Chapter 22
GC & LC
![Page 2: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/2.jpg)
22.1 Gas Chromatography -1
1. Schematic diagram
![Page 3: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/3.jpg)
22.1 Gas Chromatography -2
2. Columns : open tubular columns
![Page 4: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/4.jpg)
22.1 Gas Chromatography -3A) m.p.(gas) - s.p.
1) s.p.: solid( using adsorption) ex: SiO2
column ages: Si-O-H cause tailing peak.2) s.p.: liquid( GLC, using partition) Table 22-1
Decrease thickness of stationary phase leads to
a) Resolution (H)b) tr c) Sample capacity
![Page 5: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/5.jpg)
![Page 6: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/6.jpg)
b.p.
22.1 Gas Chromatography -4
ex
![Page 7: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/7.jpg)
22.1 Gas Chromatography -5B) The effects of column polarity on separation
Like dissolves like
(a)S.P: nonpolar
b.p. dependent
![Page 8: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/8.jpg)
22.1 Gas Chromatography -6
-one:C=O
-ol: OH
S.P. : Polar
![Page 9: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/9.jpg)
22.1 Gas Chromatography -7
C) Common solid s.p. :
a) Porous carbon : larger molecules bind more tightly than small ones, flexible molecules bind more than rigid ones
b) Molecular sieves : retain & separate small molecules : H2, O2, N2, CO2, CH4. (Fig. 22-5)
![Page 10: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/10.jpg)
22.1 Gas Chromatography -8packed column vs. open tubular column
higher resolutionlower sample capacity
![Page 11: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/11.jpg)
22.1 Gas Chromatography -9
3. Temperature programming
temp of column v.p. solute,
tr
sharpens peaksisothermal : constant temp.temp. programming (gradient) : raise the column temp. during the sepa
ration.
![Page 12: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/12.jpg)
22.1 Gas Chromatography -10
![Page 13: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/13.jpg)
4. Carrier Gas
22.1 Gas Chromatography -11
![Page 14: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/14.jpg)
22.1 Gas Chromatography -12
5. Sample Injection-1
1) gasses, liquids, or solids
vaporized, not decomposition
2) injection time bands broader
3) injected by syringe (manual or automatic injection)
![Page 15: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/15.jpg)
22.1 Gas Chromatography -12
![Page 16: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/16.jpg)
22.1 Gas Chromatography -13
5. Sample injection-2 :
4) operation
a).on-column injection (50 )℃- best for quantitative analysis- thermally sensitive compounds- low resolution
![Page 17: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/17.jpg)
22.1 Gas Chromatography -14b) split injection (350℃) (only 0.1-10% sample)
- concentrated sample
- high resolution
- dirty samples- could cause thermal decomposition
c) splitless injection (220℃) (80%)
- dilute sample
- high resolution
- solvent trapping (Tsolvent < 40℃)- cold trapping (Tsolute < 150℃)
![Page 18: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/18.jpg)
22.1 Gas Chromatography -15
5. Detectors
Qualitative analysis :mass spectrometer, IR
Quantitative analysis :area of a chromatographic peak.
![Page 19: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/19.jpg)
22.1 Gas Chromatography -16
a) Thermal conductivity detector:
-most general way
-responds to everything
-not sensitive enough for high resolution.
b) Flame ionization detector :
-most popular
-mainly responds hydrocarbons (C-H)
![Page 20: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/20.jpg)
22.1 Gas Chromatography -17
c) Electron capture detector :
- for compounds containing atoms with high electron affinities.
- sensitive for halogen, C=O, NOx, & orgaometallic compounds.
d) Other detectors : p 476
![Page 21: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/21.jpg)
22.2 Liquid Chromatography -1
1. open, gravity-feed column 2. closed column (under high pressure)
packed with micron-size particles. (HPLC)
3. stationary phase :
① adsorption : silica (SiO2xH2O), alumina (Al2O3xH2O),
② molecular exclusion, ③ ion-exchange, affinity
![Page 22: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/22.jpg)
22.2 Liquid Chromatography -2compete with ▲ for binding on s.p.
the more strongly bind to s.p.eluent strength
![Page 23: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/23.jpg)
22.2 Liquid Chromatography -3
4. Eluent strength : Table 22.2
The more polar solvent
eluent strength
tr
5. Gradient elution : increased the eluent strength during the separation in liquid chromatography.
![Page 24: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/24.jpg)
![Page 25: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/25.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -1
![Page 26: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/26.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -21. Through a closed column, and needs high
pressure.
2. s.p. particles size (s.p. m.p. faster,
i.e. C in van Deemter eqn.)
resolution
![Page 27: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/27.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -3
![Page 28: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/28.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -4
![Page 29: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/29.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -53. Stationary phase
a) Normal-phase chromatography : polar s.p. and less polar solvent. Eluent strength is increased by adding a more polar solvent.
b) Reversed-phase chromatography : low-polarity s.p. and polar solvent. Eluent strength is increased by adding a less polar solvent.
![Page 30: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/30.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -6
c) Bonded stationary phase. polar vs. nonpolar p.461
d) Optical isomersD- & L-amino acids
![Page 31: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/31.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -7
d) Optical isomers separation for ant-inflammatory drug Naproxen
![Page 32: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/32.jpg)
4. Columna) Guard columnb) Injection valve
22.3 High-Performance Liquid Chromatography (HPLC) -8
![Page 33: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/33.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -9
5. Solvents a) Isocratic elution :
elution with single solvent or a constant solvent mixture
b) Gradient elution : solvent is changed continuously from a weak eluent strength to a strong eluent strength by mixing more and more of a strong solvent to a weak solvent during the chromatography.
![Page 34: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/34.jpg)
22.3 High-Performance Liquid Chromatography (HPLC) -10
A : KH2PO4(aq)
B: CH3CN(l)
![Page 35: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/35.jpg)
22.3 High-Performance Liquid Chromatography (HPLC)-11
Reversed-phase column :eluent strength , solvent polarity , tr
![Page 36: Chapter 22 GC & LC. 22.1 Gas Chromatography -1 1.Schematic diagram](https://reader036.vdocuments.net/reader036/viewer/2022081512/56649db45503460f94aa5467/html5/thumbnails/36.jpg)
22.3 High-Performance Liquid Chromatography (HPLC)-12
The gradient can be used to resolve all peaks by reducing the time from 2 h to 38 min.