complexpermittivitymeasurementmethodofhigh ... · complexpermittivitymeasurementmethodofhigh...

4
Complex Permittivity Measurement Method of High Loss Materials Using Cylindrical Cavity Resonator in Millimeter-wave Band A. Tameishi, T. Kamijo, Y. Suzuki, A. Kik, M. Taki Department of Electrical and Electronic Engineering Tokyo Metropolitan University 1-1 Minami-Osawa, Hachioji, Tokyo 192-0397, Japan Email:[email protected] K. Sasaki National Institute of Information and Communications Technology 4-2-1 Nukui-Kitamachi, Koganei, Tokyo 184-3795, Japan Abstract—We proposed a new method to measure the complex permittivity of small biological tissue in millimeter-wave band using the TE 01n mode of a cylindrical cavity resonator. We examined the validity of the proposed measurement technique using an agar sample. The agar sample was maded by mixing pure water 98 % and agar powder 2 %. We prepared agar samples of different thickness ranging from 0.3 to 0.8 mm, and measured the complex permittivity between the 32 GHz – 35 GHz band for each sample. It was assumed that the agar samples have approximately the same complex permittivity as pure water. We compared the measured value of agar sample with the theoretical values of water. As a result, when the thickness of the agar sample of 0.57 mm or less, the measured values approximately agreed with the theoretical values of water. KeywordsComplex permittivity; Cavity resonator; Millimeter wave; biological tissue. I. I NTRODUCTION In recent years, the use of millimeter waves (30 GHz – 300 GHz) is increasing. In millimeter-wave band, the majority of energy is absorbed at the surface of the human body. Therefore, there are concerns about the influence of millimeter waves on the eye or skin. Numerical dosimetry is often used as a mean to study the effects of electromagnetic waves on human body [1]. Since there are few reports about the permittivity of biological tissues in millimeter wave band, the extrapolated values of permittivity from values at lower frequencies reported by Gabriel et al. [2] are often used for numerical calculation in millimeter-wave band. Since the relax- ation frequency of water is higher than 20 GHz, exterapolated values may not necessarily provide relevant esitimation. The complex permittivity of biological tissues has been measured with various methods such as the coaxial probe method and the free space method. However, it is difficult to measure the complex permittivity of some biological tissues using the existing methods. One example is the cornea covering the eye. The coaxial probe method [3], [4], the free-space method [5]– [8], and the waveguide penetration method [9] have been used for measurement of complex permittivity of high loss materials including biological tissues. The cornea has a diameter about 10 mm with a thickness of several hundred μm. It is difficult to measure its complex permittivity. The coaxial probe method usually that the material is semi-infinite homogeneous medium. So it is difficult for this method to measure thin materials such as cornea. In the free space method, measurement of small sample is difficult. In the waveguide penetration method, the measurement is performed by inserting the sample into small tube. Therefore, this method is suitable for the measurement of liquids, but it is not suitable for the measurement of solid samples because it is difficult to insert the material into the thin tube. In order to measure the complex permittivity of small and thin tissues with high loss such as the cornea, we propose a new measurement method that uses TE 01n mode cylindrical cavity resonator. Cavity resonator is suitable for measurement of solid with small volume. Originally, resonator methods are used for measuring complex permittivity measurement of low loss materials [10], [11]. In this study, in order to measure the complex permittivity of small size and high loss samples we place the sample at the end face of the resonator where the electric field is small. We will examine the validity of the proposed method by using a solid agar sample. II. TE 01n MODE CYLINDRICAL CAVITY RESONATOR A. Theory We insert a disc-shaped sample of thickness t on the end face of a cylindrical cavity resonator with length L and radius a as shown in Fig. 1. In this case, the following equation is satisfied from the bound- ary conditions of the TE 01n mode (n represents the number of resonant modes in the z-axis direction) electromagnetic field at boundary of Region 1 (air) and Region 2 (sample) in the resonator. tanh ˙ γ z2 t ˙ γ z2 + tanh ˙ γ z1 (L - t) ˙ γ z1 =0. (1) ˙ γ z1 and ˙ γ z2 are the propagation constants in the z-axis direction in the region 1 and the region 2. They are represented by the following equations. ˙ γ z1 2 = - { ˙ ω 2 ε 0 μ 0 - k 2 c } (2) ˙ γ z2 2 = - { ˙ ω 2 ε 0 ˙ ε r μ 0 - k 2 c } , (3) where ε 0 and μ 0 represent the permittivity and permeability of vacuum respectively. ˙ ε r = ε r - ′′ r represents the relative EMC’14/Tokyo Copyright 2014 IEICE 15P2-A5 541

