1

Comparison of Plastics used in Tissue Equivalent Proportional

Counters (TEPCs)

Tyler Collums, Eric Benton, Mohammad Rafiqul Islam, and Arthur Lucas

E. V. Benton Radiation Physics Laboratory Oklahoma State University

Page 2: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

2

A-150 tissue equivalent plastic was first described by Shonka in 1958

Developed to be tissue equivalent for photons and neutrons

Based off the International Commission of Radiological Units (ICRU) definition of muscle tissue

Has been commonly used as tissue equivalent material for energetic protons and heavy ions.

Introduction: A-150 Tissue Equivalent Plastic

Page 3: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

3

Goals of this Study To experimentally compare A-150 with four other alternative plastics to determine

their differences in response to energetic proton and heavy ion beams Nylon Acrylic Polystyrene Polyethylene

To compare all five plastics as well as the ICRU definition of muscle through Monte Carlo simulations in radiation fields using FLUKA

Based on these comparisons, to determine if any of the alternative materials may be an acceptable substitute for tissue for use in gas filled detectors

Difficult to produce due to the different melting temperatures of nylon and polyethylene

Not commercially common and more expensive due to its specialized use Lacks the structural integrity of other more common plastics

Problems with A-150

Page 4: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

4

Outline

Differences in materials Composition Attenuation coefficients, stopping powers, and cross sections

Detector Design Proton beam results

Experimental Simulation

Heavy ion results – experimental Conclusions and future work

Page 5: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

5

Acrylic

H8%

C60%

O32%

Polystyrene

H8%

C92%

ICRU Muscle

H10%

C13%

N3%

Others1%

O73%

Nylon

H11%

C67%

O12%

N10%

Polyethylene

H14%

C86%

A-150

H10%

C76%

O6%

N4%

Ca2%

F2%

Mass Percentages

ICRU Muscle A-150 Nylon Acrylic Polystyrene Polyethylene

H 10.2 10.2 10.6 8.1 7.8 14.4

C 12.3 76.8 67.2 60.0 92.2 85.6

O 72.9 5.9 12.3 32.0

N 3.5 3.6 10.0

Ca 0.007 1.8

F 1.7

Na 0.08

Mg 0.02

P 0.2

S 0.5

K 0.3

Page 6: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

6

Polystyrene

H50%

C50%

Nylon

H60%

C32%

O4%

N4%

Atomic Percentages

ICRU Muscle A-150 Nylon Acrylic Polystyrene Polyethylene

H 63.3 58.5 59.7 53.3 50 66.7

C 6.4 37.1 31.8 33.3 50 33.3

O 28.5 2.1 4.4 13.3

N 1.56 1.5 4.1

Ca 0.001 0.3

F 0.5

Na 0.02

Mg 0.01

P 0.04

S 0.10

K 0.05

ICRU Muscle

H63%C

6%

O29%

N2% Others

<1%

A-150

H59%

C37%

O2%

F1%

N1%

Ca<1%

Acrylic

H54%C

33%

O13%

Polyethylene

H67%

C33%

Page 7: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

7

Mass Attenuation Coefficients

Energy (MeV)10-4 10-3 10-2 10-1 100 101 102 103 104 105 106

Mas

s A

ttenu

atio

n C

oeffi

cien

t (cm

2 /g)

10-3

10-2

10-1

100

101

102

103

104

105

106

107 A-150 X 101

Nylon X 100.8

Acrylic X 100.6

Polystyrene X 100.4

Polyethylene X 100.2

ICRU Muscle

Energy (MeV)10-4 10-3 10-2 10-1 100 101 102 103 104 105 106

Mas

s A

ttenu

atio

n C

oeffi

cien

t (cm

2 /g)

10-3

10-2

10-1

100

101

102

103

104

A-150NylonAcrylicPolystyrenePolyethyleneICRU Muscle

Data taken from the National Institute of Standards and Technology (NIST): XCOM Photon Cross Section Database

Page 8: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

8

Proton Stopping Powers

Data taken from the National Institute of Standards and Technology (NIST): pstar - Stopping-Power and Range Tables for Protons

Energy (MeV)10-4 10-3 10-2 10-1 100 101 102 103 104 105

Sto

ppin

g P

ower

(keV

/µm

)

0

20

40

60

80

100

120A-150NylonAcrylicPolystyrenePolyethyleneICRU Muscle

Page 9: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

9

Total Neutron Cross Sections

Energy (MeV)10-1210-1110-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 100 101 102

Cro

ss s

ectio

n (b

arns

)

10-2

10-1

100

101

102

103

104

Carbon Nitrogen Oxygen Calcium Fluorine Magnesium Phosphorus Sulfur Potassium Sodium Hydrogen

Data taken from the Brookhaven National Laboratory: National Nuclear Data Center (NNDC)

Page 10: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

10

Total Neutron Cross Sections

Energy (MeV)10-1210-1110-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 100 101 102

Cro

ss s

ectio

n (b

arns

)

10-1

100

101

102

103

104

A-150 Nylon Acrylic Polystyrene Polyethylene ICRU Muscle

Energy (MeV)10-1210-1110-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 100 101 102

Cro

ss s

ectio

n (b

arns

)

10-1

100

101

102

103

104

105

A-150 X 102.5 Nylon X 102

Acrylic X 101.5

Polystyrene X 101

Polyethylene X 100.5 ICRU Muscle

Page 11: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

11

Design Fabricated at the OSU Radiation

Physics Laboratory Follows the Benjamin design

(Benjamin 1968) Two inch outer diameter 3 mm ionization cavity wall thickness Single wire anode +1400 V 173 Torr (10 μm of tissue) Inside of ionization chamber sphere is

coated with colloidal graphite (Aerodag G) to create a conductive surface

Built in preamplifier (Cremat CR-110) Contained in air tight canister (Zero

mfg.) Calibrated with an 241Am source

Page 12: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

12

Design (cont.)

