    {Diagnostics}{Production}{HealthImaging}{ExternalAndInternal} Publication No. DG4826-126AUG04

    Supersedes

    28JUL03

    DIAGNOSTICS

    for the

    KodakDirectViewCR 950 SYSTEM

    Service Code: 4826

    Important

    Qualified service personnel must repair this equipment.

    H196_0500GC

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    2/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    3/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    4/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    5/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    6/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    7/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    8/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    9/206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    10/206

    DIAGNOSTICS Using the Diagnostics

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    11/206

    G OS CS Us g t e ag ost cs

    26AUG04

    DG4826-1

    Page11 of 206

    Section 1: Using the Diagnostics

    System Service Diagnostics Screen

    Important

    For systems with Software Version 4.1, you must have a Session ID for access to theService Functions and Diagnostics. See SERVICE BULLETIN 843.

    1 At the main menu, touch:

    [Service Functions]

    [ENTER]

    [Diagnostics]

    2 Does the System Service Diagnostics screen display?

    3 Check that the TX and RX LEDs are blinking on the MASTER CENTRAL PROCESSINGUNIT BOARD A2 and on the NETWORK CARD in the computer.

    4 At the main menu, touch:

    [Service Functions]

    [ENTER]

    [EXIT TO DESKTOP]

    5 Select My Computer>Control Panel>Network>Protocols.

    6 Double-click [TCP/IP Protocol].

    7 Select [ADAPTER 1].

    Yes No

    Advance to the test or procedure. Continue with Step 3.

    DIAGNOSTICS Using the Diagnostics

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    12/206

    g g

    26AUG04

    DG4826-1

    Page12 of 206

    8 Check the values for the NETWORK CARD that is connected to the MASTER CENTRALPROCESSING UNIT BOARD A2:

    IP Address 192.168.0.1

    Subnet Mask 255.255.255.0

    Default Gateway 192.68.0.99

    DIAGNOSTICS Using the Diagnostics

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    13/206

    26AUG04

    DG4826-1

    Page13 of 206

    Functions of the System Service Diagnostics Screen

    Table 1 Component Tests

    BUTTON Description

    [Individual ComponentControl]

    Allows you to actuate and check the status of: STEPPER, ALIGNMENT CAMS

    MAIN BELT STEPPER

    LOAD/UNLOAD BAR

    ALIGNMENT CAMS

    ELEVATOR, SIZE CHANGER

    MAIN CAM, SCANNER

    MAIN CAM MOTOR

    SLOW SCAN CONTROLLER

    LASER DIODE

    ERASE LAMPS

    [Component Control: Step] Allows you to check the status of the components byexecuting the steps:1. Advancing the CASSETTE to SLOT 1

    2. Loading the CASSETTE

    3. Loading the PLATE4. Scanning the PLATE

    5. Erasing the PLATE

    6. Ejecting the PLATE

    7. Ejecting the CASSETTE

    DIAGNOSTICS Using the Diagnostics

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    14/206

    26AUG04

    DG4826-1

    Page14 of 206

    Table 2 Calibration

    BUTTON Description

    [View Collector Profiles]

    Important

    For the procedure, see Calibration for the CR 950SYSTEM.

    Displays each COLLECTOR PROFILE for each size ofCASSETTE, including:

    number of steps from HOME to START OF SCAN

    OFFSET - number of pixels advanced from START tothe diagnostic image

    Amplitude - number of pixels from OFFSET to lastpixel of the diagnostic image

    PMT 1 - 2: correction value of each PMT for gainadjustment

    DIAGNOSTICS Using the Diagnostics

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    15/206

    26AUG04

    DG4826-1

    Page15 of 206

    BUTTON Description

    [Scan Calibration]

    Caution

    All COVERS must be installed when using this option. For the procedure, see Calibration for the CR 950

    SYSTEM.

    Completes the calibration of:

    GALVO

    PMT 1

    PMT 2

    Starts a COLLECTOR PROFILE for all sizes ofSTORAGE PHOSPHOR PLATES.

    Allows calibration of the PLATES exposed at 15 - 25 mR.

    Displays the date and time of the last successfulcalibration.

    Provides the ability to:

    make a backup of the current calibration files from theSCANNER to the computer

    restore from the computer backup

    remove the existing calibration

    DIAGNOSTICS Using the Diagnostics

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    16/206

    26AUG04

    DG4826-1

    Page16 of 206

    Table 3 Component Adjustments

    BUTTON Description

    [Set Extraction Bar Home]

    Important

    For the procedure, see EXTRACTION BAR HOMEPOSITION.

    Allows you to set and check the home position of theEXTRACTION BAR:

    [FINE NUDGE UP]

    [FINE NUDGE DOWN] [HOOKS UP]

    [HOOKS DOWN]

    [COARSE NUDGE UP]

    [COARSE NUDGE DOWN]

    [SAVE] - sets the selected home position

    [CANCEL] - resets the home position to the last recordedvalue

    [Extraction Bar Positions] Allows you to move the EXTRACTION BAR to: [HOME]

    [UPPER PLATE GUIDE ROLLER]

    [LOWER PLATE GUIDE ROLLER]

    [ERASE]

    [RETURN TO HOME]

    [NUDGE UP]

    [NUDGE DOWN]

    DIAGNOSTICS Using the Diagnostics

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    17/206

    26AUG04

    DG4826-1

    Page17 of 206 [CAM Position Actuation]

    Important

    For the test, see DUPLEX CAM.

    Allows you to: check the status of the CASSETTE HANDLING AY by

    rotating the DUPLEX CAM to:

    [UNPINCH POSITION]

    [HOOKS UP POSITION]

    [PINCH POSITION]

    [UNLATCH POSITION]

    [LATCH POSITION]

    [ADVANCE CASSETTE to SLOT 1]

    [LOAD CASSETTE]

    [EJECT CASSETTE]

    [HOME SLOW SCAN]

    [HOME CAM]

    reset the DUPLEX CAM after jams:

    [CLEAR CASSETTE] [RELEASE CASSETTE]

    [Slow Scan CurrentCalibration] Important

    For the procedure, see Calibration for the CR 950SYSTEM.

    Completes the calibration of the SLOW SCAN MOTOR forthe correct current at all 3 phases of the MOTOR.

    BUTTON Description

    DIAGNOSTICS Using the Diagnostics

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    18/206

    26AUG04

    DG4826-1

    Page18 of 206 [Slow Scan Preload

    Adjustment]Moves the SLOW SCAN AY. This allows you to:

    remove the EXTRACTION BAR

    make an adjustment to the SLOW SCAN AY

    [Upper Transport Mechanism] Allows you to: move the MAIN BELT

    [MOVE number of times]

    [INDEX number of times]

    [Full Revolution]

    [Jog Fine]

    [Jog Coarse]

    move the ELEVATOR up or down

    engage or disengage the LOAD/UNLOAD BAR

    move the SLOW SCAN to the home position

    move the MAIN CAM to the home position

    BUTTON Description

    DIAGNOSTICS Using the Diagnostics

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    19/206

    26AUG04

    DG4826-1

    Page19 of 206

    Table 4 Tests

    BUTTON Description

    [Light Test]

    Important

    For the test, see Light Test.

    Checks for excessive light to the PMTs during the loadingand ejecting of a CASSETTE.

    Allows you to:

    actuate or deactuate the high voltage of the PMTs

    AUTO-ZERO the PMTs

    Checks the signal for noise.

    [Light Test] [Advanced] Checks for excessive light to the PMTs during the loadingof a PLATE.

    Allows you to energize and de-energize:

    GALVO

    LASER DIODE

    ERASE LAMPS

    DIAGNOSTICS Using the Diagnostics

    26AUG04

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    20/206

    26AUG04

    DG4826-1

    Page20 of 206 [Slow Scan Current and

    Velocity Tests] Caution

    An increase in current might indicate that a bind is

    occurring in the SLOW SCAN AY.

