Development of a Device to Measure
Constancy of Output, Symmetry and
Energy of a Clinical Linear Accelerator
Using Low-Cost Photodiodes
Mark Magness, Velindre Cancer Centre
![Page 2: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/2.jpg)
Monitoring the beam
![Page 3: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/3.jpg)
Monitoring the beam
• Output – how much radiation is
being delivered?
![Page 4: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/4.jpg)
Monitoring the beam
• Output – how much radiation is
being delivered?
• Symmetry – the shape of the
beam being delivered.
![Page 5: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/5.jpg)
Monitoring the beam
• Output – how much radiation is
being delivered?
• Symmetry – the shape of the
beam being delivered.
• Energy – How deep is the beam
being delivered?
![Page 6: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/6.jpg)
Monitoring the beam
• Output – how much radiation is
being delivered?
• Symmetry – the shape of the
beam being delivered.
• Energy – How deep is the beam
being delivered? <2%
![Page 7: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/7.jpg)
Current Daily Constancy Device
![Page 8: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/8.jpg)
Varian CTB CN-732-A
![Page 9: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/9.jpg)
Why Photodiodes?
• Using diodes instead of chambers in the
device would eliminate the need for HT.
• Existing radiation diodes are prohibitively
expensive to have one device per machine.
• Normal Photodiodes are designed to detect
photons and have an obvious plane of
sensitivity to position perpendicular to the
beam.
• Small, light package, no assembly.
![Page 10: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/10.jpg)
3 types of Diode
1N4001 VTB8440B BPW34
![Page 11: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/11.jpg)
Radiation Hardening
• 64.5cm from target (Shadow Tray
Mount)
• 10mm Buildup
• 40 x 40 field
• 6MeV electrons (Adjusted to deliver
1200Mu/min)
![Page 12: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/12.jpg)
2.5kGy All Readings are in % relative to the initial reading
99
100
101
102
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Diode 1N4001
97
98
99
100
101
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Diode VTB8440B
90
110
130
150
170
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Exposure No.
Diode BPW34
![Page 13: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/13.jpg)
5.2kGy
106
107
108
109
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Diode 1N4001
46
47
48
49
50
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Diode VTB8440B
2940
2960
2980
3000
3020
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Exposure No.
Diode BPW34
All Readings are in % relative to the initial reading
![Page 14: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/14.jpg)
6.5kGy
136
137
138
139
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Diode 1N4001
46.5
47.5
48.5
49.5
50.5
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Diode VTB8440B
2600
2620
2640
2660
2680
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Exposure No.
Diode BPW34
All Readings are in % relative to the initial reading
![Page 15: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/15.jpg)
Diode Vs Chamber
• 25 readings Vs chamber
• Raw diode % variance from mean =
±0.23%
• After comparing against thimble chamber
and subtracting beam variance the actual
diode variance = ±0.16%
![Page 16: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/16.jpg)
Symmetry Prototype
![Page 17: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/17.jpg)
CNC PCB Milling
![Page 18: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/18.jpg)
The Block
![Page 19: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/19.jpg)
![Page 20: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/20.jpg)
Full Prototype
![Page 21: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/21.jpg)
Software
![Page 22: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/22.jpg)
![Page 23: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/23.jpg)
Final Device
![Page 24: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/24.jpg)
Device Aspects
• Uses Low-cost photodiodes to measure Output,
Symmetry and Energy
• Internal temperature sensor for temperature
correction of readings, reduces user error
• Micro-switch interlocked block eliminates spurious
readings through incorrect build-up usage
• Software Based Front End
• All readings and calibration factors are recorded
both locally and backed up to a network database,
reducing chance of loss
![Page 26: Development of a Device to Measure Constancy of Output, … M Magness.pdf · 2015-10-27 · Constancy of Output, Symmetry and Energy of a Clinical Linear Accelerator Using Low-Cost](https://reader034.vdocuments.net/reader034/viewer/2022050603/5faad7105a6fa864c0405413/html5/thumbnails/26.jpg)
Acknowledgments
Machine Maintenance, VCC
• Tim Register – Liaising with heads of departments regarding budgets, training etc.
• Stephen Hayes – Original Concept, Analogue Circuit Design
• Andrew Bevan – Temperature Sensor test and selection, Thermocouple circuit design
• Anthony Frost – Interface box design and build
Mechanical Workshop, VCC
• Andrew Edwards – Help with development of CNC PCB Milling
• Robin Hymers – Device Block and Enclosure Design and Build
Medical Physics, VCC
• Richard Jarvis – Developing the correct specification for the device
• All the other physicists - For answering my random questions, taking comparison readings, etc.