development and human factors analysis of neuronavigation vs. augmented reality a. pandya 1,2, m.r....
Post on 21-Dec-2015
215 views
TRANSCRIPT
![Page 1: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/1.jpg)
Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality
A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2
1 Electrical and Computer Engineering Department, 2 Biomedical Engineering Department
3 Computer Science DepartmentWayne State University
Detroit, Mi.
![Page 2: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/2.jpg)
Augmented Reality (AR) VS
Neuronavigation (VR)
Research Question:
Does AR offer any Advantages over VR?
![Page 3: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/3.jpg)
Augmented Reality Vs. Neuronavigation
(Its not enough to develop technology, we must test it.)
Augmented Reality
Neuronavigation (VR)
Movie
![Page 4: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/4.jpg)
Client-Server Implementation
TrackedVideo
Stream
3D Models or Sensor
Data
Robotic Kinematic
Server
Object Registration
Software
Image GuidanceClients
Augmented RealityClients
Camera Calibration
Data
Robotic Hardware
Mounted Camera System
![Page 5: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/5.jpg)
Real-time Augmentation of Objects in a Phantom Skull
![Page 6: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/6.jpg)
Number of Errors AR vs. NNError Comparison
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
1 3 5 7 9 11 13 15 17 19 21
Subject Number
Nu
mb
er
of
Err
ors
Errors in NN
Errors in AR
Average STD P value
AR 1.04 1.020.06846
NN 1.71 1.27
![Page 7: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/7.jpg)
AR Takes Less Time Time Comparison
0
2
4
6
8
10
12
14
1 3 5 7 9 11 13 15 17 19 21
Subject Number
Tim
e T
ak
en
(m
in)
Time for NN
Time for AR
Average STD P value
AR 5.38 2.740.00013
NN 8.47 1.90
![Page 8: Development and Human Factors Analysis of Neuronavigation vs. Augmented Reality A. Pandya 1,2, M.R. Siadat 3, G. Auner 1,2, M. Kalash 1, R.D. Ellis 2 apandya@ece.eng.wayne.edu](https://reader035.vdocuments.net/reader035/viewer/2022070323/56649d635503460f94a45a75/html5/thumbnails/8.jpg)
Summary: Implementation of a prototype augmented robotics. Performed a verification study of system accuracy. Implemented a Neuronavigation system. Performed a verification study of the system accuracy. 21 Subjects study to determine performance
enhancements of AR vs. Neuronavigation
Future Work: Expand the HF study to include more complex
procedures. Implement AR on active medical robotic Investigate the use of registered sensors. Investigate the use of 3D ultrasound with AR.
It is not enough to develop technology, it must be tested for performance enhancement with the end-user in mind.