cav® 4.2 operator's manual - cannon instrument

86
CANNON® CAV® 4.2 Automated Dual-Bath Kinematic Viscometer ASTM D445 CAV® 4.2 Operator’s Manual P/N 68.0505

Upload: others

Post on 14-Apr-2022

10 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON® CAV® 4.2

Automated Dual-Bath Kinematic Viscometer ASTM D445

CAV® 4.2 Operator’s Manual

P/N 68.0505

Page 2: CAV® 4.2 Operator's Manual - CANNON Instrument
Page 3: CAV® 4.2 Operator's Manual - CANNON Instrument

CAV® 4.2 Operator’s Manual

CANNON Instrument Company® 2139 High Tech Road

State College, PA 16803

Page 4: CAV® 4.2 Operator's Manual - CANNON Instrument

ii CANNON®

CAV® 4.2 Operator's Manual

Copyright Copyright © 2020 CANNON Instrument Company®. All rights reserved.

Trademarks CANNON® the CANNON® logo and CAV are registered trademarks of CANNON Instrument Company®.

Brookfield® is a registered trademark of Brookfield® Engineering Laboratories, Inc.

Watlow® is a registered trademark of Watlow® Electric Manufacturing Company.

Contact Address: CANNON Instrument Company® 2139 High Tech Road State College PA 16803, USA

Phone: 1-814-353-8000; 1-800-676-6232

Fax: 1-814-353-8007

Website: www.cannoninstrument.com

Email: Sales: [email protected] Service: [email protected]

Page 5: CAV® 4.2 Operator's Manual - CANNON Instrument

iii CANNON®

CAV® 4.2 Operator's Manual

Contents Overview........................................................................................................................... 1

Related Documents ...................................................................................................................... 1 Specifications / Compliances ........................................................................................................ 2 Notes/Cautions/Warnings ............................................................................................................. 3 Safety Precautions ........................................................................................................................ 3

Getting Started ................................................................................................................. 5 Unpack and Inspect ...................................................................................................................... 5 Setup the CAV 4.2 ........................................................................................................................ 8

Connect Hoses ........................................................................................................................ 10 Fill Baths .................................................................................................................................. 13 Connect Cables ....................................................................................................................... 16

Set Bath Temperature (CAV 4.2 GUI) ........................................................................................... 17 Calibrate Temperature (CAV 4.2 GUI) ......................................................................................... 18 Set Vacuum and Pressure (CAV 4.2 GUI) ..................................................................................... 21 Adjust the Default Wash Settings (CAV 4.2 GUI) ........................................................................ 23 Run a Wash (CAV 4.2 GUI) .......................................................................................................... 24 Train Sensors (CAV 4.2 GUI) ....................................................................................................... 25 Run a Sample Test (CAV 4.2 GUI) ............................................................................................ 26 Review and Export Data (CAV 4.2 GUI) ........................................................................................27

Connecting with VISCPRO II ............................................................................................. 28 Connect Equipment .................................................................................................................... 29 Install VISCPRO II ........................................................................................................................ 29 Set Up Network Adapter ............................................................................................................. 31 Configure Firewall Rules .............................................................................................................. 33 Configure USB Power Settings ....................................................................................................35 CAV Registry Settings ................................................................................................................. 36

Operation ....................................................................................................................... 38 CAV 4.2 GUI ................................................................................................................................ 38 VISCPRO® II Software Interface ................................................................................................. 42 Manage User Accounts (VISCPRO II / CAV 4.2 GUI) .................................................................... 44

Page 6: CAV® 4.2 Operator's Manual - CANNON Instrument

iv CANNON®

Log In ...................................................................................................................................... 44 Change Password ................................................................................................................... 45 Setup and Edit User Accounts ................................................................................................. 45 Logout .................................................................................................................................... 45

Connect VISCPRO II to the CAV 4.2 ............................................................................................ 46 Set Bath Temperature (VISCPRO II) ............................................................................................ 46 Calibrate Temperature (VISCPRO II) ........................................................................................... 47 Set Vacuum and Pressure (VISCPRO II) ....................................................................................... 47 Train Sensors (VISCPRO II) .......................................................................................................... 48 Run a Wash (VISCPRO II)............................................................................................................. 48 Run a Sample Test (VISCPRO II) .................................................................................................. 49 View Test Results (VISCPRO II) ................................................................................................... 52 Control Preheaters ...................................................................................................................... 52 Control Sample Cup Heaters ....................................................................................................... 53

Advanced Setup .............................................................................................................. 55 Calibrate Viscometer Tube (VISCPRO II) ..................................................................................... 55 Method/Test Settings ................................................................................................................. 56

Print Method Settings ............................................................................................................. 56 Edit method settings ............................................................................................................... 56

Adjust Default Wash Settings ..................................................................................................... 60 Create a Standard CAV Data Table Analysis................................................................................ 61

Maintenance and Service ................................................................................................. 64 Update CAV 4.2 Software and Firmware ..................................................................................... 64

Tools and Materials Required .................................................................................................. 64 Prepare the USB Flash Drive ................................................................................................... 64 Replace the UI Updater ........................................................................................................... 64 Update the UI Software .......................................................................................................... 65 Update the Controller Firmware ............................................................................................. 65 Update the VISCPRO II software ............................................................................................. 66

Change Viscometer Tube ............................................................................................................ 67 Replace Power Supply ................................................................................................................ 68 Adjust Solvent Dosing Chamber Sensors .................................................................................... 69

Warranty ........................................................................................................................ 71 Products Limited Warranty ..........................................................................................................71

Page 7: CAV® 4.2 Operator's Manual - CANNON Instrument

v CANNON®

Reagent and Chemical Warranty ................................................................................................. 71 Returning a Product to CANNON ................................................................................................ 71

Required Information ..............................................................................................................72 Hazardous Materials ................................................................................................................72 Shipping Notification ...............................................................................................................72

Replacement Parts List..................................................................................................... 73

List of Figures Figure 1 — Complete CAV 4.2 System .............................................................................................. 8 Figure 2 — Tubing connections ....................................................................................................... 10 Figure 3 — Solvent inlets ................................................................................................................. 11 Figure 4 — Temperature baths ....................................................................................................... 14 Figure 5 — Temperature bath details ............................................................................................... 15 Figure 6 — Bath top (fill hole) ......................................................................................................... 16 Figure 7 — Main screen .................................................................................................................... 17 Figure 8 — Temperature settings screen Figure 9 — Field entry screen ....................................... 18 Figure 10 — Thermometer access holes .......................................................................................... 19 Figure 11 — Temperature calibration screen ................................................................................... 19 Figure 12 — Bath screen Figure 13 — Services menu .................................................................... 21 Figure 14 — Pressure/vacuum adjustment screen ........................................................................... 22 Figure 15 — Settings menu Figure 16 — Wash screen .................................................................. 23 Figure 17 — Main screen (wash button)........................................................................................... 24 Figure 18 — Train screen ................................................................................................................. 25 Figure 19 — Methods screen Figure 20 — Field entry screen ........................................................ 26 Figure 21 — Closed network settings .............................................................................................. 28 Figure 22 — Network connections ................................................................................................... 31 Figure 23 — Connection properties .................................................................................................. 31 Figure 24 — TCP/IP properties Figure 25 — Advanced settings .................................................... 32 Figure 26 — Main Menu Screen ...................................................................................................... 38 Figure 27 — Main screen ................................................................................................................. 40 Figure 28 — Bath screen ................................................................................................................. 41 Figure 29: VISCPRO Interface Window ........................................................................................... 42 Figure 31 — Solvent Dosing Chamber ............................................................................................. 69

Page 8: CAV® 4.2 Operator's Manual - CANNON Instrument

vi CANNON®

List of Tables Table 1 — Related Documents .......................................................................................................... 1 Table 2 — Specifications and compliances ........................................................................................ 2 Table 3 — Packing list (main components) ........................................................................................ 5 Table 4 — Packing list (installation) .................................................................................................. 5 Table 5 — Packing list (tools) ............................................................................................................ 6 Table 6 — Packing list (consumables) ............................................................................................... 7 Table 7 — Packing list (networking) .................................................................................................. 7 Table 8 — Packing list (spares) .......................................................................................................... 7 Table 9 — Tools and materials required ............................................................................................ 9 Table 10 — Silicone bath fluid selection .......................................................................................... 13 Table 11 — Tools and materials required ......................................................................................... 18 Table 12 — Tools and materials required ........................................................................................ 21 Table 13 — Tools and materials required ......................................................................................... 24 Table 14 — Tools and materials required ........................................................................................ 25 Table 15 — Tools and materials required ......................................................................................... 26 Table 16 — Tools and materials required ........................................................................................ 29 Table 17 — VISCPRO II registry entries ............................................................................................ 30 Table 18 — CAV 4.2 registry entries ................................................................................................ 36 Table 19 — GUI Screen Details ........................................................................................................ 39 Table 20 — VISCPRO Window Details ............................................................................................. 43 Table 21 — Default user accounts ................................................................................................... 44 Table 22 — Preheater Settings........................................................................................................ 52 Table 23 — Sample Cup Heater Settings (may vary between instruments, this is a starting point) .. 53 Table 24 — Method Settings: Test .................................................................................................. 57 Table 25 — Method Settings: Wash ................................................................................................ 57 Table 26 — Method Settings: Advanced ......................................................................................... 58 Table 27 — Method Settings: Drop Criteria ..................................................................................... 60 Table 28 — CAV Data Table Fields .................................................................................................. 61 Table 30 — CAV 4.2 Replacement Parts .......................................................................................... 73

Page 9: CAV® 4.2 Operator's Manual - CANNON Instrument

vii CANNON®

Page 10: CAV® 4.2 Operator's Manual - CANNON Instrument
Page 11: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Overview 1

CAV® 4.2 Operator's Manual

Overview The CANNON CAV 4.2 is an automated, high throughput, dual-bath kinematic viscometer for ASTM D445. The CAV 4.2 combines CANNON® quality and reliability with modern design and unique features to enhance lab productivity, reduce costs, and improve data quality.

