cholinesterase fs hitachi 717 - grupo biosys · hitachi 717 chemistry settings temperature: ......

121
845 1401 10 02 03 February 2004/1 HITACHI 717 Chemistry settings Temperature: 37 °C PROGRAM 2: CHEMISTRY PARAMETERS TEST CHE ASSAY CODE 5 30 50 SAMPLE VOLUME (μl) 4 R 1 VOLUME (μl) 200 100 NO R 2 VOLUME (μl) 50 20 NO WAVELENGTH (nm) 0 405 CALIBR. METHOD **2 0 0 STD.1 CONC.-POS. 0 1 STD.2 CONC.-POS. 0 0 STD.3 CONC.-POS. 0 0 STD.4 CONC.-POS. 0 0 STD.5 CONC.-POS. 0 0 STD.6 CONC.-POS. 0 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 1 PROZONE LIMIT 0 EXPECTED VALUE 3930 11500 PANIC VALUE 0 - 25000 INSTRUMENT FACTOR 1.0 #) Data entry by the user *) Enter calibration or standard value and position **) Factor must be checked by a calibration serum CHOLINESTERASE FS Order information Cat. No. 1 1401 .. .. Notes 1. Please refer to the package insert for Cholinesterase FS for detailed information about the test on the following: Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding - Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison Reference Ranges Literature 2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided. 3. Manufactured by DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany

Upload: buiduong

Post on 19-Jul-2018

244 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1401 10 02 03

February 2004/1

HITACHI 717 Chemistry settings Temperature: 37 °C

PROGRAM 2: CHEMISTRY PARAMETERS TEST CHE ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 0 – 405 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 1 PROZONE LIMIT 0 EXPECTED VALUE 3930 – 11500 PANIC VALUE 0 - 25000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position **) Factor must be checked by a calibration serum

CHOLINESTERASE FS Order information Cat. No. 1 1401 .. .. … Notes 1. Please refer to the package insert for Cholinesterase FS for

detailed information about the test on the following:

Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding

- Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison

Reference Ranges Literature

2. The stability of the reagent on board the analyser is at least

one month provided that contamination and evaporation are avoided.

3. Manufactured by

DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany

Page 2: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2701 10 02 10 October 2009/3

ALAT(GPT) FS (IFCC mod.) Cat. No 1 2701.........

Notes 1. Please refer to the package insert for ALAT(GPT)

FS (IFCC Mod.) for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST GPT ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 200 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 6000 – 1 PROZONE LIMIT 0 EXPECTED VALUE # – # PANIC VALUE 0 - 600 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [ALAT] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [ALAT] ASSAY/TIME/POINTS [Rate A] [10] [20] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [200] [0] [00108] [#]

DECREASE [7] [0] [0] [1] [0] [0] R2 [0] [0] [00108] [#] INCREASE [25] [0] [0] [1] [0] [0] R3 [50] [0] [00108] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00108] [#] ABSORBANCE LIMIT [6000] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [ALAT] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [ALAT] REPORT NAME [ALAT] DATA MODE [Aktiv]

TEST NAME [ALAT] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [ALAT] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [15] [15] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 3: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2701 10 02 10 October 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST ALAT # TEST NAME ALAT UNIT U/L

DATA MODE ON BOARD REPORT NAME ALAT

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 600 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST ALAT

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 20 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 7 0 0

S.VOL (INCREASE) 25 0 0

ABS. LIMIT 6000] 1:DECREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 108 #

R2 0 0 108 #

R3 50 0 108 #

R4 0 0 108 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name GPT 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 15 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 6000 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 42 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 4: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0220 10 02 10 September 2006/3

ALBUMIN FS Cat. No 1 0220.........

Notes 1. Please refer to the package insert Albumin FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST ALB ASSAY CODE 1 – 8 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 350 – 100 – No R 2 VOLUME (µl) 0 – 20 – No WAVELENGTH (nm) 700 – 600 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 EXPECTED VALUE 3.5 – 5.2 PANIC VALUE 0.2 – 6.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [ALB] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [ALB] ASSAY/TIME/POINTS [1 point] [10] [6] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [660]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [250] [0] [00377] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00377] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00377] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00377] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [ALB] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [ALB] REPORT NAME [Albumin] DATA MODE [Aktiv]

TEST NAME [ALB] [] CONTROL INTERVAL UNIT [g/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [ALB] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [2] [2] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 5: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0220 10 02 10 September 2006/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST ALB # TEST NAME ALB UNIT g/dl

DATA MODE ON BOARD REPORT NAME Albumin

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

3.5 - 5.2 3.5 - 5.2

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 6.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST ALB

ASSAY CODE 1 point 3 WAVELGTH (SUB / MAIN)

700 / 660

ASSAY POINTS 6 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 377 #

R2 0 0 377 #

R3 0 0 377 #

R4 0 0 377 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name ALB 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 5 5 Assay Point 1 10 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 2.5 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 3.5 55 Expect. Value (H) 5.0 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 6: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0242 10 02 10 August 2008/3

ALBUMIN IN URINE CSF/FS (Microalbumin) Serum application

Cat. No 1 0242.........

Notes 1. Please refer to the package insert for Albumin in

Urine/CSF FS *Microalbumin for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST S-ALB ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3** R 1 VOLUME (µl) 350 – 20 – NO R 2 VOLUME (µl) 70 – 20 – NO WAVELENGTH (nm) 0– 570 CALIBR. METHOD 3 – 4 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999.9 DUPLICATE LIMIT 10000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – # PANIC VALUE # – # INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) For Hitachi 717 calibrators, controls and samples must be prediluted 1+20 before analysing

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [S-Alb] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [S-Alb] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [10] [2] [190] [1] [0] [0] R1 [250] [0] [00005] [#]

DECREASE [10] [2] [190] [1] [0] [0] R2 [0] [0] [00005] [#] INCREASE [10] [5] [190] [1] [0] [0] R3 [50] [0] [00005] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00005] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [S-Alb] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [S-Alb] REPORT NAME [S-Alb] DATA MODE [Aktiv]

TEST NAME [S-Alb] [] CONTROL INTERVAL UNIT [g/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [S-Alb] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [10] [10] [10] [10] [10] [10]

DIL. SAMPLE VOLUME [2] [2] [2] [2] [2] [2] DILUENT VOLUME [190] [190] [190] [190] [190] [190]

#) data entry by the user *) enter calibration or standard value and position

Page 7: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0242 10 02 10 August 2008/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST S-Alb # TEST NAME S-ALB UNIT g/l

DATA MODE ON BOARD REPORT NAME ALBUMIN

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

- -

STD CON POS S.VOL. PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 10 2 190 # (1)

(2) * # 10 2 190 # (2)

(3) * # 10 2 190 (3)

(4) * [# 10 2 190 (4)

(5) * # 10 2 190 (5)

(6) * # 10 2 190 (6)

TEST S-Alb

ASSAY CODE 2 Point end 10 WAVELGTH (SUB / MAIN)

/ 570

ASSAY POINTS 15 - 31 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 10 2 190

S. VOL (DECREASE) 10 2 190

S.VOL (INCREASE) 10 5 190

ABS. LIMIT 0 INCREASE

PROZONE LIMIT 32000 UPPER

REAGENT R1 250 0 005 #

R2 0 0 005 #

R3 50 0 005 #

R4 0 0 005 #

CALIB. TYPE [6: Spline] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999.9

SPAN 0 DUPLICATE LIMIT 10000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name S-ALB 2 Assay Code (Mthd) 2 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 0 10 Wave Leng. (MAIN) 570 11 Sample Volume 3** 12 R1 VOLUME 350 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer **) For Hitachi 902 calibrators, controls and samples

must be prediluted 1+20 before analysing

Page 8: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0242 10 52 10 August 2008/4

ALBUMIN IN URINE/CSF FS (Microalbumin) Urine/CSF application Cat. No 1 0242.........

Notes 1. Please refer to the package insert for Albumin in

Urine/CSF FS *Microalbumin for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST M-ALB ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 12 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 700 – 405 CALIBR. METHOD 3 – 4 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999.9 DUPLICATE LIMIT 10000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – # PANIC VALUE 3 – 350 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [M-Alb] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [M-Alb] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [12] [0] [0] [1] [0] [0] R1 [250] [0] [00005] [#]

DECREASE [6] [0] [0] [1] [0] [0] R2 [0] [0] [00005] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [50] [0] [00005] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00005] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [M-Alb] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [M-Alb] REPORT NAME [M-Alb] DATA MODE [Aktiv]

TEST NAME [M-Alb] [] CONTROL INTERVAL UNIT [mg/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [M-Alb] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [12] [12] [12] [12] [12] [12]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 9: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0242 10 52 10 August 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST M-Alb # TEST NAME M-ALB UNIT mg/l

DATA MODE ON BOARD REPORT NAME MICROALBUMIN

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

22 - 29 22 - 29

TECHNICAL LIMIT CLASS I CLASS 2

- -

STD CON POS S.VOL. PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 12 0 0 # (1)

(2) * # 12 0 0 # (2)

(3) * # 12 (3)

(4) * [# 12 (4)

(5) * # 12 (5)

(6) * # 12 (6)

TEST M-Alb

ASSAY CODE 2 Point end 10 WAVELGTH (SUB / MAIN)

800 / 415

ASSAY POINTS 15 - 31 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 12 0 0

S. VOL (DECREASE) 6 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 0 INCREASE

PROZONE LIMIT 32000 UPPER

REAGENT R1 250 0 005 #

R2 0 0 005 #

R3 50 0 005 #

R4 0 0 005 #

CALIB. TYPE [Spline] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999.9

SPAN 0 DUPLICATE LIMIT 10000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name MALB 2 Assay Code (Mthd) 2 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 415 11 Sample Volume 12 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 10: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0401 10 02 10 October 2009/4

ALKALINE PHOSPHATASE FS DGKC Cat. No 1 0401.........

Notes 1. Please refer to the package insert for Alkaline

Phosphatase FS DGKC for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST ALP ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 660 – 405 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 13000 – 0 PROZONE LIMIT 0 EXPECTED VALUE 0 – 258 PANIC VALUE 0 - 800 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [AP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [AP] ASSAY/TIME/POINTS [Rate A] [10] [20] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [200] [0] [00084] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00084] [#] INCREASE [8] [0] [0] [1] [0] [0] R3 [50] [0] [00084] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00084] [#] ABSORBANCE LIMIT [11000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [AP] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [AP] REPORT NAME [Alkaline Phosphatase] DATA MODE [Aktiv]

TEST NAME [AP] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [AP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [4] [4] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 11: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0401 10 02 10 October 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST AP # TEST NAME AP UNIT U/L

DATA MODE ON BOARD REPORT NAME Alk. Phos

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 258 0 - 258

TECHNICAL LIMIT CLASS I CLASS 2

0 - 800 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST AP

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

700 / 415

ASSAY POINTS 20 - 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 8 0 0

ABS. LIMIT 11000 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 200 0 084 #

R2 0 0 084 #

R3 50 0 084 #

R4 0 0 084 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name ALP 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 415 11 Sample Volume 4 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 13000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 64 55 Expect. Value (H) 306 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 12: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0441 10 02 10 October 2009/6

ALKALINE PHOSPHATASE FS IFCC 37 °C Cat. No 1 0441.........

Notes 1. Please refer to the package insert for Alkaline

Phosphatase FS IFCC for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 8 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems

GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST ALP ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 660 – 405 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 - 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 - 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 13000 – 0 PROZONE LIMIT 0 EXPECTED VALUE # – # PANIC VALUE 0 - 1400 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [AP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [AP] ASSAY/TIME/POINTS [Rate A] [10] [20] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [200] [0] [00158] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00158] [#] INCREASE [8] [0] [0] [1] [0] [0] R3 [50] [0] [00158] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00158] [#] ABSORBANCE LIMIT [13000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [AP] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [AP] REPORT NAME [Alkaline Phosphatase] DATA MODE [Aktiv]

TEST NAME [AP] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [AP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [4] [4] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 13: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0441 10 02 10 October 2009/6

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST AP # TEST NAME AP UNIT U/L

DATA MODE ON BOARD REPORT NAME Alk. Phos

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED

VALUE

CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 1400 -

STD CO

N

PO

S

S.VO

L.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST AP

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

700 / 415

ASSAY POINTS 20 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 8 0 0

ABS. LIMIT 13000] 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 200 0 158 #

R2 0 0 158 #

R3 50 0 [158 #

R4 0 0 158 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name ALP 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 415 11 Sample Volume 4 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 13000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 42 55 Expect. Value (H) 141 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 14: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0501 10 02 10 October 2009/5

-AMYLASE CC FS Cat. No 1 0501.........

