c40 manual

112
I Table of Contents Chapter 1 Overview......................................................................................................... 1 1.1 Safety Instructions ............................................................................................... 1 1.1.1 Symbols ..................................................................................................... 1 1.1.2 Labels ........................................................................................................ 3 1.1.3 Biohazard Considerations .......................................................................... 3 1.1.4 Electrical Safety ......................................................................................... 4 1.1.5 Electromagnetic compatibility .................................................................... 6 1.2 Components......................................................................................................... 9 1.2.1 Appearance................................................................................................ 9 1.2.2 Control Panel ............................................................................................11 1.2.3 Probes ......................................................................................................11 1.3 Equipment Installation and Debugging .............................................................. 12 1.3.1 Power Connection.................................................................................... 12 1.3.2 Probe Connection .................................................................................... 12 1.3.3 Probe Bracket, Cable Hook and Gel Cup Installation .............................. 13 1.3.4 Installation Environment ........................................................................... 14 Chapter 2 System Controls and Display Functions ................................................... 15 2.1 System Controls................................................................................................. 15 2.1.1 Routine functional keys ............................................................................ 15 2.1.2 System functional keys ............................................................................ 17 2.1.3 Image-mode keys .................................................................................... 17 2.1.4 Image functional keys .............................................................................. 18 2.1.5 Body Markers and Text keys .................................................................... 19 2.1.6 Record functional keys............................................................................. 20 2.1.7 Other functional keys ............................................................................... 21 2.2 Display function.................................................................................................. 22 2.2.1 Screen Interface ...................................................................................... 22 2.2.2 Monitor adjustment .................................................................................. 23 Chapter 3 Instrument Operating .................................................................................. 24 3.1 Precautions ........................................................................................................ 24 3.2 Turning the System on ....................................................................................... 24 3.3 System Settings ................................................................................................. 25 3.3.1 General Settings ...................................................................................... 26 3.3.2 Measurement Settings ............................................................................. 29 3.3.3 Setting Image Processing ........................................................................ 31

Upload: raul-apolinar

Post on 07-Apr-2015

1.543 views

Category:

Documents


153 download

TRANSCRIPT

Page 1: C40 Manual

I

Table of Contents Chapter 1 Overview......................................................................................................... 1

1.1 Safety Instructions ............................................................................................... 1

1.1.1 Symbols ..................................................................................................... 1

1.1.2 Labels ........................................................................................................ 3

1.1.3 Biohazard Considerations.......................................................................... 3

1.1.4 Electrical Safety ......................................................................................... 4

1.1.5 Electromagnetic compatibility .................................................................... 6

1.2 Components......................................................................................................... 9

1.2.1 Appearance................................................................................................ 9

1.2.2 Control Panel ............................................................................................11

1.2.3 Probes ......................................................................................................11

1.3 Equipment Installation and Debugging .............................................................. 12

1.3.1 Power Connection.................................................................................... 12

1.3.2 Probe Connection .................................................................................... 12

1.3.3 Probe Bracket, Cable Hook and Gel Cup Installation .............................. 13

1.3.4 Installation Environment........................................................................... 14

Chapter 2 System Controls and Display Functions ................................................... 15

2.1 System Controls................................................................................................. 15

2.1.1 Routine functional keys............................................................................ 15

2.1.2 System functional keys ............................................................................ 17

2.1.3 Image-mode keys .................................................................................... 17

2.1.4 Image functional keys .............................................................................. 18

2.1.5 Body Markers and Text keys.................................................................... 19

2.1.6 Record functional keys............................................................................. 20

2.1.7 Other functional keys ............................................................................... 21

2.2 Display function.................................................................................................. 22

2.2.1 Screen Interface ...................................................................................... 22

2.2.2 Monitor adjustment .................................................................................. 23

Chapter 3 Instrument Operating .................................................................................. 24

3.1 Precautions ........................................................................................................ 24

3.2 Turning the System on ....................................................................................... 24

3.3 System Settings ................................................................................................. 25

3.3.1 General Settings ...................................................................................... 26

3.3.2 Measurement Settings ............................................................................. 29

3.3.3 Setting Image Processing ........................................................................ 31

Page 2: C40 Manual

II

3.3.4 Version Information.................................................................................. 32

3.4 Selecting Probe and Frequency......................................................................... 32

3.4.1 Selecting Probe ....................................................................................... 32

3.4.2 Selecting Frequency ................................................................................ 32

3.5 Selecting Examination Type............................................................................... 33

3.6 New Patients...................................................................................................... 34

3.7 Switch Image Mode ........................................................................................... 35

3.7.1 B Mode .................................................................................................... 35

3.7.2 2B Mode .................................................................................................. 36

3.7.3 4B Mode .................................................................................................. 38

3.7.4 B/M Mode ................................................................................................ 39

3.8 Adjusting Image Parameter................................................................................ 41

3.8.1 Adjusting image brightness and contrast ................................................. 41

3.8.2 Gain Adjusting.......................................................................................... 42

3.8.3 Changing the Dynamic Range ................................................................. 42

3.8.4 Frame Correlation.................................................................................... 43

3.8.5 Edge Enhancement and Image Smoothing ............................................. 43

3.8.6 Gray Scale ............................................................................................... 44

3.8.7 Changing the Density............................................................................... 44

3.8.8 Adjusting Focus ....................................................................................... 45

3.8.9 Adjusting imaging depth........................................................................... 45

3.8.10 Adjusting image width ............................................................................ 45

3.8.11 Image Flip .............................................................................................. 46

3.8.12 Image Zoom........................................................................................... 47

3.8.13 Changing the M Sweep Speed .............................................................. 47

3.8.14 Harmonic Imaging.................................................................................. 47

3.8.15 Changing Power .................................................................................... 48

3.9 Annotation and Arrows ....................................................................................... 48

3.9.1 Annotation................................................................................................ 48

3.9.2 Arrows...................................................................................................... 49

3.10 Body Markers................................................................................................... 50

3.11 Biopsy Guideline .............................................................................................. 51

3.12 CINE ................................................................................................................ 53

3.13 Image Managing and Printing .......................................................................... 54

3.13.1 Image Managing .................................................................................... 54

3.13.2 Printing................................................................................................... 56

Chapter 4 Measurements and Reports........................................................................ 57

4.1 Overview of measurement ................................................................................. 57

Page 3: C40 Manual

III

4.1.1 Special Measurement Menu .................................................................... 57

4.1.2 Measurement Results .............................................................................. 58

4.1.3 Measurement Report ............................................................................... 58

4.2 General measurement — B-mode ..................................................................... 60

4.2.1 Distance Measurement ............................................................................ 60

4.2.2 Area/CIRC Measurement......................................................................... 63

4.2.3 Volume Measurement .............................................................................. 65

4.2.4 Angle Measurement ................................................................................. 68

4.2.5 Histogram Measurement.......................................................................... 69

4.3 General measurement — M-mode..................................................................... 71

4.3.1 Distance................................................................................................... 71

4.3.2 Time......................................................................................................... 72

4.3.3 Slope........................................................................................................ 74

4.3.4 Heart rate................................................................................................. 74

4.4 Gynecology Measurements ............................................................................... 75

4.4.1 Uterus Measurement ............................................................................... 76

4.4.2 Cervix Measurement................................................................................ 76

4.4.3 Ovary Measurement ................................................................................ 77

4.4.4 Follicle Measurement............................................................................... 77

4.4.5 Gynecologic Report ................................................................................. 78

4.5 Obstetrics Measurements .................................................................................. 79

4.5.1 Obstetrics Input........................................................................................ 79

4.5.2 Fetus Selection ........................................................................................ 80

4.5.3 Obstetrics Measurements ........................................................................ 81

4.5.4 Amniotic fluid index .................................................................................. 84

4.5.5 Fetal heart rate......................................................................................... 85

4.5.6 Neck Transparency .................................................................................. 85

4.5.7 Fetal physiology score ............................................................................. 85

4.5.8 Obstetrics Report ..................................................................................... 87

4.6 Orthopedics Measurements ............................................................................... 88

4.6.1 Orthopedics Report.................................................................................. 89

4.7 Urology Measurements ...................................................................................... 90

4.7.1 PSAD Measurement ................................................................................ 91

4.7.2 Prostate Volume Measurement................................................................ 91

4.7.3 Residual Urine Volume Measurement...................................................... 92

4.7.4 Urology Report......................................................................................... 92

4.8 Small Part Measurements.................................................................................. 93

4.8.1 Small Parts Measurement........................................................................ 94

Page 4: C40 Manual

IV

4.8.2 Small Parts Measurement Report ............................................................ 94

Chapter 5 Cleaning and Maintenance ......................................................................... 96

5.1 Cleaning............................................................................................................. 96

5.1.1 Cleaning the Main Unit............................................................................. 96

5.1.2 Cleaning the Probe .................................................................................. 97

5.2 Maintenance ...................................................................................................... 98

5.2.1 Main Unit Maintenance ............................................................................ 99

5.2.2 Caring Probe.......................................................................................... 101

5.2.3 Caring Coupling Agent ........................................................................... 102

Chapter 6 Technical Description................................................................................ 103

6.1 Technical Data ................................................................................................. 103

6.2 System Requirements...................................................................................... 106

6.2.1 Power supply Requirements .................................................................. 106

6.2.2 Connection with Peripheral Devices ...................................................... 106

6.2.3 Operating Environment Requirements................................................... 106

6.3 Transport and Storage ..................................................................................... 107

6.3.1 Requirement for Transport and Storage Environment............................ 107

6.3.2 System Transport................................................................................... 107

6.3.3 Probe Transport ..................................................................................... 107

6.4 Disposal of Packing Materials .......................................................................... 107

6.5 Statement on Consumables ............................................................................. 108

Page 5: C40 Manual

1

Chapter 1 Overview

1.1 Safety Instructions

Do not operate the ultrasound imaging system until you fully understand the safety

considerations and procedures presented in this manual.

1.1.1 Symbols

Table1-1 is provided for your identification of important symbols located in labels on the

ultrasound imaging system and probes.

Table 1-1 Symbols

Symbol Description

Type B applied part

Protected against the effects of temporary immersion

in water

Caution: Risk of electric shock. Do not open.

Caution: Refer to user’s manual.

Monitor contrast control

Monitor brightness control

Protective Grounding

Page 6: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

2

Equipotential grounding connection

USB interface

Video output

On (mains power supply)

OFF (mains power supply)

Alternating Current (AC)

Date of manufacture

Manufacturer

Indicates separate collection for electrical and electronic

equipment in compliance with the Waste Electrical and

Electronic Equipment (WEEE) Directive

Trademark of Landwind

Page 7: C40 Manual

Chapter 1 Overview

3

1.1.2 Labels

Figure 1-1 Schematic Diagram of Label Position

Figure 1-2 Identification Label

1.1.3 Biohazard Considerations

When using the system, please pay attention to the following biosafety-related points:

Caution: For neonatal head imaging, Landwind recommends that you exercise

special care during neonatal cephalic scanning to avoid possible damage to the

posterior region of the eye. The ultrasound energy emitted by the probe easily

Page 8: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

4

penetrates the fontanels of the infant.

Caution: Landwind makes every effort to manufacture safe and effective probes.

You must take all necessary precautions to eliminate the possibility of exposing

patients, operators, or third parties to hazardous or infectious materials. These

precautions should be considered in the use of any application that may indicate the

need for such care, and during endocavity scanning; during biopsy or puncture

procedures; or hen scanning patients with open wounds.

Caution: To eliminate the possibility of exposing patients, operators, or third parties

to hazardous or infectious materials, handle dangerous or infectious materials all

along according to the medical specifications for handling biohazard castoffs.

Caution: The assessment of biological effects of diagnostic ultrasound on humans is

a subject of ongoing scientific research. This system and all diagnostic ultrasound

procedures should be used for valid reasons, for the shortest period of time, and at

the lowest mechanical and thermal indices necessary to produce clinically acceptable

images.

