development of tantalum/tungsten boated electrothermal...
TRANSCRIPT
Development of Tantalum/Tungsten Boated Electrothermal Vaporization-Flame Atomic
Absorption Spectrometry and its Application for Quantitative Elemental Analysis
byYoung-Soo Cho
DEPARTMENT OF CHEMISTRYGRADUATE SCHOOL
CHANGWON NATIONAL UNIVERSITY
Development of Tantalum/Tungsten Boated Electrothermal Vaporization-Flame Atomic Absorption
Spectrometry and its Application for Quantitative Elemental Analysis
byYoung-Soo Cho
Under the Direction ofProfessor Yong-Ill Lee
A thesis Submitted to the committee of the Graduate School of Changwon National University in
partial Fulfillment of the requirements for the degree of Master of Science
2001. 12.December 2001
Approved by the committee of the Graduate School of
Changwon National University in Partial fulfillment of
the requirements for the degree of Master of Science
Thesis Committee : Chang-Soon Lee
Tae-Jin Won
Yong-Ill Lee
DEPARTMENT OF CHEMISTRY
GRADUATE SCHOOL
CHANGWON NATIONAL UNIVERSITY
- iv -
TABLE OF CONTENTS
LIST OF TABLES ········································································································· v
LIST OF FIGURES ······································································································ vi
1 INTRODUCTION ······································································································· 1
2 EXPERIMENTAL ······································································································· 9
2.1 ETV Chamber and Y-type connector ··················································· 9
2.2 Instrumentation ······························································································ 10
2.3 Reagents ············································································································ 11
2.3.1. Standard solution ················································································· 11
2.3.2. Solid sample ··························································································· 11
2.4 Operating procedures ················································································· 12
3 RESULTS AND DISCUSSION ········································································ 14
3.1 Optimization of the Parameters ···························································· 14
3.1.1 Optimization of ETV current ·························································· 14
3.1.2 Effect of argon carrier gas flow rate ······································· 15
3.2 Optimization for analysis of solid sample ······································· 16
3.2.1 The characterization of ETV current ········································ 17
3.2.2 Argon carrier gas flow rate efficiency ····································· 17
3.3 Analytical Performance ·············································································· 18
3.4 Application of Standard addition method for solid
materials of NIST, NIES ············································································ 19
4 CONCLUSION ·········································································································· 22
LITERATURE CITED ······························································································· 23
ABSTRACT ··················································································································· 26
ACKNOWLEDGMENTS ···························································································· 60
- v -
LIST OF TABLES
Table 1. Instrumental condition of Flame Atomic Absorption
Spectrometry(A) and Electrothermal vaporization(B).
(Cu, Cd, Mn, Pb, and Zn with tantalum, and Co, Ni
with tungsten). ························································································ 28
Table 2. Physical properties of Mn, Cu, Zn, Cd , Pb, Co,
and Ni. ········································································································ 30
Table 3. Optimized ETV current of Mn, Cu, Zn, Cd , Pb,
Co, and Ni. ······························································································· 31
Table 4. Detection limits (3σblank, n=5), coefficient value,
and precision for Zinc, Cobalt, and Nickel of
ETV-FAAS. ······························································································· 32
Table 5. The reference concentration values of (A) NIST
and(B) NIES samples. ········································································· 33
Table 6. The analytical results of Mn, Cu, Zn, Cd and Pb
in food, biological and environmental samples with
linearity and relative standard deviation value in
percentage. ······························································································· 34
- vi -
LIST OF FIGURES
Fig. 1 Energy level diagrams for (a) absorption and (b) emission.
······························································································································ 2
Fig. 2 The scheme of electrothermal vaporization flame atomic
absorption spectrometry used in this study(ETV-FAAS).
···························································································································· 35
Fig. 3 The lighten filament of (A) Tantalum, and (B) Tungsten
at the vaporizing current. ····································································· 36
Fig. 4 The increasement of peak absorbance of Cu(A), Cd(B),
Mn(C), Pb(D), Zn(E) with tantalum (10ppm, 5μL) and Co(F),
Ni(G) with tungsten (10ppm, 3μL) according to the drying
and vaporizing current change. ······················································· 37
Fig. 5 The efficiency of argon carrier gas flow rate for ETV-
FAAS. ············································································································· 44
Fig. 6 Mg absorbance characteristic of ETV current in RM 8432
(Corn starch) ······························································································· 45
Fig. 7 The efficiency of argon carrier gas flow rate of Mg
in corn starch. ···························································································· 46
Fig. 8 Absorbance spectra and calibration curves of the element
(A) Zn, (B) Co, and (C) Ni. ································································· 47
Fig. 9 Absorbance spectra and calibration curves of the element
(A)Zn in RM 8414 and RM 8415, (B) Mn in RM 8415 and
RM 8418, and (C) Cu in RM 8414 and RM 8415 by
standard addition method. ···································································· 50
Fig. 10 Absorbance spectra and calibration curves of the element
(A) Mn and (B) Zn in NIES No. 3 by standard addition
method. ······································································································· 56
Fig. 11 Absorbance spectra and calibration curves of the element
(A) Pb and (B) Cd in NIES No. 8 and No. 10 by standard
addition method. ···················································································· 58
- 1 -
1 INTRODUCTION
Solids analysis utilizing a system to detect atomic
absorption or emission line continues to be a significant area
of research. A number of instrumental configurations were
used for the introduction of either slurried or solid samples
into flames, graphite furnaces and plasmas. In the absence of
external radiation, some analytes can be stimulated by
collisional processes, or by electrical or chemical energy,
and can emit photons when excited species return to
lower-energy states. The absorption and emission, which
are the most fundamental sections of many spectroscopic
phenomena including chemiluminescence, photoluminescence,
reflection and scattering, occur when external
electromagnetic radiation impinges upon a collection of
analyte species (atom, molecules, or ions) in a sample.
Absorption and emission process are illustrated in fig.
1-(a) and (b), respectively. The horizontal line labeled E0
corresponds to the lowest, or ground-state, energy and the
others labeled E1 and E2 are two higher-energy electronic
levels of the atom. Absorption of the incident photons by the
analyte, which is shown in fig. 1-(a), promotes the analyte
to an excited state. There are allowed transitions and
forbidden transitions. In the process are also transitions
between excited states (non resonant) but these are not
used in atomic absorption, thus restricting the number of
transitions (wavelengths in the ultraviolet and visible range)
that can be used for each element. Different atoms need
- 2 -
different energies to be excited. This energy can be supplied
in many forms like flame and heating by electric current.
In fig. 1-(b), the promoted atom returns to the ground
state, emitting a photon whose frequency and wavelength.
The lights are often used as an emission source like plasma.
(a) (b)
E0 E0
E1
E2
E1
E2
(a) (b)
E0 E0
E1
E2
E1
E2
Fig. 1 Energy level diagrams for (a) absorption and (b)
emission.
In atomic absorption, the radiation is measured before
and after absorption and the amount absorbed is proportional
to the concentration of the analyte. Beer-Lambert law (a
combination of Beer's law and Lambert's law) mathematically
described the absorbance of light passing through a sample
solution (in atomic absorption-the atom population in the
flame) as a function of the length of the optical path through
the sample (length of the flame) and the concentration of the
absorbing species (ground state atoms).
