development of computer controlled eyes for patient simulator

1
Development of Computer Controlled Eyes for Patient Simulator Silviu. C. Diaconu 1 , Emadaldine A. Elsamadicy 1 , Anas Othman 1 , Azreena A. Shukri 1 Sponsors: Ray Booker 2 , Matthew Weinger, M.D. 2 1 Vanderbilt University School of, Engineering, 2 Center for Experiential Learning and Assessment ACCOMPLISHMENTS LCD SCREENS (LILIPUT 212GL-18NP) 3D EYE (DX STUDIO) 3D Studio Max is used to model the anatomical structures of the eye, while DX studio was used as the real-time rendering engine for controlling the animations. OUR EXPERIMENTAL SETUP Prototype: The anatomical and physiological responses of the eye were successfully modeled and animated such that they can be controlled by an external keyboard with the following controls: full range eye movement, blinking, pupil dilation /constriction, as well as varying degrees of eye redness. The two LCD screens were integrated within the limited space of the SimMan eye socket. The signal from the computer to the screens maintain a minimal time delay of (~0.5 sec). The screens properly displays the correct image in regards to both size and resolution. Developmental Costs: Figure 4: Redness intensity changes after keyboard command. Figure 3: The eye model controlled to look left after input from keyboard. 3D EYE (3D STUDIO MAX) Figure 2: Rendered image of the 3D eye model. Model includes eye lids, pupil, iris and sclera. Figure 1: Wire model of the eye using editable polygons to construct the eye ball and surrounding. The Lilliput screen: Embedded video circuitry. • Size: 1.7”X2” . Resolution: 116,160 (dots). • Cost: $60. Limitations : •FFC cable – limits the separation between board and LCD. •Circuit board size – large due to integrated audio functions. •Backlight - generates heat. Item Quantity Total Cost Lilliput 212GL-18NP 2 $120.00 Ultimate XP Pro: PC to TV 1 $70.00 3D-Studio Max (Trial Version) 1 $0.00 DX-Studio (Trial Version) 1 $0.00 End Mill and Aluminum Plate 1 $37.00 Labor Hours for Cutting 10 $75.00 FINAL PROTOTYPE 1 $302.00 Function Keyboard/Mouse Command Eye Movement Right click mouse and drag Redness Intensity Q = Normal, W = Mild, E = Severe Dilation/Constriction S = Normal, A = Dilation, D = Constriction Blinking X = Normal, Z = Slow, C = Fast Sleep/Wake O = Wake, P = Sleep 1.8” LCD screen 3D Studio Max Symmetrical Video Feed 3 D O b j e c t 3D Real Time Animation Engine VGA Adapter Mapped Keyboard/Mouse C o n t r o l R e n d e r e r Hardware Software ACKNOWLEDGEMENTS A special thanks to Mr. Ray Booker and Dr. Matthew Weinger for providing the original concepts, advice, and knowledge throughout the design process. Ronald Reiserer of VIBRE was an invaluable resource in the dissection of the SimMan head. Trauma care physicians rely heavily on human patient simulators (SimMan) to expose them to real-life medical scenarios. Realism is the most critical aspect of the learning process. However, an ongoing problem with SimMan is its fixed eyes. Physicians depend on various physiological responses of the eye to determine level of consciousness, degree of brain trauma, and responses to administered drugs. SimMan’s stationary eyes take away the realism and effectiveness from the simulation. Therefore, we have developed a cost-effective realistic eye simulator that has the ability to incorporate a range of physiological responses including movement, blinking, pupil dilation, and redness. INTRODUCTION Mechanical eye simulators • Expensive ($600 ~ $2000). • Limited functions: movement and blinking. • High maintenance due to electrical and mechanical malfunctions. Stand-alone eye simulators • Expensive ($2500). • Display anatomical eye features. • Too large for SimMan head. CURRENT EYE SIMULATORS •Size – Has to fit into SimMan’s eye sockets (1’’ diameter) •Response Time Externally controlled with minimal time delay. •Functionality – Speeds of functions must be controlled •Movement (Full Range) •Pupil dilation/constriction (1-10 mm) •Redness (varying intensity) •Blinking (Eyelids) PERFORMANCE CRITERIA We propose that by using micro-display technology we can implement 3D graphics integrated with a real-time rendering engine to model the detailed anatomical features and animate the physiological responses of the eye. OUR SOLUTION PROPOSAL •Create realistic eye features that can be computer controlled. •Integration of 3D computer graphics with micro- displays. •Create a cheaper alternative option for eye GOALS SIMULATION TECHNOLOGIES PROGRAM CELA

Upload: deliz

Post on 24-Feb-2016

43 views

Category:

