df series manual · 2017-07-11 · atesteo gmbh df series manual – version 1.7 – 2017/06 2/48...

48
DF Series Manual

Upload: others

Post on 05-Jul-2020

20 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

DF Series Manual

Page 2: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 2/48

DF Series Manual Version 1.7 2017/06

ATESTEO GmbH

Konrad-Zuse-Str. 3

52477 Alsdorf

T +49 (0) 2404 9870-0

F +49 (0) 2404 9870-59

[email protected]

www.atesteo.com

Page 3: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 3/48

Inhalt

1 Introduction ..................................................................................................................................... 5

1.1 Manufactures .................................................................................................................................... 5

2 Safety Instructions ......................................................................................................................... 6

3 Quickstart guide ............................................................................................................................. 7

3.1 Unpacking and installation ................................................................................................................ 7

3.2 Webbrowser...................................................................................................................................... 8

3.3 Network connection .......................................................................................................................... 9

3.4 Network settings [Windows 7] ........................................................................................................ 10

3.5 Proxy configuration [Windows 7] .................................................................................................... 13

3.6 Web interface.................................................................................................................................. 16

4 Mounting ........................................................................................................................................ 17

4.1 Placement ....................................................................................................................................... 18

5 Website guide ............................................................................................................................... 20

5.1 Homepage ...................................................................................................................................... 21

5.1.1 Graph overview .............................................................................................................................. 22

5.1.2 Navigation menu ............................................................................................................................. 23

5.2 Signalbar ......................................................................................................................................... 24

5.3 Newdata .......................................................................................................................................... 25

5.4 Power supply .................................................................................................................................. 26

5.5 Alarm settings ................................................................................................................................. 27

5.6 Torquemeter ................................................................................................................................... 28

5.7 Measuring1 ...................................................................................................................................... 29

5.8 Analog settings ............................................................................................................................... 30

5.9 Frequency settings+ ....................................................................................................................... 31

5.10 Filter settings .................................................................................................................................. 31

5.11 CAN settings ................................................................................................................................... 32

5.11.1 CAN state ....................................................................................................................................... 33

5.11.2 CAN configuration .......................................................................................................................... 34

5.12 State ............................................................................................................................................... 36

5.13 Ethernet settings ............................................................................................................................. 37

5.14 General settings ............................................................................................................................. 38

5.15 Service page ................................................................................................................................... 38

6 Technical specs ............................................................................................................................ 39

6.1 Pinallocations.................................................................................................................................. 39

6.1.1 X770 Power supply / Frequency output ......................................................................................... 39

6.1.2 X771 Analogue / CAN / Alarm / Input ............................................................................................. 40

6.1.3 X772 Ethernet ................................................................................................................................. 42

6.1.4 X773 Central cable ......................................................................................................................... 42

6.2 Mechanical data ............................................................................................................................. 43

6.2.1 Dimension rotor DF standard type [mm] ........................................................................................ 44

Page 4: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 4/48

6.2.2 Dimension stator ............................................................................................................................. 46

6.2.3 Dimension TCUIV ........................................................................................................................... 46

Page 5: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 5/48

1 Introduction

Dear customer,

before leaving our company every unit is tested by extensive function and quality examinations, which

guarantee that the system complies with the stated specifications. Nevertheless should there be any problem,

please contact us.

Before shipping a system, the serial number of each component of your configuration is registered by our

company, so that an individual and short-term support can be guaranteed. It is understood that we will inform

you about innovations and modifications of the system

1.1 Manufactures

ATESTEO GmbH

(Hereinafter referred to as manufacturer)

Konrad-Zuse-Str. 3

52477 Alsdorf

T +49 (0) 2404 9870-0

F +49 (0) 2404 9870-59

[email protected]

www.atesteo.com

Page 6: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 6/48

2 Safety Instructions

The Manual must be read carefully before start-up, maintenance work or any other work on the torque

measuring system. Prerequisite for the safe and proper handling of the equipment knows all safety instructions

and safety regulations of the attachment.

Every safeguard needs to be correctly mounted and fully functional before any start-up.