    For the test, see Slow Scan Current and VelocityTests.

    Drives the SLOW SCAN AY to the limits in bothdirections.

    Displays for the SLOW SCAN AY:

    speed variation

    current

    Start of Scan

    End of Scan

    [Laser Power]Laser Warning

    Prevent direct exposure to the laser beam.

    For the test, see Laser Power.

    Allows you to:

    check the power of the LASER DIODE set the power values for a new LASER DIODE

    store and restore the laser calibration file

    BUTTON Description

    DIAGNOSTICS Using the Diagnostics

    26AUG04

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    21/206

    26AUG04

    DG4826-1

    Page21 of 206

    BUTTON Description

    [Galvo Test]

    Important

    For the test, see Galvo Test.

    Allows you to: set the Amplitude of the GALVO

    check the response of the LASER DIODE

    [PMT Test]

    Important

    For the test, see PMT Test.

    Checks the functions of the 2 PMTs.

    [Galvo Plot]

    Important

    For the test, see GALVO PLOT.

    Allows you to: check:

    response of the GALVO

    automatic gain control - AGC

    2.5 V reference set the GALVO without an OSCILLOSCOPE

    DIAGNOSTICS Using the Diagnostics

    26AUG04

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    22/206

    26AUG04

    DG4826-1

    Page22 of 206

    Table 5 Logs

    BUTTON Description

    [Error Log]

    Important

    For the procedure, see Checking the Error Logs.

    Allows you to check: error code and description

    date and time

    CASSETTE ID

    scan count

    [Error Frequency Log]

    Important

    For the procedure, see Checking the Error Frequency Log.

    Allows you to: check:

    frequency of each error code

    date and time of last error

    collate error codes by:

    date

    error number

    frequency

    DIAGNOSTICS Using the Diagnostics

    26AUG04

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    23/206

    26AUG04

    DG4826-1

    Page23 of 206 [Actuation Log]

    Important

    For the procedure, see Checking the Actuation Logs.

    Allows you to check: SCAN CYCLES - this value cannot be reset

    BATCH ERASE CYCLES - the number of times that theBatch Erase function was used

    TOTAL POWER-ON HOURS

    LASER DIODE SCAN CYCLES

    LASER HOURS (hours and minutes)

    LASER HOURS LAST CLEARED

    [Export Log Files]

    Important

    For the procedure, see Storing the Error Logs on a Disk.

    Records onto a FLOPPY DISKETTE: error logs

    Actuation Logs

    parameters of the SLOW SCAN

    BUTTON Description

    DIAGNOSTICS Using the Diagnostics

    26AUG04

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    24/206

    DG4826-1

    Page24 of 206 [Clear Error Log]

    Important

    For the procedure, see Clearing the Error Logs.

    Clears all the errors from the error logs.

    Displays the date when the error logs were last cleared.

    [Import Files]

    Important

    If the value is 7000, the home position is not adjusted

    automatically. Set the value manually.

    For the procedure, see Installing Files from a Disk.

    Records files from a FLOPPY DISKETTE onto the CR 950SYSTEM:

    Actuation Logs

    parameters of the SLOW SCAN

    home position of the SLOW SCAN

    BUTTON Description

    DIAGNOSTICS Using the Diagnostics

    26AUG04Table 6 Repetitive Cycles

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    25/206

    DG4826-1

    Page25 of 206

    Table 6 Repetitive Cycles

    BUTTON Description

    [AutoLooper] Automatically processes the CASSETTE to check if it isreliable.

    Completes a test cycle includes scanning and erasing.

    Processes all the CASSETTES on the LOAD BELT.

    Uses all sizes of the CASSETTES.

    Disables the automatic checking Light Leak feature,allowing operation with the PANELS removed.

    Displays status screens for the SENSORS.

    Select [GIVEN NUMBER OF TIMES] to set the numberof times to scan for each CASSETTE.

    [Transport] Cycles the MAIN BELT.

    Uses a CASSETTE to test:

    MAIN BELT operation of loading

    Displays status of the SENSORS at each step.

    DIAGNOSTICS Using the Diagnostics

    26AUG04Table 7 Miscellaneous

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    26/206

    DG4826-1

    Page26 of 206

    Table 7 Miscellaneous

    BUTTON Description

    [Display Software Versions]

    Important

    For the procedure, see Displaying the Software Versions.

    Displays the software version number of: CR 950 SYSTEM

    MASTER CENTRAL PROCESSING UNIT BOARD

    MOTION SYSTEM CONTROL BOARD BOOT

    MOTION SYSTEM CONTROL BOARD

    SLOW SCAN CONTROLLER BOARD BOOT

    SLOW SCAN CONTROLLER BOARD FIXED APP

    SLOW SCAN CONTROLLER BOARD PRODUCT APP

    DIGITIZER BOARD

    MEDICAL IMAGE MANAGER

    DIAGNOSTICS

    [Download to Circuit Boards]

    Important

    For the procedure, see Downloading Software to the

    BOARDS.

    Allows you to download the necessary software to all theBOARDS from the HARD DRIVE.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826 1 BUTTON D i ti

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    27/206

    DG4826-1

    Page27 of 206 [System Self Test]

    Important

    For the test, see System Self Tests

    .Checks the operation of the MASTER CENTRALPROCESSING UNIT BOARD A2 by running the followingtests:Completes the following diagnostic tests:

    GALVO BOARD COMMUNICATION - checks the

    connection of the CABLE between:

    GALVO BOARD A4

    MASTER CENTRAL PROCESSING UNIT BOARD A2

    GALVO REFERENCE - reads the GALVO AGCCHANNEL connected to the reference voltage of theGALVO

    GALVO MOTOR - reads the voltage of the GALVO AGCto check for problems with the GALVO

    GALVO SWEEP - checks the functions of reading andwriting of the REGISTERS on the GALVO BOARD A4

    LASER OFFSET

    PMT BOARD TEST - checks the connection of theCABLE between:

    MASTER CENTRAL PROCESSING UNIT BOARD A2

    PMT/DAS BOARD A5

    MASTER CENTRAL PROCESSING UNIT BOARD

    MEMORY TEST - checks for errors in the reading andwriting of data to each of the image buffers

    BUTTON Description

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826 1 Diagnostic Tests

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    28/206

    DG4826-1

    Page28 of 206

    Diagnostic Tests

    Light Test

    1 On the System Service Diagnostic screen, touch [Light Test].

    2 Touch [TEST].

    3 Wait for the test to complete.

    4 Check that PASS displays in AUTO-ZERO.

    5 Touch [START GRAPH].

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    29/206

    DG4826-1

    Page29 of 206

    6 Check that the value is approximately 4 - 5. Use [ZOOM IN].

    7 Touch [HV ON].

    8 Check that the value increases to approximately 50 - 51.

    9 Select:

    [ZOOM OUT]

    [CLEAR/STOP]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1 Slow Scan Current and Velocity Tests

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    30/206

    DG4826 1

    Page30 of 206

    Slow Scan Current and Velocity Tests

    1 On the System Service Diagnostic screen, touch [Slow Scan Current and VelocityTests].

    2 Insert the 35 x 43 CASSETTE into the CR 950 SYSTEM.

    3 Touch:

    [LOAD 35 x 43 CASSETTE]

    [START CURRENT TEST]

    4 Wait for the test to complete.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    31/206

    Page31 of 206

    5 Check that the value is:

    5 -180 between Start of Scan and End of Scan

    < 660 between End of Scan and End of Travel

    no spikes > 30

    6 If the results are not correct, check the error logs. Use Checking the Error Logs.

    7 Touch [START VELOCITY TEST].

    8 Wait for the test to complete.

    9 Check that the results between Start of Scan and End of Scan are below 90 andstable.

    10 If the results are not correct, check the error logs. Use Checking the Error Logs.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1 11 Touch [EJECT THE CASSETTE].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    32/206

    Page32 of 206 12 Remove the CASSETTE.

