chapter 1 introductionshodhganga.inflibnet.ac.in/bitstream/10603/23472/3/03... · 2018. 7. 9. · 2...

23
1 CHAPTER 1 INTRODUCTION 1.1 Cancer and radiotherapy Cancer is a term used for disease in which abnormal cells divides without control and are able to invade other tissue. The main treatment modalities include surgery, chemotherapy and radiotherapy. Radiotherapy or radiation oncology is the medical use of ionizing radiation, generally as part of cancer treatment to control or kill malignant cells. Ionizing radiation works by damaging the deoxy ribo nucleic acid (DNA) of exposed tissue leading to cellular death. Aim of radiotherapy is to give maximum radiation dose to tumor while minimizing radiation to normal tissues to reduce the complications. To spare normal tissues, shaped radiation beams are aimed from several angles of exposure to intersect at the tumor, providing a much larger absorbed dose there than in the surrounding healthy tissue. The three main divisions of radiation therapy are external beam radiation therapy or teletherapy, brachytherapy or sealed source radiation therapy, and systemic radioisotope therapy or unsealed source radiotherapy. The differences relate to the position of the radiation source: external is outside the body, brachytherapy uses sealed radioactive sources placed precisely in the area under treatment, and systemic radioisotopes are given by infusion or oral ingestion.

Upload: others

Post on 22-Oct-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

  • 1

    CHAPTER 1

    INTRODUCTION

    1.1 Cancer and radiotherapy

    Cancer is a term used for disease in which abnormal cells divides

    without control and are able to invade other tissue. The main treatment

    modalities include surgery, chemotherapy and radiotherapy. Radiotherapy

    or radiation oncology is the medical use of ionizing radiation, generally as

    part of cancer treatment to control or kill malignant cells. Ionizing radiation

    works by damaging the deoxy ribo nucleic acid (DNA) of exposed tissue

    leading to cellular death. Aim of radiotherapy is to give maximum

    radiation dose to tumor while minimizing radiation to normal tissues to

    reduce the complications. To spare normal tissues, shaped radiation beams

    are aimed from several angles of exposure to intersect at the tumor,

    providing a much larger absorbed dose there than in the surrounding

    healthy tissue. The three main divisions of radiation therapy are external

    beam radiation therapy or teletherapy, brachytherapy or sealed source

    radiation therapy, and systemic radioisotope therapy or unsealed source

    radiotherapy. The differences relate to the position of the radiation source:

    external is outside the body, brachytherapy uses sealed radioactive sources

    placed precisely in the area under treatment, and systemic radioisotopes are

    given by infusion or oral ingestion.

  • 2

    1.2 Radiotherapy machines and radiation treatment techniques

    Therapeutic external radiation is given to the patients using the

    equipments such as ortho-voltage units, deep-therapy x-ray machines, tele-

    cobalt units and linear accelerators (linacs) which produce megavoltage x-

    rays (figure1.1 and figure1.2). In treating with linear accelerators having

    multileaf collimator (MLC), the treatment volume can be shaped to

    conform to the tumor volume through beam shaping (figure1.3) and

    shielding of normal tissues and critical organs.

    Figure 1.1 A medical linear accelerator

  • 3

    Figure 1.2 Medical linear accelerator and its components

    Figure 1.3 Beam shaping with multi leaf collimators

  • 4

    The technique of radiotherapy with beam shaping and shielding of

    normal structures surrounding tumor volume is called as 3-dimensional

    conformal radiation therapy (3-D CRT). Intensity modulated radiation

    therapy (IMRT) is an advanced form of 3-D CRT (figure1.4). In IMRT,

    customized radiation dose is intended to maximize tumor dose while

    simultaneously protecting the surrounding normal tissue. The transition of

    radiotherapy from IMRT to volumetric modulated arc therapy (VMAT)

    made treatment of cancer easier & beneficial (figure1.5). In VMAT, three

    parameters are changing simultaneously - Gantry speed of linac, MLCs

    shape, and dose rate, but in IMRT there is no movement of gantry during

    the treatment and there is no dose rate variation.

