chapter 11.humidification equipment

43
Chapter 11 Humidification Equipment General Considerations Terminology Humidity is a general term used to describe the amount of water vapor in a gas. It may be expressed several ways. Absolute Humidity Absolute humidity is the mass of water vapor present in a volume of gas. It is commonly expressed in milligrams of water per liter of gas. Humidity at Saturation The maximum amount of water vapor that a volume of gas can hold is the humidity at saturation. This will vary P.297 with the temperature. The warmer the temperature, the more water vapor can be held in a gas. Table 11.1 shows the absolute humidity of saturated gas at various temperatures. At a body temperature of 37°C, it is 44 mg H 2 O/L. TABLE 11.1 Water Vapor Pressure and Absolute Humidity in Moisture-saturated Gas Temperature °C mg H 2 O/L mm Hg 0 4.84 4.58 1 5.19 4.93 2 5.56 5.29 3 5.95 5.69 4 6.36 6.10

Upload: adham08

Post on 20-Jul-2016

14 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: Chapter 11.Humidification Equipment

Chapter 11 Humidification Equipment General Considerations Terminology Humidi ty is a general term used to describe the amount of water vapor in a gas. I t

may be expressed several ways.

Absolute Humidity Absolute humidi ty is the mass of water vapor present in a volume of gas. It is

commonly expressed in mill igrams of water per l i te r of gas.

Humidity at Saturation The max imum amount of water vapor that a volume of gas can ho ld is the humid ity

at satura tion. This wi l l vary

P.297

wi th the tempera ture. The warmer the temperature, the more water vapor can be

held in a gas. Table 11.1 shows the abso lute humidi ty of saturated gas at various

temperatures. At a body tempera ture of 37°C, i t is 44 mg H2O/L.

TABLE 11.1 Water Vapor Pressure and Absolute Humidity in Moisture-saturated Gas

Temperature °Cmg H2O/L mm Hg

0 4.84 4.58

1 5.19 4.93

2 5.56 5.29

3 5.95 5.69

4 6.36 6.10

Page 2: Chapter 11.Humidification Equipment

5 6.80 6.54

6 7.26 7.01

7 7.75 7.51

8 8.27 8.05

9 8.81 8.61

10 9.40 9.21

11 10.01 9.84

12 10.66 10.52

13 11.33 11.23

14 12.07 11.99

15 12.82 12.79

16 13.62 13.63

17 14.47 14.53

18 15.35 15.48

19 16.30 16.48

Page 3: Chapter 11.Humidification Equipment

20 17.28 17.54

21 18.33 18.65

22 19.41 19.83

23 20.57 21.07

24 21.76 22.38

25 23.04 23.76

26 24.35 25.21

27 25.75 26.74

28 27.19 28.35

29 28.74 30.04

30 30.32 31.82

31 32.01 33.70

32 33.79 35.66

33 35.59 37.73

34 37.54 39.90

Page 4: Chapter 11.Humidification Equipment

35 39.57 42.18

36 41.53 44.56

37 43.85 47.07

38 46.16 49.69

39 48.58 52.44

40 51.03 55.32

41 53.66 58.34

42 56.40 61.50

Relative Humidity Relat ive humidity, or percent saturation , is the amount of water vapor at a

part icula r tempera ture expressed as a percentage of the amount that wou ld be he ld

i f the gas were satura ted.

Water Vapor Pressure Humidi ty may also be expressed as the pressure exerted by water vapor in a gas

mixture. Table 11.1 shows the vapor pressure of water in saturated gas a t various

temperatures.

Inter-relationships I f a gas saturated with water vapor is heated, its capaci ty to hold moisture

increases and i t becomes unsaturated (has <100% relat ive humid ity). Its absolu te

humidi ty remains unchanged. Gas tha t is 100% satura ted at room temperature and

warmed to body temperature wi thout addit ional humidi ty wi l l absorb water by

Page 5: Chapter 11.Humidification Equipment

evaporat ion f rom the surface of the respiratory tract mucosa unti l i t becomes

sa turated.

I f gas sa turated wi th water vapor is coo led , it wil l condense (rain out) water. The

absolute humid ity wil l fall, but the relat ive humidi ty wi l l remain at 100%.

I f inspired gas is to have a rela tive humidi ty o f 100% at body temperature , it must

be maintained at body temperature af te r leaving the humidif ier or heated above

body tempera ture at the humidif ier and allowed to coo l as i t f lows to the patient.

Coo ling wi l l result in condensat ion (ra in out) in the breathing system.

The specif ic heat of gas is low. As a consequence, i t quick ly assumes the

temperature of the surround ing environment. Inhaled gases quick ly approach body

temperature , and gases in corrugated tubes rap idly approach room temperature.

The heat of vaporizat ion of water is relat ively high. Evaporat ion of water, therefore ,

requires cons iderably more heat than warming of gases . Likewise , condensation of

water y ields more heat than cool ing of gases.

Considerations for Anesthesia Water is intent ional ly removed from medical gases so tha t gases delivered f rom the

anesthesia mach ine are dry and at room tempera ture. As gases f low to the a lveol i,

inspired gas is brought to body temperature (ei ther by heating or cool ing) and 100%

rela tive humidi ty (ei ther by evaporat ion or condensation). In the unintubated

patient, the upper respiratory tract (especial ly the nose) funct ions as the princ ipa l

heat and moisture exchanger (HME). During normal nasal breathing, the

temperature in the upper trachea is between 30°C and 33°C, wi th a

P.298

rela tive humidi ty of approximately 98%, providing a water content of 33 mg/L (1).

Tracheal tubes and supraglott ic ai rway devices bypass the upper a irway, modifying

the patte rn of heat and mois ture exchange so tha t the tracheobronchial mucosa

must assume a greater role in heating and humid ifying gases.

Effects of Inhaling Dry Gases The importance of humid if icat ion in anesthesia remains uncertain . It is of g reates t

benef it in ped iatric pat ients, pa tients at increased risk for developing pulmonary

complica tions , and long procedures.

Damage to the Respiratory Tract

Page 6: Chapter 11.Humidification Equipment

As the respiratory mucosa dries and i ts temperature drops, secretions thicken,

c il iary funct ion is reduced, surfactant act iv ity is impaired, and the mucosa becomes

more suscept ib le to injury (2,3). If secret ions are not cleared, ate lec tasis o r ai rway

obstruct ion can resul t . Thickened plugs may prov ide loc i for infec tion. Dry gases

can cause bronchoconstric t ion, further compromis ing respiratory funct ion.

Humidifying gases may decrease the incidence of respira tory complications

(coughing and breath holding) associated wi th an inhalat ion induct ion (4,5).

There is no agreement about the minimum humidity necessary to prevent

patho log ical changes. Recommendations have ranged f rom 12 to 44 mg H2O/L

absolute humid ity (2,3,6,7,8 ,9). The dura tion of exposure is important. I t is unlike ly

that a brief exposure to dry gases wil l damage the tracheobronchial tree. As time

increases, the like lihood tha t significant tracheobronchial damage wi l l become

greater.

Body Heat Loss Body temperature is lowered as the ai rways bring the inspired gas into thermal

equil ibrium and sa turate it wi th water. The use of a humidif ica tion dev ice can

decrease the heat loss tha t occurs during anesthesia and may provide some heat

input (10 ,11,12,13,14,15,16,17). Contro ll ing inspired gas temperature and humidi ty

is not an ef f ic ien t method of maintaining body temperature. Means to increase body

temperature are discussed in Chapter 31.

Absorbent Desiccation The effects of dry absorbent on the composi tion of inspired gases are discussed in

Chapter 9. Whi le HMEs preserve patien t heat and humidi ty , they also may

contribute to absorbent desicca tion (18).

Tracheal Tube Obstruction Thickened secretions in a tracheal tube increase i ts res is tance and can result in

comple te obstruction (19 ,20,21,22,23,24).

Consequences of Excessive Humidity An increased water load can cause c i l iary degenerat ion and paralysis, pulmonary

edema, al tered alveolar-arterial oxygen gradien t, decreased v ital capacity and

compliance, and a decrease in hematocri t and serum sodium (25).

Sources of Humidity

Carbon Dioxide Absorbent

Page 7: Chapter 11.Humidification Equipment

The reac tion of absorbent with carbon diox ide l iberates water (Chapter 9). Water is

also conta ined in the absorbent granu les . Since the react ion is exothermic, heat is

produced. I f the absorbent granu les desiccate , they may react wi th certain

anesthet ics and produce extreme heat. This is discussed in detail in Chapter 9.

Exhaled Gases There is some rebreath ing in the tracheal tube, the supraglott ic a irway dev ice, and

the connections to the breathing sys tem. A lmos t half of the humidi ty in expired gas

is preserved in this manner (26).

In systems that allow rebreathing of exhaled gases (Chapter 8 ), the humidi ty and

temperature of the insp ired gases depend on the relat ive proport ions of f resh and

expired gases. This wil l depend on the sys tem and the fresh gas f low. As the f resh

gas f low is increased, the inspired temperature and humidi ty are reduced. Use of

the sys tem by a previous patient wil l increase the humidi ty .

Moistening the Breathing Tubes and Reservoir Bag Rinsing the ins ide of the breathing tubes and reservoir bag wi th water before use

increases the inspired humidi ty (27).

Low Fresh Gas Flows Using low fresh gas f lows wi th a ci rcle breathing sys tem wi ll conserve mois ture.

This is d iscussed in more detail in Chapter 9.

