carolyn atkins, kiranmayee kilaru, brian ramsey, mikhail gubarev, steve o’dell, will w. zhang....
DESCRIPTION
Introduction Horizontal chamber Vertical chamber Summary, conclusion and future workTRANSCRIPT
![Page 1: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/1.jpg)
Overview of Differential Deposition at NASA MSFC
Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang.
AXRO 2015, Prague, Czech Republic, 8th December 2015
![Page 2: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/2.jpg)
Presentation outline
Introduction Horizontal chamber Vertical chamber Summary, conclusion and future
work
![Page 3: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/3.jpg)
Presentation outline
Introduction Horizontal chamber Vertical chamber Summary, conclusion and future
work
![Page 4: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/4.jpg)
Introduction to Differential Deposition
![Page 5: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/5.jpg)
5. X-ray testing
Differential Deposition process loop
2. Correction profile and simulations
4. Metrology1. Metrology
3. Perform correction
![Page 6: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/6.jpg)
Proof of concept
0 5 10 15 20 25 30
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
desired profileprofile before correctionprofile after correctionsimulated profile
Axial position along mirror (mm)
Profi
le h
eigh
t (µ
m)
Slope improvement from 12 to 7 arcsec RMS
K.Kilaru et al, Optical Engineering 50(10), 106501 (October 2011)
Initial proof of concept was performed upon miniature optics intended for radio-nuclide imaging using an existing RF sputtering system.
Form Talysurf data demonstrated an improvement in slope from 12 to 7 arc-sec.
Accurate metrology was the limiting factor in this investigation.
RF sputtering chamberMedical imaging optics
![Page 7: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/7.jpg)
Vertical Chamber• DC magnetron sputtering
source• Designed for segmented
glass optics and large diameter full shell optics.
Horizontal Chamber• DC magnetron sputtering
source• Designed for full shell
electroformed optics
Concept developmentFollowing on from the proof of concept investigation a grant was awarded to further the research to astronomical X-ray optics.
![Page 8: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/8.jpg)
Presentation outline
Introduction Horizontal chamber
- research conducted by Dr. Kiranmayee Kilaru Vertical chamber Summary and conclusion
![Page 9: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/9.jpg)
Horizontal Chamber
X-ray optic held in a rotating and translating collet
Mask
X-ray optic
Magnetron head
Horizontal differential deposition chamber
Magnetron head within the optic
• Developed for full shell electroformed optics.• The optic translates and rotates relative to a stationary
sputtering/magnetron head.• The rotation and translation are encoder to ensure accurate
positioning.• Maximum shell diameter and length: 0.25m and 0.6m
respectively.• 1 inch (25.4mm) diameter sputtering/magnetron head
Work performed by Dr Kiranmayee Kilaru
![Page 10: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/10.jpg)
Metrology - VLTP
• Vertical Long Trace Profiler• 1 mm spatial interval• New 2D camera and
modified software• Established procedures to
obtain repeatability of <100 Angstroms
• Multiple measurements - 4 scans per meridian with 4 realignment scans
![Page 11: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/11.jpg)
Initial experimentation: single meridian correction
0 20 40 60 80 100 120 140 160 180-2000
-1500
-1000
-500
0
500
1000
1500
mm
Ang
stro
ms
Profile pre- & post- correction
pre-correctionpost1-correctionpost2-correction
complete profile only higher-order frequencies0
1
2
3
4
5
6
7
8Calculated HPD
7.08
3.68
2.87
6.71
2.95
1.86
arc-
seco
nds
pre-correctionpost1-correctionpost2-correction
0 20 40 60 80 100 120 140 160 180-800
-600
-400
-200
0
200
400
600
800
mmA
ngst
rom
s
Higher-frequencies of profile
pre-correctionpost1-correctionpost2-correction
Initial experimentation was performed on single meridians from Con-X replicas.
These results demonstrate the successful application of Differential Deposition upon astronomical X-ray optics.
The HPD values are calculated solely from VLTP metrology data.
![Page 12: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/12.jpg)
Correction of optics for X-ray testing
• Optical prescription: parabola-hyperbola, nominal 140mm diameter, 580mm in length.
• Correction applied to both the parabolic and hyperbolic sections of the shell.