    Figure 1.4 3DCRT and IMRT comparison

  • 5

    Figure 1.5 IMRT and VMAT comparison

    1.3 Importance of accuracy in dose planning and delivery in IMRT

    It is well understood in radiation therapy that the dose-response

    curves are quite steep and there is clinical evidence that a small change

    (5%) in the dose to target volume can result in a change in the tumor

    control probability (ICRU 1976). Along the same argument, similar dose

    change may also result in a sharp change in the incidence and severity of

    radiation-induced morbidity, especially for serial critical structures such as

    spinal cord, optic chiasm, and brain stem. For IMRT and VMAT the

    accuracy in dose planning and delivery is even more important because

    even a small displacement of the delivered dose distribution can result

    changes in doses that exceed the tolerance values for critical organs and

    seriously under dose the tumor volume. Based on clinical evidence on

    effective and excessive dose levels, the consensus in radiation therapy

    community is that the dose delivered to the tumor volume should be within

    5% of the prescribed dose. Therefore, the guiding principle in establishing

    quality assurance (QA) test procedures and in defining tolerance limits for

  • 6

    IMRT process is to minimize the overall uncertainty in delivered dose to

    less than 5%.

    1.4 Overall process of IMRT, sources of errors and importance of

    patient specific QA

    The IMRT process comprises of several steps: treatment setup,

    patient immobilization, computed tomography (CT) image acquisition,

    inverse treatment planning, plan acceptance, plan verification, and the

    actual treatment delivery. With multiple steps involved, there remains a

    large potential for random and systematic errors at each step along the way.

    Of these, the systematic errors are the most significant since they can have

    a huge impact on the final treatment outcome. In a clinical scenario

    therefore, every attempt should be made to reduce such systematic errors.

    The first logical step is to analyze the uncertainties in both the components

    of IMRT QA process – pertaining to machine (machine specific QA) and

    pertaining to individual patient treatment (patient specific QA). In IMRT,

    patient specific QA plays a crucial role because of the complexities

    involved in treatment plan, dose calculation and treatment delivery.

    1.5 Test tools and methods in IMRT patient specific QA

    Commonly followed IMRT QA methods include point dose

    measurements using a small volume ion chamber, planar dose

    measurements using a film or a 2-D array detector, portal dosimetry etc

    [(wagter et al. (2004)]. In advanced QA systems, fluence measured by a 2-

    D array detector can be used to calculate dose to a 3-D volume. This makes

    possible comparison between treatment planning systems (TPS) calculated

    dose volume histogram (DVH) and QA system’s DVH.

  • 7

    1.5.1 Point dose measurement

    Absorbed dose determination using calibrated ionization chambers

    in combination with a well established dosimetry protocol, such as the

    international atomic energy agency (IAEA) protocol, are generally

    assumed to be the gold standard in radiation dosimetry. Under reference

    conditions, the estimated combined standard uncertainty in the

    determination of absorbed dose in high energy photon beams amounts to

    about 1.5%.

    As an initial step, during the commissioning of IMRT, the solid

    water phantom with ion chamber (figure1.6) has to be scanned and the

    image set has to be imported to the treatment planning system. Patient

    verification plans can be created in the treatment planning system for the

    absolute dosimetric measurement by exporting the patient specific IMRT

    plans on the image set of IMRT phantom, which saved in the treatment

    planning system. After the 3-D dose calculation, the dose at a reference

    depth in the phantom can be measured from the TPS created verification

    plan. These plans will be executed in the linear accelerator. To measure the

    absolute dose at the reference point, the IMRT water equivalent phantom

    should place on the treatment couch and the ion chamber has to be inserted

    at the level of reference point depth. The measured dose for each IMRT

    fields at reference point is then compared with the TPS calculated absolute

    dose at the same point and the % of variation will be calculated [Mijnheer et

    al. (2008)].

  • 8

    Figure 1.6 Solid water phantom, ion chamber and electro

    meter for absolute point dose measurements

    1.5.2 Film dosimetry

    Radiographic films have been employed almost since the discovery

    of X-rays to measure radiation dose. The use of radiographic films are

    relatively easy, quick and cheap and therefore very often applied for many

    applications in radiotherapy. It provides data with a high resolution and a

    permanent record of the 2-D dose distribution in the plane of irradiation.

