cardiac output measurement
TRANSCRIPT
![Page 1: Cardiac output measurement](https://reader037.vdocuments.net/reader037/viewer/2022100503/575091401a28abbf6b9cb747/html5/thumbnails/1.jpg)
CLINICAL MEASUREMENT
Cardiac output measurementMahesh Prabhu
AbstractMeasurement of cardiac output provides an estimate of whole body
perfusion, oxygen delivery and ventricular function. This article aims to
provide an overview of cardiac output measurement techniques, with
an emphasis on their principles of operation and limitations. Cardiac
output can be measured in terms of volume displacement or velocity of
blood flow. Although thermodilution with a pulmonary artery catheter
has been used extensively, less invasive technologies such as Doppler
echocardiography, pulse contour analysis, impedance plethysmography
and Fick partial rebreathing are increasingly used. The advantages and
limitations of each technique are different and each technique may find
a specific niche in the armamentarium of the clinician.
Keywords Doppler; Fick principle; indicator dilution; pulse contour
analysis; thermodilution
Definition and significance
Cardiac output is the volume of blood pumped by the heart per minute, and is the product of the heart rate and stroke volume. The stroke volume of the ventricle is determined by the interactions between its preload, contractility and afterload.
Clinical indicators such as capillary refill, urine output, serum lactate, core–peripheral temperature gradient and level of consciousness are inaccurate and nonspecific, especially with large variations in cardiac output. The measurement of cardiac output provides an estimate of wholebody perfusion, oxygen delivery and ventricular function. Systemic vascular resistance cannot be measured directly; however, monitoring cardiac output allows a better understanding of the causes of blood pressure variation.
Methods of measurement
The output of the ventricle is pulsatile and intermittent. By estimating the velocity or volume of blood through the aorta and integrating this volume with time, flow (i.e. cardiac output) can be calculated. Thus, cardiac output can be measured in terms of volume displacement or velocity. ‘Gold standard’ techniques for
Mahesh Prabhu, FRCA, is Consultant Cardiothoracic Anaesthetist at
the Freeman Hospital, Newcastle upon Tyne. He qualified from the
University of Bombay, India, and trained at Cambridge, Peterborough
and Newcastle upon Tyne. His interests include basic science,
cardiothoracic anaesthesia and intensive care.
ANAESTHESIA AND INTENSIVE CARE MEDICINE 8:2 6
cardiac output measurement, such as aortic electromagnetic or ultrasound transittime flowmetry, are highly invasive (Table 1).
Measurement of volume displacementThe Fick principle states that the total uptake or release of a substance by an organ is the product of the blood flow through that organ and the arteriovenous concentration difference of the substance. Thus, oxygen uptake in the lungs is the product of cardiac output and the difference in oxygen concentration in the arterial and venous blood in the lungs.
The Fick principle method is limited by invasive monitoring and the need for steadystate respiratory and haemodynamic conditions. However, portable metabolic monitors have been adapted for paediatric use.
Indicator dilution techniques: an indicator, when mixed completely into a unit volume of constantly flowing blood, can be used to identify the volume of blood displaced in a given time provided that the indicator remains in the system between the two points of injection and measurement.
Dye dilution – an inert dye such as indocyanin green (ICG) is injected into the central or peripheral vein. Arterial concentration is measured using a calibrated cuvette densitometer. Plotting the concentration of dye against time on a semilogarithmic scale and extrapolating the primary curve with exponential decay will create an indicator dilution curve (Figure 1). Recirculation of the dye elevates the baseline and limits the number of measurements. Pulse dye densitometry technique uses a pulse dye densitogram analyser with a finger or nose probe. To eliminate recirculation, the device automatically calibrates to zero ICG plasma concentration when a new measurement starts. A dye densitogram is displayed on a computer screen, and cardiac output can be derived from the area under the curve.