Caution: To ensure the safety of examined patients, do not scan the same part of an

examined patient consecutively for a long period of time.

Caution: When making ultrasonic scanning, the instrument shall be used with

ultrasound coupling agent, and no biologic compatibility problem is found. However,

to ensure safety, only qualified ultrasound coupling agent in compliance with related

standard shall be used.

Caution: Do not use the system if irregular and unusual repeated video refreshes

occur. Discontinuous scanning frequency is a hardware fault, please remove it before

use.

1.1.4 Electrical Safety

The system is designed in line with the safety standard of EN 60601-1, and use Type B

isolation for the patient part. To ensure safety to the greatest extent, please pay attention

Page 9: C40 Manual

Chapter 1 Overview

5

to the following Cautions:

Caution: Improper ground of the instrument may result in the danger of electric

shock. The only means to prevent electric shock is to use three-phase cable and

plug. The power cable shall be inserted into a grounded socket, and the ground

wire should not be moved away or removed.

Caution: Only cable hooks provided by Landwind can be used.

Caution: Do not attempt to modify the instrument. If it is really necessary to do so,

please notify us or our agent.

Caution: Operators of the instrument shall be qualified service-trained

professionals or shall use the instrument under the supervision of qualified

professionals.

Caution: To make electrical or mechanical connection between this instrument and

the equipment of other manufactures, confirm that it complied with the

IEC60601-1-1:1995 standard, and recheck the leakage current and other safety

indices of the entire set of system, to avoid possible hazard arising from leakage

current overlapping.

Caution:Upon the expiration of the service life of the system, please notify your

dealer or Landwind to handle properly.

Caution: To avoid electrical shock and damage to the ultrasound system, turn off

and unplug the equipment from the AC power outlet before cleaning and

disinfecting.

Caution: Do not open the enclosure of this unit, as some components inside it

carry high voltages that may cause hazards. If it is really necessary to do so, please

contact your dealer or Landwind to handle properly.

Caution: Do not use the instruments in an environment with combustible gases or

anesthetic to avoid explosions.

Page 10: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

6

Caution: Do not pour any fluid onto the system surfaces, as fluid seepage into the

electrical circuitry may cause excessive leakage current or system failure.

Caution: This unit must be serviced once every 12 months to ensure its safety and

functionality. Electric safety test must be performed at periodic intervals according

to the local safety regulations or as needed.

Caution: To avoid the possibility of static shock and damange to the instrument,

avoid the use of aerosol spray cleaners on the monitor screens.

Caution: The power supply of the instrument can only be connected to a special

power inteface. Never connect an unspecified device to the port of this instrument.

Pay attention to correct connection of the grounding wire; otherwise, electric shock

may easily occur.

Caution: Do not use the system in damp or wet environments. Keep it away from

inflammables and explosives.

Caution: First check this unit and its parts every time before you use it. Timely

replace any components with sign of damage.

Caution: Use the probe with care. Do not use damaged probes. Do not use probes

with cracks or surface crimps, since such probes might have their insulation

damaged ,which easily causes electric shock to patients.

1.1.5 Electromagnetic compatibility

Portable Ultrasound Diagnostic Scanner has passed the tests that confirmed its

compliance with electromagnetic compatibility (EMC) requirements of IEC

60601-1-2:2001+A1:2004. It is CISPR 11: Group 1 Class А.

Caution: Special EMC requirements are imposed on medical equipment. The

following should be taken into account when installing and operating such

equipment.

Strong electromagnetic fields induced over the air or the wires by portable and mobile

communication devices or powerful sources of radio-frequency emission can cause

Page 11: C40 Manual

Chapter 1 Overview

7

failures in Portable Ultrasound Diagnostic Scanner operation. Such failures are

manifested in impaired quality of ultrasound images produced by the system or in

distortion thereof, in distortion of measurement results, instability of the system operation

and other abnormal manifestations.

In the event that such disturbance is detected when operating the system, its source shall

be localized and the following measures shall be taken.

— To detect the disturbance source, disconnect the nearby devices and turn them on

again.

— Having detected the interference-emitting device, relocate or re-orient it.

— Increase the separation distance between the interference emitting device and

Portable Ultrasound Diagnostic Scanner .

— Do not use the equipment with operating frequencies close to frequencies of Portable

Ultrasound Diagnostic Scanner.

— The personnel in charge of monitoring the medical devices shall be trained to identify

potential problems associated with electromagnetic emission.

— Use of cellular phones and computers shall be restricted in the areas where

interference-sensitive devices are located.

— Electromagnetic emission shall be removed or eliminated technically, for example,

through shielding.

Caution: Use of probes, cables and accessories other than stated in the operational

documentation can entail RF emission enhancement or degradation of Portable

Ultrasound Diagnostic Scanner stability.

Declaration:

All cables, probes and other accessories, whose compliance with requirements of IEC

60601-1-2:2001+A1:2006 36.201 – EMISSION and 36.202 –IMMUNITY is declared by

the manufacturer.

Page 12: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

8

Preferable operation conditions and electromagnetic compatibility levels of Portable

Ultrasound Diagnostic Scanner are given in Tables 1-2 ,1-3.

The system is designed to be operated in the below electromagnetic environment.

Table 1-2 EMISSION

EMISSION(EN 60601-1-2:2001+A1:2006)

Test Type Basic Standard

Mains terminal disturbance voltage

Radiated disturbance

EN 55011:1998+A1:1999+A2:2002

Harmonic current emissions EN 61000-3-2:2006

Voltage fluctuations & flicker EN 61000-3-3:1995+A1:2001+A2:2005

Table 1-3 IMMUNITY

IMMUNITY(EN 60601-1-2:2001+A1:2006)

Test Type Basic Standard

Electrostatic discharge immunity EN 61000-4-2:1995+A1:1998+A2:2001

Radiated, radio frequency

electromagnetic field immunity EN 61000-4-3:2006

Electrical fast transient/burst EN 61000 4-4:2004

Surge immunity EN 61000-4-5:2006

Page 13: C40 Manual

Chapter 1 Overview

9

Immunity to conducted disturbances

Induced by RF fields EN 61000-4-6:1996+A1:2001

Voltage dips and short interruptions

immunity EN 61000-4-11:2004

Power frequency magnetic field immunity EN61000-4-8:1993+A1:2001

1.2 Components

1.2.1 Appearance

1 Cable hook 2. Coupling agent case

3. Probe bracket 4. Keyboard

5. Contrast/brightness adjustment 6. Monitor

Figure 1-3 Front Left View of the Ultrasonic Diagnosis System

Page 14: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

10

1. Output Video port 2. USB port 3. Power switch

4. RS-232-C serial port 5. Equipotential grounding connection 6. Fuse case

7. VGA terminal 8. Power input jack 9. LAN port

10. Remote terminal 11. Probe port

Figure 1-4 Rear View of the Ultrasonic Diagnosis system

Page 15: C40 Manual

Chapter 1 Overview

11

1.2.2 Control Panel

This section describes the overview of the control panel of the instrument.

Figure 1-5 Overview of Operation Panel

1.2.3 Probes

Multi-frequency capability is available for all imaging probes. You can connect two probes

to the ultrasound imaging system, with one probe being the active probe. The

configuration of probes depends on your purchase contract.

Model Type Description

35C60H convex probe center frequency: 3.5 MHz, R=60mm

65C13H convex probe center frequency: 6.5 MHz, R=13mm

75L40H linear probe center frequency: 7.5 MHz, L=40mm

Page 16: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

12

1.3 Equipment Installation and Debugging

1.3.1 Power Connection

This section describes how to connect the 100-240V~ 50/60Hz power supply for the unit.

Please ensure a stable power supply for the equipment. The power interface card is

located on the back of the instrument. Check the power supply of the unit to ensure that it

is within the specified range. Connect one end of the power cable to the power input jack

(8) (see Figure 1-4), and the other to the power supply.

Caution: Do not connect any device to the external power outlet on your own.

Caution: Confirm the power voltage is the same as the nominal voltage of the

instrument before connection.

Caution: This instrument adopts two fuses. Do not change fuses on your own;

otherwise, it may cause the danger of electric shock.

This equipment includes two other fuses, which are only used by professional

equipment maintenance personnel.

1.3.2 Probe Connection

Operating procedure;

1. Turn the probe’s latch to the open position and plug the probe into the slot

(Figure 1-6).

2. Turn the latch clockwise to the locked position.

Figure 1-6 Connecting a probe

Page 17: C40 Manual

Chapter 1 Overview

13

Caution: Before probe connection, check whether latch is in open position. Do not

insert or pull out the connector in switch-on status to avoid damage to the probe.

Once the probe is connected the main unit, do not insert or pull out it at will.

Caution: Never force a probe into place. If you feel any resistance, reposition the

probe connector and try again.

Caution: Before disconnecting a probe, be sure to open the connector by turning the

latch to the open position.

1.3.3 Probe Bracket, Cable Hook and Gel Cup Installation

1. Probe Bracket and Gel Cup Installation.

The installation of Probe bracket an gel cup adopts the method of directly inserting

the position bore and locked, shown as follow:

Figure 1-7 Probe Bracket and Gel Cup Installation.

2. Cable Hook Installation

The installation of Cable Hook adopts the method of revolved to lock , shown as

follow:

Page 18: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

14

Figure 1-8 Cable Hook Installation

1.3.4 Installation Environment

The instrument shall be installed in a stable place far from electric fields, devices with

strong magnetic fields and high-voltage devices. Avoid direct sunlight on the monitor

screen and keep a dark indoor environment to facilitate in image observation.

Furthermore, keep a ventilated and dustproof environment.

Page 19: C40 Manual

15

Chapter 2 System Controls and Display Functions

2.1 System Controls

2.1.1 Routine functional keys

Trackball

It is the main tool of screen interface and is for selection and location. Trackball can be

used in conjunction with other keys, when the machine is in different states, the trackball

has different functions. Under normal circumstances, trackball controls the cursor position.

Set

It is a Multi-function key, used in conjunction with the trackball. Its function changes in the

different state automatically, such as setting the cursor location, positioning location map,

noting the location or selecting menu item, increasing the parameter value.

Exit

Exit the current operation mode. It also used to switch the endpoints in DiST-LINE and Ellipse measurement.

FREEZE

Freeze or unfreeze image

Page 20: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

16

Gain

It is a Multi-function key which can adjust the total gain of the system, switch the input

method and measurement items and turn page to view the measurement results.

Adjust the time gain compensation

Menu

Open or close the menu of real-time image.

Meas

Enter the conventional measurement function.

Probe

Select and switch to probes.

Page 21: C40 Manual

Chapter 2 System Controls and Display Functions

17

2.1.2 System functional keys

New Patient

Press this key to open the new patient dialog box to input the information of new patient.

Press this key again to close the dialog box.

Exam

Press this key to open exam dialog box to preset examination subjects. Press this key

again to close the dialog box.

Setup

Enter or exit the main menu system settings dialog box.

2.1.3 Image-mode keys

B Mode

Switch to single B-mode.

2B Mode

Switch to 2B -mode.

Page 22: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

18

4B Mode

Switch to 4B-mode.

B/M Mode

Switch to B/M-mode or single M-mode.

2.1.4 Image functional keys

Zoom

Start or close B-mode image zoom function

Depth up

Increase scanning depth.

Depth down

Reduce scanning depth.

Focus up

Page 23: C40 Manual

Chapter 2 System Controls and Display Functions

19

Rise up focal zone position.

Focus down

Adjust down focal zone position.

Cine

Start the movie playback function.

2.1.5 Body Markers and Text keys

Mark

Press this key to enter the position-work state, and the body markers and the location of

the scanning probe can be input at the designated location of the screen. Press the key

again to exit this mode.

