2
HM8018 R Measurement Functions: L, C, R, Θ, Q/D, |Z| R Basic Accuracy 0.2% R 5 Measurement Frequencies: 100Hz, 120Hz, 1kHz, 10kHz, 25kHz R Max. Resolution: 0.001Ω, 0.001pF, 0.01µH R 2- and 4-Wire Measurement, parallel and serial Mode 25kHz LCR-Meter HM8018 Option HZ18 Kelvin Test Lead Mainframe HM8001-2 required for Operation Option HZ19 SMD Test Tweezers

Upload: pho-duc-nam

Post on 13-Sep-2015

2 views

Category:

Documents


1 download

DESCRIPTION

HM8018

TRANSCRIPT

  • HM8018

    R MeasurementFunctions:L,C,R,,Q/D,|Z|

    R BasicAccuracy0.2%

    R 5MeasurementFrequencies:100Hz,120Hz,1kHz,10kHz,25kHz

    R Max.Resolution:0.001,0.001pF,0.01H

    R 2-and4-WireMeasurement,parallelandserialMode

    25kHz LCR-Meter HM8018

    Option HZ18 Kelvin Test Lead

    Mainframe HM8001-2 required for Operation

    Option HZ19 SMD Test Tweezers

  • HM8018E/100812 C&E Subject to change without notice HAMEG Instruments GmbH DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08

    HAMEGInstrumentsGmbHIndustriestr.6D-63533MainhausenTel+49(0)61828000Fax+49(0)[email protected]

    25 kHz-LCR-Meter HM8018Alldatavalidat23Cafter30minuteswarm-up.

    Measurement functionsMeasuringmodes: R, L, C, , Q/D, |Z|Equivalentcircuits: serial, parallelMeasuringmethod: 2-wire, 4-wireMeasuringranges: R: 0.001 99.9 M

    C: 0.001 pF99.9 mF L: 0.01 H9,999 H Q: 0.000199.9 D: 0.00019.9999 : (-180.00)(+180.00)

    Basicaccuracy: 0.2 %Measuringfrequencies: 100 Hz, 120 Hz, 1 kHz, 10 kHz, 25 kHzFreq.Accuracy: 100 ppm (except 120 Hz: 120.2 Hz 100 ppm)Measuringvoltage: 0.5 Vrms 10 % (unloaded)Measuringrate: 2 measurements/secondRangechanging: automatic, manualDCBiasvoltage: 1 V 10 %Zerosetting: Open/short circuit compensationCompensationlimits: Short: R