    There are, however, many parameters influencing the film irradiation, film

    processing and data analysis procedure that determine the accuracy of the

    final result. Simultaneous dosimetric measurements in more than one point

    have become an important need for quality assurance in modern

    radiotherapy. Such measurements are traditionally performed with

    radiographic films as a two-dimensional detector. However, their

    application is not straightforward due to many factors of influence on the

    optical density, such as energy and spectral composition, depth, field size,

    orientation, and processing conditions [Kapulsky (2002)]. Additionally, the

    increasing number of IMRT patients suggests the use of faster and more

  • 9

    efficient dosimetric tools. Finally, many hospitals are aiming towards a so-

    called “digital hospital”, where film-processing machines for traditional

    silver-halide films will not be available or easily accessible in the near

    future [Wiezorek et al. (2005)].

    Radiochromic films, which are self-developing, almost tissue

    equivalent and therefore shows little energy and directional dependence,

    represent an alternative to radiographic films but their use is still limited

    because they were until recently rather expensive [(wagter et al. (2004)].

    Other factors include cumbersome film handling, sensitivity variations

    across the film, and the low sensitivity to ionizing radiation doses typically

    used in radiation oncology. This prevented their use in external beam

    therapy, and their dominant application in radiation oncology was limited

    to brachytherapy. Figure1.7 shows the film dosimetry set up for IMRT.

    Figure 1.7 Film dosimetry set up for IMRT

  • 10

    1.5. 3 Portal dosimetry

    On-line electronic portal imaging devices (EPIDs) have been

    developed for acquiring megavoltage images during patient treatment.

    Megavoltage images, obtained in digital format with such a device, are

    then used for further analysis, mainly for determining set-up errors. The

    image information can, however, also be related to the dose delivered to

    the EPID, yielding dose information in a plane instead of in one or few

    points. EPIDs can serve for several purposes during the verification

    process of IMRT and are used: 1) to verify the leaf position either during

    static (step-and-shoot) or dynamic MLC (sliding window) techniques; 2)

    to check the correct transfer of the leaf sequencing file to the treatment

    machine; and 3) to measure the combined mechanical and dosimetric

    performance of the treatment unit [Prisciandaroo et al. (2004), Fielding et

    al. (2004), Yang et al. (2004)]. More recently the uses of flat panel

    imagers based on amorphous silicon (aSi) are becoming more popular for

    their use as 2-D dosimeters. Most new accelerators are nowadays

    equipped with aSi1000 EPID and it can therefore be expected that the use

    of these devices for IMRT verification will increase in the future. Several

    approaches have been described for the use of EPIDs for pre-treatment

    verification of IMRT delivery [Warkentin et al. (2003), Vieira et al.

    (2004), Budgel et al. (2005), Monti et al. (2006), Winkler et al. (2006)].

    More recently various groups have developed methods to translate EPID

    images into 2-D primary fluence maps, which are then used as input in a

    TPS to recalculate 3-D dose distributions using CT data of a phantom or

    patient. Generally these approaches are able to reconstruct the 3-D dose

    distribution in phantoms with a high accuracy.

  • 11

    Figure 1.8 Linac with inbuilt aSi1000 EPID (black arrow)

    The aSi1000 portal imager (figure1.8) is the most recent detector

    used for portal dosimetry. It is a flat panel X-ray imager with large area

    active matrix readout structure and is made up of phosphor or photo

    conductor. This detector is having four major parts- 1mm Cu build up

    plate, a scintillating Phosphor screen, Image forming sensitive layer and

    associated electronics. The Cu build up plate absorb the incident photons

    and emits recoil electrons and also it shields the scintillation screen from

    the scattered radiation. The recoil electrons from the build up plate are

    absorbed by the scintillating phosphor screen and convert it into visible

    light. The image forming layer is a 512 X 384 matrix deposited on a glass

    substrate. Here each pixel in the matrix is having 0.784 mm pitch and

    consists of aSi-n-i-p photo cathode to integrate the incoming light in charge

    capture and a thin film transistor (TFT). The associated electronics with the

    TFT switches enables the charge capture readout. The image acquisition

    system with fast readout electronics enable up to 30 frames per second is a

  • 12

    major difference in aSi1000 EPID. The resolution of aSi1000 EPID is upto

    0.39 mm.

