Considerations for Considerations for Sampling Wet, Sampling Wet, High Pressure, and High Pressure, and Supercritical Supercritical Natural Gas Natural Gas Presented by: Shannon M. Bromley

Upload: doque

Post on 06-Mar-2018

216 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

Considerations for Considerations for Sampling Wet, Sampling Wet,

High Pressure, and High Pressure, and Supercritical Supercritical Natural GasNatural Gas

Presented by: Shannon M. Bromley

Page 2: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

2

IntroductionIntroduction

The key to analyzer performance and accurate analysis is proper sample conditioning system (SCS) design and operation.

If designed incorrectly, it will result in the largest source of problems for the analyzer and compromise the integrity of the sample.

The sample conditioning system consists of all components which contact the sample.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 3: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

3

Sample Conditioning System TasksSample Conditioning System Tasks

a,f

b

c

d,f

e

© 2009 A+ Corporation, LLC. All rights reserved.

a) Extract a representative sample of the gas phaseb) Reduce (regulate) the pressurec) Transport the sample to the analyzer without changing its

compositiond) Protect the analyzer (analyzer “safety net”)e) Control flow ratef) Remove contaminants

Page 4: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

4

Sample integrity MUST be maintained Sample integrity MUST be maintained during the sample conditioning process.during the sample conditioning process.

Preserve sample integrity by:Removing contaminants (solids and liquids), when present, in the pipeline.Reducing (regulating) the pressure without traversing the phase envelope.Maintaining the sample in the gas phase at all times after extraction.

At least 30°F of de-saturation is required by the API 14.1 standard to keep away from the hydrocarbon dew point (HDP).

© 2009 A+ Corporation, LLC. All rights reserved.

Page 5: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

5

In order to design, operate, and troubleshoot sample In order to design, operate, and troubleshoot sample conditioning systems one must know some basic conditioning systems one must know some basic Chemistry & Physics related to sample conditioning.Chemistry & Physics related to sample conditioning.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 6: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

6

Vapor/Liquid EquilibriumVapor/Liquid Equilibrium

© 2009 A+ Corporation, LLC. All rights reserved.

Gas / VaporPhase

Liquid Phase

What Happens?

Pressure

Decrease

Gas Phase Composition

Altered

Page 7: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

7

Vapor/Liquid EquilibriumVapor/Liquid Equilibrium

What Happens?

© 2009 A+ Corporation, LLC. All rights reserved.

Pressure

Increase

Gas / VaporPhase

Liquid PhaseGas Phase

Composition Altered

Page 8: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

8

Key Point:Key Point:

When liquid is entrained in the source gas, changes in either the pressure or the temperature will change the gas phase composition.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 9: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

9

The first step in sampling natural gas is The first step in sampling natural gas is the extraction processthe extraction process

The most severe problems are encountered during sample extraction.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 10: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

10

What is a What is a ““representativerepresentative”” sample?sample?

Industry standards are not absolutely clear on the definition of “representative” when liquid is present in the source gas.

There is also a lack of clear consensus in the industry on this subject.

There is NO industry standard that covers the sampling of wet or supercritical gases.

How we define a representative sample is the key to how the sample is conditioned.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 11: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

11

How do the industry standards define what a How do the industry standards define what a ““representativerepresentative”” sample is?sample is?

The standards imply that a representative natural gas sample when liquid is present in the source consist of:

“The gas phase at the prevailing pressure and temperature of the source gas at the point of sampling.”

© 2009 A+ Corporation, LLC. All rights reserved.

Page 12: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

12

Entrained liquid must be separatedEntrained liquid must be separated

No technology is available for extracting a gas sample containing a representative amount of entrained liquid

Making matters more difficult is the fact that liquid is constantly changing forms in the pipeline.

This is backed by the API 14.1 standard (Appendix B, Section B.3 - “Multi-phase flow”)

It states, in summary, that current sampling technology is not sufficiently advanced to obtain a sample representing both phases.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 13: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

13

When liquids are present, we CANNOT:When liquids are present, we CANNOT:

Heat tracing (or ambient warmer than gas)

Gas warmer than ambient

Can’t heat Can’t cool

© 2009 A+ Corporation, LLC. All rights reserved.

Can’t

drop pressure

externall

y

Can’t drop pressure

internally

Can’t ignore the liquid!

Page 14: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

14

Since we cannot ignore the liquid, what Since we cannot ignore the liquid, what can we do?can we do?