    Laser Power

    Laser Warning

    Prevent direct exposure to the laser beam.

    1 On the System Service Diagnostic screen, touch [Laser Power].

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1 2 Touch:

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    33/206

    Page33 of 206 [LASER ON]

    [READ]

    3 Check Laser Power is 28 - 32 mW.

    4 Is the value correct?

    Galvo Test

    1 On the System Service Diagnostic screen, touch [Galvo Test].

    Yes No

    Continue with the next test. 1. Check the connections to theLASER DRIVER BOARD A17.

    2. Install a new OPTICAL

    MODULE.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    P

    2 Touch:

    http://ar4826_1.pdf/http://ar4826_1.pdf/http://ar4826_1.pdf/http://ar4826_1.pdf/
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    34/206

    Page34 of 206 [START GALVO]

    [LASER ON]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    35/206

    Page35 of 206

    Laser Warning

    Prevent direct exposure to the laser beam.

    3 Open the TOP COVER.

    4 Insert one sheet of paper into the CR 950 SYSTEM between:

    FOLD MIRROR

    COLLECTOR

    5 Observe the laser beam on the sheet of paper.

    6 Touch [-100] until the Set Amplitude value is 400.

    H196_1801HC

    TOP COVER

    H196_1801HCA

    COLLECTOR

    paper

    laser beam FOLD MIRROR

    CR 950 SYSTEM

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page

    7 Observe the laser beam on the sheet of paper.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    36/206

    Page36 of 206 8 Touch [+100] until the Set Amplitude value is 3000.

    9 Observe the laser beam on the sheet of paper.

    10 Does the laser beam change size correctly?

    PMT Test

    1 On the System Service Diagnostic screen, touch [PMT Test].

    2 Touch [START PMT TEST].

    3 Check that PASSED displays for both PMT 1 and PMT 2.

    Set Amplitude Laser Beam

    400 short

    3000 long

    Yes No

    Continue with the next test. Check the Error Codes for the

    cause of the malfunction.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page

    4 If one of the PMTs displays FAILED, install a new PMT. See PMTs and COLLECTORAY

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    37/206

    Page37 of 206

    AY.

    5 If both PMTs display FAILED, check the connections between:

    PMT BOARD

    DAS BOARD

    GALVO PLOT

    Important

    The Galvo Plot screen has more Cassette Size options with Software Version 4.1.

    1 On the System Service Diagnostic screen, touch [Galvo Plot].

    2 Touch [START GALVO PLOT].

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    38/206

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page Normal GALVO PLOT

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    39/206

    g39 of 206

    4 Check the values:

    5 Are the values correct?

    GALVO PLOT normal

    AGC 10.0 V 0.1

    2.5 vRef 2.5 0.1

    Yes No

    Continue with the next test. Adjust the GALVO BOARDuntil theGALVO PLOT is normal.

    GALVO PLOT with low HOOK GALVO PLOT with excessive HOOK

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page

    System Self Tests

    1 On the System Service Diagnostic screen touch [System Self Test]

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    40/206

    40 of 206 1 On the System Service Diagnostic screen, touch [System Self Test].

    2 Touch [START MCPU SELF TEST].

    3 Wait approximately 20 seconds for the test to complete.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page41 f 206

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    41/206

    41 of 206

    4 Observe the results for each test.

    5 If any of the tests display FAILED, check the error logs. Use Checking the Error Logs.

    6 See the Error Codesto determine the possible cause and action for the problem.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page42 of 206

    Diagnostic Procedures

    Checking the Error Logs

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    42/206

    42 of 206 Checking the Error Logs

    1 On the System Service Diagnostic screen, touch [Error Log].

    2 Touch either:

    [Sort By Date],

    [Sort By Error Code], or

    [Sort By Cassette ID]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page43 of 206

    Checking the Error Frequency Log

    1 On the System Service Diagnostic screen, touch [Error Frequency Log].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    43/206

    43 of 206 1 On the System Service Diagnostic screen, touch [Error Frequency Log].

    2 Touch either: [Sort By Date],

    [Sort By Error Code], or

    [Sort Error Frequency]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page44 of 206

    Checking the Actuation Logs

    1 On the System Service Diagnostic screen, touch [Actuation Log].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    44/206

    44 of 206 y g , [ g]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page45 of 206

    Storing the Error Logs on a Disk

    1 On the System Service Diagnostic screen, touch [Export Log Files].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    45/206

    y g [ p g ]

    2 Insert a disk into the DRIVE.

    3 Touch either:

    [COPY ERROR LOGS TO FLOPPY],

    [COPY ACTUATION LOGS TO FLOPPY], or

    [COPY SLOWSCAN PARAMETERS FILE TO FLOPPY]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page46 of 206

    Clearing the Error Logs

    1 On the System Service Diagnostic screen, touch [Clear Error Log].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    46/206

    2 Touch [CLEAR ERROR LOG].

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page47 of 206

    Installing Files from a Disk

    1 On the System Service Diagnostic screen, touch [Import Files].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    47/206

    2 Insert the disk into the DRIVE.

    3 To install the ACTUATION LOGS to the CR 950 SYSTEM, touch [COPY ACTUATIONLOGS FROM FLOPPY].

    Important

    If the value is 7000, the home position is not adjusted automatically. Set the value manually.

    4 To install the parameters of the SLOW SCAN, touch:

    [GET SLOWSCAN HOME POSITION FROM FLOPPY]

    [UPDATE SLOWSCAN PARAMETER FILE FROM FLOPPY]

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page48 of 206

    Displaying the Software Versions

    1 On the System Service Diagnostic screen, touch [Display Software Versions].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    48/206

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page49 of 206

    2 Check that each BOARD has the correct communication:

    Message Description Action

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    49/206

    The version numberis in the field.

    The BOARD isResponding.

    Check that the newest and correctversion of the software is installed.

    All fields are blank. The diagnostics is notoperating. 1. Do a shutdown of the computer.2. De-energize the CR 950

    SYSTEM.

    3. Energize the CR 950 SYSTEM.

    Not Responding

    Important

    The message is normalduring initialization.

    Failure of communicationsto the BOARD.

    1. Check the connections to theBOARD.

    2. Download the software to theBOARD.

    3. If you cannot successfullydownload the software to theBOARD, install a new BOARD.

    Not Reporting The BOARD has a failure of

    communication becauseanother BOARD is NotResponding.

    1. Check the BOARD that is Not

    Responding.

    2. Do the test again.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page50 of 206

    Downloading Software to the BOARDS

    1 On the System Service Diagnostic screen, touch [Download to Circuit Boards].

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    50/206

    Caution

    Downloading the SSC BOOT can cause damage to the SLOW SCAN CONTROLLERBOARD A6. Only download the SSC BOOT if necessary.

    DIAGNOSTICS Using the Diagnostics

    26AUG04

    DG4826-1

    Page51 of 206

    2 Touch either:

    [DOWNLOAD MCPU APPLICATION],

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    51/206

    [DOWNLOAD MSC APPLICATION],

    [DOWNLOAD SSC BOOT and APPLICATIONS],

    [DOWNLOAD SSC APPLICATIONS], or

    [DOWNLOAD SSC PARAMETERS]

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page52 of 206

    Section 2: Error Codes

    Overview

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    52/206

    Overview

    The first 2 digits of the error code identifies the subsystem or component with the error.Example: 10003

    The 1 and the 0 indicate that the error occurred in the subsystem BAR CODE.