Notes 1. Please refer to the package insert for -Amylase

CC FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST AMY ASSAY CODE 5 – 40 – 50 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 200 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 660 – 405 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 10000 – 0 PROZONE LIMIT 0 EXPECTED VALUE 0 – 100 PANIC VALUE 0 - 2000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [AMY] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [AMY] ASSAY/TIME/POINTS [Rate A] [10] [25] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [200] [0] [00570] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00570] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [50] [0] [00570] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00570] [#] ABSORBANCE LIMIT [10000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [AMY] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [AMY] REPORT NAME [Amylase] DATA MODE [Aktiv]

TEST NAME [AMY] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [AMY] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [4] [4] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 15: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0501 10 02 10 October 2009/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST AMY # TEST NAME AMY UNIT U/l

DATA MODE ON BOARD REPORT NAME Amylase

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # -

TECHNICAL LIMIT CLASS I CLASS 2

0 - 2000 -

STD CO

N

PO

S

S.VO

L.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST AMY

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

700 / 415

ASSAY POINTS 25 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 10000 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 200 0 570 #

R2 0 0 570 #

R3 50 0 570 #

R4 0 0 570 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name AMY 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 28 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 415 11 Sample Volume 5 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 10000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 220 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 16: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7102 10 02 10 October 2009/6

APOLIPOPROTEIN A1 FS Cat. No. 1 7102.........

Notes 1. Please refer to the package insert for Apoliporotein

A1 FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST APOA ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 2 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 700 – 570 CALIBR. METHOD 3 – 1 – 4 STD.1 CONC.-POS. 0 - 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * - * STD.4 CONC.-POS. * - * STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 999 DUPLICATE LIMIT 500 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 110 – 190 PANIC VALUE 0.2 - 250 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [APOA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [APOA] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [250] [0] [00168] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00168] [#] INCREASE [4] [0] [0] [1] [0] [0] R3 [50] [0] [00168] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00168] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [APOA] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[4] SPAN POINT [4]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [APOA] REPORT NAME [Apolipoprotein A1] DATA MODE [Aktiv]

TEST NAME [APOA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [APOA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [0] [0]

CONCENTRATION # # # # POSITION # # # #

SAMPLE VOLUME [2] [2] [2] [2] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 17: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7102 10 02 10 October 2009/6

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST APOA # TEST NAME APOA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME APO A1

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

110 - 170 120 - 190

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 250 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) * # 2 0 0 # (3)

(4) * # 2 0 0 # (4)

(5) # 0 0 # (5)

(6) # 0 0 # (6)

TEST APOA

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

700 / 570

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 4 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 168 #

R2 0 0 168 #

R3 50 0 168 #

R4 0 0 168 #

CALIB. TYPE [LOGIT LOG 4P] 4 4 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name APOA 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 570 11 Sample Volume 2 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Logit-Log 4P 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 2500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 110 55 Expect. Value (H) 190 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 18: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7112 10 02 10 May 2009/4

APOLIPOPROTEIN B FS Cat. No 1 7112.........

Notes 1. Please refer to the package insert for Apoliporotein

B FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST APOB ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 700 – 340 CALIBR. METHOD 3 – 1 – 4 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * - * STD.4 CONC.-POS. * - * STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 999 DUPLICATE LIMIT 500 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 75 – 155 PANIC VALUE 0.3 - 250 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [APOB] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [APOB] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00151] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00151] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [50] [0] [00151] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00151] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [APOB] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[4] SPAN POINT [4]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [APOB] REPORT NAME [Apolipoprotein B] DATA MODE [Aktiv]

TEST NAME [APOB] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [APOB] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [0] [0]

CONCENTRATION # # # # POSITION # # # #

SAMPLE VOLUME [3] [3] [3] [3] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 19: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7112 10 02 10 May 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST APOB # TEST NAME APOB UNIT mg/dl

DATA MODE ON BOARD REPORT NAME APO B

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

80 - 155 75 - 150

TECHNICAL LIMIT CLASS I CLASS 2

0.3 - 250 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) * # 3 0 0 # (3)

(4) * # 3 0 0 # (4)

(5) # 0 0 # (5)

(6) # 0 0 # (6)

TEST APOB

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

700 / 340

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 151 #

R2 0 0 151 #

R3 50 0 151 #

R4 0 0 151 #

CALIB. TYPE [LOGIT LOG 4P] 4 4 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name APOB 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 340 11 Sample Volume 3 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Logit-Log 4P 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 2500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 45 55 Expect. Value (H) 125 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 20: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2601 10 02 10 March 2010/4

ASAT(GOT) FS (IFCC mod.) Cat. No 1 2601.........

Notes 1. Please refer to the package insert for ASAT(GOT)

FS (IFCC mod) for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST GOT ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 - 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 6000 – 1 PROZONE LIMIT 0 – 0 EXPECTED VALUE # – # PANIC VALUE 0 - 700 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [ASAT] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [ASAT] ASSAY/TIME/POINTS [Rate A] [10] [20] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [200] [0] [00105] [#]

DECREASE [7] [0] [0] [1] [0] [0] R2 [0] [0] [00105] [#] INCREASE [25] [0] [0] [1] [0] [0] R3 [50] [0] [00105] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00105] [#] ABSORBANCE LIMIT [6000] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [ASAT] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [ASAT] REPORT NAME [ASAT] DATA MODE [Aktiv]

TEST NAME [ASAT] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [ASAT] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [15] [15] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 21: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2601 10 02 10 March 2010/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST ASAT # TEST NAME ASAT UNIT U/L

DATA MODE ON BOARD REPORT NAME ASAT

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 700 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST ASAT

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 20 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 7 0 0

S.VOL (INCREASE) 25 0 0

ABS. LIMIT 6000] 1:DECREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 105 #

R2 0 0 105 #

R3 50 0 105 #

R4 0 0 105 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code ATTENTION This kit is for use with the original application. If you wish to use this included DiaSys application please contact your supplier for additional bar code labels.

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name GOT 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 15 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 6000 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 37 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 22: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7012 10 02 10 July 2009/5

ANTISTREPTOLYSIN O FS Cat. No 1 7012.........

Notes 1. Please refer to the package insert for Antistreptolysin

O FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at

least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST ASO ASSAY CODE 2 – 32 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 0 – 600 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 - 0 PROZONE LIMIT 32000 - 1 EXPECTED VALUE 0 - 200 PANIC VALUE 7 - 800 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [ASO] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [ASO] ASSAY/TIME/POINTS [2 point end] [10] [19] [30] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [600]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00037] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00037] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [50] [0] [00037] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00037] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [ASO] CALIBRATION TYPE [SPLINE] [ ]

CALIBRATION POINTS

[6] SPAN POINT [3]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [ASO] REPORT NAME [Antistreptolysin O] DATA MODE [Aktiv]

TEST NAME [ASO] [] CONTROL INTERVAL UNIT [IU/ml] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [ASO] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [3] [3] [3] [3] [3] [3]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 23: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7012 10 02 10 July 2009/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST ASO # TEST NAME ASO UNIT IU/ml

DATA MODE ON BOARD REPORT NAME ASO

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 200 0 -

TECHNICAL LIMIT CLASS I CLASS 2

7 - 800 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) * # 3 0 0 # (3)

(4) * # 3 0 0 # (4)

(5) * # 3 0 0 # (5)

(6) * # 3 0 0 # (6)

TEST ASO

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 600

ASSAY POINTS 19- 30 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 037 #

R2 0 0 037 #

R3 50 0 037 #

R4 0 0 037 #

CALIB. TYPE [SPLINE] 6 3 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name ASO 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 3 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Logit-Log 4P 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 2500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 200 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 24: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0950 10 02 10 August 2008/5

BICARBONATE FS Cat. No 1 0950.........

Notes 1. Please refer to the package insert for Bicarbonate

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

3 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CO2 ASSAY CODE 2 – 6 – 42 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 300 – 100 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 505 – 405 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 200 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 1 PROZONE LIMIT -32000 – 0 EXPECTED VALUE 22 – 29 PANIC VALUE 0 – 50 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CO2] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CO2] ASSAY/TIME/POINTS [2 point end] [10] [2] [25] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [505] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [300] [0] [00156] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00156] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00156] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00156] [#] ABSORBANCE LIMIT [0] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [CO2] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CO2] REPORT NAME [Bicarbonate] DATA MODE [Aktiv]

TEST NAME [CO2] [] CONTROL INTERVAL UNIT [mmol/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [AP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 25: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0950 10 02 10 August 2008/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CO2 # TEST NAME CO2 UNIT mmol/l

DATA MODE ON BOARD REPORT NAME CO2

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

22 - 29 22 - 29

TECHNICAL LIMIT CLASS I CLASS 2

0 - 50 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CO2

ASSAY CODE 2 Point end 10 WAVELGTH (SUB / MAIN)

505 / 415

ASSAY POINTS 2 - 25 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 DECREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 300 0 156 #

R2 0 0 156 #

R3 0 0 156 #

R4 0 0 156 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CO2 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 5 6 Assay Point 2 30 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 505 10 Wave Leng. (MAIN) 415 11 Sample Volume 3.0 12 R1 VOLUME 300 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 200 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 22 55 Expect. Value (H) 29 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 26: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0821 10 02 10 October 2008/4

BILIRUBIN AUTO DIRECT FS Cat. No 1 0821.........

Notes 1. Please refer to the package insert for Bilirubin Auto

Direct FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST BIDI ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 20 R 1 VOLUME (µl) 240 – 100 – 0 R 2 VOLUME (µl) 60 – 20 – 0 WAVELENGTH (nm) 660 – 546 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0.0 – 0.3 PANIC VALUE - 10.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [DBIL] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [DBIL] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [660] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [10] [0] [0] [1] [0] [0] R1 [240] [0] [00294] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [00294] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [60] [0] [00294] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00294] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [DBIL] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [DBIL] REPORT NAME [Direct Bilirubin] DATA MODE [Aktiv]

TEST NAME [DBIL] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [DBIL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [10] [10] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 27: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0821 10 02 10 October 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST DBIL # TEST NAME DBIL UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Bilirubin Direct

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 0.2 0 - 0.2

TECHNICAL LIMIT CLASS I CLASS 2

0 - 10.0 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 10 0 0 # (1)

(2) * # 10 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST DBIL

ASSAY CODE 2 point end 10 WAVELGH (SUB / MAIN)

660 / 546

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 10 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 294 #

R2 0 0 294 #

R3 60 0 294 #

R4 0 0 294 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name DBIL 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 546 11 Sample Volume 28 12 R1 VOLUME 280 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0.0 55 Expect. Value (H) 0.25 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 28: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0811 10 02 10 February 2007/3

BILIRUBIN AUTO TOTAL FS Cat. No 1 0811.........

Notes 1. Please refer to the package insert for Bilirubin Auto

Total FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST BILI ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 6 R 1 VOLUME (µl) 240 – 100 – 0 R 2 VOLUME (µl) 60 – 20 – 0 WAVELENGTH (nm) 660 – 546 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0.1 – 1.2 PANIC VALUE 0.1 – 30.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [TBIL] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [TBIL] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [660] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [6] [0] [0] [1] [0] [0] R1 [240] [0] [00269] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00269] [#] INCREASE [12] [0] [0] [1] [0] [0] R3 [60] [0] [00269] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00269] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [TBIL] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [TBIL] REPORT NAME [Total Bilirubin] DATA MODE [Aktiv]

TEST NAME [TBIL] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [TBIL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [6] [6] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 29: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0811 10 02 10 February 2007/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST TBIL # TEST NAME TBIL UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Bilirubin Total

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.1 - 1.2 0.1 - 1.2

TECHNICAL LIMIT CLASS I CLASS 2

0 - 30.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 6 0 0 # (1)

(2) * # 6 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST TBIL

ASSAY CODE 2 point end 10 WAVELGH (SUB / MAIN)

660 / 546

ASSAY POINTS 15 - 31 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 6 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 12 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 269 #

R2 0 0 269 #

R3 60 0 269 #

R4 0 0 269 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name TBIL 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 546 11 Sample Volume 7 12 R1 VOLUME 280 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0.0 55 Expect. Value (H) 1.0 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 30: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1802 10 02 10 August 2008/3

COMPLEMENT C3c FS Cat. No 1 1802.........

Notes 1. Please refer to the package insert for Complement

C3c FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST C3c ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) – 340 CALIBR. METHOD 3 – 4 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 90 – 180 PANIC VALUE 1 – 500 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [C3] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [C3] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [250] [0] [00036] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00036] [#] INCREASE [8] [0] [0] [1] [0] [0] R3 [50] [0] [00036] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00036] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [C3] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[6] SPAN POINT [4]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [5000] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [C3] REPORT NAME [Complement C3c] DATA MODE [Aktiv]

TEST NAME [C3] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [C3] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [4] [4] [4] [4] [4] [4]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 31: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1802 10 02 10 August 2008/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST C3C # TEST NAME C3C UNIT mg/dl

DATA MODE ON BOARD REPORT NAME C3C

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

90 - 180 # - #

TECHNICAL LIMIT CLASS I CLASS 2

1 - 500 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) * # 4 0 0 # (3)

(4) * # 4 0 0 # (4)

(5) * # 4 0 0 # (5)

(6) * # 4 0 0 # (6)

TEST C3C

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 340

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 036 #

R2 0 0 036 #

R3 50 0 036 #

R4 0 0 036 #

CALIB. TYPE SPLINE 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name C3 2 Assay Code (Mthd) 2 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 340 11 Sample Volume 4 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 32: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1812 10 02 10 August 2008/3

COMPLEMENT C4 FS Cat. No 1 1812.........