Upload: doliem

Post on 20-Apr-2018

217 views

Category:

Documents


4 download

TRANSCRIPT

Complex Permittivity Measurement Method of HighLoss Materials Using Cylindrical Cavity Resonator

in Millimeter-wave Band

A. Tameishi, T. Kamijo, Y. Suzuki, A. Kik, M. TakiDepartment of Electrical and

Electronic EngineeringTokyo Metropolitan University

1-1 Minami-Osawa, Hachioji, Tokyo 192-0397, JapanEmail:[email protected]

K. SasakiNational Institute of Information and

Communications Technology4-2-1 Nukui-Kitamachi, Koganei, Tokyo 184-3795, Japan

Abstract—We proposed a new method to measure the complexpermittivity of small biological tissue in millimeter-wave bandusing the TE01n mode of a cylindrical cavity resonator. Weexamined the validity of the proposed measurement techniqueusing an agar sample. The agar sample was maded by mixingpure water 98 % and agar powder 2 %. We prepared agarsamples of different thickness ranging from 0.3 to 0.8 mm, andmeasured the complex permittivity between the 32 GHz – 35 GHzband for each sample. It was assumed that the agar samples haveapproximately the same complex permittivity as pure water. Wecompared the measured value of agar sample with the theoreticalvalues of water. As a result, when the thickness of the agar sampleof 0.57 mm or less, the measured values approximately agreedwith the theoretical values of water.

Keywords—Complex permittivity; Cavity resonator; Millimeterwave; biological tissue.

I. INTRODUCTION

In recent years, the use of millimeter waves (30 GHz –300 GHz) is increasing. In millimeter-wave band, the majorityof energy is absorbed at the surface of the human body.Therefore, there are concerns about the influence of millimeterwaves on the eye or skin. Numerical dosimetry is often usedas a mean to study the effects of electromagnetic waveson human body [1]. Since there are few reports about thepermittivity of biological tissues in millimeter wave band,the extrapolated values of permittivity from values at lowerfrequencies reported by Gabriel et al. [2] are often used fornumerical calculation in millimeter-wave band. Since the relax-ation frequency of water is higher than 20 GHz, exterapolatedvalues may not necessarily provide relevant esitimation. Thecomplex permittivity of biological tissues has been measuredwith various methods such as the coaxial probe method andthe free space method. However, it is difficult to measurethe complex permittivity of some biological tissues using theexisting methods. One example is the cornea covering the eye.The coaxial probe method [3], [4], the free-space method [5]–[8], and the waveguide penetration method [9] have been usedfor measurement of complex permittivity of high loss materialsincluding biological tissues. The cornea has a diameter about10 mm with a thickness of several hundred µm. It is difficultto measure its complex permittivity. The coaxial probe methodusually that the material is semi-infinite homogeneous medium.

So it is difficult for this method to measure thin materials suchas cornea. In the free space method, measurement of smallsample is difficult. In the waveguide penetration method, themeasurement is performed by inserting the sample into smalltube. Therefore, this method is suitable for the measurementof liquids, but it is not suitable for the measurement of solidsamples because it is difficult to insert the material into thethin tube.