Text

Press this key to enter the text-work state, and input texts at the designated location of the

screen. Press the key again to exit this mode.

Arrow

Page 24: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

20

Press this key to enter the arrow-work state, make arrow mark at the designated location

of the screen. Press the key again to exit this mode.

Clear

Press this key to clear the measurement results, body marks, arrows and text messages

on the screen.

2.1.6 Record functional keys

Report

Open or close the report.

Review

Open or close the image reviewing mode.

Save

Save images.

Print

Page 25: C40 Manual

Chapter 2 System Controls and Display Functions

21

Activation the video printing function.

2.1.7 Other functional keys

Calculating key

Press this key to enter the special measurement state.

Biopsy

Enter or exit the puncture biopsy line display mode.

Backspace key

Press this key can delete a character from right to left when in text input state.

Shift

It is used to access characters on the upper right corner of the keyboard.

Caps

It used to access capital or lowercase letters.

Page 26: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

22

2.2 Display function

2.2.1 Screen Interface

1 Trademark of Landwind;

2 patient information area: Name of hospital, Name of patient, Number of medical record, exam, etc;

3 Image information area: Total gain (G), Dynamic range (D), Frame correlation (FC), Frame rate (FR),

Edge enhancement and image smoothing (EE);

4 Date & Time display area;

5 USB logo: U-disk is inserted to the system;

6 Gray bar;

7 Horizontal scale: For images in B mode, it indicates horizontal distance (the distance between

different point is 10mm), and the scale varies with the magnification ratio of images. For images in M

mode, it indicates scanning time (the scanning time between different points is 1 second), and the

scale varies with the scanning speed of the M mode.

8 Focus position indicator;

9 Vertical scale: Indicates the vertical distance of an image (the distance between different points is

10mm), and the scale varies with the magnification ration of the image;

10 TGC curve display area;

11 Current active image display depth;

12 Quick button; 13 Cine display indicator;

14 Image display area;

15 Direction Logo: L indicates probe direction, corresponding to the small protrusion on the probe

handle;

16 System working mode;

18 17 16

14

13

12

11

10

9

8

7 6

5 4 3 1 2

15

Page 27: C40 Manual

Chapter 2 System Controls and Display Functions

23

17 Imaging mode;

18 Probe information display area;

Figure 2-1 Layout of Screen Operation Interface

Note on EMC: Using this instrument in strong electromagnetic areas near radio stations or

similar units may cause visual interference upon the monitor screen. However, after

proper design and test, the instrument can withstand similar interferences without any

permanent damage.

2.2.2 Monitor adjustment

After starting the instrument, adjust the brightness and contrast of the monitor or you may

make such adjustment during check at any time. For detailed adjustment, refer to 3.8.1

section.

Page 28: C40 Manual

24

Chapter 3 Instrument Operating

3.1 Precautions

Make sure that the AC power connector plug for the ultrasound system is a three-prong

grounded plug, and ensure that the AC power in the plug is 100-240V~ 50/60Hz.

Check all power suppliers and connection cables to make sure no such damage as

scratch and breakage.

Check the probes and connecting parts to make sure no such damage as scratch and

desquamate.

Illustrations are added in appropriate places of this chapter, and please note that the

actual product shall apply if the details of illustrations differ from the actual product.

3.2 Turning the System on

Make sure that one end of the power cable is inserted into the power input jack on the

power interface card (refer to Figure 1-4) of the instrument, and the other end is inserted

into the AC power jack. Turn on the main power switch (for its position, refer to Figure 1-4.

Press to turn on, and press to turn off), the system startup screen appears. The

system shows the main screen, displaying B-mode images as default (as shown in Figure

3-12).

Please check whether the instrument is started normally, and check the images and

operation panel to make sure there is no abnormality. If no fault occurs, apply coupling

agent on the transmission surface of the probe, and make preparation for patient

examination.

Caution: Please stop operation if any fault is found, and contact your supplier for

troubleshooting.

Caution: Black boot time is about 30 seconds to respond, which is a normal

phenomenon, please be patient.

Page 29: C40 Manual

Chapter 3 Instrument Operations

25

Caution: To ensure long-term stable and reliable work of the instrument, do not turn

on the instrument immediately after power-off. Please wait for at least one minute.

3.3 System Settings

System Settings are used to set the power-on operating environment, status, and

configuration parameters of different examination modes. Preset values are saved in the

memory of the machine so that it will not get lost upon power failure and will ensure the

system runs at the expected state each time when powered on. Reset default setting can

restore the parameters in factory.

Press the 【Setup】 key on the keyboard to access the system settings. The system

settings dialog is displayed on the screen, and the General Settings menu appears by

default, as shown in Figure 3-1.

Setting items in the dialog include General Settings, Measurement Settings, Image

Processing Settings and Version information.

Figure 3-1 main menu of System Setup

Page 30: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

26

3.3.1 General Settings

General Settings include Date And Time, Time Format, TGC Autohide, Default Port, Cine

Playback Speed, Hospital Name, System Language, System Environment, Network

Status, Exam Settings, Report mode and Biopsy Settings.

3.3.1.1 Setting Date And Time

Move cursor to Date And Time item, and then press the 【Set】 key to click it, then the

time setting screen appears. After adjusting system time, press to save setting

and press (or press【Setup】key) to exit the system setting.

Figure 3-2 Date and Time Settings

3.3.1.2 Setting Time Format

Move cursor to the Time Format item, and then press the 【Set】key to click it, then move

cursor to your favorite date format and press the 【Set】key to click it. Press to

save setting and press (or press【Setup】key) to exit the system setting.

Figure 3-3 Setting Time Format

3.3.1.3 Setting TGC Autohide

The system supports two modes of displaying TGC curve: AutoHide and NoneAutoHide.

Make a choice according to yourself. The detail is the same as 3.3.1.2.

Page 31: C40 Manual

Chapter 3 Instrument Operations

27

3.3.1.4 Setting Default Port

Choose probe port I or II as the startup probe by default by yourself. The detail is the same

as 3.3.1.2.

3.3.1.5 Setting Cine Playback Speed

The system supports 1/1, 1/2, 1/4, 1/8 four kinds of speeds to choose, the detail is the

same as 3.3.1.2.

3.3.1.6 Setting Hospital Name

Move cursor to the hospital name setting item, and then press the【Set】key to click it.

Input a name in the hospital name text box. English is the default language. After inputting

hospital name, press to save setting and press (or press【Setup】key)

to exit the system setting.

Figure 3-4 Setting Hospital Name

3.3.1.7 Choosing System Language

The system supports two language modes: English and Chinese. The detail is the same

as 3.3.1.2 section.

3.3.1.8 Modifying Network Status

Move cursor to the network setting item, and then press the 【Set】key to click it. The

screen will appear shown as Figure 3-5. Setting network according to your actual need,

and then press to save setting and press (or press【Setup】key) to

exit the system setting.

Page 32: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

28

Figure 3-5 Network Status

3.3.1.9 Exam Settings

The system supports the exams including Abdomen, Gynecology, Obstetrical, Small Part,

Orthopaedy, Urology and Emergency. After choosing, the system will initialize the current

selected examination type which is the default exam when you restart the system.

3.3.1.10 Setting Report Mode

The system supports two report modes: Conventional Report and Specialist Report. The

detail is the same as 3.3.1.2.

3.3.1.11 Biopsy Settings

In this menu, you can set the biopsy direction, start point, biopsy angle and depth prompt.

The detail is the same as 3.3.1.2.

Figure 3-6 Biopsy Settings

Page 33: C40 Manual

Chapter 3 Instrument Operations

29

3.3.2 Measurement Settings

Measurement Settings includes Meas Setup, EFW Formula, HR Cycles and GA

Referance (see Figure 3-7).

Operation procedures:

1. Move the cursor to the (▼) and press the 【Set】 key to choose the default

measurement method.

2. After choosing, move the cursor to the button to save setting.

3. Repeat steps 1-2 if you want to amend other items. Press the button or

【Setup】key if you want to exit measurement menu.

Table 3-1 gives the conventional measurement items and its measurement methods.

Table 3-2 lists the formula for calculating fetal weight and gestational age reference.

Figure 3-7 the Menu of Measurement Settings

Page 34: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

30

Table 3-1 General measurement setting

Measurement item Measurement method

Default Area/Circ Method Ellipse, Trace

Default Distance Method Line, Trace

Default Distance Unit cm, mm

Default Weight Unit Gram, Ounce, Pound

Table 3-2 EFW Formula and GA Referance

EFW and gestational age

measurement item Measurement formula

EFW Formula Tokyo, Osaka, Hadlock1, Hadlock2, Hadlock3,

Hadlock4, Shepard, Merz1, Hansmann, Campbell

GS Tokyo, Hellman, Rempen, Hansmann, China

CRL Tokyo, Hadlock, Nelson, Robinson, Rempen,

Hansmann, Osaka, China

BPD Tokyo, Hadlock, Jeanty, Kurtz, Merz, Hansmann,

Rempen, Chitty(0-I), Osaka, China

FL Tokyo, Hadlock, Jeanty, Hohler, Merz, Hansmann,

Chitty, Osaka, China

HUM Jeanty, Merz

HC Hadlock, Merz, Hansmann, Chitty(pltd)

AC Hadlock, Merz

FTA Osaka

Page 35: C40 Manual

Chapter 3 Instrument Operations

31

FT Mercer

OFD Hansmann

CER Hill

3.3.3 Setting Image Processing

Image Processing Settings can preset IP values. The items include Dynamic Range

(30-150dB, with a step of 6dB), Frame correlation (0-7 adjustable), M Scan Speed (0-3

adjustable), Edge Enhance (0-6 adjustable), Line Density (0 -2), witch is shown in Figure

3-8.

Figure 3-8 Image Processing Settings

The detail is as follows:

Press button, then you can input IP name, dynamic range, frame correlation,

M scan speed, edge enhance and line density. Press button to save a new

set of IP.

Page 36: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

32

Select any existing IP in the combo box, and press button, then input the

new name, and press button to save the setting.

Select any existing IP in the combo box, and press button to delete a set of

existing IP value.

Select any existing IP in the combo box, and press button to apply a set of

existing IP value. At this point, both the image area of information and real-time

images will change.

Press button or【Setup】key to exit System Setup.

3.3.4 Version Information

To see the current software and Firmware version.

3.4 Selecting Probe and Frequency

3.4.1 Selecting Probe

Two ports and names of corresponding probes info is shown in the probe info area. The

current port is highlight, and its work frequency is shown.

Press 【Probe】to switch probes connected to the system. When first pressing it, switch

the current port to another port. Every time pressing it again, switch the current port to

another port. If another port doesn’t have a probe, it will not switch.

When the system is connected to two probes, select probe port activated when start up by

system setting. Please refer to 3.3.1.4 section.

3.4.2 Selecting Frequency

Operating procedures is as follows:

1. Under real-time status, press the 【Menu】key and the main menu will be displayed in

nether left of the screen.

2. Move the cursor or rotate【Gain】knob to select the PROBE FREQ menu item.

Page 37: C40 Manual

Chapter 3 Instrument Operations

33

3. Press 【Set】key or 【Gain】knob to choose this menu.

4. Rotate【Gain】 knob to change probe frequency, and press 【Set】 key or 【Gain】

knob to finish it.

5. Press Exit item to exit the main menu.

Figure 3-9 the main menu of B/M mode

3.5 Selecting Examination Type

Press 【Exam】key to enter exam menu and select examination type. The examination

types of the system include Abdomen, Gynecology, Obstetrical, Small Part, Emergency,

Orthopaedy and Urology. After choosing, the name of the current selected examination

type will be displayed on the main screen.

Page 38: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

34

Figure 3-10 the menu of the exam

3.6 New Patients

Press 【New Patient】key to enter new patients menu. Input name, sex, age and other

personal information and automatically generate ID. Patient’s name is the required

information.