Notes 1. Please refer to the package insert for Complement

C4 FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST C4 ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 7 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) – 340 CALIBR. METHOD 3 – 4 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 10 – 40 PANIC VALUE 0.6 - 90 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [C4] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [C4] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [7] [0] [0] [1] [0] [0] R1 [250] [0] [00032] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00032] [#] INCREASE [14] [0] [0] [1] [0] [0] R3 [50] [0] [00032] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00032] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [C4] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[6] SPAN POINT [4]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [5000] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [C4] REPORT NAME [Complement C4] DATA MODE [Aktiv]

TEST NAME [C4] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [C3] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [7] [7] [7] [7] [7] [7]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 33: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1812 10 02 10 August 2008/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST C4 # TEST NAME C4 UNIT mg/dl

DATA MODE ON BOARD REPORT NAME C4

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

10 - 40 # - #

TECHNICAL LIMIT CLASS I CLASS 2

0.6 - 90 -

STD CON POS S.VOL. PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 7 0 0 # (1)

(2) * # 7 0 0 # (2)

(3) * # 7 0 0 # (3)

(4) * # 7 0 0 # (4)

(5) * # 7 0 0 # (5)

(6) * # 7 0 0 # (6)

TEST C4

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 340

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 7 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 032 #

R2 0 0 032 #

R3 50 0 032 #

R4 0 0 032 #

CALIB. TYPE SPLINE] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name C4 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 340 11 Sample Volume 7 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 34: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1130 10 02 10 September 2009/3

CALCIUM AS FS Cat. No 1 1130.........

Notes 1. Please refer to the package insert for Calcium AS

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CALC ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 100 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 700 – 660 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 100 – 1 EXPECTED VALUE 8.6 – 10.3 PANIC VALUE 0.04 – 25.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CA] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [660]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00381] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00381] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00381] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00381] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CA] REPORT NAME [Calcium] DATA MODE [Aktiv]

TEST NAME [CA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 35: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1130 10 02 10 September 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CA # TEST NAME CA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Calcium

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

8.6 - 10.3 8.6 - 10.3

TECHNICAL LIMIT CLASS I CLASS 2

0 - 25.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CA

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 660

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 381 #

R2 0 0 381 #

R3 0 0 381 #

R4 0 0 381 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CAL 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 35 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 660 11 Sample Volume 3 12 R1 VOLUME 300 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 600 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 8.6 55 Expect. Value (H) 10.3 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 36: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1121 10 02 10 September 2006/3

CALCIUM CPC FS Cat. No 1 1121.........

Notes 1. Please refer to the package insert for Calcium CPC

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 7 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CALC ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 240 – 100 – 0 R 2 VOLUME (µl) 60 – 20 – 0 WAVELENGTH (nm) 660 – 570 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 8.6 – 10.3 PANIC VALUE 0.2 – 20.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CA] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [660] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [5] [0] [0] [1] [0] [0] R1 [240] [0] [00380] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00380] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [60] [0] [00380] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00380] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CA] REPORT NAME [Calcium] DATA MODE [Aktiv]

TEST NAME [CA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [5] [5] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 37: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1121 10 02 10 September 2006/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CA # TEST NAME CA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Calcium

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

8.6 - 10.3 8.6 - 10.3

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 20.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 5 0 0 # (1)

(2) * # 5 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CA

ASSAY CODE 2 point 10 WAVELGTH (SUB / MAIN)

660 / 570

ASSAY POINTS 15 - 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 5 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 380 #

R2 0 0 380 #

R3 60 0 380 #

R4 0 0 380 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CA 2 Assay Code (Mthd) 2 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 570 11 Sample Volume 5 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 8.6 55 Expect. Value (H) 10.4 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 38: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1181 10 02 10 March 2010/5

CALCIUM P FS Cat. No 1 1181.........

Notes 1. Please refer to the package insert for Calcium P FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 8 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CALC ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 240 – 100 – 0 R 2 VOLUME (µl) 60 – 20 – 0 WAVELENGTH (nm) 0 – 660 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 500 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 8.6 – 10.3 PANIC VALUE 0.1 – 25.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CA] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [660]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [240] [0] [00264] [#]

DECREASE [3] [0] [0] [1] [0] [0] R2 [0] [0] [00264] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [60] [0] [00264] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00264] [#] ABSORBANCE LIMIT [0] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CA] REPORT NAME [Calcium] DATA MODE [Aktiv]

TEST NAME [CA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 39: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1181 10 02 10 March 2010/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CA # TEST NAME CA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Calcium

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

8.6 - 10.3 8.6 - 10.3

TECHNICAL LIMIT CLASS I CLASS 2

0.1 - 25.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CA

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

800 / 660

ASSAY POINTS 15 - 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 1:DECREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 264 #

R2 0 0 264 #

R3 60 0 264 #

R4 0 0 264 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CA 2 Assay Code (Mthd) 2 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 0 10 Wave Leng. (MAIN) 660 11 Sample Volume 3 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 8.6 55 Expect. Value (H) 10.4 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 40: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1401 10 02 10 October 2008/5

CHOLINESTERASE FS Cat. No 1 1401.........

Notes 1. Please refer to the package insert for

Cholinesterase FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 8 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CHE ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 0 – 405 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 1 PROZONE LIMIT 0 EXPECTED VALUE 3930 – 11500 PANIC VALUE 0 - 20000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CHE] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CHE] ASSAY/TIME/POINTS [Rate A] [10] [21] [30] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [200] [0] [00434] [#]

DECREASE [3] [0] [0] [1] [0] [0] R2 [0] [0] [00434] [#] INCREASE [5] [0] [0] [1] [0] [0] R3 [50] [0] [00434] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00434] [#] ABSORBANCE LIMIT [0] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CHE] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CHE] REPORT NAME [Cholinesterase] DATA MODE [Aktiv]

TEST NAME [CHE] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [AP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [4] [4] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 41: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1401 10 02 10 October 2008/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CHE # TEST NAME CHE UNIT U/l

DATA MODE ON BOARD REPORT NAME CHE

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 20000 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CHE

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

/ 415

ASSAY POINTS 21 - 30- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 5 0 0

ABS. LIMIT 0 1:DECREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 200 0 434 #

R2 0 0 434 #

R3 50 0 434 #

R4 0 0 434 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CHE 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 22 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 0 10 Wave Leng. (MAIN) 415 11 Sample Volume 4 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Decrease 51 Prz. Limit 32000 52 Prz. Limit (U/D) Higher 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 42: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1200 10 02 10 September 2009/3

CHLORIDE FS Cat. No 1 1200.........

Notes 1. Please refer to the package insert for Chloride FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 7 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CL ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 600 – 480 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 –1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 EXPECTED VALUE 95 – 105 PANIC VALUE 1 - 130 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CL] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CL] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [480]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00379] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00379] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00379] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00379] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CL] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CL] REPORT NAME [Chloride] DATA MODE [Aktiv]

TEST NAME [CL] [] CONTROL INTERVAL UNIT [mmol/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 43: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1200 10 02 10 September 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CL # TEST NAME CL UNIT mmol/l

DATA MODE ON BOARD REPORT NAME Chloride

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

95 - 105 95 - 105

TECHNICAL LIMIT CLASS I CLASS 2

1 - 130 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CL

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 480

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 379 #

R2 0 0 379 #

R3 0 0 379 #

R4 0 0 379 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CL 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 35 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 600 10 Wave Leng. (MAIN) 480 11 Sample Volume 3 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 98 55 Expect. Value (H) 107 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 44: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1300 10 02 10 September 2005/2

CHOLESTEROL FS Cat. No 1 1300.........

Notes 1. Please refer to the package insert for Cholesterol

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CHOL ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 100 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 700 – 505 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 - 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0 – 200 PANIC VALUE 3 - 750 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CHOL] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CHOL] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00263] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00263] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00263] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00263] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CHOL] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CHOL] REPORT NAME [Cholesterol] DATA MODE [Aktiv]

TEST NAME [CHOL] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CHOL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 45: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1300 10 02 10 September 2005/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CHOL # TEST NAME CHOL UNIT mg/dl

DATA MODE ON BOARD REPORT NAME CHOLESTEROL

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 200 0 - 200

TECHNICAL LIMIT CLASS I CLASS 2

3 - 750 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CHOL

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 505

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 250 0 263 #

R2 0 0 263 #

R3 0 0 263 #

R4 0 0 263 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CHOL 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 35 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 505 11 Sample Volume 3.0 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 2000 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 220 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 46: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1641 10 02 10 May 2010/2

CK-MB FS Cat. No 1 1641.........

Notes 1. Please refer to the package insert for CK-MB FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least 6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CKMB ASSAY CODE 5 – 40 – 50 SAMPLE VOLUME (µl) 10 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT (INC/DEC) 9000 – 0 PROZONE LIMIT 0 EXPECTED VALUE 0 – 24 PANIC VALUE 0 - 2000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CK-MB] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CK-MB] ASSAY/TIME/POINTS [Rate A] [10] [25] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [10] [0] [0] [1] [0] [0] R1 [200] [0] [00060] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [00060] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [50] [0] [00060] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00060] [#] ABSORBANCE LIMIT [9000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [higher] Calibration

TEST [CK-MB] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CK-MB] REPORT NAME [CK-MB] DATA MODE [Aktiv]

TEST NAME [CK-MB] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CK-MB] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [10] [10] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 47: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1641 10 02 10 May 2010/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CKMB # TEST NAME CKMB UNIT U/L

DATA MODE ON BOARD REPORT NAME CK-MB

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 2000 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 10 0 0 # (1)

(2) * # 10 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CKMB

ASSAY CODE RATE A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 25 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 10 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 9000 2:INCREASE

PROZONE LIMIT 0 HIGHER

REAGENT R1 200 0 060 #

R2 0 0 060 #

R3 50 0 060 #

R4 0 0 060 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CKMB 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 28 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 10 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 9000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 24 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 48: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1651 10 02 10 October 2009/6

CK-MB FS Cat. No 1 1651.........

Notes 1. Please refer to the package insert for CK-MB FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CKMB ASSAY CODE 5 – 40 – 50 SAMPLE VOLUME (µl) 10 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 9000 – 0 PROZONE LIMIT 0 EXPECTED VALUE 0 – 24 PANIC VALUE 0 - 2000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CK-MB] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CK-MB] ASSAY/TIME/POINTS [Rate A] [10] [25] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [10] [0] [0] [1] [0] [0] R1 [200] [0] [00060] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [00060] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [50] [0] [00060] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00060] [#] ABSORBANCE LIMIT [9000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [higher] Calibration

TEST [CK-MB] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CK-MB] REPORT NAME [CK-MB] DATA MODE [Aktiv]

TEST NAME [CK-MB] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CK-MB] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [10] [10] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 49: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1651 10 02 10 October 2009/6

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CKMB # TEST NAME CKMB UNIT U/L

DATA MODE ON BOARD REPORT NAME CK-MB

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 2000 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 10 0 0 # (1)

(2) * # 10 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CKMB

ASSAY CODE RATE A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 25 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 10 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 9000 2:INCREASE

PROZONE LIMIT 0 HIGHER

REAGENT R1 200 0 060 #

R2 0 0 060 #

R3 50 0 060 #

R4 0 0 060 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CKMB 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 28 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 10 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 9000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 24 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 50: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1601 10 02 10 October 2009/6

CK-NAC FS Cat. No 1 1601.........

Notes 1. Please refer to the package insert for CK-NAC FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CK ASSAY CODE 5 – 35 – 50 SAMPLE VOLUME (µl) 7 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 9000 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE # – # PANIC VALUE 0 - 1100 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CK] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CK] ASSAY/TIME/POINTS [Rate A] [10] [22] [30] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [7] [0] [0] [1] [0] [0] R1 [200] [0] [00057] [#]

DECREASE [3] [0] [0] [1] [0] [0] R2 [0] [0] [00057] [#] INCREASE [14] [0] [0] [1] [0] [0] R3 [50] [0] [00057] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00057] [#] ABSORBANCE LIMIT [9000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [higher] Calibration

TEST [CK] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CK] REPORT NAME [CK] DATA MODE [Aktiv]

TEST NAME [CK] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CK] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [7] [7] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 51: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1601 10 02 10 October 2009/6

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CK # TEST NAME CK UNIT U/L

DATA MODE ON BOARD REPORT NAME CK-NAC

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 1100 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 7 0 0 # (1)

(2) * # 7 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CK

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 22 - 30- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 7 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 14 0 0

ABS. LIMIT 9000 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 200 0 057 #

R2 0 0 057 #

R3 50 0 057 #

R4 0 0 057 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CK 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 24 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 7 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 9000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 24 55 Expect. Value (H) 190 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 52: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1711 10 02 10 March 2008/4

CREATININE FS Cat. No 1 1711.........