Thermodilution – is an intermittent technique, which is widely accepted as the clinical reference technique. An injection of cold injectate into the right atrium causes a decrease in blood temperature, which is measured by a thermistor in the pulmonary artery. The decrease in temperature is inversely proportional to the dilution of the injectate (i.e. cardiac output). The cardiac output can be derived from the modified StewartHamilton conservation of heat equation. The pulmonary artery catheter (PAC) is attached to a cardiac output computer that displays a curve and calculates the output and derived indices automatically. The accuracy of this technique can be improved by injecting boluses of 10 ml icecold saline at a constant, consistent rate using a closedloop injectate system throughout the ventilatory cycle for 2 minutes and rejecting any uneven curves. Since there is no recirculation peak or elevation of baseline, the measurements can be repeated.
Factors that may cause this technique to be inaccurate include intracardiac shunts, tricuspid regurgitation, cardiac arrhythmias, abnormal respiratory pattern and low cardiac output. Other drawbacks include invasiveness of the PAC, the concomitant morbidity and mortality and possible lack of costeffectiveness.
Continuous cardiac output (CCO) catheters work on the basis of the thermodilution principle, with a randomly pulsed heating coil placed in the right ventricle. The temperature change is detected in the pulmonary artery by a rapidresponse thermistor and a computer produces a continuously updated, timeaveraged
3 © 2007 Elsevier Ltd. All rights reserved.
![Page 2: Cardiac output measurement](https://reader037.vdocuments.net/reader037/viewer/2022100503/575091401a28abbf6b9cb747/html5/thumbnails/2.jpg)
CLINICAL MEASUREMENT
Comparison of methods of cardiac output measurement
Method Invasiveness Continuous
measurement
Validated by
trials in all
groups of
patients
Accuracy
(compared with
thermodilution
intermittent bolus)
Skills
required
Limitations
Thermodilution
intermittent bolus
Pulmonary artery,
central venous
canulation
No Yes Considered ‘gold
standard’
Cognitive,
technical skills
Intracardiac shunts,
arrhythmias, abnormal
respiratory pattern
Thermodilution
continuous
Pulmonary artery,
central venous
cannulation
Yes Yes Overestimates
by 4.3%
Cognitive,
technical
skills
As above
Pulse dye
densitometry
Central venous
canulation
No No Underestimates
(bias 0.42 litre/min,
LOA ± 1.91 litre/min)
Iodine allergy,
presence of an aortic
counterpulsation
device, severe renal
disease or liver
dysfunction
Lithium dilution Arterial, central
venous/peripheral
cannulation
No Yes Underestimates
by 5%
Cognitive Intracardiac shunts,
cannot be used in
patients on lithium
therapy
Oesophageal
Doppler
Oesophagus Yes Yes Underestimates by
10–15% (bias 0.24
litre/min, LOA +2 to
–1.5 litre/min)
Positioning
of probe
User dependent,
needs sedation,
physiological
assumptions
Transoesophageal
echo
Oesophagus No Yes Underestimates
by 7% (bias 0.5 ±
1 litre/min)
Specialist
expertise
User dependent,
needs sedation
PulseCO Arterial Yes No Cognitive Needs calibration,
dependent on
waveform
Pulse waveform
cardiac output
Arterial, central
venous cannulation
Yes Yes Overestimates
by 6% (bias 0.38
litre/min)
Hemodynamic
instability, severe
arrhythmia and
major changes in
SVR may affect
accuracy
Inert gas
throughflow
Double-lumen
endobronchial
intubation
Yes Yes Underestimates
by 5.2% (bias 0.68
litre/min, LOA +1.18
to –1.48 litre/min)
Double-lumen
intubation
needed
Lung separation is
vital
Impedance
plethysmography
Non-invasive Yes No Variable Electrocautery,
mechanical
ventilation and
surgical manipulation,
perioperative fluid
shifts
Non-invasive
cardiac output
measurement
Endotracheal
intubation, arterial
cannulation
Yes No Overestimates
by 10% (bias ± 1.8
litre/min)
Controlled
mechanical
ventilation
Stable CO2
elimination essential,
not real-time, poor
correlation in lung
disease
LOA, limits of agreement; SVR, systemic vascular resistance
Table 1
ANAESTHESIA AND INTENSIVE CARE MEDICINE 8:2 64 © 2007 Elsevier Ltd. All rights reserved.
