Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained (MQDS) Microscopic Slides No more lights off in the lab! No more microscope! Kumiko Isse, M.D., Ph.D. Demetris Lab Department of Pathology, Division of Transpla Thomas E. Starzl Transplantation Institu University of Pittsburgh Medical Center

Upload: isabell-bobbitt

Post on 14-Dec-2015

218 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Comparison of High Resolution Whole Slide Imaging (WSI)

vs Conventional Fluorescence Microscopy

for Viewing and Analyzing Multiplex Quantum Dot Immunostained

(MQDS) Microscopic Slides

No more lights off in the lab!No more microscope!

Kumiko Isse, M.D., Ph.D.Demetris Lab

Department of Pathology, Division of TransplantationThomas E. Starzl Transplantation InstituteUniversity of Pittsburgh Medical Center

Page 2: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

What is Quantum Dots (Qdots)?

Page 3: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Traditional Fluorescence Markers vs Qdots (1)

Page 4: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Adapted from 25 SEPTEMBER 1998 VOL 281 SCIENCE Adapted from Invitrogen.com

Rhodamine

Qdot

Qdot

Rhodamine

1H

3min

Traditional Fluorescence Markers vs Qdots (2)

Page 5: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

• No photo-bleaching• Wide Stokes’ shift•Narrow emission spectra •Permanent•Multiple staining

• Expensive• Special filter

Qdots

Page 6: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Enable pathologists to contribute to the molecular revolution in medicine by merging traditional morphologic examination with multiple markers to precisely characterize specific cell types and investigate intra-cellular signaling pathways

•Time consuming•Complicated protocol

Multiple Staining

Page 7: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

• Panoramic overview of tissue at low magnification

• Distribution, localization and cell-cell interactions visible

• Can unlock decades of human biology/pathology from paraffin blocks

• Connect to conventional morphology (H&E)

• Immediate sample collection and triage

• Complicated to analyze• Artifacts (wrinkles, bubbles,

dust, scratches, etc.)

Tissue Staining, not Flow Cytometry

Page 8: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

• Avidin Block• Biotin Block• Non Serum Protein Block• 1st Primary Antibody• 1st Biotinylated Secondary Antibody• 1st Streptavidin Qdot

• Extra Blocking for each segment

• Amplify Signals by 2-step immunohistochemical staining

• Avidin Block• Biotin Block• Non Serum Protein Block• 2nd Primary Antibody• 2nd Biotinylated Secondary Antibody• 2nd Streptavidin Qdot

• Avidin Block• Biotin Block• Non Serum Protein Block• 3rd Primary Antibody• 3rd Biotinylated Secondary Antibody• 3rd Streptavidin Qdot

• Repeat for additional stainings

• Before starting the panel staining, titration of antibodies will be done by immunohistochemistry to decide the staining order

• Requires the best antigen retrieval for all antibodies that you are going to use

• Antigen retrieval

Protocol for Multiplex Quantum Dot Immunostaining (MQDS)

Page 9: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

CD3 Ab + Biotynilated anti-rabbit + streptavidin Qdot CD3 Ab + Anti-rabbit Qdot

Comparison of 2-Step and 3-Step IHC

2-Step 3-Step

Page 10: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Nuance420nm

720nm

unmix

420nm 720nm

Captured Image Pseudocolor Image

Unmixed Grayscale Individual Images

Subtract AF by program

Page 11: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Skeletal Muscle 2.36 Liver 1.84

Small Intestine 1.38Heart 1.51

Frozen Liver 1.21 Colon 1.03

Lung 1.0

Same AF pattern in different tissues

Autofluorescence (AF) in Different Tissue

Page 12: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

+ =

+ =

LCA

CD68 CD3 CD68CD3

CD4 CD8 CD4CD8

LCACD68CD3CD4CD8

Page 13: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Physical problems:• Unpleasant usually isolated

environment• Limited availability

=

Data problems:• Multiple layered image with lower

opacity is unclear• Individual colors need to be saved

separately

Mechanical problems:• Repeated training often needed• Precise adjustment of settings needed for good quality images• Suboptimal at low magnification

Disadvantage of Traditional Fluorescent Microscopes

Page 14: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Qdot Filter

• Total 12 slides are scanned in one time• 10-20min by 20x lens, 20-40min by 40x lens

for biopsy size tissue• 2.2GB with 80% compression of JPG• Using filters specific for Qdots

Inside of the scannerZeiss/3D Histech scanner

What is Whole Slide Imaging (WSI)?