Notes 1. Please refer to the package insert for Creatinine FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 4 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CREA ASSAY CODE 5 – 30 – 37 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 570 – 505 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. *– * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 4500 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0.6 – 1.3 PANIC VALUE 0.2 – 15.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CREA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CREA] ASSAY/TIME/POINTS [2 point rate] [10] [19] [24] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [570] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [200] [0] [00400] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [00400] [#] INCREASE [30] [0] [0] [1] [0] [0] R3 [50] [0] [00400] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00400] [#] ABSORBANCE LIMIT [4500] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [CREA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CREA] REPORT NAME [Creatinine] DATA MODE [Aktiv]

TEST NAME [CREA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CREA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [15] [15] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 53: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1711 10 02 10 March 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CREA # TEST NAME CREA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME CREATININE

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.9 - 1.3 0.6 - 1.1

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 15.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CREA

ASSAY CODE 2 POINT

RATE

10 WAVELGTH (SUB / MAIN)

570 / 505

ASSAY POINTS 19 - 24- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 30 0 0

ABS. LIMIT 4500 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 400 #

R2 0 0 400 #

R3 50 0 400 #

R4 0 0 400 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CREA 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 26 7 Assay Point 3 10 8 Assay Point 4 15 9 Wave Leng. (SUB) 570 10 Wave Leng. (MAIN) 505 11 Sample Volume 15 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 4500 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0.5 55 Expect. Value (H) 1.1 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 54: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1711 10 02 10 March 2008/4

CREATININE FS Compensated Method To use the Creatinine FS Compensated Methodon the Hitachi 912, 911 and 902 the Instrument factor ’b’ must be changed as follows; When the units are mg/dL

Instrument Factor (Y=aX+b) b= - 0.3 When the units are µmol/L

Instrument Factor (Y=aX+b) b= - 27 To use the Creatinine Compensated Method on the Hitachi 717 go to Parameter Job, Chapter 7, Compensated Test and input the following formula; When the units are mg/dL [ (Crea test Code) ] [ - ] [ 0.3 ] [ ; ] When the units are µmol/L [ (Crea test Code) ] [ - ] [ 27 ] [ ; ] Note: Ensure that when using the Creatinine Compensated Method the relevant assigned Calibrator value is selected

HITACHI 717

Chemistry settings PROGRAM 2: CHEMISTRY PARAMETERS TEST CREA ASSAY CODE 5 – 30 – 37 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 570 – 505 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. *– * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 4500 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0.6 – 1.3 PANIC VALUE 0.2 – 15.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CREA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CREA] ASSAY/TIME/POINTS [2 point rate] [10] [19] [24] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [570] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [200] [0] [00400] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [00400] [#] INCREASE [30] [0] [0] [1] [0] [0] R3 [50] [0] [00400] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00400] [#] ABSORBANCE LIMIT [4500] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [CREA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CREA] REPORT NAME [Creatinine] DATA MODE [Aktiv]

TEST NAME [CREA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [-0.30]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CREA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [15] [15] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 55: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1711 10 02 10 March 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CREA # TEST NAME CREA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME CREATININE

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b -0.3

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.9 - 1.3 0.6 - 1.1

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 15.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CREA

ASSAY CODE 2 POINT

RATE

10 WAVELGTH (SUB / MAIN)

570 / 505

ASSAY POINTS 19 - 24- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 30 0 0

ABS. LIMIT 4500 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 400 #

R2 0 0 400 #

R3 50 0 400 #

R4 0 0 400 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CREA 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 26 7 Assay Point 3 10 8 Assay Point 4 15 9 Wave Leng. (SUB) 570 10 Wave Leng. (MAIN) 505 11 Sample Volume 15 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 4500 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0.5 55 Expect. Value (H) 1.1 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) -0.3 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 56: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1759 10 02 10 September 2005/2

CREATININE PAP FS Cat. No 1 1759.........

Notes 1. Please refer to the package insert for Creatinine

PAP FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CREA ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 6 R 1 VOLUME (µl) 250 – 100 – NO R 2 VOLUME (µl) 125 – 100 – NO WAVELENGTH (nm) 700 – 546 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. *– * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0.6 – 1.3 PANIC VALUE # – # INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [Crea] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [Crea] ASSAY/TIME/POINTS [2 Point End] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [6] [0] [0] [1] [0] [0] R1 [250] [0] [00063] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00063] [#] INCREASE [18] [0] [0] [1] [0] [0] R3 [125] [0] [00063] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00063] [#] ABSORBANCE LIMIZ [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [Crea] KALIBRATIONS TYP [Linear] [ ]

CALIBRATION POINTS [2] SPAN POINT [2] WEIGHT [0]

AUTOMATIC CALIBRATION AUTO TIME OUT AUTO CHANGE

BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS. LIMIT [-32000] - [32000]

Range TEST [Crea] REPORT NAME [Creatinin] DATA MODE [Aktiv]

TEST NAME [Crea] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [Crea] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [6] [6] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 57: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1759 10 02 10 September 2005/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CREA # TEST NAME CREA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME CREATININE

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.9 - 1.3 0.6 - 1.1

TECHNICAL LIMIT CLASS I CLASS 2

# - # -

STD CON POS S.VOL. PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 6 0 0 # (1)

(2) * # 6 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST CREA

ASSAY CODE 2 POINT END 10 WAVELGTH (SUB / MAIN)

700 / 546

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 6 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 18 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 250 0 063 #

R2 0 0 063 #

R3 125 0 063 #

R4 0 0 063 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CRE 2 Assay Code (Mthd) 2 point end 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 546 11 Sample Volume 6 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 17 R2 Bottle Size Small 18 R3 VOLUME 125 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0.9 55 Expect. Value (H) 1.3 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 58: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7045 10 42 10 August 2008/5

CRP U-hs High sensitive (hs) application

Cat. No 1 7045.........

Notes 1. Please refer to the package insert for CRP U-hs for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

8 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CRPS ASSAY CODE 2 – 31 – 50 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 150 – 20 – NO R 2 VOLUME (µl) 150 – 20 – NO WAVELENGTH (nm) 800 – 505 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 500 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – 5 PANIC VALUE 0.05 – 20 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CRP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CRP] ASSAY/TIME/POINTS [2 point end] [10] [19] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [150] [0] [00218] [#]

DECREASE [10] [0] [0] [1] [0] [0] R2 [0] [0] [00218] [#] INCREASE [30] [0] [0] [1] [0] [0] R3 [150] [0] [00218] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00218] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CRP] CALIBRATION TYPE [SPLINE] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CRP] REPORT NAME [CRP] DATA MODE [Aktiv]

TEST NAME [CRP] [] CONTROL INTERVAL UNIT [mg/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CRP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [15] [15] [15] [15] [15] [15]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 59: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7045 10 42 10 August 2008/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CRP # TEST NAME CRP UNIT mg/l

DATA MODE ON BOARD REPORT NAME CRP US

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.0 - 5 0.0 - 5

TECHNICAL LIMIT CLASS I CLASS 2

0. 05 - 20 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.00 # 15 0 0 # (1)

(2) # # 15 0 0 # (2)

(3) # # 15 0 0 # (3)

(4) # # 15 0 0 # (4)

(5) # # 15 0 0 # (5)

(6) # # 15 0 0 (6)

TEST CRP

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

800 / 505

ASSAY POINTS 19- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 10 0 0

S.VOL (INCREASE) 30 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 150 0 218 #

R2 0 0 218 #

R3 150 0 218 #

R4 0 0 218 #

CALIB. TYPE [SPLINE] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 5000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CRP 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 32 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 800 10 Wave Leng. (MAIN) 505 11 Sample Volume 17 12 R1 VOLUME 170 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 170 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 0.5 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 60: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7045 10 32 10 August 2008/5

CRP U-hs Universal (U) application Cat. No 1 7045.........

Notes 1. Please refer to the package insert for CRP U-hs for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

8 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CRPS ASSAY CODE 2 – 31 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 150 – 20 – NO R 2 VOLUME (µl) 150 – 20 – NO WAVELENGTH (nm) 800 – 505 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 500 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – 5 PANIC VALUE 0.3 – 350 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CRP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CRP] ASSAY/TIME/POINTS [2 point end] [10] [19] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [150] [0] [00218] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00218] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [150] [0] [00218] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00218] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CRP] CALIBRATION TYPE [SPLINE] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CRP] REPORT NAME [CRP] DATA MODE [Aktiv]

TEST NAME [CRP] [] CONTROL INTERVAL UNIT [mg/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CRP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [3] [3] [3] [3] [3] [3]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 61: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7045 10 32 10 August 2008/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CRP # TEST NAME CRP UNIT mg/l

DATA MODE ON BOARD REPORT NAME CRP DY

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.0 - 5 0.0 - 5

TECHNICAL LIMIT CLASS I CLASS 2

0. 3 - 350 -

STD CO

N

PO

S

S.VO

L.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.00 # 3 0 0 # (1)

(2) # # 3 0 0 # (2)

(3) # # 3 0 0 # (3)

(4) # # 3 0 0 # (4)

(5) # # 3 0 0 # (5)

(6) # # 3 0 0 (6)

TEST CRP

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

800 / 505

ASSAY POINTS 19- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 150 0 218 #

R2 0 0 218 #

R3 150 0 218 #

R4 0 0 218 #

CALIB. TYPE [SPLINE] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 5000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CRP 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 32 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 800 10 Wave Leng. (MAIN) 505 11 Sample Volume 3 12 R1 VOLUME 170 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 170 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 0.5 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 62: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7002 10 02 10 August 2008/4

CRP FS Cat. No 1 7002.........

Notes 1. Please refer to the package insert for CRP FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST CRP ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 700 – 340 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 500 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – 0.5 PANIC VALUE 0.2 – 25.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [CRP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [CRP] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [250] [0] [00704] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [00704] [#] INCREASE [30] [0] [0] [1] [0] [0] R3 [50] [0] [00704] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00704] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [CRP] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [CRP] REPORT NAME [CRP] DATA MODE [Aktiv]

TEST NAME [CRP] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [CRP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [15] [15] [15] [15] [15] [15]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position n

Page 63: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7002 10 02 10 August 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST CRP # TEST NAME CRP UNIT mg/dl

DATA MODE ON BOARD REPORT NAME CRP

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0.0 - 0.5 0.0 - 0.5

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 25.0 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.00 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) * # 15 0 0 # (3)

(4) * # 15 0 0 # (4)

(5) * # 15 0 0 # (5)

(6) * # 15 0 0 # (6)

TEST CRP

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

700 / 340

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 30 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 704 #

R2 0 0 704 #

R3 50 0 704 #

R4 0 0 704 #

CALIB. TYPE [LOGIT LOG 4P] 6 6 0 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name CRP 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 340 11 Sample Volume 15 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Spline 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 0.8 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 64: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7059 10 02 10 August 2008/3

FERRITIN FS Cat. No 1 7059.........

Notes 1. Please refer to the package insert for Ferritin FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST FERR ASSAY CODE 2 – 27 – 42 SAMPLE VOLUME (µl) 8 R 1 VOLUME (µl) 160 – 20 – NO R 2 VOLUME (µl) 80 – 20 – NO WAVELENGTH (nm) 800– 570 CALIBR. METHOD 3 – 4 – 5 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. 0 – 0 SD LIMIT 999 DUPLICATE LIMIT 32000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 32000 – 0 PROZONE LIMIT -32000 – 1 EXPECTED VALUE 30 – 300 PANIC VALUE 5 - 1000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [FER] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [FER] ASSAY/TIME/POINTS [2 point end] [10] [18] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [8] [0] [0] [1] [0] [0] R1 [160] [0] [00164] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00164] [#] INCREASE [16] [0] [0] [1] [0] [0] R3 [80] [0] [00164] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00164] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [-32000] [0] [lower] Calibration

TEST [FER] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[5] SPAN POINT [5]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [FER] REPORT NAME [Ferritin] DATA MODE [Aktiv]

TEST NAME [FER] [] CONTROL INTERVAL UNIT [µg/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [FER] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [0]

CONCENTRATION # # # # # POSITION # # # # #

SAMPLE VOLUME [8] [8] [8] [8] [8] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 65: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7059 10 02 10 August 2008/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST FER # TEST NAME FER UNIT µg/l

DATA MODE ON BOARD REPORT NAME FERRITIN

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

30 - 300 10 - 160

TECHNICAL LIMIT CLASS I CLASS 2

5 - 1000 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0 # 8 0 0 # (1)

(2) * # 8 0 0 # (2)

(3) * # 8 0 0 # (3)

(4) * # 8 0 0 # (4)

(5) * # 8 0 0 # (5)

(6) (6)

TEST FER

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

800 / 570

ASSAY POINTS 18- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 8 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 16 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT -32000 LOWER

REAGENT R1 160 0 164 #

R2 0 0 164 #

R3 80 0 164 #

R4 0 0 164 #

CALIB. TYPE [SPLINE] 5 5 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 32000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name Ferr 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 20 6 Assay Point 2 32 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 800 10 Wave Leng. (MAIN) 570 11 Sample Volume 10 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 100 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 34 Calib. Pos. 6 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit -32000 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 10 55 Expect. Value (H) 300 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 66: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2801 10 02 10 October 2009/3

GAMMA-GT FS (Szasz mod.) Cat. No 1 2801.........