In order to measure the complex permittivity of small andthin tissues with high loss such as the cornea, we propose anew measurement method that uses TE01n mode cylindricalcavity resonator. Cavity resonator is suitable for measurementof solid with small volume. Originally, resonator methods areused for measuring complex permittivity measurement of lowloss materials [10], [11]. In this study, in order to measurethe complex permittivity of small size and high loss sampleswe place the sample at the end face of the resonator wherethe electric field is small. We will examine the validity of theproposed method by using a solid agar sample.

II. TE01n MODE CYLINDRICAL CAVITY RESONATOR

A. Theory

We insert a disc-shaped sample of thickness t on the endface of a cylindrical cavity resonator with length L and radiusa as shown in Fig. 1.In this case, the following equation is satisfied from the bound-ary conditions of the TE01n mode (n represents the number ofresonant modes in the z-axis direction) electromagnetic fieldat boundary of Region 1 (air) and Region 2 (sample) in theresonator.

tanh ˙γz2t

˙γz2+

tanh ˙γz1(L− t)

˙γz1= 0. (1)

˙γz1 and ˙γz2 are the propagation constants in the z-axisdirection in the region 1 and the region 2. They are representedby the following equations.

˙γz12 = −

{ω̇2ε0µ0 − k2c

}(2)

˙γz22 = −

{ω̇2ε0ε̇rµ0 − k2c

}, (3)

where ε0 and µ0 represent the permittivity and permeabilityof vacuum respectively. ε̇r = ε′r − jε′′r represents the relative

EMC’14/Tokyo

Copyright 2014 IEICE

15P2-A5

541

Fig. 1. Dimensions used for TE01n cavity resonator with high loss sample

complex permittivity of the sample in the region 2. ω̇ is thecomplex resonance angular frequency, and is expressed in (4).

ω̇ = 2πf

{1 + j

1

2

(1

Q− 1

Q0

)}(4)

Here, Q0 is the quality factor of the empty resonator. f and Qare resonant frequency and quality factor respectively, whenthe sample is inserted in the resonator. kc represents the wavenumber in the radial direction of the cylindrical surface of theresonator. kc is represented by (5).

kc =ρ′01a

=3.832

a(5)

In (5), ρ′01 represents the first zero of the derivative of thebessel function J ′

0(ρ′).

B. Measurement system and the cavity resonator

We measured the complex permittivity using the TE01n

mode of a cylindrical cavity resonator. Figure 2 is the blockdiagram of the measurement system. Figure 3 is a photographof the resonator. Figure 4 shows the sample insertion holderwhich is placed on the bottom surface of the resonator. Theresonator is as long as about 496 mm in length to obtain highresolution of frequency. As a result the resonant frequenciesappear at the interval of about 200 MHz as shown in Fig.5 owing to the large length of the resonator. The complexpermittivity is obtained at this frequency interval.

TM11 mode and TE01 mode are degenerate with each other.In order to attenuate the TM11 mode, we used the resonatorof spiral structure on the sidewall. This resonator method wasoriginally developed as a method for measuring the complexpermittivity of low-loss materials by placing the sample in thecenter of resonator where the electric field is maximal [10],[11]. In this study, we place the sample at the end face ofthe resonator, where the electric field is minimal, to allowmeasurement of high-loss material.

The cavity resonator method has an advantage in the smallsample volume required to measure the complex permittivity.This merit is desirable for measurement of cornea of rabbit,whose volume is very small.

Fig. 2. Block diagram of measurement system

Fig. 3. Photograph of measurement system

Fig. 4. Photograph of the sample insertion holder

Fig. 5. The resonance curves between 33 GHz and 34 GHz (n = 68− 72)

EMC’14/Tokyo

Copyright 2014 IEICE

15P2-A5

542

C. Experimental Method

First, we measure the resonance frequency f0 and qualityfactor Q0 of the empty reasonator. Then the sample of discshape is inserted with the sample insertion holder as shown inFig. 4. Then we measure the resonant frequency f and qualityfactor Q with the sample inserted. Finally, we substitute theobtained values in the expressions (2) – (4), and calculate thecomplex permittivity using equation (1).