![Page 3: Cardiac output measurement](https://reader037.vdocuments.net/reader037/viewer/2022100503/575091401a28abbf6b9cb747/html5/thumbnails/3.jpg)
CLINICAL MEASUREMENT
(every 3–6 minutes) value for cardiac output. This reduces workload for healthcare staff and also decreases procedural complications. This system is not truly continuous as it takes a variable time to update.
Lithium dilution – is an invasive technique, and is based on the principle of indicator dilution. Lithium is safe, nontoxic, rapidly redistributed and does not have any firstpass effect. Lithium chloride (0.002–0.004 mmol/kg, maximum cumulative dose 3 mmol) is injected into a central or peripheral vein. Arterial plasma concentrations are measured by withdrawing blood across a lithiumselective electrode at a rate of 4 ml/min. Cardiac output can be calculated from lithium dose and the area under the primary concentration–time curve before recirculation. This technique cannot be used in patients on lithium therapy or atracurium.
Inert gas throughflow: this technique measures continuous cardiac output using differential Fick method. A doublelumen endobronchial tube is used to achieve functional separation of the two lungs. Nitrous oxide is administered exclusively to one lung. The continuous transfer of inert gas from one lung to the other through the circulation is termed ‘through flow’. Simultaneous measurement of fresh gas and mixed expiratory concentration in each lung allows pulmonary blood flow to be calculated. Smooth, synchronous ventilation of both lungs as well as correct placement of the doublelumen tube is essential for accuracy.
Non-invasive cardiac output measurement (NICO): this technique, also known as differential carbon dioxide Fick partial rebreathing method, derives pulmonary capillary blood flow from the ratio of change in carbon dioxide elimination to change in endtidal carbon dioxide (EtCO2), in response to a brief period of partial rebreathing. The production of carbon dioxide (VCO2) is calculated from minute ventilation and the instantaneous
Indicator dilution curve
Source: Hinds C J, Watson D. Intensive care. Philadelphia:
W B Saunders, 1997.
Time
1
1
2
3
4
5
2
4
3
5
Transit timeInjection
Log
co
nce
ntr
ati
on
of
dye
Primary curve
Recirculation peak
Extrapolation of primary decay curve
Elevation of baseline secondary to circulation of dye
Appearance time
Area under
the curve
Figure 1
ANAESTHESIA AND INTENSIVE CARE MEDICINE 8:2 6
carbon dioxide content, and the arterial carbon dioxide content is estimated from EtCO2. Cardiac output is proportional to the change in carbon dioxide elimination divided by the change in EtCO2 resulting from rebreathing. This technique assumes that the mixed venous CO2 content and deadspace are constant. An intrapulmonary shunt correction factor derived from fractional inspired oxygen concentration and pulse oximetry is added to the final equation. The NICO monitor should be used only whilst the patient is under fully controlled mechanical ventilation. Arterial blood samples are required for shunt estimation.
Measurement of blood velocity and flowDoppler: with the Doppler technique blood flow measurements are usually taken at the aorta, from either the transthoracic or transoesophageal approach. Blood velocity is calculated from the frequency shift of reflected ultrasound waves using the Doppler principle. The ultrasonic beam (continuous wave or pulse wave Doppler) is positioned parallel to the direction of blood flow to display the maximum blood velocity signal. The crosssectional area (CSA) of the aorta is measured or computed. The integral of velocity versus time (i.e. velocity time integral (VTI)) during a cardiac cycle, the CSA and a correcting factor based on anatomical and mathematical assumptions is used to calculate stroke volume (stroke distance) and cardiac output. Limitations of this technique include incorrect probe positioning and the possibility of having to anaesthetize the patient.
