cem humox lit 0609 eta
DESCRIPTION
xTRANSCRIPT
-
BULLETIN P-460BULLETIN P-460BULLETIN P-460BULLETIN P-460BULLETIN P-460
CEM/Humox
The CEM/Humox Moisture in Flue Gas Analyzer isdesigned to measure net oxygen and moisturecontent in flue gas and process applications. TheO2 and moisture data is used to correct emissions,assure product quality, or minimize stack corro-sion. The analyzer indicates moisture, oxygen-wetand oxygen-dry concentrations in percent byvolume.
MEASUREMENT OF MOISTURE IN FLUE GAS TOCALCULATE EMISSIONS ON A DRY OR WET BASIS
Principle of Operation:Principle of Operation:Principle of Operation:Principle of Operation:Principle of Operation: Zirconium oxide - onesensor for hot wet sample; one sensor for drysample.
SENSORS
SERIES 2000 CONTROL UNIT
Display: Display: Display: Display: Display: Four-line x 20-character vacuum fluorescent.Displays combinations of % moisture, oxygen (wet or dry),time and date, cell temperature (wet or dry), cooler tempera-ture, user programmable text, thermocouple mV (wet or dry)or cell mV (wet or dry). Password protection, programmablepressure compensation and context-sensitive help are alsoprovided.
Recorder Output: Recorder Output: Recorder Output: Recorder Output: Recorder Output: Three isolated linear current outputs.Each output can be 4-20 mA, 0-20 mA, 20-4 mA or 20-0 mAand is fully scalable. Hold or track during calibration andselect degree of damping. Maximum load 1200 ohms.
Alarms: Alarms: Alarms: Alarms: Alarms: Four independent sensor alarms, each high or lowselectable. One alarm can be allocated to % moisture,calibrate or verify. Set relays to energize or de-energize onalarm.
Contact RatingContact RatingContact RatingContact RatingContact Rating: 1 A, 30V max. noninductive load, AC or DC
Diagnostics: Diagnostics: Diagnostics: Diagnostics: Diagnostics: Watchdog timer and service alarms. Systemtest for A/D, RAM, EEPROM and keypad. Display line 4reserved for full-text error and diagnostic messages.Twenty-entry event log.
Communications: Communications: Communications: Communications: Communications: RS-485 2-way addressable.
Operating RangeOperating RangeOperating RangeOperating RangeOperating RangeOxygen: 0.1% to 100%Moisture: 5% to 85% by volume
AccuracyAccuracyAccuracyAccuracyAccuracyOxygen: 2% of reading or 0.1% absoluteMoisture: 5% of reading or 1% absolute, whichever isgreater
Response Time: Response Time: Response Time: Response Time: Response Time: > 4 seconds at 2 scfh over one decade
Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: 400F (204C)
Sample Flow: Sample Flow: Sample Flow: Sample Flow: Sample Flow: 0.1 to 1.0 L/min (0.6 L/min is recommended)
EnvironmentEnvironmentEnvironmentEnvironmentEnvironmentAmbient Temperature: -18C to 50C (0F to 122F)Relative Humidity: 10% to 90%, noncondensing
Power Requirements: Power Requirements: Power Requirements: Power Requirements: Power Requirements: 115 VAC, 10%, 47-63 Hz; 230 VAC,10%, 47-63 Hz; 313 VA max.
Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Use calibration gases @ 0.1to 1.0 L/min. (0.6 L/min recommended)
Zero Gas: From 0.1 to 10% O2, balance N2Span Gas: Minimum one decade above zero gas(10 times greater)
OPTIONALOPTIONALOPTIONALOPTIONALOPTIONAL
An integrated version is available containing wet and drysensors together with sample conditioning in a singleenclosure which mounts directly to the process.
CEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/O22222sensors and Series 2000 Control Unitsensors and Series 2000 Control Unitsensors and Series 2000 Control Unitsensors and Series 2000 Control Unitsensors and Series 2000 Control Unit
-
wwwwwwwwwwwwwww.thermox.com.thermox.com.thermox.com.thermox.com.thermox.com 2003, by AMETEK, Inc. All rights reserved.