Measurement shafts mounted to the torque meter must be properly designed, so that critical bending speed is

avoided.

Exclusively qualified laborers are allowed to do maintenance work on any electrical components.

Note

Improper use and handling as well as constructional changes will invalidate the EC declaration of

conformity.

Page 7: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 7/48

3 Quickstart guide

3.1 Unpacking and installation

The package contains the following terms:

1. Torque meter (Rotor)

2. DF Stator

3. TCU IV (Torque Control Unit)

4. Central cable

5. 12-pin connector

6. 16 pin connector

Follow the next instructions to connect the system.

► Mount torque meter and antenna as described in the 2 Mounting guide

► Connect DF-Stator and TCU IV with the central cable

► Connect the 12-pin and 16-pin connectors to the TCU IV

Note

Pin allocations are specified under 4 TECHNICAL SPECS

Important

Power supply pins are mandatory. Power supply rating: 24V±1V/1A

1. 2. 3.

Page 8: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 8/48

3.2 Webbrowser

One of the following web browsers is required, to use the web interface:

http://www.mozilla.org/en-US/firefox/all/a

https://www.google.com/intl/de/chrome/browser/index.html?standalone=1#eula

http://windows.microsoft.com/de-de/internet-explorer/ie-11-worldwide-languages

If none of the browsers is installed yet, follow the instruction after downloading the browser.

Note

Please use the latest version of the browser

Chrome

Internet

Explorer

Firefox

Page 9: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 9/48

3.3 Network connection

Before startup ensure that the TCU IV is connected properly. Please check:

► Central cable of Receiver

► Correct Position of Receiver

► Power supply connection

Requires:

Patch cable with RJ45 connector

Three options to connect to network:

1. Connect in domain network:

Connect the TCU IV using the patch cable with a

free network socket.

2. Connect with network-to-USB adapter:

The adapter needs to be installed on the

computer. Now connect TCU IV with a patch

cable to the adapter.

3. Connect directly

Connect with a patch cable directly from the

Ethernet socket of the computer.

Page 10: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 10/48

3.4 Network settings [Windows 7]

Network settings need to be changed, if option one or two is selected.

1. Open Control panel

2. Navigate to Network and Sharing Center

3. Open the Status window of the connection

4. To open Connection Properties click on

5. Select Internet Protocol Version and click on

6. Enter the following values

a. IP address: 172.16.86.2

b. Subnet mask: 255.255.255.0

Note

Administrator rights are required

Page 11: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 11/48

Page 12: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 12/48

Page 13: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 13/48

3.5 Proxy configuration [Windows 7]

1. Open Control Panel

2. Navigate to Internet Properties

3. Choose the Connections tab

4. Click on

5. Navigate to Proxy Settings by clicking

6. Input the new proxy exception

a. Proxy exception: 172.16.86.*

Page 14: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 14/48

Page 15: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 15/48

4. LAN Settings

5. Proxy Settings

Page 16: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 16/48

3.6 Web interface

The web address of the TCU IV is needed to open the web interface.

The web address contains of "torque20-" and [SERIAL NUMBER].

Use the following IP address to open the web interface:

http://172.16.86.3,

Or use the following web address (requires a DNS-Server):

http://torque20-[SERIAL NUMBER]

Page 17: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 17/48

4 Mounting

Mounting torque meter with input side to drive train.

Type Amount of screws Tightening torque

DF1 6x M8 12.9 43 Nm

DF2 8x M10 12.9 84 Nm

DF3 8x M12 12.9 145 Nm

DF4 8x M14 12.9 235 Nm

DF5 8x M16 12.9 365 Nm

Page 18: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 18/48

4.1 Placement

The DF stator must be positioned under the measuring flange that the output flange of the rotor covers the

black area of the stator.

Note:

Radial air-gap between receiver and torque-meter: 3mm +1mm /-2mm. Axial displacement of

black area: ±1mm

Outer edge of the flange

Page 19: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 19/48

The following images show the placement options with and without the speed option.

Optimal 6mm

Page 20: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 20/48

5 Website guide

LOGIN

Web interface is opened as described in the quick start guide.