Echocardiography allows the measurement of cardiac output by twodimensional imaging (measurement of ventricular volumes) or Doppler. The velocity of blood flow across the aortic/mitral valve or the left ventricular outflow tract is measured with a pulsewave Doppler. With twodimensional echocardiography, the CSA of the particular structure is determined. Stroke volume (SV) can be determined using the CSA and the stroke distance travelled by the column of blood at a given anatomicial site.
SV = VTI × CSA
Echocardiography can also be used to define indices of diastolic dysfunction. Twodimensional echocardiography usually underestimates volumes of the left ventricle (LV). Transoesophageal echocardiography is repeatable but is user dependent and requires the patient to be anaesthetized.
PulseCO haemodynamic monitor: the PulseCO system (LiDCO Ltd) provides beattobeat calculation of cardiac output by analysis of peripheral arterial pulse waveform using an autocorrelation algorithm to convert the analogue pressure waveform into nominal stroke volume. Changes in blood volume in the arterial tree are measured over the range of arterial distensibility. This provides an estimate of the volume of blood flowing out of the arterial circuit during a period nearly equal to diastole. This volume is equivalent to stroke volume. This technique is comparable to thermodilution and provides an early warning of haemodynamic changes. However, it needs calibration by the lithium dilution technique and may be limited by damping of the waveform or fluctuations in temperature and vascular resistance.
5 © 2007 Elsevier Ltd. All rights reserved.
![Page 4: Cardiac output measurement](https://reader037.vdocuments.net/reader037/viewer/2022100503/575091401a28abbf6b9cb747/html5/thumbnails/4.jpg)
CLINICAL MEASUREMENT
Pulse contour cardiac output: cardiac output is computed by analysis of the area under the systolic portion of the arterial pressure waveform, from the enddiastole to the end of the ejection phase; this corresponds to stroke volume. A continuous pulse waveform contour analysis is obtained using a long arterial catheter (with thermistor) placed in the femoral, axillary or brachial artery and connected to a pulse contour device (PiCCO, PULSION Medical Systems). A beattobeat analysis of cardiac output, averaged for 30 seconds, is displayed. Central venous cannulation is required for calibration using transcardiopulmonary thermodilution technique. Changes in aortic impedance caused by haemodynamic instability, severe arrhythmia, and major changes in systemic vascular resistance may interfere with accuracy.
Impedance plethysmography: measurement of changes in transthoracic electrical impedance due to ejection of blood into the ascending aorta during each cardiac cycle allows the assessment of stroke volume. An alternating current of low amplitude and high frequency is introduced and simultaneously sensed by two sets of electrodes placed around the neck and thorax. By using mathematical models, alterations in thoracic impedance can be used to derive stroke volume and cardiac output.
ANAESTHESIA AND INTENSIVE CARE MEDICINE 8:2
Electrocautery, mechanical ventilation, surgical manipulation, perioperative fluid shifts and pulmonary oedema can cause distortion of impedance signal. Cardiac output measurement is unpredictable in a perioperative and critical care environment. ◆
FurTher reADIng
Botero M, Kirby D, Lobato EB, et al. Measurement of cardiac output
before and after cardiopulmonary bypass: comparison among aortic
transit-time ultrasound, thermodilution, and noninvasive partial CO2
rebreathing. J Cardiothorac Vasc Anesth 2004; 18: 563–72.
Bursten A, Soni N. Oh’s Intensive care manual, 5th edn. Philadelphia:
Butterworth Heinemann, 2003.
Cholley B, Payen D. Noninvasive techniques for measurements of
cardiac output. Curr Opin Crit Care 2005; 11: 424–9.
Dark PM, Singer M. The validity of trans-esophageal Doppler
ultrasonography as a measure of cardiac output in critically ill
adults. Intensive Care Med 2004; 30: 2060–6.
Roizen M, Berger D, Gabel R. Practice guidelines for pulmonary artery
catheterization: an updated report by the American Society of
Anesthesiologists. Task force on pulmonary artery catheterization.
Anesthesiology 2003; 99: 988–1014.
66 © 2007 Elsevier Ltd. All rights reserved.