• Correction was performed in 3 sections comprising of 3 meridians each. Each meridian was separated by 24°.
• The optic was measured at MSFC’s 100m long Stray Light Facility.
• The optic can be tip-tilted and translated to obtain the optimum focus.
• Before and after correction measurements were performed using both the Stray Light Facility and the VLTP.
![Page 13: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/13.jpg)
Correction 1 – X-ray testing and metrology
• VLTP metrology pre and post coating estimated an improvement from 15.7arc-sec to 9.2arc-sec.
• A 5mm slit was used to perform this correction, which targeted all spatial frequencies and customised for each meridian. • X-ray testing
highlighted an improvement from 15.7arc-sec HPW to 10.7 arc-sec HPW.
• Upon analysis the discrepancy between the expected and actual values was caused by the overlap region
![Page 14: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/14.jpg)
Correction 2 – metrology
0 100 200-100
-50
0
50
100Slope
[mm]
[mic
rora
dian
s]
0 100 200
-1
0
1
2
Height
[mm]
[mic
rons
]
0 100 200
-500
0
500
Hitmap
[mm]
[nm
]
0
5
10
15
20 17.29
4.54
[HP
D in
arc
sec
s]
Estimated Performance
prepost
prepost
prepost
prepost
0 100 200-100
0
100
200Slope
[mm]
[mic
rora
dian
s]
0 100 200-2
-1
0
1
2
3
Height
[mm]
[mic
rons
]
0 100 200-400
-200
0
200
400
600
Hitmap
[mm]
[nm
]
0
5
10
1512.33
4.25
[HP
D in
arc
sec
s]
Estimated Performance
prepost
prepost
prepost
prepost
Single meridian parabolic segment
Single meridian hyperbolic segment
• Second correction used the same ART shell from correction 1, but on a different azimuthal region.
• 3 meridians at 24° separation were corrected.
• Corrections customised for each meridian were used.
• When considering the overlap regions between meridians, an improvement of 7.9 arc-sec was estimated post correction.
![Page 15: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/15.jpg)
Correction 2 – X-ray testing
Higher frequency- correction
Key point
A factor of >2 improvement has been observed in X-rays from a single stage of correction.
Further results and information can be found in the paper:
Kilaru, K. et al., “Progress in differential deposition for improving the figures of full-shell astronomical grazing incidence x-ray optics”, Proc. of SPIE 9603, 96031F, 2015
![Page 16: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/16.jpg)
Presentation outline
Introduction Horizontal chamber Vertical chamber Summary, conclusion and future
work
![Page 17: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/17.jpg)
Challenges for segmented glass correctionSEGMENTED GLASS OPTICS Shape – segmented, leading
to a less rigid structure. Material – borosilicate,
lightweight, but easily deformed.
Mounting – point fixture, where the optics are held at discreet points.
NI-CO REPLICATED OPTICS Shape – full shell, rigid
structure. Material – Ni-Co alloy, heavier
but less prone to deformation.
Mounting – collet fixture, where the optic is held uniformly around its circumference.
![Page 18: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/18.jpg)
Vertical chamber hardware
• 1m in diameter, 1.3m in height.• Vertical translation provided by a portable tower which has a travel of
~680mm in the vertical direction.• The 360° rotation is provided by an annular platform. • Optical encoders are used on both the translation and rotation to ensure
accurate positioning and feedback.• A 1 inch DC magnetron is to sputter the material for the differential
deposition process.
Translation tower
Optic
Mask and slit
Kinematic mountTip/tilt
stage
Ceramic
Magnetron
The slumped glass optics presented have been developed by NASA GSFC.
![Page 19: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/19.jpg)
Metrology – Form Talysurf
Photo courtesy of Lewis Moore, NASA tribology group
Due to the requirement for accurate alignment and repeatability a kinematic mount has been developed to ensure accurate positioning of the optic relative to the profiler’s stylus between coatings.
The final profile, which is used to define the required correction, is composed of 2 sets of 5 repeat measurements with the optic being repositioned between each set.
Typically the accuracy of alignment between placements is within 10/20 microns.
![Page 20: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/20.jpg)
Initial experimentation: Optic 2, coating 1
Initial experimentation used an aluminium fixture which held the optic along its azimuthal length.
Successful corrections were obtained from targeting small localised regions.