I t= I 0 eεlc
- 3 -
A= log I 0 / I t=εlc
,where A is the absorbance (or optical density), I0 the
incident radiation power, It the transmitted radiation power, ε
is absorptivity (absorption coefficient at the wavelength used
for the analysis), l the length of the absorption path, and c
the concentration of absorbing atoms. This equation means
that when analyzing the same type of atom (e.g. Cu) in
samples of unknown concentrations and standard solutions of
know concentration, where the absorptivity remains the same
and the absorption path (length of flame) remains the same,
the absorbance will be a linear function of the Cu
concentration. It should be linear but it is not always linear
throughout all the range of concentration. Hence, one cannot
rely upon mathematical calculations and analytical
measurements are made using calibration curves (though
work is essentially limited to the range where the calibration
lines are not too curved). There is no need to know the
values of ε and l as atomic absorption is a comparative
technique. One measures a set of standards of known
concentrations, prepares an absorbance vs. concentration
line and for each absorbance reading of samples with
unknown concentrations, finds the respective concentration
(in modern spectrometers this is done by the instrument's
computer). Atomic absorption spectrometers read the amount
of incident light without the sample and with the sample
(after absorption), compare them, and show the results in
absorbance units.[1]
Atomic absorption spectrometry (AAS) has been the most
- 4 -
widely used method which is analytical technique for the
determination of minor, trace, and ultratrace metals in a
widely variety of complex samples such as environmental,
geological, and biological samples. AAS has also become
used in the analytical laboratory due to their many desirable
factors of low detection limits in the range of a few ppm,
ppb or less (atomization by flame and electrothermal
atomization), high sensitivity and selectivity, widespread
availability, modest cost per sample after initial set-up,
simultaneous or sequential multi-elemental determination,
good precision and accuracy, and ease of operation in an
analysis. The trace metal analysis can be carried out in the
presence of many other elements, usually, having an
advantage which makes the process simpler and saves a lot
of time and errors.
However, limiting factor in atomic spectroscopic
techniques, being universally accepted as an absolute method
of metal determination, is a sample introduction process and
system. A greater awareness for the development of sample
introduction process and technique to analyze various
samples has led to a subsequent increase in research
activity in this area. Ideally, the sample introduction system
for atomic spectroscopy would reproducibly and efficiently
transfer the sample to the atomization or excitation stage. It
should produce no interferences, be reproducible and
independent of the sample type, be universal for all atomic
spectroscopic techniques and have no memory or carry-over
effect.
- 5 -
Electrothermal vaporization (ETV) is relatively new and
has proved to be a very useful technique for sample
introduction in various atomic spectrometry techniques
because of the higher transport efficiency of dry aerosols to
the atomizer, micro sampling capability, and the potential for
direct solid analysis.[2-3] The most common substrate used
for ETV is pyrolitically coated graphite. In a typical ETV
experiment, the sample is deposited on a graphite surface;
the process is consisted of drying, pyrolyzing, and
vaporizing. The gas-phase analyte is the interrogated by
atomic absorption spectrometry or the vaporized species are
transferred into the flame for the subsequent AAS detection.
The sample is always vaporized from the graphite surface,
and the interaction of various analytes with the graphite
substrate contributes to the sensitivity, the detection limits,
and the shape of the analytical signal.[4-5]
There were attempts to introduce matrix modifiers
which are included metallic materials for stabilizing high
temperature platform furnace, preventing analytes from
volatilizing, and increasing the volatility of sample matrix
.[6-12] However, it noted that chemical modifier should be
highly pure, and not caused spectral or chemical
interference.
In addition, refractory metal filaments such as
tungsten[13-15] or tantalum[16] had been used as furnace
materials instead of pyrolitic coated graphite. A boat-shaped
tantalum or tungsten filament vaporizer normally
manufactured belong to this group and have been applied as
- 6 -
sample introduction part to AAS. The attraction of these
filaments is to be seen in their relatively low price, high
reproducibility, and physical property such as electrical
resistant due to high temperature. They form a simple,
inexpensive and efficient ETV device mounted in a small
pyrex glass apparatus, and sample volumes of up to 50 μL
can be loaded onto the boat with glass syringe.
Sample preparation is one of the critical factors
determining the quality of the analysis results. The analytes
were already largely dissolved in water, and required
minimum treatment for the subsequent stages of the chemical
analysis. If the analysis of solids are included in extended
scope, the extraction of the species of interest should have
to be considered.[17] In solid sample analysis, sample
preparation is often the most time consuming step and is
also considered involves some potential problems, such as
incomplete dissolution, precipitation of insoluble analyte and
loss of some elements during the heating. There has been
interest in and proposals for the analysis of ultratrace
metals in food, biological, and environmental
samples.[18-21] These kinds of solid samples are mostly
pretreated by concentrated acid which are HNO3, HF, HClO4,
HCl and H2O2, which are considered of the risk of sample
contamination. Recently microwave-assisted sample
dissolution has been employed extensively for shortening the
time required for sample dissolution, as well as to avoid
analyte losses and contamination. However, this commercial
microwave oven for analytical tasks has some main
- 7 -
drawbacks which are expensive, the short lifetime of the
digestion vessels operated under high pressures and
temperatures, the acid content left after the digestion, and
sample throughput is not very high.[22] The ideal method
for the analysis of solid samples is to eliminate the sample
dissolution, minimize the sample preparation and improve the
analytical results. Slurry sampling has been extensively used
for the analysis of solids[23-30] in order to simplify sample
preparation and to avoid some problems associated with
dissolution procedures. The success of slurry sampling
depends on some variables, such as the variance of the
sampled analyte mass, particle numbers and sizes[31]
present in the injected volumes, analyte homogeneity,
suspension medium, slurry concentration, stirring method and
sampling depth.
Calibration, standard addition, and internal standard
method have been tried for the determination of trace
element in solid samples, but standard addition method is
applied much more than others.
Standard addition methods are particularly useful for
analyzing complex samples in which the likelihood of matrix
effects is substantial. A standard addition method can take
several forms. One of the most common forms involves
adding one or more increments of a standard solution to
sample aliquots of the same size. This process is often
called spiking the sample. Each solution is then diluted to a
fixed volume before measurement. It should be noted that
when the amount of sample is limited, standard additions can
- 8 -
be carried out by successive introductions of increments of
the standard to a single measured volume of the unknown.
Measurements are made on the original sample and on the
sample plus the standard after each addition. In most
versions of the standard addition method, the sample matrix
is nearly identical after each addition, the only difference
being the concentration of the analyte or, in cases involving
the addition of an excess of an analytical reagent, the
concentration of the reagent. All other constituents of the
reaction mixture should be identical because the standards
are prepared in aliquots of the sample.[32]
The present work introduces a development of a tantalum
and tungsten filamented-ETV system connected to the
conventional acetylene/air FAAS system. Also it describes
evaluation of the performance of the system for the direct
determination of trace metal ions including linearity,
precision and detection limit. Standard addition method was
used for evaluating the quantitative capability of ETV on the
determination of Cu, Cd, Mn, Pb, and Zn in standard food
and environmental samples, and the analytical results were
compared to the certified reference values.