    Before using the EPID for clinical purpose, the dosimetric

    calibration and characteristics study of the portal imager has to be

    performed [Berger et al. (2006), McDermott et al (2006), Greer et al.

    (2007)]. For the IMRT patient specific QA, verification plans are creating

    in treatment planning system using PDIP (portal dose image prediction)

    algorithm. To measure the delivered dose the aSi1000EPID has to be

    placed at the calibrated distance from the source. The verification plan is

    then executed in linac through the networking platform and control

    console. The measured and TPS predicted planar doses for individual fields

    can be compare using the portal dosimetry analysis tool (figure1.9).

    Figure 1.9 Portal dosimetry work flow

  • 13

    1.5.4 2-D array system

    The major benefit of dosimeter arrays are their simple handling by

    connecting them to a computer and the availability of on-line information.

    Two-dimensional (2-D) arrays are more practical as they allow the

    verification of a planar fluence or dose distribution. During the last years

    several systems became commercially available for 2-D dosimetry. The

    most commonly utilized dosimetric principles are ionization in air or

    ionization in semiconductor material, but other principles such as

    scintillation have been applied as well. Advantages of ionization chambers

    are the simple calibration, practically no dead time, which allows real time

    measurement, and no (significant) effect of radiation damage. In general,

    dosimetric properties are governed by the physics principle of the detector.

    The most important ones are dose linearity, energy dependence, directional

    dependence, dose rate dependence, source to detector distance (SDD) field

    size response and temperature response. The various commercially

    available 2-D arrays show differences in the number of detectors, detector

    spacing, detector shape, effective point of measurement, water-equivalent

    build-up layer, backscatter layer, and maximum field size covered. Most of

    these systems can be used for absolute dose measurement after appropriate

    individual calibration procedures to correct for response variations across

    the array. 2-D arrays mounted on the gantry enable IMRT verification at

    gantry angles identical to the ones applied in treatment plans. For such

    procedures detector misalignments and influences of gravity need to be

    considered carefully and corrected for if present.

    The commercial 2-D arrays for dosimetric purposes come with their

    inherent evaluation software. It is generally possible to import calculated

    dose distributions from a planning system and to perform 1-D dosimetric

    analysis of profiles or a 2-D gamma evaluation using data of the whole

  • 14

    array. The major limitation of 2-D array is it’s limited number of detectors,

    which impairs measurements in high dose gradient regions and in small

    fields. The usefulness of a gamma evaluation, based on dosimetric

    information with a limited spatial resolution is therefore questionable.

    Obviously the limited spatial resolution of the 2-D array influences the

    effectiveness of the verification at some points. For that reason it is

    recommended to combine the results of multiple measurements in which

    the array has been replaced over a small distance. Recently transmission-

    type radiation detectors have been developed that can be positioned on the

    radiation entrance side of the patient. These detectors are multi wire or

    multi-strip ionization chambers connected to a multi-channel electrometer.

    They are designed to be placed in dedicated holders or in standard

    accessory holders of the linear accelerator. As a consequence the spatial

    resolution depends on the mounting distance. Because of their negligible

    attenuation, transmission-type 2-D detectors can be permanently installed

    on accelerators primarily used for IMRT. They enable on-line monitoring

    of beam characteristics or leaf settings with and without the patient in

    place. However, when using such detectors the characteristics of a certain

    device need to be taken into account, including its specific influences on

    the overall QA procedures and dosimetry logistics. Moreover, besides the

    advantage of offering on-line information, 2-D detectors have the potential

    to increase the overall efficiency for IMRT QA. In addition, these tools can

    also be used for QA of linear accelerators used for conventional treatments,

    such as measurement of leaf position, output constancy, beam symmetry

    and field flatness.

