Broadband UV LED measurements at NIST G. P. Eppeldauer, V. B. Podobedov, C. C. Cooksey NIST, Gaithersburg Sensor Science Division CORM Conference, Troy, NY July 31 – Aug 2, 2017

Upload: others

Post on 02-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Broadband UV LED measurements at NIST

G. P. Eppeldauer, V. B. Podobedov, C. C. CookseyNIST, Gaithersburg

Sensor Science Division

CORM Conference, Troy, NYJuly 31 – Aug 2, 2017

Page 2: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Outline

• Problem: There is no standard for broadband UV measurements. Results are large errors

• Solution of problem: Use broadband (standardized) measurement procedure

To simplify procedure: Use irradiance meters of constant (flat) spectral responsivity: Filtered Si

Low-NEP pyroelectric

Measure integrated irradiance from LED sources

Page 3: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Broadband measurement procedure

A broadband UV measurement procedure, earlier developed at NIST, is used where the source distribution and meter response functions are designed (or selected) such that output signal is not affected by the changes of these functions.

The NIST suggested standard broadband measurement procedure is:The spectral response of the UV meters is chosen such, that the spectral product of the source-distribution and meter-responsivity will produce signals with differences (errors) less than the required measurement-uncertainty when different UV meters (models) and/or different UV sources are used.

Page 4: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Integrated irradiance measurement

When a reference UV-meter with flat-response is used:

sref(l) = s = constant,

where the unit of i is A, the unit of s is A cm2/W,

and the obtained unit for is W/cm2.

lll

dEsiiref )(

s

iE

E

l

lll d)()( refref sEi

Measurement equation:

Standard source is not needed!

E

is measured by the reference UV-meter

where E(l) is the spectral irradiance of the source, sref(l) is the spectral responsivity of the reference UV-meter, and l is the wavelength.

Page 5: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Development of reference irradiance meters1) Filtered-Si UV meters

for LED-365 sources and the 345-440 nm range

In the UV-365 meters multilayer thin-film and glass filters are used. They are stable but their shapes are different. The response changes are close to 10 % !They will be used as field-level reference meters.

0.05

0.055

0.06

0.065

0.07

345 350 355 360 365 370 375 380

BCDEFG

Wavelength [nm]

0

0.05

0.1

0.15

0.2

0.25

320 340 360 380 400 420 440

C

D

Wavelength [nm]

+/- 7.5 % response changes

Page 6: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

2) Flat-responsivity pyroelectric detector of low-NEPfor high-level (reference) standard

The absorptance of the black detector-coating, which is proportional to the response of the detector, is equal to 1–reflectance (measured) if the transmittance of the coating is negligibly small. Irradiance responsivity tie point(s) can convert the flat relative response curve into absolute spectral irradiance responsivity.

Page 7: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Flat spectral irradiance responsivity of the reference pyroelectric detector

Irradiance responsivity tiepoint using 660 nm LED against Si-trap detector. Spectral irradiance responsivity uncertainty is 0.25 % (k=1).

Realization of a new UV-VIS irradiance responsivity scale

Relative spectral response from spectral reflectance measurement

0.9

0.92

0.94

0.96

0.98

1

860

880

900

920

940

0.3 0.4 0.5 0.6 0.7 0.8

Wavelength [m]

Const. +/- 0.2% 250 to 400 nm

Comparison to FEL

Page 8: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

UV LED broadband measurements using the constant-response pyroelectric detector

ILT E275P (1 mW flux) E = 0.71 W/cm2 at 19.8 mAThorlabs M265L3 (10 mW) E = 80 W/cm2 at 300 mAThorlabs M285L4 (45 mW) E = 0.44 mW/cm2 at 325 mATenzi 365 E = 2.20 mW/cm2 at 400 mALED Engin LZ1 365 (1 W) E = 48.6 mW/cm2 at 1.0 AOld blue LED 400 nm E = 0.14 mW/cm2 at 400 mA

Integrated irradiance, E, at 40 cm; Uncert. <1% (k=2):

0

2 103

4 103

6 103

8 103

1 104

240 260 280 300 320 340

Wavelength [nm]

TENZI 365Hex homogenizer

Thorlabs mounted deep-UV LEDs

f /1 UV-lens added for collimation

Thorlabs deep-UV LEDs

Page 9: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Variable output (0.5 - 7 mW/cm2) LED-365 irradiance sourcefor Army non-destructive crack-inspection at 40 cm distance

LED ENGIN, Model LZ1-10UV00-0000, LED clusterwith integrated glass lens, 1 W output-flux at 1A.

2” diam. and 2” fL UV lens to LED distance is 33 mm

LED temperature (to 25.0 oC) and feed-current are stabilized

Page 10: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Variable output (0.5 - 7 mW/cm2) LED-365 irradiance sourcecharacteristics

368.0

368.5

369.0

369.5

370.0

370.5

0 0.2 0.4 0.6 0.8 1

LED current [A]

0

50

100

150

0 0.2 0.4 0.6 0.8 1

LED current [A]

0

5000

10000

15000

20000

300 320 340 360 380 400 420 440

Integrated irradiance: 48.6 mW/cm2

at 1.0 A LED feeding current and source - detector distance of 40 cm

Spatial nonuniformity of target spot of -12 mm to +12 mm (for both X and Y) at 40 cm: +/- 5 %;for -5 to +5 mm spot: +/- 0.4 %

Spectrometer measured spectralpower distribution at 0.1 A LEDfeeding current.

For current changefrom 50 mA to 1 Apeak changes 1.5 nm

Integrated irradiance can be used for UV hazardevaluations as well !

0

0.2

0.4

0.6

0.8

1

-40 -30 -20 -10 0 10 20 30 40

XY

Distance from center [mm]

Page 11: CORM2017 broadb UV LED meas (short) (Eppeldauer) · %urdgedqg phdvxuhphqw surfhgxuh $ eurdgedqg 89 phdvxuhphqw surfhgxuh hduolhu ghyhorshg dw 1,67 lv xvhg zkhuh wkh vrxufh glvwulexwlrq

Conclusions

The NIST developed broadband UV measurement procedure has been simplified to measure integrated-irradiance from UV LEDs simply with low uncertainty

A broadband UV-LED measurement capability has been developed at NIST including development of flat-response UV irradiance meters:

o Filtered Si (for field applications in UV-A)o Low-NEP pyroelectric (standard detector for 250 nm to 750 nm)

The pyroelectric standard (without a source standard) can be used to• calibrate UV-A meters and to• measure UV-VIS LEDs for integrated (broadband) irradiance