- 9 -
2 EXPERIMENTAL
2.1 ETV Chamber and Y-type connector
Electrothermal vaporization assembly (ETV) was
respectively designed and manufactured of pyrex glass
chamber. The improved vaporizer unit (78.7 mm long) is
contained within an argon-filled enclosure. Tantalum and
tungsten boats which are fabricated from 0.025 and 0.05
mm-thick foil (Aldrich Chemical Co., USA) are mounted on
copper electrode terminals supported by a cylindrical pyrex
glass base (30 mm o.d) with a tangential gas inlet port and a
joint. This assembly is surrounded by a cylindrical pyrex
glass manifold. The conical top of the manifold contains two
ports. One port allows delivery of the sample to the tantalum
boat vaporizer surface, while the other permits the vaporized
sample to be swept by the injector gas into the flame.
The Y-type tube mentioned of transport efficiency in the
literature[16] was equipped between the flame and ETV
chamber for the connector. The right branch was connected
to the top of the ETV chamber, and the left to the auxiliary
oxidant gas inlet hose. Vaporized sample was mixed with the
auxiliary oxidant gas, and flew to the flame, then. To
decrease analyte loss in the transport process, the flexible
silicon tube was used as the transport tubing from the ETV
to the flame through Y-type connector. Using this design,
the sample vapor was directly penetrated to the transport
tube from vaporization chamber and then swept into the
flame by the stream of argon.
- 10 -
2.2 Instrumentation
The ETV-FAAS shown in fig. 2 was constructed around
the following components, and the instrumental conditions
were shown in table 1. The system can be distinguished to
three parts, sample introduction, Y-type connector, and the
spectrometry. When the liquid or slurry samples loaded on
the metal filament were desolvated and vaporized in the
ETV, they were mixed with argon carrier gas. With auxiliary
gas, vaporized mixtures flew through Y-tube into the flame
and were analyzed. The atomic absorption measurements
were carried out using a Perkin Elmer 3300 atomic
absorption spectrometry (Perkin Elmer Co., Norwalk, CT)
with acetylene-air flame and a special hollow cathode lamp
for each metal element. Before all measurements were
performed, background correction of the hollow cathode lamp
was made with the spectrometry itself.
Tantalum (17×8mm) and tungsten (17×5mm) filament
were heated by supplying the electric current through
copper electrodes (88.5mm in length) by dc power supply
operated in constant voltage mode (max 50 V and 100 A,
Korea Switching Co., Seoul, Korea). The operating current is
supplied in step-mode which is changed automatically from
low to high value, and is optimized with respect to the
maximum peak height using the aqueous standard solutions.
Typical operating currents used in the experiments are 12~
20 A (about 170~480℃) up to 32 s for desolvation and 36~
68 A (about 1600~3500℃) up to 10 s for vaporization,
depending on the analyte element and sample matrix. The
- 11 -
current was controlled with a programmable current-timer
unit system designed in our laboratory.
The flow of argon carrier gas remained spiral in the
vaporizer chamber by forcing argon to flow tangentially
around the wall of the glass cone. The tangential flow cools
the glass wall and centers the vaporized samples with the
argon flow that remains spiral in the chamber. The argon
carrier gas flow rate was controlled with a Dwyer flow
meter and was typically 0.5~2.5 liters/min.
2.3 Reagents
2.3.1. Standard solution
Stock standard solutions of various elements (Cu, Cd,
Mn, Pb, Zn, Co, and Ni) of 1000 mg/L were prepared from
Aldrich atomic absorption standards.
Working standard solutions were prepared daily by
appropriate dilution of the respective stock standard
solutions. All solutions were prepared using high purity
deionized water (resistivity 18.2 MΩcm-1) from Milli-Q water
purification system (Millipore Corp., Molsheim, France).
2.3.2. Solid sample
The applicability of the method to real samples was
demonstrated by the analysis of the National Institute of
Standards & Technology (NIST, U.S.A) and the National
Institute for Environmental Studies (NIES, Japan) which are
RM 8414 (bovine muscle), 8415 (whole egg powder), 8418
(wheat gluten), and 8432 (corn starch) of NIST and No. 3
- 12 -
(chlorella), No. 8 (vehicle exhaust particulates), and No.
10-c (flourished rice) of NIES.
The powder samples were weight with microbalance and
were diluted with deionized water. They were warm up in
water for 1~2 hr with stirred, then the fine particles turned
to the slurry state. For the analysis of Mn and Zn in RM
8418 and NIES No. 3, the solid samples were pretreated
with acid digestion method because they were not dissolved
in deionized water at all. Nitric acid, hydrogen peroxide, and
chloric acid of analytical grade were added with 1~3 mL,
and the samples were digested in microwave oven with high
pressure and temperate. No. 8 and No. 10-c were used to
analysis directly of Pb and Cd. Cd for rice sample was
mixed with 50% (w/w) with standard solution of known
concentration, but Pb for vehicle sample was mixed with
0.01% (w/w) because of a high concentration of analyte.
These samples were dried in vacuum oven for 2 days at 5
0℃. Dried solid samples were grinded for their homogeneity.
The slurry sample was prepared with diluted ratio by
standard addition method which is known of the
concentration of added standard solution. At least the
sampling volume was 0.3 mL with micropipette which of the
volume was controlled digitally.
2.4 Operating procedures
For the manipulation of ETV-FAAS, the whole procedure
is as follows. Before the micro volume samples were loaded,
the metal filament was cleaned at the same current for
- 13 -
vaporizing with argon continuous flow. 3~10 microliter
standard solution samples loaded onto the filament through
the sample injection port using a Hamilton syringe were
initially desolvated by passage of approximately 16~20 A
current through the filament for 32 s. In the case of the
slurry sample prepared for a direct analysis of solid, 2~10
microliter was loaded onto the filament. All samples were
dried at 16 A for 32 s. After the drying step of the slurry
sample, dark brown ash were remained on the filament. At
this moment, the white smoke was found to fly through the
top of chamber. The argon carrier gas was about 1.5
liters/min rate during the desolvation and ashing. Final
vaporization of both aqueous standards and the slurry sample
was carried out by instantaneously raising the current to 52
A. At this time, adequate argon flow rate was preserved at
about 1.5 liters/min. The gaseous samples were flew into the
flame on the burner head mixed with the argon and
acetylene as auxiliary gas in the state of turning on flame,
and analyzed by FAAS.
- 14 -
3 RESULTS AND DISCUSSION
3.1 Optimization of the Parameters
Key operating parameters for ETV-FAAS, such as ETV
current and argon carrier gas flow rate were investigated for
optimization studies. As the vaporization and the excitation
of the analytes realized in two separate steps, the ETV
parameters were optimized successfully. For the most of the
sample in the state of solution and slurry, it found that 3~
10 microliter of stock solution and 2~10 microliter of solid
sample were adequate for these studies.
3.1.1 Optimization of ETV current
In principles, the volatilization of the sample was
performed by the resistive heating of the tantalum filament.
The filament was heated by supplying the electric current
through copper electrodes by dc power supply operated in
the constant voltage mode. The operating current was
optimized with respect to the maximum peak height for
various elements of the aqueous standard solution according
to the different volatility of elements. The physical and
chemical properties of the elements are shown in Table 2.
To optimize the ETV current, argon gas flow rate was set at
1.5 liter/min. The drying step was set at current of 12, 16 A
for tantalum boat and 16, 20 A for tungsten filament with the
holding time of 32 s. The vaporizing step was changed from
36 A to 68 A increasingly.
Fig. 3 shows the filament of tantalum and tungsten
- 15 -
lightening. Their currents were maximized at 52 and 64 A
respectively. The light of tantalum filament are much
brighter than that of tungsten.