    1 Use Table 8to help identify the component with the error.

    2 Advance to the error code to diagnose the error.

    Table 8 Subsystem Identification

    ID No. Subsystem

    10 BAR CODE

    11 LATCH/UNLATCH CAM

    12 SLOW SCAN AY

    13 ERASE LAMPS

    14 POWER SUPPLY15 DATA PATH

    16 Calibration

    18 UPPER CONTROL

    19 Self Test for the MASTER CENTRAL PROCESSING UNIT

    22 - 32 MOTION SYSTEM CONTROL BOARD A1

    Windows2000 Subsystem

    47 Shutdown Management

    49 DISK MANAGER

    50 DICOM

    55 Installation

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page53 of 206

    BAR CODE

    10003 - READER did not read the CASSETTE BAR CODE

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    53/206

    Cause 1: The CASSETTE has a bar code that you cannot read or it does not have a barcode.

    Action: Install the correct bar code.

    Cause 2: The internal BAR CODE READER malfunctioned.

    Action: 1. Observe that the BAR CODE READER is not blocked.

    2. Check the connections to the BAR CODE READER.

    3. If the BAR CODE READER emits a beep when you insert a CASSETTE, but

    the CASSETTE does not load, check CONNECTOR RJ45 between the BARCODE READER and the MASTER CENTRAL PROCESSING UNIT BOARDA1.

    4. If the BAR CODE READER does not emit a beep when you load aCASSETTE, check CONNECTOR RJ45 between the BAR CODE READERand the MOTION SYSTEM CONTROL BOARD A1.

    5. Open the FRONT DOOR.6. Close the FRONT DOOR to initialize the MOTION SYSTEM CONTROL

    BOARD A1.

    7. Touch:

    [Diagnostics]

    [Component Control: Step] [Read CASSETTE ID]

    8. If the CASSETTE ID does not display, do again with another CASSETTE.

    9. If the CASSETTE ID does not display, install a new BAR CODE READER.

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page54 of 206

    LATCH/UNLATCH CAM

    11001 - DUPLEX CAM did not move to the Home Position

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    54/206

    11002 - DUPLEX CAM did not move to PINCH position

    11003 - DUPLEX CAM did not move to CLAMP/UNLATCH position

    11004 - DUPLEX CAM did not move to UNCLAMP/LATCH position

    11005 - DUPLEX CAM did not move to UNPINCH position

    11006 - CAM OUTER SENSOR did not detect position of DUPLEX CAM

    11007 - CAM INNER SENSOR did not detect Home Position for DUPLEXCAM

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page55 of 206

    11009 - DUPLEX CAM: Movement failure

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    55/206

    11010 - DUPLEX CAM: Software command failure on the MSC BD

    11014 - DUPLEX CAM: Timeout Error

    11015 - Move command failure because SLOW SCAN not at home position

    11016 - MSC BOARD A1: Software Error for DUPLEX CAM

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Description: The DUPLEX CAM received a command to move that was not correct.

    Cause 1: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The DUPLEX CAM did not move in the time allowed.

    Cause: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause 1: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The MOTION SYSTEM CONTROL BOARD A1 did not provide the RTXCTIMER. The TIMER determines the timeouts for the motion of the

    DUPLEX CAM.

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page56 of 206

    11017 MSC BOARD A1 S f E f DUPLEX CAM

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    56/206

    11017 - MSC BOARD A1: Software Error for DUPLEX CAM

    11018 - DUPLEX CAM did not move to next position

    11019 - DUPLEX CAM did not move to PINCH position

    11020 - DUPLEX CAM did not move to UNLATCH position

    Description: The MOTION SYSTEM CONTROL BOARD A1 did not set up thecommand to move for the DUPLEX CAM.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The DUPLEX CAM malfunctioned.Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause 1: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

    Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 1: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause 2: The DUPLEX CAM malfunctioned.

    Action: Do the checkout for the CASSETTE HANDLING AY.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    57/206

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page58 of 206

    SLOW SCAN AY

    12001 - SLOW SCAN EXTRACTION BAR did not move to home position

    Cause 1: The EXTRACTION BAR malfunctioned

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    58/206

    12002 - EXTRACTION BAR REFERENCE SENSOR S9 did not actuate

    Cause 1: The EXTRACTION BAR malfunctioned.

    Action: 1. Remove any obstructions in the path of the EXTRACTION BAR.

    2. Check the SLOW SCAN MOTOR.

    3. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    4. Check the EXTRACTION BAR HOME POSITION.

    Cause 2: The EXTRACTION BAR REFERENCE SENSOR S9 malfunctioned.

    Action: Do the checkout for the EXTRACTION BAR SENSOR S9.

    Cause 1: The EXTRACTION BAR REFERENCE SENSOR S9 operates correctly, butanother malfunction occurred.

    Action: 1. Remove any obstructions from the path of the EXTRACTION BAR.

    2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 2: The EXTRACTION BAR REFERENCE SENSOR S9 malfunctioned.

    Action: Do the checkout for the EXTRACTION BAR SENSOR S9.

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page59 of 206

    12003 - PLATE PRESENT SENSOR S5 did not detect PLATE

    Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunctionoccurred.

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    59/206

    Action: 1. Check:

    a. PLATE was removed from the CASSETTEb. SLOW SCAN MOTOR operates correctly

    2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    3. Remove any obstructions from the path of the EXTRACTION BAR.

    Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.Action: 1. Check the status of the PLATE PRESENT SENSOR S5 in the diagnostics.

    2. Check that LED DS29 changes status on the MOTION SYSTEM CONTROLBOARD A1.

    3. Check the power to the PLATE PRESENT SENSOR S5.

    4. Check the alignment of the EMITTER and DETECTOR on the PLATEPRESENT SENSOR S5.

    DIAGNOSTICS Error Codes26AUG04

    DG4826-1

    Page60 of 206

    12004 - PLATE PRESENT SENSOR S5 did not actuate

    Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunctionhas occurred.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    60/206

    Action: 1. Check:

    a. PLATE was removed from the CASSETTEb. SLOW SCAN MOTOR operates correctly

    2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    3. Remove any obstructions from the path of the EXTRACTION BAR.

    Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.Action: 1. Check the status of the PLATE PRESENT SENSOR S5 in the diagnostics.

    2. Check that LED DS29 changes status on the MOTION SYSTEM CONTROLBOARD A1.

    3. Check the power to the PLATE PRESENT SENSOR S5.

    4. Check the alignment of the EMITTER and DETECTOR of the PLATEPRESENT SENSOR S5.

    Cause 3: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page61 of 206

    12009 - Caution! SLOW SCAN MOTOR drawing too much power

    Cause: A malfunction occurred in one or more of the following components: COIL BOARD

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    61/206

    12012 - SLOW SCAN AY: Timeout Error

    12013 - Software Error in SLOW SCAN Caused Error in MSC BD

    EXTRACTION BAR REFERENCE SENSOR S9

    SLOW SCAN AYAction: 1. Check for an obstruction.

    2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The PLATE did not move within the correct time.

    Cause: A malfunction occurred in one or more of the following components: COIL BOARD A7

    SLOW SCAN AY

    Action: 1. Check for an obstruction.

    2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: A command for the PLATE that is not correct was made by the systemsoftware.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: 1. Load the software for the MOTION SYSTEM CONTROL BOARD A1.

    2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page62 of 206

    12014 - MSC BOARD A1: Software Error for SLOW SCAN

    Description: This error might occur during the diagnostics if the service person startsto move the PLATE before the PLATE stops moving.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    62/206

    12100 - SSC BOARD A6: Software Error

    12101 - SLOW SCAN had a velocity malfunction during scan

    Cause: The SLOW SCAN AY malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The SLOW SCAN AY will not move.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: 1. De-energize the CR 950 SYSTEM.

    2. Energize the CR 950 SYSTEM.

    3. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The SLOW SCAN AY exceeded the speed limit when moving the SCANPLATE.

    Cause 1: The speed of the SLOW SCAN exceeded the limit. A band might appear on theimage.