Coaxial Breathing Circuits Coaxial ci rc le systems, when combined wi th low f lows , wi l l inc rease the humid ity

more quickly than a system with two separate l imbs (2), bu t th is is not very eff ic ien t

in terms of heat or humidi ty improvement (28).

The Bain system (see Chapter 8) is a coax ial version of the Mapleson D. It does not

meet opt imal

P.299

humidif ica tion requirements because of the high fresh gas f low required (26).

Heat and Moisture Exchangers An HME conserves some exhaled water and heat and returns them to the patien t in

the inspired gas. Many HMEs also perform bacterial /v iral f i l t ra tion and prevent

inhalat ion of small part ic les. The HME is also known as a condenser humid if ie r,

Swedish nose, art i fic ial nose, nose humidif ier, passive humidif ier, regenerat ive

Page 8: Chapter 11.Humidification Equipment

humidif ier, moisture exchanger, and vapor condenser. When combined wi th a f i l te r

for bac teria and v i ruses, i t is cal led a heat and moisture exchanging f i l ter (HMEF).

Two international s tandards on HMEs have been published (29,30).

Description HMEs are disposable devices wi th the exchanging medium enclosed in a plast ic

housing. They vary in s ize and shape. Typical ones are shown in Figure 11.1. Each

has a 15-mm female connec tion port at the patient end and a 15-mm male port at

the other end. The patient port may also have a concentric 22-mm male f i tt ing

(Fig .11 .1C,D). There may be a port to attach the gas sampling l ine for a respiratory

gas monitor (Fig. 11 .1B,D) or an oxygen line. One type uti l izes a ceramic heat ing

element with a water input port , a membrane, and an aluminum grid that vaporizes

the water (31).

The dead space of HMEs varies. Pediatric and neonatal HMEs wi th low dead space

are avai lable (32 ,33). Mos t modern HMEs are one of two types, as shown in Table

11.2.

Hydrophobic Hydrophobic HMEs have a hydrophobic membrane with smal l pores. The membrane

is pleated to increase the surface area. A pleated hydrophobic membrane provides

moderately good inspired humidi ty . The performance of this type of HME may be

impaired by h igh ambient temperatures (1,21).

Hydrophobic HMEs are eff ic ient bacterial and v iral f i l ters (34,35,36,37,38,39). A

pleated hydrophobic f i l te r wi l l consistent ly prevent the hepat it is C v i rus from

passing whi le a hygroscop ic f i l ter may be inef fective (40). They a llow the passage

of water vapor but not l iquid water a t usual venti lato ry pressures (34 ,36). They are

associated wi th small inc reases in res istance even when wet (12,41,42).

Hygroscopic Hygroscopic HMEs conta in a wool , foam, or paperl ike material coated with

moisture-retain ing chemicals. The medium may be impregnated with a bacteric ide

(43). Composite hygroscopic HMEs consist of a hygroscopic layer plus a layer of

thin , nonwoven f iber membrane that has been subjected to an e lectrical f ie ld to

increase i ts polari ty. This improves fi l trat ion e ff ic iency and hydrophobic ity.

Most studies have shown that compos ite hygroscopic HMEs are more eff icient a t

moisture and tempera ture conservat ion than hydrophobic ones

(1,22,44,45,46,47,48,49,50,51,52,53). They wil l lose their ai rborne f i l t ra tion

Page 9: Chapter 11.Humidification Equipment

ef fic iency if they become wet, and microorganisms held by the f i l te r medium can be

washed through the device. Their res istance can inc rease great ly when they

become wet (42).

Indications An HME can be used to increase inspired heat and humidi ty during both short- and

long-term ventila tion. HMEs may be especially usefu l when transport ing intubated

patients, because transport vent i la tors frequently have no means for humid ifying

inspired gases.

An HME can be used to supply supp lemental oxygen to an intubated patien t or

patient wi th a supraglott ic airway by connecting oxygen tubing to the gas sampling

port (54,55,56).

Contraindications Contraindicat ions inc lude pat ients with thick , cop ious, or bloody secre tions and

patients with a leak tha t prevents exhaled gas f rom travers ing the passive

humidif ier (e.g., bronchopleuralcutaneous f is tula or leaking or absent tracheal tube

cuff ). HMEs should be used with caut ion when weaning a patient from respiratory

support (57,58).

Factors Affecting Moisture Output

Heat and Moisture Exchanger Type Composi te hygroscop ic HMEs have bette r heat and moisture exchanging properties

than do hydrophobic ones.

Initial Humidity Increasing the humidi ty in the gas entering the HME f rom the breathing sys tem wi ll

increase the inspired humidi ty (59).

Inspiratory and Expiratory Flows The faster that gas passes through the HME, the less time there is to evaporate or

deposi t moisture, so a la rge t idal volume may cause the humidi ty of the inspired

gas to fa ll (45,48,59,60,61).

System Continuity A leak around the tracheal tube wi l l resul t in decreased inspired humidi ty (32,62).

Use The HME selec ted should be of an appropriate size fo r the patien t's tidal volume. If

a small HME is used in a large pat ient, the HME wi ll be ineff ic ient (63). Connecting

Page 10: Chapter 11.Humidification Equipment

more than one HME in series wil l improve perfo rmance (64 ,65). Care must be taken

that the uni ts a re pushed f irmly together and that the increase in dead space is not

P.300

excess ive for that particular patien t. Added dead space is especially important in

smal l patien ts.

View Figure

Figure 11.1 Heat and moisture exchangers. A,E: Straight variety. B: Right angle HME with port for aspiration of respiratory gases on the breathing system side. C: The flexible tube attached to the HME extends the distance between the patient and the breathing system and allows the angle between the breathing system and the patient to be altered. Because this HME has significant dead space, it should be used only with high tidal volumes and controlled ventilation with monitoring of inspired and exhaled carbon dioxide. D: Hydrophobic HME with respiratory gas aspiration port. (Pictures C, D, and E courtesy of Gibeck Respiration, Pall Biomedical Products Corp. and ARC Medical Inc.)

An HME shou ld be v isible and accessib le at all times in o rder to de tec t

contamination or d isconnection. The greates t inspired relative humidity occurs wi th

the HME posi tioned next to the tracheal tube, mask, o r suprag lott ic ai rway device .

Some gas moni tors (Chapter 22) a re part icula rly sensit ive to water. I f the sampl ing

l ine is on the machine s ide of the HME, the amount of moisture to which the

moni tor is exposed wi l l be reduced.

An HME can be used with any breathing system. With the Mapleson sys tems, dead

space can be reduced by

P.301

ut i l izing the gas sampl ing port as the fresh gas inlet (66). Oxygen can be

administered through the gas sample port on the HME (55,56). An HME may be

used for pat ien ts who have a tracheostomy (67).

TABLE 11.2 Comparison of Hygroscopic and Hydrophobic Heat and Moisture Exchangers

Page 11: Chapter 11.Humidification Equipment

Type Hygroscopic Hydrophobic

Heat and moisture exchanging efficiency

Excellent Good

Effect of increased tidal volume on heat and moisture exchange

Slight decrease Significant decrease

Filtration efficiency when dry Good Excellent

Filtration efficiency when wet Poor Excellent

Resistance when dry Low Low

Resistance when wet Significantly increased

Slightly increased

Effect of nebulized medications Greatly increased resistance

Little effect

An HME may be used as the so le source of humidi ty o r may be combined wi th

another source such as an unheated humidif ier (68) but should not be used wi th a

heated humidif ier.

I f a nebu lizer o r metered-dose inhaler (Chapter 7) is used to deliver medication, i t

should be inserted between the HME and the pat ient or the HME removed f rom the

c i rcui t during aerosol treatment.

An HME shou ld be replaced i f contaminated with secretions.

Advantages HMEs are inexpensive, easy to use, small, l ightweight, re liable, s imple in design,

and s ilent in operation. They have low resistance when dry. They do not require

water, an external source of energy, a temperature monitor, or alarms. There is no

danger of overhydra tion, hyperthermia, sk in or respira tory tract burns, or electrical

shock. Their use may inc rease the correlat ion between esophageal and core

Page 12: Chapter 11.Humidification Equipment

temperatures (69). They act as a barrie r to large partic les, and some are eff ic ient

bacterial and v iral f i l ters, al though their role in reducing nosocomial infec tions

remains controvers ial . They may reduce problems caused by humidi ty in the

breathing sys tem such as obs truct ion of l ines and venti lato r mal-funct ion (70).

Disadvantages The main disadvantage of HMEs is the limi ted humidity tha t these dev ices can

deliver. The ir contribution to temperature preservation is not sign if icant.

Tempera ture management is discussed in Chapter 31. Ac tive heating and

humidif ica tion are more effective than an HME in retaining body heat, allev ia ting

thick sec retions, and preventing tracheal tube blockage

(19,21,22,23,71,72,73,74,75). The d if fe rence is more apparent af te r intubation

last ing for several days.

Placing an HME between the breathing system and the pat ient increases dead

space. This may necessi ta te an increase in tida l volume and can lead to dangerous

rebreathing (76). I t also increases the work of breathing during both inspirat ion and

exhalation (77).

Hazards

Excessive Resistance The use of an HME increases resistance, although usual ly it is not a major

component of the to ta l work of breathing (78). Resis tance increases with use

(9,33,35,38,58,78,79,80,81,82,83,84,85,86). Heavy v iscous sec retions can greatly

increase resistance. An HME should no t be used with a heated humidif ier, as this

can cause a dangerous increase in resistance. Nebulized medication increases the

resistance of hygroscopic HMEs (34,38).