The optimized ETV current is presented in Fig. 4. It is
the typical peak height variation of Cu, Cd, Mn, Pb, Zn, Co,
and Ni responses as a function of ETV current with 4 s
holding time. The variation of ETV currents from 36 A to 56
A by tantalum and from 56 A to 68 A by tungsten were
investigated and the all elements absorption signal reached
their maximum at 52 A and 64 A. The absorbance peak of
Co and Ni was not shown using tantalum boat but tungsten
because of the difference of electrical resistivity between
the metal filaments. Therefore, it can be guessed that the
temperature of the tungsten is higher than that of the
filament.
The effect of ETV current on the background signal was
evaluated because of depending on the amount of tantalum
and tungsten ablated from the filament during vaporization,
and the background intensity was little observed on
increasing the current.
From this evaluation, seven elements were chosen and
the optimized currents were determined from seven
elements. The optimized currents for these seven elements
are listed in Table 3.
3.1.2 Effect of argon carrier gas flow rate
The argon carrier gas flow rates of Co and Ni elements,
which were introduced to the ETV unit to carry the
- 16 -
vaporized sample into the flame, must be one of the most
important parameters in ETV-FAAS because of the transport
efficiency. The argon carrier gas flow rate from 1.0 to 2.5
liter/min was used. To optimize the argon carrier gas flow
rate, ETV current was set at 64 A for 4 s.
Fig. 5 shows the effect of the argon carrier gas flow
rate in a range of 1.0 to 2.5 liter/min on the peak height in
due consideration of background for Co and Ni loaded onto
tungsten filament (3 ㎕ of 10 μg/mL). Each element
absorption signal initiated with increasing of argon carrier
gas flow rate, reaches its maximum at 2.0 liter/min due to
the transport efficiency and decreases at higher flow rate.
For later investigations, argon carrier gas flow rate was held
constant at 2.0 liter/min for Co and Ni. Though higher gas
flow rates above 1.5 liters/min introduce more analyte into
the flame, the peak height was actually decreased in the
high flow rate by reducing the residence time of the analyte
in flame. In contrast, at low flow rates below 1.0 liters/min,
the signal peak becomes lower and broader in profile
because of increase diffusion of analyte in carrier gas.
3.2 Optimization for analysis of solid sample
To analyze trace elements in solid sample,
characterization must be performed about ETV current and
argon gas flow rate which give an effect for maximum
absorption signal intensity. These parameters were optimized
and 5μL of the slurry sample was tested in this study.
- 17 -
3.2.1 The characterization of ETV current
Corn starch was soluble in deionized water and was
warm up in water bath for 2 hr with stirring. 5 μL of slurry
corn starch sample was loaded onto tantalum boat-shape
filament by using cylinder. Before applying the ETV system
for solid analysis, it has been guessed that the sample like
corn starch can include complex matrix. Therefore, the study
of ETV current was performed varying vaporization current
from 32 A to 40 A.
Fig. 6 shows the Mg characteristic absorbance according
to the different ETV current. Slurry sample loaded on the
tantalum filament was not only desolvated but also ashed at
12 and 16 A for 32 s. Dried solid sample was dark brown
colored and all were transported to the flame at each
vaporization current. Mg element, which was performed at
285.2 nm as high concentration in corn starch, showed a
little bit higher absorbance at 52 A though the values were
similar.
3.2.2 Argon carrier gas flow rate efficiency
This carrier gas are employed for preventing the
filament oxidation and transporting the gaseous sample
efficiently which makes it possible that the peak broadening
can be avoided and the loss of sample also can be reduced.
That is to say, this parameter for highly efficient sample
transportation is very important.
Fig. 7 shows the effect of carrier gas flow rate for Mg
in corn starch slurry sample. The flow rate was changed
- 18 -
from 0.5 to 2.5 liter/min, and the sample volume was needed
about 5 μL repeating the drying (32 s) and vaporizing (4 s)
step. The Mg absorbance increased linearly and showed the
maximum flow rate at 1.5 liter/min. The average precision of
4.1 % RSD (relative standard deviation) was calculated from
five consecutive measurements of the peak height in this
experiment and the improved precision was obtained over
1.5 liter/min.
3.3 Analytical Performance
Under the selected operating conditions of ETV current
and argon carrier gas flow rate, linear calibration graphs
were obtained by using a constant volume. Fig. 8 presents
the absorbance spectra and the calibration graphs of Zn, Co,
and Ni with a volume of 10 and 3 μL in ranges of 5~100 ng
for Zn, 1.25~30 ng for Co, and 2.5~30 ng for Ni. The other
elements except Zn were tested by drawing calibration curve
in the literature [16], and cobalt and nickel which have
higher boiling points were employed to test the efficiency of
new filament tungsten in this study.
The numerical values were listed in Table 4 with a
coefficient linearity (r), reproducibility for five consecutive
firings yielded a %RSD, and the detection limits in the form
of absolute weight, ng. The standard deviation was
calculated from five consecutive measurements of the
absorbance using the peak height of each element. The
relatively good precision and lower detection limits may be
attributed to the stability of the whole system developed
- 19 -
which enabled efficient transport of the analyte aerosol,
optimization of gas flow rate, and proper temperature
programming for vaporization of analyte. The detection limits
were defined as the concentration of analyte that produces a
signal three times the standard deviation of the blank signal.
The blank signal was measured at the analytial line of the
element used in this work.
3.4 Application of Standard addition method for solid
materials of NIST, NIES
As mentioned in the experimental section, standard solid
materials were diluted with deionized water and were also
warm up in water bath. Only RM 8418 (Wheat gluten) was
pretreated by acid digestion method because this material
was not soluble in water 2~3 mLof the concentrated pure
acid such as mixed fluoric acid, chloric acid, hydrogen
peroxide and nitric acid may be used, and high pressure was
provided with polymer vessel in microwave oven for one and
a half hour. After digestion of the solid sample with acid,
the vessel should be cooled and the pressure in the vessel
should be drained. This is a tedious procedure taking very
long time for sample pretreatment.
With prepared sample through the upper method of solid
treatment, the quantitative analysis was performed with
standard addition method. The standard addition method has
been widely applied for the purpose of reducing matrix
effect which induces decrease of absorbance signal. It is
difficult to match sample complex matrix exactly if
- 20 -
calibration curve is used as a method for analysis of
unknown sample. However, the standard addition method can
be solved this problem because of adding the
concentration-known standard solutions, in the order of
increasing solution volume of the same concentration or
keeping different concentration of the same volume. When
the sample and standard solution are mixed, the complex
matrix are matched well.
In Table 5, the solid samples used in this study are
listed which are used in this study with the certificated
values. Most elements are copper, manganese, and zinc. Mg
in RM 8432 (corn starch) that was used for making sure of
the analysis ability by standard addition method. The solids
from RM 8414 to 8418 are food and biological samples which
are bovine muscle powder, whole egg powder, and wheat
gluten. NIES series were used as standard food and
environmental sample flourished rice, chlorella and vehicle
exhaust particulates.
The calibration curve and absorbance peak are shown in
fig. 9 and fig. 10 which are distinguished to NIST and NIES
samples. Blank test was performed for all elements to
determine absorbance degree by blank solvent, so that the
blank absorbance peak was removed from all spectra. The
calibration curve and linear-fitted line were graphed by
origin program, and the analysis value was calculated from
the graph considering dilution ratio. Following equation was
constructed for the calculation of unknown concentration of
solid samples.