    Action: Do the Slow Scan Current and Velocity Tests.

    Cause 2: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    63/206

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page64 of 206

    12108 - SLOW SCAN stopped during scan

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    64/206

    12109 - SLOW SCAN stopped during scan

    12110 - SLOW SCAN did not communicate with the MSC

    12111 - SLOW SCAN did not move

    12112 - SLOW SCAN did not move

    BOARD A7.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The SLOW SCAN AY did not move because of a communication errorwith the SLOW SCAN CONTROLLER BOARD A1.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page65 of 206

    12113 - SSC BOARD A6: NVRAM malfunctioned

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    65/206

    12115 - SSC BOARD A6: Software Error

    12116 - SSC BOARD A6 did not communicate with MSC

    12120 - SLOW SCAN did not move

    BOARD A7.

    Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The SLOW SCAN AY did not stop moving before the next command tomove. This error indicates a communication error with the SLOW SCANCONTROLLER BOARD A6.

    Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page66 of 206

    12121 - SLOW SCAN did not move

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    66/206

    12122 - SLOW SCAN COIL BOARD A7 malfunctioned

    12123 - SSC BOARD A6: Software Error

    12124 - SSC BOARD A6: Software Error

    12125 - SLOW SCAN COIL BOARD A7 malfunctioned

    BOARD A7.

    Description: The SLOW SCAN AY stopped or did not move at all.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The MOTOR CONTROLLER for the SLOW SCAN AY malfunctioned.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: An error occurred during the initialization the SLOW SCAN AY.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL

    BOARD A7.

    Description: The SLOW SCAN AY stopped or did not move at all.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page67 of 206

    12126 - Caution! SLOW SCAN MOTOR is overheated

    Description: The SLOW SCAN MOTOR was using excessive power and is too hot.

    Cause: Malfunction in one or more of:COIL BOARD A7

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    67/206

    12130 - SSC BOARD A6: Software Error

    12131 - SSC BOARD A6 did not communicate with MSC

    12132 - SSC BOARD A6 did not communicate with MSC

    COIL BOARD A7

    EXTRACTION BAR REFERENCE SENSOR S9

    SLOW SCAN AY

    Action: 1. Check for an obstruction.

    2. Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 2: The MOTION SYSTEM CONTROLLER BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page68 of 206

    12201 - SSC BOARD A6: Programming Error

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    68/206

    12202 - 12260 - SSC BOARD A6: Memory Error

    BOARD A7.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Error

    Code Description

    12202 Error occurred during downloading to the SLOW SCAN CONTROLLER BOARD A6.

    12203 Memory error on the SLOW SCAN CONTROLLER BOARD A6.

    12204 Memory error on the SLOW SCAN CONTROLLER BOARD A6.

    12216 Communication error between the SLOW SCAN CONTROLLER BOARD A6 and theMOTION SYSTEM CONTROL BOARD A1.

    12217 Communication error between the SLOW SCAN CONTROLLER BOARD A6 and theMOTION SYSTEM CONTROL BOARD A1. Error EM17 - Receive Buffer Overflow.

    12218 Error during downloading of code or parameters for the SLOW SCANCONTROLLER BOARD A6.

    12219 Error during downloading of code or parameters for the SLOW SCANCONTROLLER BOARD A6.

    12220 Error EM20 - New Bootstrap Program, Download all Boot and Program Modules.12221 Malfunction of the SLOW SCAN CONTROLLER BOARD A6.

    12222 Checksum error occurred during downloading of code to the SLOW SCANCONTROLLER BOARD A6.

    12223 Checksum error occurred during downloading of code to the SLOW SCANCONTROLLER BOARD A6.

    12228 Error occurred during downloading of the code to the SLOW SCAN CONTROLLERBOARD A6.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page69 of 206

    12230 Error EM30 - Host Flash Download Copy Boot Memory Error.

    12240 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM40 Illegal Record T pe

    Error

    CodeDescription

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    69/206

    Error EM40 - Illegal Record Type.

    12241 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM41 - Invalid Extended Address.

    12242 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM42 - Invalid Erase Block ID.

    12243 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM43 - Receive Record Checksum Error.

    12244 Error when initializing. Error EM44 - File Sum of Record Checksum Error, DownloadProgram.

    12245 Program error. Error EM45 - Boot Checksum Write Error.

    12246 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM46 - FIX_APP Code Block Checksum Write Error.

    12247 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM47 - Code Page Checksum Write Error.

    12248 Error EM48 - Parameters Save Flash Error.

    12250 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM50 - Flash Write Address Out Of Range.

    12251 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM51 - Flash Program Error: Erase.

    12252 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM52 - Flash Program Error: Low Voltage.

    12253 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM53 - Flash Program Error: Program.

    12254 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM54 - Flash Program Error: Sequence.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page70 of 206

    12255 Error occurs during downloading to the SLOW SCAN CONTROLLER BOARD A6.Error EM55

    12260 Error recording the parameters to the SLOW SCAN CONTROLLER BOARD A6

    Error

    CodeDescription

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    70/206

    12896 - SSC BOARD A6: Unexpected Reset

    12901 - SSC BOARD A6: Unexpected Retry

    12902 - SSC BOARD A6: Unexpected Retry

    12260 Error recording the parameters to the SLOW SCAN CONTROLLER BOARD A6.

    Error EM60 - Parameters Save Flash Error.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The PLATE was not scanned. The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The SLOW SCAN AY did not move to the ERASE position. Thecommand is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page71 of 206

    12903 - SSC BOARD A6: Unexpected Retry

    Description: The SLOW SCAN AY did not move to the home position. The commandis executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    71/206

    12904 - SSC BOARD A6: Unexpected Retry

    12905 - SSC BOARD A6: Unexpected Retry

    12906 - SSC BOARD A6: Unexpected Retry

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The COARSE NUDGE UP command did not occur. The command isexecuted again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COIL

    BOARD A7.

    Description: The COARSE NUDGE DOWN command did not occur. The command is

    executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The FINE NUDGE UP command did not occur. The command isexecuted again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page72 of 206

    12907 - SSC BOARD A6: Unexpected Retry

    Description: The COARSE NUDGE DOWN command did not occur. The command isexecuted again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    72/206

    12908 - SSC BOARD A6: Unexpected Retry

    12911 - SSC BOARD A6: Unexpected Retry

    12913 - SSC BOARD A6: Unexpected Retry

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The PLATE did not move to the CASSETTE. The command is executedagain.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The SOFT RESET command did not occur. The command is executed

    again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The EXTRACTION BAR did not move to the mechanical adjustmentposition one. The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page73 of 206

    12914 - SSC BOARD A6: Unexpected Retry

    Description: The command MOVE TO EXTRACTION BAR REFERENCE SENSORdid not occur. The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    73/206

    12915 - SSC BOARD A6: Unexpected Retry

    12930 - SLOW SCAN: Unexpected Retry during Current Test

    12931 - SLOW SCAN: Unexpected Retry during Velocity Test

    Cause e S O SC CO O O 6 a u ct o ed

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The command MOVE TO THE PLATE GUIDE POSITION did not occur.The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The system was not able to move to the position selected in the

    diagnostics. The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The system was not able to move to the position selected in thediagnostics. The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page74 of 206

    12941 - Unexpected Retry during Communication to MSC BD

    Description: A message could not be sent to the SLOW SCAN CONTROLLERBOARD A6. The command is executed again.

    Cause: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    74/206

    ERASE LAMPS

    13001 - 13008 - ERASE LAMP PAIRS

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Code Log Message13001 ERASE LAMP PAIR No. 1 malfunctioned

    13002 ERASE LAMP PAIR No. 2 malfunctioned

    13003 ERASE LAMP PAIR No. 3 malfunctioned

    13004 ERASE LAMP PAIR No. 4 malfunctioned

    13005 ERASE LAMP PAIR No. 5 malfunctioned

    13006 2 ERASE LAMP PAIRS malfunctioned

    13007 3 ERASE LAMP PAIRS malfunctioned

    13008 4 ERASE LAMP PAIRS malfunctioned

    Cause: One or more ERASE LAMP PAIRS malfunctioned.