With a Mapleson system, increased resis tance may cause divers ion of f resh gas

down the expiratory l imb (87,88,89).

High resis tance may resul t in suffic ient back pressure to prevent the low ai rway

pressure alarm f rom be ing activa ted if there is a disconnec tion between the pat ient

and the HME (90,91).

I f increased resistance is suspec ted during control led vent ilation, the peak pressure

should be measured wi th and without the HME in p lace. Spontaneously breathing

patients should be observed for s igns o f increased work of b reathing .

Airway Obstruction

Page 13: Chapter 11.Humidification Equipment

An HME can become obs tructed by f luid, b lood, secretions, a manufac turing defect,

or nebulized drugs (18,92,93,94,95,96,97,98,99,100,101,102). Parts may become

detached and block the breathing system (103). The HME's weight may cause the

tracheal tube to kink.

I f an HME is used for long-term ventilat ion , tracheal tube occlusion may occur

(20,21,22,23,24,73).

Inefficient Filtration Liquid can break through a hygroscop ic HME, result ing in poor f i l trat ion (34,36).

P.302

Foreign Particle Aspiration Parts of the HME may become detached (64,104,105,106,107). The parts may then

be inhaled by the pat ient.

Rebreathing The HME dead space may cause excessive rebreathing, especial ly wi th small tidal

volumes. Special low-volume dev ices are availab le for pedia trics. Even these smal l

devices may be too la rge for infants under 15 kg (108). HMEs should not be used

for mask vent ilation in smal l infants (109).

Leaks and Disconnections Adding an HME to a breathing system increases the potent ia l fo r disconnect ions

and leaks (18 ,92,110,111,112).

Hypothermia Patient warming is discussed in Chapter 31. HMEs are a means to conserve

temperature , but the amount of heat p reserved by th is method is smal l

(15,113,114,115).

Dry Carbon Dioxide Absorbent HMEs wi l l decrease the amount of humidity available to the absorbent. The ex tent

to wh ich HMEs wil l lead to absorbent desiccat ion is unclear. The effec t of dry

carbon d ioxide absorbents on the produc tion of Compound A and carbon monoxide

is discussed in Chapter 9. In add it ion, dry absorbent wi l l absorb some volati le

agents. This can impede anesthetic induct ion with these agents (18).

Humidifiers

Page 14: Chapter 11.Humidification Equipment

A humidif ier (vaporizer o r vaporizing humidif ier) passes a stream of gas over water

(pass -over), ac ross wicks dipped in water (b low-by), o r through water (bubble or

cascade). Humidif iers may be heated or unheated.

Unheated Most unheated humidif iers are disposable, bubble-th rough devices that are used to

increase the humidi ty in oxygen suppl ied to patients v ia a face mask or nasa l

cannula. They cannot deliver more than about 9 mg H2O/L.

Heated Heated humidif ie rs incorporate a device to warm the water in the humidif ier. Some

also heat the inspiratory tube.

Description Humidification Chamber The humid if icat ion chamber contains the l iquid water. It may be d isposable or

reusable . A clear chamber makes i t easy to check the water level . Some humidifiers

have an in tegral or remote reservoir that supplies l iqu id water to the humid if icat ion

chamber (Fig. 11.2).

Heat Source Heat may be supplied by heated rods immersed in the water o r a pla te at the

bottom of the humidif icat ion chamber (Fig. 11.3).

Inspiratory Tube The insp iratory tube conveys humidif ied gas f rom the humidif ier ou tlet to the

patient. If i t is not heated, the gas wil l cool and lose some of i ts moisture as i t

t ravels to the patien t. This water wil l co llect in the inspiratory tub ing. A water trap

wi l l be necessary to collec t the condensed water.

Heating or insulating the inspiratory tube allows more prec ise control of the

temperature and humidity del ivered to the pat ient and avoids moisture rainout

(116,117). A heated wire ins ide the inspira tory tubing is mos t of ten used (Fig.

11.3). I t should ex tend as c lose to the patient connect ion as possible. Disposab le

wi res in preassembled disposable breathing systems are available. A reusable wire

must be inserted manual ly into the insp iratory tube by using a draw wire (118). If

the gas temperature de livered to the patient is set above the temperature of the

humidif ica tion chamber outlet , less than 100% rela tive humidi ty wi l l be del ivered

(119,120,121,122).

Page 15: Chapter 11.Humidification Equipment

Temperature Monitor(s) Most heated humidif iers have a means to measure the gas temperature at the

patient end of the breath ing system. Usually , there are temperature sensors in the

water reservoir o r in contact wi th the heater pla te to activate alarms and shut off

heater power when necessary.

Thermostat

Servo-controlled Units A servo-controlled unit automatically regulates power to the heating element in the

humidif ier in response to the temperature sensed by a probe near the pat ient

connec tion or the humidifier outlet (123,124,125). These devices are equipped wi th

alarms to warn of high temperature. Often, there are two thermos ta ts so that if one

fails , the other wil l cu t off the power before a dangerous temperature is reached.

Nonservo-controlled Units A nonservo-controlled un it p rov ides power to the heat ing element according to the

se tt ing of a contro l, irrespec tive of the delivered tempera ture. I t may inc lude a

servo-control led ci rcuit , but the servo-control led uni t controls the heater rather than

the del ivered tempera ture (123,125).

Controls Most humidif iers a llow temperature selec tion at the end of the delivery tube or at

the humidi ficat ion chamber out let. Some al low less than 100% relative humidi ty to

be de livered. Some models genera te saturated vapor only at a preset tempera ture

(8).

Alarms Alarms may warn when the temperature at the patient end of the c i rcuit deviates

f rom the set temperature by a f ixed amount, when the temperature probe is no t in

place, when the heater wire is not connected, when the water level in the

humidif ica tion chamber is low, or when the ai rway tempera ture probe does not

sense an increase in temperature wi thin a certa in t ime af te r the humid if ie r is turned

ON. A low temperature a la rm wil l help to detect problems wi th the heater element.

I t

P.303

also may be the means of detect ing lack of gas f low in the c i rcu it (126).

Page 16: Chapter 11.Humidification Equipment

View Figure

Figure 11.2 Heated humidifier with separate water reservoir. (Picture courtesy of Hudson RCI.)

Action Some humidif iers heat the gas to a temperature exceed ing the desired patient

ai rway tempera ture (superheat ing) so that the cooling that occurs as i t f lows to the

patient wi l l resul t in the desired temperature at the patient connection. Coo ling in

the tube wil l resul t in water ra ining out in the tube. In other humidif iers,

temperature increases as i t passes th rough the inspiratory tube so tha t gas wi th

less than 100% relative humidi ty is del ivered.

View Figure

Figure 11.3 Heated humidifier. Heat is supplied from a heated plate below the humidification chamber. The heating wire at the left fits inside the delivery tube.

Page 17: Chapter 11.Humidification Equipment

I f the delivery tube is not heated, the tempera ture wil l d rop as i t f lows to the

patient. The magnitude of the drop depends on many factors , including ambient

temperature ; gas f low; and the length, diameter, and thermal mass of the breathing

system. Cool ing can be reduced by shortening or insula ting the del ivery tube or by

us ing higher inspiratory f lows. I f the gas is saturated at the humidif ier outlet , the

temperature drop wi l l cause water vapor condensation (rainout) to occur. A water

trap wi l l be necessary to col lec t the condensed water.

Standard Requirements An international and a U.S. standard on humidif iers have been pub lished (127,128).

They contain the fol lowing prov is ions.

P.304

• Humidif iers must be capable of p roducing an output of at least 10 mg H2O/L.

Those in tended for use wi th patien ts whose supraglottic ai rways have been

bypassed must be capab le of producing an output of at least 33 mg H2O/L.

• The average temperature at the delivery tube out le t shal l not f luc tuate by

more than 2°C from the se t temperature af te r a s ta te of thermal equil ibrium

has been established following a change in gas f low or set temperature. If

the measured gas tempera ture differs f rom the set temperature by more than

the range specif ied by the manufacturer, an ala rm must be activated.

• The volume of l iquid ex iting the humidif ie r shal l no t exceed 1 mL/minute or

20 mL/hour for humidif iers intended for use wi th neonates or 5 mL/minute or

20 mL/hour for all other humidif iers.

• I f the humid if ie r is heated, the gas temperature a t the delivery tube outlet

shall not exceed 41°C or the gas temperature a t the humidif ier out let shall be

indicated continuously and the tempera ture-measuring dev ice shal l ac tivate

an alarm when the temperature exceeds 41°C. The humidif ie r shal l interrupt

heating when the measured gas temperature exceeds 41°C.

• The accessible surface temperature of the del ivery tube mus t not exceed

44°C within 25 cm of the patien t connection port.

• When the humidifier is ti l ted 20° from its normal operating posit ion, there

shall be no water sp il led into the breathing system.

Page 18: Chapter 11.Humidification Equipment

• All calibrated contro ls and ind ica tors sha ll be accurate to wi thin 5% of their

full -scale values, except for temperature displays and contro ls. The

measured gas temperature shall be accura te to ±2°C.

• The direct ion of f low must be marked on humidif iers with f low-direc tion–

sensi tive components .

Use In the c i rc le system, a heated humidif ie r is placed in the inspira tory limb

downstream of the unidirect ional valve by using an accessory breath ing tube. A

heated humidif ier must not be placed in the exp iratory l imb (129,130). I f a fi l ter is

used in the breathing system, i t mus t be placed upstream of the humidif ie r to

prevent i t f rom becoming c logged.