Page 39: C40 Manual

Chapter 3 Instrument Operations

35

Figure 3-11 New Patient

3.7 Switch Image Mode

The system provides five scanning modes: B, 2B, 4B, B/M and M. Press the image-mode

keys to switch over under real-time status.

In the following, the convex probe will be taken as an example to describe the individual

mode.

3.7.1 B Mode

Press 【B】key to activate B Mode, and the image area will display B Mode image, default

mode of the system.

Page 40: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

36

Figure 3-12 B Mode

3.7.2 2B Mode

Press 【2B】key to switch to 2B image mode.

The operating procedure is as follows:

1. Press the 【2B】key, and a B-mode image appears on the left of the screen (See

Figure 3-13(a)).

2. Press the【2B】key again to activate the right of the screen, and the screen displays

two B-mode images in parallel. Where, the left one is a frozen image, and the right

one is a real-time image. “L” symbol appears on the upper part of a real-time image

(See Figure 3-13(b)).

3. You can continuously press the【2B】key to switch between these two images.

4. Under real-time status, press【Depth Up】or【Depth Down】key to adjust the depth

of real-time images;

Page 41: C40 Manual

Chapter 3 Instrument Operations

37

5. Press the【FREEZE】key to freeze an image, and press the【FREEZE】key again to

unfreeze the image;

6. In real-time mode, press the mode switching key in a mode instead of the 2B mode

can switch to another mode.

(a)

(b)

Figure 3-13 2B Mode

Page 42: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

38

3.7.3 4B Mode

Under real-time status, press the 【4B】key to switch to 4B-mode images.

Figure 3-14 4B Mode

Operating procedure:

1. Press the 【4B】mode key continuously to display four B-mode images (as shown in

Figure 3-14). Only one image is in real time at the same time.

2. Press the 【4B】 key for four times, the screen will display four B-mode images in

parallel, including one real-time image, and other three frozen images. “L” symbol

appears on the upper part of the real-time image.

3. In the 4B mode, press the【4B】key repeatedly to switch over between the four images.

4. Press the 【FREEZE】 key to freeze an image, and press the【FREEZE】key again

to unfreeze the image.

Page 43: C40 Manual

Chapter 3 Instrument Operations

39

5. In real-time mode, press the mode switching key in a mode instead of the 4B mode

can switch to another mode.

3.7.4 B/M Mode

Under real-time status, press the 【B/M】key to switch to BM-mode images.

Operating procedure:

1. Pressing the 【B/M】key continually, up B down M, left B right M and single M modes

will appear(shown as figure 3-15). A dotted line (i.e., sampling line) appears in the

B-mode imaging area. You may use the trackball to move the sampling line; the

M-mode displays the movement of the tissue on the sampling line in this position.

2 Press the 【FREEZE】 key to freeze the B-mode and M-mode images at the same

time, and press the 【FREEZE】key once again to unfreeze the B-mode and M-mode

images.

3 The system provides four scanning speeds (0-3). Press 【Menu】key to enter the main

menu and setting it in the M Mode scanning speed menu (Refer to 3.8.13 section).

4 In the real-time status, press the mode switching key instead of the B/M mode can

switch to another mode.

Page 44: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

40

(a) Up B Down M Mode

(b) Left B Right M Mode

Page 45: C40 Manual

Chapter 3 Instrument Operations

41

Rotate the brightness dial to appropriate brightness. Too bright result in weak

contrast, and too dark result in misty image.

Rotate the contrast dial to appropriate contrast. For clear profile, the contrast

shall be bigger; but for features of objects ,it shall be small.

(c) Full M Mode

Figure 3-15 B/M Mode

3.8 Adjusting Image Parameter

3.8.1 Adjusting image brightness and contrast

3.8.1.1 Adjusting image brightness

3.8.1.2 Adjusting image contrast

Page 46: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

42

3.8.2 Gain Adjusting

3.8.2.1 Adjusting the Total Gain

Adjust the 【Gain】 knob to control the total gain of the entire image to obtain satisfying

images. Clockwise rotation may increase the gain, and anticlockwise rotation may

decrease the gain. You cannot adjust gain in freezing status.

The total gain ranges from 0 dB to 100 dB, with the minimum adjustable step of 5. The

gain value directly appears on the upper part of the screen. When checking 2B or 4B

Mode with many images, it will only display the value of activated image.

3.8.2.2 Adjusting the TGC

To compensate for weak signals or over-bright signals at various depths, you can adjust

Time Gain Compensation (TGC) using the TGC slide controls. The TGC provides six slide

controls for adjusting image quality under real-time status.

Use the system presets to turn on the display of the TGC curve. The curve provides an

on-screen representation of the TGC settings.(Refer to 3.3.1.3 section)

In B/M-mode, TGC adjustments affect both the B-mode image and the M-mode sweep.

The TGC slide controls have no effect during freezing.

To adjust the TGC slide controls: Move a slide control to the right to increase a setting or

to the left to decrease a setting.

3.8.3 Changing the Dynamic Range

You can adjust the overall contrast resolution of a B-mode image and the overall contrast

resolution of an M-mode sweep. For B-mode and M-mode, the dynamic range values are

30 dB to 150 dB with a step of 6dB. The dynamic range value displays in the imaging

parameters area at the top of the screen. In 2B mode or 4B mode with more than one

image, the value displays only for the active image.

Operating procedure:

Page 47: C40 Manual

Chapter 3 Instrument Operations

43

1. Under real-time status, press the 【Menu】key, and the main menu appears on the

lower part of the screen.

2. Move the track ball to place the cursor on the option DYNAMIC RANGE; Press the

【Set】 key to set parameter values or use the 【Gain】 knob to adjust parameters,

and press 【Gain】 knob to choose it.

3. Adjust parameter by 【Gain】 knob.

4. Press 【Menu】key and it will disappear. Directly exit the adjustment status or move

the cursor to the EXIT option and then press 【Set】 to exit.

3.8.4 Frame Correlation

The frame correlation adjustment involves superposing and averaging B-mode images in

neighboring frames to eliminate noise from the images and produce clearer image details.

The frame correlation range is from 0 to 7, totaling 8 levels. 0 means no frame correlation,

while 7 means the largest frame correlation. The current frame correlation value is also

displayed on the menu item.

Operating procedure:

1. Under real-time status, press the 【Menu】key, and the main menu appears on the

lower part of the screen.

2 Move the track ball to place the cursor on the option FRAME CORR. Press the 【Set】

key or press 【Gain】 knob to choose it.

3 Adjust parameter by rotating【Gain】 knob.

4 Press 【Menu】key or move the cursor to the EXIT option and then press 【Set】to

close menu.

3.8.5 Edge Enhancement and Image Smoothing

Edge enhancement highlights image profile so that the border of tissue structure of an

image can be displayed more clearly, while image smoothing has the reverse effect.

Page 48: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

44

When adjusting, the number displayed on the side of the menu item is the edge

enhancement strength, which is shown as Table 3-3.

Table 3-3 Edge Enhancement and Image Smoothing

EE6 strong

EE5 medium

EE4 light Edge Enhancement

EE3 close

EE3 close

EE2 light

EE1 medium Image Smoothing

EE0 strong

The detail refers to 3.8.3 section.

3.8.6 Gray Scale

3.8.6.1 Setting Gray Map

The system supports 16 kinds of Gray Map (0-15 adjustable). It does not work when

choose OFF option.

The detail refers to 3.8.3 section.

3.8.6.2 Setting Gray Reject

The system supports 16 kinds of Gray Reject (1-16 adjustable). It does not work when

choose OFF option). The detail refers to 3.8.3 section.

3.8.7 Changing the Density

When in B Mode real-time status, press number key【3】can adjust the density, that is

consequently displaying high density, middle density and low density image. Increasing

density could improve image detail, but decrease frame rate.

Density sign differentiates high, middle and low density. H means high density, M means

Page 49: C40 Manual

Chapter 3 Instrument Operations

45

middle density, and L means low density.

3.8.8 Adjusting Focus

The system includes 64 groups of focus combinations. Each focus combination comprises

different positions and number of focuses. The focus control keys(【Focus Up】or 【Focus

Down】) and numeric keys(【5】or 【6】)can be used to adjust image focus. Under B mode

real-time status, focus adjustment brings clearer images in the target area. In B/M mode

and single M mode, the system does not support multi-focus.

Increasing the number of focus can gain higher resolution in different depth, at the same

time; the number of focus can affect frame rate of the image: the more focus ,the lower

frame rate.

For visual display, the focus display mode on the screen is displayed on the left or right of

the vertical scale line by the or symbol.

Operating procedure:

Under real-time status, press numeric keys【5】or【6】to decrease or increase the

number of focus. The range is from 1 to 4. When overstep this step, pressing 【5】

or【6】 will not change the number. B/M Mode has only one focus.

Under real-time status, press【Focus up】and【Focus down】can adjust the position

of the focus. Every kind of focus number can have 16 kinds of focus combination.

When the focus is in the highest or lowest position, press this key again, the position

of the focus will keep the same.

3.8.9 Adjusting imaging depth

Increase or decrease imaging depth by【Depth Up】and【Depth Down】.

3.8.10 Adjusting image width

The system provides images with four steps width.

Operating procedure:

Page 50: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

46

1. Under real-time status, press【Menu】to display the main menu.

2. Move the trackball or turn【Gain】to place the cursor in the IMAGE WIDTH item.

3. Press【Set】key or【Gain】key to choose this menu item.

4. Image narrowing, broaden, restore default and exit buttons will be displayed in the

bottom of the screen.

5. Press numeric keys 【3】or 【4】to make images narrowing or broadening. Press

numeric key【5】to restore image width.

6. Press numeric key 【6】to exit this adjusting mode.

3.8.11 Image Flip

Figure 3-16 Image Flip

Image flip can be adjusted on the menu, including U/D flip, L/R flip, and B/W flip.

Page 51: C40 Manual

Chapter 3 Instrument Operations

47

Operating procedure:

Under real-time status (except single M mode ), press numeric key 【1】, then the

image will flip 180° along vertical direction.

In B Mode real-time status, press numeric key 【2】, then the image will flip 180° along

level direction.

Under real-time status, press 【Menu】key to choose B/W FLIP item to achieve

black-white inversion of the image. The detail refers to 3.8.3 section.

3.8.12 Image Zoom

Single B mode images under real-time status have zoom function. Measurement,

annotation, body-mark are available in the zoomed image.

Operating procedure:

1. Pressing【Zoom】key or pressing【4】key, the zoom frame will appear.

2. Rotate 【Gain】knob to adjust the size of the zoom frame (two sizes available) so as

to magnify the image twice or three times.

3. Move the trackball to place zoom frame in the concerned area of the image.

4. Press 【Zoom】key or press【4】key to zoom selected area.

5. Press 【Zoom】key or press【4】key to exit zoom mode.

3.8.13 Changing the M Sweep Speed

The system provides four kinds of scanning speed: 0-3. 0 expresses the lowest speed,

while 3 expresses the fastest speed. The detail refers to 3.8.3 section.

3.8.14 Harmonic Imaging

The function of Harmonic Imaging is available for low-frequency probe (The system

supports the probe of 35C60H.). The detail refers to 3.8.3 section.

Page 52: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

48

3.8.15 Changing Power

The power ranges from 0 to 100, with step of 5. The detail refers to 3.8.3 section.

3.9 Annotation and Arrows

3.9.1 Annotation

Press【Text】key to active annotation function.

Operating procedure:

1. Move the cursor to the wanted position, and then press 【Set】to confirm input

position.

2. After inputting needed text, press 【Enter】key to confirm it.

3. Repeat step 1-2 if you want to input other characters continuously.

4. Press 【Text】key again to exit the annotation function, and the annotations are still

displayed on the images.

