Testing- and Calibration Technology

Calibration Software EasyCal

Pressure and Temperature Measurement 

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 2/44

Example sheet from demo-version

Advantages demo-version:- Interactive presentation with Powerpoint

- no admin rights are necessary

- with explanation text and red marks will lead through the calibration

procedure

Screenshot demo-version of a calibration

procedure with software EasyCal

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 3/44

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 4/44

 Application: Calibration of a pressure gauge with accuracy 1,0%.

 As reference item there is a CPH 6200 with accuracy 0,2%.

I nterface cable

Test item accuracy 

1,0%

Calibration item

accuracy 0,2%

Test-pump to

generate pressure

Semi-automated computer aided calibration of 

pressure gauges with EasyCal

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 5/44

Semi-automated computer aided calibration of 

pressure gauges with EasyCal

General: A calibration is done by a direct comparison between the measured values of the

calibration object (test item) and those of the reference or working standard, which is directly or 

indirectly traceable to a national standard.

Calibration procedure:

The pressure values of the Test Cycle will be calculated by the Software EasyCal.*

Each time the recommended pressure value has to be set via a pressure source (test pump)

acc. to the display of the test item. (Notice: the pressure should not over-shoot

to detect the hysteresis of the test item) .

Is the respective pressure set, user informs the software via a mouse-click/Enter-key and the

value of the reference/CPH 6200 is read in via the RS-232/USB interface and stored next to the

recommended pressure value, which is currently shown by the test item. After all pressure values are likewise produced the software automatically do the evaluation

(calculation of uncertainty and hysteresis), stores the data in an  ACCESS®-Databank and a

printable Certificate is produced, which can be designed by the user (own Logo, etc).

* if needed the recommended pressure values can be changed.

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 6/44

Set Powerpoint to <presentation>

Please press the red circles with your mouse button

Semi-automated computer aided calibration of 

pressure gauges with EasyCal

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 7/44

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 8/44

56789

03.12.2004

1. step: Please fill in the basic data

12345678

Sample comp.

Sample departm.

Sample street

12345

Sample city

Sample country

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 9/44

2. step: Choice of a reference instrument

Select the

reference

instrument

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 10/44

Select the

calibration item

How to set up a new

calibration item, will be

shown in the appendix

3. step: Choice of a calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 11/44

990

5

Select the

calibration

method

4. step: Data for calibration procedure

The values of measurement points will be defined

automatically. The generation of these points are

influenced by the range and accuracy of the test

item.

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 12/44

5. step: Calibration procedure Calibration circuit:

Calibration

method: Read

measurement

value on test item.The set value is

set on the test

item and is

compared with the

actual value of the

reference item.

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 13/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 14/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 15/44

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

5. step: Calibration procedure Calibration circuit:

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 16/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 17/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 18/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 19/44

defined

hold-time

runs down

5. step: Calibration procedure

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 20/44

5. step: Calibration procedure

defined

hold-time

runs down

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 21/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 22/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 23/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 24/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 25/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 26/44

5. step: Calibration procedure Calibration circuit:

Generate pressure

with test pump. If 

set value is

reached oncalibration item

(test item), press

the OK button

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 27/44

5. step: Calibration procedure

 All measurement

values of the test

item have the

required

accuracy of 1,0%

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 28/44

5. step: Calibration procedure

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 29/44

Notes can be add,

which will be shownin the calibration

report.

6. step: Store measurement data and create test certificate

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 30/44

6. step: Store measurement data and create test certificate

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 31/44

6. step: Store measurement data and create test certificate

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 32/44

Here could your 

company logo be

placed

 Your 

company

logo

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 33/44

If the test certificate is created,

the calibration procedure is finished

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 34/44

Appendix

1.) Fully automated calibration with EasyCal

2.) Installation of a new calibration item

3.) The calibration item does not match the calibration class

4.) Product Support

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 35/44

1.) Fully-automated computer aided calibration of pressure

transmitters with EasyCal.

Up to the 5. step (the calibration procedure), the fully-automated calibration runs

analog to the semi-automated procedure, but the set values are generated

automatically by the pressure controller and the measurement values are stored.

The calibration report is also created automatically.

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 36/44

Pressure supply

(110% FS)

Test item

multimeter 

(with shunt)

PC with EasyCalIEEE-interface cableCPC 8000

Calibration procedure:

The CPC 8000 will control automatically all pressure steps (one after another), recommended / calculated by the Software

EasyCal.

Is one pressure controlled and stable the software will read the signal of the test item via the multimeter and will give the

signal to start to control the next pressure.

 After all pressure values are likewise produced the software automatically do the evaluation (calculation of uncertainty and

hysteresis), stores the data in an  ACCESS®-Databank and a printable Certificate is produced.

1.) Fully-automated computer aided calibration of pressure

transmitters with EasyCal.

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 37/44

2.) Installation of a new calibration item

Window calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 38/44

to essential characteristics

2.) Installation of a new calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 39/44

to output signal

2.) Installation of a new calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 40/44

to specified

test cycle

2.) Installation of a new calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 41/44

The values of the measuring points

are created automatically to chosen

measurement range.

The quantity of the measuring points

are created to chosen accuracy

2.) Installation of a new calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 42/44

New calibration item

is installed and can

be chosen

User interface: calibration item

2.) Installation of a new calibration item

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 43/44

The calibration

item does not

match thecalibration class of 

1%. It has a

tolerance of 1.1%.

The calibration has

to be stopped and

the calibration itemhas to be adjusted.

 After adjustment,

the calibration item

can be calibrated

again.

3.) The calibration item does not match the calibration class

8/8/2019 Cal Soft. Overview

http://slidepdf.com/reader/full/cal-soft-overview 44/44

For product order please contact:

Heike Scherf 

GB-PI-PK

Tel 09372 / 132-473

Fax 09372 / 132-217

E-Mail: [email protected]

Corinna Wengerter 

GB-PI-PK

Tel 09372 / 132-9986

Fax 09372 / 132-217

E-Mail: [email protected]

For product information please contact:

Christian Elbert

GB-PI-PK

Tel 09372 / 132-299

Fax 09372 / 132-217E-Mail: [email protected]

4.) Product Support