In Mapleson systems (Chapter 8), the humid if ie r is usually placed in the f resh gas

supply tube (131,132). Using a large-diameter tubing and plac ing the humidif ier

near the end of the tube wil l decrease condensation. The delivery tube temperature

probe may be placed ei ther between the fresh gas supp ly tube and the T-piece or

between the T-piece and the pat ient.

The humid if ie r must be lower than the patient to avoid the risk of water running

down the tubing in to the pat ient. The condensate mus t be drained periodical ly o r a

water trap inserted in the most dependent part of the tubing to prevent blockage or

aspirat ion.

The heater wire in the delivery tube should no t be bunched, but s trung evenly along

the length of the tube. The de livery tube should no t rest on other surfaces or be

covered with sheets , blankets , or other materials . A boom arm or tube tree may be

used for support (133).

Advantages Most heated humidif iers are capable of delivering sa turated gas at body

temperature or above, even with high f low rates. A heated humidif ier can produce

more effect ive humidif icat ion than an HME (74,134). Some (but not all ) can be used

for spontaneously breath ing and tracheotomized pat ients (67).

Disadvantages Humidif iers are bu lky and somewhat complex. These devices involve high

maintenance cos ts, electrical hazards , and increased work (temperature control ,

ref i l l ing the reservoir, draining condensate , c lean ing , and steri l ization). Their use is

associated wi th higher cos ts than HMEs.

Page 19: Chapter 11.Humidification Equipment

Compared wi th c irculat ing water and forced-air warming (Chapter 31), the heated

humidif ier offers relat ively l i tt le protect ion against heat loss during anesthes ia

(135).

Hazards Infection Bac terial growth can occur in water stored in a reservoir or the condensate in the

delivery tube. The use of a heated c ircui t reduces the amount of condensate, which

may reduce the infect ion risk.

Breathing System Problems Reported breathing system problems include st icking valves, leaks , disconnec tions ,

incorrect connec tions , obs tructed f resh gas l ine or inspiratory l imb, no ise, and

c logged f i l ters and HMEs (136,137,138,139,140,141).

The delivery tubing may mel t, resu lt ing in an obstruction or leak

(118,142,143,144,145,146,147,148,149,150). Fires have been reported

(123,151,152). A charred breathing sys tem may resul t in fumes entering the

patient 's lungs. Overheat ing of breath ing c ircui ts with melt ing may be caused by

defec ts in or damage to the heated wire; bunching of the heated-wire coi ls within

the breathing system; elec trical incompatibil i ty between the heated-wire breathing

c i rcui t and the humidi fier (153); opera ting the device outside the specified range of

f lows or minute volumes; or covering the del ivery tube wi th sheets, blankets, or

other materials (133,144,147). Problems may occur if a c ircui t that is not from the

humidif ier manufacturer is used. Electrical connectors for heated-wi re c ircui ts tha t

are physical ly compatible may not be electrically interchangeable (133). When

venti lation is in terrupted, as when cardiac bypass is begun, the humid if ier should

be turned OFF. Without a gas a ir f low, the temperature at the pat ient end wi l l drop

and the heating element may increase i ts ou tput, causing the attached breathing

c i rcui t to mel t .

P.305

Adding a humidif ie r may change the breathing system volume and compliance

s ignif icant ly (154). This can result in less accurate small tidal volume del ivery.

Venti la to rs that ca lculate the sys tem compliance and gas compression mus t

perform their checkout procedure wi th the humidif ier in place s ince the humidif ier

can affect vent ilator accuracy.

Page 20: Chapter 11.Humidification Equipment

Water Aspiration There is danger of l iquid water entering the trachea and drowning the pat ient or

causing a burn in the respiratory tract. These risks can be decreased by instal l ing a

water trap in both the inspiratory and exha la tion sides in the mos t dependent

port ion of the breathing tube, draining condensate frequently, and placing the

humidif ier and breathing tubes below the patien t.

Overhydration A heated humidi f ier can produce a posi tive water balance and even overhydration .

Al though mos t anesthetics are of suff icient ly short duration that this is no t a

s ignif icant problem, i t can be a prob lem wi th infants.

Thermal Injury Delivering overheated gases into the trachea can cause hyperthermia or damage to

the tissues l in ing the tracheobronchial tree. Skin burns have been reported f rom

administering heated oxygen nasally and when cont inuous posi t ive airway pressure

was de livered (125,155,156). Burns can also occur when t issue is in contact with

heated breathing ci rcui ts (157,158).

Overheating inspired gas may be caused by omitting , misplacing, d islodging, or not

fully inserting the ai rway temperature probe or by turn ing the humidif ier ON wi th a

low gas f low (123,155,156,159). A temporary increase in inspired gas tempera ture

may occur following a period of interrupted f low or an inc reased f low rate (160).

Increased Work of Breathing A heated humidi f ier inc reases resistance (26,161,162). Most cannot be used wi th

spontaneously breathing pat ien ts.

Monitoring Interference A humidif ier may add enough resis tance to prevent a low a irway pressure alarm

f rom being ac tivated if the sensor is upstream of the humidif ie r (91). Some f low

sensors are a ffected by condensat ion, produc ing a false posit ive alarm (163).

Pressure and f low monitoring are discussed in Chapter 23.

High humidi ty can cause prob lems wi th s idestream (asp irat ing) respiratory gas

moni tors (Chapter 22).

Altered Anesthetic Agents Halothane may be altered if i t passes through a humidif ie r whose heating e lement

is in di rect contact wi th the gas at a temperature of 68°C or higher (164).

Equipment Damage or Malfunction

Page 21: Chapter 11.Humidification Equipment

Venti la to rs are sensit ive to rainout caused by water condensation. Signs inc lude

increased res is tance to exhalation, inaccura te pressure and volume measurements ,

autocycl ing, and vent ilator shutdown (70,165). In o rder to prevent these problems ,

a water trap should be used i f water is l ikely to condense in the breathing system.

These need to be inspected regu larly and emptied as needed. The humidif ier needs

to be lower than the patient and venti lato r.

Nebulizers Description A nebu lizer (aerosol generator, a tomizer, nebuliz ing humid if ier) emits water in the

form of an aerosol mist (water vapor plus part iculate water) (166). The most

commonly used are the pneumatical ly d riven (gas-driven, jet , high pressure,

compressed gas) and u ltrasonic nebulizers. Both can be heated. In addi tion to

providing humidif ica tion, nebul izers may be used to deliver drugs to the breathing

system.

A pneumatic nebulizer works by pushing a jet of high-pressure gas into a l iquid,

inducing shearing forces and breaking the water up in to f ine part ic les. An ul trasonic

nebulizer produces a f ine mis t by subject ing the l iqu id to a h igh-f requency,

electrically driven ultrason ic resonator. The frequency of osci l la tion determines the

s ize of the droplets. There is no need for a driv ing gas. Ul trasonic nebulizers create

a denser mist than pneumatic ones (124).

Use Because a high f low of gas must be used wi th a pneumatic nebul izer, it shou ld be

placed in the fresh gas l ine. An ul trasonic nebulizer can be used in the fresh gas

l ine or the inspiratory l imb.

Hazards Nebulized drugs may obstruct an HME or f i l ter in the breathing sys tem (95,96,167).

Overhydat ion can occur. If the droplets a re not warmed, hypothermia may resul t .

Infection can be transmitted because mic roorganisms can be suspended in the

water droplets.

There are reported cases where a nebulizer was connected di rectly to a trachea l

tube wi thout provis ion for exhalation (168). In one case, th is resu lted in a

pneumothorax.

Advantages

Page 22: Chapter 11.Humidification Equipment

Nebulizers can del iver gases saturated with water without heat and, if desi red, can

produce gases carrying more water.

Disadvantages Nebulizers are somewhat costly. Pneumatic nebu lizers require high gas f lows.

Ul trasonic nebul izers require a source of electric ity and may present electrical

hazards. There may be considerable water deposi tion in the tubings, requiring

f requent d raining, water traps in both the insp iratory and exha lat ion tubes, and

posing the dangers of water d raining into the pat ient or blocking the tub ing.

P.306

References 1. Mart in C, Papazian L, Perrin G, et al . Preservat ion of humidi ty and heat of

respira tory gases in patients wi th a minute venti lat ion greater than 10 L/min. Crit

Care Med 1994;22:1871–1876.

[Med line Link]

2. Branson R, Campbell R, Dav is K, e t al . Anaesthesia c i rcuits , humidity output,

and mucoci l iary structure and func tion. Anaes th Intens Care 1998;26:178–183.

[Med line Link]

3. Wi ll iams R, Rankin N, Smith T, et al . Relat ionship be tween the humidi ty and

temperature of inspired gas and the function of the ai rway mucosa. Crit Care Med

1996;24:1920–1909.

[Fu ll text Link]

[CrossRef]

[Med line Link]

4. Van Heerden PV, Wi lson IH, Marshall FPF, e t al . Effec t of humidif ication on

inhalat ion induct ion with isof lurane. Br J Anaesth 1990;64:235–237.

[CrossRef]

[Med line Link]

5. Cregg N, Wall C, Green D, et al . Humidif icat ion reduces cough ing and breath-

holding during inhalat ion induction wi th isoflurane in children. Can J Anaes th

1996;43:1090–1094.

[Med line Link]

6. Chatburn RL, Primiano FP. A rationa l bas is for humidi ty therapy. Respir Care

1987;32:249–254.

Page 23: Chapter 11.Humidification Equipment

7. Branson R, Campbell R, Chatburn R, et al. Humid if ication during mechanical

venti lation . Respir Care 1992;37:887–890.