5. Press 【Clear】key to clear all annotations and exit the text function.

Page 53: C40 Manual

Chapter 3 Instrument Operations

49

Figure 3-17 Annotation Function

3.9.2 Arrows

Press【Arrow】key to activate arrows. A solid arrow is displayed in the image area ,which

can be rotated 360° along eight directions.

Operating procedure:

1. Turn 【Gain】knob to change arrow direction.

2. Move the trackball to adjust the position of the arrow.

3. Press 【Set】key to confirm entering an arrow.

4. Repeat step 1-3 if you want to continue.

5. Press 【Arrow】again to close arrow function, and the arrows are still displayed on the

images.

6. Press 【Clear】key to clear all the arrows and close the arrow function.

Figure 3-18 Arrow Function

Page 54: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

50

3.10 Body Markers

Markers are graphics that display on-screen to indicate the anatomical structure under

evaluation and to indicate the orientation of the probe to the structure. Every examination

subject will have fixed body markers.

Operating procedure:

1. Press 【Mark】key to activate body markers function. Corresponding body markers

will be displayed on the bottom of the screen.

2. Select body marker using numeric keys 【1~8】. If there are more than eight body

markers, pressing numeric key 【9】can turn to next page, while 【0】 to the previews

page. A small pictograph like probe will be displayed on the screen.

(a)

Page 55: C40 Manual

Chapter 3 Instrument Operations

51

(b)

Figure 3-19 Body Markers

3. Rotate 【Gain】knob to adjust probe pictograph’s direction, and press 【Set】 key to

confirm it.

4. Press【Clear】key to clear body markers. Repeat 1-3 to reselect body marker.

3.11 Biopsy Guideline

Caution: Needle biopsy is highly dangerous for both patients and operating

doctors. The use of puncture rack and its accessories shall strictly follow the correct

needling sequence to avoid unnecessary risks to patients.

Caution: The biopsy guideline displayed on the equipment monitor is for reference

only. During biopsy, aspiration, puncture and related operations, the user shall

confirm the correct needling position.

Caution: Biopsy, aspiration and puncture are intervention operations with high

risks. Operators shall strictly stick to the principle of operation of sterilization.

The biopsy guideline is used to help the operator to move the probe to an appropriate

Page 56: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

52

position during puncture operation so that the end of the probe with the puncture rack is

located in the ideal needling position. The biopsy guideline can only be used under

real-time status.

Operating procedure:

1. In single B mode real-time status, press the【Biopsy】key. The screen displays a

biopsy guideline.

2. Press numeric keys【1】and【2】to adjust offset of the biopsy guideline.

3. Press numeric keys【3】and【4】to adjust angle of the biopsy guideline.

4. Press numeric key【5】to restore the default settings.

5. Press numeric key【6】to display the depth of the biopsy guideline.

6. Press the【Biopsy】key again to exit biopsy function.

Figure 3-20 Biopsy Guideline

Page 57: C40 Manual

Chapter 3 Instrument Operations

53

3.12 CINE

During B mode real-time imaging, the system places the most recently acquired images

as well as image currently on screen into a CINE memory buffer. This memory buffer can

store 512 frames. Film indicator will be displayed on right bottom of the image, which tells

us current frame and saved frame number.

Operating procedure:

1. Press 【Cine】key or 【FREEZE】key to enter CINE mode under real-time status.

2. Rotate 【Gain】knob to play forward or backward manually.

3. Press 【Gain】key to play automatically, and press 【Gain】key again to stop.

4. Press 【FREEZE】key again to exit CINE mode, and enter real-time image status.

Figure 3-21 CINE

Page 58: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

54

3.13 Image Managing and Printing

3.13.1 Image Managing

3.13.1.1 Image Storage

In the status of image freezing, press 【Save】key to store images.

3.13.1.2 Image Browsing and Exporting

Press 【Review】key to enter image browsing window.

Figure 3-22 Image Reviewing

Picture layout, 1×1,2×2 two formats available

Zoom the selected image ( Double-clicking the selected image can

also achieve full-screen zoom function .). Press【Exit】key to exit this

mode.

Page 59: C40 Manual

Chapter 3 Instrument Operations

55

View the previous image

View the next image

Delete all images

Delete the selected image

Print the selected image

Export all images

Export the selected image

Caution: You’d better select optional 512M Netac U265 or 1G Netac U208 provided

by Landwind. The system only supports FAT32 file-system. Before using your U-Disk,

please format it in your computer. After inserting an U-Disk, mark will be

displayed on the upper right of the screen, then you can store images in the U-Disk.

After storing, move the cursor to this USB mark, and press 【Set】key, then the U-Disk

information and operating dialog box will appear, click button and after a

tip box appeared, extract the U-Disk. System update is only used by the engineer

authorized by Landwind. Do not operate it casually, or accept the consequences.

Page 60: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

56

3.13.2 Printing

3.13.2.1 Printing Using a Laser Printer

Portable Ultrasound Diagnostic Scanner supports HP LaserJet P2015 printer.

Laser printer connection

Connect the data line of laser printer to the USB port of the system.

Printing Images and Reports

Open【Review】window, choose an image an click icon to print the

selected image.

Open【Report】menu, and click button to print a report.

3.13.2.2 Printing Using a Video Printer

Video printer connection

Connect one end of the video printer to the video output port of the system, and

the other end to the remote port of the system.

Printing an Image

We recommend only printing an image when freezing.

To print an image, please press the【Print】key.

Page 61: C40 Manual

57

Chapter 4 Measurements and Reports

4.1 Overview of measurement

The system provides the operation methods with general measurement (B-mode general

measurement, M-mode general measurement), and special measurement (abdominal

measurement, obstetrical measurement, orthopedics measurement, urology

measurement and small parts measurement).

4.1.1 Special Measurement Menu

The system’s special measurement is equipped with a special measuring menu. Take the

obstetrics measurement menu as an example to introduce the use of the special

measurement menu.

1.Title 2.Measurements(which have sub-menu)

3. Measurements 4.Exit the menu

5.Return the main menu

Figure 4-1 Special Measurement Menu-Obstetrical

In special measurement mode, press【2】key to enter obstetrics measurement menu, as

shown in Figure 4-1. Move cursorl and move the cursor to a measurement item, then

press【Set】key. If the there is logo on the item, you can enter the sub-menu; if there’s

not, you can enter the appropriate measurement mode. Choose RETURN to return the

Page 62: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

58

main menu of obstetrics. Choose EXIT to exit the main menu of obstetrics.

4.1.2 Measurement Results

Measuring results are displayed on the left side of the screen. If there are more measuring

information, the arrow will appear to carry out instructions. In measurement condition, you

can use 【Gain】knob to turn the page up and down to view the measuring results, as

shown in Figure 4-2.

Figure 4-2 the displayed screen of the measuring results

4.1.3 Measurement Report

The measurement report is used to record the measurement results. A measurement

report contains the report name, patient information, measurement and calculation results,

doctor diagnosis and more. The user can edit and print measurement reports. This system

provides general measurement report and exam-specific measurement report.

The exam-specific measurement reports include gynecological report, obstetrical report,

orthopedic report, urological report and small report.

Page 63: C40 Manual

Chapter 4 Measurements and Reports

59

General measurement report:

Figure 4-3 General measurement report

Operating procedure:

Click and buttons to select the appropriate

diagnostic image.

Input the diagnostic results in the column of DIAGNOSE.

Click button to print report.

Click button or press【Report】key to exit the report mode.

Page 64: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

60

4.2 General measurement — B-mode

The B-mode includes distance measurement, area/perimeter measurement, volume

measurement, angle measurement and histogram function.

Under real-time status, press 【Meas】key into the general measurement, or press

【FREEZE】key, and then press the numeric key 【5】 to enter the general measurement,

as shown in Figure 4-4.

Figure 4-4 B-mode General Measurement

4.2.1 Distance Measurement

Distance measurement is representative of the measurement on B-mode.

Press the numeric key【1】to enter the distance measurement mode. repeatedly pressing

【1】key can switch over between two measurement modes: line and trace.

4.2.1.1 DIST-LINE Measurement

Function: Measure the distance between two points in Ultrasound image.

Page 65: C40 Manual

Chapter 4 Measurements and Reports

61

Figure 4-5 DIST-LINE Measurement

Measurement:

1. Repeatedly press 【1】key to switch over the DIST-LINE measurement, at this

time a “+” cursor appears on the screen;

2. Use the trackball to move the “+” cursor to the starting point of the

measurement. press 【Set】key to determine the starting point;

3. Use the trackball to move the cursor, at this time between the starting point

and the cursor will always be a dotted line connection, the screen shows

real-time measurement result on the left;

4. Use the trackball to move the cursor to the end of the measurement.(Press

【Exit】key to switch the endpoints of the segments ) Press 【Set】 key, and

the measurement result will be finally confirmed, which means the

measurement finishes;

5. Repeat steps 2 to 4 to continue a next beeline measurement. The screen can

display a maximum of four measurement results. If it does beyond four,

Page 66: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

62

subsequent measurement results will automatically replace the previous ones.

6. Press【Clear】key to clear the measurement data and exit this mode.

Caution: An image can be measured four times at maximum, with different cursor

forms for each measurement.

4.2.1.2 DIST-TRACE Measurement

Function: Calculate the distance manually with the trackball.

Figure 4-6 DIST-TRACE Measurement

Measurement:

1. Repeatedly press 【1】key to switch over the DIST-TRACE Measurement, at this

time the “+” cursor appears on the screen;

2. Use the trackball to move the cursor to the starting point of the measurement,

and press【Set】key;

3. Use the trackball to make the cursor move along the edge of the target mobile,

outline the trace to determine the end point and then press【Set】key, at this

time real-time results will display on the left of the screen;

Page 67: C40 Manual

Chapter 4 Measurements and Reports

63

4. Repeat steps 2 to 3 to continue the next track measurement;

5. Press 【Clear】key to clear the measurement data and exit this mode.

4.2.2 Area/CIRC Measurement

4.2.2.1 Ellipse method

Function: Measure the circumference and area of a closed area by using ellipse approach

Operating procedures:

Figure 4-7 Ellipse Measurement

Measurement:

1. Repeatedly press【2】key to switch over the Ellipse method measurement

mode, at this time a “+” cursor appears on the screen;

2. Move cursorl and choose the starting point of the long axis. Press 【Set】key

to confirm the starting point;

3. Move cursorl to move the cursor to the end of the long axis (Press【Exit】key

to switch the endpoints of the elliptical axes), and then press 【Set】key to

Page 68: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

64

confirm the end;

4. Move cursorl and choose the start-stop point of the short axis. Press 【Set】

key to confirm the starting and end point, at this time, the result shows on the

left of the screen ;

5. Repeat steps 3 to 4 to continue the Ellipse measurement, at most 4 groups;

6. Press 【Clear】key to clear the measurement data and exit this mode.

4.2.2.2 Trace method

Function: The trace measurement is used to measure the area and circumference

manually with the trackball.

Figure 4-8 Trace Measurement

Measurement:

1. Repeatedly press【2】key to switch over the trace measurement mode, at this

time a “+”cursor appears on the screen;

Page 69: C40 Manual

Chapter 4 Measurements and Reports

65

2. Move the trackball to choose the starting point of the track, and press【Set】

key;

3. Move the trackball to choose the end point of the track, and press【Set】key, at

this time, the system will automatically connect the start and the end of the

track in a circle, at the same time, on the left of the screen will display the

results;

4. Repeat steps 2 to 3 to continue the trace measurement, at most 4 groups;

5. Press【Clear】key to clear measurement data and exit this mode.

4.2.3 Volume Measurement

4.2.3.1 3-LINE

Function: Calculate the volume with three distance measurements, volume measurement

need two images: longitudinal and transverse.