Related Documents

Table 1 — Related Documents

Title Reference Number

Standard Test Method for Kinematic Viscosity of Transparent and Opaque Liquids (and Calculation of Dynamic Viscosity)

ASTM D445

Standard Specifications and Operating Instructions for Glass Capillary Kinematic Viscometers

ASTM D446

Petroleum products -- Transparent and opaque liquids -- Determination of kinematic viscosity and calculation of dynamic viscosity

ISO 3104

Glass capillary kinematic viscometers -- Specifications and operating instructions

ISO 3105

Petroleum products -Transparent and opaque liquids - Determination of kinematic viscosity and calculation of dynamic viscosity

IP 71

Page 12: CAV® 4.2 Operator's Manual - CANNON Instrument

2 | CAV 4.2 Operation Manual

Specifications / Compliances

Table 2 — Specifications and compliances

Specifications Details

Model CAV 4.2

Methodology ASTM D445/D446, ISO 3104/3105, IP 71

Applications Formulated oil analysis, hydraulic oil analysis, additive analysis, marine fuel testing, base stock analysis, light and heavy fuel testing, waxes/paraffin, crude oil testing, glycols

Dimensions (W × D × H) 36 cm × 66 cm × 72 cm (14.25 in × 26 in × 28.5 in)

Weight 63 kg (140 lb)

Operational temperature range

20 °C to +100 °C (± 0.01 °C); without a TE bath cooler, the minimum temperature is +5 °C above ambient temperature Up to 150 °C (± 0.03 °C) with integrated high temperature bath Down to 15 °C (± 0.03 °C) with integrated TE bath cooler

Maximum throughput 24 tests per hour

Automated sample capacity 28 positions (2 × 14 positions)

Minimum sample/solvent volume required

8 mL sample/15 mL solvent per test

Viscosity range 0.5 mm²/s to 10,000 mm2/s in 100-fold increments (depending on viscometer tubes selected). Fast run tubes are also available.

Timing resolution 0.01 seconds (timing accuracy to ± 0.001 seconds)

Power requirements 100 VAC, 50/60 Hz, 1200 W 115 VAC, 50/60 Hz, 1200 W 230 VAC, 50/60 Hz, 1200 W

Operating conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II, Pollution Degree 2

Compliance CE Mark: EMC Directive (2004/108/EC); Low Voltage Directive (2006/95/EC); ROHS

Page 13: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Overview 3

Notes/Cautions/Warnings Notes, caution, and warnings are used in the manual to call an operator’s attention to important details prior to performing a procedure or step. Read and follow these important instructions. Failure to observe these instructions may void warranties, compromise operator safety, and/or result in damage to the CAV 4.2.

Notes provide more information about the content that follows.

Cautions alert the operator to conditions that may damage equipment.

Warnings alert the operator to conditions that may cause injury.

Safety Precautions Please observe the following general safety precautions for proper and safe operation of the CAV 4.2.

• Only qualified personnel should operate the CAV 4.2.

• Make sure that you read and understand all operating instructions and safety precautions listed in this manual before installing or operating your unit. If you have questions regarding instrument operation or documentation, contact CANNON Instrument Company.

• Do not deviate from the installation, operation, or maintenance procedures described in this manual. Improper use of the CAV 4.2 may result in a hazardous situation and may void the manufacturer’s warranty.

• Handle and transport the unit with care. Sudden jolts or impacts may cause damage to components.

• Observe all warning labels. Never remove warning labels.

• Never operate damaged or leaking equipment.

Page 14: CAV® 4.2 Operator's Manual - CANNON Instrument

4 | CAV 4.2 Operation Manual

• Unless procedures specify otherwise, always turn off the unit and disconnect the mains cable from the power source before performing service or maintenance procedures, or before moving the unit.

• Refer all service and repairs to qualified personnel.

~MAINS

The ~MAINS symbol indicates the connections for the AC power supply. The AC power input must match the electrical specifications of the instrument.

Never operate the equipment with a damaged MAINS AC power cable.

Use only the manufacturer-supplied MAINS AC power cable. This cable must be inserted into a receptacle with a protective earth ground.

(O) The (O) symbol indicates the OFF position for the electrical switches for your unit.

Page 15: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 5

CAV® 4.2 Operator's Manual

Getting Started This chapter covers all steps necessary to set up, calibrate, and run a sample on the CAV 4.2. Procedures are intended to be performed in the order written. These are basic, simplified procedures that will familiarize an operator with the instrument. More detailed instructions are provided in other chapters.

Unpack and Inspect Unpack and inspect the complete CANNON CAV 4.2 and all accessories as soon as they are received.

Caution: Many components are fragile. Use caution when moving and handling the CAV 4.2 and accessories.

1. Carefully unpack the entire contents of the shipment.

2. Referring to the packaging list, verify that all materials were received.

3. Inspect materials for defects and shipping damage. Contact your CANNON representative to resolve any issues.

Table 3 — Packing list (main components)

Description Part Number

CAV 4.2 68.0229

Sample carousel (2) 68.0049 for use with glass vials 68.0092 for use with metal vials

MAINS AC power cable Varies by locale

Dostmann digital thermometer 61.5600

Thermometer probes (2) 61.5601

CAV 4.2 Operator’s Manual 68.0505

Table 4 — Packing list (installation)

Page 16: CAV® 4.2 Operator's Manual - CANNON Instrument

6 | CAV 4.2 Operation Manual

Description Part Number So

lven

t

1⁄8" blue FEP tubing (25 ft) 81.2185

Brass tee fittings (2) 81.0007

Bottle caps for GL-38 bottles (4) 65.3201

20 micron filters (4) 65.0047

Was

te

1⁄2" white nylon tubing (15 ft) 62.1500.3

Brass elbow 68.0309 1⁄2" white nylon elbows (2) 62.1500.2 1⁄2" white nylon tees (2) 62.1500.4

Exhaust 6 mm blue FEP tubing (15 ft) 81.2102

Table 5 — Packing list (tools)

Description Part Number 1⁄8" Allen wrench 27.1340 3⁄16" Allen wrench 68.0464

T-wrench 81.2222

Glass funnel 81.3112

Rubber stoppers (3) 11.3115

Tubing cutter 81.1074

Yellow Tygon tubing (2 ft) 71.4.16

Page 17: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 7

Table 6 — Packing list (consumables)

Description Part Number

Sample test oils varies

Glass vials (144 vials) (optional) Metal sleeves (14)

9717-V01: glass vials 68.0455: metal sleeves

Silicone bath fluid (2 L) Varies by application

Table 7 — Packing list (networking)

Description Part Number

CAT cables, red (2) 61.3378

USB to Ethernet adapter 68.0496

8-port network switch 68.0495

USB cable 68.0508

USB flash drive 17.5121

Table 8 — Packing list (spares)

Description Part Number

6 A fuses (4) 28.0851

12 A fuses (4) 25.2455

Thermal fuse 68.0439

Hose fittings (2) 81.2101

Compression connectors (4) 81.2096

Brass 3⁄8" to 1⁄4" reducer fittings (2) 61.3566

O-rings (4) 61.3567

Page 18: CAV® 4.2 Operator's Manual - CANNON Instrument

8 | CAV 4.2 Operation Manual

Setup the CAV 4.2 Initial setup of the CAV 4.2 involves connecting the hoses and cables and filling the baths with silicone bath fluid.

Figure 1 — Complete CAV 4.2 System

Page 19: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 9

Table 9 — Tools and materials required

Description Hose connections

Complete CAV 4.2 1⁄8" blue FEP tubing (25 ft)

MAINS AC power cable 1⁄2" white nylon tubing (15 ft)

Waste receiver Bottle caps for GL-38 bottles (4)

Solvent 20 micron filters (4) 7⁄8" wrench 3⁄16" Allen wrench

Glass funnel Rubber stopper

Silicone bath fluid (2 L) Tubing cutter

Glass or metal vials

Optional

Networking Brass tee fittings (2)

CAT cables, red (2) Brass elbow

USB to ethernet adapter 1⁄2" white elbows (2)

8-port network switch 1⁄2" white tees (2)

6 mm blue FEP tubing (15 ft)

Page 20: CAV® 4.2 Operator's Manual - CANNON Instrument

10 | CAV 4.2 Operation Manual

Connect Hoses Assemble the CAV 4.2 on a secure, level surface with sufficient room behind the instrument to maintain a clear airflow and allow for unimpeded connection of the solvent inlet and waste outlet tubing.

Figure 2 — Tubing connections

Solvent inlets

Waste outlets

Exhaust ports

Page 21: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 11

1. Place the CAV 4.2 on a solid, level workbench. The workbench must be located near a MAINS AC power connection capable of providing sufficient current (10 A at 115 VAC) to the CAV 4.2. Depending on the materials tested, solvent used, and laboratory regulations, the work area may also need to be equipped with a ventilation system.

2. Place the waste receiver beneath the CAV 4.2. In order to promote proper drainage, the top of the waste receiver must be lower than the waste outlets on the back of the instrument.

3. Place the solvent bottles alongside or beneath the CAV 4.2.

4. Measure and cut the 1⁄8" blue FEP tubing to use as the solvent tubing. Each piece of tubing must run from the solvent inlet well to the bottom of a solvent bottle.

5. Connect the solvent tubing to the CAV 4.2:

a. Remove the cover plate from the solvent inlet well.

b. Thread the tubing through the solvent inlets on the back of the lid.

c. Unscrew the appropriate compression connector and slide it onto the tubing.

d. Slide the tubing onto the flared inlet pipe and hand tighten the compression connector.

Figure 3 — Solvent inlets

6. Insert the solvent tubing into the solvent bottle:

Solvent inlet well

Solvent tubing

Inlet pipes with compression connectors

B A A B

Page 22: CAV® 4.2 Operator's Manual - CANNON Instrument

12 | CAV 4.2 Operation Manual

a. Thread the tubing through the bottle cap.

b. Screw a 20 micron filter onto the end of the tubing.

c. Screw the bottle cap onto the solvent bottle.

7. Repeat steps 4 through 6 for each piece of solvent tubing.

Note: Each piece of tubing must run from the middle of the CAV 4.2 into the waste receiver. 1⁄2" white nylon elbows and tees are provided to enable combining both waste tubes into one tube before entering the waste receiver. A brass elbow is also provided to connect the waste tubing to the waste receiver.

8. Measure and cut the 1⁄2" white nylon tubing to use as the waste tubing.

Caution: Do not overtighten the waste outlet compression nut.

9. For each bath, connect the waste tubing to the CAV 4.2:

a. Unscrew the brass compression nut from the waste outlet fitting.

b. Thread the tubing through the compression nut.

c. Slide the ferrule onto the tubing. Position it 1⁄2" from the end of the tubing.

d. Insert the tubing into the outlet fitting until the ferrule is seated.

e. Tighten the compression nut by hand.

f. Using a 7⁄8" wrench, tighten the compression nut to create a good seal.

10. Connect the waste tubing to the waste receiver.

11. Measure, cut, and install the 6 mm blue FEP tubing as exhaust tubing.

Page 23: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 13

Fill Baths Each temperature bath must be filled with Dow Corning 200 silicone fluid prior to operation. The viscosity of the fluid depends upon the temperature of the bath during most operations:

Table 10 — Silicone bath fluid selection

Bath Operating Range Recommended Bath Fluid Viscosity

20 °C to 100 °C 10 cSt

101 °C to 150 °C 20 cSt

Page 24: CAV® 4.2 Operator's Manual - CANNON Instrument

14 | CAV 4.2 Operation Manual

Figure 4 — Temperature baths

Warning: Follow all lab policies regarding the safe handling and disposal of all hazardous materials.

1. Open the plumbing lid.

2. Raise the display hood by pulling it forward, up, and then back in an arcing motion until the hood rests on top of the frame. Once raised, the latches will keep the hood in position.

3. Insert the 3/16" Allen wrench into each drain plug access hole and verify that the drain plug is closed (clockwise to close).

Left bath Right bath

Display hood

Plumbing lid

Page 25: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 15

Figure 5 — Temperature bath details

4. Insert the funnel into the thermometer hole in the top of the bath to be filled.

5. Slowly pour a small amount of silicone fluid into the bath. Check for leaks. Continue filling until the bath fluid reaches the cold fill level (marked with a thin black/red line on the glass), approximately 0.9 L.

6. Repeat steps 3 through 5 for the second bath.

Cold fill level

Drain plug access holes

Page 26: CAV® 4.2 Operator's Manual - CANNON Instrument

16 | CAV 4.2 Operation Manual

Figure 6 — Bath top (fill hole)

Connect Cables

1. Connect the MAINS AC power cord to the CAV 4.2.

2. Plug the power cord into the MAINS AC outlet at the work station.

3. (optional) Connect the CAV 4.2 to the network switch with one CAT cable.

4. (optional) Connect the network switch to the PC with a CAT cable and USB adapter.

Fill hole

Page 27: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 17

Set Bath Temperature (CAV 4.2 GUI) Set the temperature of the baths before calibrating the instrument.

1. From the main screen, click the target temperature button for the appropriate bath. The temperature settings screen opens.

Figure 7 — Main screen

2. Click the Target Temperature field. A field entry screen opens.

Target temperature

Current temperature

Target temperature

Current temperature

Page 28: CAV® 4.2 Operator's Manual - CANNON Instrument

18 | CAV 4.2 Operation Manual

Figure 8 — Temperature settings screen Figure 9 — Field entry screen

3. Enter the target temperature, then click the green checkmark to submit it and return to the temperature settings screen.

4. Click Save and then Back to return to the main screen.

5. Wait for the current temperature to match the target temperature.

6. Repeat steps 1 through 5 for the second bath.

Calibrate Temperature (CAV 4.2 GUI) The temperature calibration of the CAV 4.2 should be verified whenever the instrument is moved, bath fluid is refilled, or a new temperature is to be used. The instrument is calibrated by inserting a thermometer through an access hole in the display hood, measuring the actual temperature of the bath fluid, and entering that temperature into the CAV 4.2 software.

Table 11 — Tools and materials required

Description

Assembled CAV 4.2

Dostmann digital thermometer

Thermometer probe

1. If it is raised, lower the display hood and plumbing lid.

Page 29: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 19

2. Carefully insert a thermometer through the access hole in the display hood and into the bath until the tip of the thermometer is in the middle of the capillary portion of the viscometer tube. The thermometer passes through the same hole in the top of the bath unit that was used to fill the unit with fluid.

Figure 10 — Thermometer access holes

3. Wait for the thermometer to adjust to the temperature of the bath fluid. Read the reference temperature.

4. From the main screen, click the current bath temperature button for the appropriate bath. The temperature calibration screen opens.

5. Click the Reference Temperature field. A field entry screen opens.

6. Enter the reference temperature measured in step 3, then click the green checkmark to submit it and return to the temperature calibration screen.

7. Click Calibrate, wait for the system to save the adjustment, and then click Back to return to the main screen.

8. Remove and clean the thermometer.

9. Repeat steps 2 through 8 for the second bath.

Thermometer access holes

Figure 11 — Temperature calibration screen

Page 30: CAV® 4.2 Operator's Manual - CANNON Instrument

20 | CAV 4.2 Operation Manual

Page 31: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 21

Set Vacuum and Pressure (CAV 4.2 GUI) Set the vacuum and pressure of the CAV 4.2 before running the instrument.

Table 12 — Tools and materials required

Description

Assembled CAV 4.2

Rubber stoppers

1. From the main screen, swipe the screen with left or right to move to the appropriate bath screen.

Figure 12 — Bath screen Figure 13 — Services menu

Page 32: CAV® 4.2 Operator's Manual - CANNON Instrument

22 | CAV 4.2 Operation Manual

2. On the bath screen, click on the Service button. The services menu opens.

3. Click Adjust Pressure/Vacuum. Follow the instructions on the screen.

4. Click Back to return to the bath screen.

5. Repeat steps 1 through 4 for the second bath.

Figure 14 — Pressure/vacuum adjustment screen

Page 33: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 23

Adjust the Default Wash Settings (CAV 4.2 GUI) Verify and adjust the default wash settings before running a wash for the first time.

Note: The CAV 4.2 wash settings are defined in the method settings of the last run test. When you start the CAV 4.2 from a complete shutdown, it has no knowledge of the last run test and uses the default wash settings.

1. From the main screen, swipe the screen with left or right to move to the appropriate bath screen.

2. On the bath screen, click the Settings button. The settings menu opens.

Figure 15 — Settings menu Figure 16 — Wash screen

3. Click Default Wash. The Wash screen opens.

4. Verify and adjust the wash values as necessary. To change the wash value, you must be logged onto the CAV 4.2 with appropriate privileges.

5. Click Save when you are done adjusting the settings.

6. Click Back to return to the bath screen.

7. Repeat steps 1 through 6 for the second bath.

Page 34: CAV® 4.2 Operator's Manual - CANNON Instrument

24 | CAV 4.2 Operation Manual

Run a Wash (CAV 4.2 GUI) Wash solvent through the CAV 4.2 prior to training sensors or running a test.

Table 13 — Tools and materials required

Description

Assembled CAV 4.2

Sample vials

1. Insert an empty vial into position 1 of the carousel for the bath to be washed.

2. From the main screen, click the wash button for the appropriate bath. The instrument immediately starts the wash cycle.

Figure 17 — Main screen (wash button)

3. When the wash cycle is complete, repeat steps 1 and 2 for the second bath.

Wash button Wash button

Page 35: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 25

Train Sensors (CAV 4.2 GUI) Train the sensors whenever a CAV 4.2 bath is to operate at a new temperature. Once trained, the calibration values for a particular temperature are saved and will be recalled if the CAV 4.2 is set to run at that temperature again. Retraining for a particular temperature is only necessary when the viscometer tube is changed or if sensor performance changes over time.

Table 14 — Tools and materials required

Description Description

Assembled CAV 4.2 Sample test oil

Sample vials Sample preparation materials

Notes: The ubbelohde viscometer tubes in each bath may have different viscosity limits. Temperature calibration must be performed first.

1. Insert an empty vial into position 1 of the carousel for the bath to be trained.

2. Fill a vial with a test solution that has a viscosity approximately 3 times higher than the lower viscosity limit of the ubbelohde tube. Insert the vial into position 2.

3. From the main screen, click the tube button for the appropriate bath. The train screen opens.

4. Click Start Training.

5. When training is complete, repeat steps 1 through 4 for the second bath.

Note: Do not check the Prewash or Trip Points Only boxes.

Figure 18 — Train screen

Page 36: CAV® 4.2 Operator's Manual - CANNON Instrument

26 | CAV 4.2 Operation Manual

Run a Sample Test (CAV 4.2 GUI) This test verifies system calibration and provides a technician with basic training in operating the CANNON CAV 4.2.

Table 15 — Tools and materials required

Description Description

Assembled CAV 4.2 Sample test oil

Sample vials Sample preparation materials

1. Following your laboratory practices, prepare a 10 mL sample of test oil in the provided sample vials, and then insert the sample vial into position 1 of the appropriate carousel.

2. From the main screen, click the circle representing position 1 on the appropriate bath. The methods screen opens.

Figure 19 — Methods screen Figure 20 — Field entry screen

3. Click Verify Known KV. The field entry screen open.

4. Enter the sample ID, then click the green checkmark to submit it. Another field entry screen opens.

5. Enter the viscosity value (in cSt) of the test oil, then click the green checkmark to submit it and return to the main screen.

6. Click the run button .

Page 37: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Getting Started 27

7. When the test is complete, click the Results button to review the test results.

Review and Export Data (CAV 4.2 GUI) Once a test completes, the results can be exported to a USB drive in CSV format.

1. From the main screen, click the Results button.

2. Review the results.

3. Click the Export button. A field entry screen opens.

4. Enter the sample filename, then click the green checkmark to submit it. A Save As window opens.

5. Select the location to which to save the data file, then click Save.

Page 38: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Connecting with VISCPRO II 28

CAV® 4.2 Operator's Manual

Connecting with VISCPRO II The CAV 4.2 ships from CANNON ready for deployment on a dedicated, closed network.1 The following graphic displays the default settings for the closed network.

Figure 21 — Closed network settings

1 For other network configurations, please contact the CANNON service department.

Page 39: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Connecting with VISCPRO II 29

Connect Equipment All CAV systems and the PC (with USB 2.0 to Ethernet adapter) connect to the network switch using standard CAT 5e/6 cables.

Table 16 — Tools and materials required

Description Description

Assembled CAV 4.2 CAT cables, red (1 + 1 per CAV 4.2)

USB flash drive with VISCPRO II installation software and drivers

USB to Ethernet adapter

PC with Windows® 10 8-port network switch

1. Start and log onto the PC.

2. Plug the USB to Ethernet adapter into a USB 2.0 port on the PC. Windows® will attempt to automatically install the required drivers. If necessary, a copy of the driver is provided on the USB flash drive.

3. Connect a CAT cable from the Ethernet adapter to the network switch.

4. Connect CAT cables from each CAV 4.2 to the network switch.

Install VISCPRO II There are no configurable options in the VISCPRO installation routine.

1. Uninstall any previous version of VISCPRO software.

2. Insert the USB drive into an open USB port on the PC.

3. Run USB Flash Drive > VISCPRO > setup.exe.

Page 40: CAV® 4.2 Operator's Manual - CANNON Instrument

30 | CAV 4.2 Operation Manual

The following registry keys are set to default values during installation of VISCPRO. You cannot edit them from within the VISCPRO software. The only way to change these values is by directly editing the registry.

Caution: Editing registry entries may be hazardous to the health of your Windows® installation.

Table 17 — VISCPRO II registry entries

Registry Key Default value Description

EthernetPortNumber 1982 TCP port on which VISCPRO listens

EthernetServerIPAddress 10.0.1.2 IP address of the Ethernet connection that VISCPRO uses

MaxAllowedInstruments 4 Maximum number of instrument to which VISCPRO will connect. VISCPRO polls for instruments in sequential order of the InstrumentAddress. Once the maximum number of allowed instrument has been reached, it stops polling.

MaxPollResponseWaitTime 2000 (2 seconds)

Maximum time in milliseconds that VISCPRO waits for an instrument response when polling.

MaxResponseWaitTime 20000 (20 seconds)

Maximum time in milliseconds that VISCPRO waits for an instrument response.

Page 41: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Connecting with VISCPRO II 31

Set Up Network Adapter

1. Open Network and Sharing Center on the PC.

2. Click Change adapter settings. The Network Connections window opens. Right-click the device associated with the USB adapter, and rename the connection USB To Ethernet Connection.

Figure 22 — Network connections

3. Right-click the connection and select Properties. The Connection Properties window opens.

Figure 23 — Connection properties

4. Click Configure. The Controller Properties window opens.

5. Select the Power Management tab. Uncheck Allow the computer to turn off this device to save power. Click OK.

6. Right-click the connection and select Properties. The Connection Properties window opens.

Page 42: CAV® 4.2 Operator's Manual - CANNON Instrument

32 | CAV 4.2 Operation Manual

7. Uncheck Internet Protocol Version 6 (TCP/IPv6).

8. Double-click Internet Protocol Version 4 (TCP/IPv4). The Internet Protocol (TCP/IP) Properties window opens.

Figure 24 — TCP/IP properties Figure 25 — Advanced settings

9. Set the IP address to 10.0.1.2, the subnet mask to 255.255.255.0, and the default gateway to 10.0.1.1, then click OK to close the Internet Protocol (TCP/IP) Properties window.

10. Click OK to close the Properties window.

11. In the top Network Connections menu bar, click Advanced > Advanced Settings.

12. Move the USB to Ethernet Connection to the last connection position. Click OK.

Page 43: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Connecting with VISCPRO II 33

Configure Firewall Rules The Windows® Firewall on the PC on which VISCPRO is installed must be configured to permit inbound, outbound, and ping communications on port 1982 (default).

1. If necessary, close VISCPRO.

2. Open Control Panel > Windows Firewall > Advanced Settings. This opens the Windows Firewall with Advanced Security window.

3. Configure an inbound rule:

• In the upper left corner, click Inbound Rules.

• In the upper right corner under Actions, click New Rule.

• Select Port. Click Next.

• Select TCP and Specific local ports. Enter 1982 in the port field.

• Select Allow the connection. Click Next.

• Check all boxes: Domain, Private, and Public. Click Next.

• In the Name field, enter Viscpro_IN. Click Finish.

4. Configure an outbound rule:

• In the upper left corner, click Outbound Rules.

• In the upper right corner under Actions, click New Rule.

• Select Port and click Next.

• Select TCP and Specific local ports. Enter 1982 in the port field.

• Select Allow the connection. Click Next.

• Check all boxes: Domain, Private, and Public. Click Next.

• In the Name field, enter Viscpro_Out. Click Finish.

5. Configure on inbound rule to respond to ping requests:

• In the upper left corner, click Inbound Rules.

• In the upper right corner, under Actions, click New Rule.

• Select Custom. Click Next.

Page 44: CAV® 4.2 Operator's Manual - CANNON Instrument

34 | CAV 4.2 Operation Manual

• Select All programs and click Next.

• Change Protocol Type to ICMPv4.

• Click Customize.

• Select Specific ICMP types.

• Check Echo Request and click OK. Click Next.

• Select both Any IP address boxes. Click Next.

• Select Allow the connection. Click Next.

• Check all boxes: Domain, Private, and Public. Click Next.

• In the Name field, enter Ping In. Click Finish.

6. Configure on outbound rule to send ping requests:

• In the upper left corner, click Outbound Rules.

• In the upper right corner, under Actions, click New Rule.

• Select Custom. Click Next.

• Select All programs and click Next.

• Change Protocol Type to ICMPv4.

• Click Customize.

• Select Specific ICMP types.

• Check Echo Request and click OK. Click Next.

• Select both Any IP address boxes. Click Next.

• Select Allow the connection. Click Next.

• Check all boxes: Domain, Private, and Public. Click Next.

• In the Name field, enter Ping Out. Click Finish.

7. Close the Windows Firewall with Advanced Security window.

Page 45: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Connecting with VISCPRO II 35

Configure USB Power Settings The Windows® Firewall on the PC on which VISCPRO is installed must be configured to constantly provide power to the USB port for the USB to Ethernet adapter.

Hints:

• Under Control Panel > Power Options, select Change plan settings for your power plan. Select Change advanced power settings:

• Under the Hard Disk >Turn off hard disk after, change the setting to Never.

• Under the USB settings > USB selective suspend setting, change the setting to Disabled.

• Under Device Manager > Universal Serial Bus controllers, double-click on the appropriate USB hub, select the Power Management tab, and uncheck Allow the computer to turn off this device to save power.

Page 46: CAV® 4.2 Operator's Manual - CANNON Instrument

36 | CAV 4.2 Operation Manual

CAV Registry Settings The following registry keys are set up at CANNON prior to shipping and should not need to be adjusted for the initial installation. However, when reconfiguring a network or adding an additional CAV 4.2 to your setup, you may need to change the Instrument Address and Instrument ID to avoid device collision. These values and the Remote Client Name can be editing from within the CAV 4.2 GUI. Other values can only be changed by directly editing the registry.

Caution: Editing registry entries may be hazardous to the health of your Windows® installation.

Table 18 — CAV 4.2 registry entries

Registry Key CAV 4.2 GUI Default value

Description

EthernetPortName Remote Client Name2

10.0.1.2 IP address of the Ethernet connection to which the instrument connects

InstrumentAddress Instrument Address2

1-4 Numeric address for the instrument

Instrument ID Instrument ID2

4 Instrument name

Serial Number Serial Number3

Text string Instrument serial number

EthernetPortNumber Not available

1982 TCP port on which VISCPRO listens

2 On either bath screen, click Settings > Instrument Settings to access these entries. 3 On either bath screen, click Settings > Setup to access this entry.

Page 47: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Connecting with VISCPRO II 37

Page 48: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 38

CAV® 4.2 Operator's Manual

Operation The CAV 4.2 can be operated in either local or remote mode through a pair of software programs. In Local Mode, the CAV 4.2 GUI (graphical user interface) software program controls the instrument, while in Remote Mode a separate computer with the VISCPRO II software program installed controls it. Each of these programs is independent of the other and maintains its own data set, where a data set consists of a run of tests, test results and error logs. Common instrument settings, method templates, and equipment status are shared between the programs.

The operating status of the CAV 4.2 is shown on the left-most button of the main screen menu of the CAV 4.2 GUI. This button toggles between Local Mode and Remote Mode. A green icon indicates that the instrument is connected to VISCPRO II. In Remote Mode, the Users, Results, and Errors action buttons are grayed to indicate that instrument operation cannot be managed from the CAV 4.2 GUI.

Figure 26 — Main Menu Screen

CAV 4.2 GUI The CAV 4.2 GUI is the software program shown on the touch-screen embedded in the display hood of the CAV 4.2. The Getting Started chapter of this manual introduces the operator to the basic usage of the CAV 4.2 GUI. This chapter provides more in-depth instructions.

Page 49: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 39

Table 19 — GUI Screen Details

Area Button Action

Main screen menu

Local / Remote Mode Toggles between control modes.

Users Opens login menu.

Results Opens local results log.

Errors Opens local errors log.

Help Opens help menu.

Bath screen menu

Settings Opens settings menu.

Tube Calibration Opens tube calibration menu.

Methods Opens methods menu.

Services Opens services.

Carousel Positions 1 – 14 Each position displays the title of the sample. Click to configure a sample. Click center of carousel for bath options.

Screen sliders Left / right arrows Slides between screens.

Status

Sensors Displays the status of the bath sensors. Red – unavailable Green – covered; Gray – uncovered Click to train sensors.

TE Cooler Displays the status of the TE cooler. Click to turn on/off the TE cooler.

Current temperature

Displays the current bath temperature. Click to calibrate temperature.

Target temperature

Displays the target temperature. Click to set target temperature.

Activity message Displays current activity.

Preheater temperature Displays current preheater temperature.

Actions Run / Stop Click to start or stop a run.

Page 50: CAV® 4.2 Operator's Manual - CANNON Instrument

40 | CAV 4.2 Operation Manual

Area Button Action

Wash Click to run a wash.

Pause Click to pause a run or wash.

Figure 27 — Main screen

Bath 1

Status Status

Main screen menu

Actions

Bath 2

Act

ions

Carousel Carousel

Slider Slider

Page 51: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 41

Figure 28 — Bath screen

Bath screen menu

Carousel

Slider

Page 52: CAV® 4.2 Operator's Manual - CANNON Instrument

42 | CAV 4.2 Operation Manual

VISCPRO® II Software Interface VISCPRO II is a Windows® software program comprising the general interface application and instrument/analysis-specific modules. Figure 29 shows a typical VISCPRO II interface with an open instrument window, analysis log, application specific status window, and general information status windows.

Figure 29: VISCPRO Interface Window

Page 53: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 43

Table 20 — VISCPRO Window Details

Area Detail Click to open a window for …

VISCPRO II Menu

Main … managing the VISCPRO application.

Analyses … displaying, configuring, and managing analysis logs.

Configure … configuring the settings of attached instruments.

Service … servicing an attached instrument.

Window … managing the open windows.

Help … managing VISCPRO help.

CAV 4.2 Instrument Window Use this window to view the status of the instrument and manage sample test runs.

Analysis Log Displays a specific analysis report.

Information Status Bar

General Click to open the general information status window.

VI Matching Click to open the VI Matching information status window.

<instrument> Click to open a status window for a particular instrument.

Page 54: CAV® 4.2 Operator's Manual - CANNON Instrument

44 | CAV 4.2 Operation Manual

Manage User Accounts (VISCPRO II / CAV 4.2 GUI) VISCPRO II and CAV 4.2 GUI user accounts are independent of each another. The VISCPRO II accounts are associated with the entire software installation and are not instrument specific.

Both VISCPRO II and the CAV 4.2 GUI start with three default accounts:

Table 21 — Default user accounts

Username Privilege level Privileges

CANNON Instrument Company Inc Service Access to all instrument and security parameters. Call CANNON for the current login and password

Manager Manager Access to advanced configuration and reporting options

Unknown Technician Access to operational features required for testing samples and routine operations

VISCPRO II displays the username and privilege level of the active user in the title bar of the main window. If no user is logged in, VISCPRO II displays the default account (Unknown: Technician). The CAV 4.2 GUI does not display login status.

Log In Log in as soon as possible to avoid problems. You can log in as another user even if you are already logged in.

Note: The initial password for the Manager account is blank.

1. For VISCPRO II: click Main > Log In… The Log In window opens. For CAV 4.2 GUI: click Users. The Login Menu window opens.

2. Select the appropriate user. Type the corresponding password.

3. Click OK.

Page 55: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 45

Change Password In order to change a password, you must be logged in as the user whose password you are changing. However, in VISCPRO II a user with a higher privilege level may assign a new password by editing a user account.

1. For VISCPRO II: click Main > Change Password … The Change Password window opens. For CAV 4.2 GUI: click Users > Change Manager Password. Because new accounts cannot be created using the CAV 4.2 GUI, this is the only option.

2. Type your new password into the appropriate fields.

3. Click OK or the green checkmark.

Setup and Edit User Accounts (VISCPRO II only) Service level users can setup and edit manager and technician accounts. Manager level users can setup and edit only technicians. The default account may not be edited.

1. Click Main > Update User Information … The Change User List window opens.

2. Select the appropriate user. To create a new user, type a unique user name.

3. To assign a password to a user, type the password into the appropriate fields.

4. To assign a privilege level, select the appropriate level in the drop-down box.

5. To save changes click Add or Update User.

6. To close the Change User List window, click Done.

7. To delete the user record from the database, click Remove User.

Logout It is not necessary to logout in order to change user accounts. Logout does not close VISCPRO II. It changes the login to the default (Unknown: Technician) account.

1. For VISCPRO II: click Main > Logout. For CAV 4.2 GUI: click Users > Logout.

Page 56: CAV® 4.2 Operator's Manual - CANNON Instrument

46 | CAV 4.2 Operation Manual

Connect VISCPRO II to the CAV 4.2 VISCPRO II and the CAV 4.2 must be set up and communicating prior to attempting to establish a connection. Refer to Connecting with VISCPRO II for instructions.

1. Start VISCPRO II on the PC.

2. On the CAV 4.2 GUI, touch the Local Mode/Remote Mode button to toggle the instrument state. The icon will turn green to indicate that the CAV 4.2 has found VISCPRO II.

3. In the VISCPRO II Menu, click Main > Poll for Instruments. If an instrument is located, an associated message box will be added to the status bar on the bottom of the VISCPRO II window.

Note: CAV 4.2 are grouped according to the Ethernet port number on which they communicate. Since VISCPRO II can only communicate on one Ethernet port at a time, there will be only one group of CAV 4.2 instruments.

4. Click Main > View Instrument and select the CAV 4.2 instrument group. The CAV 4.2 Instrument Window opens.

Caution: When multiple instruments (CAV 2100/2200, miniAV, CAV 4.2) are connected to the same instance of VISCPRO II, verify that all instruments have completed their current operations before closing VISCPRO II.

Set Bath Temperature (VISCPRO II)

1. In the VISCPRO II Menu, select Configure > [instrument group] > [instrument] > Tray Settings: Temperature.

2. Adjust the Temperature Units, Target Temperature, and Temperature Tolerance as necessary.

3. Click Save; wait for the system to save the changes, and then click Close.

4. Wait for the bath to adjust to the new temperature.

Page 57: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 47

Note: Temperature calibration, vacuum and pressure adjustment, and sensor training can easily be performed directly from the CAV 4.2 GUI. Refer to Getting Started for more information.

Calibrate Temperature (VISCPRO II) Calibrate the temperature of the CAV 4.2 whenever the instrument is moved, a new bath temperature is to be used, or the bath fluid is refilled. The instrument is calibrated by inserting a thermometer through an access hole in the display hood, measuring the actual temperature of the bath fluid, and entering that temperature into the CAV 4.2 software.

1. If it is raised, lower the display hood and plumbing lid.

2. Carefully insert a thermometer through the access hole in the display hood and into the bath until the top of the thermometer is in the middle of the capillary portion of the viscometer tube. The thermometer passes through the same hole in the top of the bath unit that was used to fill the unit with fluid.

3. Wait for the thermometer to adjust to the temperature of the bath fluid. Read the reference temperature.

4. In the VISCPRO II Menu, select Service > [instrument group] > [instrument] > Temperature Calibration. The Temperature Calibration window opens.

5. Enter the reference temperature measured in step 3 in the Reference Temperature field.

6. Click Calibrate; wait for the system to save the adjustment, and then click Done to close the winow.

7. Remove and clean the thermometer.

8. Repeat steps 2 through 7 for the second bath if necessary.

Set Vacuum and Pressure (VISCPRO II) Set the vacuum and pressure of the CAV 4.2 before running the instrument.

1. In the VISCPRO II Menu, select Service > [instrument group] > [instrument] > Adjust Vacuum/Pressure.

2. Follow the instructions on the screen to set the vacuum, then the pressure for each bath.

Page 58: CAV® 4.2 Operator's Manual - CANNON Instrument

48 | CAV 4.2 Operation Manual

Train Sensors (VISCPRO II) Train the sensors whenever the CAV 4.2 bath is to operate at a new temperature. Once trained, the calibration values for a particular temperature are saved and will be recalled if the CAV 4.2 is set to run at that temperature again. Retraining for a particular temperature is only necessary when the viscometer tube is changed or if sensor performance changes over time.

Note: Sensor training uses Test and Advanced settings from the default method, but Default Wash settings.

1. Insert an empty vial into position 1 of the carousel for the bath to be trained.

2. Fill a vial with a test solution that has a viscosity approximately 3 times higher than the lower viscosity limit of the ubbelohde tube. Insert the vial into position 2.

3. In the VISCPRO II Menu, select Service > [instrument group] > [instrument] > Train viscometer tube sensors. The Train viscometer tube sensors window opens.

4. (Optional) Check Prewash to run a wash cycle prior to training sensors.

5. (Optional) Check Trip Points Only to skip adjusting the power levels of the sensors.

6. Click Start.

7. When training is complete, repeat this procedure for the second bath if necessary.

Run a Wash (VISCPRO II) When necessary, washes may be run independent of a sample test.

1. In the VISCPRO II Menu, select Service > [instrument group] > [instrument] > Wash viscometer tube.

2. Select the tube to wash.

3. Select whether to use wash parameters from the last sample run or use default wash parameters.

4. The initial cup evacuation can be bypassed by checking the appropriate box.

5. Click Begin Wash.

Page 59: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 49

Run a Sample Test (VISCPRO II) This procedure details how to run and configure various kinematic viscosity tests on the CAV 4.2 using VISCPRO II software.

1. Verify that the CAV 4.2 is running and has achieved steady-state temperature in each bath to be used.

• The status window displays IDLE. During warmup, SEEKING TEMPERATURE will be displayed.

• The current temperature and target temperature icons are green, indicating that the bath has been temperature calibrated at that temperature. If any other condition is indicated, refer to Calibrate temperature.

• The tube sensor icons are all grey, indicating that the sensors are trained and uncovered. If any other condition is indicated, refer to Train sensors.

2. Connect VISCPRO II to the CAV 4.2:

a. Start VISCPRO II on the PC.

b. On the CAV 4.2 GUI, touch the Local Mode/Remote Mode button to toggle the instrument state. The icon will turn green to indicate that the CAV 4.2 has found VISCPRO II.

c. In the VISCPRO II Menu, click Main > Poll for Instruments. If an instrument is located, an associated message box will be added to the status bar on the bottom of the VISCPRO II window.

d. Click Main > View Instrument and select the CAV 4.2 instrument group. The CAV 4.2 Instrument window opens.

Note: Do not overfill the vials or your sample may overflow when the vial is raised into position and the viscometer tube displaces some sample.

3. Follow your laboratory practices to prepare 10 mL samples and load the vials into the carousel. 10 mL is half a vial.

4. Populate the sample run in VISCPRO II:

Page 60: CAV® 4.2 Operator's Manual - CANNON Instrument

50 | CAV 4.2 Operation Manual

a. In the CAV 4.2 Instrument window, click on the tab corresponding to the desired CAV 4.2 tray.

b. Left-click on the first tray position that is populated with a sample. The entire entry highlights blue.

c. Left-click the Sample ID field and enter a short description for the sample. Press Enter. The Sample Action field is populated with the default action ("Method") associated with the viscometer tube.

If the default action is Verify Known KV, the Sample Configuration window opens automatically. Enter the kinematic viscosity value (cSt) of the sample as the Parameter Value of Check Standard KV. When finished, click Save, and then Close.

d. Continue entering data in the remaining positions. Press ESC to stop entering data.

5. To edit or customize the sample action for a particular tray position, right-click anywhere on the row to open the editing menu.

• Shift-left-click—to select a contiguous range of entries.

• Ctrl-left-click—to select multiple entries.

• Configure—enables editing of all values for the selected entry. These localized changes do not affect the default values of the sample action ("Method").

• Sample Actions—selecting a new action (method template) will replace the existing action. All localized changes will be lost.

• Delete—deletes the selected entry. All following positions are decreased one place.

• Insert At—inserts a new entry into the carousel at the active position. All following positions are advanced one place.

• Copy All Selected Samples—copies selected entries to the clipboard.

• Paste Into Selected Samples—pastes all entries that were copied to the clipboard into the carousel, only at the highlighted positions. This is not an "insert" action. Existing entries are overwritten, not moved.

Page 61: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 51

Note: Other actions—several additional actions are available from the right-click editing menu: • Sequential Sample IDs—use this action to populate sequential

Sample ID fields. • Clear All Trays—clears all entries from all trays • Save Tray to Archive—saves the current tray (carousel) to an

archive file. Each saved tray is associated with a specific bath. You cannot save a tray for bath 1 and load it into bath 2. However, you can copy the complete set of entries from one tray to another.

• Restore Tray from Archive—restores a previously saved tray. • Delete Tray from Archive—opens the archive to allow deleting

saved trays.

6. Click Run. The Select Trays window opens.

7. Select the trays to run. Click OK. The CAV 4.2 begins running the selected trays. Additional samples may be added to the trays until the run completes.

• Pause—opens a selection window. Pausing a tray "Now" immediately halts all actions for that tray. If a test is halted prior to the start of the wash cycle, the data for that sample is discarded.

• Resume—opens a selection window to resume operation for selected trays.

• Abort—opens a selection window to abort all activity for selected trays.

8. When each test completes, the measured KV is shown in the Sample Action field. Once the entire test run finishes, the tray pauses. No additional samples may be added to the run. Click Resume to clear all data for the active tray.

Page 62: CAV® 4.2 Operator's Manual - CANNON Instrument

52 | CAV 4.2 Operation Manual

View Test Results (VISCPRO II) Data obtained during sample testing is temporarily displayed in the Sample Action field and permanently stored in the data base, which may be viewed by opening an analysis package.

Tests must finish running before they can be viewed. The entire tray does not need to have finished, but individual results are only available in the database after they have completely finished running.

1. In the VISCPRO II Menu, select Analyses > View Analysis. The Choose Analysis window opens.

2. Open Basic Package and select CAV Data Table.

3. Select the Open Select Analysis Configuration radio button.

4. Select an analysis and click OK.

Control Preheaters The preheaters are associated with the sample test. Preheater settings can be modified by editing the configuration of a specific sample test or editing the test method template.

Table 22 — Preheater Settings

Parameter name Description

Preheat Type No preheat Minimum preheat starts the preheat cycle as soon as the previous sample begins running and runs for at least the set preheat time. Serialized preheat starts the preheat cycle after the previous sample finishes and runs for exactly the set preheat time.

Minimum Preheat Time This is the time that the preheater is applied to the sample.

Preheat Temperature Units Select either Fahrenheit or Celsius.

Preheat Temperature Fahrenheit: 0 °F to 437 °F; Celsius: 0 °C to 225 °C The heater temperature is determined by the Preheat Temperature setting of the sample that is in the preheat position, even if that sample is set to No preheat. Set the temperature to 0 (zero) to turn off the heater.

Page 63: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Operation 53

Control Sample Cup Heaters

Note: You must be logged onto the GUI or into VISCPRO as a manager in order to change tray/tube settings.

The sample cup heaters assist in maintaining sample temperature during a run and in improving the efficiency of the wash cycle. They are associated with the tray/tube settings.

• For VISCPRO II: select Configure > [instrument group] > [instrument] > Tray Settings: Tube. The Tray Settings window opens.

• Select the correct tube, and enter the appropriate number of duty cycle into the Sample Cup Heater Duty Cycle, Parameter Value field. Click Save and then Close.

• For CAV 4.2 GUI: navigate to the appropriate Bath Screen, click Settings > Tube.

• Enter the appropriate number of duty cycle into the Sample Cup Heater Duty Cycle field. Click Save and then Close.

Table 23 — Sample Cup Heater Settings (may vary between instruments, this is a starting point)

Bath temperature

Sample cup heater duty cycle

Bath temperature

Sample cup heater duty cycle

< 40 °C 0 80 °C to 90 °C 60%

40 °C to 50 °C 20% 90 °C to 100 °C 70%

50 °C to 60 °C 30% 100 °C to 110 °C 80%

60 °C to 70 °C 40% 110 °C to 120 °C 90%

70 °C to 80 °C 50% > 120 °C 100%

Page 64: CAV® 4.2 Operator's Manual - CANNON Instrument
Page 65: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Advanced Setup 55

CAV® 4.2 Operator's Manual

Advanced Setup

Note: You must be logged into the CAV4.2 GUI or VISCPRO as a manager in order to perform advanced procedures.

Calibrate Viscometer Tube (VISCPRO II) Viscometer tube calibration values are associated with specific temperatures, so calibration must be performed when operating a CAV 4.2 bath at a new temperature. Once a tube is calibrated at a particular temperature, the C and E constants are saved by the instrument and will be recalled if that tube is set to run at that temperature again.

1. Assemble a set of standard oils with known kinematic viscosities. Use at least different two standards for each bulb to be calibrated. The CAV 4.2 is designed to run "thin" standards (those with viscosities less than 10X the lower limit of the viscometer tube) in two bulbs, while "thick" standards (those with viscosities greater than 10x the lower limit of the viscometer tube) run in only the lower bulb.

2. Follow the procedures in Run a sample test to prepare, load, and run all of the calibration standards as method Verify Known KV.

3. When all tests have completed, from the VISCPRO II Menu, select Configure > [instrument group] > [instrument] > Tube Calibration. The Tube Calibration window opens.

4. Select the tube and the bulb to calibrate.

5. Select the date range from which to retrieve appropriate test data.

6. Select the sample entries to use for calibration. You must select at least two entries or the software will not recalculate the constants.

• Shift-left-click—to select a contiguous range of entries.

• Ctrl-left-click—to select multiple entries.

7. Click Calculate new constants. The new C and E constants display in their respective fields.

• Evaluate the accuracy of the new calibration constants for each standard by comparing the Known Viscosity with the Calculated Viscosity. This difference is displayed as a percentage

Page 66: CAV® 4.2 Operator's Manual - CANNON Instrument

56 | CAV 4.2 Operation Manual

in the % Difference column. You can repeat the calibration by selecting other sample entries.

• Check Limit E Constant before calculating constants to limit the E constant to no more than 10000.

8. When the new calibration constants are acceptable, click Update bulb to save the C and E constants and calibration data to the local database and update the C & E constants in the registry of the CAV 4.2.

Method/Test Settings Method templates are stored in the registry of the CAV 4.2. When an instance of VISCPRO II connects with a CAV 4.2, those method templates are pushed to VISCPRO II. Any changes made to the method templates in VISCPRO II are synchronized with the CAV 4.2. No method templates are permanently stored with VISCPRO II.

All method templates have the same set of fields. However, when the Verify Known KV method is applied to a sample test, the Check Standard KV value is added to the information set for that test. The Verify Known KV method must be used when running sample test for calibration.

Print Method Settings

• For VISCPRO II: select Configure > [instrument group] > [instrument] > Print Instrument Settings. The Select Methods to Print window opens.

• Select the methods to print. Shift-left-click—to select a contiguous range of entries; Ctrl-left-click—to select multiple entries.

• Click OK. The print dialog box opens.

• For CAV 4.2 GUI: navigate to the appropriate Bath Screen, click Methods.

• Select the method. Click Export, and follow the on-screen instruction to export the method settings as a text file to a USB drive.

Edit method settings

• For VISCPRO II: select Configure > [instrument group] > [instrument] > Method Settings. The Select Methods window opens.