The login password must be entered. By default the password is: admin.

Important

The password can be changed in the menu item Settings. Protect your device against

unauthorized access!

Page 21: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 21/48

5.1 Homepage

The login leads to the homepage.

Device information:

This overview of currently connected devices

includes information about the name of the TCU

and the serial number of the stator and the

torque sensor. The device name is editable in

the Settings menu.

Measurement features.

The measuring signal is set to zero by pushing

the button. The test signals are

enabled or disabled by the buttons.

Page 22: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 22/48

Test signal [controller] generates a full-scale deflection on every system output independent of the measured

inputs.

Test signal [torque sensor] generates a half-scale deflection at the torque sensor of torque input 1 and 2. The

deflection is generated by an offset jump of the first amplifiers of the respective measurement chain.

Note:

Test signals must be disabled before measuring

5.1.1 Graph overview

Diverse signals are displayed in the main part of the home page. These signals graphically represent Torque

input 1, Torque input 21, Acceleration input

1, the Speed input

1 and the Temperature of the torque sensor.

Graphs are automatically scaled according to their nominal value.

Note

Signals are not displayed in real-time. This may lead to delayed representations.

1 Depending on the configuration of the current system

Page 23: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 23/48

5.1.2 Navigation menu

If the navigation bar is closed, click on the company logo to open it again.

Page 24: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 24/48

5.2 Signalbar

The signal bar is shown on the right-hand side of the website.

General states

The “General” tab contains information about the transmission of

the system. The LED “’Transmission quality” provides information

about transmission quality of measured data from torque meter to

torque control unit. The “Sensor supply” LED indicates the status of

the torque sensor supply voltage.

Alarm states

Alarm values are displayed in the “Alarm” section. Red indicates

that threshold is exceeded. Alarms can be reset by pressing

button.

Alarm limits can be set in the “Alarm” settings menu.

Measured data

Below the alarm states the measured data of the different system

inputs Torque input 1, Torque input 22, Acceleration input

1, Speed

input1, Power supply torque sensor, Power supply controller and

torque sensor Temperature are displayed numerical.

Note

Signals are not displayed in real-time. This may lead to delayed representations.

Important

Check the supply voltage and the orientation between torque sensor and stator-antenna to guarantee an optimal transmission quality.

A

Optimal (green) | Okay (yellow) | Bad (red) B

Transmitting (green) | No Transmitting (gray) C

Value below Threshold (green) | Value above Threshold (red)

2 Depending on the configuration of the current system

A

B

C

Page 25: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 25/48

5.3 Newdata

In the case that another torque meter is connected, the adjusted values of the torque meter and the TCUIV are

compared. If the values are different following dialog appears:

By pressing the button, the values of the torque meter are transferred to the TCU IV.

The outputs are adjusted accordingly.

Page 26: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 26/48

5.4 Power supply

In the menu “Power supply” the supply voltage of the torque sensor will be adjusted. The power-switch

activates and deactivates the power supply. On activation the last adjusted voltage-value will be reactivated.

Because the inductively transmitted power depends on the gap between torque sensor and stator-antenna the

supply voltage has to be readjusted after change of position. The optimal supply voltage of the torque sensor is

9.5 V +/- 0.5 V. There are 2 methods for adjusting:

1. By pressing the button, the optimal supply voltage is automatically set.

2. The torque sensor voltage is set manually with the slider. Changes are applied in real-time. The further

the slider is moved to the right, the higher the torque sensor supply voltage gets.

In the following section, the influence of the torque sensor voltage to the signal quality is shown:

Sensor voltage LED Sensor supply Description

9,5 V ± 0,5 V Green Optimal supply voltage

9,5 V ± 1,0 V Yellow Supply voltage is ok. Possibly interrupted

transmission, if the supply voltage vary to much

during the operation of the test bench

9,5 V ± > 1,0 V Red Poor supply voltage. Possibly interrupted

transmission, possibly invalid measurement values

Important

The optimal torque sensor supply voltage should be at 9.5V. Power supply will be deactivated and set back to zero, if a critically value is adjusted, in order to prevent damage of inductive power supply components.

