Page 1: Ccnpwt user manual

USER HANDBOOK

CLOSED CYCLIC

NEGATIVE PRESSURE WOUND THERAPY

www.npwtindia.com

[email protected]

Page 2: Ccnpwt user manual

CCNPWT provides a Closed Sterile & Moist Wound healing environment

- Reduces bacterial burden

- Reduces oedema

- Enhances dermal perfusion

- Promotes micro angiogenesis

- Stimulates cell proliferation

- Promotes granulation tissue formation.

- Studies suggest mechanical stretching may result in increased mitosis (Cell Replication) Controlled Negative Pressure Therapy decreases wound volume approximating wound edges, helps uniformly draw wounds closed

- Promotes continuous wound debridement by removing wound micro debris & bacteria colonized wound output / interstitial fluid allowing tissue decompression

- Protects wound from re-infection, cross infection or spreading further infection

- Promotes Skin Graft/Flap uptake (The sterile flexible foam dressing gently presses the skin graft or flap ensuring full contact with the wound bed & simultaneously removes wound output ensuring moist healing environment)

Page 3: Ccnpwt user manual

SOFTWARE SETTINGS

Switch on the CCNPWT System, You will see the home screen showing

SELF TEST MESSAGE (SEE: snap shot image 1).

After self test system will try achieving the target pressure set already or during previous therapy session. Pump will stop once the actual pressure equals target pressure, when it will fall again pump will restart, process will continue like this (SEE: snap shot image 2).

To Change pressure settings click the menu button, select the setting as in (SEE: snap shot image 3) by pressing UP / DOWN switch, the < in the left will move up/Down

To select the desired setting press ENTER Switch the arrow will turn >, (SEE: snap shot image 4) this shows the desired setting has been selected, by pressing UP / DOWN switch you can change the pressure or other settings. After doing the changes again press ENTER switch to save and then the MENU switch to exit and restart therapy.

Page 4: Ccnpwt user manual

DISPLAY FRONT PANEL

(VTG1901)

LCD DISPLAY

(4 Lines)

Power Indicator

LED (GREEN)

Battery Indicator RED LED

MENU Switch

DOWN Cursor Switch

UP Cursor Switch

ENTER Switch

Page 5: Ccnpwt user manual

REAR VIEW

VACUUM OUTLET

POWER SOCKET /

MAIN SUPPLY 220V / 50 - 60Hz

MAINS ON/OFF

SWITCH

WITH GLOW LIGHT INDICATOR

Page 6: Ccnpwt user manual

REAR VIEW

BATTERY ON / OFF

SWITCH

PRESS 0 for OFF PRESS I for ON

CANISTER HOLDER SEE IMAGE in NEXT PAGE

MACHINE SERIAL NO. LABEL

CUSTOMER CARE NO.

Page 7: Ccnpwt user manual

ACCESSORIES

INSTALLATION STEP

Open Canister Pack

& hook it in the

canister holder

(As shown in the above

picture)

Open Filter Pack

Connect the female

connector

of filter with the male

vacuum outlet in the

system

Connect the distal end

of the filter silicon tube

with the L Connector of

the canister as shown

in the picture

(push & glide the L

Connector inside the

silicon tube)

Page 8: Ccnpwt user manual

AFTER ALL SETUP

CONNECT THE MALE CONNECTOR FROM CANISTER & FEMALE CONNECTOR FROM

DRESSING/DRAINAGE TUBE TO COMPLETE THE VACUUM CIRCUIT (as shown in fig 2)

Fig. 2

Fig. 1

Fig. 3

CROSS CHECK AFTER

COMPLETE INSTALLATION

Page 9: Ccnpwt user manual

DRESSING INSTALLATION STEPS

Page 10: Ccnpwt user manual

TROUBLE SHOOTING

PRESSURE NOT ACHIEVING TO TARGET PRESSURE

Unplug the filter tube and clamp & see Actual Pressure = Target Pressure if OK

If NOT Check the VACUUM PORT Connector or it could be a machine problem

Proceed to Next Step

Unplug the other end of the filter and clamp & see if Actual Pressure = Target Pressure if OK

If NOT, possibly filter is malfunctioning

Proceed to Next Step

Clamp the Canister tube & see Actual Pressure = Target Pressure if OK

If NOT, possibly canister is leaking or faulty.

