CT physics and instrumentation
Lecture (10)
Radiation Dosimtery
RSSI 471
Prepared by
Mr. Essam Mohammed Alkhybari
![Page 2: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/2.jpg)
Mr. Essam Mohammed AlkhybariRadiological science and Medical
Imaging DepartmentLecturer in Nuclear Medicine streamPrince salman universityE-mail: [email protected]
Staff contact information:
![Page 3: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/3.jpg)
Radiation quantity and their unitsBiological effectsCT dosimtery measurementCT and radiation riskEstimation Effective doseDiagnostic reference levels
Objectives:
![Page 4: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/4.jpg)
Radiation dose quantities are described in numerous ways:
1. Exposure 2. Absorbed dose3. Equivalent dose 4. Cumulative Equivalent dose 5. Effective dose
Radiation quantities and their units
![Page 5: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/5.jpg)
Definition:
The amount of energy ionized in air per unit of mass
Number of ions produced in air by photonConventional unit Rontjen (R) SI coulomb per kilogram (C/kg)
1- Exposure:
![Page 6: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/6.jpg)
2- Absorbed dose:
Definition:The amount of energy deposited per unit of material (e.g., human tissue), is called the absorb dose.SI unit gray (Gy)Conventional unit rad
![Page 7: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/7.jpg)
Definition:It is used to assess how much biological damage is expected from the absorbed dose from specific type of radiation based on radiation quality factor (x-ray. Gamma ray, alpha …etc) Different types of radiation have different damaging
properties.The unit for the quantity equivalent dose is the Sievert (Sv)Equivalent dose (in Sv) = absorbed dose (in Gy) x radiation
weighting factor(Quality factor, RBE)
3- Equivalent dose:
![Page 8: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/8.jpg)
Radiation Weighting factors
![Page 9: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/9.jpg)
Definition:Total energy absorbed by patient’s body per examination or procedure.SI unit mGy× cmConventional unit:mrad× cm
4- Cumulative radiation dose:
![Page 10: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/10.jpg)
Definition:
Probability of a harmful effect from radiation exposure depends on what part or parts of the body are exposed, and
A method is required to permit comparison of the risks when different organs are irradiated
A tissue weighting factor is used to take into account that some organs are more sensitive to radiation than others.
When an equivalent dose to an organ is multiplied by the tissue weighting factor for that organ the result is the effective dose to that organ.
The unit of effective dose is the sievert (Sv). convetional unit: rem
5- Effective Dose
![Page 11: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/11.jpg)
Biological effect: hazard
One of the reasons why technologists and radiologists should have a clear understanding of the dose in CT relates to radiation biological effect.
Can be classified as:1. Stochastic2. Deterministic (non-
stochastic)
![Page 12: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/12.jpg)
What makes ionizing Radiation dangerous?
Biological effect – hrs, days, years following exposure, next generation, not at all, depends on bonds broken.
![Page 13: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/13.jpg)
1.Compare system2.Compare protocol3.Estimate Patient risk
Why measurement dose is important ?
![Page 14: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/14.jpg)
Tools for measurement:
Pencil ionisation chamber- 100 mm length (Measurements free-in-air)
Standard dosimetry phantom (with pencil ionization chamber) using head and body phantom
Alternatives: TLD, solid state detectors.
![Page 15: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/15.jpg)
CT radiation measurement:
Measurement or estimation the radiation dose quantity in CT can be performed by using:
1. CT Dose Index (CTDI) or Multiple Scan average Dose (MSAD)
2. Dose Length Product (DLP)
![Page 16: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/16.jpg)
MSAD:
Definition: average of absorbed dose from series of slices
Because the early CT examinations consist of a series of “ stop-and-go” scan (slices), MSAD was the dose descriptor for use in a clinical situation at that time.
The MSAD concept will be highlighted for historical reasons only.
![Page 17: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/17.jpg)
CTDI:
The CTDI is the special quantity used to express radiation dose in CT.
When the appropriate factors are applied to convert the measured phantom CTDI to an actual patient scan, the CTDI is a reasonable estimation of the actual absorbed dose to the patient.
![Page 18: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/18.jpg)
CTDI :
CT Dose Index (CTDI) is the measure of ionizing radiation exposure per slice of data acquisition.
CTDI is the total energy absorbed within a dose profile deposited within one nominal collimation
CTDI= area( mGy)/ T (slice thickness mm)
![Page 19: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/19.jpg)
Dose at a single point: CTDI100 :
100 (CTDI100) is used to denote the measurement length.
Allowed calculation of the index for 100 mm along the length of entire pencil ionization chamber
Don’t normally want to measure dose at a point
We want dose in area or volume
![Page 20: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/20.jpg)
Weighted CTDIw:
CTDIw is a weighted average of the CTDI at the center and periphery of the phantom
CTDIw represents the average radiation dose over the x and y direction
CTDIW = Weighted avg. of center (1/3)+ peripheral (2/3) contributions of dose.
![Page 21: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/21.jpg)
Volume CTDIvol :
CTDIvol (or CTDI volume) represents the dose for
a specific scan protocol which takes into account
gaps and overlaps between the radiation dose
profile from consecutive rotations of the x-ray
source.
CTDIvol is the approximate average radiation dose
over x, y, and z axis of the patient
CTDIvol is similar to CTDIw but also includes the
effect of pitch on the radiation dose.
CTDIVOL= CTDIW/pitch
![Page 22: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/22.jpg)
CT radiation measurement: dose length product (DLP):
Definition (DLP):
The DLP is a Practical Quantity for Expressing the Total Radiation Energy Deposited in the Body and to estimate radiation risk
DLP= CTDIVOL x scan length
(mGy.cm) A way to relate scan to risk
Estimate stochastic radiation risk
![Page 23: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/23.jpg)
CT Dose Descriptors
![Page 24: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/24.jpg)
Effective Dose from DLP on console of CT scanner
![Page 25: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/25.jpg)
CT and Risk:
25
Exposure (in phantom)
CTDI (dose in phantom per slice)
Length of scan and pitch
DLP
Effective dose
Risk
![Page 26: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/26.jpg)
![Page 27: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/27.jpg)
Definition of DRLs:
Dose levels in medical radiodiagnostic practices or, in the case of radio-pharmaceuticals, levels of activity, for typical examinations for groups of standard-sized patients or standard phantoms for broadly defined types of equipment. These levels are expected not to be exceeded for standard procedures when good and normal practice regarding diagnostic and technical performance is applied."
Diagnostic Reference Level(DRL):
![Page 28: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/28.jpg)
DRLs
![Page 29: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/29.jpg)
The main purposes of establishing DRLs in CT are to limit the number of unjustifiably high-dose examinations by promoting good practice and implementing the ALARA principle
The levels are set at approximately the 75th percentile of these measured data, meaning that the procedures are performed at most institutions with doses at or below the reference level.
If such doses are found to exceed the corresponding reference dose(75%), possible causes should be investigated and corrective action taken accordingly, unless the unusually high doses could be clinically justified.
DRLs:
![Page 30: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/30.jpg)
DRLs can be established at different levels:
1) DRLs are established by a country at a national level.
2) Local DRLs are established by a hospital or group of hospitals to monitor local practice.
DRLs:
![Page 31: CT physics and instrumentation Lecture (10) Radiation Dosimtery RSSI 471 Prepared by Mr. Essam Mohammed Alkhybari](https://reader036.vdocuments.net/reader036/viewer/2022062517/56649ebe5503460f94bc7955/html5/thumbnails/31.jpg)
The impact of this new technology on established DRLs needs to be investigated , especially if changes in protocols, procedures, or equipment are affected
DRLs: