XR-3 Calibration Engineer Manual

1

XR-3 PC Console Engineer Manual

Revision History

Model : XR3 PC Console Rev Date Description Remarks

1.0 2008.05.28 Birthday of the manual

1.1 2008.06.19 Error solution added 2.2.3 The error coming before the program start

Page 2: Calibration Manual

XR-3 Calibration Engineer Manual

2

1. Before you start 1.1 Required PC specification 1.1.1 IBMPC 1.2 USB to RS232 Converter

1.3 USB to RS232 Converter & Generator connection

2. Program installation 2.1 Converter installation : 2.2 XR-3 PC Console installation 2.2.1 Auto setup 2.2.2 Manual Setup

2.2.3 The error coming before the program start

3. Program configuration and start 3.1 Configuration 3.1.1 Operation Part 3.1.1.1 kV,mA,Time Set adjust 3.1.1.2 Bucky Select 3.1.1.3 Focal Select 3.1.1.4 AEC 3.1.1.5 MC(Mass chest) 3.1.1.6 Tube Select 3.1.1.7 Stator Speed 3.1.1.8 Fluoroscope Mode 3.1.2 Set up part 3.1.2.1 Tube Setup Mode 3.1.2.2 System information 3.1.2.3 Bucky Type 3.1.2.3 System check 3.2 Start 3.2.1 Program run 3.2.2 Port check 3.2.3 Power ON 3.2.3 Exposure Key

4. Calibration 4.1 Mode entrance 4.2 Calibration Window configuration 4.2.1 Data input window 4.2.2 Level shift 4.2.3. Data Display 4.2.4. Feedback display

4.2.5 Calibration scope 4.3 Calibration start 4.3.1 Data change 4.3.2 Focal change 4.3.3 Tube change 4.3.4. Tube 2 Calibration

Page 3: Calibration Manual

XR-3 Calibration Engineer Manual

3

1. Before you start Before Calibration, prepare

1.1 Required PC specification 1.1.1 IBMPC

System Pentium MMX 200Mhz or higher or, compatible system

Operating system Windows XP Service Pack2 or higher

USB Interface USB 1.1 or 2,0 (USB 2.0 recommended)

1.2 USB to RS232 Converter - Converter of PC and Generator for data communication 1.3 Connector between USB to RS232 Converter and Generator

2. Program installation 2.1 Converter installation :

Converter Manual is provided with Converter and since the following explanation is quoted from the contents that are offered by manufacturer, there could be some differences according to manufacturer of Converter.

Page 4: Calibration Manual

XR-3 Calibration Engineer Manual

4

After finishing installation, it should be set up within Com5 in the Changes COM Port

※ If it is set up above Com6, PC Console could not operate correctly.

USB-Serial Cable User Guide 1. Product Features

2. System Requirements

3. Driver Installation (WIN2000)

4. Changes COM Port

5. Setting Up the RS232 Serial Device

6. Uninstalling the Driver Program

Page 5: Calibration Manual

XR-3 Calibration Engineer Manual

5

1. Product Features USB Specification Rev. 1.1 compliant

Support the RS232 Serial interface

Over 1Mbps data transfer rate

Support remote wake-up and power management

2. System Requirements IBM PC 486DX4-100 MHz CPU or higher or compatible system

Available USB port

CD-ROM drive

Windows 2000

3. Driver Installation (WIN2000) Follow the steps below to install Window 2000 driver of USB-Serial cable: 1. Power on your computer and make sure that the USB port is enabled and working properly.

2. During installation , please don‟t link USB-Serial cable with your computer . 3. Double click „ setup.exe ‟ , then it will start install

4. After installation , click „ OK ‟ 5. Plug in the USB-Serial cable into the USB port and run the Found New Hardware Wizard to assist you in

setting up the new device. Click Next to continue.

Page 6: Calibration Manual

XR-3 Calibration Engineer Manual

6

6. click Next to continue. Select the “ Search for a suitable driver for my device(recommend) “

7. Windows will detect the driver and Click Next to continue installation.

Page 7: Calibration Manual

XR-3 Calibration Engineer Manual

7

8. Click Finish to continue and let Windows copy the needed files to your hard disk.

9. You can to check the Device Manager and see the Prolific USB to Serial Com Port)

.

Page 8: Calibration Manual

XR-3 Calibration Engineer Manual

8

4. Changes COM Port

Under Windows 98SE , you can change the COM Port via setcom.exe sofeware

7. Under Windows Me 、 Windows 2000 、 Windows XP, please kindly following the steps as below. :

(1) Please click on the Device Manager\ Devices by connection \ ports (COM & LPT). ( images 1)

( images 1)

(2) Doubly click on “Prolific USB-to- serial Comm port(COM3)” .And it could go to „ properties'. (images

2)

(images 2)

Page 9: Calibration Manual

XR-3 Calibration Engineer Manual

9

(3) Please choose "Port Settings ", and then click on "Advanced". Please kindly choose the needed com, and click on" ok”. (images 3)

(images 3)

Notice:

If the Modem could not link with the new port, and run smoothly when changed the port, please kindly unload Modem, and re-install it. When Changing the port, the peripheral equipment could not shift to the new port, and run smoothly. Under this situation, please kindly unload the equipment and install again.

5. Setting Up the RS232 Serial Device Follow the steps below to connect your RS232 Serial Device (ie: Modem) to the USB port of your PC: 1. Turn off your Modem. Plug the DB 9 connector of the USB-Serial cable into your Modem. Turn on the

Modem afterwards. 2. Turn on your computer and plug in the USB connector of the USB-Serial cable into the USB port. 3. If you have already installed a Modem device before, click Start, Settings, Modems. Open Modems

Properties, select the Modem, click Properties. In General tab, please change the Port value to USB to Serial Port (COM3).

4. After finishing these processes, you can use USB to Serial Bridge Cable to connect RS232 Serial 56K FAX Modem and execute the Dial-up fuction to connect Internet, and send the FAX with the FAX software(ie: Microsoft FAX) properly.

Page 10: Calibration Manual

XR-3 Calibration Engineer Manual

10

6. Uninstalling the Driver Program If you want to remove the USB-Serial cable driver program, you can uninstall it by following the steps below: 1. Unplug the USB-Serial cable from your PC. 2. Run the DRemover98_2K.exe program from the software CD-ROM, then reboot the PC.

Page 11: Calibration Manual

XR-3 Calibration Engineer Manual

11

2.2 XR-3 PC Console installation 2.2.1 Auto setup

Page 12: Calibration Manual

XR-3 Calibration Engineer Manual

12

Run Cancel

Page 13: Calibration Manual

XR-3 Calibration Engineer Manual

13

Run

Cancel

Run Cancel

Finish

Page 14: Calibration Manual

XR-3 Calibration Engineer Manual

14

2.2.2 Manual Setup - After Auto setup, if the Program doesn‟t operate, you should install it manually.

Copy all files of DLL folder into “C:\WINDOWS\system32”

Page 15: Calibration Manual

XR-3 Calibration Engineer Manual

15

2.2.2 The error coming before the program start - When installing the program, the error might be occurred according to installation specification of PC System 2.2.2.1 Class Not Registered After press the button of Start and Execution on the operating display line of Windows in turns, input the following order

It is ok In the case that “Regsvr32.exe” file is not on the PC, please copy “regsv32a.exe” of “Util” Folder on the provided Program into “C:\Windows\system32\” and run it

Page 16: Calibration Manual

XR-3 Calibration Engineer Manual

16

2.2.2.2. File copy problem for Manual setup - In the case that oleaut32 or asycfilt item, it would be occurred

Copy the rest of file except for the upper two kinds of items

Page 17: Calibration Manual

XR-3 Calibration Engineer Manual

17

3. Program configuration and start 3.1 Configuration 3.1.1 Operation Part 3.1.1.1 kV,mA,Time Set adjust When choosing Fast in the case of kV, it would be changed into 10kVstep 3.1.1.2 Bucky Select -

3.1.1.3 Focal Select

Operation

Part

Set UP

Part

Page 18: Calibration Manual

XR-3 Calibration Engineer Manual

18

3.1.1.4 AEC

- Use AEC with ION Chamber

3.1.1.5 MC(Mass chest)

3.1.1.6 Tube Select

3.1.1.7 Stator Speed

3.1.1.8 Fluoroscope Mode

kV 의 경우 Fast 선택 시 10kVstep 으로 변화 된다.

3.1.2 Set up part : If it is possible, don’t be changed into Factory set mode in the field voluntarily

3.1.2.1 Tube Setup Mode - Establish min,max about each Tube kV mA - It is not displayed higher or lower than established value.

