chandra calibration status
DESCRIPTION
Chandra Calibration Status. ACIS Gain correction files for epochs 30, 31 and 32 were released in CALDB 3.4.1 (Sept 2007), 3.4.2 (Dec. 2007) and 3.4.3 (March 2008). The blank sky background data sets were reprocessed with the latest cti-corrected calibration products and released in CALDB 3.4.1 - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/1.jpg)
Chandra Calibration StatusChandra Calibration Status
ACISACIS
1.1. Gain correction files for epochs 30, 31 and 32 were released in CALDB Gain correction files for epochs 30, 31 and 32 were released in CALDB 3.4.1 (Sept 2007), 3.4.2 (Dec. 2007) and 3.4.3 (March 2008). 3.4.1 (Sept 2007), 3.4.2 (Dec. 2007) and 3.4.3 (March 2008).
2.2. The blank sky background data sets were reprocessed with the latest The blank sky background data sets were reprocessed with the latest cti-corrected calibration products and released in CALDB 3.4.1cti-corrected calibration products and released in CALDB 3.4.1
3.3. Updated gain correction files for data taken at a focal plane temperature Updated gain correction files for data taken at a focal plane temperature of T=-110 C with the BI chips (S1 and S3) were released in CALDB of T=-110 C with the BI chips (S1 and S3) were released in CALDB 3.4.33.4.3
![Page 2: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/2.jpg)
HRCHRC
An updated HRC-I de-gap corrections table derived from the An updated HRC-I de-gap corrections table derived from the AO8 Capella raster scan was released in CALDB 3.4.1. This AO8 Capella raster scan was released in CALDB 3.4.1. This improves image reconstruction for off-axis sources. improves image reconstruction for off-axis sources.
Web Pages
1. A tag index of all presentations given at the Chandra calibration workshops is now on-line at cxc.harvard.edu/ccw/tags
![Page 3: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/3.jpg)
ACIS and HETG Calibration ProjectsACIS and HETG Calibration Projects
Develop a set of cti-corrected calibration products
for CC-mode observations.
ACIS
Flight
Grades
![Page 4: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/4.jpg)
Comparison of TE and CC mode grade distributions
![Page 5: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/5.jpg)
Comparison between ECS data taken in TE and CC mode
![Page 6: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/6.jpg)
Comparison between Mn-K line profile in TE and CC mode
![Page 7: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/7.jpg)
Future work on CC mode calibrationFuture work on CC mode calibration
1. The CC mode cti-correcter only works when all flight grades other than 255 are telemetered. This will require a new default SI mode for ACIS data taken in CC mode.
2. At present, CC mode observations do not telemeter grade 7 events, so a separate QE for CC mode must be developed for all CC mode data taken until a new SI mode is implemented.
![Page 8: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/8.jpg)
Plot from the IACHEC Meeting comparing MOS, PN, and ACIS spectral Plot from the IACHEC Meeting comparing MOS, PN, and ACIS spectral fitting results in the 2-7 keV band for a sample of 7 clusters.fitting results in the 2-7 keV band for a sample of 7 clusters.
HRMA Calibration Projects
![Page 9: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/9.jpg)
Comparison of ACIS derived temperatures in a broad band, a hard band and Comparison of ACIS derived temperatures in a broad band, a hard band and from the H-like to He-like Fe K alpha line ratio.from the H-like to He-like Fe K alpha line ratio.
![Page 10: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/10.jpg)
Residuals in the Abell 2029 spectrum assuming the gas temperature is given Residuals in the Abell 2029 spectrum assuming the gas temperature is given by the Fe line ratio (kT=7.9 keV).by the Fe line ratio (kT=7.9 keV).
![Page 11: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/11.jpg)
Two corrections have been applied to the predictions of the raytrace code Two corrections have been applied to the predictions of the raytrace code since XRCF.since XRCF.
Empirical XRCF correction
HRMA overlayer of 22A
![Page 12: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/12.jpg)
Sensitivity of derived cluster temperatures on the depth of the HRMA Sensitivity of derived cluster temperatures on the depth of the HRMA overlayer without the empirical XRCF correction.overlayer without the empirical XRCF correction.
![Page 13: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/13.jpg)
Spectra fitting results with a HRMA effective area model without the XRCF Spectra fitting results with a HRMA effective area model without the XRCF empirical correction and a depth of 20A for the overlayer.empirical correction and a depth of 20A for the overlayer.
![Page 14: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/14.jpg)
Spectra fitting results with a HRMA effective area model without the XRCF Spectra fitting results with a HRMA effective area model without the XRCF empirical correction and a depth of 20A for the overlayer.empirical correction and a depth of 20A for the overlayer.
Comparison with XMM-Newton
![Page 15: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/15.jpg)
Fit to the continuum source at XRCF with a variable depth for Fit to the continuum source at XRCF with a variable depth for the overlayer on each shellthe overlayer on each shell
![Page 16: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/16.jpg)
![Page 17: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/17.jpg)
Things to doThings to do
Determine the depth of the overlayer required to match the SSD continuum measurement for each shell.
Apply the XRCF derived overlayer depths for each shell to the in-flight HRMA effective area model.
Adjust the HETG gratings transmission efficiency and the HRC-S QE accordingly.
Validate the in-flight HRMA effective area model with gratings and cluster data.
![Page 18: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/18.jpg)
HRC-I and LETG Calibration ProjectsHRC-I and LETG Calibration Projects
Generate high spatial resolution, time-dependent gain corrections for the HRC-SGenerate high spatial resolution, time-dependent gain corrections for the HRC-S
Median PHA (sum of all pre-amps) vs. energy on HRC-I
![Page 19: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/19.jpg)
Median PHA at C-Ka vs. position on central HRC-S chip
![Page 20: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/20.jpg)
Median SAMP (sum of 6 pre-amps) at C-Ka vs. position on
central HRC-S chip
![Page 21: Chandra Calibration Status](https://reader035.vdocuments.net/reader035/viewer/2022062322/56814799550346895db4cc95/html5/thumbnails/21.jpg)
Time dependence of HRC-S gain