• Select the tube and method to edit.

Page 67: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Advanced Setup 57

• Click OK. The Method Settings dialog box opens.

• For CAV 4.2 GUI: navigate to the appropriate Bath Screen, click Methods.

• Select the method to edit.

Table 24 — Method Settings: Test

Parameter name Description

Wet Tube Runs an initial, unrecorded sample drop to coat the inside of the tube with the sample material.

Soak Time Units Time unit in seconds, minutes, or hours

Bulb1 Soak Time Time to thermal soak a sample that draws only into bulb 1.

Bulb2 Soak Time Time to thermal soak a sample that draws only into bulb 2.

Bulb3 Soak Time Time to thermal soak a sample that draws only into bulb 3.

Preheat Type No preheat Minimum preheat starts the preheat cycle as soon as the previous sample begins running and runs for at least the set preheat time. Serialized preheat starts the preheat cycle after the previous sample finishes and runs for exactly the set preheat time.

Minimum Preheat Time This is the time that the preheater is applied to the sample.

Preheat Temperature Units Select either Fahrenheit or Celsius.

Preheat Temperature Fahrenheit: 0 °F to 437 °F; Celsius: 0 °C to 225 °C The heater temperature is determined by the Preheat Temperature setting of the sample that is in the preheat position, even if that sample is set to No preheat. Set the temperature to 0 (zero) to turn off the heater.

Table 25 — Method Settings: Wash

Setting Description

Bulb1 Solvent A Washes Number of solvent A washes for samples run in only bulb 1

Bulb1 Solvent B Washes Number of solvent B washes for samples run in only bulb 1

Page 68: CAV® 4.2 Operator's Manual - CANNON Instrument

58 | CAV 4.2 Operation Manual

Setting Description

Bulb1 Solvent C Washes Number of solvent C washes for samples run in only bulb 1

Bulb2 Solvent A Washes Number of solvent A washes for samples run in bulbs 1/2

Bulb2 Solvent B Washes Number of solvent B washes for samples run in bulbs 1/2

Bulb2 Solvent C Washes Number of solvent C washes for samples run in bulbs 1/2

Bulb3 Solvent A Washes Number of solvent A washes for samples run in bulbs 1/2/3

Bulb3 Solvent B Washes Number of solvent B washes for samples run in bulbs 1/2/3

Bulb3 Solvent C Washes Number of solvent C washes for samples run in bulbs 1/2/3

Maximum Solvent A Fill Time Max time (sec) to fill the solvent bulb with solvent A

Maximum Solvent B Fill Time Max time (sec) to fill the solvent bulb with solvent B

Maximum Solvent C Fill Time Max time (sec) to fill the solvent bulb with solvent C

Solvent A Injection Time Time (sec) that solvent A will be injected after bulb empties

Solvent B Injection Time Time (sec) that solvent B will be injected after bulb empties

Solvent C Injection Time Time (sec) that solvent C will be injected after bulb empties

Cup Empty Time Time (sec) to empty cup of sample at start of wash cycle

Tube Tip Soak Time Time (sec) for tip to soak after solvent injection

Solvent Evacuation Time Time(sec) to empty cup of solvent after soaking

Final Air Dry Time Time(sec) to force air dry A-Pipe at end of wash cycle

Wash B-Pipe Specify if B-Pipe should be washed

B-Pipe Solvent A Washes Number of solvent A washes for B-Pipe

B-Pipe Solvent B Washes Number of solvent B washes for B-Pipe

B-Pipe Solvent C Washes Number of solvent C washes for B-Pipe

B-Pipe Dry Time Time(sec) to force air dry B-Pipe

Table 26 — Method Settings: Advanced

Page 69: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Advanced Setup 59

Setting Description

Maximum Bulb 1 Fill Time Maximum time (sec) to fill bulb 1

Minimum Bulb 1 Select Time Time (1/10 sec) used to determine whether the sample will run in bulb 1 or 2. Bulb 1 will be selected if the efflux exceeds this time

Minimum Bulb 2 Fill Time Maximum time (sec) to fill bulb 2

Maximum Bulb Empty Time Maximum time (sec) for the sample to drop between the target sensor and the bottom sensor

Tube Empty Time Percentage of the bulb 1 efflux time that is used to empty the tube after the bottom sensor becomes uncovered

Bulb 1 Overdraw Time Time (sec) that the sample is pulled past the target sensor during a bulb 1 sample fill

Bulb 2 Overdraw Time Time (sec) that the sample is pulled past the target sensor during a bulb 2 sample fill

Bulb 3 Overdraw Time Time (sec) that the sample is pulled past the target sensor during a bulb 3 sample fill

Sample Draw Delay Time (sec) after the sample cup is raised before sample is drawn into the tube

Bulb 1 Vent Drain Time Time (sec) for a bulb 1 sample between when the tube vent is opened to atmosphere and the sample efflux starts

Bulb 2 Vent Drain Time Time (sec) for a bulb 2 sample between when the tube vent is opened to atmosphere and the sample efflux starts

Bulb 3 Vent Drain Time Time (sec) for a bulb 2 sample between when the tube vent is opened to atmosphere and the sample efflux starts

Page 70: CAV® 4.2 Operator's Manual - CANNON Instrument

60 | CAV 4.2 Operation Manual

Table 27 — Method Settings: Drop Criteria

Setting Description

Maximum Determinations Maximum number of efflux determinations that will be run when trying to match a set

Required Determinations Number of valid efflux determination needed to produce a matched set

Maximum % Difference Maximum percent difference between efflux determinations in order to be considered for inclusion in a matched set

Maximum Efflux Difference Maximum efflux difference between efflux determinations in order to be considered for inclusion in a matched set

Drop Criteria Mode Determines how the efflux determination will be selected for inclusion in a matched set

Use Percent Difference Stipulates that a positive match between two determination is dependent upon the percent difference between the two

Use Efflux Difference Stipulates that a positive match between two determinations is dependent upon the efflux difference between the two

Use Either If 'Use Percent Difference' and 'Use Efflux Difference' are both set, then, if true, a match between two determinations is made if either criteria is met; else if false, both criteria must be met to produce a matched set.

Adjust Default Wash Settings

• For VISCPRO II: select Configure > [instrument group] > [instrument] > Tray Settings: Default Wash. The Tray Settings: Default Wash window opens.

• Select the tube to edit.

• For CAV 4.2 GUI: navigate to the appropriate Bath Screen, click Settings > Default Wash.

Page 71: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Advanced Setup 61

Create a Standard CAV Data Table Analysis An analysis is used to view detailed data stored in the VISCPRO II database.

1. Select Analysis > View Analysis. The Choose Analysis window opens.

2. Open Basic Package.

3. Click on CAV DATA Table.

4. Select Define and Open New Configuration. Click OK. The CAV Analysis Configuration window opens.

5. On the Date Filter tab, set the appropriate range for the data to be displayed.

6. On the Sample Filter tab, set the filters on the type of data to be displayed.

7. On the Report tab, select the data items to display:

Table 28 — CAV Data Table Fields

Setting Description

Sample ID Descriptive name given to the sample

INV Invalid test indication

TOL (Tolerance) Acceptable difference in multiple test results for the test

Date Date the test was run

Time Time the test was run

Tube S/N Tube serial number

Tray Bath number

Smple Index Sample position

Bulb Bulb in which the sample ran

Technician The technician logged into VISCPRO when the test was run

Instrument Type Instrument type (CAV, CCS, etc.)

Instrument ID Instrument identification, from Instrument Settings

Page 72: CAV® 4.2 Operator's Manual - CANNON Instrument

62 | CAV 4.2 Operation Manual

Table 28 — CAV Data Table Fields

Setting Description

Comment Displays comments associated with the sample

Method Name Name of the method used to run the sample

KV Average kinematic viscosity of the sample

% KE % contribution of the kinetic energy correction to the measured viscosity

Known KV Value entered for the known KV of the sample

% Diff KV % difference between the measured KV and the known KV

Temp Average bath temperature

T1-T4 Times for the valid effluxes used in the calculations

% Diff (Max efflux time – Min efflux time) / average efflux time

Abs. Diff Max efflux time – Min efflux time

{1,10} Efflux %d Adds the first 10 efflux times to the analysis

{1,20} Efflux %d Adds the first 20 efflux times to the analysis

Avg Efflux Average efflux time calculated using valid effluxes

C C constant of calibration

E E constant of calibration

SUS Saybolt Furol Seconds calculation

Display Kinematic Viscosity Data

Check to display samples not run with Verify Known KV method

Display KV Check Data Check to display sample run with Verify Known KV method

Max KV % Difference When the % Diff KV exceeds this value, an "X" is displayed in the TOL column

8. Click OK. You will be prompted to save the analysis.

• If you choose YES, the analysis will be displayed after it is saved.

Page 73: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Advanced Setup 63

• If you choose NO, the analysis will still be displayed.

Page 74: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Maintenance and Service 64

CAV® 4.2 Operator's Manual

Maintenance and Service Update CAV 4.2 Software and Firmware Software and firmware code updates may occasionally need to be manually installed on the CAV 4.2 user-interface, controller boards, and controlling PCs. There are four types of code that can be installed:

• UI updater: this software is actually the complete UI updater program (UpdateUI.exe) that runs on the CAV 4.2. It is used to install the UI software and firmware updates. It is updated by copying it over the existing UpdateUI.exe.

• CAV 4.2 GUI software: this software is the CAV 4.2 user-interface. To update the CAV 4.2 GUI software, you must be logged out of the user-interface and running the UI updater instead.

• Firmware updates: this code controls the baths, sample handlers, and valve controls. It is installed directly onto the respective controller boards using the UI updater.

• VISCPRO II: this software is installed on the controlling PC.

Tools and Materials Required The following tools and materials are required to update the software and firmware:

• #2 Philips screwdriver • USB flash drive • Software/firmware updates • USB keyboard • USB Type A-Male to Mini B cable

(6ft)

Prepare the USB Flash Drive A USB flash drive must be loaded with the software and firmware to be updated.

