BULLETIN P-460 BULLETIN P-460 BULLETIN P-460 BULLETIN P-460 BULLETIN P-460 CEM/Humox The CEM/Humox Moisture in Flue Gas Analyzer is designed to measure net oxygen and moisture content in flue gas and process applications. The O 2 and moisture data is used to correct emissions, assure product quality, or minimize stack corro- sion. The analyzer indicates moisture, oxygen-wet and oxygen-dry concentrations in percent by volume. MEASUREMENT OF MOISTURE IN FLUE GAS TO CALCULATE EMISSIONS ON A DRY OR WET BASIS Principle of Operation: Principle of Operation: Principle of Operation: Principle of Operation: Principle of Operation: Zirconium oxide - one sensor for hot wet sample; one sensor for dry sample. SENSORS SERIES 2000 CONTROL UNIT Display: Display: Display: Display: Display: Four-line x 20-character vacuum fluorescent. Displays combinations of % moisture, oxygen (wet or dry), time and date, cell temperature (wet or dry), cooler tempera- ture, user programmable text, thermocouple mV (wet or dry) or cell mV (wet or dry). Password protection, programmable pressure compensation and context-sensitive help are also provided. Recorder Output: Recorder Output: Recorder Output: Recorder Output: Recorder Output: Three isolated linear current outputs. Each output can be 4-20 mA, 0-20 mA, 20-4 mA or 20-0 mA and is fully scalable. Hold or track during calibration and select degree of damping. Maximum load 1200 ohms. Alarms: Alarms: Alarms: Alarms: Alarms: Four independent sensor alarms, each high or low selectable. One alarm can be allocated to % moisture, calibrate or verify. Set relays to energize or de-energize on alarm. Contact Rating Contact Rating Contact Rating Contact Rating Contact Rating: 1 A, 30V max. noninductive load, AC or DC Diagnostics: Diagnostics: Diagnostics: Diagnostics: Diagnostics: Watchdog timer and service alarms. System test for A/D, RAM, EEPROM and keypad. Display line 4 reserved for full-text error and diagnostic messages. Twenty-entry event log. Communications: Communications: Communications: Communications: Communications: RS-485 2-way addressable. Operating Range Operating Range Operating Range Operating Range Operating Range Oxygen: 0.1% to 100% Moisture: 5% to 85% by volume Accuracy Accuracy Accuracy Accuracy Accuracy Oxygen: ± 2% of reading or 0.1% absolute Moisture: ± 5% of reading or ±1% absolute, whichever is greater Response Time: Response Time: Response Time: Response Time: Response Time: > 4 seconds at 2 scfh over one decade Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: 400°F (204°C) Sample Flow: Sample Flow: Sample Flow: Sample Flow: Sample Flow: 0.1 to 1.0 L/min (0.6 L/min is recommended) Environment Environment Environment Environment Environment Ambient Temperature: -18°C to 50°C (0°F to 122°F) Relative Humidity: 10% to 90%, noncondensing Power Requirements: Power Requirements: Power Requirements: Power Requirements: Power Requirements: 115 VAC, ±10%, 47-63 Hz; 230 VAC, ±10%, 47-63 Hz; 313 VA max. Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Use calibration gases @ 0.1 to 1.0 L/min. (0.6 L/min recommended) Zero Gas: From 0.1 to 10% O 2 , balance N 2 Span Gas: Minimum one decade above zero gas (10 times greater) OPTIONAL OPTIONAL OPTIONAL OPTIONAL OPTIONAL An integrated version is available containing wet and dry sensors together with sample conditioning in a single enclosure which mounts directly to the process. CEM/Humox system consists of two CEM/O CEM/Humox system consists of two CEM/O CEM/Humox system consists of two CEM/O CEM/Humox system consists of two CEM/O CEM/Humox system consists of two CEM/O 2 sensors and Series 2000 Control Unit sensors and Series 2000 Control Unit sensors and Series 2000 Control Unit sensors and Series 2000 Control Unit sensors and Series 2000 Control Unit

Upload: esganfia

Post on 08-Nov-2015

213 views

Category:

Documents


1 download

DESCRIPTION

x

TRANSCRIPT

  • BULLETIN P-460BULLETIN P-460BULLETIN P-460BULLETIN P-460BULLETIN P-460

    CEM/Humox

    The CEM/Humox Moisture in Flue Gas Analyzer isdesigned to measure net oxygen and moisturecontent in flue gas and process applications. TheO2 and moisture data is used to correct emissions,assure product quality, or minimize stack corro-sion. The analyzer indicates moisture, oxygen-wetand oxygen-dry concentrations in percent byvolume.