1. Separate the liquid from the gas phase at line conditions of pressure and temperature.

Use a membrane tipped probe inserted vertically into the sample source to accomplish this task.

2. Take steps to “de-saturate” the sample.

Drop the pressure

Increase the temperature

© 2009 A+ Corporation, LLC. All rights reserved.

Page 15: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

15

Caution:Caution: Improper Sequence of eventsImproper Sequence of events

If steps to de-saturate (pressure reduction and heating) the sample were taken BEFORE liquid was separated at line conditions, the following are likely to occur:

Liquids ( an arbitrary amount) would become vaporizedVapor composition is altered (ie-richer)

BTU value is increased

Physical constants calculated from the composition are incorrect

Flow (volume) calculations are incorrect

Monetary value assigned to Natural Gas is incorrect

© 2009 A+ Corporation, LLC. All rights reserved.

Page 16: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

16

Phase Separation MembranePhase Separation Membrane

Phase Separation Membrane

Liquid

Gas Flows Through Freely

Is Separated

Chemically inert, NOT chemically selective© 2009 A+ Corporation, LLC. All rights reserved.

Page 17: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

17

GPA 2166 & Membrane SeparationGPA 2166 & Membrane Separation

Section 7.3.3“An acceptable method for removal of unwanted contaminants is a membrane filter inserted directly in the line.”

Appendix B.1.8To effectively separate unwanted liquids, the device must be operating at flowing temperature and pressure conditions.

© 2009 A+ Corporation, LLC. All rights reserved.© 2008 A+ Corporation, LLC. All rights reserved.

Page 18: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

18

Membrane Tipped ProbeMembrane Tipped Probe

Membrane

GasFlow

CoalescedLiquid

© 2009 A+ Corporation, LLC. All rights reserved.

Page 19: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

19

Topic

THE USE OF PHASE DIAGRAMS THE USE OF PHASE DIAGRAMS IN THE DESIGN & OPERATION IN THE DESIGN & OPERATION OF A NATURAL GAS SYSTEMOF A NATURAL GAS SYSTEM

© 2009 A+ Corporation, LLC. All rights reserved.

Page 20: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

20© 2009 A+ Corporation, LLC. All rights reserved.© 2008 A+ Corporation, LLC. All rights reserved.

1. Liquid Region

3. Phase Envelope“Multi‐Phase Region”

Critical Point

Cricondenbar

2. Vapor Region

Cricondentherm

4.  Supercritical Fluid Region

A

B

C

D

E

Page 21: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

21

DEFINING THE DEFINING THE ““PRACTICALPRACTICAL””TEMPERATURETEMPERATURE

© 2009 A+ Corporation, LLC. All rights reserved.

Page 22: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

22

Insertion Pressure Regulators are not Insertion Pressure Regulators are not 100% effective in preventing J100% effective in preventing J--T coolingT cooling

Note that insertion pressure regulators are designed to “offset” or prevent excessive Joule-Thomson (J-T) cooling during pressure reduction.

This is accomplished by having the pressure reduction valve located in a section of the sample probe which is inside of the pipeline. This allows the sample gas to be “heat sinked” to the process gas.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 23: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

23

Insertion Pressure Regulators are not 100% Insertion Pressure Regulators are not 100% effective in preventing Jeffective in preventing J--T cooling T cooling (continued)(continued)

Experience has shown the actual “practical”temperature achieved during pressure reduction with an insertion regulator to be near the midpoint of the adiabatic (no heat transfer) and isothermal (without temperature change) temperature points.

The “practical” temperature point will be used in this discussion for guidance in the design and operation of sample systems.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 24: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

24© 2009 A+ Corporation, LLC. All rights reserved.

·Critical Point

Page 25: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

25

Topic

PROBLEMS ENCOUNTERED WHEN PROBLEMS ENCOUNTERED WHEN SAMPLING & CONDITIONING WET, SAMPLING & CONDITIONING WET, HIGHHIGH--PRESSURE, AND PRESSURE, AND SUPERCRITICAL NATURAL GAS & SUPERCRITICAL NATURAL GAS & SOLUTIONSSOLUTIONS

© 2009 A+ Corporation, LLC. All rights reserved.

Page 26: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

26

Problem # 1Problem # 1

How does one extract & condition a “representative” sample when liquid in any form is present in the gas sample (i.e.- “wet”gas)?