1. On a PC, download the update packages from the CANNON website Software & Downloads page.

2. Unpack and copy the update files to the USB flash drive. Only one set of files can be on the USB flash drive during an update.

Replace the UI Updater

1. From the GUI, click Help > Upgrade UpdateUI.

Page 75: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Maintenance and Service 65

2. Plug the prepared USB flash drive into the USB connection.

3. Follow the instructions on the screen.

Update the UI Software

Note: A video of this procedure is available at https://vimeo.com/139739339.

1. From the GUI, click Help > Upgrade Instrument. Click OK. The baths are placed into Bootloader mode (orange lights), and the Windows logon screen displays.

2. Plug the prepared USB flash drive into the USB connection

3. Click Updates. This runs the UpdateUI.exe program, which opens into the update shell.

4. Click Update Software. A File Open dialog opens.

5. Select and run the setup.exe program in the CAV UI update directory on the E: drive.

6. Follow the instructions provided by the setup program.

7. Click Save And Exit. The CAV unit will reboot several times, but the bath lights will remain orange, indicating that they are still in Bootloader mode.

8. After the CAV finishes its rebooting routine, power cycle the unit by completely turning it off with the power switch on the hood, waiting 10 seconds, then turning it on.

Update the Controller Firmware

Warning: Be careful not to touch any moving parts or live wires in the CAV 4.2.

Note: A video of this procedure is available at https://vimeo.com/139737507.

1. Slide the trays out and remove the carousels.

2. With the Philips screwdriver, remove the screws securing the corner panels.

3. Remove the corner panels. Remove the center panel. Raise the lid. Push the trays in.

4. From the GUI, click Help > Upgrade Instrument. A dialog box asks you to confirm your action.

Page 76: CAV® 4.2 Operator's Manual - CANNON Instrument

66 | CAV 4.2 Operation Manual

5. Click OK. The baths are placed into Bootloader mode (orange lights), and the Windows logon screen displays.

6. Click Updates. The update shell opens.

7. Plug the prepared USB flash drive into the USB connection

8. Click Update Firmware.

9. Select the type of controller to update. Windows Explorer opens.

10. Using the GUI touch-screen, select and open the firmware update in the firmware update directory on the E: drive.

11. Remove the USB flash drive.

12. To update a sample handler or valve controller, connect a USB cable between the USB connection and the appropriate controller board. Both bath controllers can be updated without connecting them with a USB cable.

13. Repeat steps 7 through 10 as necessary to update all applicable controller boards.

14. Click Save And Exit.

15. Power cycle the unit by completely turning it off with the power switch on the hood, waiting 10 seconds, then turning it on.

Update the VISCPRO II software Run the VISCPRO II installation program. Follow all instructions in the installation program.

Page 77: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Maintenance and Service 67

Change Viscometer Tube

Warning: Do not drain the silicone bath fluid until it has cooled off to a safe temperature.

1. Turn off the CAV 4.2 and allow the bath to cool down.

2. Drain the silicone bath fluid into a clean container and save for refilling the unit.

3. Raise the hood.

4. Note the solvent line connections. Gently unscrew the solvent lines from the top of the viscometer tube.

5. Disconnect the sensor cable by gently pulling straight up.

6. Unscrew the thumb screws until they spring free of the bath top.

7. Gently draw the viscometer tube straight up and out of the bath.

8. Inspect the new viscometer tube for defects, especially small chips on the top and bottom of the tube. These can cause vacuum leaks and problems with solvent contamination.

9. Clean the bottom of the new tube and apply a light coating of silicone bath fluid.

10. Gently lower the viscometer tube into the bath and press it into the cone of the tube tip in the bottom of the bath. The viscometer tube must pass through 2 o-rings in the cone and seat gently onto the top of the metal tube tip. Failure to seat the viscometer tube properly may result in poor performance.

11. Press the top bracket into position and evenly tighten the thumb screws.

12. Verify that the bath drain set screw is tight and refill the bath with silicone bath fluid.

13. Connect the solvent lines and sensor cable.

14. Turn the CAV 4.2 on.

15. In Tube Settings, enter the name of the new tube.

16. If necessary, adjust and calibrate the temperature of the bath. If the temperature has not changed, temperature calibration is not necessary.

Page 78: CAV® 4.2 Operator's Manual - CANNON Instrument

68 | CAV 4.2 Operation Manual

Note: Sensor training data is associated with the temperature setting of a bath. When a tube is changed and the bath temperature is either kept the same or changed to a temperature at which a tube was previously trained, the CAV will display that the sensors are trained. This is correct. The sensors must be trained whenever a tube is changed.

17. Train the sensors.

18. Verify the Default Wash and Method Settings, especially Verify Known KV, for the bath in which the tube was changed.

19. Calibrate the new viscometer tube.

Replace Power Supply The CAV 4.2 power supply is a switching regulated power supply designed to handle 90 – 240 VAC. To accommodate different locales, only the MAINS AC power cable needs to be changed.

Caution: The power supply is easy to replace. The side panels do not need to be removed to access the power supply attachment point. It is not connected via cable. Instead, a mating coupler is attached to a bulkhead in the chassis.

1. Turn off the CAV 4.2 and unplug it from the MAINS AC power.

2. Remove the power cable.

3. Remove the four screws securing the power supply assembly to the chassis.

4. Firmly pull the power supply assembly from the chassis.

Page 79: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Maintenance and Service 69

Adjust Solvent Dosing Chamber Sensors Each CAV 4.2 bath has its own solvent dosing chamber, mounted on the side panel. Optical sensors on the inlet and the outlet of the dosing chasing must be adjusted to "see" the solvents that are drawn past them.

When the chamber is full of solvent, adjust the sensors by turning the set screws counter-clockwise until the associated green light turns on. When running multiple solvents, verify each solvent, always adjusting by turning the screws counter-clockwise.

Figure 30 — Solvent Dosing Chamber

Top sensor adjustment

Bottom sensor adjustment

Page 80: CAV® 4.2 Operator's Manual - CANNON Instrument
Page 81: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Warranty 71

CAV® 4.2 Operator's Manual

Warranty Products Limited Warranty In addition to other manufacturers’ warranties, CANNON Instrument Company (“the Company”) warrants all products (other than reagents and chemicals) delivered to and retained by their original purchasers to be free from defect in material and workmanship for one year from the date of the Company’s invoice to the purchaser. For a period of one year from the date of such invoice, the Company will correct, either by repair or replacement at the Company’s sole discretion, any defect in material or workmanship (not including defects due to misuse, abuse, abnormal conditions or operation, accident or acts of God, or to service or modification of the product without prior authorization of the Company) without charge for parts and labor. The determination of whether any product has been subject to misuse or abuse will be made solely by the Company.

The Company shall not be liable for any special, incidental, or consequential damages, or any damage to plant, personnel, equipment or products, directly or indirectly resulting from the use or misuse of any product. Representations and warranties made by any person, including dealers and representatives of the Company, which are inconsistent, in conflict with, or in excess of the terms of this warranty shall not be binding upon the Company unless placed in writing and approved by an officer of the Company.

Reagent and Chemical Warranty CANNON Instrument Company (“the Company”) warrants all reagents and chemicals sold by the Company and delivered to and retained by their original purchasers to conform to the weight, specifications and standards stated on the package. The Company will, at its sole option, either replace or refund the price (net of freight, handling charges and taxes), of any reagent or chemical sold by the Company which does not conform to such weight, specifications and standards upon the prompt return of the unused portion. Except for replacement or refund of the net price, the Company shall not be liable for any damages occurring as a consequence of the failure of any reagent or chemical sold by the Company to conform to the weight, specifications and standards stated on the package.

Returning a Product to CANNON Before returning a CANNON product for repair or service, make every attempt to identify the problem. If, after careful checking, the problem remains unidentified or unsolved, telephone

Page 82: CAV® 4.2 Operator's Manual - CANNON Instrument

72 | CAV 4.2 Operation Manual

CANNON Instrument Company (or the local service agent) to consult with a product specialist. If the specialist cannot recommend a simple solution or repair, CANNON will authorize the return of the product through the issuance of a Return Authorization number (RA).

CANNON Telephone Number 814-353-8000

CANNON Fax Number 814-353-8007

Products returned to CANNON must be carefully packed. Ship prepaid to the following address:

CANNON Instrument Company

ATTN: Return Authorization # __________

2139 High Tech Road

State College, PA 16803 USA

Please include the following:

Required Information

• The Return Authorization number (RA).

• The name and telephone number of the person at your company to contact regarding the product.

• Shipping and billing instructions for the return of the product to your location.

• A detailed explanation of the reason for the return. If the product is not covered by warranty, the customer will be provided with an estimate of the repair costs and asked for approval before any repairs are made. The customer will be required to issue a purchase order for the cost of the repairs.

Hazardous Materials Please contact CANNON before returning a product that could possibly contain hazardous material.

Shipping Notification Products returned without CANNON’s prior authorization will not be accepted. The customer may be billed a testing fee if a product is returned to CANNON and found to be working properly.

Page 83: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® | Replacement Parts List 73

CAV® 4.2 Operator's Manual

Replacement Parts List Table 29 — CAV 4.2 Replacement Parts

Description Part Number

Power supply 68.0004

Sample carousel for glass vials 68.0049

Sample carousel for metal vials 68.0092

SBC module 68.0238

MAINS AC power cable 74.2110 (US standard)

Power supply fuses Varies

Surge protector Varies

Page 84: CAV® 4.2 Operator's Manual - CANNON Instrument

74 | CAV 4.2 Operation Manual

Page 85: CAV® 4.2 Operator's Manual - CANNON Instrument
Page 86: CAV® 4.2 Operator's Manual - CANNON Instrument

CANNON Instrument Company® 2139 High Tech Road

State College, PA 16803