    There are different types of 2-D array systems available now in

    market for the IMRT patient specific QA. ImatriXX 2-D array system of

    IBA dosimetry, map check 2-D arrays of sun nuclear dosimetry system, 2-

  • 15

    D arrays of PTW dosimetry system etc are the commonly using 2-D array

    systems.

    ImatriXX 2-D array verification-process

    The ImatriXX 2-D array system consists of 1020 parallel plate ion chamber

    arranged in a 32x32 grid, with an inter detector spacing of 7.619 mm. Each

    detector is having a diameter of 4.5 mm, height 5 mm and chamber volume

    0.02 cc. To compare the TPS calculated planar dose with the measured

    planar dose the ImatriXX 2-D array system has to place on the treatment

    couch of the linac with the detector level at 100cm from the source.

    Sufficient backscatter is placing below the detector and build up is placing

    above the ImatriXX detector (figure1.10).

    Figure1.10 ImatriXX 2-D array system- measurement set up

    The TPS created verification plans will be executed in the linac. The

    measured and TPS calculated planar dose comparison can be performed by

    the IMRT QA software (figure1.11). The gamma evaluation can be

  • 16

    performed and the % of pixels passing the specified gamma criteria can be

    obtained for every individual IMRT fields.

    Figure1.11 ImatriXX 2-D array system – work flow

    1.5. 6 Compass - DVH based verification system

    Compass is a dose verification system of IBA and it is used in

    combination with MatriXX-Evolution detector (1020 pixels) for pre-

    treatment verification of conformal IMRT plans and with MatriXX -

    Evolution and gantry angle sensor for the verification of rotational plans as

    well as transmission detector (1600) pixels for online verifications of

    IMRT plans and rotational plans. Compass can determine the 3-D dose

    distribution in the patient anatomy, based on the measured beam intensity

    and it determines the fluence for all segments in a beam. As this quantity

    cannot be directly measured, Compass does first a calculation of the

    expected response of electrical signal for each segment based on detector

    pixel response, linac and detector models. After the measurement, expected

  • 17

    and delivered responses are compared. The residual response is then used

    for computation of the really delivered fluence. The dose computation in

    Compass is a second independent step in which the resulting dose to the

    patient is determined based on a collapsed cone super position algorithm.

    For the commissioning of the Compass 3-D verification system with

    MatriXX-Evolution detector, the same data of TPS commissioning

    (profiles, depth dose curves, output factors, absolute dose measurements)

    are used. The primary quantity determined by Compass is the fluence for

    each segment. Discrepancies in delivery can be visualized as difference in

    the response patterns. The fluence determined in Compass is then used as

    input for the dose computation with the collapsed cone algorithm. For the

    conformal or IMRT plans, the Dicom RT plan, Dicom RT dose, Dicom RT

    structure and Dicom CT information etc are to be transferred to the

    Compass verification system. The Compass will compute using collapsed

    cone algorithm and compare the TPS calculated and compass calculated

    DVHs (dose volume histograms) and provide the differences.

    1.6 Tolerance limits and Action levels for IMRT verification

    Tests for IMRT verification can be separated into those for

    verification of equipment for IMRT delivery, verification of IMRT

    treatment planning, and verification of patient-specific IMRT techniques,

    i.e., of the combined planning and delivery process of that particular

    patient treatment based on relative as well as absolute dosimetry. Different

    approaches exist for the comparison of sets of measured and calculated

    dose distributions [Mijenheer et al (2008)]. Each of these approaches needs

    well-defined criteria for acceptance of a plan and procedures if these

    criteria are not met. Tolerance and action levels can be used. These

    quantities can be defined in the following way: whenever a parameter is

  • 18

    found in the range below the tolerance level, the equipment is suitable for

    high quality radiation therapy. If, however, a parameter exceeds the action

    level, it is essential that appropriate actions be taken as soon as possible.

    Consequently, tolerance levels are appropriate limits for performance

    specification and for acceptance testing procedures, while action levels

    might be regarded as more relevant values for use in ongoing quality

    control activities. If a parameter has a value between the tolerance limit

    and the action level, the responsible physicist will generally decide to

    continue with the treatment until a suitable moment for further

    investigation occurs. If such an investigation is not possible, then high

    quality treatments should no longer be performed with such equipment.