Proceed to Next Step

Connect the dressing tube with the canister tube & Clamp the dressing tube & see Actual Pressure = Target Pressure if OK

If NOT, possibly the canister & dressing tube joint is leaking or not properly connected.

then CHECK DRESSING SEAL

Page 11: Ccnpwt user manual

ERROR – NO DISPLAY

Check the MAINS Cable / POWER

Cable is connected to the power

socket & is switched ON.

See the POWER SWITCH in the

CCNPWT System is RED GLOWING

See the BATTERY SWITCH in the

CCNPWT System is ON (PRESSED I)

(ON POWER)GREEN LED

(ON BATTERY) RED LED

IF ONLY RED is ON, that

shows/indicates the system is working

on battery

IF ONLY GREEN is ON, that shows/indicates the system is on MAIN supply

IF BOTH RED & GREEN is ON, that shows/indicates the system is working

on MAINS supply & battery is charging. In absence of MAIN supply power

sytem will automatically switch over to battery standby. (IDEAL CONDITION)

ENSURE the RED & GREEN LED/LIGHT in the front panel or in the left side of

the LCD Display is glowing every time

Page 12: Ccnpwt user manual

ERROR : BLOCKAGE / CANISTER FULL

Ensure the canister gets replaced between 450ml -500ml accumulation of fluid

or this may choke the filters or even can damage the system pump (if continued further or for longer period).

IF canister is not full remove the filter tube & run the system.

IF OK, then connect the filter & CHECK for any blockage in the filter.

IF OK connect the canister without connecting the dressing tube to CHECK any tube blockage or tube folds/kink.

IF OK, there could be drainage tube blockage or mismatch hole with the dressing/drape sealed.

Check with a needle prick in the dressing (Close to Drainage Tube junction point).

OR CONTACT A CCNPWT REPRESENTATIVE / SUPPORT CENTRE

Page 13: Ccnpwt user manual

CCNPWT WOUND DRAINAGE KIT

SMALL (10 x 7.5 x 3 cm) CAT NO. 9005D

MEDIUM (18 x 12.5 x 3 cm) CAT NO. 9004D

LARGE (26 x 15 x 3 cm) CAT NO. 9002D

EXTRA LARGE (60 x 30 x 2 cm) CAT NO. 9007D

ABDOMINAL KIT CAT NO. 9009D

PVA (15 x 10 x 1 cm) CAT NO. 9008D

OTHER CONSUMABLES

DRAINAGE TUBING CAT NO. 9018D

CANISTER 500ml (Disposable) CAT NO. 9010D

with aqua gel pack

FILTER CAT NO. 111D

TRANSPARENT PU DRAPE CAT NO. 9017D

CONTENTS OF DRESSING KIT

DRESSING FOAM

TRANSPARENT PU DRAPE

DRAINAGE TUBING

DISPOSABLE MEASURING SCALE

Page 14: Ccnpwt user manual

ALL CCNPWT CONSUMABLES ARE SINGLE USE

TO ENSURE MAXIMUM PROTECTION & STERILITY

STERILITY : GAMMA RADIATED

SHELF LIFE : 3 YEARS

CERTIFICATIONS : CE / GMP / ISO 13485:2003

24 x 7 CUSTOMER CARE : +91 9650440444

MANUFACTURED BY or ON BEHALF OF

TRIAGE HEALTHCARE TECHNOLOGIES

REGD. OFF.: 17, Aashirwaad Appts.

Prop. No. 886-C/8, Mehrauli

New Delhi-110030 (INDIA)

WORKS: H-41, Shahjahanpur Indl. Area,

Distt. Alwar, Rajasthan (INDIA)

CUSTOMER CARE: +91 9650440444 / 9971722111

Email : [email protected]

www.npwtindia.com

USAGE : STRICTLY BY PHYSICIAN OR WOUND CARE NURSE‘s ADVICE ONLY