Documents


0 download

DESCRIPTION

Development of Computer Controlled Eyes for Patient Simulator Silviu . C. Diaconu 1 , Emadaldine A. Elsamadicy 1 , Anas Othman 1 , Azreena A. Shukri 1 Sponsors: Ray Booker 2 , Matthew Weinger , M.D. 2 - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Development of Computer  Controlled Eyes for Patient Simulator

Development of Computer Controlled Eyes for Patient SimulatorSilviu. C. Diaconu1, Emadaldine A. Elsamadicy1, Anas Othman1, Azreena A. Shukri1

Sponsors: Ray Booker2, Matthew Weinger, M.D.2

1 Vanderbilt University School of, Engineering, 2Center for Experiential Learning and Assessment

ACCOMPLISHMENTS

LCD SCREENS (LILIPUT 212GL-18NP)

3D EYE (DX STUDIO)

3D Studio Max is used to model the anatomical structures of the eye, while DX studio was used as the real-time rendering engine for controlling the animations.

OUR EXPERIMENTAL SETUP

Prototype: The anatomical and physiological responses of the eye were successfully modeled and animated such that they can be controlled by an external keyboard with the following controls: full range eye movement, blinking, pupil dilation /constriction, as well as varying degrees of eye redness. The two LCD screens were integrated within the limited space of the SimMan eye socket. The signal from the computer to the screens maintain a minimal time delay of (~0.5 sec). The screens properly displays the correct image in regards to both size and resolution.

Developmental Costs:

Figure 4: Redness intensity changes after keyboard command.

Figure 3: The eye model controlled to look left after input from keyboard.

3D EYE (3D STUDIO MAX)

Figure 2: Rendered image of the 3D eye model. Model includes eye lids, pupil, iris and sclera.

Figure 1: Wire model of the eye using editable polygons to construct the eye ball and surrounding.

The Lilliput screen:• Embedded video circuitry.• Size: 1.7”X2” .• Resolution: 116,160 (dots).• Cost: $60.

Limitations:• FFC cable – limits the separation between board and LCD.•Circuit board size – large due to integrated audio functions.•Backlight - generates heat.

Item Quantity Total CostLilliput 212GL-18NP 2 $120.00

Ultimate XP Pro: PC to TV 1 $70.003D-Studio Max (Trial Version) 1 $0.00

DX-Studio (Trial Version) 1 $0.00End Mill and Aluminum Plate 1 $37.00

Labor Hours for Cutting 10 $75.00FINAL PROTOTYPE 1 $302.00Function Keyboard/Mouse Command

Eye Movement Right click mouse and dragRedness Intensity Q = Normal, W = Mild, E = Severe

Dilation/Constriction S = Normal, A = Dilation, D = ConstrictionBlinking X = Normal, Z = Slow, C = Fast

Sleep/Wake O = Wake, P = Sleep

1.8” LCD screen

3D Studio Max

Symmetrical Video Feed

3D O

bject

3D Real Time Animation

Engine

VGA AdapterMapped Keyboard/Mouse

Control

Renderer

Hardware Software

ACKNOWLEDGEMENTSA special thanks to Mr. Ray Booker and Dr. Matthew Weinger for providing the original concepts, advice, and knowledge throughout the design process. Ronald Reiserer of VIBRE was an invaluable resource in the dissection of the SimMan head.

Trauma care physicians rely heavily on human patient simulators (SimMan) to expose them to real-life medical scenarios. Realism is the most critical aspect of the learning process. However, an ongoing problem with SimMan is its fixed eyes. Physicians depend on various physiological responses of the eye to determine level of consciousness, degree of brain trauma, and responses to administered drugs. SimMan’s stationary eyes take away the realism and effectiveness from the simulation. Therefore, we have developed a cost-effective realistic eye simulator that has the ability to incorporate a range of physiological responses including movement, blinking, pupil dilation, and redness.

INTRODUCTION

Mechanical eye simulators • Expensive ($600 ~ $2000).• Limited functions: movement and blinking.• High maintenance due to electrical and mechanical malfunctions.

Stand-alone eye simulators• Expensive ($2500). • Display anatomical eye features.• Too large for SimMan head.

CURRENT EYE SIMULATORS

•Size – Has to fit into SimMan’s eye sockets (1’’ diameter)•Response Time – Externally controlled with minimal time delay.•Functionality – Speeds of functions must be controlled•Movement (Full Range)•Pupil dilation/constriction (1-10 mm)•Redness (varying intensity)•Blinking (Eyelids)

PERFORMANCE CRITERIA

We propose that by using micro-display technology we can implement 3D graphics integrated with a real-time rendering engine to model the detailed anatomical features and animate the physiological responses of the eye.

OUR SOLUTION PROPOSAL

•Create realistic eye features that can be computer controlled.•Integration of 3D computer graphics with micro-displays.•Create a cheaper alternative option for eye simulators.•Allows for simple adjustments of quality and basic eye functions.

GOALS

SIMULATION TECHNOLOGIES PROGRAMC E L A