CAIRE Portable Units FILLING PROCEDURES 1. 5. 6. 4. 8. 7. 2. 3. The portable unit should be filled 15 minutes prior to intended use. Clean the reservoir and portable unit fill connections to remove moisture and other contaminants with a lint free cloth. Do not disturb or depress the poppet on the reservoir or portable while drying the fill connectors. Also, be sure to keep the portable in the upright position at all times while wiping its fill connector. Position the portable unit above the reservoir’s Quick Disconnect Valve and align it with the indentation on top of the reservoir. Press down on top of the portable unit to engage with the reservoir’s Quick Disconnect Valve. Once the portable unit is engaged with the reservoir it can be filled. The unit is full when the hissing sound changes in tone and a dense, white vapor is seen coming from under the shroud of the reservoir. When this is observed, close the vent valve by raising its lever. Begin filling the portable unit by simply pulling its vent valve handle into its open position while continuing to apply pressure on the portable unit to maintain its connection. NOTE: FOR HELIOS PORTABLE UNITS ONLY: After 20-30 seconds, close and reopen the vent valve one or more times. This will help prevent the vent valve from sticking in the open position. Your portable unit is filled and ready to use! When disconnecting the portable unit, simply hit the disconnect plunger on the reservoir. This will disengage the portable from the reservoir. If the portable unit freezes to the reservoir while filling, simply close the vent valve handle and leave it attached. Once the fill connectors have thawed, the units can then be easily disengaged. If the portable unit or reservoir leak liquid oxygen from the poppet, reattach the portable to the reservoir and allow the connectors to warm sufficiently to melt the ice. Top Fill Models ML-LOX0004 E Questions? Please call our Technical Support Team: 800.482.2473 or email us at: [email protected]

Upload: others

Post on 23-Sep-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CAIRE Portable Units - maverickoxygen.com · 9/25/2015  · CAIRE Portable Units FILLING PROCEDURES 1. 4. 5. 6. 7. 8. 2. 3. The portable unit should be filled 15 minutes prior to

CAIRE Portable Units FILLING PROCEDURES

1.

5. 6.4.

8.7.

2. 3.

The portable unit should be filled 15 minutes prior to intended use. Clean the reservoir and portable unit fill connections to remove moisture and other contaminants with a lint free cloth. Do not disturb or depress the poppet on the reservoir or portable while drying the fill connectors. Also, be sure to keep the portable in the upright position at all times while wiping its fill connector.

Position the portable unit above the reservoir’s Quick Disconnect Valve and align it with the indentation on top of the reservoir. Press down on top of the portable unit to engage with the reservoir’s Quick Disconnect Valve. Once the portable unit is engaged with the reservoir it can be filled.

The unit is full when the hissing sound changes in tone and a dense, white vapor is seen coming from under the shroud of the reservoir. When this is observed, close the vent valve by raising its lever.

Begin filling the portable unit by simply pulling its vent valve handle into its open position while continuing to apply pressure on the portable unit to maintain its connection. NOTE: FOR HELIOS PORTABLE UNITS ONLY: After 20-30 seconds, close and reopen the vent valve one or more times. This will help prevent the vent valve from sticking in the open position.

Your portable unit is filled and ready to use!

When disconnecting the portable unit, simply hit the disconnect plunger on the reservoir. This will disengage the portable from the reservoir.

If the portable unit freezes to the reservoir while filling, simply close the vent valve handle and leave it attached. Once the fill connectors have thawed, the units can then be easily disengaged.

If the portable unit or reservoir leak liquid oxygen from the poppet, reattach the portable to the reservoir and allow the connectors to warm sufficiently to melt the ice.

Top Fill Models

ML-LOX0004 E

Questions?

Please call our Technical Support Team:

800.482.2473

or email us at:

[email protected]

Page 2: CAIRE Portable Units - maverickoxygen.com · 9/25/2015  · CAIRE Portable Units FILLING PROCEDURES 1. 4. 5. 6. 7. 8. 2. 3. The portable unit should be filled 15 minutes prior to

Portable Unit Filling Procedures 1. Clean the reservoir AND portable unit fill

connectors to remove moisture and other contaminants with a clean dry towel.

2. Position the portable above the quick disconnect valve and align it with the indentation at the top. Press down on top of portable to engage the unit.

3. Once the unit is engaged, filling can begin by pulling down the vent valve into its open position while applying pressure to top of unit to keep it in place.

4. After 20-30 seconds of filling, open and close the valve a few times…this will help keep the valve from sticking.

5. Unit is full when hissing sound changes in tone and a dense white vapor is observed. Close the fill lever by lifting it up.

6. Disengage the unit by hitting the disconnect plunger on the reservoir.

7. If the units freeze together, close the vent valve and wait till the connectors have thawed. They will then be easy to separate.