    Tolerance and action levels should now be defined for the various tests to

    compare measured with calculated dose distributions. The most often

    applied dose evaluation techniques comprise a direct comparison of dose

    differences (%dose difference), a comparison of distance to agreement

    (DTA) between measured and calculated dose distributions, and a

    combination of these two parameters: the gamma evaluation method.

    In the gamma evaluation, doses in the TPS calculated plan and the

    measured at the same pixel position are compared, and the difference as a

    percentage of the plan value is the percentage difference. In high gradient

    regions, one looks for the distance between a pixel in the plan and a pixel

    in the measured distribution, that have the same dose. This is the distance

    to agreement (DTA).

    Differences of about 5% are generally significant for IMRT

    verification [Mijenheer et al. (2008)]. Deviations larger than ± 5% should

    therefore firstly result in a review of the complete dosimetric procedure

    taking into account the various factors influencing the comparison result. If

    no explanation for the observed discrepancy can be given, the

  • 19

    measurement may be repeated. A possible recommendation might then be

    that a tolerance limit of ± 3% and an action level ± 5% should be applied

    for these types of point dose verifications. When the number of comparison

    points is large, simple methods of reporting deviations between dose

    measurements and calculations will collapse, and a method of compiling

    these deviations into a single number is required as a pass-fail criteria.

    Other methods have therefore been proposed, e.g., the use of the quantity

    “confidence limit” by Venselaar et al. (2001). The confidence limit is

    based on the average deviation between measurements and calculations for

    a number of data points in a comparable situation, and the standard

    deviation (SD) of the average of the differences. The confidence limit is

    then defined as the sum of the average deviation and 1.5 SD. The factor 1.5

    was based on experience and a useful choice in clinical practice. A

    multiplicative factor of 1.96 instead of 1.5 has later been proposed by Palta

    et al. (2003) for having 5% of the individual points exceeding the tolerance

    level. For both the verification of individual beams, as well as for the

    verification of patient-specific “hybrid” plans, Palta et al. (2003) proposed

    the set of values of confidence limits and action levels for IMRT

    treatments. An IMRT treatment plan should not be used clinically if the

    measured dose difference is more than the value given as the action level,

    which serves therefore as a pass-fail criterion. Application of the gamma

    evaluation method for selecting action levels is still in a development stage.

    Careful statistical analysis of patient specific verification data might

    reveal systematic uncertainties valid for the whole patient group. The

    statistical analysis of the results of a routine QA programme, possibly

    applied to a set of patients treated according to a class solution, can be very

    useful in defining appropriate tolerance/action levels taking into account

    the special aspects of IMRT relevant for a specific clinic. Currently no

    recommendations for 3-D dose evaluation are available and are therefore

  • 20

    urgently needed. Biological considerations, combined with the clinical

    experience from the 3-DCRT era, may be required to develop tolerance and

    action levels for the evaluation of 3-D dose distributions for an individual

    patient.

    If a gamma evaluation exceeds a certain action level for a chosen

    combination of dose-difference and DTA criteria, then possible reasons for

    discrepancies such as variation in phantom positioning and linac

    performance should first be investigated. If these experimental

    uncertainties are within accepted values, then it might be useful to repeat

    the experiment to confirm the observed discrepancies. If the same areas of

    the gamma maps fail the tolerance criteria again, then these areas should be

    compared with the corresponding regions in the patient dose distribution,

    and the implications of such a failure should be discussed with the

    responsible physicist and radiation oncologist. For each patient a decision

    should then be made if a new plan has to be generated or if the differences

    are clinically acceptable. It should be noted that such patient-specific

    action levels depend on many decisive factors, including the position and

    size of the area that failed to pass the evaluation criteria, the dose level in

    the PTV (planning target volume) or OAR (organ at risk), and the

    sensitivity of the plan for movement. Furthermore, gamma evaluation is

    currently mainly restricted to the dose delivered to the PTV, whereas the

    dose in an OAR is equally important. Extension of decision protocols

    including OAR is therefore urgently needed. The situation becomes even

    more complicated if dose distributions are evaluated in 3-D. It should be

    noted that tolerance limits and action levels have proven to be very useful

    in everyday quality control of accelerators, but some parameters are not

    easily and quickly corrected or repaired and some may almost be

    impossible or very expensive to restore. On very rare occasions, it might

    therefore be justified to use the radiation equipment clinically, even if an

  • 21

    action level has been exceeded. The decision to clinically use a treatment

    unit, in spite of the fact that an action level has been exceeded, has to be

    discussed thoroughly and documented for every treatment method.