    Action: Install a new ERASE LAMP PAIR.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page75 of 206

    13009 - All ERASE LAMPS malfunctioned

    Cause 1: The power to the ERASE LAMPS malfunctioned.

    Action: 1. De-energize the ERASE LAMPS.

    2. Check for 120 V AC across the AC side of RELAY K1.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    75/206

    3. If 120 V AC is not indicated at RELAY K1, check:

    ERASE LAMP AY is fully closed

    FUSE F1

    TERMINAL BLOCK TB1

    condition of the CONNECTOR P1/J1 on the ERASE LAMP AY

    condition of the INTERLOCK SWITCH S8

    wires to the POWER SUPPLY PS1

    4. If 120 V AC is indicated on RELAY K1, connect the MULTIMETER to the ACside of the RELAY.

    5. Energize and de-energize the ERASE LAMPS. With the LAMPS energized, you

    should measure < 2 V AC.6. If the 120 V AC does not change to a Lo signal, check the DC side of RELAY

    K1:

    connect the MULTIMETER TL-4114 negative to the violet wire

    connect the MULTIMETER TL-4114 positive to the blue wire

    check for 4.5 - 5.0 V DC on K1 with the ERASE LAMPS energized7. Check the CONNECTOR J10 on the MOTION SYSTEM CONTROL BOARD

    A1.

    Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page76 of 206

    13010 - ERASE LAMP PAIRS: Timeout Error

    13011 - Command to Erase did not occur

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    76/206

    13011 - Command to Erase did not occur

    13012 - MSC BOARD A1: Invalid Command for ERASE LAMPS

    13013 - Failure to allocate RTXC timer during Erase Process

    13899 - General Erase Error

    POWER SUPPLY

    14001 - SLOW SCAN: Malfunction of 24 V Power

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The software did not provide the RTXC TIMER during the ERASEfunction.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The 24 V DC CIRCUIT BREAKER of the SLOW SCAN is open.

    Cause: Malfunction of the 24 V DC power distribution.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    77/206

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page78 of 206

    14005 - Front DOOR Open

    Cause 1: The INTERLOCK SWITCH S8 malfunctioned.

    Action: 1. Check the wires to the INTERLOCK SWITCH CONNECTOR P5/J5.

    2. Install a new INTERLOCK SWITCH S8.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    78/206

    14006 - DUPLEX CAM: Initialization Error

    Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The PLATE PRESENT SENSOR S5 is blocked when the SLOW SCANAY is in the home position.

    Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunctionhas occurred. This error occurs when initializing, either when the CR 950 SYSTEMis energized or when a mechanical subsystem initializes.

    Action: Check: PLATE was removed from the CASSETTE

    SLOW SCAN MOTOR operates correctly

    obstruction in the path of the EXTRACTION BAR

    Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.

    Action: Check: status of the PLATE PRESENT SENSOR S5 in the diagnostics

    LED DS29 changes status on the MOTION SYSTEM CONTROL BOARD A1

    power to the PLATE PRESENT SENSOR S5

    alignment of the EMITTER and DETECTOR of the PLATE PRESENT SENSORS5

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    79/206

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page80 of 206

    14012 - UPPER CONTROL INTERLOCK was pressed while the CASSETTEwas advancing

    Cause: The DEAD STOP INTERLOCK S24 actuated.

    Action: Check: operation of the DEAD STOP INTERLOCK S24

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    80/206

    14013 - Initialization error at the UPPER CONTROL AY

    14019 - Invalid Software Version detected for this Model

    DATA PATH

    15001 - MCPU BOARD A2: Fatal Error in Software

    15002 - MCPU BOARD A2: Fatal Error in Software

    operation of the DEAD STOP INTERLOCK S24

    SENSOR S22 in SLOT 8 of the UPPER CONTROL AY

    Cause: A CASSETTE is in SLOT 0 or SLOT 8 when the CR 950 SYSTEM is energized.

    Action: Remove the CASSETTE.

    Cause: The MSC BOARD has reported a model number that is not consistent with thecurrent version of MCPU BOARD software.

    Action: Download the correct version of MCPU and MSC BOARD software. Do theprocedure Downloading Software to the BOARDS.

    Description: A PSOS system error occurred that is not possible to repair. The EventLog includes the PSOS system error number, module, and line number.

    Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Description: A system software error occurred. The Event Log will include the PSOSsystem error number, module, and line number.

    Cause: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    81/206

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page82 of 206

    15006 - Communications Error between MCPU BD A2 and GALVO BD A4

    Description: The process of the DMA for the MASTER CENTRAL PROCESSINGUNIT BOARD A2 did not complete within the expected time.

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    82/206

    15007 - Communications Error between MCPU BD A2 and GALVO BD A4

    15008 - Start of Scan Timeout

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.Cause 2: The DIGITIZER BOARD A3 malfunctioned.

    Action: Do the checkout for the DIGITIZER BOARD A3.

    Cause 3: The GALVO BOARD A4 malfunctioned.

    Action: Do the checkout for the GALVO BOARD A4.

    Description: The process of the DMA for the MASTER CENTRAL PROCESSINGUNIT BOARD A2 did not complete within the expected time.

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Cause 2: The DIGITIZER BOARD A3 malfunctioned.

    Action: Do the checkout for the DIGITIZER BOARD A3.

    Cause 3: The GALVO BOARD A4 malfunctioned.

    Action: Do the checkout for the GALVO BOARD A4.

    Description: A timeout occurred for the START OF SCAN.

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Cause 2: The DIGITIZER BOARD A3 malfunctioned.

    Action: Do the checkout for the DIGITIZER BOARD A3.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page83 of 206

    15012 - Communications Error between MCPU BOARD A2 and NT

    Cause 3: The GALVO BOARD A4 malfunctioned.

    Action: Do the checkout for the GALVO BOARD A4.

    Description:Windows2000 did not return a response to a message within a given

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    83/206

    time.

    Cause 1: The TCP/IP connection malfunctioned.

    Action: Check the TCP/IP connection to the computer. See Step 5, Page 11.

    Cause 2: Failure of the Windows2000 software.

    Action: De-energize and energize the CR 950 SYSTEM.

    Cause 3: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page84 of 206

    15013 - Image Quality Warning

    Description: The image quality might be wrong. The MASTER CENTRAL PROCESSINGUNIT BOARD A2 received a system error from the MOTION SYSTEMCONTROL BOARD A1 and was automatically repaired. If a CASSETTE was inthe CR 950 SYSTEM when the error occurred, the TOUCH SCREEN displays

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    84/206

    15014 - LASER: Calibration did not occur

    15015 - PLATE: Size calibration did not occur

    a message about a problem with the image.Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause 3: A light leakage occurred.

    Action: Do the Light Test.

    Cause 4: The INTERLOCK SWITCH S8 malfunctioned. The DOOR might be open duringscan.

    Action: Check the INTERLOCK SWITCH S8.

    Cause: The LASER DIODE malfunctioned.

    Action: Do the calibration for the LASER DRIVER BOARD A17- use Optics Module.

    Cause: The CR 950 SYSTEM does not have the calibration data for all sizes of PLATES.Action: 1. Do the Calibration for the CR 950 SYSTEM.

    2. Do the checkout MASTER CENTRAL PROCESSING UNIT BOARD A2.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page85 of 206

    15016 - Laser power high

    Cause: The LASER DIODE in the DETECTOR is backward.