Page 27: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 27/48

5.5 Alarm settings

Alarm thresholds for speed3, acceleration

1 and torque can be adjusted.

Torque input limit

An alarm for the torque signal can be set. Alarm is triggered when threshold is exceeded. In dual range

systems only the large measurement range1 is observed.

Speed input limit1

An alarm for the speed signal can be set. Alarm is triggered when threshold is exceeded.

Acceleration input limit1

An alarm for the acceleration signal can be set. Alarm is triggered when threshold is exceeded.

3 Depending on the configuration of the current system

Page 28: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 28/48

5.6 Torquemeter

Important

Improper values can falsify measurements or in worst case damage the measurement device. Proper values can be found in the test report.

Output switch4

Output signals can be swapped. The output swap affects the following outputs:

4 Depending on the configuration of the current system

Effects of the output swap

Output State

Website output Not swapped

Frequency output1 Swapped

Analog output

(voltage)1(*)

Swapped

Analog output

(current)1(*)

Swapped

CAN output1 Not swapped

Page 29: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 29/48

(*)

Only output 1 and output 2

Input settings

Rated torque and sensitivity of the measurement system can as well as

Torque input 1 and Torque input 21 be adjusted.

5.7 Measuring1

The Measuring Menu appears if the system measures acceleration. Data input is performed as described in

the previous chapter.

Page 30: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 30/48

5.8 Analog settings

Voltage5 range of analog outputs can be adjusted.

Current1 range of analog outputs can be adjusted.

Output calibration

Outputs must not be calibrated by an untrained user. Incorrect values can falsify the measurement.

Outputs are calibrated at the factory and usually do not need to be recalibrated.

5 Depending on the configuration of the current system

Page 31: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 31/48

5.9 Frequency settings+

Diverse output ranges can be set in the menu item Frequency. Output 1 to 21 can be configured.

Output calibration

Outputs must not be calibrated by an untrained user. Incorrect values can falsify the measurement.

Outputs are calibrated at the factory and usually do not need to be recalibrated.

For further information please contact our support.

5.10 Filter settings

The filter-settings influences the input signal.

The filter is a digital IIR filter 1st order and is related to the torque input channels. For each input channel, the

filter can be activated and the cut-off frequency can be set in the corresponding dropdown-box. The following

cut-off frequencies between 1 Hz and 4000 Hz are supported:

1 Hz, 10 Hz, 50 Hz, 100 Hz, 150 Hz, 250 Hz, 500 Hz, 1000 Hz, 2000 Hz and 4000 Hz.

The button is clicked to save the settings.

Page 32: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 32/48

5.11 CAN settings

CAN communication can be configured in the CAN menu.

Scaling factor

Received torque values need to be divided by the scale factor manually by the user.

Page 33: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 33/48

5.11.1 CAN state

CAN transmission is enabled and disabled by toggling the button.

The “CAN State” tab also contains information about the current state of the CAN bus. The different states are

explained in the following sections.

MODULE_ACTIVE: The CAN bus is working without any significant problems. The receive error counter (RX)

and the transmit error counter (TX) are < 128.

MODULE_PASSIV: The CAN bus works, however, a transmission or reception error occurred. TX or RX is >

127. In case that no more errors occurred, the counters are decremented and the status changes to

MODULE_ACTIVE. Otherwise, the bus should be checked.

BUS_OFF: The CAN module has been disconnected due to many transmission errors (TX > 255).Check the

CAN settings and perform a CAN reset.

INIT_ERROR: The CAN module cannot connect to the CAN bus. Check the CAN settings and perform a CAN

reset.

Page 34: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 34/48

NOTE: CAN State can be reset by switching the output off and on again.