- 21 -
Analytical results = (A- x)B
×(S+W)S
, where A is the y-axial absorbance obtained from
calibration curve, x is the absorbance of blank test, B is
equal to the slope magnitude, and S, W are the weight of
solid sample and deionized water.
The analysis of Pb, Co, Ni and Cd in NIES No. 8 and No.
10-c was applied with solid samples directly which are not
diluted or digested. Rice sample in complex matrix turned
into the dark brown ashes at drying current, 16A. Fig. 11
shows the absorbance peak and the calibration curves for Pb
and Cd. Tantalum-boat made it possible to analysis of solid
samples because the shape prevented them from being
dispersed. However, it was impossible to the analysis of Co
and Ni in NIES No. 8 as tungsten filament was not so
concave but flat that the powder was dispersed.
Table 6 summarizes the analysis of each element for
accuracy of the elemental determination. The accuracy of
the elemental determinations is evaluated by their standard
deviation from the reference value of the certified samples.
It follows from the results that the ETV system in the
present work has high accuracy and precision for the
quantitative analysis of Cu, Mn, and Zn in biological samples
of NIST , and Cd, Pb in NIES series as an environmental
solid.
- 22 -
4 CONCLUSION
Through the development of ETV-FAAS, an improved
ETV system using a boat and thin layer tantalum and
tungsten filament as vaporizer was successfully optimized to
give a high degree of reliability and flexibility. Argon carrier
gas of 1.5 and 2.0 liter/min, drying current of 16 and 20 A,
and vaporizing current of 40~52 and 64 A for copper,
cadmium, lead, manganese, zinc, cobalt, and nickel with
tantalum and tungsten, respectively, were achieved as
optimized conditions used for this system.
Not only the good detection limits for zinc, cobalt, and
nickel were obtained and adequate, and but also was
precision observed with this technique for its direct coupling
between ETV device and flame on burner head. That is,
detection limits in the range of 0.16 to 1.48 ng and good
precision of 3.19~3.87 %RSD for zinc, cobalt, and nickel
were obtained.
For a direct analysis of powder-type solid, the slurry
samples diluted with deionized water and undiluted powder
samples were prepared. They were mixed with a constant
volume standard solution to reduce matrix effect, and could
be analyzed in the ETV-FAAS system. The analysis results
were satisfied to the certificated values compared the
reference. The average relative standard deviation was 2.8~
5.5%RSD, and linearity coefficient showed highly good result.
- 23 -
LITERATURE CITED
[1] Determination of Trace Elements, ed. by Z. B. Alfassi.-
Rehovot (Israel): Balaban Publ.; Weinheim; New York; Basel;
Cambridge; Tokyo: VCH, 1994, p. 154-155.
[2] Z. Szitanyi and C. Nemes, Microchemical Journal, 54,
246(1996).
[3] G. Galbacs and F. Vanhaecke, Microchemical Journal, 54,
272(1996).
[4] E. A. Piperaki, N. S. homaidis, and I. Denis, J. Anal. At.
Spectro., 14, 1901(1999).
[5] J. Murphy, G. Schlemmer, I. L. Shuttler, P. Jones, and S.
J. Hill, J. Anal. At. Spectro., 14, 1593(1999).
[6] L. L. Shuttler, M. Feuerstein and G. Schlemmer, J. Anal.
At. Spectro., 7, 1299(1992).
[7] S. Sachsenberg, T. Klenke, W. E. Krumbein, H. J.
Schellnhuber and E. Zeeck, Analytica Chimica Acta, 279,
241(1993).
[8] E. Bulska and W. Jedral, J. Anal. At. Spectro., 10,
49(1995).
[9] H. Li, B. Gong and K. Matsumoto, Anal. Chem., 68,
2277(1996).
[10] S. Imai, N. Hasegawa, Y. Nishiyama, Y. Hayashi, and K.
Saito, J. Anal. At. Spectro., 11, 601(1996).
[11] P. B. Barrera, M. C. B. Alonso, J. M. Piñeiro, C. G.
Sixto, and A. B. Barrera, Spectrochimica Acta B, 51,
1235(1996).
[12] M. A. Alvarez, N. Carrión, and H. Gutiérrez,
- 24 -
Spectrochimica Acta B, 51, 1121(1996).
[13] T. ernohorský and S. Kotrlý, J. Anal. At. Spectro., 10,
155(1995).
[14] I. Marawi, L. K. Olson, J. S. Wang, and J. A. Caruso, J.
Anal. At. Spectro., 10, 7(1995).
[15] Y. Zhou, P. J. Parsons, K. M. Aldous, P. Brockman, and
W. Slavin, J. Anal. At. Spectro., 16(2001).
[16] Y. I. Lee, J. K. Kim, K. H. Kim, Y. J. Yoo, G. H. Back,
and S. C. Lee, Microchemical Journal, 60, 231(1998).
[17] Environmental Analysis, ed. by J. D. Barnes; JOHN
WILEY & SONS Publ.; Chichester; New York; 1994, p. 141.
[18] A. Marco, Z. Arruda, M. Gallego, and M. Valcárcel, J.
Anal. At. Spectro., 10, 55(1995).
[19] F. V. Haecke, S. Boonen, L. Moens, and R. Dams, J.
Anal. At. Spectro., 10, 81(1995).
[20] D. D. T. Phuong, E. Tatár, I. Varga, G. Záray, E. Cseh,
and F. Fodor, Microchemical Journal, 51, 145(1995).
[21] N. J. Miller-Ihli, J. Agric. Food Chem., 44, 2675(1996).
[22] É. C. Lima, F. Barbosa Jr., F. J. Krug, M. M. Silva, and
M. G. R. Vale, J. Anal. At. Spectro., 15, 995(2000).
[23] K. C. Friese, V. Krivan, and O. Schuierer,
Spectrochimica Acta part B, 51, 1223(1996).
[24] U. Schäffer, and V. Krivan, Spectrochimica Acta part B,
51, 1211(1996).
[25] Y. E. Freedman, D. Ronen, and G. L. Long, Environ. Sci.
Technol., 30, 2270(1996).
[26] C. C. Vique, P. L. Teissedre, M. T. Cabanis, and J. C.
Cabanis, J. Agric. Food Chem., 45, 1808(1997).
- 25 -
[27] M. Resano, M. Verstraete, F. Vanahaecke, L. Moens, A.
V. Alphen, and E. R. Denoyer, J. Anal. At. Spectro., 15,
389(2000).
[28] P. Viñas, M. P. Martínez, and M. H. Córdoba, J. Agric.
Food Chem., 48, 5781(2000).
[29] P. B. Barrera, T. P. Alvite, M. C. B. Aonso, A. B.
Barrera, J. A. Cocho, and J. M. F. Bermúdez, J. Anal. At.
Spectro., 15, 435(2000).
[30] C. Santos, F. A. Moreno, I. Lavilla, and C. Bendicho, J.
Anal. At. Spectro., 15, 987(2000).
[31] Q. Y. Chao, J. Z. Cheng, Z. Yune, and H. Bin, J. Anal.
At. Spectro. , 10, 455(1995).
[32] Principles of Instrumental Analysis (fifth edition), ed.
Douglas A. Skoog, F. James Holler, and Timothy A. Nieman;
Saunders Golden Sunburst Series; Philadelphia; 1998, p.
15-16.