    Action: Check: LASER REFERENCE VOLTAGE in the PMT/DAS BOARD A5

    LASER DIODE is installed correctly

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    85/206

    15017 - PLATE: Calibration did not occur

    15018 - There was an error reading or writing to the PCMCIA card

    Important

    A malfunction of the PCMCIA CARD is the last possible cause. Check all othermalfunctions before installing a new PCMCIA CARD.

    15040 - Communications Error Between MCPU and the PC

    Cause: The customer scanned a PLATE of a size that was not recognized.

    Action: 1. Do the Calibration for the CR 950 SYSTEM.

    2. Use the diagnostics to complete the size calibration of the PLATE.

    Cause 1: The PCMCIA CARD operates correctly, but another malfunction occurred.

    Action: Check the error logs for the malfunction. Use Checking the Error Logs.

    Cause 2: The PCMCIA CARD malfunctioned.

    Action: Install a new PCMCIA CARD.

    Description: The TCP/IP SOCKET COMMUNICATIONS were not completed or didnot occur between the MASTER CENTRAL PROCESSING UNITBOARD A2 and the host computer with Windows2000.

    Cause 1: The shutdown of the CR 950 SYSTEM was not completed in the correctsequence.

    Action: Use the correct shutdown sequence to de-energize the CR 950 SYSTEM.

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    86/206

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page87 of 206

    Calibration

    16001 - Calibration Error: Did not find START OF SCAN position

    Description: During the calibration, the CR 950 SYSTEM did not locate the STARTSCAN position within the given time.

    C 1 Th f th PLATE i t t

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    87/206

    16002 - Calibration Error: Did not find Offset

    Cause 1: The exposure of the PLATE is not correct.Action: Use the DOSIMETER to check exposure of the PLATE.

    Cause 2: Calibration of the CR 950 SYSTEM is not correct.

    Action: Do the Calibration for the CR 950 SYSTEM.

    Cause 3: The alignment of the optics is not correct.

    Action: Check the adjustment: FOLD MIRROR

    GALVO BOARD

    Cause 4: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Cause 5: The PMT/DAS BOARD A5 malfunctioned.

    Action: Do the checkout for the SLOW SCAN CONTROLLER BOARD A6 and COILBOARD A7.

    Description: The GALVO did not detect the Offset.Cause 1: The exposure of the PLATE is not correct.

    Action: Use the DOSIMETER to check the exposure of the PLATE.

    Cause 2: The GALVO is not aligned correctly.

    Action: Do:

    adjustment of the GALVO BOARD checkout for the GALVO BOARD A4

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page88 of 206

    16003 - Calibration Error: Did not find Amplitude

    Cause 3: The PLATE has scratches on the edges.

    Action: Check the PLATE for scratches and scan another PLATE.

    Check the edges and the degree of reflection.

    D i ti Th GALVO did t d t t th A lit d

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    88/206

    16005 - Calibration Error: Routine did not receive image in the timeallowed

    Description: The GALVO did not detect the Amplitude.

    Cause 1: The exposure of the PLATE is not correct.

    Action: Use the DOSIMETER to check the exposure of the PLATE.

    Cause 2: The GALVO is not aligned correctly.

    Action: Do:

    adjustment of the GALVO BOARD

    checkout for the GALVO BOARD A4.

    Cause 3: The PLATE has scratches on the edges.

    Action: Check: PLATE for scratches and scan another PLATE

    edges and the degree of reflection

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Cause 2: The DIGITIZER BOARD A3 malfunctioned.

    Action: Do the checkout for the DIGITIZER BOARD A3.

    Cause 3: The GALVO BOARD A4 malfunctioned.

    Action: Do the checkout for the GALVO BOARD A4.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page89 of 206

    16006 - Calibration Error: Did not obtain image

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Cause 2: The DIGITIZER BOARD A3 malfunctioned.

    Action: Do the checkout for the DIGITIZER BOARD A3

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    89/206

    16010 - Possible Light Leakage

    16011 - Imaging System Error occurred during Initialization

    Action: Do the checkout for the DIGITIZER BOARD A3.Cause 3: The GALVO BOARD A4 malfunctioned.

    Action: Do the checkout for the GALVO BOARD A4.

    Description: Large quantities of light occurred during the AUTO-ZERO.

    Cause: The DOOR is not closed, a PANEL is loose, or a light leakage occurred in theCASSETTE HANDLING SYSTEM.

    Action: 1. Do the Light Testto locate the source of the light leakage.

    2. Repair the light leakage.

    Cause: The PMT/DAS BOARD A5 malfunctioned.

    Action: 1. Check the CABLE between the PMT BOARD and the DAS BOARD.

    2. Do:

    checkout for the GALVO BOARD A4

    System Self Tests

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page90 of 206

    16012 - PMT/DAS BOARD A5: Calculation Error occurred duringInitialization

    Description: The value for one or more of the following parameters did not match thespecifications when the CR 950 SYSTEM initialized: Fine or CoarseOffset or Fine or Coarse Ratio.

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    90/206

    16013 - Imaging System contains noise beyond the limit

    Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.

    Action: Do the checkout for the MASTER CENTRAL PROCESSING UNIT BOARD A2.

    Cause 2: The PMT/DAS BOARD A5 malfunctioned.

    Action: Do the checkout for the PMT/DAS BOARD A5.

    Description: The peak-to-peak signal was high when the CR 950 SYSTEM was setto AUTO-ZERO.

    Cause 1: The PMT/DAS BOARD A5 malfunctioned.

    Action: Do the checkout for the PMT/DAS BOARD A5.

    Cause 2: Noise in the POWER SUPPLY PS1.Action: Do the checkout for the POWER SUPPLY PS1.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page91 of 206

    16014 - Imaging System did not zero or calibration of the PMT did notoccur

    Cause 1: The DOSIMETER malfunctioned.

    Action: Check that the DOSIMETER operates correctly.

    Cause 2: The exposure on the PLATE is not correct during calibration

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    91/206

    16016 - Malfunctioning PMT

    16019 - Calibration could not find offset, rotate GALVO counterclockwise

    16020 - Calibration could not find offset, rotate GALVO clockwise

    Cause 2: The exposure on the PLATE is not correct during calibration.Action: Do the calibration with another PLATE.

    Cause 3: The X-RAY TUBE malfunctioned.

    Action: Expose the PLATE using an X-RAY TUBE you did not use before.

    Cause 4: The Imaging System did not auto zero.

    Action: Do the checkout for: PMT/DAS BOARD A5.

    GALVO BOARD A4.

    Description: A PMT does not match the requirements of the diagnostic tests.

    Cause: The PMT/DAS BOARD A5 malfunctioned.

    Action: Do the checkout for the PMT/DAS BOARD A5.

    Cause: The calibration could not locate the Offset.

    Action: Rotate the GALVO counterclockwise.

    Cause: The calibration could not locate the Offset.

    Action: Rotate the GALVO clockwise.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page92 of 206

    16021 - Malfunctioning PMT1: Gain out of range

    Description: For a given high voltage, the low gain of PMT1 might prevent it fromreaching the correct signal level.

    Cause 1: The exposure on the PLATE is not correct during calibration.

    Action: Do the Calibration for the CR 950 SYSTEM.

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    92/206

    16022 - Malfunctioning PMT2: Gain out of range

    Cause 2: The alignment of the optics is not correct.

    Action: Do the adjustment: FOLD MIRROR

    GALVO BOARD

    Cause 3: The PMT malfunctioned.Action: Install a new PMT.

    Cause 4: The PMT/DAS BOARD A5 malfunctioned.

    Action: Do the checkout for the PMT/DAS BOARD A5.

    Description: For a given high voltage, the low gain of PMT 2 might prevent it fromreaching the correct signal level.

    Cause 1: The exposure on the PLATE is not correct during calibration.

    Action: Do the Calibration for the CR 950 SYSTEM.

    Cause 2: The alignment of the optics is not correct.

    Action: Do the adjustment: FOLD MIRROR

    GALVO BOARD

    Cause 3: The PMT malfunctioned.

