Czerny-Turner Spectrometer − Analysis of Spectral Resolution

Optical Metrology > Spectrometer

Task/System Illustration

www.LightTrans.com2

power

?

Highlights

• high-performance analysis of complex optical systems• full vectorial analysis of gratings by using rigorous algorithm (FMM)

www.LightTrans.com3

Specification: Light Source

www.LightTrans.com4

here

Parameter Description / Value & Unittype plane wavewavelengths 534.5 nm − 535.5 nm in steps of 10 pmpolarization linear in x-direction (0°)

Specification: Apertures

www.LightTrans.com5

here

here

Parameter Description / Value & Unitwidth of entrance aperture 500 µmwidth of detector aperture 649 µm

Specification: Parabolic Mirrors

www.LightTrans.com6

Parameter Description / Value & Unittype parabolic mirrormaterial ideal high-reflective materialfocal length 1 mdiameter 20 mmtilt angle 5°reflectance 100%

here

Specification: Grating

www.LightTrans.com7

Parameter Description / Value & Unitgrating period 833 nmgrating height 282.4 nm (optimized for -1st order efficiency)blaze angle 18.7°grating material silver (Ag)substrate material silver (Ag)

here

gratingheight

grating periodblaze angle

Specification: Detectors

www.LightTrans.com8

a, b

Position Modeling Technique Detector / Analyzerfull system

3D ray tracing 3D ray tracing system visualization

a field tracing power detection

Result: 3D Ray Tracing

www.LightTrans.com9

source screen gratingdetectoraperture

parabolicmirror 1

entrance aperture

parabolicmirror 2

for better illustration thedistances and focal length of the mirrors were reduced to 100 mm

Result: Variation of Wavelength

www.LightTrans.com10

detector aperture

image

by varying the wavelength, the image of the entrance aperture is scanned over the detector aperture:

Result: Variation of Wavelength

www.LightTrans.com11

detector aperture

image

by varying the wavelength, the image of the entrance aperture is scanned over the detector aperture:

detected power:

Result: Resolution of Monochromator

www.LightTrans.com12

detected power:

50%

𝛥𝛥𝛥𝛥FWHM

𝛥𝛥0

definition of spectral resolution:

𝐴𝐴 =𝛥𝛥0𝛥𝛥𝛥𝛥

𝛥𝛥0 = 535 nm 𝛥𝛥𝛥𝛥FWHM = 0.43 nm

spectral resolution: 𝐴𝐴 =1244