cmut and pmut: new technology platform for medical ultrasound

33
CMUT and PMUT: New Technology Platform for Medical Ultrasound Rob van Schaijk November 2018

Upload: others

Post on 31-Oct-2021

123 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: CMUT and PMUT: New Technology Platform for Medical Ultrasound

CMUT and PMUT: New Technology Platform for Medical UltrasoundRob van Schaijk

November 2018

Page 2: CMUT and PMUT: New Technology Platform for Medical Ultrasound

MUT introduction

Page 3: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Medical ultra-sound imaging

3

Cart based systems

Portable

Hand-held

Endoscopic and intravascular

SystemsProbes and transducers

Linear array• Sound waves in straight line• Rectangular image• Often high frequency vascular probes

Curvilinear array• Wider footprint• Lower frequency for abdominal imaging

Phased array• Electronically steered beam • Small array for e.g. cardiac imaging

Endoscopic and intravascular• Full panoramic view• Doppler for blood flow dynamics• IVUS operates in high frequency range

Page 4: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Medical ultra-sound imaging application range

4

Page 5: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Bulk piezo versus MUT

5

The MEMS US revolution:• High volume production• Eliminate (manual) assembly• Low-cost platform multiple applications• Miniaturization catheters• Higher frequencies• 3D imaging compatible• Enter consumer market

Todays ultrasound imaging:• Based on piezo-ceramics• Difficult to manufacture• No volume production• Labor intensive expensive• Reserved for professional use

Page 6: CMUT and PMUT: New Technology Platform for Medical Ultrasound

CMUT versus PMUT

6

Both CMUT and PMUT are promising new technologies for US medical imaging (1MHz-40MHz)

– PMUT better in lower frequency range

• Products in < 1MHz range (gesture detection, fingerprint sensor,..)

– CMUT better suited for higher frequency range

• Advantage in catheters

Ceramic Piezo CMUT PMUT

Pros • Proven technology • Integration with electronics (on top of ASIC) 3D imaging

• High bandwidth and resolution

• Suited for high frequency (catheters)

• Integration with electronics (flip-chip on ASIC) 3D imaging

• No bias voltage needed• Low frequency range

(>40kHz)

Cons • Labor intensive and expensive

• Miniaturization• High frequencies

• High bias voltage (~100V)• Only for high frequencies

(> 1MHz)

• Challenging material (deposition, properties,lead,..)

Page 7: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Market data and forecast

7

According to Yole 2018 (*)

• PMUT will be mainly used in fingerprint sensing and gesture detection

• CMUT will be mainly used in medical ultrasound imaging

• Strong growth expected for handheld imaging device strong focus on cost

Ultrasound Sensing Technologies for Medical, Industrial, and Consumer Applications; Yole report 2018(*)

Page 8: CMUT and PMUT: New Technology Platform for Medical Ultrasound

PMUT history

8

Ultrasound Sensing Technologies for Medical, Industrial, and Consumer Applications; Yole report 2018

Research Foundry US module

• Leti• VTT• Berkeley• Imec

• STMicroelectronics

• Silex• Global Foundries• Fujifilm Dimatex

• ?

Page 9: CMUT and PMUT: New Technology Platform for Medical Ultrasound

CMUT history

9

Ultrasound Sensing Technologies for Medical, Industrial, and Consumer Applications; Yole report 2018

Research Foundry US module

• Stanford• Fraunhofer• VTT• Leti• Imec

• Philips• Global Foundries• Micralyne• Silex

• Hitachi• Verasonics

Kolo• Butterfly• Vermon

Page 10: CMUT and PMUT: New Technology Platform for Medical Ultrasound

10

MEMS US transducers

CMUT PMUT

Low Temp.Sacrificial layer

High Temp.Wafer bonding

Low Temp.PVD sputtering

High Temp.CSD Sol-gel

• Possible on CMOS ASIC (low temp.)

• Uniformity issue and stress in layer stack

• Uniform, but needs flat and smooth wafers

• High temperature process not compatible with CMOS

• Sputtering is mature and fast (expensive equipment)

• Possible on CMOS ASIC (low temp.)

• Often: (Sc doped) AlN

• Cost effective• Most common PZT• Lower throughput• Uniformity issue• Not common in

semiconductor industry

Image: Yongqiang Qiu et al., Sensors 2015, 15(4), pg. 8020-8041

Page 11: CMUT and PMUT: New Technology Platform for Medical Ultrasound

CMUT technology platform

Page 12: CMUT and PMUT: New Technology Platform for Medical Ultrasound

12

CMUTCapacitive Micromachined Ultrasound Transducer

A replacement for piezo-based ultrasound in the medical domain

• CMUT is fabricated by IC technology• Parallel plate capacitor on membrane• Transmits and receives ultrasound at 1 - 50MHz • Collapse mode: the membrane touches the cavity bottom• An RF-voltage makes the membrane vibrate

Advantages:• Robust design, large volume & low cost, high level of integration• Miniaturization & high frequency, lead free for disposable applications

12

Typically 100 μm

Page 13: CMUT and PMUT: New Technology Platform for Medical Ultrasound

CMUT modular technology platform

CMUT in general offers a lot of design freedom

Wafer

Membrane/

cavity

Bare Si wafer

ASICSubstrate

150mm 200mm

Membrane thickness

Pitch

Gap height

Dielectric

< 1 µm

Variable

< 5 µm

# cMUTs / die

Variable

< 500 µmDiameter

SiO2 SiN

Page 14: CMUT and PMUT: New Technology Platform for Medical Ultrasound

14

CMUT design

• FEM and analytical model• Validated for a wide range of frequencies

14

Membrane

Electrostatic Force

3 MHz, =280 m

20 MHz, =30 m

9 MHz,= 135 m

Frequency vs diameter

100

Diameter [m] Fr

equ

ency

[M

Hz]