[Med line Link]

8. Shel ly MP, Lloyd GM, Park GR. A review of the mechanisms and methods of

humidif ica tion of inspired gases. Intens ive Care Med 1988;14:1–9.

[CrossRef]

[Med line Link]

9. Branson RD, Dav is K Jr., Campbell RS, et al . Humidif icat ion in the in tensive care

unit . Prospective study o f a new protocol u ti l izing heated humid if ication and a

hygroscop ic condenser humid if ier. Chest 1993;104:1800–1805.

[CrossRef]

[Med line Link]

10. Haslam KR, Nielsen CH. Do passive heat and moisture exchangers keep the

patient warm? Anesthesiology 1986;64:379–381.

11. Pf lug AE, Aasheim GM, Foster C, et al . Prevention of post-anaesthesia

sh ivering. Can Anaesth Soc J 1991;25:41–47.

12. Hedley RM, Allt -Graham J . Heat and mois ture exchangers and breathing f i l ters.

Br J Anaesth 1994;73:227–236.

[CrossRef]

[Med line Link]

13. Kulkarni P, Webster J, Carli F. Body heat transfer during hip surgery using

ac tive core warming. Can J Anaesth 1995;42:571–576.

[Med line Link]

14. Johansson A, Lundberg D, Luttropp HH. The effec t of heat and mois ture

exchanger on humidity and body tempera ture in a low f low anaesthesia sys tem.

Acta Anaesthesiol Scand 2003;47:564–568.

[CrossRef]

[Med line Link]

15. Gregorini P , Cang ini D. Control of body temperature during abdominal aort ic

surgery. Acta Anaesthesiol Scand 1996;40:187–190.

[Med line Link]

16. Frank SM, Hesel TW, El-Rahmany HK, et al . Warmed humidif ied inspired

oxygen accelerates postoperat ive rewarming. J Clin Anes th 2000;12:283–287.

[CrossRef]

[Med line Link]

Page 24: Chapter 11.Humidification Equipment

17. Goldberg ME, Epstein R, Rosenblum F, et al. Do heated humidif iers and heat

and moisture exchangers prevent temperature drop during lower abdominal

surgery? J Cl in Anesth 1992;4:16–20.

18. Lawes EG. Hidden hazards and dangers associa ted wi th use of HME/f i l ters in

breathing c i rcuits . Their effec t on toxic metabolite produc tion, pulse oximetry and

ai rway res is tance. Br J Anaes th 2003;91:249–264.

[Fu ll text Link]

[CrossRef]

[Med line Link]

19. Jaber S, Pigeot J, Fodil R, et al . Long-term effects of dif fe rent humidif icat ion

systems on endotracheal tube patency. Evaluat ion by the acous tic ref lection

method. Anes thesiology 2004;100:782–788.

[Fu ll text Link]

[CrossRef]

[Med line Link]

20. Cohen IL, Weinberg PF, Fein IA, et al. Endotracheal tube occlusion associated

wi th the use of heat and mois ture exchangers in the intensive care uni t. Crit Care

Med 1988;16:277–279.

[CrossRef]

[Med line Link]

21. Roustan JP, Kienlen J, Aubas P, et al. Comparison of hydrophob ic heat and

moisture exchangers with heated humidif ie r during prolonged mechanical

venti lation . In tensive Care Med 1992;18:97–100.

[CrossRef]

[Med line Link]

22. Vil lafane MC, Cinnel la G, Lofaso F, e t al . Gradual reduction of endotracheal

tube diameter during mechanical vent ilation v ia d if fe rent humidi ficat ion devices.

Anesthesiology 1996;85:1341–1349.

[Fu ll text Link]

[CrossRef]

[Med line Link]

23. Mart in C, Perrin G, Gevaudan MJ, et al. Heat and mois ture exchangers and

vaporis ing humidif ie rs in the intensive care uni t . Ches t 1990;97:144–149.

[CrossRef]

[Med line Link]

Page 25: Chapter 11.Humidification Equipment

24. Turner DAB, Wright EM. Eff ic iency of heat and mois ture exchangers .

Anaesthesia 1987;42:1117–1118.

[CrossRef]

[Med line Link]

25. Wil l iams R. The effects of excessive humidi ty . Resp ir Care 1998;4 :215–228.

26. Carson KD. Humidi fica tion during anes thesia. Resp Care Clin N Am

1998;4:281–299.

27. Chase HF, Tro tta R, Ki lmore MA. Simple methods for humidifying

nonrebreath ing anesthes ia gas systems. Anes th Ana lg 1962;41:249–256.

[Fu ll text Link]

[CrossRef]

[Med line Link]

28. Grousset D, Mazerol les M, Duterque D, et al. Evaluat ion of the thermic support

provided by a coaxial tube under mechan ical venti lation . Anes thesiology 2003;99:

A605.

29. International S tandards Organizat ion. Anaesthet ic and respiratory equipment—

heat and moisture exchangers (HMEs) fo r humidify ing respired gases in humans.

Part 1: HMEs for use with min imum tidal volumes of 250 ml (ISO 9360-1). Geneva,

Switzerland: Author, 2000.

30. International S tandards Organizat ion. Anaesthet ic and respiratory equipment–

heat and moisture exchangers (HMEs) fo r humidify ing respired gases in humans.

Part 2: HMEs for use with tracheostomized patients hav ing minimum t idal volumes

of 250 ml (ISO 9360-2). Geneva, Switzerland: Author, 2001.

31. Thomachot L , Viv iand X, Boyadkoev O, et al. The combination of a heat and

moisture exchanger and a Booster: a cl in ical and bacteriologica l evaluation over 96

h. Intensive Care Med 2002;28:147–153.

[CrossRef]

[Med line Link]

32. Gedeon A , Mebius C, Palmer K. Neonatal hygroscopic condenser humidif ier.

Crit Care Med 1987;15:51–54.

[CrossRef]

[Med line Link]

33. Wilk inson KA, Cranston A, Hatch DJ, e t al . Assessment of a hygroscopic heat

and moisture exchanger fo r paediatric use. Anaesthesia 1991;46:296–269.

[CrossRef]

[Med line Link]

Page 26: Chapter 11.Humidification Equipment

34. Hedley RM, Allt -Graham J . A comparison of the fi l trat ion propert ies of heat and

moisture exchangers. Anesthesia 1992;47:414–420.

[CrossRef]

[Med line Link]

35. Berry AJ, Nolte FS. An al ternative strategy fo r infec tion control of anes thesia

breathing c i rcuits : a laboratory assessment of the Pal l HME f i l ter. Anesth Analg

1991;72:651–655.

[Fu ll text Link]

[CrossRef]

[Med line Link]

36. Lee MG, Ford JL, Hunt PB, et al . Bacterial retention propert ies of heat and

moisture exchange f i l te rs. Br J Anaes th 1992;69:522–525.

[CrossRef]

[Med line Link]

37. Lei tjen DTM, Re jger VS, Mouton RP. Bacterial contaminat ion and the effect of

f il te rs in anaesthetic ci rcui ts in a s imulated patient mode. J Hosp Infect

1992;21:51–60.

[CrossRef]

[Med line Link]

38. Vandenbroucke-Grau ls CMJE, Teeuw KB, Bal lemans K, et a l. Bacterial and v i ral

removal eff iciency, heat and moisture exchange propert ies of four f i l tration devices.

J Hosp Infect 1995;29:45–56.

[CrossRef]

[Med line Link]

39. Gallagher J , Strangeways JEM, Al lt -Graham J. Contamination control in long-

term ventila tion. A c linical s tudy using a heat-and-moisture-exchanger f i l ter.

Anaesthesia 1987;42:476–481.

[CrossRef]

[Med line Link]

40. Lloyd G, Howells J, Lidd le C, et al . Barriers to hepati tis C transmiss ion wi thin

breathing sys tems: eff icacy of a pleated hydrophobic f i l te r. Anaesth In tens Care

1997;25:235–238.

[Med line Link]

41. Wilkes AR. Evaluat ion of heat and moisture exchangers. Anaes thesia

1991;46:888.

[CrossRef]

Page 27: Chapter 11.Humidification Equipment

[Med line Link]

42. Turnbull D, F isher PC, Mi l ls GH, et al . Performance of breathing fi l ters under

wet condit ions: a laboratory evaluat ion. Br J Anaesth 2005;94:675–682.

[Fu ll text Link]

[CrossRef]

[Med line Link]

43. Anonymous . Evaluat ion report: heat and mois ture exchangers . J Med Eng

Technol 1987;11:117–127.

[Med line Link]

44. Branson RD, Hurst JM. Labora tory evaluation of mois ture output of seven

ai rway heat and moisture exchangers. Resp Care 1987;32:741–745.

45. Cigada M, Elena A, So lca M, e t al . The ef fic iency of twelve heat and moisture

exchangers: an in v itro evaluat ion. Intensive Care World 1990;7 :98–101.

46. Jackson C, Webb AR. An evalua tion of the heat and moisture exchange

performance of four vent i lator c i rcui t f i l te rs. Intens ive Care Med 1992;18:264–268.

[CrossRef]

[Med line Link]

47. Sottiaus T, Mignole t G, Damas P, et al . Comparative evalua tion of three heat

and moisture exchangers during short-term pos toperative mechanica l venti la tion.

Chest 1993;104:220–223.

[CrossRef]

[Med line Link]

48. Eckerbom B, L indholm C-E. Performance evaluation of s ix heat and mois ture

exchangers according to the Draft Internat ional Standard (ISO/DIS 9360). Acta

Anaesthesiol Scand 1990;34:404–409.