    MEASUREMENT OF MOISTURE IN FLUE GAS TOCALCULATE EMISSIONS ON A DRY OR WET BASIS

    Principle of Operation:Principle of Operation:Principle of Operation:Principle of Operation:Principle of Operation: Zirconium oxide - onesensor for hot wet sample; one sensor for drysample.

    SENSORS

    SERIES 2000 CONTROL UNIT

    Display: Display: Display: Display: Display: Four-line x 20-character vacuum fluorescent.Displays combinations of % moisture, oxygen (wet or dry),time and date, cell temperature (wet or dry), cooler tempera-ture, user programmable text, thermocouple mV (wet or dry)or cell mV (wet or dry). Password protection, programmablepressure compensation and context-sensitive help are alsoprovided.

    Recorder Output: Recorder Output: Recorder Output: Recorder Output: Recorder Output: Three isolated linear current outputs.Each output can be 4-20 mA, 0-20 mA, 20-4 mA or 20-0 mAand is fully scalable. Hold or track during calibration andselect degree of damping. Maximum load 1200 ohms.

    Alarms: Alarms: Alarms: Alarms: Alarms: Four independent sensor alarms, each high or lowselectable. One alarm can be allocated to % moisture,calibrate or verify. Set relays to energize or de-energize onalarm.

    Contact RatingContact RatingContact RatingContact RatingContact Rating: 1 A, 30V max. noninductive load, AC or DC

    Diagnostics: Diagnostics: Diagnostics: Diagnostics: Diagnostics: Watchdog timer and service alarms. Systemtest for A/D, RAM, EEPROM and keypad. Display line 4reserved for full-text error and diagnostic messages.Twenty-entry event log.

    Communications: Communications: Communications: Communications: Communications: RS-485 2-way addressable.

    Operating RangeOperating RangeOperating RangeOperating RangeOperating RangeOxygen: 0.1% to 100%Moisture: 5% to 85% by volume

    AccuracyAccuracyAccuracyAccuracyAccuracyOxygen: 2% of reading or 0.1% absoluteMoisture: 5% of reading or 1% absolute, whichever isgreater

    Response Time: Response Time: Response Time: Response Time: Response Time: > 4 seconds at 2 scfh over one decade

    Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: Max. Inlet Temperature: 400F (204C)

    Sample Flow: Sample Flow: Sample Flow: Sample Flow: Sample Flow: 0.1 to 1.0 L/min (0.6 L/min is recommended)

    EnvironmentEnvironmentEnvironmentEnvironmentEnvironmentAmbient Temperature: -18C to 50C (0F to 122F)Relative Humidity: 10% to 90%, noncondensing

    Power Requirements: Power Requirements: Power Requirements: Power Requirements: Power Requirements: 115 VAC, 10%, 47-63 Hz; 230 VAC,10%, 47-63 Hz; 313 VA max.

    Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Calibration Gas Requirements: Use calibration gases @ 0.1to 1.0 L/min. (0.6 L/min recommended)

    Zero Gas: From 0.1 to 10% O2, balance N2Span Gas: Minimum one decade above zero gas(10 times greater)

    OPTIONALOPTIONALOPTIONALOPTIONALOPTIONAL

    An integrated version is available containing wet and drysensors together with sample conditioning in a singleenclosure which mounts directly to the process.

    CEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/OCEM/Humox system consists of two CEM/O22222sensors and Series 2000 Control Unitsensors and Series 2000 Control Unitsensors and Series 2000 Control Unitsensors and Series 2000 Control Unitsensors and Series 2000 Control Unit

  • wwwwwwwwwwwwwww.thermox.com.thermox.com.thermox.com.thermox.com.thermox.com 2003, by AMETEK, Inc. All rights reserved.