Experimental dataTrend

10

110 100 1000

Page 15: CMUT and PMUT: New Technology Platform for Medical Ultrasound

15

Finish membrane & close etch holes

Silicon wafer

ASIC

Hole

Sac

Top electrode

Bottom electrode & Via

Top view of a CMUT – on - ASIC

Cross section

PhilipsMEMS Foundry

Page 16: CMUT and PMUT: New Technology Platform for Medical Ultrasound

16

MembraneEtch hole(sealed)

Electrodes Vacuum gap

Planarization layer

ASIC

Page 17: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Collapse mode operation

0 5 10 15 20 250

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

Frequency (MHz)

Pre

ss

ure

Non-Collapse CollapseBias

Advantages: High performance

Challenges: Stress sensitive Charging driftLifetime issues

Page 18: CMUT and PMUT: New Technology Platform for Medical Ultrasound

18

Automatic electrical characterization on wafer level

Page 19: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Packaging: acoustic interface

• CMUT requires a dedicated window, optimized for its application.

• FEM - optimization on attenuation, impedance and mechanical properties

Key to output characteristic: pressure and bandwidth

5 10 15 20 25 30 350

1

2

3

4

5

Frequency [MHz]S

pe

ctr

um

RTV

Poly-Butadiene Rubber (PBR)

0 5 10 15 200

5

10

15

Frequency [MHz]

Att

enu

atio

n [

dB

/mm

]

PBR

RTV

Page 20: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Low frequency example: CMUT imaging probe

20

1 7

-4

0

Frequency [MHz]

-150 VDC

-130 VDC

-120 VDC

50 70RF [V]

Pre

ssu

re a

.u.

2 pulse @ 2.5 MHz

Pressure Spectrum

0 10

-12

-20Pre

ssu

re [

dB

(n

orm

)]

20 cmFetal heart: 4 chambers

• fetal imaging, 24 weeks phantom• 20 cm penetration

CMUT CM5

PZT

PZT

CMUT

Page 21: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Collapse mode: frequency agilityExample: cMUT based Forward Looking Inter Cardiac Echo (FLICE)

21

Penetration modeBias low 6MHz

General modeBias nominal 10MHz

Resolution modeBias high 14MHz

• Image from inside the hart (aortic valve)• cMUT frequency tuning 6 14 MHz enables zooming

2x2 mm aperture

Page 22: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Live 3D ultrasound image

22

• Monolithically integrated CMUT–on-ASIC

• Test array 6x6 mm with 2000 individual elements

• Each element is one membrane

CMUT

Via

Page 23: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Miniaturization challenge: smart catheters

Page 24: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Approximately 10% of the western population will at some moment in their life be treated in a cath lab

Simplifying complex procedures to help you focus on your patients

Our vision:• Seamlessly integrated systems that make image guided therapy simpler, faster and more

cost-effective• Creating a unique, uncluttered, radiation free lab experience • Clinical specialists that become your partner to bring new live image guiding and sensing

innovation to life, enable new therapies

Page 25: CMUT and PMUT: New Technology Platform for Medical Ultrasound

Smart catheters

25

Todays Ultrasound transducers:• Obsolete technology• Analog instruments (expensive)• Many expensive (coaxial) wires• Point solutions • Not integrated in the Cath Lab

Next generation smart catheter:• State of the art technology• in-tip AD conversion• Open MEMS technology platforms• MUT with flex-to-rigid (F2R) interconnect• High speed serial interface• Standardized connector (e.g. USB type)• Fully integrated in Cath Lab infrastructure

Page 26: CMUT and PMUT: New Technology Platform for Medical Ultrasound

State-of-art: Volcano Eagle Eye IVUS catheterØ 1.2 mm catheter, 64 piezo elements around circumference

26

Transducer region

5 radially placed ASICs

Scanner

Page 27: CMUT and PMUT: New Technology Platform for Medical Ultrasound

• Seamless cMUT-interconnect integration

• Simplifies interventional device manufacturing

• Enables advanced functionality

Flex-to-Rigid (F2R) interconnect platform

Interconnect built into device:

27

Page 28: CMUT and PMUT: New Technology Platform for Medical Ultrasound

fully processed silicon substrate

1st polyimide layerinterconnect

island 1 island 2

2nd polyimide layerhard etch mask

island 1 island 2

island 1 island 2

micro flexcable

tab tab

fully processed silicon substrate

Standard Si process flows: Schematic F2R process flow

Page 29: CMUT and PMUT: New Technology Platform for Medical Ultrasound

29

Technology demonstrator As manufactured in wafer

29

Page 30: CMUT and PMUT: New Technology Platform for Medical Ultrasound

3030

Page 31: CMUT and PMUT: New Technology Platform for Medical Ultrasound

31

Ø 2 mm

Page 32: CMUT and PMUT: New Technology Platform for Medical Ultrasound

To evaluate CMUT performance and develop CMUT applications

32

Contact: [email protected]

Page 33: CMUT and PMUT: New Technology Platform for Medical Ultrasound

33

Thank you!