Page 15: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

x40

Digital x20

Digital x10

Digital x100

Page 16: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

12bit vs 8bit

Page 17: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

•Permanent data•Share the same slide with many people at once

•Observe anytime, anywhere, portable.

•No need to reserve microscope•Easy surveillance and analysis•Preservation of context and detailed morphological information

•Saves space in your lab

•Large data•Mechanical problems•Requires lot of adjustment

•Cost

Advantage of WSI

Page 18: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Digitally Preserving and Sharing the World’s Cultural Heritage

Page 19: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

WSI : HLADRCK19CD31 (3D HISTECH/ CRi Pannoramic Viewer)

Page 20: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

5 fields from liver, all portal tracts in the biopsy using three different antibodies + DAPI = 4 colorsTotal 19 cases------over 400 images

Data Obtained From WSI

Page 21: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

CD31 in x4 CD31 in x4 CD31 in x20 CD31 in x20Nuance Mirax Nuance Mirax

0

5

10

15

20

25

30

%CD31 signal

Microscope x4 Microscope x20

WSI Digital x4 WIS Digital x20

CD31

Adapted from Zeiss “Microscopy From The Very Beginning”

Microscope vs WSI

Page 22: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Unfocused Areabecause of hardware

Shifting Problem Mechanical adjustment Layer adjustment in the software

Digitally subtract AF

Multiple focus points Dark field condenser

DAPI DAPI + Q705

Disadvantage of WSI

Limited Abs numbers because of AF and DAPI

Page 23: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

http://www.farsight-toolkit.org/wiki/Main_Page

FarSight__ Developed by Dr. Badri Roysam

Page 24: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

FarSight__Nucleus Editor HLADRIL10TGFbDAPI

Page 25: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

HLADR expression and HLADR +TGFβ+/- cell numbers

Data Obtained From FarSight

Page 26: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

N=3 N=8 N=8 N=4 N=10 N=10

X40 magnification, unknown field size x50 magnification, 230x350μm2

Vδ1+CD3+Vδ2+CD3+Vδ1+2+CD3+

N=8 N=10

Problem of FarSight or Human??

Page 27: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

H&E after Qdot multiple staining Fluorescence signal

H&E Qdot multiple staining

Combitnation of H&E and MQDS

Page 28: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

• Eosin has wide spectrum( - - - - - - - Eosin)• Eosin is strong Acid• Hematoxylin is strong Base

Eosin emission spectrum on the top of Qdots

pH Ranges for Qdot® Nanocrystals

pH Recommendations >9 Not recommmended- Qdot® nanocrystals start to self-aggregate/clump.

(Qdot® nanocrystals are not degraded by basic pH. )

>6 to <9 Qdot® nanocrystals most optimal stability in this pH range.

>5 to <6 Marginal stability is shown in this range

<4 Not recommended- The polymer will dissociate; exposed core/shell will start to dissociate.

Page 29: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

CD31CD34aSMA

MQDS H&EWSI

Page 30: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

• Conventional Method

•Bottleneck because of limited resource (location and time)•Microscope often difficult to use•Not portable•Limited triage to automated image analysis •Looking at a tree and not the forest

•Easy to use, portable•Stronger signal at lower magnification•Combination of FL and Bright Field•Direct connection to automated image analysis

• Unfocused areas• Autofluorescence• Adjustments needed

Microscope WSI

Page 31: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

• Analyze whole slide image Automated whole slide image analysis using a selected region of interest (ROI)

• Better performance Hardware – computer, scanner, filter Software – imaging, analysis

Future Plan

Page 32: Comparison of High Resolution Whole Slide Imaging (WSI) vs Conventional Fluorescence Microscopy for Viewing and Analyzing Multiplex Quantum Dot Immunostained

Thank you!Rensselaer Polytechnic InstituteRoysam Lab

Dr. Badri RoysamKedar Grama

Demetris Lab

Dr. DemetrisJohn Lunz IIISusan SpechtYoshiaki MizuguchiNatasha CorbittEnrico Pegolo

MIDI and system Consultant

Andrew Lesniak

RHS Lab

Lisa ChedwickLori PerezTrevor BenyackEleck WaltonTraci Ondik

ISH Lab

Kathy Cieply