[Med line Link]

49. Bickler PE, Sessler DI. Eff iciency of ai rway heat and moisture exchangers in

anesthet ized humans. Anesth Analg 1990;71:415–418.

[Fu ll text Link]

[CrossRef]

[Med line Link]

50. Mebius C. Heat and moisture exchangers with bac terial f il te rs : a labora tory

evaluation . Acta Anaes thesiol Scand 1992;36:572–576.

[Med line Link]

Page 28: Chapter 11.Humidification Equipment

51. Mart in C, Papazian L , Perrin G, et al . Performance evaluation of three

vaporizing humidif ie rs and two heat and moisture exchangers in pa tients with

minute venti lat ion > 10 L /min. Chest 1992;102:1347–1350.

[CrossRef]

[Med line Link]

52. Branson RD, Hurst JM. Labora tory evaluation of mois ture output of seven

ai rway heat and moisture exchangers. Respir Care 1987;37:741–745.

53. Sottiaux T, Mignole t G, Damas P, et al . Comparative evalua tion of three heat

and moisture exchangers during short-term pos toperative mechanica l venti la tion.

Chest 1993;104:220–223.

[CrossRef]

[Med line Link]

54. Chui PT, Poon M. Us ing the Pal l HME f i l te r as an oxygen therapy dev ice in

recovery room. Anaesth In tens Care 1996;24:514.

[Med line Link]

55. Wilkes AR, Vaughan RS. The use of breathing sys tem fi l te rs as oxygen-de livery

devices. Anaes thesia 1999;54:552–558.

[Fu ll text Link]

[CrossRef]

[Med line Link]

56. Orme RMLE, Will iams M. Supp lementary oxygen and the laryngeal mask

ai rway—evaluation of a heat-and-moisture exchanger. Anaesth In tens Care

1999;27:509–511.

[Med line Link]

57. Pelosi P , Solca M, Ravagnan I , et a l. Effects of heat and moisture exchangers

on minute vent i lat ion, vent ilatory drive, and work o f breathing during pressure-

support venti lation in acute respiratory failure. Cri t Care Med 1996;24:1184–1188.

[Fu ll text Link]

[CrossRef]

[Med line Link]

58. Boots RJ, George N, Faoagali JL, e t al . Doub le-heater-wire c i rcu its and heat-

and-moisture exchangers and the risk of vent i la tor-associa ted pneumonia . Cri t Care

Med 2006;34:687–693.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 29: Chapter 11.Humidification Equipment

59. Wilkes AR. The moisture-conserv ing perfo rmance of breathing sys tem f i l te rs in

use wi th s imulated c i rc le anaes thesia breathing systems. Anaesthesia 2004;59:

271–277.

[Fu ll text Link]

[CrossRef]

[Med line Link]

60. Branson RD, Campbell RS, David K J r. Effect of expira tory f low on moisture

output of passive humidif iers as measured by the ISO 9360 standard . Respir Care

1987;42: 960–964.

61. Wilkes AR. The moisture-conserv ing perfo rmance of breathing sys tem f i l te rs

during the f i rs t three minutes of simulated use. Anaes thesia 2004;59:265–270.

[Fu ll text Link]

[CrossRef]

[Med line Link]

62. Til l ing SE, Hayes B. Heat and moisture exchangers in arti f ic ial vent i lat ion. Br J

Anaesth 1987;59:1181–1188.

[CrossRef]

[Med line Link]

63. Gedeon A , Mebius C. The hygroscopic condenser humidif ier. A new device for

general use in anaesthesia and intensive care . Anaesthesia 1979;34:1043–1047.

[CrossRef]

[Med line Link]

64. Anonymous . Heat and mois ture exchangers. Heal th Devices 1983;12:155–167.

65. Shanks CA, Sara CA. A reappraisal of the mul t ip le gauze heat and moisture

exchanger. Anaesth Intens Care 1973;1:428–432.

[Med line Link]

66. Jerwood DC, Jones SEF. HME f i lter and Ayre 's piece. Anaesthes ia

1995;50:915–916.

[CrossRef]

[Med line Link]

67. Thomachot L , Viv iand X, Arnaud S, et al. Preservat ion of humidi ty and heat of

respira tory gases in spontaneously breathing tracheotomized patients. Acta

Anaesth Scand 1998;42:841–844.

[Med line Link]

Page 30: Chapter 11.Humidification Equipment

68. Suzukawa M, Usuda Y, Numata K. The effects on sputum charac teris t ics of

combining an unheated humid if ie r with a heat-moisture exchanging f il ter. Respir

Care 1989;34:976–984.

69. Siegel MN, Gravenstein N. Passive warming of airway gases (art if ic ia l nose)

improves accuracy of esophageal temperature moni toring . J Clin Moni t 1990;6 :89–

92.

[Med line Link]

70. Cantil lo J, Domsky R, Gra tz I , et al . Venti lato ry failures wi th the Datex-Ohmeda

7900 SmartVent. Anesthesiology 2002;96:766–768.

[Fu ll text Link]

[CrossRef]

[Med line Link]

71. Bissonnette B, Sessler DI, LaFlamme P. Passive and act ive inspired gas

humidif ica tion in infants and children. Anesthesio logy 1989;71:350–354.

[Fu ll text Link]

[CrossRef]

[Med line Link]

72. Bissonnette B, Sessler DI. Passive or ac tive inspired gas humidif ica tion

increases thermal steady-state temperatures in anesthetized infants.

Anesthesiology 1989;69:783–787.

73. Misset B, Escudier B, Rivara D, et al . Heat and moisture exchanger vs heated

humidif ier during long-term mechanical venti lation. A prospect ive randomized

study. Ches t 1991;100:160–163.

[CrossRef]

[Med line Link]

74. Edwards EA, Byrnes CA. Humidif icat ion diff icul ties in two tracheostomized

ch ildren. Anaesth In tens Care 1999;27:656–658.

[Med line Link]

75. Nakagawa NK, Macchione M, Petrol ino HM, e t al . Effec ts of a heat and mois ture

exchanger and a heated humidifier on respiratory mucus in pa tients undergoing

mechanical vent i la tion. Cri t Care Med 2000;28:312–317.

[Fu ll text Link]

[CrossRef]

[Med line Link]

76. Mason DG, Edmondson L, McHugh P. Humidif ier-induced hypercarbia .

Anaesthesia 1987;42:672–673.

Page 31: Chapter 11.Humidification Equipment

[CrossRef]

77. Nishimura M, Mishi jima MK, Okada T, et a l. Comparison of f low-resistive work

load due to humidifying devices. Chest 1990;97:600–604.

[CrossRef]

[Med line Link]

P.307

78. Johnson PA, Raper RF, Fisher MM. The impact of heat and mois ture

exchanging humidif iers on work of b reath ing . Anaesth Intens Care 1995;23:697–

701.

[Med line Link]

79. Ploysongsang Y, Branson R, Rashk in MC, e t al . Pressure f low charac teris t ics of

commonly used heat-moisture exchangers. Am Rev Respir Dis 1988;138:675–678.

[Med line Link]

80. Ploysongsang Y, Branson RD, Rashkin MC, e t al . Effec t of f lowrate and

durat ion of use on the pressure drop ac ross s ix a rt if ic ial noses. Respir Care

1989;34:902–907.

81. Manthous CA, Schmidt GA. Resist ive pressure of a condenser humidifier in

mechanical ly vent i lated patients. Crit Care Med 1994;22:1792–1795.

[Med line Link]

82. Costigan SN, Snowdon SL. Breathing system f i lters can affec t the performance

of anaes thetic moni to rs . Anaesthesia 1993;48:1015–1016.

[Med line Link]

83. Chiaranda M, Verona L, Pinamonti O, et a l. Use of heat and moisture

exchanging (HME) f i l ters in mechanical ly vent i lated ICU patients: inf luence on

ai rway f low-res is tance. Intensive Care Med 1993;19:462–466.

[CrossRef]

[Med line Link]

84. Mart inez FJ, Pie tchel S, Wise C, et al. Inc reased resistance of hygroscopic

condenser humidif ie rs when using a c losed c i rcui t suc tion system. Cri t Care Med

1994;22:1668–1673.

[Med line Link]

85. Schumann W, Schumann CM, Fuchs J, e t al . Sudden upper ai rway obstruc tion

due to inv isible rain-out in the heat and moisture exchange f i l ter. B r J Anaes th

2002;89:335–336.

Page 32: Chapter 11.Humidification Equipment

[Fu ll text Link]

[CrossRef]

[Med line Link]

86. Wilkes AR. The abi li ty of breathing sys tem f i l te rs to prevent l iquid

contamination of b reathing systems: a laboratory s tudy. Anaesthes ia 2002;57:33–

39.

[Fu ll text Link]

[CrossRef]

[Med line Link]

87. Da Fonseca JMG, Wheeler DW, Pook JAR. The effect of a heat and mois ture

exchanger on gas f low in a Mapleson F breathing system during inhalat ional

induction. Anaes thesia 2000;55:571–573.

[Fu ll text Link]

[CrossRef]

[Med line Link]

88. Goddard JM, Bennett HR. F il te rs and Ayre 's T-piece. Anaes thesia

1996;51:605.

[Fu ll text Link]

[CrossRef]

[Med line Link]

89. Marcus R, Wandless J, Thompson J . Filters and Ayre's T-piece. Paedia tr

Anaesth 1999;9:93.