    1.7 Motivation for research and purpose of the thesis

    Quality assurance in IMRT is mainly founded on quantitative

    comparisons between computed and/or measured dose distributions.

    Differences between measurement and calculation are principally caused

    by errors in treatment planning, patient positioning, treatment delivery and

    radiation dose measurement technique. However a simple agreement

    between the two distributions cannot be held as a proof of satisfying

    quality. Indeed the distributions that are compared may contain individual

    uncertainties or bias such that the agreement seen is a chance coincidence.

    This consideration may serve as an argument to include many degrees of

    freedom in the QA measurement process, i.e. in terms of measurement

    points in the comparison, volume of detectors, resolution of detectors, type

    or complexity of plans, number of fields or arcs, movement of carriage,

    inclusion or exclusion of couch in the plan, mode of delivery etc. The

    proposed QA tolerance values found in some reports for these parameters

    should therefore be considered as general recommendations and these may

    not be always achievable in a clinical scenario.

    Such guidelines for the IMRT QA are given in the American

    Association of Physicists in Medicine- Task Group - 119 (AAPM-TG 119)

    report [(Ezzell et al. (2009)] and in the European Society of Radiotherapy

    and Oncology (ESTRO) guidelines [(Mijenheer et al. 2008)]. But a

    comprehensive report that considers all possible influencing parameters or

    factors that affect QA results such as complexity of plans, degree of

    modulation, numbers of target volumes, numbers of fields or arcs,

  • 22

    movement of MLC carriage, inclusion or exclusion of treatment couch in

    the plan, mode of delivery etc., is still to come. Because of the nature of

    complexity and the many steps involved, an optimized stringent QA

    protocol is essential to rule out all possible uncertainties and errors. This

    research work mainly aimed to study the influence of different factors on

    the optimization of patient specific QA in IMRT to adopt a stringent QA

    protocol.

    1.8 Aims and objectives

    The main objectives of this research work were as follows:

    - To assess the beam stability of a Varian high energy medical linear

    accelerator; quantify MLC positional errors using ionometric

    gravity test and dynalog file analysis.

    - To establish the characteristics of the different QA systems such as

    aSi1000 based portal dosimetry and ImatriXX 2-D array system.

    - To study the angular response of the portal dosimetry and ImatriXX

    2-D array system.

    - To optimize the patient specific QA measurements for IMRT and

    VMAT through the statistical analysis of the results.

    - To examine the relationship and significance of different parameters

    such as detector systems, types of plans, their complexity, number

    of targets, movement of MLC carriage, number of fields or arcs,

    inclusion or exclusion of couch insert etc., on the selection of pass-

    fail criteria and action levels in patient specific QA.

    - To examine whether the IMRT QA guidelines and published reports

    can be used for the VMAT QA analysis also or separate

    optimization is needed or not.

    - To validate the Compass 3-D verification system and to optimize

    patient specific QA for IMRT.

  • 23

    1.9 Hypothesis

    Null Hypothesis (H0)

    The null hypothesis was that the patient specific QA results are not

    influenced by the QA systems, type of plans, complexity of plans, number

    of fields or arcs, number of target volumes, movement of carriage and

    inclusion or exclusion of couch insert and there is no need for institutional

    local optimization of QA for IMRT and VMAT.

    Alternate Hypothesis (H1)

    The alternate hypothesis was that the patient specific QA results are

    influenced by the QA systems, type of plans, complexity of plans, number

    of fields or arcs, number of target volumes, movement of carriage and

    inclusion or exclusion of couch insert and institutional local optimization

    of QA for IMRT and VMAT is needed.