5.11.2 CAN configuration

The general CAN transmission can be configured. Selected values need to fit the values of the current receiver

system. ► Baud (250kbit, 500kbit, 1Mbit)

► Transmit interval (between 1 and 1000 ms)

► Data format (Intel, Motorola)

► Identifier-Length (11 Bit, 29 Bit)

Send configuration

Page 35: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 35/48

CAN messages are formatted (dependent on configurations) in the following way:

Intel Data byte 0-3 Data byte 4-7

Identifier D0 D1 D2 D3 D4 D5 D6 D7

CAN-Identifier Torque 1 x Factor Torque 2 x Factor

CAN-Identifier Speed 0

Motorola

Identifier D3 D2 D1 D0 D7 D6 D5 D4

CAN-Identifier Torque 1 x Factor Torque 2 x Factor

CAN-Identifier Speed 0

Four CAN messages at most can be configured. CAN Messages can be enabled and disabled. Messages one,

two and three can be manually chosen and adjusted. The fourth message cannot be configured. It is reserved

for watchdog and state and the transmit interval is fixed at 1000ms.

Receive configuration

The receiving identifier for CAN message can be chosen. The following CAN messages can be received:

Messages

Type Hex Dec

Zero 0x4B1 1201

Test signal (on) 0x4B2 1202

Test signal (off) 0x4B3 1203

MD1/MD2 0x4B5 1205

MD2/MD1 0x4B6 1206

State reset 0x4BB 1211

Request state 0x4BC 1212

Power supply (off) 0x514 1300

Power supply (on) 0x515 1301

Alarm reset 0x578 1400

Request Ethernet settings 0xD05 3333

Page 36: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 36/48

The command must be included in the first 4 bytes [data bytes 0-3]. While receiving, distinction is made

between Motorola and Intel. A response message is sent if a message is successfully received. The response

massage is formatted in the following way:

Response message

Identifier Data byte [0-3] Data byte [4-7]

receive identifier +1 last command state

5.12 State

The status word is a 16-bit long data word, and is structured in the following way:

STATE DESCRIPTION Bit

ER

RO

R S

TA

TE

S

alarm rx aborted transmission 15

alarm md exceeded the torque threshold 14

alarm n exceeded the speed threshold 13

alarm acc exceeded the acceleration threshold 12

ME

AS

UR

EM

EN

T S

TA

TE

S

signal quality poor signal quality 11

current error shutdown due overcurrent 10

current warning excessive current. current error in 4 minutes 9

md2/md1 channel switch is active 8

zero zero adjustment was performed 7

test mode torque sensor test signal is on 6

test mode TCU test signal is on 5

GE

NE

RA

L S

TA

TE

S

- reserved 4

- reserved 3

- reserved 2

- reserved 1

power power supply is active 0

Page 37: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 37/48

5.13 Ethernet settings

Relevant adjustments for embedded measurement system in the intranet can be configured.

Important

Improper settings can break the device. In some cases the device cannot be reconfigured! In

that case the torque control unit must be reprogrammed in factory. Administration must be

consulted before configuring to get the proper settings.

If the network settings of the TCU have been forgotten, the settings can be queries via the

following CAN command:

Messages

Type Hex Dec

Request Ethernet settings 0xD05 3333

The response of the TCU has the following structure:

Type IP Address Subnet Mask

Byte i Byte 0 Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7

Content

(HEX)

AC 10 56 2 FF FF FF 0

Result

(DEC)

172 16 86 3 255 255 255 0

Page 38: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 38/48

5.14 General settings

A password and a custom name for the measurement system can be adjusted.

Following characteristics must be fulfilled: ► Total length between 4 and 8 signs

► At least one upper-case letter

► At least one lower-case letter

► At least one digit

Important

Write down your password and keep it in a safe place. The password cannot be set back to default!

5.15 Service page

The up to date firmware version of the device and manufacturer’s contact details can be found on the

Service page.

For service purposes, it’s also possible to export a list of system parameters.