Notes 1. Please refer to the package insert for Gamma-GT FS

(Szasz mod.) for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at

least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST GGT ASSAY CODE 5 – 30 – 50 SAMPLE VOLUME (µl) 8 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 660 – 405 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 8000 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE # – # PANIC VALUE 0 - 1200 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [GGT] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [GGT] ASSAY/TIME/POINTS [Rate A] [10] [20] [30] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [8] [0] [0] [1] [0] [0] R1 [200] [0] [00479] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00479] [#] INCREASE [16] [0] [0] [1] [0] [0] R3 [50] [0] [00479] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00479] [#] ABSORBANCE LIMIT [8000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [GGT] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [GGT] REPORT NAME [Gamma GT] DATA MODE [Aktiv]

TEST NAME [GGT] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [GGT] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [8] [8] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 67: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2801 10 02 10 October 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST GGT # TEST NAME GGT UNIT U/l

DATA MODE ON BOARD REPORT NAME Gamma-GT

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 1200 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 8 0 0 # (1)

(2) * # 8 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST GGT

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

700 / 415

ASSAY POINTS 20- 30 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 8 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 16 0 0

ABS. LIMIT 8000 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 200 0 479 #

R2 0 0 479 #

R3 50 0 479 #

R4 0 0 479 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name GGT 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 415 11 Sample Volume 8 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 8000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 7 55 Expect. Value (H) 50 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 68: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2411 10 02 10 March 2009/1

GLDH FS DGKC Cat. No 1 2411.........

Notes 1. Please refer to the package insert for GLDH FS

DGKC for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST GLDH ASSAY CODE 5 – 28 – 44 SAMPLE VOLUME (µl) 20 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 8000 – 1 PROZONE LIMIT 0 EXPECTED VALUE 0– 7 PANIC VALUE 0 - 120 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [GLDH] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [GLDH] ASSAY/TIME/POINTS [Rate A] [10] [18] [27] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [30] [0] [0] [1] [0] [0] R1 [200] [0] [#] [#]

DECREASE [5] [0] [0] [1] [0] [0] R2 [0] [0] [#] [#] INCREASE [40] [0] [0] [1] [0] [0] R3 [50] [0] [#] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [#] [#] ABSORBANCE LIMIT [8000] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [GLDH] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [GLDH] REPORT NAME [GLDH] DATA MODE [Aktiv]

TEST NAME [GLDH] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [GLDH] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [30] [30] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 69: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2411 10 02 10 March 2009/1

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST GLDH # TEST NAME GLDH UNIT U/L

DATA MODE ON BOARD REPORT NAME GLDH

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 7 0 - 5

TECHNICAL LIMIT CLASS I CLASS 2

0 - 120 -

STD CON POS S.VOL. PRE. DIL. VOL CODE LOT QUALITATIVE NO

(1) 0.0 # 30 0 0 # (1)

(2) * # 30 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST GLDH

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 18 - 27- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 30 0 0

S. VOL (DECREASE) 5 0 0

S.VOL (INCREASE) 40 0 0

ABS. LIMIT 8000] 1:DECREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 # #

R2 0 0 # #

R3 50 0 # #

R4 0 0 # #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name GLDH 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 31 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 30 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 8000 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 7 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 70: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2500 10 02 10 February 2008/4

GLUCOSE GOD FS Cat. No 1 2500.........

Notes 1. Please refer to the package insert for Glucose GOD

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST GLUC ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 100 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 700 – 505 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 200 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 EXPECTED VALUE 70 – 115 PANIC VALUE 1 - 400 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [GLUC] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [GLUC] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00249] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00249] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00249] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00249] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [GLUC] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [GLUC] REPORT NAME [Glucose] DATA MODE [Aktiv]

TEST NAME [GLUC] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [GLUC] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 71: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2500 10 02 10 February 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST GLUC # TEST NAME GLUC UNIT mg/dl

DATA MODE ON BOARD REPORT NAME GLucose

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

70 - 115 70 - 115

TECHNICAL LIMIT CLASS I CLASS 2

1 - 400 -

STD CON POS S.VOL. PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST GLUC

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 505

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 250 0 249 #

R2 0 0 249 #

R3 0 0 249 #

R4 0 0 249 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name GLU 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 35 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 505 11 Sample Volume 2.5 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 75 55 Expect. Value (H) 115 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 72: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2511 10 02 10 March 2009/4

GLUCOSE HEXOKINASE FS Cat. No 1 2511.........

Notes 1. Please refer to the package insert for Glucose

Hexokinase FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST GLUC ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 240 – 100 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 - 0 EXPECTED VALUE 70 – 115 PANIC VALUE 2 - 900 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [GLUC] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [GLUC] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [376] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [240] [0] [00248] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00248] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [60] [0] [00248] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00248] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [GLUC] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [GLUC] REPORT NAME [Glucose] DATA MODE [Aktiv]

TEST NAME [GLUC] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [GLUC] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 73: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2511 10 02 10 March 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST GLUC # TEST NAME GLUC UNIT mg/dl

DATA MODE ON BOARD REPORT NAME GLucose

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

70 - 115 70 - 115

TECHNICAL LIMIT CLASS I CLASS 2

2 - 500 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST GLUC

ASSAY CODE 2 point 10 WAVELGTH (SUB / MAIN)

376 / 340

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 240 0 248 #

R2 0 0 248 #

R3 60 0 248 #

R4 0 0 248 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code ATTENTION This kit is for use with the original application. If you wish to use this included DiaSys application please contact your supplier for additional bar code labels.

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name GLU 2 Assay Code (Mthd) 2 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 3 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 75 55 Expect. Value (H) 115 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 74: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3329 10 02 10 January 2010/7

HbA1c FS Cat. No 1 3329.....

Notes 1. Please refer to the package insert for oneHbA1c for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 8

weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

Attention: Immediately after HbA1c measurement cleaning of cuvettes with NaOH-D solution is necessary. Please, refer to the information of the analyzer manufacturer. **This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST HbA1c ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 180 – 20 – 0 R 2 VOLUME (µl) 90 – 20 – 0 WAVELENGTH (nm) 0 – 660 CALIBR. METHOD 3 – 4 – 5** STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. 0 – 0 SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 32000 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE # – # PANIC VALUE # – # INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Spline function

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [HbA1c] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [HbA1c] ASSAY/TIME/POINTS [1 point ] [10] [20] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [660]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [180] [0] [00027] [#]

DECREASE [4] [0] [0] [1] [0] [0] R2 [60] [0] [00027] [#] INCREASE [4] [0] [0] [1] [0] [0] R3 [30] [0] [00027] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00027] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [HbA1c] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[5] SPAN POINT [5]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [1000] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [HbA1c] REPORT NAME [HbA1c] DATA MODE [Aktiv]

TEST NAME [HbA1c] [] CONTROL INTERVAL UNIT [ # ] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [#] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [DBIL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [0]

CONCENTRATION # # # # # POSITION # # # # #

SAMPLE VOLUME [4] [4] [4] [4] [4] [0]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 75: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3329 10 02 10 January 2010/7

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST HbA1c # TEST NAME HbA1c UNIT #

DATA MODE ON BOARD REPORT NAME Hemoglobin A1c

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

# - # -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) * # 4 0 0 # (3)

(4) * # 4 0 0 # (4)

(5) * # 4 0 0 # (5)

(6) (6)

TEST HbA1c

ASSAY CODE 1 Point 10 WAVELGH (SUB / MAIN)

0 / 660

ASSAY POINTS 20 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 4 0 0

S.VOL (INCREASE) 4 0 0

ABS. LIMIT 0 INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 180 0 027 #

R2 60 0 027 #

R3 30 0 027 #

R4 0 0 027 #

CALIB. TYPE [SPLINE] 5 5 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE BLANK

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name HbA1c 2 Assay Code (Mthd) 1 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 23 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 0 10 Wave Leng. (MAIN) 660 11 Sample Volume 4.0 12 R1 VOLUME 180 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 60 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 30 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Non-Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 34 Calib. Pos. 6 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 76: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3521 10 02 10 September 2005/2

HDL-C IMMUNO FS Cat. No 1 3521.........

Notes 1. Please refer to the package insert for HDL-C

Immuno FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST HDL-C ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 240 – 20 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 700 – 600 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 35 – 80 PANIC VALUE 1 - 180 INSTRUMENT FACTOR 1.00

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [HDL] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [HDL] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [600]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [240] [0] [00035] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00035] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [60] [0] [00035] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00035] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [HDL] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [HDL] REPORT NAME [HDL] DATA MODE [Aktiv]

TEST NAME [HDL] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [HDL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 77: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3521 10 02 10 September 2005/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST HDL # TEST NAME HDL UNIT mg/dl

DATA MODE ON BOARD REPORT NAME HDL

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

35 - 80 35 - 80

TECHNICAL LIMIT CLASS I CLASS 2

1 - 180 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST HDL

ASSAY CODE 2 point end 10 WAVELGH (SUB / MAIN)

700 / 600

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 240 0 035 #

R2 0 0 035 #

R3 60 0 035 #

R4 0 0 035 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name HDL 2 Assay Code (Mthd) 2 point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 3 12 R1 VOLUME 280 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 0 54 Expect. Value (L) 35 55 Expect. Value (H) 85 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 78: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7202 10 02 10 January 2009/4

IMMUNOGLOBULIN A FS Cat. No 1 7202.........

Notes 1. Please refer to the package insert for

Immunoglobulin A FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST IGA ASSAY CODE 2 – 24 – 42 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 350 – 20 – NO R 2 VOLUME (µl) 70 – 20 – NO WAVELENGTH (nm) – 570 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE # – # PANIC VALUE 30 – 900 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [IGA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [IGA] ASSAY/TIME/POINTS [2 point end] [10] [15] [26] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [250] [0] [00777] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00777] [#] INCREASE [4] [0] [0] [1] [0] [0] R3 [50] [0] [00777] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00777] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [IGA] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [IGA] REPORT NAME [IGA] DATA MODE [Aktiv]

TEST NAME [IGA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [IGA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [2] [2] [2] [2] [2] [2]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 79: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7202 10 02 10 January 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST IGA # TEST NAME IGA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME IGA

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

30 - 900 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) * # 2 0 0 # (3)

(4) * # 2 0 0 # (4)

(5) * # 2 0 0 # (5)

(6) * # 2 0 0 # (6)

TEST IGA

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 570

ASSAY POINTS 15- 26 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 4 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 777 #

R2 0 0 777 #

R3 50 0 777 #

R4 0 0 777 #

CALIB. TYPE [LOGIT/LOG] 6 6 1

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name IgA 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 30 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 570 11 Sample Volume 3 12 R1 VOLUME 350 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 80: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7239 10 02 10 October 2009/3

IMMUNOGLOBULIN E FS Cat. No 1 7239.........

Notes 1. Please refer to the package insert for

Immunoglobulin E FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST IGE ASSAY CODE 2 – 30 – 40 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 200 – 20 – NO R 2 VOLUME (µl) 100 – 20 – NO WAVELENGTH (nm) 800 – 570 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 10000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) -32000 – 0 PROZONE LIMIT -32000 – 0 EXPECTED VALUE 0 – 100 PANIC VALUE 10 – 1000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [IGE] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [IGE] ASSAY/TIME/POINTS [2 point end] [10] [18] [24] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [5] [0] [0] [1] [0] [0] R1 [200] [0] [00432] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00432] [#] INCREASE [5] [0] [0] [1] [0] [0] R3 [100] [0] [00432] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00432] [#] ABSORBANCE LIMIT [32000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [-32000] [0] [Lower] Calibration

TEST [IGE] CALIBRATION TYPE [SPLINE] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [IGE] REPORT NAME [IGE] DATA MODE [Aktiv]

TEST NAME [IGE] [] CONTROL INTERVAL UNIT [IU/ml] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [IGE] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [5] [5] [5] [5] [5] [5]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 81: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7239 10 02 10 October 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST IGE # TEST NAME IGE UNIT IU/ml

DATA MODE ON BOARD REPORT NAME IGE

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 100 0 - 100

TECHNICAL LIMIT CLASS I CLASS 2

10 - 1000 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 5 0 0 # (1)

(2) * # 5 0 0 # (2)

(3) * # 5 0 0 # (3)

(4) * # 5 0 0 # (4)

(5) * # 5 0 0 # (5)

(6) * # 5 0 0 # (6)

TEST IGE

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

800 / 570

ASSAY POINTS 18- 24 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 5 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 5 0 0

ABS. LIMIT 32000 2:INCREASE

PROZONE LIMIT -32000 LOWER

REAGENT R1 200 0 432 #

R2 0 0 432 #

R3 100 0 432 #

R4 0 0 432 #

CALIB. TYPE [SPLINE] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 32000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name IgE 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 20 6 Assay Point 2 27 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 570 11 Sample Volume 5 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Small 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 100 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit -32000 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 100 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 82: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7212 10 02 10 January 2009/4

IMMUNOGLOBULIN G FS Cat. No 1 7212.........