    Action: Install a new PMT.

    Cause 4: The PMT/DAS BOARD A5 malfunctioned.

    Action: Do the checkout for the PMT/DAS BOARD A5.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page93 of 206

    UPPER CONTROL

    18001 - SIZE CHANGER: error

    Description: The STEPPER MOTOR exceeded the time allowed to move.

    Cause: The SIZE CHANGER HOME SENSOR malfunctioned.

    Action: a. Check the CABLES and CONNECTORS to the SIZE CHANGER HOME

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    93/206

    18002 - UPPER CONTROL AY: timeout error

    18007 - UPPER CONTROL AY: Wrong command error

    18008 - UPPER CONTROL AY: Move error

    Action: a. Check the CABLES and CONNECTORS to the SIZE CHANGER HOMESENSOR.

    b. Replace the SIZE CHANGER HOME SENSOR.

    Cause: Size .

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The UPPER CONTROL AY exceeded the time allowed to move.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The UPPER CONTROL MOTION MECHANISM malfunctioned.

    Action: Check the error logs for the cause of the malfunction. See Checking the ErrorLogs.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page94 of 206

    18010 - Invalid STEPPER MOTOR error

    Description: The command to move the STEPPER MOTOR is not within the limits ofthe MOTOR in the UPPER CONTROL AY.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    94/206

    18011 - STEPPER MOTOR did not clear HOME SENSOR

    18012 - Failure to do the setup of the SENSOR INTERRUPT

    18014 - STEPPER MOTOR: jog timeout error

    18013 - Invalid DC MOTOR INTERRUPT

    Cause 1: The SLOT 8 SENSOR S22 is actuated when it should be open.

    Action: Check the SLOT 8 SENSOR S22.

    Cause 2: The SLOT 0 SENSOR S20 is actuated when it should be open.

    Action: Check the SLOT 0 SENSOR S20.

    Cause 3: The CASSETTE PRESENT SENSOR S19 is actuated when it should be open.

    Action: Check the CASSETTE PRESENT SENSOR S19.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The command to move the DC MOTOR is not within the limits of theMOTORS in the UPPER CONTROL AY.

    Cause: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page95 of 206

    18015 - Failure to advance CASSETTE

    Cause 1: The CASSETTE was removed.

    Action: Insert the CASSETTE again.

    Cause 2: The LOAD SENSORS 1 - 8 did not detect the CASSETTE.

    Action: 1. Check the LOAD SENSORS 1 - 8.

  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    95/206

    18016 - No CASSETTES to advance

    18017 - Cannot advance because the CASSETTE is in SLOT 8

    2. Check the test points on the TOP SENSOR BOARD A11.

    Description: The UPPER CONTROL AY received a command to advance aCASSETTE but the LOAD STATION SENSORS indicate that no

    CASSETTES are available.

    Cause: No CASSETTES are detected in SLOTS 1 - 8.

    Action: 1. Insert a CASSETTE.

    2. Check the SLOT SENSORS 1 - 8.

    3. Check the test points for the TOP SENSOR BOARD A11.

    Cause: The CASSETTE is in SLOT 8.

    Action: 1. Remove the CASSETTE.

    2. Check the SENSOR SLOT 8.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page96 of 206

    18018 - MAIN BELT: Home position failure

    Description: The MAIN BELT STEPPER MOTOR did not move to the home position.

    Cause 1: The MAIN BELT SENSOR S21 malfunctioned.

    Action: Check the MAIN BELT SENSOR S21.

    Cause 2: The MAIN BELT is not correctly aligned.

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    96/206

    18019 - LOAD BAR: Home position failure

    18020 - UNLOAD BAR: Home position failure

    y gAction: Check the alignment of the COG BELTS.

    Cause 3: The STEPPER MOTOR DRIVER BOARD A12 malfunctioned.

    Action: Do the checkout for the STEPPER MOTOR DRIVER BOARD A12.

    Cause 4: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The LOAD BAR STEPPER MOTOR did not move to the home position.

    Cause 1: Cannot determine the home position of the LOAD BAR.

    Action: Check the adjustment of the LOAD BARS.Cause 2: The STEPPER DRIVER BOARD 13 malfunctioned.

    Action: Do the checkout for the STEPPER MOTOR DRIVER BOARD A13.

    Cause 3: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The UNLOAD BAR did not move to the home position.

    Cause 1: Cannot determine the UNLOAD BAR home position.

    Action: Check the adjustment of the UNLOAD BARS.

    Cause 2: The STEPPER MOTOR DRIVER BOARD A14 malfunctioned.Action: Do the checkout for the STEPPER MOTOR DRIVER BOARD A14.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page97 of 206

    18024 - ALIGNMENT CAM: Clear timeout error

    Description: The ALIGNMENT CAM did not move to the CLEAR POSITION in thecorrect time.

    Cause 1: A mechanical bind occurred.

    Action: Check the adjustment of the ALIGNMENT CAMS. See ALIGNMENT CAMS /BELTS.

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    97/206

    18025 - ALIGNMENT CAM: Engage timeout error

    Cause 2: The ALIGNMENT CAM CLEAR SENSOR malfunctioned.

    Action: Check the ALIGNMENT CAM CLEAR SENSOR. See ALIGNMENT CAMS /BELTS.

    Cause 3: The DC ALIGNMENT CAM MOTOR M5 malfunctioned.

    Action: 1. Check the CONNECTORS J1/P1.

    2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The ALIGNMENT CAM did not move to the ENGAGE POSITION in theavailable time.

    Cause 1: A mechanical bind occurred.

    Action: Check the adjustment of the ALIGNMENT CAMS. See ALIGNMENT CAMS /BELTS.

    Cause 2: The ALIGNMENT CAM ENGAGE SENSOR malfunctioned.

    Action: Check the ALIGNMENT CAM ENGAGE SENSOR. See ALIGNMENT CAMS /

    BELTS.

    Cause 3: The DC ALIGNMENT CAM MOTOR M5 malfunctioned.

    Action: 1. Check the CONNECTORS J1/P1.

    2. Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    DIAGNOSTICS Error Codes

    26AUG04

    DG4826-1

    Page98 of 206

    18026 - ELEVATOR: Up timeout error

    Description: The ELEVATOR did not move to the up position in the available time.

    Cause 1: The UPPER ELEVATOR SENSOR S12 malfunctioned.

    Action: Check the UPPER ELEVATOR SENSOR S12.

    Cause 2: The COUPLING from the DC MOTOR to the SHAFT is broken.A ti I t ll COUPLING

    http://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdfhttp://d/Jobs/CR%20Service%20CDs_7-04/CR%20950%20CD%2007-04/CR%20950/ar4826_1~1.pdf
  • 8/9/2019 Diagnostics for Kodak DirectView CR 950 System 26AUG04

    98/206

    18027 - ELEVATOR down timeout error

    Action: Install a new COUPLING.

    Cause 3: A mechanical bind occurred.

    Action: Check the alignment of the ELEVATOR.

    Cause 4: The DC MOTOR malfunctioned.

    Action: Check the CABLE to the DC MOTOR.

    Cause 5: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.

    Action: Do the checkout for the MOTION SYSTEM CONTROL BOARD A1.

    Description: The ELEVATOR did not move to the down position in the available time.Cause 1: The LOWER ELEVATOR SENSOR S11 malfunctioned.

    Action: Check the LOWER ELEVATOR SENSOR S11.

    Cause 2: The COUPLING from the DC MOTOR to the SHAFT is broken.

    Action: Install a new COUPLING.

    Cause 3: A mechanical bind occurred in the ELEVATOR.Action: Check the alignment of the ELEVATOR.

    Cause 4: The DC MOTOR in the ELEVATOR malfunctioned.

    Action: Check the connections to the DC MOTOR.

    Cause 5: The