Page 39: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 39/48

6 Technical specs

6.1 Pinallocations

6.1.1 X770 Power supply / Frequency output

12-pin connector, type M23

Pin Signal Description Cable color (cable optional)

Cross-Section in mm

1 F2_out- *1 Frequency output ch. 2 – RS422 White 0,25

2 F2_out+ *1 Brown 0,25

3 N2_out+ *1 Speed pulses output track 2 – RS422 Gray 0,25

4 N2_out- *1 Pink 0,25

5 N1_out+ *1 Speed pulses output track 1 – RS422 Blue 0,25

6 N1_out- *1 Red 0,25

7 F1_out- Frequency output ch. 1 – RS422 Yellow 0,25

8 F1_out+ Green 0,25

9 IP-reset_in Reset IP-Configuration – TTL 3.3 or TTL 5.0 White 0,5

10 Power+ Power supply 24 - 30 V / 1 A

Green 0,5

11 Power- Yellow 0,5

12 Digital GND Ground connection of digital signals Brown 0,5

*1 Optional – depending on the configuration of the system Frequency outputs The frequency output F1 and F2 represents torque output 1 and 2. The frequency span is declared by the web interface. The maxima represent positive and negative rated torque of the respective channel. When the output switch is activated, the two frequency outputs are switched. The frequency outputs must be connected to RS422 receivers. Connecting an output to ground will damage the TCU. It is irrelevant whether it is a positive or a negative output signal. Speed-pulses output The speed-pulses-outputs N1 and N2 represent every single pulse of speed track 1 and 2. The amount of pulses per rounds is dependent on the number of increments parameter of the speed ring. The increments numbers of both tracks are identical and phase shifted by

90°. The speed-pulses output must be connected to RS422 receivers. Connecting an output to ground will damage the TCU. It is irrelevant whether it is a positive or a negative output signal. IP-reset_in The IP reset signal resets the IP configuration of the Ethernet interface to the factory setting. The factory settings are listed in chapter x. For security reasons, the following procedure must be used for recovery:

Page 40: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 40/48

1. Turn on the IP reset signal.

2. Turn on the TCU.

3. Wait until the red LED flashes and the green LED is off.

4. Wait until the red and green LEDs light up constantly.

5. Turn off the IP reset signal. The TCU automatically restarts, with the reset IP

configuration. Power supply Connect the positive and negative power pins with an external power supply. The power supply must have a supply voltage between 24 and 30 Volt and must be able to supply 1 A constant current.

6.1.2 X771 Analogue / CAN / Alarm / Input

16-pin connector, type M23

Pin Signal Description Cable color (cable optional)

Cross-Section in mm

1 Test_in Activates controller test signal White 0,25

2 Zero_in Zero balance input – TTL 3.3 or TTL 5.0

Brown 0,25

3 Digital GND Ground connection of digital signals Green 0,25

4 Digital GND Yellow 0,25

5 CAN_H CAN HIGH connection Grey 0,25

6 CAN_L CAN LOW connection Pink 0,25

7 An4_out *1 Galvanic isolated analog voltage output

Blue 0,25

8 An2_out *1 Galvanic isolated analog voltage output

Red 0,25

9 An3_out *1 Galvanic isolated analog voltage output

Black 0,25

10 An1_out *1 Galvanic isolated analog voltage output

Purple 0,25

11 Alarm-MD_out Alarm torque limit – open-collector Grey/Pink 0,25

12 Analog GND Ground connection of analog signals Red/Blue 0,25

13 Alarm-N_out Alarm speed limit – open-collector White/Green 0,25

14 Output-switch_out / Err-state_out

State of output-switch – open-collector / State of system-errors – open-collector. Depending on software version