Figure 4-9 3-LINE Measurement

Measurement:

Page 70: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

66

1. Repeatedly press 【3】key to switch over 3-line measurement mode, at this

time a “+”cursor appears on the screen;

2. Use Section 4.2.1.1 DIST-LINE measurement, in a different plane to measure

the length of the segments, at this time, the actual measurement value of the

volume will be displayed on the left of the screen ;

3. Repeat step 2 to continue the 3-line measurement, at most 4 groups;

4. Press【Clear】key to clear measurement data and exit this mode.

4.2.3.2 1-LINE

Function: Calculate volume with one distance measurement.

Figure 4-10 1-LINE Measurement

Measurement:

1. Repeatedly press 【3】key to switch over 1-line measurement mode, at this

time a “+”cursor appears on the screen;

2. Use Section 4.2.1.1 DIST-LINE measurement to measure the length of a

segment , at this time, the actual measurement value of the volume will be

Page 71: C40 Manual

Chapter 4 Measurements and Reports

67

displayed on the left of the screen ;

3. Repeat step 2 to continue one distance measurement, at most 4 groups;

4. Press【Clear】key to clear measurement data and exit this mode.

4.2.3.3 LINE+ELLIPSE

Measurement:

1. Repeatedly press 【3】key to switch over LINE + ELLIPSE measurement mode,

at this time a “+”cursor appears on the screen;

2. Using section 4.2.1.1 DIST-LINE measurement to draw a line;

3. Then using section 4.2.2.1 Ellipse measurement to determine an oval; at this

time, the actual measurement value of the volume will display on the left of the

screen;

4. Repeat steps 2 to 3 to continue Distance + Ellipse measurement, at most 4

groups;

5. Press【Clear】key to clear measurement data and exit this mode.

Figure 4-11 LINE+ELLIPSE Measurement

Page 72: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

68

4.2.3.4 2-ELLIIPSE

Measurement:

1. Repeatedly press 【3】key to switch over Ellipse + Ellipse measurement mode,

mode, at this time a “+”cursor appears on the screen;

2. Using section 4.2.2.1 Ellipse measurement to draw two ellipses in different

planes, at this time, the actual measurement value of the volume will display

on the left of the screen;

3. Repeat step 2 to continue Ellipse + Ellipse measurement, at most 4 groups;

4. Press【Clear】key to clear measurement data and exit this mode.

Figure 4-12 2-ELLIPSE Measurement

4.2.4 Angle Measurement

Function: Measure the inclination of two lines in a plane in the ultrasonic image.

Measurement:

1. Press 【1】key to switch over the angle measurement mode, at this time a “+”

Page 73: C40 Manual

Chapter 4 Measurements and Reports

69

cursor appears on the screen;

2. Use the trackball to move the cursor to the starting point of the measurement,

and press【Set】key;

3. Move cursor to move the cursor to the end point. Press【Set】key again to

confirm the end point. By now, a segment will appear;

4. Move cursor to choose another segment. Press【Set】key to confirm the

starting and end point , at this time, the real-time measurement result of the

angle between the two segments displays on the left of the screen;

5. Repeat steps 2 to 4 to continue the next angle measurement;

6. Press 【Clear】key to clear the measurement data and exit this mode.

Figure 4-13 Angle Measurement

4.2.5 Histogram Measurement

Function: Calculate the gray distribution of ultrasonic echo signals in the specified region.

Measure the interest region by using rectangle and show the results by histogram.

Page 74: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

70

Caution: Histogram Measurement cannot used in B/M mode.

Measurement:

1. Press the number key【5】into the histogram measurement mode, at this time

a “+” cursor appear in the center of the screen;

2. Move the trackball and press 【Set】key to determine an acme of a rectangle;

3. Move the trackball , and the screen will display a rectangular box of samples;

4. Press【Set】to determine the vertex of the rectangle diagonally, and determine

the histogram measurements region. The histogram results show in the

middle of the screen;

5. Repeat steps 2 to 4 to continue histogram measurement;

6. Press【Clear】key to clear measurement data and exit this mode.

Figure 4-14 Histogram Measurement

The significance of the measurement result values as follows:

total: the sum of the data in the measuring region;

Page 75: C40 Manual

Chapter 4 Measurements and Reports

71

mean: average;

most: the maximum value of the measuring area;

ratio: the maximum percentage of the measuring region;

sd: standard deviation.

4.3 General measurement — M-mode

M-mode general measurement includes distance, time, slope and heart rate

measurements in B/M mode or single M mode real-time state. First press 【Meas】key,

then press number key 【4】 into M mode general measurement, or first press【FREEZE】

key, and then press numeric keys 【5】and【4】into the M mode general measurement.

Figure 4-15 M-mode General Measurement

4.3.1 Distance

Function: Measure the vertical distance between the two points in M mode image.

Measurement:

1. Repeatedly press【4】key to switch over distance measurement mode, at this

Page 76: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

72

time a “-” cursor appears on the screen;

2. Using the trackball to move the “-” cursor to the starting point of the

measurement, and press【Set】key, then the starting point for measuring

shows the baseline and the short horizontal line marker “-”, at this time, the

cursor can only move up and down, and the measurement results in real-time

display on the left of the screen;

3. Move the cursor up and down to the end of measurement, press 【Set】key,

and the final result shows on the left of the screen. The system automatically

enters the next measurement;

4. Repeat steps 2 to 3 to continue distance measurement, at most 4 groups.

5. Press【Clear】key to clear measurement data and exit this mode.

Figure 4-16 Distance Measurement in M mode

4.3.2 Time

Function: Measure the time interval between two points of the image in M mode.

Measurement:

Page 77: C40 Manual

Chapter 4 Measurements and Reports

73

1. Repeatedly press【4】key to switch over the time measurement mode, at this

time a vertical long-dashed line appeared on the screen;

2. Move the long-dashed line with Trackball to the starting point of the

measurement, press【Set】key, and the starting point for measuring displays a

vertical line. The cursor can only move horizontally to determine the starting

measurement, at the same time, the second long-dashed line appears on the

screen;

3. Move the cursor with Trackball, and the real-time measurement result will be

displayed on the bottom of the screen. Move the cursor to the end of the

measurement, and press【Set】key, then the final measurement results show

on the left of the screen. The system automatically enter the next

measurement after this measurement;

4. Repeat steps 2 to 3 to continue the time measurement, at most 4 groups;

5. Press【Clear】key to clear measurement data and exit this mode.

Figure 4-17 Time Measurement in M mode

Page 78: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

74

4.3.3 Slope

Function: Calculate the slope between two points by measuring their time and distance in

M mode.

Measurement: Please refer to section 4.2.1.1 DIST-LINE measurement.

Figure 4-18 Slope Measurement in M mode

4.3.4 Heart rate

Function: Measure the time interval between N times HR Cycle in M mode, and calculate

the number of heartbeats per minute.

Measurement:

1. Press【5】key into the heart rate measurement mode, at this time a long –dashed

broken line appears on the screen;

2. Use the trackball to move the long-dashed broken line to the starting point of the

measurement, press 【Set】key, where will display measurement base line and another

long-dashed line. At this time, the cursor can only be moved left and right, and the

measurement result will appear on the left of the screen real-time.

Page 79: C40 Manual

Chapter 4 Measurements and Reports

75

3. Move the cursor left and right to the end point of the measurement, press【Set】key,

and the final measurement result will appear on the left of the screen. After the

measurement, the system will enter next measurement automatically.

4. Press 【Clear】key to clear all measurement data.

5. Repeat steps 2-3 to measure heart rate continually.

Figure 4-19 Heart Rate Measurement in M mode

4.4 Gynecology Measurements

This section describes the system of the Gynecological measurement functions, including

Uterus (length, width, depth, endometrium), cervix (length, width, depth), the left ovary

(length, width, depth), the right ovary (length, width, depth), left-follicle (left-follicle1,

left-follicle2, left-follicle3, left-follicle4), right-follicle (right-follicle1, right-follicle2,

right-follicle3, right-follicle4).

Gynecology measurement only works in the B mode; there are two methods to enter the

Gynecology measurement, as follows:

Page 80: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

76

In real-time state, press【Calc】key to enter the special measurement, and then

press 【1】key to open the menu of gynecology measurement;

In real-time state, press 【FREEZE】 and【 6】 key to enter the special

measurement, and then press【1】key to open the menu of gynecology

measurement.

Figure 4-20 the main menu of Gynecology Measurement

4.4.1 Uterus Measurement

It is to measure uterus length, width and depth(The system automatically calculates the

size of the uterus)and endometrial thickness. Their measurement methods are distance

measurement in B mode (please refer to Section 4.2.1).

Figure 4-21 the menu of uterus

4.4.2 Cervix Measurement

It is to measure cervix length, width and depth, and then the system automatically

calculates the size of the cervix. The measurement methods are distance measurement in

B mode (please refer to the Section 4.2.1).

Page 81: C40 Manual

Chapter 4 Measurements and Reports

77

Figure 4-22 the menu of cervix

4.4.3 Ovary Measurement

It is to measure Ovary length, width and depth, and then the system automatically

calculates the volume of the ovary, The measurement methods are distance

measurement in B mode (please refer to the Section 4.2.1).

Figure 4-23 the menu of ovary

4.4.4 Follicle Measurement

Ovary can be measured within four follicles (follicle1, follicle2, follicle3, follicle4). After

measuring the diameter of the follicle, the system can automatically calculate the volume

of the follicle, The measurement methods are distance measurement in B mode (please

refer to Section 4.2.1).

Figure 4-24 the menu of follicle

Page 82: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

78

4.4.5 Gynecologic Report

In the gynecological measurement mode, press 【Report】 on the control panel to enter

the GYN report, as shown in Figure 4-25.

Refer to Section 4.1.3 General Measurement Report for details of the gynecological

measurement reports.

Figure 4-25 Gynecology Measurement

Page 83: C40 Manual

Chapter 4 Measurements and Reports

79

4.5 Obstetrics Measurements

This section describes the system of the obstetrics measurement function, including OB

measurements, AFI, FHR, NT, OB input, biophysical profile score and so on.

There are two methods to enter the obstetrics measurement, as follows:

In real-time state, press【Calc】key to enter the special measurement, and then

press 【2】key to open the menu of obstetrics measurement;

In real-time state, press 【FREEZE】 and【 6】 key to enter the special

measurement, and then press【 2】 key to open the menu of obstetrics

measurement.

Figure 4-26 the main menu of Obstetrics Measurement

4.5.1 Obstetrics Input

Obstetrics input items include the last menstrual period(calendar), the date of

ovulation(calendar), the number of pregnancies, the number of partition and the number

of abnormal pregnancies.

The data of delivery have been calculated from the last menstrual period and the date of

ovulation.

Operating procedure:

Click button to open the date choosing box (see Figure 4-28).

Click the triangle icons on the to choose year, and click the triangle icons

Page 84: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

80

on the to choose month.

Move cursor and press 【Set】to choose day.

Click button to confirm the date and exit the box, and then the

dates of LMP or ovulate and their EDD are displayed automatically, as shown in

Figure 4-27.

Figure 4-27 Obstetrics Input Box

Figure 4-28 choosing date

4.5.2 Fetus Selection

This system can measure up to 6 fetuses. After typing the number of fetus in the obstetrics

input box, the serial numbers of fetus can be switch in the fetus changing box. Choose to

enter the measurement of the fetus, and the serial number of the fetus is displayed on the

left side of the screen. Fetus No.1 is the default measurement .

Page 85: C40 Manual

Chapter 4 Measurements and Reports

81

Figure 4-29 selecting fetus number

4.5.3 Obstetrics Measurements

Figure 4-30 the menu of Obstetrics Measurement

Click anyone of the obstetrics measurements and enter the appropriate measurement

mode. Then the system automatically calculates the gestational age, which displayed in

the left side of the screen. Table 4-1 gives the methods and application scopes of the

measurements.