Page 13: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

13

Proton Beam Experiment ProCure Proton Therapy Center in Oklahoma City Three beam energies:

87 MeV 162 MeV 222 MeV

Low flux beam: ~450 particles cm-2 s-1

Page 14: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

14

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

222 MeV protons

162 MeV protons

87 MeV protons

A-150 Nylon Acrylic Polystyrene Polyethylene

Experimental Results

Page 15: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

15

A-150NylonAcrylicPolystyrenePolyethyleneICRU Muscle

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

A-150 + 1Nylon + 0.8Acrylic + 0.6Polystyrene + 0.4Polyethylene + 0.2ICRU Muscle

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.5

1.0

1.5

2.0

2.5

87 MeV Protons FLUKA Simulation

Page 16: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

16

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.5

1.0

1.5

2.0

2.5

A-150NylonAcrylicPolystyrenePolyethyleneICRU Muscle

A-150 + 1Nylon + 0.8Acrylic + 0.6Polystyrene + 0.4Polyethylene + 0.2ICRU Muscle

162 MeV Protons FLUKA Simulation

Page 17: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

17

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

0.01 0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.5

1.0

1.5

2.0

2.5

222 MeV Protons FLUKA Simulation

A-150NylonAcrylicPolystyrenePolyethyleneICRU Muscle

A-150 + 1Nylon + 0.8Acrylic + 0.6Polystyrene + 0.4Polyethylene + 0.2ICRU Muscle

Page 18: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

18

Heavy Ion Beam Experiment

Heavy Ion Medical Accelerator in Chiba (HIMAC) at the National Institute for Radiological Sciences (NIRS) in Japan

Five beams: 150 MeV/amu He 290 MeV/amu C 490 MeV/amu Si 500 MeV/amu Ar 500 MeV/amu Fe

Low flux field (~1000 particles cm-2 s-1)

Page 19: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

19 y (keV/m)

1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

0.1 1 10 100

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

1 10 100 1000y

D(y

) (no

rmal

ized

uni

ts)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

150 MeV/amu He

290 MeV/amu C

490 MeV/amu Si

A-150 Nylon Acrylic Polystyrene Polyethylene

Experimental Results

Page 20: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

20

y (keV/m)

1 10 100 1000

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

y (keV/m)

1 10 100 1000

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

500 MeV/amu Ar 500 MeV/amu Fe

A-150 Nylon Acrylic Polystyrene Polyethylene

Experimental Results (cont.)

Page 21: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

21

Average Lineal Energy

∑=

=

=

==max

max

0

0)(

yy

y

yy

yF

y

yfyy

∑=

=

=

==max

max

0

0

2

)(

)(

yy

y

yy

yD

yfy

yfyy

( )∫∞

=0

dyyfyyF

( )∫∞

=0

21 dyyfyy

yF

Dy (keV/m)

0.01 0.1 1 10 100 1000

y D

(y) (

norm

aliz

ed u

nits

)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

222 MeV Protons 162 MeV Protons 87 MeV ProtonsHeliumCarbonSiliconArgonIron

Frequency Averaged

Lineal Energy

Dose Averaged

Lineal Energy

H He C Si

Ar Fe

A-150 TEPC

Page 22: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

22 LET (keV/µm)

0.1 1 10 100 1000

y d (ke

V/µ

m)

0

50

100

150

200

250

300TheoreticalA-150 ExperimentNylon ExperimentAcrylic ExperimentPolystyrene ExperimentPolyethylene ExperimentAverage of Plastics SimulationICRU Muscle Simulation

H He

Si

C

Ar

Fe

Dose Averaged Lineal Energies

Page 23: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

23

Conclusions

Nylon Acrylic Polystyrene Polyethylene 18.2% 6.6% 9.9% 10.8%

Average Experimental Percent Difference (yd values) with A-150 Tissue Equivalent Plastic

Acrylic showed the most similar response to that of A-150 However, simulation showed a difference of less than 1% between all five

plastics for each proton beam This would imply that the larger differences between experimental values

are more likely due to difference in geometry and fabrication of the detector and not the response of the plastics

Page 24: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

24

Future Work

Complete heavy ions simulations

Irradiations in neutrons field Experimental Simulation

Balloon flight

Page 25: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

25

CPU

MCA AmplifierCircuit

Hard Drive

High Voltage Power Supply

Integrated components CPU MCA HV power supply Battery Linear Amplifier

Weight: ~2.4 kg Size:

15X20X10 cm3 CPU container 9 cm diameter, 10 cm tall TEPC

Flight Version

Page 26: Comparison of Plastics used in Tissue Equivalent Proportional …wrmiss.org/workshops/seventeenth/Collums.pdf · 2013. 5. 3. · 2 A-150 tissue equivalent plastic was first described

26

Acknowledgements

NIRS Yukio Uchihori, Nakahiro Yasuda, Hisashi

Kitamura, Satoshi Kodaira HIMAC staff

ProCure Yuanshui Zheng IBA Staff

OSU machine shop staff