D. Determination of the dimensions of the resonator

The resonator used in the experiments is designed to be14 mm inner diameter and 496 mm in length. However, sincethe cavity inside has a spiral structure, 14 mm inner diameteris not a correct value. Therefore, we determined the resonatorlength L and its effective inner diameter a.The resonant frequency of the empty resonator f0 satisfies thefollowing equation.

4π2ε0µ0 · f20 =

(ρ01a

)2

+(nπL

)2

(6)

We measured the resonance frequency for several resonancemodes n. Then we determined a and L by using the leastsquares method, with a and L as unknowns in equation (6).As a result, the values a = 14.083 mm and L = 496.56 mmwere obtained.

III. INVESTIGATION OF THE VALIDITY OF THEMEASUREMENT METHOD

A. Experimental sample

We investigated the validity of the measurement methodby measuring a sample with known permittivity. A goodsample for use in investigation is a sample having a complexpermittivity close to that of the biological tissues and can befreely adjusted in thickness. Water is a sample that satisfiesthese requirements. However, there are some problems in theexperiment because water is a liquid. The first problem isthat the resonator must be kept in a vertical position. Thesecond problem is that the meniscus effect is being formed.The thickness of the sample becomes not uniform due tomeniscus formation. Therefore we used a sample of agar whichis solidified by mixing the agar powder with pure water (98%pure water, 2% agar powder) instead of water.

We prepared agar samples of different thickness t rangingfrom 0.3 mm to 0.8 mm, and measured the complex permit-tivity for the 32 GHz – 35 GHz band for each sample. Thethickness of the agar sample was measured using a micrometer.The mesured complex permittivity of agar was compared withthe literature values of the complex permittivity of pure water[12]. Since 98% of the sample are composed of pure water,agar was assumed to have approximately the same complexpermittivity as pure water.

B. Experimental results and discussion

Figures 6 and 7 show the measurement results for theresonant frequency f and quality factor Q of agar. Theroom temperature during measurement was 25.6 ◦C. Measuredresonant frequency and quality factor were compared withthe theoretical value for pure water. The theoretical values

were calculated by reference to the values of the complexpermittivity of water by Kaaze et al. [12] and (1) – (4).When the thickness of the agar sample of 0.57 mm or less,the measured values approximately agreed with the theoreticalvalues. Figures 8 and 9 show the measurement results forthe relative permittivity of agar (the average of the results inthe sample of 0.57 mm or less). Solid line is the complexpermittivity of water at the same temperature. Differencebetween measured value and theoretical value were up to 8%in the real part and up to 15% in the imaginary part. Thecause of the difference could be attributed to the errors in themeasurement of sample thickness, the inaccurate levelness ofthe sample face and the difference of the complex permittivitybetween the agar and pure water.

The result of the experiment shows that the complexpermittivity of agar sample of 0.57 mm or less in thicknesscan be measured. However, measurement is difficult for thesamples with a thickness of 0.68 mm or more. As samplebecomes thicker, electromagnetic waves do not penetrate intothe end of the sample where the wall of the resonator lies. Inthis case the resonance due to reflection at the surface of thesample could prevail over the resonance due to the reflectionat the wall of the resonator. Different formula of estimatingcomplex permittivity should be derived in order to measurethe complex dielectric constant of a sample of 0.57 mm ormore in thickness.