Page 86: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

82

Table 4-1 Obstetrics Measurements

Measurement Gestational age in table Method

GS Tokyo、Hellman、 Rempen、Hansmann、ChinaDistance

Measurement

CRL Tokyo、Hadlock、Nelson、Robinson、Rempen、

Hansmann、Osaka、China

Distance

Measurement

BPD Tokyo、Hadlock、Jeanty、Kurtz、Merz、

Hansmann、Rempen、Chitty(O-I)、 Osaka、

Distance

Measurement

FL Tokyo、Hadlock、Jeanty、Hohler、Merz、

Hansmann、Chitty、Osaka、China

Distance

Measurement

HUM Jeanty,Merz Distance

Measurement

HC Hadlock,Merz ,Hansmann,Chitty(pltd) Circumference

Measurement

AC Hadlock、Merz Circumference

Measurement

FTA Osaka Area

Measurement

FT Mercer Distance

Measurement

OFD Hansmann Distance

Measurement

CER Hill Distance

Measurement

TTD Fetal weight Distance

Measurement

Page 87: C40 Manual

Chapter 4 Measurements and Reports

83

APTD Fetal weight Distance

Measurement

For all the obstetrics measurements, if it is set up a fetus weight formula in measurement

settings and all the measurements corresponding to this formula has been measured, the

system will calculate the gestational age automatically. Table 4-2 gives every fetus weight

formula required for the measurement of parameters. Please set with the information from

Table 4-2.

Table 4-2 Obstetrics Formula

Formula Name Measurements

Tokyo BPD、APTD、TTD、FL

Osaka BPD、FL、FTA

Hadlock1 AC、FL

Hadlock2 AC、FL、BPD

Hadlock3 AC、FL、HC

Hadlock4 AC、BPD、FL、HC

Shepard AC、BPD

Merz1 AC、BPD

Hansmann BPD、TTD

Campbell AC

Taking the Osaka formula as an example, the operation steps are detailed below.

Page 88: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

84

1. Press 【Setup】key to enter the menu of system settings, and move cursor to open

the menu of measurement settings.

2. Choose the Osaka formula in EFW Formula Box, and confirm the settings.

3. Enter the obstetrics measurement mode and measure the items of BPD, FL and

FTA.

4. Then system calculates the fetus weight automatically, and the results are displayed

on the left side of the screen, as shown in Figure 4-31.

Figure 4-31 the Fetal weight measurement with “Osaka” formula

4.5.4 Amniotic fluid index

In the 4B mode, after four times distance measurement (refer to Section 4.2.1), the

system automatically calculates the amniotic fluid index.

Page 89: C40 Manual

Chapter 4 Measurements and Reports

85

Figure 4-32 AFI

4.5.5 Fetal heart rate

Fetal heart rate can only be measured in the B/M or M mode. For the measurement

method, please refer to Section 4.3.4 heart rate.

4.5.6 Neck Transparency

The measurement method of neck transparency (NT) is distance measurement in B mode,

referring to Section 4.2.1.

4.5.7 Fetal physiology score

Fetal physiology score is on the basis of the fetal ultrasound images, experiment or

measurement methods to get the fetal development indicators, and then speculate the

state of the fetus by a certain standard score.

The system provides criteria, which is based on the “Vintzileos” formula, as shown in

Table 4-3. Table 4-4 gives the description of the total score.

Figure 4-33 Biophysical Profile Score

Page 90: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

86

Table 4-3 Fetal Scoring Standard(Vintzileos Formula)

Fetal development

indicators 0 Score 1 Score 2 Score

Observation time

Remarks

Response

fetal heart

rate to

acceleration

FHR FHR

Increase ≤1

FHR increase

≥15/min,

duration≥15S,

1~4 times

FHR

increase≥15/min,

duration ≥15S,

≥5 times

20 minutes

Fetal

movement FM

No

movement 1~2 times ≥3 times 30 minutes

Fetal

breathing

movement

FBM No FBM, or

duration≤30S

FBM ≥1,

duration

30-60S

FBM≥1,

duration≥60S 30 minutes

Fetal tone FT

Limbs

Stretched,

without

bending,

hands open

≥1 times,

stretching

and bending

movement of

limbs of

vertebral

column

≥1 times,

stretching and

bending

movement of

limbs of vertebral

column

30 minutes

The scores

obtained

from fetal

response

experiment

can be

entered

into the

system

through

the

interface.

Placental

grading PL

III-level

placenta

Back wall of

placenta, the

placenta

difficult to

assess

classification

≤II-level placenta

According to the

audio-visual placenta to

the placenta mature

situations as 4 levels: level

0, I, II, III

The depth

of amniotic

fluid

AF AF<1cm 1cm≤AF≤2cm AF>2cm Obtained by measuring

B-mode images

Page 91: C40 Manual

Chapter 4 Measurements and Reports

87

Table 4-4 Description of the Total Score

Total Score Development condition

7-12 score Normal

3-6 score Equivocal

0-2 score Abnormal

4.5.8 Obstetrics Report

In the obstetrical measurement mode, press 【Report】 on the control panel to enter the

obstetrics report, as shown in Figure 4-34.

Refer to Section 4.1.3 General Measurement Report for details of the obstetrical

measurement reports.

Page 92: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

88

Figure 4-34 Orthopedics Report

4.6 Orthopedics Measurements

This section describes the orthopedics measurement function of the system, including BL,

ARL and IL. They can only be measured in the B mode.

Page 93: C40 Manual

Chapter 4 Measurements and Reports

89

Figure 4-35 the main menu of Orthopedics Measurement

There are two methods to enter Orthopedics measurements, as follows:

In real-time state, press【Calc】key to enter the special measurement, and then

press 【3】key to open the menu of orthopedics measurement;

In real-time state, press 【FREEZE】 and【 6】 key to enter the special

measurement, and then press【3】key to open the menu of orthopedics

measurement.

BL, ARL and IL are three lines, by the identification of BL and ARL, their included angle α

can be measured; by the identification of BL and IL, their included angle β can be

measured. The measurements of α and β are displayed on the left of the screen. The

measurements of BL, ARL and IL refer to Section 4.2.4.

4.6.1 Orthopedics Report

In the orthopedic measurement mode, press 【Report】 on the control panel to enter the

ORTH report, as shown in Figure 4-36.

Refer to Section 4.1.3 General Measurement Report for details of the orthopedic

measurement reports.

Page 94: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

90

Figure 4-36 Orthopedics Report

4.7 Urology Measurements

This section describes the system of the Urology measurement function, including

prostate volume (length, width, and depth), residual urine volume (length, width, depth)

and PSAD measurement. The prostate volume and residual urine volume can only be

measured on the B mode.

Page 95: C40 Manual

Chapter 4 Measurements and Reports

91

Figure 4-37 the main menu of Urology Measurement

There are two methods to enter Urology measurement, as follows:

In real-time state, press【Calc】key enter the special measurement, and then

press 【4】key to open the menu of Urology measurement;

In real-time state, press 【FREEZE】 and【 6】 key to enter the special

measurement, and then press 【 4 】 key to open the menu of Urology

measurement.

4.7.1 PSAD Measurement

The value of PSAD (serum prostate specific antigen density) can be calculated by PSA

( serum prostate specific antigen) and prostate volume. Firstly, Input the value of PSA in

urology PSA input box, then measure the prostate volume with the methods described in

section 4.7.2. Finally, the system automatically calculates the value of PSAD, and the

result is displayed on the left side of the screen.

Figure 4-38 PSA Input Box

4.7.2 Prostate Volume Measurement

It is to measure prostate length, width and depth, and then the system automatically

calculates the volume of the prostate. The method is the distance measurement (refer to

Section 4.2.1).

Page 96: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

92

Figure 4-39 the menu of Prostate Volume Measurement

4.7.3 Residual Urine Volume Measurement

By measuring the length, width and thickness diameter, the system automatically

calculates the volume of residual urine. The method is distance measurement (refer to

Section 4.2.1).

Figure 4-40 the menu of RUV Measurement

4.7.4 Urology Report

In the urological measurement mode, press 【Report】 on the control panel to enter the

URO report, as shown in Figure 4-41.

Refer to Section 4.1.3 General Measurement Report for details of the urological

measurement reports.

Page 97: C40 Manual

Chapter 4 Measurements and Reports

93

Figure 4-41 Urology Report

4.8 Small Part Measurements

This section describes the system of the small part measurement functions, including the

left side of the small part (length, width, depth) and the right side of the small part (length,

width, depth).

Figure 4-42 the main menu of Small Part Measurement

Page 98: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

94

Small parts measurement only in the B mode, there are two methods to enter the small

parts measurement, as follows:

In real-time state, press【Calc】key to enter the special measurement, and then

press 【6】key to open the menu of small parts measurement;

In real-time state, press 【FREEZE】 and【 6】 key to enter the special

measurement, and then press【6】key to open the menu of small parts

measurement.

4.8.1 Small Parts Measurement

It is to measure the small parts’ length, width and depth, and then the system

automatically calculates the volume of the small parts. The method is distance

measurement (refer to Section 4.2.1).

Figure 4-43 the menu of both sides of small parts measurement

4.8.2 Small Parts Measurement Report

In the small parts measurement mode, press 【Report】 on the control panel to enter the

SMT report, as shown in Figure 4-44.

Refer to Section 4.1.3 General Measurement Report for details of the small parts

measurement reports.

Page 99: C40 Manual

Chapter 4 Measurements and Reports

95

Figure 4-44 Small Parts Measurement Report

Page 100: C40 Manual

96

Chapter 5 Cleaning and Maintenance

5.1 Cleaning

Caution: Please turn off the power before cleaning the unit; otherwise, it may

cause electric shock and equipment damage.

Caution: Avoid any fluid from entering the keyboard or monitor of the equipment;

otherwise, it may cause electric shock or equipment damage.

5.1.1 Cleaning the Main Unit

To clean the surface of the ultrasonic diagnosis system:

1. Power off the ultrasound system and disconnect it from the power source.

2. Use a clean gauze pad or lint-free cloth, lightly moistened with a mild

detergent, to wide the surface of the ultrasound system. Take particular care

to clean the areas near the trackball and the slide controls. Ensure these

areas are free of gel and any other visible residue.

3. After cleaning, use a clean, lint-free cloth to dry the surface.

4. After cleaning, reconnect the ultrasound system to the power supply.

To clean the surface of probe bracket and coupling agent cup:

Use neutral cleaning agent to clean the surfaces of the probe brackets and

coupling agent cup, and then use a piece of lint-free cloth to dry them.

To clean the cables and connector of the diagnosis system:

Use neutral cleaning agent to clean the surfaces of the probe brackets and

coupling agent cup, and then use a piece of lint-free cloth to dry them.

Page 101: C40 Manual

Chapter 5 Cleaning an Maintenance

97

5.1.2 Cleaning the Probe

Caution: To avoid electrical shock and damage to the system, disconnect the

probe prior to cleaning or disinfecting.

Caution: Ensure surfaces of the probe are always disinfected during procedures

requiring disinfection.

Caution: To minimize the risk of cross contamination and infectious diseases,

clean and disinfect the transducer for entocoele and surgical operations thoroughly

after use each time.

Caution: The outer surfaces of an endocavity or intraoperative probe should be

checked to ensure there are no unintended rough surfaces, sharp edges, or

protrusions that may cause a safety hazard.

Caution: Probes are sensitive devices——irreparable damage may occur if they

are dropped, knocked against other objects, cut, or punctured. Do not attempt to

repair or replace any part of a probe.

Caution: Do not use probe sheaths containing oil-based coatings or petroleum-or

mineral oil-based ultrasonic coupling agents. Use only water-based ultrasonic

coupling agent.

Caution: Take extreme care when handling or storing probes. They must not be

dropped, jarred, or knocked against other objects. Do not allow probes to come into

contact with any sharp-edged or po9inted object.