- 26 -
ABSTRACT
Development of Tantalum/Tungsten Boated
Electrothermal Vaporization-Flame Atomic
Absorption Spectrometry and its Application
for Quantitative Elemental Analysis
by Young-Soo Cho
Department of Chemistry
Graduate School, Changwon National University
매트릭스가 복잡한 시료 속에 존재하는 미량의 중금속을 보다
쉽고 정확하게 분석하기 위한 방법으로서 전열증기-원자흡수분광기
가 널리 사용되어지고 있다. 전열증기화 장치는 원자분광분석기기의
시료도입장치로써 원자화장치로의 높은 시료전송효율과 미량시료에
대한 분석 가능성, 고체시료의 직접 분석에 대한 가능성을 가지고 있
으며, 특히, 시료에 포함되어 있는 용매에 대해 전원공급장치에서 제
공되는 전류세기를 조절하여 선택적인 탈용매화가 가능하므로 원자흡
수스펙트럼을 얻는데 있어서 방해가 되는 요인들을 잠재적으로 제거
할 수 있는 장점을 가지고 있다.
본 연구에서는 탄탈륨과 텅스텐을 이용한 전열증발장치를 직접
제작하여 미량원소를 분석하는데 있어서 최적조건을 얻기 위해 시료
를 전송하는 아르곤 기체의 최적유속과 각각의 금속원소들이 가지는
- 27 -
물리적 성질(녹는점과 끓는점)에 기초하여 시료의 탈용매화와 증기화
에 필요한 전류에 대해서 특성연구를 수행하 다. 이 실험을 통해 얻
은 최적 조건 하에서 미량 중금속 (Cu, Cd, Mn, Pb, Zn, Co, Ni)에
대한 검정곡선(Calibration curve)을 작성하여 선형성(Linearity)과
정 도(Precision), 검출한계(Detection limits)를 구할 수 있었다.
3∼10μL의 금속표준용액을 이용하여 Zn(5∼100 ng),
Co(1.25∼30 ng), Ni(2.5∼30 ng)의 범위 내에서 검정곡선을 작성
한 결과, 직선을 의미하는 1의 값에 가까운 선형계수값을 나타냄으로
써 선형성이 매우 좋고, 실험결과들을 통계적으로 처리하여 상대표준
편차를 백분율로 나타내었을 때 4% 이내임을 보 으며, 절대검출한
계가 각각 0.16, 1.48, 0.93 ng 임을 알 수 있었다.
표준물 첨가법에 의해 제조된 고체시료를 실제 정량분석에 적
용한 결과, 고체시료 속에 존재하는 Cu, Cd, Mn, Zn, Pb에 대한 결
과들이 모두 표준시료를 제조하는 기관에서 제공한 기준분석수치 내
에 포함됨을 확인할 수 있었다.
위의 최적화 연구를 통하여 전열증발장치를 이용한 불꽃원자흡
수분광법이 실제 매트릭스가 복잡한 고체시료의 분석에 적합하다는
것을 확인하 다.
- 28 -
Table 1. Instrumental condition of Flame Atomic Absorption
Spectrometry (A) and Electrothermal vaporization (B). (Cu,
Cd, Mn, Pb, and Zn with tantalum, and Co, Ni with tungsten).
(A)
Element Cu Cd Mn Pb Zn
Wavelength (nm) 324.8 228.8 279.5 283.3 213.9
Slit width (nm) 0.7 0.7 0.2 0.7 0.7
C2H2 : Air (L/min) 2 : 10 2 : 10 2 : 10 2 : 10 2 : 10
Lamp (mA) 15 4 20 10 15
Element Co Ni
Wavelength (nm) 240.7 232.0
Slit width (nm) 0.2 0.2
C2H2 : Air (L/min) 2 : 10 2 : 10
Lamp (mA) 25 30
- 29 -
(B)
Element Cu Cd Mn Pb Zn
Gas Ar Ar Ar Ar Ar
Flow rate (L/min) 1.5 1.5 1.5 1.5 1.5
Drying current (A) 16 12 12, 16 12 16
Drying time (sec) 32 32 32 32 32
Vaporizing current (A) 52 44 44, 52 44 40
Vaporizing time (sec) 4 4 4 4 4
Element Co Ni
Gas Ar Ar
Flow rate (L/min) 2.0 2.0
Drying current (A) 20 20
Drying time (sec) 32 32
Vaporizing current (A) 64 64
Vaporizing time (sec) 4 4
- 30 -
Table 2. Physical properties of Mn, Cu, Zn, Cd , Pb, Co, and
Ni.
Mn Cu Zn Cd Pb
Melting
point (K)1517 1357.6 692.7 594.2 600.6
Boiling
point (K)2335 2836 1180 1040 2023
Co Ni
Melting
point (K)1768 1726
Boiling
point (K)3201 3187
- 31 -
Table 3. Optimized ETV current of Mn, Cu, Zn, Cd , Pb, Co,
and Ni.
Element ETV current (A)Mn 48Cu 56Zn 40Cd 44Pb 44Co 64Ni 64
- 32 -
Table 4. Detection limits (3σblank, n=5), coefficient value, and
precision for Zinc, Cobalt, and Nickel of ETV-FAAS.
ElementDetection limit
(ng)
Coefficient value
(r)% RSD
Zn 0.16 0.99915 3.19Co 1.48 0.99942 3.53Ni 0.93 0.99931 3.87
- 33 -
Table 5. The reference concentration values of (A) NIST and
(B) NIES samples.
(A)
RMMinor and Trace Constituents (mg/kg)
Mg Zn Mn Cu
8414 - 142±14 - 2.84±0.45
8415 - 67.5±7.6 1.78±0.38 2.70±0.35
8418 - - 14.3±0.8 -
8432 31±5 - - -
(B)
Trace Constituents (μg/g)
Pb Cd Mn Zn
NIES
No. 3- - 69±5 20.5±1.0
NIES
No. 8219±9 - - -
NIES
No. 10- 1.82±0.06 - -
- 34 -
Table 6. The analytical results of Mn, Cu, Zn, Cd and Pb in
food, biological and environmental samples with linearity and
relative standard deviation value in percentage.
Element Sample typeRSD
(%)r
Analytical
result (mg/kg)
Mn
Whole egg powder
Wheat gluten
Chlorella
4.5
5.2
4.8
0.99600
0.99995
0.97720
1.52±0.07
15.0±0.8
71.36±3.43
CuBovine muscle powder
Whole egg powder
4.5
2.8
0.99570
0.98703
2.52±0.11
2.45±0.07
Zn
Bovine muscle powder
Whole egg powder
Chlorella
3.4
4.2
4.7
0.99726
0.99988
0.99620
143.01±4.86
63.3±2.7
20.0±1.0
Cd Flourished Rice 5.5 0.99912 1.84±0.10
PbVehicle exhaust
particulates5.4 0.99999 210±11
- 35 -
T e f lo n B a s e
P y r e x G la s s D o m e
S a m p le I n le t
+ -
S a m p le O u t l e t
A u x i l i a r y O x i d a n t I n le t
F lo w M e t e rA r g o n G a s O - r in g
F u e l I n l e t
D r a in
F la m e
P o w e r S u p p ly
T a n ta l u m F i la m e n t
E le c t r o d e
Fig. 2 The scheme of electrothermal vaporization flame
atomic absorption spectrometry used in this study
(ETV-FAAS).
- 36 -
(A) (B)
Fig. 3 The lighten filament of (A) Tantalum, and (B)
Tungsten at the vaporizing current.