Brown/Green 0,25

15 Alarm-reset_in Reset all alarms – TTL 3.3 or TTL 5.0 White/Yellow 0,25

16 Aux_in Not used – TTL 3.3 or TTL 5.0 Yellow/Brown 0,25

*1 Optional – depending on the configuration of the system

Page 41: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 41/48

Test_in The test_in signal activates the controller testsignal as soon as the button is pressed for at least one second until the signal is switched off again. The test signal produces a positive full-scale at all outputs for measured values. A "TTL 3.3" or "TTL 5.0" level between test_in pin and digital GND is applied for the control. The signal is active high. Zero_in The zero_in signal activates the zeroing as soon as the button is pressed for at least one second. The zeroing is executed only once after trigging. The zeroing sets the instantaneous values of the torque inputs as a new zero points. A "TTL 3.3" or "TTL 5.0" level between zero_in pin and digital GND is applied for the control. The signal is active high. CAN The CAN interface allow the customer to receive the measured data in digital form and simultaneously send control signals to the TCU. The CAN_High and CAN_Low pins have to be connected with a 120 Ohm terminated CAN-Bus. The Identifiers are described in Chapter Fehler! Verweisquelle konnte nicht gefunden werden.. Analog_out The analog outputs 1 and 2 represents torque output 1 and 2. The analog output 3 represents speed output. The analog output 4 is not in use. The voltage span is declared by the web interface. The maxima represent positive and negative rated values of the respective channel. When the output switch is activated, output 1 and 2 are switched. The analog outputs 1-4 are single ended galvanic isolated voltage outputs with separately analog GND. Alarm-MD_out The alarm-MD output indicates that the torque threshold is exceeded. The threshold value is set in the Web Interface (see chapter Fehler! Verweisquelle konnte nicht gefunden werden.). The alarm remains active until the Alarm_Reset signal is triggered. The alarm output consists of an open collector circuit. In the active state, it connects the Alarm-MD_out pin with digital GND directly. Alarm-N_out The alarm-N output indicates that the speed threshold is exceeded. The threshold value is set in the Web Interface (see chapter Fehler! Verweisquelle konnte nicht gefunden werden.). The alarm remains active until the Alarm_Reset signal is triggered. The alarm output consists of an open collector circuit. In the active state, it connects the Alarm-N_out pin with digital GND directly. Output-switch_out / Err-state_out The function of this pin depends on the TCU firmware version:

Firmware less than V1.6.10: - Output-switch_out The output-switch output indicates the state of the output-switch. On active state the output-switch is active. Firmware greater than or equal V1.6.10: - Err-state_out The error state output indicates whether the system is disturbed. On active state the system works without faults and full transmission.

Page 42: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 42/48

The signal output consists of an open collector circuit. In the active state, it connects the output pin with digital GND directly. Alarm-reset_in The alarm-reset signal resets all alarm-signals as soon as the button is pressed for at least one second. The reset is executed only once after trigging. A "TTL 3.3" or "TTL 5.0" level between alarm-reset_in pin and digital GND is applied for the control. The signal is active high.

6.1.3 X772 Ethernet

Can be connect with a standard Kat 5e cable for diagnostics or setup. For permanent installation inside the teststand use the special connector. Hummel - 7R10400000

6.1.4 X773 Central cable

15-pin connector, type M16 – not for external periphery

Pin Signal Description Cable color

Cross-Section in mm / typ

A Power+ Supply voltage White 0,5 / straight

B Power- *2 Brown 0,5 / straight

C 7V-Power+ Green 0,5 / straight

1 Data-in+ Digital rotor data – LVDS White 0,25 / twisted

2 Data-in- Brown 0,25 / twisted

3 7V-Power- *2 Supply voltage (Reserved) NC 0,25 / twisted

4 N2- Speed-pulses track 2 Green 0,25 / twisted

5 N2+ Yellow 0,25 / twisted

6 N1- Speed-pulses track 1 Grey 0,25 / twisted

7 N1+ Pink 0,25 / twisted

8 Not used

9 Not used

10 Not used

11 Not used

12 Not used

*2 Pin B and Pin 3 are connected together in the circuit-board of the stator-antenna The following applies to double channel systems: Channel 1 corresponds to the measuring channel with the lower rated torque. Channel 2 corresponds to the measuring channel with the higher rated torque.