Optic 2 – coating 1
![Page 21: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/21.jpg)
Initial experimentation:Optic 2, coating 3
Optic 2 – coating 3
The third coating was successful in correcting the first half of the optic; however the latter half exhibited an artefact. Furthermore subsequent optics all demonstrated similar defects after 1-3 coatings.The following were investigated: stress, mounting and thermal effects.Results indicated that all three could be factors and as such the mount was redesigned with ceramic components (to match the optic’s cte) in addition to different fixture points.
![Page 22: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/22.jpg)
Ceramic mount results
![Page 23: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/23.jpg)
Ceramic mount results
These initial results from the ceramic mount demonstrate the gradual improvement of the slumped glass optic after each coating.
Further coatings upon this optic were abandoned due to loss of positioning accuracy, this has been rectified for future optics and coatings.
![Page 24: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/24.jpg)
Presentation outline
Introduction Horizontal chamber Vertical chamber Summary, conclusion and future
work
![Page 25: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/25.jpg)
Summary, conclusion and future work: Horizontal
Summary and conclusion• An improvement of a factor >2 has been measured
in X-rays from a NiCo full shell astronomical optic.• Improvements by a factor of 3 are observed in VLTP
metrology data.• Results indicate that overlap regions between
meridians and undesired scatter from the gas inlets of the magnetron mask are currently limiting the full scope for the correction.
Future work
• To perform a full 360° correction of a NiCo full shell astronomical optic.
• To use different width slits to target different spatial frequencies to further improve the resolution.
• To improve the efficiency of the Differential Deposition process through incorporating in-situ metrology.
Gas inlet
![Page 26: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/26.jpg)
Summary, conclusion and future work: Vertical
Summary and conclusion• The differences and challenges of the slumped glass
optics versus the NiCo optics have been identified.• Early results were presented which indicated
sensitivity of the glass optics to: stress, mounting and thermal effects.
• The redesigned ceramic mount demonstrated successful corrections after several coatings.
• Slumped glass optics can be corrected via the Differential Deposition method.
Future work
• To ensure consistent correction after several coating using the ceramic mount.
• To implement the single meridian correction into a global full optic correction.
![Page 27: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/27.jpg)
AcknowledgementsThe author wishes to acknowledge:
Chip Moore at the MSFC Tribology Laboratory for use of the metrology equipment.
Will Zhang at NASA’s GSFC for supplying the slumped glass optics for the experiments.
All those who helped in the construction of the vacuum chambers.
and the X-ray astronomy group at NASA’s MSFC
Thank you for your attention!
![Page 28: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/28.jpg)
![Page 29: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/29.jpg)
29
Initial Results – Stress?Stress in a thin film can be estimated using Stoney’s equation,
σf = film stressE = Young’s modulus of the substrateν = Poisson’s ratio of the substrateroc = measured change in radius of curvaturetf = film thickness
Borosilicate Glass wafer
Measured stress [MPa]
Wafer 1 119Wafer 2 175Wafer 3 172Wafer 4 248
Average measured stress = 179MPa (Tensile)
This is an active area of investigation with several options: -1. Reduce the stress through
changing the coating parameters (gas pressure, power etc.)
2. Create an FEA model of the stress to predict its effects.
3. Use the stress to correct the low order spatial frequencies within the surface.
9603-46
![Page 30: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/30.jpg)
30
Initial Results – Mounting
Coating direction
Second, the mount was flipped to reverse the direction of coating to see if there was a mechanical influence. The results indicate that there could be a mechanical effect and this warrants further investigation.
9603-46
![Page 31: Carolyn Atkins, Kiranmayee Kilaru, Brian Ramsey, Mikhail Gubarev, Steve O’dell, Will W. Zhang. AXRO 2015, Prague, Czech Republic, 8 th December 2015](https://reader034.vdocuments.net/reader034/viewer/2022042800/5a4d1b657f8b9ab0599aff77/html5/thumbnails/31.jpg)
31
Initial Results - Temperature
Thermocouples have been set up within the vacuum chamber to measure the surface temperature of different components during coating.
Options to minimise the effects of thermal mismatch include: matching the cte of the mount to that of the optic; and modelling the effects to offset them in the correction
9603-46
Gas in (Ar)
Power on
Power off
Air in