Notes 1. Please refer to the package insert for

Immunoglobulin G FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST IGG ASSAY CODE 2 – 24 – 42 SAMPLE VOLUME (µl) 2 R 1 VOLUME (µl) 350 – 20 – NO R 2 VOLUME (µl) 70 – 20 – NO WAVELENGTH (nm) – 570 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE # – # PANIC VALUE 175 – 3500 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [IGG] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [IGG] ASSAY/TIME/POINTS [2 point end] [10] [15] [26] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [350] [0] [00779] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00779] [#] INCREASE [4] [0] [0] [1] [0] [0] R3 [70] [0] [00779] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00779] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [IGG] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [IGG] REPORT NAME [IGG] DATA MODE [Aktiv]

TEST NAME [IGG] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [IGG] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [2] [2] [2] [2] [2] [2]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 83: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7212 10 02 10 January 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST IGG # TEST NAME IGG UNIT mg/dl

DATA MODE ON BOARD REPORT NAME IGG

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

175 - 3500 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) * # 2 0 0 # (3)

(4) * # 2 0 0 # (4)

(5) * # 2 0 0 # (5)

(6) * # 2 0 0 # (6)

TEST IGG

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 570

ASSAY POINTS 15- 26 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 4 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 350 0 779 #

R2 0 0 779 #

R3 70 0 779 #

R4 0 0 779 #

CALIB. TYPE [LOGIT/LOG] 6 6 1

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name IgG 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 30 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 570 11 Sample Volume 2 12 R1 VOLUME 350 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 84: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7222 10 02 10 January 2009/4

IMMUNOGLOBULIN M FS Cat. No 1 7222.........

Notes 1. Please refer to the package insert for

Immunoglobulin M FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST IGM ASSAY CODE 2 – 16 – 42 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 350 – 20 – NO R 2 VOLUME (µl) 70 – 20 – NO WAVELENGTH (nm) – 405 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE # – # PANIC VALUE 25 – 800 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [IGM] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [IGM] ASSAY/TIME/POINTS [2 point end] [10] [10] [26] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [415]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [250] [0] [00789] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00789] [#] INCREASE [4] [0] [0] [1] [0] [0] R3 [50] [0] [00789] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00789] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [IGM] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [IGM] REPORT NAME [IGM] DATA MODE [Aktiv]

TEST NAME [IGM] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [IGM] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [2] [2] [2] [2] [2] [2]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 85: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7222 10 02 10 January 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST IGM # TEST NAME IGM UNIT mg/dl

DATA MODE ON BOARD REPORT NAME IGM

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

25 - 800 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) * # 2 0 0 # (3)

(4) * # 2 0 0 # (4)

(5) * # 2 0 0 # (5)

(6) * # 2 0 0 # (6)

TEST IGM

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 415

ASSAY POINTS 10- 26 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 4 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 789 #

R2 0 0 789 #

R3 50 0 789 #

R4 0 0 789 #

CALIB. TYPE [LOGIT/LOG] 6 6 1

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name IgM 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 13 6 Assay Point 2 30 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 415 11 Sample Volume 3 12 R1 VOLUME 350 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 86: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1911 10 02 10 August 2008/4

IRON FS Ferene Cat. No 1 1911.........

Notes 1. Please refer to the package insert for Iron FS

Ferene for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST IRON ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 240 – 20 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 700 – 600 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 23 – 168 PANIC VALUE 5 - 1000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [IRON] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [IRON] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [600]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [240] [0] [00661] [#]

DECREASE [7] [0] [0] [1] [0] [0] R2 [0] [0] [00661] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [60] [0] [00661] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00661] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [IRON] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [IRON] REPORT NAME [Iron] DATA MODE [Aktiv]

TEST NAME [IRON] [] CONTROL INTERVAL UNIT [µg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [IRON] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [15] [15] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 87: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1911 10 02 10 August 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST IRON # TEST NAME IRON UNIT µg/dl

DATA MODE ON BOARD REPORT NAME Iron

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

35 - 155 23 - 165

TECHNICAL LIMIT CLASS I CLASS 2

5 - 1000 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST IRON

ASSAY CODE 2 point end 10 WAVELGH (SUB / MAIN)

700 / 600

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 7 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 661 #

R2 0 0 661 #

R3 60 0 661 #

R4 0 0 661 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name Iron 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 15 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 23 55 Expect. Value (H) 168 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 88: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4001 10 02 10 August 2008/4

LACTATE FS Cat. No 1 4001.........

Notes 1. Please refer to the package insert for Lactate FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

5 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST LAC ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 240 – 100 – 0 R 2 VOLUME (µl) 60 – 20 – 0 WAVELENGTH (nm) 800 – 340 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 200 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE # – # PANIC VALUE 1 – 120 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [LAC] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [LAC] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [800] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [240] [0] [00040] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00040] [#] INCREASE [8] [0] [0] [1] [0] [0] R3 [60] [0] [00040] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00040] [#] ABSORBANCE LIMIT [32000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [LAC] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [LAC] REPORT NAME [Lactate] DATA MODE [Aktiv]

TEST NAME [LAC] [] CONTROL INTERVAL UNIT [mg/dll] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [AP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [4] [4] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 89: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4001 10 02 10 August 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST LAC # TEST NAME LAC UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Lactate

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

1 - 120 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST LAC

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

800 / 340

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 8 0 0

ABS. LIMIT 32000 32000 1:INCREASE

PROZONE LIMIT 32000 32000 1:HIGHER

REAGENT R1 240 0 040 #

R2 0 0 040 #

R3 60 0 040 #

R4 0 0 040 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name LAC 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 800 10 Wave Leng. (MAIN) 340 11 Sample Volume 4 12 R1 VOLUME 280 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.00 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 200 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0.0 55 Expect. Value (H) 1.0 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 90: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4201 10 02 10 October 2009/3

LDH FS DGKC Cat. No 1 4201.........

Notes 1. Please refer to the package insert for LDH DGKC

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST LDH ASSAY CODE 5 – 30 – 45 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 6500 – 1 PROZONE LIMIT 0 EXPECTED VALUE 225 – 480 PANIC VALUE 0 - 1200 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [LDH] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [LDH] ASSAY/TIME/POINTS [Rate A] [10] [20] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [7] [0] [0] [1] [0] [0] R1 [200] [0] [00122] [#]

DECREASE [4] [0] [0] [1] [0] [0] R2 [0] [0] [00122] [#] INCREASE [14] [0] [0] [1] [0] [0] R3 [50] [0] [00122] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00122] [#] ABSORBANCE LIMIT [6500] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [LDH] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [LDH] REPORT NAME [LDH] DATA MODE [Aktiv]

TEST NAME [LDH] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [LDH] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [7] [7] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 91: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4201 10 02 10 October 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST LDH # TEST NAME LDH UNIT U/L

DATA MODE ON BOARD REPORT NAME LDH

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

225 - 480 225 - 480

TECHNICAL LIMIT CLASS I CLASS 2

0 - 1200 -

STD CO

N

PO

S

S.VO

L.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 7 0 0 # (1)

(2) * # 7 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST LDH

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 20 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 7 0 0

S. VOL (DECREASE) 4 0 0

S.VOL (INCREASE) 14 0 0

ABS. LIMIT 6500] 1:DECREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 122 #

R2 0 0 122 #

R3 50 0 122 #

R4 0 0 122 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name LDH 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 31 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 5 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 6500 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 225 55 Expect. Value (H) 450 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 92: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4211 10 02 10 October 2009/5

LDH FS IFCC Cat. No 1 4211.........

Notes 1. Please refer to the package insert for LDH FS IFCC

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 10 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST LDH ASSAY CODE 5 – 30 – 45 SAMPLE VOLUME (µl) 4 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD **2 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. 0 – 0 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 14000 – 0 PROZONE LIMIT 0 EXPECTED VALUE #– # PANIC VALUE 0 - 1200 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [LDH] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [LDH] ASSAY/TIME/POINTS [Rate A] [10] [20] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [4] [0] [0] [1] [0] [0] R1 [200] [0] [00080] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00080] [#] INCREASE [8] [0] [0] [1] [0] [0] R3 [50] [0] [00080] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00080] [#] ABSORBANCE LIMIT [14000] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [LDH] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [LDH] REPORT NAME [LDH] DATA MODE [Aktiv]

TEST NAME [LDH] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [LDH] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [4] [4] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 93: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4211 10 02 10 October 2009/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST LDH # TEST NAME LDH UNIT U/L

DATA MODE ON BOARD REPORT NAME LDH

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

0 - 1200 -

STD CO

N

PO

S

S.VO

L.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 4 0 0 # (1)

(2) * # 4 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST LDH

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 20 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 4 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 8 0 0

ABS. LIMIT 14000] 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 080 #

R2 0 0 080 #

R3 50 0 080 #

R4 0 0 080 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name LDH 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 31 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 5 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 14000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 135 55 Expect. Value (H) 225 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 94: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4121 10 02 10 September 2005/2

LDL-C SELECT FS Cat. No 1 4121.........

Notes 1. Please refer to the package insert for LDL-C Select

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST LDL-C ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 240 – 20 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 700 – 600 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0 – 130 PANIC VALUE 1 - 400 INSTRUMENT FACTOR 1.00

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [LDL] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [LDL] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [600]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [240] [0] [00059] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00059] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [60] [0] [00059] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00059] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [LDL] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [LDL] REPORT NAME [LDL] DATA MODE [Aktiv]

TEST NAME [LDL] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [LDL] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 95: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4121 10 02 10 September 2005/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST LDL # TEST NAME LDL UNIT mg/dl

DATA MODE ON BOARD REPORT NAME LDL Cholesterol

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 130 0 - 130

TECHNICAL LIMIT CLASS I CLASS 2

1 - 400 -

STD CO

N

PO

S

S.VO

L.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST LDL

ASSAY CODE 2 point end 10 WAVELGH (SUB / MAIN)

700 / 600

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 240 0 059 #

R2 0 0 059 #

R3 60 0 059 #

R4 0 0 059 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name LDL 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 3 12 R1 VOLUME 280 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 0 54 Expect. Value (L) 0 55 Expect. Value (H) 130 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 96: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4321 10 02 10 August 2008/4

LIPASE DC FS Cat. No 1 4321.........

Notes 1. Please refer to the package insert for Lipase DC FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST LIP ASSAY CODE 5 – 32 – 39 SAMPLE VOLUME (µl) 6 R 1 VOLUME (µl) 280 – 20 – NO R 2 VOLUME (µl) 70 – 20 – NO WAVELENGTH (nm) 700 – 570 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * – 2 STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 32000 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0 – 60 PANIC VALUE 0 - 300 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

**) Factor must be checked by a calibration serum

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [LIP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [LIP] ASSAY/TIME/POINTS [Rate A] [10] [21] [25] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [5] [0] [0] [1] [0] [0] R1 [200] [0] [00731] [#]

DECREASE [3] [0] [0] [1] [0] [0] R2 [0] [0] [00731] [#] INCREASE [9] [0] [0] [1] [0] [0] R3 [50] [0] [00731] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00731] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [LIP] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [LIP] REPORT NAME [LIPASE] DATA MODE [Aktiv]

TEST NAME [LIP] [] CONTROL INTERVAL UNIT [U/l] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [LIP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [5] [5] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 97: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4321 10 02 10 August 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST LIP # TEST NAME LIP UNIT U/L

DATA MODE ON BOARD REPORT NAME Lipase

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 60 0 - 60

TECHNICAL LIMIT CLASS I CLASS 2

0 - 300 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 5 0 0 # (1)

(2) * # 5 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST LIP

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

800 / 570

ASSAY POINTS 21 - 25- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 5 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 9 0 0

ABS. LIMIT 32000 2:INCREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 731 #

R2 0 0 731 #

R3 50 0 731 #

R4 0 0 731 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ** The given factor must be checked by a calibratior ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name LIP 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 23 6 Assay Point 2 27 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 570 11 Sample Volume 6 12 R1 VOLUME 280 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 200 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 32000 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 60 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 98: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7139 10 02 10 September 2009/5

Lp(a) 21 FS Cat. No 1 7139.........

Notes 1. Please refer to the package insert for Lp (a) 21 FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST LPA ASSAY CODE 2 – 31 – 50 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 200 – 20 – NO R 2 VOLUME (µl) 100 – 20 – NO WAVELENGTH (nm) 0 – 700 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. *- * STD.3 CONC.-POS. * - * STD.4 CONC.-POS. * - * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – 30 PANIC VALUE 3 - 130 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [LPA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [LPA] ASSAY/TIME/POINTS [2 point end] [10] [19] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [700]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [5] [0] [0] [1] [0] [0] R1 [200] [0] [00242] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00242] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [100] [0] [00242] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00242] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [-32000] [0] [lower] Calibration

TEST [LPA] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [50.0] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [32000] % [300] ABS

2-POINT [0] St SENSITIVITY LIMIT 0] - [0] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [LPA] REPORT NAME [Lipoprotein a] DATA MODE [Aktiv]

TEST NAME [LPA] [] CONTROL INTERVAL UNIT [#] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [LPA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [5] [5] [5] [5] [5] [5]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 99: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7139 10 02 10 September 2009/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST LPA # TEST NAME LPA UNIT #

DATA MODE ON BOARD REPORT NAME Lp(a)

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 30 0 - 30

TECHNICAL LIMIT CLASS I CLASS 2

3 - 130 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 5 0 0 # (1)

(2) * # 5 0 0 # (2)

(3) * # 5 0 0 # (3)

(4) * # 5 0 0 # (4)

(5) * # 5 0 0 # (5)

(6) * # 5 0 0 # (6)

TEST LPA

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 700

ASSAY POINTS 19- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 5 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT -32000 LOWER

REAGENT R1 200 0 242 #

R2 0 0 242 #

R3 100 0 242 #

R4 0 0 242 #

CALIB. TYPE [SPLINE] 6 6 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 50.0

SPAN 0 DUPLICATE LIMIT 32000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name LPA 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 0 10 Wave Leng. (MAIN) 700 11 Sample Volume 5 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Small 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 100 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Spline 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 50.0 45 Duplicate Limit 32000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 30 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 100: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4610 10 02 10 August 2008/4

MAGNESIUM XL FS Cat. No 1 4610.........