    403PDF (P460)

    One of a family of innovative process analyzer solutions from AMETEK Process Instruments. Specications subject to change without notice.

    PROCESS INSTRUMENTSPROCESS INSTRUMENTSPROCESS INSTRUMENTSPROCESS INSTRUMENTSPROCESS INSTRUMENTS

    CANADACANADACANADACANADACANADA USA - DelawareUSA - DelawareUSA - DelawareUSA - DelawareUSA - Delaware USA - PittsburghUSA - PittsburghUSA - PittsburghUSA - PittsburghUSA - Pittsburgh USA - TexasUSA - TexasUSA - TexasUSA - TexasUSA - TexasevirD nrohtihW 0579daoR tropeerF 051.dvlB etaroproC 554.E.N yaW egdirnuS 6782

    59077 XT ,notsuoH83251 AP ,hgrubsttiP20791 ED ,kraweN9H7 Y1T BA ,yraglaC0282-364-182 .hP0409-828-214 .hP0044-654-203 .hP0048-532-304 .hP1072-364-182 xaF9930-628-214 xaF4444-654-203 xaF0553-842-304 xaF

    CHINACHINACHINACHINACHINA FRANCEFRANCEFRANCEFRANCEFRANCE GERMANYGERMANYGERMANYGERMANYGERMANY MIDDLE EAST - DubaiMIDDLE EAST - DubaiMIDDLE EAST - DubaiMIDDLE EAST - DubaiMIDDLE EAST - Dubai SINGAPORESINGAPORESINGAPORESINGAPORESINGAPORE-56 .hP2502 188-4-179 .hP0 63 19 95 12 94 .hP02 96 86 03 1 33 .hP9407 6246 12 68 .hP 6484-2388

    3502 188-4-179 xaF0863 19 95 12 94 xaF92 96 86 03 1 33 xaF4507 6246 12 68 xaF Fax 65-6481-6588

    1.483.76

    5.2213.26

    2.255.72

    12.632.0

    9.1023.11

    19.0048.26

    4.9512.57

    19" RACK MOUNT

    12.030.5

    12.030.5

    10.025.4

    8.1920.8

    6.7517.14

    1/4" CompressionTube Fitting

    Sample Input1/4" CompressionTube Fitting

    Sample Output1/4" CompressionTube Fitting

    8.0020.32

    MountVertically

    12.7532.38

    2.506.35

    3.07.62

    3.07.62

    1.503.811.25

    3.17.31 dia. Holes for

    1/4" fasteners

    SENSOR

    DIMENSIONSINCHES

    CM

    EnvironmentEnvironmentEnvironmentEnvironmentEnvironmentAmbient Temp: 14 to 122F (-10 to 50C)Relative Humidity: 10% to 80%, noncondensing

    Hazardous Area: Hazardous Area: Hazardous Area: Hazardous Area: Hazardous Area: Meets NEC Class I, Division 2 withoutrequiring purge apparatus.

    Enclosure:Enclosure:Enclosure:Enclosure:Enclosure: Standard: 19" rack mount. Optional: GP (generalpurpose) wall mount, GP panel mount, weatherproof NEMA 4(IP 56) wall/panel mount, and stainless steel weatherproofNEMA 4X (IP 56) wall/panel mount. All are UL Listed for NECClass I, Division 2 areas.

    Calibration:Calibration:Calibration:Calibration:Calibration: Oxygen cell lifetime extender. Calibrate or verifycalibration. Store last calibration and verication data.Selectable calibration gas run time and process recoverytime. Timed automatic calibration with optional remotecalibration unit.

    Bias adjustment (2%) for process calibration to EPAMethod 4 if required.

    Power Requirements: Power Requirements: Power Requirements: Power Requirements: Power Requirements: Nominal 95-230 VAC,10%, 47-63 Hz; 75 VA max.

    System ComplianceSystem ComplianceSystem ComplianceSystem ComplianceSystem ComplianceEMC Directive 89/336/EECDirective 73-23/EEC

    CONTROL UNITAPPROX WEIGHT: 14 LB (6.35 KG)

    APPROX WEIGHT: 24 LB (10.9 KG)