Caution: Do not sterilize probes using hot steam, cold gas, or Ethylene Oxide

(ETO) methods. Before applying any other methods which might be recommended

by manufacturers of sterilization equipment, please contact your supplier.

Caution: Do not use abrasive cleaning agents, organic solvents such as benzene,

isopropyl alcohol, or phenol-based substances, or cleaning agents containing

organic solvents to clean or disinfect probes. These substances can damage the

probes.

Page 102: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

98

Caution: Do not use aerosol spray cleaner on the transducer, as the cleaner

liquid may enter the enclosure and damage the probe.

Operating procedure for probe cleaning and sterilization:

1. Disconnect the probe from the system..

2. Moisten a clean gauze pad with water and wipe the probe to remove any gel

or particles remaining on the probe. If water is not effective, you may use a

pre-enzyme cleaner

3. Carefully wipe the entire probe, including the cable and connector.

4. To disinfect the probe, take care to keep the cable strain relief and connector

of the probe dry while immersing the probe in an approved disinfectant.

5. Carefully follow the disinfectant manufacturer's instructions.

6. After cleaning or disinfecting, use a clean cloth to dry the probe.

5.2 Maintenance

Caution: To maintain the safety and functionality of the ultrasound system,

maintenance must be performed every 12 months. Electric safety test must be

performed at periodic intervals according to the local safety regulations or as needed.

Page 103: C40 Manual

Chapter 5 Cleaning an Maintenance

99

5.2.1 Main Unit Maintenance

5.2.1.1 External Parts Maintenance

Complete the following each day before using the system:

Visually inspect all probes. Do not use a probe which has a cracked, punctured,

or discolored casing, or a frayed cable.

Visually inspect all power cables. Do not turn on the power if a cable is frayed or

split, or shows signs of wear. Contact your dealer for replacement of the power

cable if a cable is frayed or split, or shows signs of wear.

Verify that the trackball, TGC sliding controls, and other controls on the control

panel are clean and free from gel or other contaminants.

Once the system is powered on.

Visually check the on-screen displays and lighting.

Make sure the monitor displays the current date and time.

Verify that the probe identification and indicated frequency are correct for the

active probe.

5.2.1.2 Internal parts Maintenance

1. It should be established science institution. The Internal parts maintenance should be

based on the use of the environment on a regular, generally speaking, it should not

be less than twice a year, recommendations of the late spring and the Mid-Autumn

two seasons, as more humid in the spring, summer and early autumn are hot climate,

these equipments are vulnerable to adverse effects on.

2. Internal maintenance should be chaired by engineer authorized by Landwind, never

blind to open the cover to carry out maintenance.

3. In the implementation of internal maintenance, electricity is strictly prohibited, and

pulling the power plug system to ensure that equipment and personal safety.

Page 104: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

100

4. Maintenance is completed, it is necessary to connect to the equipment and check

connector to prevent incorrect connections etc, electricity can confirmed after the

correct test machine.

5. Safeguard the process, pulled the prohibition of arbitrary changes to the wiring and

circuit parts of the location, do not free to spin the adjustable components.

6. Prohibit the use of washing internal parts, nor can of gasoline, kerosene and lacquer

thinner etc, to scrub the motor windings, commutator, as well as transformers,

plexiglass, cable and other rubber products. Alcohol and carbon tetrachloride on the

rubber, and other organic material has dissolved, and therefore can not be used to

scrub cables, rubber line, gelatin plate and mechanical spray paint, these organisms

can be used to dry the wet cloth to clean.

7. Connector on the gold-plated electrical contacts (or silver), it is strictly prohibited to

use gauze or other metal objects polished, and not in direct contact with their hands

in order to avoid corrosion caused by perspiration.

8. The proper use of tools in order to prevent the screw fixed as a result of damage to

the ultra-moment.

As manufacturers and installers of ultrasound equipment, Landwind cannot assume

responsibility for the safety properties, reliability, and/or performance of the equipment if:

Installations, extensions, readjustments, modifications, additions, or repairs are

carried out by persons not specifically authorized by Landwind.

Components that affect the safe operation of the system are replaced by parts

not authorized by Landwind.

The electrical installation of the room where the equipment is located does not

meet the power and environment requirements specified in this manual.

The equipment is not used in accordance with the operation instructions.

The system is operated by persons who do not receive good education or

Page 105: C40 Manual

Chapter 5 Cleaning an Maintenance

101

training.

Landwind suggests that you request any person that performs maintenance, or repairs, to

provide you with a certificate showing:

The nature and extent of the work performed

Changes in rated performance

Changes in working ranges

Date of service

Name of the person or firm performing the service

Signature of the person performing the service

Landwind refuses any responsibility whatsoever for repairs performed without the express

written consent of the Landwind.

5.2.2 Caring Probe

Caution: Ensure surfaces of the probe are always disinfected during procedures

requiring disinfection.

Caution: To minimize the risk of cross contamination and infectious diseases,

endocavity and intraoperative probes must be cleaned and high-level disinfected

after each use.

Caution: The outer surfaces of an endocavity or intraoperative probe should be

checked to ensure there are no unintended rough surfaces, sharp edges, or

protrusions that may cause a safety hazard.

Caution: Probes are sensitive devices——irreparable damage may occur if they are

dropped, knocked against other objects, cut, or punctured. Do not attempt to repair

or replace any part of a probe.

Caution: To avoid cable damage, do not roll the system over probe cables.

Caution: Do not use probe sheaths containing oil-based coatings or petroleum-or

Page 106: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

102

mineral oil-based ultrasonic coupling agents. Use only water-based ultrasonic

coupling agent.

Caution: Take extreme care when handling or storing probes. They must not be

dropped, jarred, or knocked against other objects. Do not allow probes to come into

contact with any sharp-edged or pointed object.

5.2.3 Caring Coupling Agent

Caution: Before use, examine the coupling agent for any material flaws.. Thinning,

bulging, or brittleness of the material indicates damage. Any product showing flaws

should not be used.

Do not store coupling agent below 5°C nor above 57°C. Coupling agent have a limited

shelf life. Before use, examine these products for any material flaws. Some packaging

may list an expiration date. Any product showing flaws, or whose expiration date has

passed, should not be used.

Page 107: C40 Manual

103

Chapter 6 Technical Description

6.1 Technical Data

Scanning mode electronic liner array, electronic convex array;

Scanning angle 25~120°;

Display depth 250mm;

Scanning depth 200mm(35C60H, B mode, 3.5MHz);

Frequency range 2.0~10MHz;

Probe configuration

35C60H(with a center frequency of 3.5MHz, R=60mm);

65C13H(with a center frequency of 6.5MHz, R=13mm);

75L40H(with a center frequency of 7.5MHz, L=40mm);

Display mode B, 2B, 4B, B/M, M;

CINE 512 frames, supporting automatic playback and manual

playback;

Image smooth 0, 1, 2, 3;

Edge enhance 3, 4, 5, 6;

Dynamic range 30~150dB, with a step of 6dB;

Frame correlation 0~7: eight adjustable ranges;

Gray reject 1~16 adjustable ranges;

Gray map 0~15 adjustable ranges;

Linear density H, M, L;

Page 108: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

104

Image width four optional width;

Focus mode Quadruple focus:16 optional focus positions;

M Sweep speed 0, 1, 2, 3 (four adjustable speed);

Image flip Left/right flip, up/down flip, polarity flip;

Zoom adjustable two levels of zoom frames;

Body Markers Each subject has fixed body markers;

Display information

Landwind trademark, Hospital, name, medical record ID,

Date/Time, probe information, IP Settings, Examination

subjects ,depth, focus, gray scale, direction symbol, etc;

Examination subjects Abdomen, Gynecology, Obstetrical, Small Part, Orthopaedy,

Urology, Emergency;

B-mode

measurements Distance, Area/CIRC, Volume, Angle, Histogram;

M-mode

measurements Time, Heart rate, Slope, Distance;

Specific Measurement:

Gynecology Uterus, Cervix, left/right ovary, left/right follicle;

Obstetrical

OB Measurements(GS、CRL、BPD、FL、HUM、HC、

AC、FTA、FT、OFD、CER、TTD、APTD), AFI(Q1、

Q2、Q3、Q4), FHR, NT, SELECT FNO, OB INPUT,

Biophysical profile score;

Orthopaedy BL、ARL、IL;

Urology Prostate VOL, PSAD, RUV;

Small parts Left, Right;

Annotation function Supporting body marker, puncture, arrow and text

Page 109: C40 Manual

Chapter 6 Technical Description

105

annotation;

Storage function Image Storage, Image Review, Report Print, etc.;

Acoustic Output parameters(B mode, 35C60H, 3.5MHz):

pra (Ma) 0.957

P (MW) 18.64

z at max. Ipi,α (cm) 4.70

deq(zb) (cm) 4.21

fawf (MHz) 3.445

X (cm) 2.81

Associated acoustic

parameter

Dim of Aaprt

Y (cm) 1.45

td (μsec) 0.332

prr (Hz) 3210

pr at max. Ipi (MPa) 1.673 Other Information

Ipa.3 at max. MI (W/ cm2) 101.42

Depth 110mm

Focal Position 50mm

Scan Width 1

Operating Control

Conditions

Line Density H

Monitor 10 inch CRT;

System VGA system, Supporting external VGA or video printer;

Cursor A combination of trackball and keyboard;

Keyboard Foldable and portable keyboard;

External interface USB, LAN, RS232, Video, VGA, Remote;

Weight Net weight: 9.5 kg; Gross weight: 12kg;

Page 110: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

106

Dimensions 522×365×420mm3

6.2 System Requirements

This section describes the power and environmental requirements for the ultrasound

diagnostic system.

6.2.1 Power supply Requirements

Power supply: 100-240V~ 50/60Hz 150VA

6.2.2 Connection with Peripheral Devices

Only the peripheral devices listed in the operation instructions are approved to use with

the ultrasound system. Any use of other devices with the system will be at the user’s risk

and may void the system warranty. On-board peripheral devices must be installed by an

authorized Landwind representative.

Caution: Landwind can only guarantee the performance and safety of the devices in

the Accessories and Options chapter. If in doubt, consult Landwind service department or

your local Landwind representative.

6.2.3 Operating Environment Requirements

Ambient temperature: 5℃~40℃

Relative humidity: ≤80%

Atmospheric pressure: 700hPa~1060hPa

Far away from equipment with strong electric and magnetic field and high

voltage

Avoiding the exposure of the screen in direct sunlight; light shielding indoors for

easy observation; good ventilation and anti-dust measures.

Page 111: C40 Manual

Chapter 6 Technical Description

107

6.3 Transport and Storage

6.3.1 Requirement for Transport and Storage Environment

Ambient temperature: -20℃~+55℃

Relative humidity: ≤93%

Atmospheric pressure: 700hPa~1060hPa

6.3.2 System Transport

When shipping the system, perform the following tasks, as appropriate.

To prepare the system for shipment over long distances or rough road:

Disconnect the probe first, and then disconnect the power cable and some

peripherals;

Repackage the system with the delivery-attached packages;

Please fix the system properly to avoid side movement of the system during

transport;

To prevent sudden jarring of the system during transport, provide shock cushions

beneath the system.

6.3.3 Probe Transport

Disconnect the probe from the system.

Place the probe correctly back into the probe box which is specially designed to

protect the probe. Do not dispose it.

6.4 Disposal of Packing Materials

Please properly keep all internal and external packing materials of the instrument

for future transportation and movement.

Please store all internal and external packing materials of the instrument

Page 112: C40 Manual

Portable Ultrasound Diagnostic Scanner User’s Manual

108

together.

6.5 Statement on Consumables

This instrument has no replaceable or consumable parts.

In the case of failure, please contact your dealer and never attempt to dismantle

the machine on your own.