[CrossRef]

[Med line Link]

90. Mil l igan KA. Disablement of a venti lato r disconnec t alarm by a heat and

moisture exchanger. Anaesthes ia 1992;47:279.

[CrossRef]

[Med line Link]

91. Slee TA, Pav lin EG. Fai lure of low pressure alarm associa ted wi th the use o f a

humidif ier. Anesthesiology 1988;69:791–793.

[Fu ll text Link]

[CrossRef]

[Med line Link]

92. Prasad KK, Chen L. Compl icat ions re lated to the use of a heat and moisture

exchanger. Anes thesiology 1990;72:958.

[Med line Link]

Page 33: Chapter 11.Humidification Equipment

93. Anonymous . Heat & moisture exchangers block breathing c i rcuit ai rway. Biomed

Safe Stand 1996;28:43.

94. Prados W. A dangerous defect in a heat and moisture exchanger.

Anesthesiology 1989;71:804.

[Fu ll text Link]

[CrossRef]

[Med line Link]

95. Anonymous . Heat/moisture exchange humid if ie rs. Technol Anesth

1991;11(8):5.

96. Anonymous . Humidif iers , heat/moisture exchange. Techno l Anes th 1991;12:5.

97. Anonymous . Hazard not ice. Anaesthesia 1994;49:563.

[CrossRef]

98. Stacey MRW, Asa i T, Wilkes A, et al. Obstruct ion of a breathing sys tem f il te r.

Can J Anaesth 1996;43:1276.

[Med line Link]

99. Walton JS, Fears R, Burt N, et a l. Intraoperat ive breathing c i rcui t obs truct ion

caused by a lbu terol nebul izat ion. Anesth Analg 1999;89:650–651.

[Fu ll text Link]

[CrossRef]

[Med line Link]

100. Peady CJ. Another report of obstruc tion of a heat and moisture exchange

f il te r. Can J Anesth 2002;49:1001.

101. Warmington A, Peck D. HME plus heated humidif ie r danger. Anaes th Intens

Care 1995;23:125.

[Med line Link]

102. Will iams DJ , Stacey MRW. Rap id and complete occlusion of a heat and

moisture exchange f i l te r by pulmonary edema (c linical report). Can J Anaesth

2002;49:126–131.

[Med line Link]

103. Anonymous . Heat and moisture exchangers and f i l te rs: components may enter

breathing sys tems and cause blockages. Heal th Dev ices Alerts 2004; 28(36):5–6.

104. Anonymous . Humidif iers, heat/moisture exchange. Technol Anesth 1985;6:8.

105. James PD, Gothard JWW. Possible hazard f rom the inserts of condenser

humidif iers. Anaes thesia 1984;39:70.

[CrossRef]

[Med line Link]

Page 34: Chapter 11.Humidification Equipment

106. Cas ta A, Houck CS. Acute intraoperative endotrachea l tube obs truct ion

associated wi th a heat and moisture exchanger in an infant. Anes th Analg

1997;84:939–940.

[Fu ll text Link]

[CrossRef]

[Med line Link]

107. Anonymous . Heat/mois ture exchange humidif iers . Technol Anesth 2004;25(3–

4):11.

108. Booker PD. Equipment and moni to ring in paediatric anaesthesia. Br J Anaes th

1999;82:78–90.

[Med line Link]

109. Lel louche F, Maggiore SM, Deye N, et al . Ef fec t of the humidif ica tion dev ice

on the work of breathing during noninvasive venti lat ion. Intensive Care Med

2002;28:1582–1589.

[CrossRef]

[Med line Link]

110. Bengtsson M, Johnson A. Fai lu re of a heat and mois ture exchanger as a cause

of disconnect ion during anaesthesia . Acta Anaesthesiol Scand 1989;33:522–523.

[Med line Link]

111. Mansor M, Chan L . Cri t ica l incident involv ing a heat and moisture exchanger

wi th attached f lex ible connector. Anaes th Intens Care 1999;27:114–115.

[Med line Link]

112. Wilkes AR, Hampson MA, Mecklenburgh JS. A method of tes ting the

compliance of 15 and 22 mm con ical connec tors on heat and moisture exchangers

to ISO 5356-1:1987. J Med Eng Technol 1997;21:147–150.

[Med line Link]

113. Eckerbom B, Lindho lm CE. Heat and moisture exchangers and the body

temperature : a preoperat ive s tudy. Ac ta Anaesthes iol Scand 1990;34:538–542.

[Med line Link]

114. Goldberg ME, Jan R, Gregg CE, et al . The heat and mois ture exchanger does

not preserve body temperature or reduce recovery time in outpatients undergo ing

surgery and anesthes ia . Anes thesiology 1988;68:122–123.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 35: Chapter 11.Humidification Equipment

115. Bickler PE, Sessler DI. Ef f ic iency of airway heat and moisture exchangers in

anesthet ized humans. Anesth Analg 1990;71:415–418.

[Fu ll text Link]

[CrossRef]

[Med line Link]

116. Branson R. Humidif icat ion for pa tients wi th arti f ic ial airways. Respir Care

1999;44:630–641.

117. Mizutani AR, Ozaki G, Rusk R. Insulated c ircui t hose improves heated

humidif ier performance in anesthesia vent i la tion ci rcuits . Anesth Analg

1991;72:566–567.

[Fu ll text Link]

[CrossRef]

[Med line Link]

118. Anonymous . Heated wires can mel t disposable breathing c ircui ts . Technol

Anesth 1989;9:2–3.

119. Miyao H, Miyasaka K, Hirokawa T, e t al . Cons iderat ion of the in ternat ional

standard for ai rway humidif icat ion using simulated secret ions in an art if ic ia l ai rway.

Respir Care 1996;41:43–49.

120. Levy H, Simpson SQ, Duval D. Hazards of humid if iers wi th heated wires. Cri t

Care Med 1993;21:477–478.

[CrossRef]

[Med line Link]

121. Gilmour IJ , Boyle MJ, Rozenberg A, e t al . The effect of heated wire ci rcuits on

humidif ica tion of inspired gases. Anesth Analg 1994;79:160–164.

[Fu ll text Link]

[CrossRef]

[Med line Link]

122. Wilkes AR. Low re lat ive humidi ty de livered by a humidif ier with a heat ing wire.

Crit Care Med 1993;21:948.

[CrossRef]

[Med line Link]

123. Anonymous . Heated humidi f iers. Technol Anesth 1987;8:1–5.

124. Anonymous . Heated humidi f iers. Heal th Dev ices 1987;16:223–250.

125. Anonymous . An overv iew of heated humidif iers. Techno l Anes th 1994;15(6):1–

5.

Page 36: Chapter 11.Humidification Equipment

126. Russell WJ, Webb RK, van der Walt JH, et al . Problems wi th venti lat ion : an

analys is of 2000 inc ident reports. Anaesth Intens Care 1993;21:617–620.

[Med line Link]

127. Inte rnational Standards Organization. Humidif iers fo r medical use—general

requirements for humid if icat ion systems (ISO 8185). Geneva, Switzerland: Author,

1997.

128. American Society fo r Testing and Materials . Standard speci f ication for

humidif iers fo r medical use—Part 1: General requirements for act ive humidif icat ion

systems (ASTM F-1690). Wes t Conshohocken, PA: Author, 2004.

129. Hawk ins C, Ross A. Unexplained humid if ie r failure. Anaesth Intens Care

1994;22: 739–740.

[Med line Link]

130. Spencer M. Unexplained humid if ier fa ilure. Anaesth Intens Care

1994;22:740.

131. Hannal lah RS, McGil l WA. A pract ical way of using heated humidif iers wi th

pedia tric T-piece systems. Anesthes iology 1983;59:156–157.

[Fu ll text Link]

[CrossRef]

[Med line Link]

132. Kovac AL, Filard i JP, Goto H. Water trap for f resh gas f low l ine o f Bain or

CPRAM circuit . Can J Anaesth 1987;34:102–103.

[Med line Link]

133. Burling ton DB. FDA safety alert. Hazards of heated-wire breathing c ircu its .

Rockvi l le, MD: Food and Drug Administration, 1993.

134. Yam PC, Carl i F. Maintenance of body temperature in e lderly pat ients who

have joint rep lacement surgery. A comparison between the heat and mois ture

exchanger and heated humidif ier. Anaesthesia 1990;45:563–565.

[CrossRef]

[Med line Link]

135. Hynson JM, Sessler DI. Intraopera tive warming therapies : a comparison of

three devices . J Clin Anes th 1992;4:194–199.

[CrossRef]

[Med line Link]

136. McNul ty S, Barringer L, Browder J. Carbon dioxide retent ion associa ted wi th a

humidif ier defect. Can J Anaesth 1987;34:519–521.

[Med line Link]

Page 37: Chapter 11.Humidification Equipment

137. Shrof f PK, Skerman JH. Humidif ier malfunction—a cause of anes thesia c i rcuit

occlusion. Anesth Analg 1988;67:710–711.

[Fu ll text Link]

[CrossRef]

[Med line Link]

138. Shampaine EL, Helfaer M. A modest proposal for improved humidif ier design.

Anesth Analg 1991;72:130–131.

[Fu ll text Link]

[CrossRef]

[Med line Link]

139. Wang J, Hung W, Lin C. Leakage of d isposable breathing c i rcui ts . J Clin

Anesth 1992;4:111–115.

[CrossRef]

[Med line Link]

140. Warmington A, Peck D. Another complicat ion of heated hose humidif ication.

Anaesth Intens Care 1994;22:740–741.