403PDF (P460)
One of a family of innovative process analyzer solutions from AMETEK Process Instruments. Specications subject to change without notice.
PROCESS INSTRUMENTSPROCESS INSTRUMENTSPROCESS INSTRUMENTSPROCESS INSTRUMENTSPROCESS INSTRUMENTS
CANADACANADACANADACANADACANADA USA - DelawareUSA - DelawareUSA - DelawareUSA - DelawareUSA - Delaware USA - PittsburghUSA - PittsburghUSA - PittsburghUSA - PittsburghUSA - Pittsburgh USA - TexasUSA - TexasUSA - TexasUSA - TexasUSA - TexasevirD nrohtihW 0579daoR tropeerF 051.dvlB etaroproC 554.E.N yaW egdirnuS 6782
59077 XT ,notsuoH83251 AP ,hgrubsttiP20791 ED ,kraweN9H7 Y1T BA ,yraglaC0282-364-182 .hP0409-828-214 .hP0044-654-203 .hP0048-532-304 .hP1072-364-182 xaF9930-628-214 xaF4444-654-203 xaF0553-842-304 xaF
CHINACHINACHINACHINACHINA FRANCEFRANCEFRANCEFRANCEFRANCE GERMANYGERMANYGERMANYGERMANYGERMANY MIDDLE EAST - DubaiMIDDLE EAST - DubaiMIDDLE EAST - DubaiMIDDLE EAST - DubaiMIDDLE EAST - Dubai SINGAPORESINGAPORESINGAPORESINGAPORESINGAPORE-56 .hP2502 188-4-179 .hP0 63 19 95 12 94 .hP02 96 86 03 1 33 .hP9407 6246 12 68 .hP 6484-2388
3502 188-4-179 xaF0863 19 95 12 94 xaF92 96 86 03 1 33 xaF4507 6246 12 68 xaF Fax 65-6481-6588
1.483.76
5.2213.26
2.255.72
12.632.0
9.1023.11
19.0048.26
4.9512.57
19" RACK MOUNT
12.030.5
12.030.5
10.025.4
8.1920.8
6.7517.14
1/4" CompressionTube Fitting
Sample Input1/4" CompressionTube Fitting
Sample Output1/4" CompressionTube Fitting
8.0020.32
MountVertically
12.7532.38
2.506.35
3.07.62
3.07.62
1.503.811.25
3.17.31 dia. Holes for
1/4" fasteners
SENSOR
DIMENSIONSINCHES
CM
EnvironmentEnvironmentEnvironmentEnvironmentEnvironmentAmbient Temp: 14 to 122F (-10 to 50C)Relative Humidity: 10% to 80%, noncondensing
Hazardous Area: Hazardous Area: Hazardous Area: Hazardous Area: Hazardous Area: Meets NEC Class I, Division 2 withoutrequiring purge apparatus.
Enclosure:Enclosure:Enclosure:Enclosure:Enclosure: Standard: 19" rack mount. Optional: GP (generalpurpose) wall mount, GP panel mount, weatherproof NEMA 4(IP 56) wall/panel mount, and stainless steel weatherproofNEMA 4X (IP 56) wall/panel mount. All are UL Listed for NECClass I, Division 2 areas.
Calibration:Calibration:Calibration:Calibration:Calibration: Oxygen cell lifetime extender. Calibrate or verifycalibration. Store last calibration and verication data.Selectable calibration gas run time and process recoverytime. Timed automatic calibration with optional remotecalibration unit.
Bias adjustment (2%) for process calibration to EPAMethod 4 if required.
Power Requirements: Power Requirements: Power Requirements: Power Requirements: Power Requirements: Nominal 95-230 VAC,10%, 47-63 Hz; 75 VA max.
System ComplianceSystem ComplianceSystem ComplianceSystem ComplianceSystem ComplianceEMC Directive 89/336/EECDirective 73-23/EEC
CONTROL UNITAPPROX WEIGHT: 14 LB (6.35 KG)
APPROX WEIGHT: 24 LB (10.9 KG)