- Set up all of two Tube In the case of R/F System - When converting Tube, choose Tube1/2 by using Tube select

Page 19: Calibration Manual

XR-3 Calibration Engineer Manual

19

- Filament Preheat : Filament preheat voltage

3.1.2.2 System information S/N : product delivery date

MIN MAX

S.kV Establishing Min kV of Small Focus

Establishing Max kV of Small Focus

L.kV Establishing Min kV of Large Focus

Establishing Max kV of Large Focus

S.mA Establishing Max mA of Small Focus

Establishing Max mA of Small Focus

L.mA Establishing Max mA of Large Focus

Establishing Max mA of Large Focus

Page 20: Calibration Manual

XR-3 Calibration Engineer Manual

20

Slice : number of photographing H.V : kV Data B.C : mA Data kV : Display really checked kV mA : Display really checked mA (now it is not realized)

3.1.2.3 Bucky Type - Choose Bucky operating Type - Spring , Motor

3.1.2.3 System check - Display the error of System State - System State : mark Ready state ,EXP state

3.2 Start Connect Converter as below 3.2.1 Program run

PC Console 파일을 실행 합니다.

PC Generator

Page 21: Calibration Manual

XR-3 Calibration Engineer Manual

21

PC Console 실행 화면

3.2.2 Port check

- Port check 를 선택하면 현재 연결되어 있는 Port 와 Generator 간 자동으로 연결됩니다.

- COM6 번부터는 연결이 잘되지 않을 수 있으니 manual Page8 을 참고하여 Com port 를

Change Port Check into COM5

3.2.3 Power ON - If you choose Power ON, the figure of red area is going up and it will be Power on

Two-way Data Counter Connection Port

Page 22: Calibration Manual

XR-3 Calibration Engineer Manual

22

Power ON screen

3.2.3 Exposure Key - Ready and EXP are utilized by keyboard and acceptable keys are like below

Tube 1 Tube 2

Radiography Radiography Fluoroscope

Ready Ctrl Ctrl Z

Exposure X X

4. Calibration

Page 23: Calibration Manual

XR-3 Calibration Engineer Manual

23

4.1 Mode entrance - Choose “ show calibration window” on the right side of screen

Calibration Window

Choose Calibration ON/Off key

Page 24: Calibration Manual

XR-3 Calibration Engineer Manual

24

Data loading with display of “Now Data Access”

Data Load screen

Page 25: Calibration Manual

XR-3 Calibration Engineer Manual

25

4.2 Calibration Window configuration

4.2.1 Data input window - After inputting kV and mA data, Data would be changed with inputting “Enter” 4.2.2 Level shift - Use it when you increase and decrease overall kV and mA Data

- After inputting “Enter”, Data would be changed

kV Data input window mA Data input window

Tube Select Tube1,Tube 2

Focal Select Small,Large

Tube display Focus display

Calibration Point mA

Calibration Point kV

Calibration Data

Page 26: Calibration Manual

XR-3 Calibration Engineer Manual

26

4.2.3. Data Display - Data value that is currently checked is displayed 4.2.4. Feedback display

- The kV mA value checked really is displayed 됩니다.

- Only HV is operated, BC is not

4.2.5 Calibration scope

- If you check Calibration Point, in the case of mA, the scope of more than “100” is displayed all to “100”. Only first “100” is the area of Calibration and the following area is unnecessary.

- As kV is same, only the first “125” is the area Calibration and the following area is unnecessary

- Only the area of red is Calibration as below

Page 27: Calibration Manual

XR-3 Calibration Engineer Manual

27

4.3 Calibration start 4.3.1 Data change

- If you choose the wanted condition of Calibration, color of window is changed and data is input into “ADJ”

- Data change : change 740750 of ADJ HV to change Data

- Data input : after inputting the data, if you choose Enter, the changed data would be saved

Page 28: Calibration Manual

XR-3 Calibration Engineer Manual

28

- Screen after change 4.3.2 Focal chage If you choose , it would be able to chang into Small Large or vice versa (it changes every time you pressed button)

Page 29: Calibration Manual

XR-3 Calibration Engineer Manual

29

4.3.3 Tube change

When you choose , it changes Tube1 Tube2 (it changes every time you pressed button)

Page 30: Calibration Manual

XR-3 Calibration Engineer Manual

30

4.3.4. Tube 2 Calibration - Radiography : progress in Focal Large - Fluoroscope : progress in Focal Small

Only kV can be Calibration and mA is progressed in Generator

Fluoroscope mA Calibration VR