[Med line Link]

141. Beards SC, Payne T. An unexpected complicat ion of heated hose

humidif ica tion. Anaes th Intens Care 1994;22:232.

[Med line Link]

142. Pat i l AR. Mel ting of anesthesia c ircui t by humidif ie r. Another cause of

“vent i lator d isconnect.” Anesth Prog 1989;36:63–65.

[Med line Link]

143. Wong DHW. Mel ted del ivery hose—a complicat ion of a heated humidif ie r. Can

J Anaesth 1988;35:183–186.

[Med line Link]

144. Wood D, Boyd M, Campbell C. Insulat ion of heated wire ci rcui ts . Anes th Analg

1992;74:471.

[Fu ll text Link]

[CrossRef]

[Med line Link]

145. Sprague DH, Maccioli GA. Disposable c i rcui t tubing melted by heated

humidif ier. Anesth Analg 1986;65:1247.

[CrossRef]

[Med line Link]

146. Anonymous . Anesthesia breathing c ircui ts . Technol Anes th 1992;13:8.

Page 38: Chapter 11.Humidification Equipment

147. Anonymous . Incompatibi li ty of disposable heated-wire breathing ci rcui ts and

heated-wire humid if ie rs. Technol Anesth 1993;14(2):4–5.

148. Webb RK, Russell WJ, Klepper I , et a l. Equipment fai lure : an analys is of 2000

incident reports. Anaes th Intens Care 1993;21:673–677.

[Med line Link]

149. Anonymous . Inappropriate Fisher & Paykel heater-wire adapter melts

Allegiance breathing ci rcui t. Health Devices 2000;29(2–3):86–87.

[Med line Link]

150. Anonymous . Anesthesia breathing c ircui ts may overheat & melt tubing . Biomed

Safe Stand 1994;24:85.

151. Anonymous . Breathing circu it heat ing component could short and cause f ire.

Biomed Safe Stand 1990;20:67–68.

152. Anonymous . Fisher & Paykel—dual-heated adul t breathing c ircui ts : r isk of fi re.

Hea lth Devices Alerts 2006;30:4–5.

153. Anonymous . Venti la tor breath ing c ircuits . Technol Anes th 2000;21:7

154. Cote CJ, Petkau AJ, Ryan JF, et al. Wasted vent i lat ion measured in v itro with

eight anes thetic ci rcu its wi th and wi thout inline humidif icat ion. Anesthesiology

1983;59:442–446.

[Fu ll text Link]

[CrossRef]

[Med line Link]

155. Anonymous . Heated humidi f iers can burn infants during CPAP. Heal th Devices

1987;16:404–406.

156. Anonymous . Heated humidi f iers can burn infants during CPAP. Technol Anesth

1988;8:7–9.

157. Anonymous . Possible burn f rom heated breathing c i rcui t. Biomed Safe Stand

1991;21:147.

158. Whi teley SM. A hazard o f heated humidif iers. Anaes thesia 1992;47:909.

[CrossRef]

[Med line Link]

159. Anonymous . Safety act ion bulletin. Anaesthes ia 1992;47:547.

[CrossRef]

160. Smith HS, Allen R. Another hazard of heated water humidif iers. Anaesthesia

1986;41:215–216.

[CrossRef]

[Med line Link]

Page 39: Chapter 11.Humidification Equipment

161. Oh TE, Lin ES, Bhatt S. Resistance of humidifiers, and inspiratory work

imposed by a vent i lator-humidif ier circuit . Br J Anaesth 1991;66:258–263.

[CrossRef]

[Med line Link]

162. Rathgeber J, Kazmaier S, Penack O, et al . Evaluat ion of heated humidif iers fo r

use on intubated patients: a comparat ive s tudy of humidifying eff ic iency, f low

resistance, and alarm func tions us ing a lung model . Intensive Care Med

2002;28:731–739.

[CrossRef]

[Med line Link]

163. Anonymous . Rainout from a F isher & Paykel heated humidif icat ion system can

shut down certain venti la tors. Technol Anesth 2002;22(9):1–2.

164. Karis JH. Al terat ion of halothane in heated humidif iers. Anesth Analg

1980;59:518.

[Fu ll text Link]

[CrossRef]

[Med line Link]

165. Anonymous . Rainout puts venti la tor-dependent patien ts at risk. Technol

Anesth 2003;23:1–2.

166. Peterson BD. Heated humidif iers . Structure and function. Respir Care Cl in N

Amer 1998;4:243–259.

167. Barton RM. Detection of expiratory antibacteria l f i l te r occlus ion. Anesth Analg

1993;77:197.

[Fu ll text Link]

[CrossRef]

[Med line Link]

168. Anonymous. Nebulizers. Technol Anesth 2002;22:10.

P.308

Questions For the fol lowing quest ions, select the correct answer.

1. Which is not an effect of inhaling dry gases?

Drying of the mucosa

Decreased compliance

Page 40: Chapter 11.Humidification Equipment

Development of loci for infection

Impairment of surfactant act iv ity

Decreased alveolar-arteria l oxygen difference

View Answer2. Where should a heated humidifier be located in the circle system? Between the exhalation tubing and the carbon dioxide absorber

Between the Y-piece and tracheal tube

Between the inspiratory tubing and the Y-p iece

Between the absorber and the inspiratory tubing

Between the Y-piece and the mask

View Answer3. If a bacteria l fi lter is used in a circ le system that has a heated humidifier, where should the fi lter be placed?

Between the Y-piece and the tracheal tube

Between the Y-piece and the inhalat ional tubing

Between the inhalat ional tubing and the humidif ier

Between the inhalat ional unid irectional valve and the humidif ier

Between the Y-piece and the exhala tion tubing

View Answer4. Which is the definition of absolute humidity? The pressure exerted by water vapor in a gas mixture

The amount o f water vapor at a part icular temperature as a percentage of the

amount that would be held if the gas were saturated

The max imum amount of water vapor that a volume of gas can ho ld

The mass of water vapor present in a volume of gas.

The humid ity of gases at body temperature

View Answer5. What w ill cause water to rain out in the breathing hose?

Using coaxial tubes in the breath ing system

Heated humidif ied gas that cools in the breathing tube

Warming cooled humidif ied gases in the breathing tube

Maintaining the temperature of heated gas in the breathing tubes

Using an ex tra-long breath ing tube

View AnswerFor the following quest ions, answer

• i f A, B, and C are correct

• i f A and C are correct

• i f B and D are correct

• is D is correct

Page 41: Chapter 11.Humidification Equipment

• i f A, B, C, and D are correct.

6. Which of the following statements about humidity and ambient conditions are true?

I f a gas saturated with water is heated, i t can hold more water

Gas that is 100% saturated at room tempera ture and warmed to body temperature

wi l l be about 60% saturated

I f a gas saturated with water is heated, the absolute humid ity remains the same

If a gas saturated with water is heated, the rela tive humidi ty inc reases

View Answer7. A result of the low specific heat of gases is Inhaled gas wi l l quick ly assume body tempera ture

Water wi l l condense in the exhalat ion s ide of a ci rcle sys tem

Gas in breath ing tubes wil l quick ly assume room tempera ture

Gas has a tendency to change temperature s lowly

View Answer8. Heat and moisture are normally lost during anesthesia because Dry gases are suppl ied f rom the anesthesia machine

Gases are heated by the body

Tracheal intubat ion bypasses normal humidif ica tion mechanisms

The trachea l tube does not act to conserve heat or moisture

View Answer9. Sources of humidity in the breathing system include Carbon dioxide absorbent

Exhaled water from a prev ious pat ient

Rebreathing of p rev iously exhaled gases

Fresh gas

View Answer10. Advantages of a hydrophobic HME over a composite hygroscopic HME include Better fi l trat ion

Less res is tance when wet

Nebulized drugs have l it t le ef fect on resistance

Better humidif ica tion

View Answer11. Using a heated breathing tube with a heated humidifier will result in Rainout in the delivery tube

A higher temperature at the Y-piece

Drying of secre tions

Page 42: Chapter 11.Humidification Equipment

A del ivered relat ive humidity of 100%

View Answer12. Hazards of heated humidifiers include

Sticking valves

Overhydra tion

Al te rat ion of anesthet ic agents

Obs truct ion of sides tream gas moni tors

View Answer13. Advantages of HMEs include Fluctuat ing tempera ture and humidi ty

Decrease in dead space

Low compliance

Decreased resistance to breathing

View Answer14. Contraindications to use of an HME include

Bloody secret ions

Pat ient temperature of less than 35°C

Bronchopleurocutaneous f is tula

Uncuffed tracheal tube

View Answer15. When using an HME, the inspired humidity can be in-creased by Use of an uncuffed tube

Increasing the minute volume

Lowering the humidi ty of gas entering the HME

Use of a hygroscopic rather than a hydrophobic HME

View AnswerP.309

16. Which patients or conditions derive the greatest benefit from humidification of inspired gases? Long anes thetics

Pat ients who are l ikely to develop pulmonary complicat ions

Pediatric pa tients

Elderly patients

View Answer17. Which of the following are respiratory complications of dry respiratory mucosa? Bronchoconstric t ion

Airway obstruction

Infection

Page 43: Chapter 11.Humidification Equipment

Atelectasis

View Answer18. Consequences of excessive humidity include which of the following? Pulmonary edema

An increased serum sodium

Ci liary degeneration

Increased pulmonary compl iance

View Answer19. Hazards associated with using a nebulizer inc lude

Overhydra tion

Fi lter obstruction

Bac teria infection

Hypothermia

View Answer