Notes 1. Please refer to the package insert for Magnesium

XL FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

3 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST MG ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 700 – 546 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 1.8 – 2.6 PANIC VALUE 0.05 –5.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [MG] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [MG] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00693] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00693] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00693] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00693] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [MG] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [MG] REPORT NAME [Magnesium] DATA MODE [Aktiv]

TEST NAME [MG] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [MG] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 101: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 4610 10 02 10 August 2008/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST MG # TEST NAME MG UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Magnesium

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

1.8 - 2.6 1.8 - 2.6

TECHNICAL LIMIT CLASS I CLASS 2

0.0 - 5.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST MG

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 546

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 693 #

R2 0 0 693 #

R3 0 0 693 #

R4 0 0 693 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 500

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name MAG 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 5 5 Assay Point 1 5 6 Assay Point 2 17 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 546 11 Sample Volume 3 12 R1 VOLUME 300 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 1.9 55 Expect. Value (H) 2.5 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 102: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 5211 10 02 10 January 2009/5

PHOSPHATE FS Cat. No 1 5211.........

Notes 1. Please refer to the package insert for Phosphate FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST PHOS ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 240 – 100 – 0 R 2 VOLUME (µl) 60 – 20 – 0 WAVELENGTH (nm) 660 – 340 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 200 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 1.8 – 4.5 PANIC VALUE 0.2 – 20.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [PHOS] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [PHOS] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [660] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [240] [0] [00394] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00394] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [60] [0] [00394] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00394] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [PHOS] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [PHOS] REPORT NAME [Phosphate] DATA MODE [Aktiv]

TEST NAME [PHOS] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [PHOS] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 103: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 5211 10 02 10 January 2009/5

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST PHOS # TEST NAME PHOS UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Phosphate

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

2.6 - 4.5 1.8 - 2.6

TECHNICAL LIMIT CLASS I CLASS 2

0.2 - 40.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST PHOS

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

660 / 340

ASSAY POINTS 15 - 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 32000 2:INCREASE

PROZONE LIMIT 32000 0 1:HIGHER

REAGENT R1 240 0 394] #

R2 0 0 394] #

R3 60 0 394] #

R4 0 0 394] #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name PHOS 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 660 10 Wave Leng. (MAIN) 340 11 Sample Volume 3 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 200 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 1.8 55 Expect. Value (H) 4.5 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 104: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7022 10 02 10 August 2008/6

RHEUMATOID FACTOR FS Cat. No 1 7022.........

Notes 1. Please refer to the package insert for Rheumatoid

Factor FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

3 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST RF ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 15 R 1 VOLUME (µl) 250 – 20 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 700 – 340 CALIBR. METHOD 3 – 4 – 6 STD.1 CONC.-POS. 0.0 - 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE 0 – 15 PANIC VALUE 0 - 500 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [RF] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [RF] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [15] [0] [0] [1] [0] [0] R1 [250] [0] [00017] [#]

DECREASE [3] [0] [0] [1] [0] [0] R2 [0] [0] [00017] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [50] [0] [00017] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00017] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [RF] CALIBRATION TYPE [Spline] [ ]

CALIBRATION POINTS

[6] SPAN POINT [3]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [RF] REPORT NAME [RF] DATA MODE [Aktiv]

TEST NAME [RF] [] CONTROL INTERVAL UNIT [IU/ml] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [RF] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [15] [15] [15] [15] [15] [15]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 105: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7022 10 02 10 August 2008/6

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST RF # TEST NAME RF UNIT IU/ml

DATA MODE ON BOARD REPORT NAME RF

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 15 0 - 15

TECHNICAL LIMIT CLASS I CLASS 2

0 - 500 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 15 0 0 # (1)

(2) * # 15 0 0 # (2)

(3) * # 15 0 0 # (3)

(4) * # 15 0 0 # (4)

(5) * # 15 0 0 # (5)

(6) * # 15 0 0 # (6)

TEST RF

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

700 / 340

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 15 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 017 #

R2 0 0 017 #

R3 50 0 017 #

R4 0 0 017 #

CALIB. TYPE [SPLINE] 6 3 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name RF 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 340 11 Sample Volume 15 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Logit-Log 4P 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 2500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) 0 55 Expect. Value (H) 15 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 106: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2311 10 02 10 March 2009/4

TOTAL PROTEIN FS Cat. No 1 2311.........

Notes 1. Please refer to the package insert for Total Protein

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 10 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST TP ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 6 R 1 VOLUME (µl) 240 – 100 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 0 – 546 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 6.6 – 8.8 PANIC VALUE 0.05 – 15.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [TP] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [TP] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [6] [0] [0] [1] [0] [0] R1 [240] [0] [00376] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00376] [#] INCREASE [12] [0] [0] [1] [0] [0] R3 [60] [0] [00376] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00376] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [TP] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [TP] REPORT NAME [Total Protein] DATA MODE [Aktiv]

TEST NAME [TP] [] CONTROL INTERVAL UNIT [g/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [TP] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [6] [6] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 107: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 2311 10 02 10 March 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST TP # TEST NAME TP UNIT g/dl

DATA MODE ON BOARD REPORT NAME Total Protein

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

6.6 - 8.8 6.6 - 8.8

TECHNICAL LIMIT CLASS I CLASS 2

0.05 - 15.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 6 0 0 # (1)

(2) * # 6 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST TP

ASSAY CODE 2 point end 10 WAVELG

H

(SUB / MAIN)

0 / 546

ASSAY POINTS 15 - 31 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 6 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 12 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 376 #

R2 0 0 376 #

R3 60 0 376 #

R4 0 0 376 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name TP 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 546 11 Sample Volume 6 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size lSmall 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 6.6 55 Expect. Value (H) 8.7 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 108: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0210 10 52 10 May 2009/4

TOTAL PROTEIN UC FS Cat. No 1 0210.........

Notes 1. Please refer to the package insert for Total Protein

UC FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST TPU ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 6 R 1 VOLUME (µl) 300 – 20 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 700 – 600 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 200 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 24 – 141 PANIC VALUE 20 – 3000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [TPU] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [TPU] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [600]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [5] [0] [0] [1] [0] [0] R1 [250] [0] [00769] [#]

DECREASE [2] [0] [0] [1] [0] [0] R2 [0] [0] [00769] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [0] [0] [00769] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00769] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [TPU] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [TPU] REPORT NAME [TPU] DATA MODE [Aktiv]

TEST NAME [TPU] [] CONTROL INTERVAL UNIT [mg/L] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [TPU] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [5] [5] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 109: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 0210 10 52 10 May 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST TPU # TEST NAME TPU UNIT mg/L

DATA MODE ON BOARD REPORT NAME Urine TP

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

24 - 141 24 - 141

TECHNICAL LIMIT CLASS I CLASS 2

24 - 3000 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0 # 5 0 0 # (1)

(2) * # 5 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST TPU

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 600

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 5 0 0

S. VOL (DECREASE) 2 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 769 #

R2 0 0 769 #

R3 0 0 769 #

R4 0 0 769 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name TPU 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 35 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 5 12 R1 VOLUME 250 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 40 55 Expect. Value (H) 80 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 110: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7252 10 02 10 September 2005/2

TRANSFERRIN FS Cat. No 1 7252.........

Notes 1. Please refer to the package insert for Transferrin FS

for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST TRF ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 350 – 20 – NO R 2 VOLUME (µl) 70 – 20 – NO WAVELENGTH (nm) – 570 CALIBR. METHOD 3 – 1 – 6 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. * – * STD.4 CONC.-POS. * – * STD.5 CONC.-POS. * – * STD.6 CONC.-POS. * – * SD LIMIT 999 DUPLICATE LIMIT 1000 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 32000 – 1 EXPECTED VALUE # – # PANIC VALUE 50 – 800 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [TRF] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [TRF] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [0] [570]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [250] [0] [00216] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00216] [#] INCREASE [4] [0] [0] [1] [0] [0] R3 [50] [0] [00216] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00216] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [TRF] CALIBRATION TYPE [LOGIT/LOG 4P] [ ]

CALIBRATION POINTS

[6] SPAN POINT [6]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [999] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [15] % [500] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [TRF] REPORT NAME [Transferrin] DATA MODE [Aktiv]

TEST NAME [TRF] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [TRF] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [#] [#] [#] [#]

CONCENTRATION # # # # # # POSITION # # # # # #

SAMPLE VOLUME [2] [2] [2] [2] [2] [2]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 111: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 7252 10 02 10 September 2005/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST TRF # TEST NAME TRF UNIT mg/dl

DATA MODE ON BOARD REPORT NAME TRF

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

# - # # - #

TECHNICAL LIMIT CLASS I CLASS 2

50 - 800 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) * # 2 0 0 # (3)

(4) * # 2 0 0 # (4)

(5) * # 2 0 0 # (5)

(6) * # 2 0 0 # (6)

TEST TRF

ASSAY CODE 2 point end 10 WAVELGTH (SUB / MAIN)

0 / 570

ASSAY POINTS 15- 31 - 0- 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 4 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 250 0 216 #

R2 0 0 216 #

R3 50 0 216 #

R4 0 0 216 #

CALIB. TYPE [LOGIT/LOG] 6 6 1

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 999

SPAN 0 DUPLICATE LIMIT 1000

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name TRF 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) - 10 Wave Leng. (MAIN) 570 11 Sample Volume 3 12 R1 VOLUME 350 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 70 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Line-graph 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 # 28 Calib. Pos. 3 # 29 Calib. Conc. 4 # 30 Calib. Pos. 4 # 31 Calib. Conc. 5 # 32 Calib. Pos. 5 # 33 Calib. Conc. 6 # 34 Calib. Pos. 6 # 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 999 45 Duplicate Limit 1000 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 32000 52 Prz. Limit (U/D) Upper 53 Prz. (End Point) 35 54 Expect. Value (L) # 55 Expect. Value (H) # 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 112: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 5710 10 02 10 September 2005/2

TRIGLYCERIDES FS Cat. No 1 5710.........

Notes 1. Please refer to the package insert for Triglycerides

FS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST TRIG ASSAY CODE 1 – 50 – 0 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 300 – 100 – NO R 2 VOLUME (µl) 0 – 20 – NO WAVELENGTH (nm) 700 – 505 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 0 – 200 PANIC VALUE 1 - 1000 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [TRIG] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [TRIG] ASSAY/TIME/POINTS [1 point] [10] [31] [0] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [505]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [3] [0] [0] [1] [0] [0] R1 [250] [0] [00006] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00006] [#] INCREASE [6] [0] [0] [1] [0] [0] R3 [0] [0] [00006] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00006] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [TRIG] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [TRIG] REPORT NAME [Triglycerides] DATA MODE [Aktiv]

TEST NAME [TRIG] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [TRIG] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [3] [3] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 113: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 5710 10 02 10 September 2005/2

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST TRIG # TEST NAME TRIG UNIT mg/dl

DATA MODE ON BOARD REPORT NAME TRIGLCERIDES

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

0 - 200 0 - 200

TECHNICAL LIMIT CLASS I CLASS 2

1 - 1000

-

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 3 0 0 # (1)

(2) * # 3 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST TRIG

ASSAY CODE 1 point 10 WAVELGTH (SUB / MAIN)

700 / 505

ASSAY POINTS 31 - 0 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 3 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 6 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 0 1:HIGHER

REAGENT R1 250 0 006 #

R2 0 0 006 #

R3 0 0 006 #

R4 0 0 006 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name TRIG 2 Assay Code (Mthd) 1 Point 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 35 6 Assay Point 2 0 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 505 11 Sample Volume 3.0 12 R1 VOLUME 300 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 0 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 500 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 50 55 Expect. Value (H) 200 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 114: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1921 10 02 10 September 2009/4

UIBC FS Cat. No 1 1921.........

Notes 1. Please refer to the package insert for UIBC FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

6 days provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST UIBC ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 18 R 1 VOLUME (µl) 240 – 20 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 700 – 600 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 120 – 470 PANIC VALUE 6 - 750 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [UIBC] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [UIBC] ASSAY/TIME/POINTS [2 point end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [600]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [18] [0] [0] [1] [0] [0] R1 [240] [0] [00785] [#]

DECREASE [6] [0] [0] [1] [0] [0] R2 [0] [0] [00785] [#] INCREASE [20] [0] [0] [1] [0] [0] R3 [60] [0] [00785] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00785] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [UIBC] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [UIBC] REPORT NAME [UIBC] DATA MODE [Aktiv]

TEST NAME [UIBC] [] CONTROL INTERVAL UNIT [µg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [UIBC] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [18] [18] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 115: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 1921 10 02 10 September 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST UIBC # TEST NAME UIBC UNIT µg/dl

DATA MODE ON BOARD REPORT NAME UIBC

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

120 - 470 120 - 470

TECHNICAL LIMIT CLASS I CLASS 2

6 - 750 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0 # 18 0 0 # (1)

(2) * # 18 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST UIBC

ASSAY CODE 2 point end 10 WAVELGH (SUB / MAIN)

700 / 600

ASSAY POINTS 15 - 31- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 18 0 0

S. VOL (DECREASE) 6 0 0

S.VOL (INCREASE) 20 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 785 #

R2 0 0 785 #

R3 60 0 785 #

R4 0 0 785 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 100

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name UIBC 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 600 11 Sample Volume 18 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 120 55 Expect. Value (H) 470 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 116: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3101 10 02 10 September 2006/3

UREA FS Cat. No 1 3101.........