Page 43: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 43/48

6.2 Mechanical data

Type DF1 DF2 DF3 DF4 DF5

Maximum rated torque

Nm 1) 50 2) 100 3) 200 4) 500

1) 200 2) 500 3) 1000

1) 1000 2) 2000 3) 3000

1) 4000 2) 5000

1) 5000 2) 10000

Alternating torque

Nm equal selected maximum rated torque

Torsional stiffness

kNm/rad

1) 115 2) 177 3) 177 4) 272

1) 431 2) 391 3) 658

1) 1080 2) 1744 3) 2304

1) 3893 2) 4466

1) 5508 2) 8463

Torsional angle

° 1) 0,02 2) 0,03 3) 0,06 4) 0,11

1) 0,03 2) 0,07 3) 0,09

1) 0,05 2) 0,07 3) 0,07

1) 0,06 2) 0,06

1) 0,05 2) 0,07

Radial stiffness

kN/mm 1) 59,3 2) 82,1 3) 82,1 4) 119

1) 168 2) 156 3) 245

1) 231 2) 374 3) 511

1) 534 2) 659

1) 559 2) 930

Axial stiffness

kN/mm 1) 254 2) 331 3) 331 4) 456

1) 530 2) 509 3) 705

1) 497 2) 711 3) 914

1) 686 2) 816

1) 517 2) 982

Max. bending moment

Nm 1) 41 2) 57 3) 57 4) 83

1) 127 2) 119 3) 273

1) 180 2) 292 3) 399

1) 427 2) 527

1) 537 2) 892

Radial load limit at run out error of 0,02mm

kN 1) 1185 2) 1642 3) 1642 4) 2372

1) 3352 2) 3128 3) 4896

1) 4622 2) 7487 3) 10210

1) 10684 2) 13185

1) 11180 2) 18591

Axial load limit

kN 1) 8 2) 10 3) 10 4) 14

1) 16 2) 15 3) 21

1) 15 2) 22 3) 28

1) 21 2) 25

1) 16 2) 30

Mass kg 1) 1,4 2) 1,4 3) 1,4 4) 1,4

1) 1,8 2) 1,8 3) 1,9

1) 2,9 2) 3 3) 3

1) 4,35 2) 4,4

1) 7,7 2) 8

Inertia kgm² 1) 0,00201 2) 0,00202 3) 0,00202 4) 0,00202

1) 0,00384 2) 0,00384 3) 0,00387

1) 0,0095 2) 0,0095 3) 0,0096

1) 0,0197 2) 0,0197

1) 0,057 2) 0,052

Balancing class

Q 2.5

Mechanical Shock (EN 60068-2-27) Vibration Loads (EN 60068-2-6)

Quantity n 1000 Frequency range Hz 1 …. 2000

Duration ms 3 Duration Hz 2.5

Acceleration m/s² 650 Acceleration m/s² 200

Page 44: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 44/48

6.2.1 Dimension rotor DF standard type [mm]

Page 45: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 45/48

Type Nominal rated torque

ØD1 without speed encoder

ØD1 with speed encoder

ØD2 B B1 B2

DF1S 50/ 100 / 200/ (500) Nm

101 107 107 45 10 20

DF2S (200)/ 500 / 1000 Nm

122 128 128 48 10 20

DF3S (1000)/ 2000/ 3000 Nm

152 158 158 49 10 20

DF4S 4/ 5 kNm 181 187 187 50 11 20

DF5S (5)/ 10 kNm 224 230 230 60 12 20

Type Nominal rated torque

ØC T1 T2 PCD 1 N PCD 2

DF1S 50/ 100 / 200/ (500) Nm

57 H6/g5 4.2 2 84 / Ø9 6x 84 / M8

DF2S (200)/ 500 / 1000 Nm

75 H6/g5 4.2 2 101,5 / Ø11 8x 101,5 / M10

DF3S (1000)/ 2000 / 3000 Nm

90 H6/g5 4.2 2.5 130 / Ø13 8x 130 / M12

DF4S 4/ 5 kNm 110 H6/g5 4.2 2.8 155,5 /Ø15 8x 155,5 / M14

DF5S (5)/ 10 kNm 140 H6/g5 4.2 3.5 196 / Ø17 8x 196 / M16

Note

Dimension will change with adapted couplings

Page 46: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 46/48

6.2.2 Dimension stator

6.2.3 Dimension TCUIV

Page 47: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 47/48

Page 48: DF Series Manual · 2017-07-11 · ATESTEO GmbH DF Series Manual – Version 1.7 – 2017/06 2/48 DF Series Manual Version 1.7 2017/06 ATESTEO GmbH Konrad-Zuse-Str. 3 52477 Alsdorf

ATESTEO GmbH

DF Series Manual – Version 1.7 – 2017/06 48/48