- 37 -
0 .0
0 .1
0 .2
0 .3
0 .4
0 .51 2 A
4 0 4 4 4 8 5 2 5 6 6 0
Ab
sorb
ance
V a p o r iz a tio n C u rre n t (A )
0 .0
0 .1
0 .2
0 .3
0 .4
4 0 4 4 4 8 5 2 5 6 6 0
1 6 A
Ab
so
rban
ce
V a p o r iz a tio n C u rre n t (A )
(A)
Fig. 4 The increasement of peak absorbance of Cu(A),
Cd(B), Mn(C), Pb(D), Zn(E) with tantalum (10ppm, 5μL) and
Co(F), Ni(G) with tungsten (10ppm, 3μL) according to the
drying and vaporizing current change.
- 38 -
0 .0 0
0 .0 5
0 .1 0
0 .1 5
0 .2 0
0 .2 5
0 .3 0
0 .3 5
1 2 A
3 6 4 0 4 4 4 8
Ab
so
rba
nc
e
V a p o r iz a t io n C u r r e n t (A )
0 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0 .0 6
0 .0 71 6 A
3 6 4 0 4 4 4 8
Ab
sorb
ance
V a p o r iz a tio n C u rre n t (A )
(B)
Fig. 4 (Continued).
- 39 -
0 .0
0 .1
0 .2
0 .3
4 0 4 4 4 8 5 2 5 6
Ab
sorb
ance
V a p o r iz a tio n C u rre n t (A )
1 2 A
0 .0 0
0 .0 8
0 .1 6
4 0 4 4 4 8 5 2 5 6
Ab
sorb
ance
V a p o r iza tio n C u rre n t (A )
1 6 A
(C)
Fig. 4 (Continued).
- 40 -
0 .00 0
0 .00 5
0 .01 0
0 .01 5
0 .02 0
0 .02 5 12 A
40 44 48
Ab
sorb
ance
V a p o riza tio n C u rre n t (A )
0 .0 0 0
0 .0 0 5
0 .0 1 0
0 .0 1 5
0 .0 2 0
0 .0 2 51 6 A
4 0 4 4 4 8
Ab
sorb
ance
V a p o r iz a tio n c u rre n t (A )
(D)
Fig. 4 (Continued).
- 41 -
0 .0
0 .1
0 .2
0 .3
0 .4
0 .5
3 6 4 0 4 4 4 8 5 2
Ab
sorb
ance
C u r re n t (A )
1 2 A
0 .0
0 .1
0 .2
0 .3
3 6 4 0 4 4 4 8 5 2
Ab
sorb
ance
C u r re n t (A )
1 6 A
(E)
Fig. 4 (Continued).
- 42 -
0 .0 0 0
0 .0 0 5
0 .0 1 0
0 .0 1 5
0 .0 2 0
0 .0 2 5
0 .0 3 0
0 .0 3 5
0 .0 4 0
0 .0 4 5
5 6 6 0 6 4 6 8
Ab
sorb
ance
C u rre n t (A )
0 .0 0 0
0 .0 0 5
0 .0 1 0
0 .0 1 5
0 .0 2 0
0 .0 2 5
0 .0 3 0
0 .0 3 5
0 .0 4 0
0 .0 4 5
5 6 6 0 6 4 6 8
Ab
sorb
ance
C u rre n t (A )
(F)
Fig. 4 (Continued).
- 43 -
0 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0 .0 6
5 6 6 0 6 4 6 8
Ab
sorb
ance
C u rre n t(A )
0 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0 .0 6
5 6 6 0 6 4 6 8
Ab
so
rba
nc
e
C u r r e n t (A )
(G)
Fig. 4 (Continued).
- 44 -
0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6
0.0160.0180.0200.0220.0240.0260.0280.0300.0320.0340.036
Ab
sorb
ance
F low -rate (L /m in)
C o N i
Fig. 5 The efficiency of argon carrier gas flow rate for
ETV-FAAS.
- 45 -
0.00
0.01
0.02
0.03
0.04
0.05
0.06
0.07
44A 48 A 52A
Ab
sorb
ance
V aporization C urren t (A)
16 A 12 A
Fig. 6 Mg absorbance characteristic of ETV current in RM
8432 (Corn starch)
- 46 -
0.5 1 .0 1 .5 2 .0 2 .50 .00
0.03
0.06
0.09
0.12
Ab
sorb
ance
F low -rate (L /m in)
Fig. 7 The efficiency of argon carrier gas flow rate of Mg
in corn starch.
- 47 -
0 2 4 6 80 .0
0 .2
0 .4
0 .6
0 .8
1 .0
1 .2
1 .4
Ab
so
rba
nc
e
T im e (s e c )
0 2 0 4 0 6 0 8 0 1 0 00 .0
0 .2
0 .4
0 .6
0 .8
1 .0
1 .2
Ab
so
rba
nc
e
C o n c e n t r a t io n (n g )
(A)
Fig. 8 Absorbance spectra and calibration curves of the
element (A) Zn, (B) Co, and (C) Ni.
- 48 -
2 4 6 80 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
Ab
sorb
ance
T im e (s e c )
0 5 1 0 1 5 2 0 2 5 3 0 3 50 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
Ab
sorb
ance
C o n c e n tra tio n (n g )
(B)
Fig. 8 (Continued).
- 49 -
2 4 6 80 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0 .0 6
0 .0 7
Ab
sorb
ance
T im e (s e c )
0 5 1 0 1 5 2 0 2 5 3 00 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0 .0 6
0 .0 7
Ab
sorb
ance
C o n c e n tra tio n (n g )
(C)
Fig. 8 (Continued).
- 50 -
2 4 60 .0 0
0 .0 4
0 .0 8
0 .1 2
0 .1 6
0 .2 0
0 .2 4
Ab
sorb
ance
T im e (s e c )
- 2 . 5 - 2 . 0 - 1 . 5 - 1 . 0 - 0 . 5 0 . 0 0 . 5 1 . 0 1 . 5 2 . 0 2 . 5 3 . 0
0 . 0 5
0 . 1 0
0 . 1 5
0 . 2 0
0 . 2 5
Ab
so
rba
nc
e
C o n c e n t r a t i o n ( p p m )
(A)
Fig. 9 Absorbance spectra and calibration curves of the
element (A)Zn in RM 8414 and RM 8415, (B) Mn in RM 8415
and RM 8418, and (C) Cu in RM 8414 and RM 8415 by
standard addition method.
- 51 -
2 4 60 .0 00 .0 50 .1 00 .1 50 .2 00 .2 50 .3 00 .3 50 .4 00 .4 50 .5 00 .5 5
Ab
so
rba
nc
e
T im e (s e c )
-4 -3 -2 -1 0 1 2
0.1
0.2
0.3
0.4
0.5
0.6
Ab
sorb
ance
C oncentra tion (ppm )
(A)
Fig. 9 (Continued).
- 52 -
2 4 60 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
Ab
sorb
ance
T im e (s e c )
- 2 - 1 0 1 2 3 4 5
0 . 0 1
0 . 0 2
0 . 0 3
0 . 0 4
Ab
so
rba
nc
e
C o n c e n t r a t io n ( p p m )
(B)
Fig. 9 (Continued).