Fig. 6. Resonant frequency as function of the thickness of agar sample

Fig. 7. Quality factor as function of the thickness of agar sample

EMC’14/Tokyo

Copyright 2014 IEICE

15P2-A5

543

Fig. 8. Real part of the relative permittivity of agar sample (T = 25.6 ◦C)

Fig. 9. Imaginary part of the relative permittivity of agar sample (T = 25.6◦C)

IV. CONCLUSION

In order to measure the complex permittivity of cornea ofrabbit eye, we proposed a measurement method that employsa cylindrical cavity resonator. We examined the validity of theproposed method by using agar sample. When the thicknessof the agar sample is 0.57 mm or less, we found that thecomplex permittivity can be measured. We will measure thecomplex permittivity of the cornea of the rabbit eye with thismethod for the next step. It should be noted that the thicknessof the cornea of the rabbit eye, the final goal of measurement,is about 0.40 mm. We consider that the proposed method isapplicable to the measurement of the complex permittivity ofthe cornea of rabbit eye.

ACKNOWLEDGMENT

This study is supported by Research on Biological EffectProject by Ministry of Internal Affairs and Communications(Japan).

REFERENCES

[1] Y. Suzuki, A. Koike, K. Sasaki, M. Taki, M. Kojima, T. Sakai, K. Wake,S. Watanabe, Y. Yamashiro, H. Sasaki and K.Sasaki, ”Investigation ondosimetry of millimeter-wave exposure considering the interaction withfluids” , IEICE Technical Report, Vol.EMCJ-110, pp.9-14, 2010.

[2] S. Gabriel, R. W. Lan and C. Gabriel ”The dielectric properties ofbiological tissues : III. Parametric models for the dielectric spectrumtissues”, Phys. Med. Biol., vol.41, pp.2271-2293, 1996.

[3] T. P. Marsland and S. Evans, ”Dielectric measurements with an open-ended probe”, Proc. IEE, vol.134, Pt. H, pp.341-349, 1987.

[4] A. Takeda, K.Takata, Y. Sato, Y. Taka and O. Fujiwara ”Measurementof complex relative permittivity of pure water and its variation withrespect to water amount using open-ended coaxial probe”, IEEJ Trans.on Fundamentals and Materials, 131(7), pp.523-528, 2011.

[5] M. N. Afsar, J. R. Birch and R. N. Clarke, G. W. Chantry, ”TheMeasurement of the Properties of Materials”, Proc. IEEE, vol.74,pp.183-199, 1986.

[6] J. Baker-Jarvis, M. D. Janezic, P. D. Domich and G. G. Geyer, ”Analysisof an open-ended coaxial probe with lift-off”, IEEE Trans. on Instrum.Meas., IM-43, pp.711-718, 1994.

[7] S. S. Stuchly, G. Gajda, L. Anderson and A. Kraszewski, ”A NewSensor for Dielectric Measurements”, Trans IEEE Instrum. Meas., IM-35, pp.138-141, 1986.

[8] K. Sasaki, H. Segawa, M. Mizuno, K. Wake, S. Watanabe and O.Hashimoto, ”Development of the complex permittivity measurementsystem for high-loss biological samples using the free space method inquasi-millimeter and millimeter wave bands”, Phys. Med. Biol., vol.58,pp.1625-1633, 2013.

[9] A. Nishikata, ”Scattering Analysis for Layered Cylindrical ObjectPerpendicularly Piercing the Wider Walls of a Rectangular Waveguideand Its Application to and Measurement”, IEEE Trans. Microw. TheoryTech, vol.57, no.6, pp.1602-1611, 2009.

[10] R. Ueno ,T. Kamijo, N. Ogasawara and T. Ini : ”Measurements of nat-ural resonance for magnetic materials at Millimeter-wave frequencies”,The 1977 IECE General Conference, No.580 (in Japanese) (1977-3).

[11] R. J. Cook, ”Microwave cavity methods”, High frequency dielectricmeasurement (Edited by J. Chamberlain et al.), IPC Science andTechnology Press Ltd., pp.12-27, 1972.

[12] U. Kaatze, ”Complex Permittivity of Water as a Function of Frequencyand Temperature”, Journal of Chemical and Engineering Data, vol.34,pp.371-374, 1989.

EMC’14/Tokyo

Copyright 2014 IEICE

15P2-A5

544