Notes 1. Please refer to the package insert for Urea FS for

detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is

6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST UREA ASSAY CODE 5 – 30 – 35 SAMPLE VOLUME (µl) 3 R 1 VOLUME (µl) 200 – 100 – NO R 2 VOLUME (µl) 50 – 20 – NO WAVELENGTH (nm) 405 – 340 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 8000 – 1 PROZONE LIMIT 0 EXPECTED VALUE 17 – 43 PANIC VALUE 2 - 300 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [UREA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [UREA] ASSAY/TIME/POINTS [Rate A] [10] [18] [22] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [415] [340]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [2] [0] [0] [1] [0] [0] R1 [200] [0] [00392] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00392] [#] INCREASE [5] [0] [0] [1] [0] [0] R3 [50] [0] [00392] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00392] [#] ABSORBANCE LIMIT [6500] [0] [decrease] TWINTEST [ ]

PROZONE LIMIT [0] [0] [lower] Calibration

TEST [UREA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [UREA] REPORT NAME [Urea] DATA MODE [Aktiv]

TEST NAME [UREA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [UREA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [2] [2] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 117: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3101 10 02 10 September 2006/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST UREA # TEST NAME UREA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Urea

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

17 - 43 17 - 43

TECHNICAL LIMIT CLASS I CLASS 2

2 - 300 -

STD CON POS S.VOL

.

PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 2 0 0 # (1)

(2) * # 2 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST UREA

ASSAY CODE Rate A 10 WAVELGTH (SUB / MAIN)

415 / 340

ASSAY POINTS 18 - 22- 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 2 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 5 0 0

ABS. LIMIT 6500] 1:DECREASE

PROZONE LIMIT 0 LOWER

REAGENT R1 200 0 392 #

R2 0 0 392 #

R3 50 0 392 #

R4 0 0 392 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name UREA 2 Assay Code (Mthd) Rate A 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 21 6 Assay Point 2 24 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 415 10 Wave Leng. (MAIN) 340 11 Sample Volume 2.5 12 R1 VOLUME 200 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 50 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 8000 50 ABS Limit (D/I) Decrease 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 10 55 Expect. Value (H) 50 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 118: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3021 10 02 10 March 2009/3

URIC ACID FS TBHBA Cat. No 1 3021.........

Notes 1. Please refer to the package insert for Uric Acid FS

TBHBA for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is at least one month provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST UA ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 240 - 100 - NO R 2 VOLUME (µl) 60 - 20 - NO WAVELENGTH (nm) 700 - 505 CALIBR. METHOD 1 - 0 - 0 STD.1 CONC.-POS. 0.0 - 1 STD.2 CONC.-POS. * - * STD.3 CONC.-POS. 0 - 0 STD.4 CONC.-POS. 0 - 0 STD.5 CONC.-POS. 0 - 0 STD.6 CONC.-POS. 0 - 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 - 0 PROZONE LIMIT 0 - 0 EXPECTED VALUE 2.3 – 8.2 PANIC VALUE 0.07 – 20.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [UA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [UA] ASSAY/TIME/POINTS [2-point-end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [6] [0] [0] [1] [0] [0] R1 [240] [0] [00398] [#]

DECREASE [3] [0] [0] [1] [0] [0] R2 [0] [0] [00398] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [60] [0] [00398] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00398] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [UA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [UA] REPORT NAME [Uric Acid] DATA MODE [Aktiv]

TEST NAME [UA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [UA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [6] [6] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 119: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3021 10 02 10 March 2009/3

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST UA # TEST NAME UA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Uric Acid

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

3.6 - 8.2 2.3 - 6.1

TECHNICAL LIMIT CLASS I CLASS 2

0.07 - 20.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 6 0 0 # (1)

(2) * # 6 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST UA

ASSAY CODE 2 point end 10 WAVELG

H

(SUB / MAIN)

700 / 546

ASSAY POINTS 15 - 31 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 6 0 0

S. VOL (DECREASE) 3 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 398 #

R2 0 0 398 #

R3 60 0 398 #

R4 0 0 398 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name UA 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 505 11 Sample Volume 4 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. 0 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 2.4 55 Expect. Value (H) 7.0 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer

Page 120: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3001 10 02 10 March 2009/4

URIC ACID FS TOOS Cat. No 1 3001.........

Notes 1. Please refer to the package insert for Uric Acid FS

TOOS for detailed information about the test on the following: Clinical Relevance, Method and Principle, Composition and Stability of the Reagents, Specimens, Calibrators and Controls, Performance Characteristics regarding Measuring Range / Specificity/Interferences / Sensitivity/Limit of Detection / Precision (Reproducibility, Repeatability) / Method Comparison / Reference Ranges / Literature

2. The stability of the reagent on board the analyser is 6 weeks provided that contamination and evaporation are avoided.

3. Manufactured by: DiaSys Diagnostic Systems GmbH

Alte Strasse 9, 65558 Holzheim, Germany.

**This application was set up and evaluated at DiaSys. It is based on the standard equipment at that time and does not apply to any equipment modifications undertaken by the manufacturer or by other persons.

HITACHI 717

Chemistry settings

PROGRAM 2: CHEMISTRY PARAMETERS TEST UA ASSAY CODE 2 – 24 – 50 SAMPLE VOLUME (µl) 5 R 1 VOLUME (µl) 240 – 100 – NO R 2 VOLUME (µl) 60 – 20 – NO WAVELENGTH (nm) 700 – 546 CALIBR. METHOD 1 – 0 – 0 STD.1 CONC.-POS. 0.0 – 1 STD.2 CONC.-POS. * – * STD.3 CONC.-POS. 0 – 0 STD.4 CONC.-POS. 0 – 0 STD.5 CONC.-POS. 0 – 0 STD.6 CONC.-POS. 0 – 0 SD LIMIT 0.1 DUPLICATE LIMIT 100 SENSITIVITY LIMIT 0 ABS.LIMIT(INC/DEC) 0 – 0 PROZONE LIMIT 0 – 0 EXPECTED VALUE 2.3 – 8.2 PANIC VALUE 0.3 – 20.0 INSTRUMENT FACTOR 1.0

#) Data entry by the user *) Enter calibration or standard value and position

HITACHI 912 – CHEMISTRY PARAMETERS Analysis

TEST [UA] ANALYZER CYCLE TIME [10 sec] DILUENT [water] [0] TEST NAME [UA] ASSAY/TIME/POINTS [2-point-end] [10] [15] [31] [0] [0] APPL.CODE [#] WAVELENGTH (SUB/MAIN) [700] [546]

SAMPLE VOLUME CLASS 1 CLASS 2 REAGENT NORMAL [5] [0] [0] [1] [0] [0] R1 [240] [0] [00397] [#]

DECREASE [1] [0] [0] [1] [0] [0] R2 [0] [0] [00397] [#] INCREASE [10] [0] [0] [1] [0] [0] R3 [60] [0] [00397] [#]

CLASS 1 CLASS 2 DECR./INCR. R4 [0] [0] [00397] [#] ABSORBANCE LIMIT [0] [0] [increase] TWINTEST [ ]

PROZONE LIMIT [32000] [0] [higher] Calibration

TEST [UA] CALIBRATION TYPE [Linear] [ ]

CALIBRATION POINTS

[2] SPAN POINT [2]

WEIGHT [0] AUTOMATIC CALIBRATION

AUTO TIME OUT AUTO CHANGE BLANK [0] St LOT [ ] SD LIMIT [0.1] SPAN [0] St BOTTLE [ ] DUPLICATE LIMIT [10] % [200] ABS

2-POINT [0] St SENSITIVITY LIMIT [-99999] - [99999] FULL [0] St S1 ABS LIMIT [-32000] - [32000]

Range TEST [UA] REPORT NAME [Uric Acid] DATA MODE [Aktiv]

TEST NAME [UA] [] CONTROL INTERVAL UNIT [mg/dl] INSTRUMENT FACTOR (y=ax+b) APPL. CODE [#] a= [1.0] b= [0.00]

SAMPLE TYPE [Ser/Pl] Reference ranges and technical limits: entry by user / see package insert Standards

TEST [UA] (STANDARD) (1) (2) (3) (4) (5) (6) CALIB. CODE [#] [#] [0] [0] [0] [0]

CONCENTRATION # # POSITION # #

SAMPLE VOLUME [5] [5] [1] [1] [1] [1]

DIL. SAMPLE VOLUME [0] [0] [0] [0] [0] [0] DILUENT VOLUME [0] [0] [0] [0] [0] [0]

#) data entry by the user *) enter calibration or standard value and position

Page 121: CHOLINESTERASE FS HITACHI 717 - Grupo BioSys · HITACHI 717 Chemistry settings Temperature: ... HITACHI 912 – CHEMISTRY PARAMETERS Analysis ... Calculated by the analyzer . 845

845 3001 10 02 10 March 2009/4

BM / HITACHI 911

04-01 CHEMISTRY PARAMETERS

TEST UA # TEST NAME UA UNIT mg/dl

DATA MODE ON BOARD REPORT NAME Uric Acid

CONTROL INT 0 INSTR. FACT. (Y=aX + b) a 1.0

b 0.0

EXPECTED VALUE CLASS 1 EXPECTED VALUE CLASS 2

AGE M F

- -

- -

3.6 - 8.2 2.3 - 6.1

TECHNICAL LIMIT CLASS I CLASS 2

0.3 - 20.0 -

STD CON POS S.VOL PRE. DIL. VOL CODE

LOT

QUALITATIVE

NO

(1) 0.0 # 5 0 0 # (1)

(2) * # 5 0 0 # (2)

(3) (3)

(4) (4)

(5) (5)

(6) (6)

TEST UA

ASSAY CODE 2 point end 10 WAVELG

H

(SUB / MAIN)

700 / 546

ASSAY POINTS 15 - 31 - 0 - 0 DILUTION 301 99

< CLASS 1 > < CLASS 2 >

S.VOL (NORMAL) 5 0 0

S. VOL (DECREASE) 1 0 0

S.VOL (INCREASE) 10 0 0

ABS. LIMIT 0 2:INCREASE

PROZONE LIMIT 32000 1:HIGHER

REAGENT R1 240 0 397 #

R2 0 0 397 #

R3 60 0 397 #

R4 0 0 397 #

CALIB. TYPE [1:LINEAR] 2 2 0

AUTOCALIB.

TIME OUT BLANK 0 SD LIMIT 0.1

SPAN 0 DUPLICATE LIMIT 200

2 POINT 0 SENSITIVITY LIMIT 0

FULL 0 SI ABS. LIMIT -3200032000

CHANGE LOT NO COMPENSATED LIMIT

BOTTLE NO

# Data entry by the user * Enter calibration or standard value ## Enter the next code

BM / HITACHI 902

INSTRUMENT SETTINGS

No. <Chemistry> 1 Test Name UA 2 Assay Code (Mthd) 2 Point End 3 Assay Code (2. Test) 0 4 Reaction Time 10 5 Assay Point 1 17 6 Assay Point 2 35 7 Assay Point 3 0 8 Assay Point 4 0 9 Wave Leng. (SUB) 700 10 Wave Leng. (MAIN) 546 11 Sample Volume 5 12 R1 VOLUME 240 13 R1 Pos. # 14 R1 Bottle Size Large 15 R2 VOLUME 0 16 R2 Pos. 0 17 R2 Bottle Size Small 18 R3 VOLUME 60 19 R3 Pos. # 20 R3 Bottle Size Small 21 Calib. Type (Type) Linear 22 Calib. Type (Wght) 0 23 Calib. Conc. 1 0.0 24 Calib. Pos. 1 # 25 Calib. Conc. 2 # 26 Calib. Pos. 2 # 27 Calib. Conc. 3 0 28 Calib. Pos. 3 0 29 Calib. Conc. 4 0 30 Calib. Pos. 4 0 31 Calib. Conc. 5 0 32 Calib. Pos. 5 0 33 Calib. Conc. 6 0 34 Calib. Pos. 6 0 35 S 1 ABS. 0 36 K Factor 10000 37 K 2 Factor 10000 38 K 3 Factor 10000 39 K 4 Factor 10000 40 K 5 Factor 10000 41 A Factor 0 42 B Factor 0 43 C Factor 0 44 SD Limit 0.1 45 Duplicate Limit 300 46 Sens. Limit 0 47 S 1 ABS Limit (L) -32000 48 S 1 ABS Limit (H) 32000 49 ABS Limit 0 50 ABS Limit (D/I) Increase 51 Prz. Limit 0 52 Prz. Limit (U/D) Lower 53 Prz. (End Point) 35 54 Expect. Value (L) 2.4 55 Expect. Value (H) 7.0 56 Instr. Fact. (a) 1 57 Instr. Fact. (b) 0 58 Key Setting #

#) Data entry by the user *) Calculated by the analyzer