associated uncertainty of measurement.

B. Wherever applicable, and when suitable for customer requirements, a statement of compliance with an identified method or procedural specification can be accepted instead of measurement results and associated uncertainties.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 9: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014

Page 9 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NAC. Only calibration certificates or reports endorsed by a recognized

accreditation body’s symbol (or which otherwise makes reference to accredited status by a specific, recognized accreditation body) that is accompanied by an indication of the type of entity accredited (e.g., “calibration laboratory”, “reference material producer”) are considered to satisfy traceability requirements. By definition, such an endorsed certificate or report will contain an appropriate statement of measurement results and/or a statement of compliance with an identified metrological specification accompanied by an appropriately defined uncertainty statement and a suitable statement of traceability.

(M6) Where measurement uncertainty analysis is applicable, A2LA re-quires laboratories to calculate measurement uncertainty in accordance with the ISO “Guide to the Expression of Uncertainty in Measurement.” These uncertainties shall be reported as the expanded uncertainty with a de-fined coverage factor, k (typically k = 2) and the confidence interval (typi-cally to approximate the 95% confidence level).

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 10: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 10 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}Compliance

CommentsY N NA(M7) If a calibration certificate or test report contains a statement of the measurement result and the associated uncertainty, then the uncertainty statement must be accompanied by an explanation of the meaning of the uncertainty statement. An example of such an explanation might be the statement “Reported uncertainties represent expanded uncertainties ex-pressed at approximately the 95% confidence level using a coverage factor of k = 2”. Statements of uncertainty which do not specify at least the cover-age factor and the confidence level are incomplete and they are inadequate for the purpose of demonstrating that measurement traceability has been achieved.

(M8) Test uncertainty ratios (TURs) must be calculated using the ex-panded uncertainty of the measurement, not the “collective uncertainty of the measurement standards.”

(M9) Implicit uncertainty statements must be accompanied by words to the effect that the uncertainty ratio was calculated using the expanded measurement uncertainty. In addition the coverage factor and confidence level must be stated.

(M10) In addition to the information required in the above sections, cal-ibration reports and certificates must contain a traceability statement.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 11: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 11 of 16

P603e: Annex- Policy on Estimating Measurement Uncertainty for Medical Laboratories

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}

ComplianceComments

Y N NA

(MU-1) The medical laboratory must comply with 5.5.1.4 and 5.3.1.4 of ISO 15189 regardless of whether a method is listed in Category 1, 2, 3 or 4(MU-2 )Categorize examinations

Document identifying examinations as Category 1, Category 2, Category 3, or Category 4.

NOTE TO ASSESSORS: _____Check here if the laboratory does not have any Category 3 methods. The remainder of MU-3 will then be considered “NA”.

(MU-3) Category 3: Well-recognized quantitative and semi-quantitative methods that are governed by FDA approved test systems, including package inserts for reagents and instrument manuals for equipment.

a. Define the measurand of the method as well as clinically significant limitations and interferences.

b. Identify the components of UM.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 12: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 12 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}

ComplianceComments

Y N NA

c. Record means long-term imprecision QC data to serve as the estimate of UM.

d. Where applicable, determine Target Uncertainty for each examination.

e. Where applicable, compare and contrast the examination UM to the Target Uncertainty.

f. For each UM determined, identify the numerical significance to reflect the UM of the method.

g. Coordinate with clinicians the availability of UM information.

h. Monitor UM estimates over time.i. Documentation must exist that identifies:

The analytical method used. The substance the method is designed to measure. What is actually measured (the measurand). Any diagnostic limitations to the method. Any cross-reacting and interfering substances that

impact the clinical interpretation of the values.j. The medical laboratory must determine their uncertainty of

measurement based on the reagent package insert performance criteria as well as the instrument performance criteria, and would typically use the internal quality control data to estimate uncertainty of measurement.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 13: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 13 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}

ComplianceComments

Y N NA

NOTE TO ASSESSORS: _____Check here if the laboratory does not have any Category 4 methods. The remainder of MU-4 will then be considered “NA”.

(MU-4) Category 4: In-house developed tests that generate quantitative data.

a. If the manufacturer does not estimate uncertainty for its calibrator, the medical laboratory must do this by conducting another uncertainty of measurement for the calibrator.

b. Define the measurand of the method as well as clinically significant limitations and interferences.

c. Identify the components of UM.d. Record means long-term imprecision QC data to serve as

the estimate of UM.e. Where applicable, determine Target Uncertainty for each

examination.f. Where applicable, compare and contrast the examination

UM to the Target Uncertainty. g. For each UM determined, identify the numerical

significance to reflect the UM of the method.h. Coordinate with clinicians the availability of UM

information.i. Monitor UM estimates over time.j. Documentation must exist that identifies:

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 14: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 14 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}

ComplianceComments

Y N NA

The analytical method used. The substance the method is designed to measure. What is actually measured (the measurand). Any diagnostic limitations to the method. Any cross-reacting and interfering substances that

impact the clinical interpretation of the values.k. For these methods, UM must be estimated using available

data, published information, and/or designed experiments.

(MU-5) Record Mean Long-Term Imprecision QC Data as UM Estimate

a. For established methods collect at least 6 months’ worth of internal QC data to calculate SD or CV

b. For new methods use at least 30 data points for each level of QC using at least 2 different lots of calibrator and reagents, where applicable.

(MU-6) The calculated UM should be evaluated to determine the significance of the measurand in impacting patient care. (MU-7) The medical laboratory must establish, based on its variation data, what represents a meaningful difference in values

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 15: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 15 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}

ComplianceComments

Y N NA

(MU-8) In many cases, UM estimates can contribute to patient care. Thus, in consult with clinicians, the medical laboratory must determine these results that could significantly impact clinical interpretation and subsequent patient management. (MU-9) Monitoring

a. For each examination with an UM estimate, an on-going monitoring program must be in place.

b. Sampling Plan

c. The medical laboratory must determine when, where, and how QC data will be collected. Generally, all QC and PT data should be used to evaluate the established estimate of UM. The UM would then be adjusted based on these data

d. The medical laboratory must determine how many measurements will be part of the sample.

e. The medical laboratory must apply statistical rules to ensure that the examinations monitored are in a state of statistical control (exhibiting only common cause variation).

(MU-10) Actions to maintain currency of UM

a. The medical laboratory must define actions it will take when examinations with UM concern are not functioning within a state of statistical control

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Page 16: C105 – GENERAL CHECKLIST: A2LA POLICY ON ... · Web viewAccredited clinical laboratories are required to meet the following additional requirements as contained in P905 - A2LA Metrological

American Association for Laboratory Accreditation

C920 - General Checklist - A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing

Document Issued:March 27, 2014Page 16 of 16

Requirement

{RESERVED FOR A2LA ASSESSORS ONLY}

ComplianceComments

Y N NA

b. This includes root cause analysis and subsequent corrective action.

c. The medical laboratory must define how it will communicate and document changes in the estimate of UM when imprecision actually improves, with regard to a smaller standard deviation or coefficient of variation.

DOCUMENT REVISION HISTORY

DATE DESCRIPTION03/27/2014 Document issued.

L:\Medical-15189 – 900 Series\15189 Checklists\C920 - General Checklist: A2LA Metrological Traceability and Measurement Uncertainty for ISO 15189 Clinical Laboratory Testing