concept generation mechanical actuator for magnetic resonance elastography of the lumbar spine...
TRANSCRIPT
Concept Generation
Mechanical Actuator for Magnetic Resonance Elastography of the Lumbar Spine
Overview
Background Information
Project Scope
Metrics
Design Process
AcknowledgementsOur team would like to thank our sponsor, Dr. Cortes, our
advisor, Dr. Wang, along with the BMEG design staff
Prototype
Final Design
Validation
Christian Aurup, Caitlin Banks, Brittany King & Kenneth Warren
The project scope is to design a functioning MRI compatible mechanical apparatus with the ability to displace a patient’s lumbar intervertebral discs to measure disc mechanical properties. An end goal is to have the apparatus available for use in a research facility.
Dr. Cortes has partnered with the Spine Biomechanics Lab in using Magnetic Resonance Elastography (MRE) to vibrate intervertebral spine tissue in-vitro. Vibrating the soft tissue creates a displacement that can be measured in an MRI scan. This displacement assesses the degradation and mechanical properties of the tissue.
Research was conducted on different types of actuators that are currently being used with MRE to vibrate and measure different types of tissue. Our sponsor had used a copper solenoid coil, which relied on the magnetic waves generated by the MRI machine to induce a vibrational movement (Figure 1). An amplifier and waveform generator was connected as a power source. We followed his basic design of the copper solenoid coil for our design.
The complete device consists of the mechanical actuator with its housing, a body RF coil, wave generator, and amplifier. The actuator and housing rests under the patient’s lumbar spine and on top of the MRI table. The wave generator and amplifier are outside of the MRI room, connected to the actuator via a cable. Four coils, each with 700 turns, are connected on opposite corners by two long arms. Each arm is connected to each other by a center rod, which holds two back supports and a rectangular frame with two contact points. Three different heights were created for the contact point, depending on tissue thickness variability. The magnetic current of the MRI machine creates a torque in the coils, causing them to vibrate. The center framing acts as
Once this was decided, calculations were performed to see how many coils and turns of copper wire were need. Safety and force generated had to be considered.
0 10 20 30 40 50 60 7070
75
80
85
90
95
Coil Temperature Test
Time (min)
Tem
pera
ture
(°F)
Load cell testing was done, where 125lbs was applied to each back support without any deformation. For the temperature testing (shown left), the coils were in parallel with 8V and 0.2A of current
pine
Design
Figure 1
Actuator TypeSolenoid actuatorPiezoelectric actuator
Motor actuatorAcoustic actuatorHydraulic actuatorPneumatic actuator
Force DeliveryLever mechanism
“See-saw” mechanism
Eccentric disks Rotating camPistons
Coil TypeSingle, R/F Coil
Body Coil
Testing
x
y
Displacement X Direction
Displacement Y Direction
Displacement Z Direction
the see-saw mechanism, and causes the contact points to hit the spine, causingintervertebral disc displacement.