© 2009 A+ Corporation, LLC. All rights reserved.

Page 27: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

27© 2009 A+ Corporation, LLC. All rights reserved.

Page 28: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

28

In order to arrive at a solution, one In order to arrive at a solution, one must determine the following:must determine the following:

Determine what type of probe should be usedDetermine if internal or external pressure regulation is requiredDetermine if heating is required

In order to make these determinations, a phase diagram for the gas composition to be extracted should be utilized.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 29: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

29

Solution for Solution for ““wetwet”” gasgas

1) Separate liquid inside of the pipelineA “phase separation” membrane tipped probe may be used for this purpose.

Removing liquid externally is not practical because line conditions would have to be maintained EXACTLY as they are in the pipeline.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 30: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

30© 2009 A+ Corporation, LLC. All rights reserved.

Page 31: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

31© 2009 A+ Corporation, LLC. All rights reserved.

Page 32: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

32

Solution for Solution for ““wet gaswet gas”” (continued)(continued)

2) Plot a new phase diagram using ONLY the GAS PHASE composition to determine if an insertion regulator can be used or if a heated regulator is required.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 33: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

33© 2009 A+ Corporation, LLC. All rights reserved.

Page 34: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

34© 2009 A+ Corporation, LLC. All rights reserved.

Page 35: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

35© 2009 A+ Corporation, LLC. All rights reserved.

Page 36: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

36© 2009 A+ Corporation, LLC. All rights reserved.

Page 37: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

37

Problem # 2Problem # 2

How does one extract & condition a “representative” sample from a supercritical or high pressure natural gas source, even though it does not contain entrained liquid?

© 2009 A+ Corporation, LLC. All rights reserved.

Page 38: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

38© 2009 A+ Corporation, LLC. All rights reserved.© 2008 A+ Corporation, LLC. All rights reserved.

1. Liquid Region

3. Phase Envelope“Multi‐Phase Region”

Critical Point

Cricondenbar

2. Vapor Region

Cricondentherm

4.  Supercritical Fluid Region

A

B

C

D

E

Page 39: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

39

In order to arrive at a solution, one In order to arrive at a solution, one must determine the following:must determine the following:

Determine what type of probe should be usedDetermine if internal or external pressure regulation is requiredDetermine if heating is required

In order to make these determinations, a phase diagram for the gas composition to be extracted should be utilized.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 40: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

40© 2009 A+ Corporation, LLC. All rights reserved.

Page 41: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

41

Solution for extracting & conditioning the Solution for extracting & conditioning the supercritical natural gas sample on previous supercritical natural gas sample on previous slideslide

Extract the sample under conditions which ensure the pressure and temperature changes do not allow the sample to change from its supercritical state.

Transfer the sample to an external heated multi-stage regulator through a sample line heated at least 30°F above the cricondentherm temperature.

Multiple stages of pressure regulation, with re-heating of the sample gas between stages is recommended

© 2009 A+ Corporation, LLC. All rights reserved.

Page 42: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

42© 2009 A+ Corporation, LLC. All rights reserved.

Page 43: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

43

Sampling High Pressure GasSampling High Pressure Gas

Although the focus of the last example was on sampling a supercritical gas (also high pressure), the considerations are the same for a high pressure gas. The following will need to be determined:

Determine what type of probe should be usedDetermine if internal or external pressure regulation is requiredDetermine if heating is required

In order to make these determinations, a phase diagram for the gas composition to be extracted should be utilized.

© 2009 A+ Corporation, LLC. All rights reserved.

Page 44: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

44

SummarySummary

The recommendations in this presentation are based on experience in similar circumstances.The sample extraction process for high pressure, wet, or supercritical gases is the most critical step in the analytical process.

Sample will be invalid if this step is performed incorrectly

Another key step is pressure regulation© 2009 A+ Corporation, LLC. All rights reserved.

Page 45: Considerations for Sampling Wet, High Pressure, and ... · PDF fileConsiderations for Sampling Wet, High Pressure, ... reduction with an insertion regulator to be ... HIGH-PRESSURE,

45

SummarySummary

Design or operation of a sample system for wet, high pressure or supercritical natural gas requires a basic knowledge of thermodynamics Understanding phase diagrams is a “must”for making sample conditioning decisions.The API 14.1 standard is a good reference even though its scope is limited to gases at or above their hydrocarbon dew point.