- 53 -
2 4 60 .0 0
0 .0 1
0 .0 2
0 .0 3
Ab
sorb
ance
T im e (s e c )
- 1 . 0 - 0 . 5 0 . 0 0 . 5 1 . 0 1 . 5 2 . 0 2 . 5 3 . 0 3 . 5
0 . 0 0 5
0 . 0 1 0
0 . 0 1 5
0 . 0 2 0
0 . 0 2 5
0 . 0 3 0
0 . 0 3 5
Ab
so
rba
nc
e
C o n c e n t r a t i o n ( p p m )
(B)
Fig. 9 (Continued).
- 54 -
2 4 60 .0 0
0 .0 1
0 .0 2
0 .0 3
Ab
sorb
ance
T im e (s e c )
- 0 . 6 - 0 . 3 0 . 0 0 . 3 0 . 6 0 . 9 1 . 2 1 . 5
0 . 0 1
0 . 0 2
0 . 0 3
Ab
so
rba
nc
e
C o n c e n t r a t io n ( p p m )
(C)
Fig. 9 (Continued).
- 55 -
2 4 60 .0 0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
Ab
sorb
ance
T im e (s e c )
- 0 . 4 - 0 . 2 0 . 0 0 . 2 0 . 4 0 . 6 0 . 8 1 . 0 1 . 2
0 . 0 1
0 . 0 2
0 . 0 3
0 . 0 4
0 . 0 5
Ab
so
rba
nc
e
C o n c e n t r a t i o n ( p p m )
(C)
Fig. 9 (Continued).
- 56 -
2 4 6 80 .0 0 0
0 .0 0 5
0 .0 1 0
0 .0 1 5
0 .0 2 0
0 .0 2 5
0 .0 3 0
Ab
sorb
ance
T im e (s e c )
- 2 - 1 0 1 2 3 4 5 6 7
0 . 0 0 5
0 . 0 1 0
0 . 0 1 5
0 . 0 2 0
0 . 0 2 5
0 . 0 3 0
0 . 0 3 5
Ab
so
rba
nc
e
C o n c e n t r a t i o n ( p p m )
(A)
Fig. 10 Absorbance spectra and calibration curves of the
element (A) Mn and (B) Zn in NIES No. 3 by standard
addition method.
- 57 -
2 4 6 80 .0
0 .2
0 .4
0 .6
0 .8
Ab
so
rba
nc
e
T im e (s e c )
- 1 . 0 - 0 . 5 0 . 0 0 . 5 1 . 0 1 . 5 2 . 0
0 . 2
0 . 4
0 . 6
0 . 8
Ab
so
rba
nc
e
C o n c e n t r a t i o n ( p p m )
(B)
Fig. 10 (Continued).
- 58 -
2 4 60 .0
0 .1
0 .2
0 .3
0 .4
0 .5
0 .6
0 .7
0 .8
0 .9
1 .0
1 .1
1 .2
1 .3
1 .4
Ab
so
rba
nc
e
T im e (s e c )
- 2 - 1 0 1 2 3 4 5 6 7
0 .2
0 .4
0 .6
0 .8
1 .0
1 .2
1 .4
Ab
so
rba
nc
e
C o n c e n t r a t io n ( p p m )
(A)
Fig. 11 Absorbance spectra and calibration curves of the
element (A) Pb and (B) Cd in NIES No. 8 and No. 10 by
standard addition method.
- 59 -
2 4 60 .0
0 .1
0 .2
0 .3
0 .4
0 .5
0 .6
0 .7
0 .8
Ab
sorb
ance
T im e (s e c )
- 1 . 0 - 0 . 5 0 . 0 0 . 5 1 . 0 1 . 5 2 . 0 2 . 5 3 . 0
0 . 1
0 . 2
0 . 3
0 . 4
0 . 5
0 . 6
0 . 7
Ab
so
rba
nc
e
C o n c e n t r a t i o n ( p p m )
(B)
Fig. 11 (Continued).
- 60 -
ACKNOWLEDGMENTS
어느덧 제게도 학부시절부터 대학원까지 화학과와 맺은 인연의 시
간들을 정리할 때가 다가왔습니다. 참으로 많은 경험들을 통해서 제
자신이 좀더 나은 방향으로 발전될 수 있었던 소중한 시간들이었다고
느끼고 있습니다. 이제 그 동안 제게 도움을 주신 모든 분들께 이 지
면을 빌어 감사의 인사를 전하고자 합니다.
먼저, 부족함이 많은 저에게 연구뿐만 아니라 여러 가지 면에서
신경을 써주시고 도움을 주신 이용일 교수님께 진심으로 감사드립니
다. 그 동안 제게 해주신 긍정적이면서도 발전적인 조언들이 마음 속
깊이 새겨져 잊혀지지 않을 듯 싶습니다. 그리고 화학과의 제일 어른
이시고 학생들 뒤에서 묵묵히 지켜봐 주시는 백건호 교수님과 따뜻하
고 자상한 마음으로 학생들을 격려해 주시는 이창순 교수님, 이민주
교수님, 투박하지만 정겨운 어투로 편안하게 대해주시는 안철진 교수
님, 소탈한 모습이 잘 어울리시는 원태진 교수님, 지금은 바쁜 연구활
동 때문에 미국에 계시는 멋스러운 유 재 교수님, 넉넉한 웃음이 좋
아 보이시는 신동수 교수님께 감사드립니다.
오랜 실험실 생활을 통하여 제가 즐겁게 학교생활을 할 수 있도록
도움을 주신, 그리고 조언을 아끼지 않으시던 김미경 박사님과 자신
의 일에 항상 최선을 다하는 강신봉 선배님, 의욕으로 가득한 모습이
인상적이던 유학생 1호 김재국 선배님, 호탕한 웃음 때문에 인기가
많은 최종수 선배님, 장난을 많이 쳐도 기분좋게 받아주던 김 욱 선
배님, 차분함과 꼼꼼함으로 똘똘 뭉친 임재민 선배님과 그의 아내가
되어 타국에서도 현명함으로 잘 지낼 박성하 선배님, 항상 웃음 짓고
다닌다 하여 방 이로 통하던 조효현 선배님, 특이한 유머감각으로
많은 사람들에게 웃음을 제공하던 김 주 선배님, 말수는 적지만 마
- 61 -
음이 참 따뜻한 이선 선배님, 방장으로서 최선을 다하고 있는 김상
득 선배님, 새내기 대학원생으로서 더욱 연구에 정진할 이원배 선배
님과 이수경 후배님, 갓 실험을 시작하여 서투름이 많을 강종필 선배
님과 학부동기인 박승건, 이태희 학우님께 감사드립니다.
같은 시기에 대학원에 진학하여 여러모로 신경을 많이 써주었고,
사회에 나가서도 자신의 위치에서 변함없이 최선을 다하여 생활할 변
기환 선배님과 한상윤 선배님, 그리고 같은 사무실에 있으면서 여러
가지로 도움을 많이 준 박주희 선배님께 진심으로 감사드립니다. 또
한, 언제나 제 뒤에서 묵묵히 지켜보면서 격려해 준 우 희 님, 조미
희 님, 신희경 님에게도 감사드립니다.
마지막으로, 이 날까지 올바른 정신으로 성실하고 정직하게 살도
록 가르쳐주신 부모님과 동생을 아끼는 마음으로 질책도 마다하지 않
던 언니, 오빠께 감사드립니다. 항상 건강하시고 행복하시기만을 바랄
뿐입니다. 그리고 저에 대한 굳은 믿음이 실망으로 바뀌지 않도록 더
욱 열심히, 부지런히 살겠습니다.