KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions Keep for future reference! DIAPHRAGM LIQUID PUMP NF 600, NF 1.600 Operating and Installation Instructions Read and observe these operating and installation instructions! An additional letter prefixing the NF model code is a country-specific designation and has no technical relevance. Contents Page 1. About this document ................................................................. 2 2. Use ........................................................................................... 3 3. Safety ....................................................................................... 4 4. Technical data .......................................................................... 6 5. Assembly and function ............................................................. 8 6. Installation and connection ....................................................... 9 7. Operation ................................................................................ 14 8. Servicing ................................................................................. 16 9. Troubleshooting ...................................................................... 21 10. Spare parts and accessories .................................................. 23 11. Decontamination declaration .................................................. 24 KNF Flodos AG Wassermatte 2 6210 Sursee, Switzerland Tel. +41 (0)41 925 00 25 Fax +41 (0)41 925 00 35 www.knf-flodos.ch [email protected] NF 600 KP DC 24V - / PMLxxxx / PLxxxx [Ch. 1] NF / UNF 600 / 1.600 KP / KT / TT [Ch. 4] DC / DCB / AA Supply voltage [Ch. 4]

Upload: others

Post on 29-Sep-2020

14 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions Keep for future reference!

DIAPHRAGM LIQUID PUMP

NF 600, NF 1.600

Operating and Installation Instructions Read and observe these operating and installation instructions! An additional letter prefixing the NF model code is a country-specific designation and has no technical relevance.

Contents Page 1. About this document ................................................................. 2 2. Use ........................................................................................... 3 3. Safety ....................................................................................... 4 4. Technical data .......................................................................... 6 5. Assembly and function ............................................................. 8 6. Installation and connection ....................................................... 9 7. Operation ................................................................................ 14 8. Servicing ................................................................................. 16 9. Troubleshooting ...................................................................... 21 10. Spare parts and accessories .................................................. 23 11. Decontamination declaration .................................................. 24

KNF Flodos AG Wassermatte 2 6210 Sursee, Switzerland Tel. +41 (0)41 925 00 25 Fax +41 (0)41 925 00 35 www.knf-flodos.ch [email protected]

NF 600 KP DC 24V

- / PMLxxxx / PLxxxx [Ch. 1]

NF / UNF

600 / 1.600

KP / KT / TT [Ch. 4]

DC / DCB / AA Supply voltage [Ch. 4]

Page 2: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 About this document

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 2

1. About this document 1.1. Use of the operating and installation instruc-

tions The operating and installation instructions are part of the pump.

Pass on the operating and installation instructions to the next owner.

Customer-specific project pumps (pump models which begin with "PL" or "PML") may differ from the operating and installation in-structions.

In the case of project pumps, take note of any additionally agreed specifications.

1.2. Symbols and markings Warning

WARNING

This symbol indicates a potential danger. It also indicates the possible consequences of failure to observe the warning. The signal word (i.e. "Warn-ing") indicates the level of danger. This specifies measures for avoiding the danger

and the consequences of failure to implement these measures.

Danger levels

Signal word Meaning Consequences if not observed DANGER warns of immedi-

ate danger Death or serious injuries and/or serious material damage are the consequence.

WARNING warns of possible danger

Death or serious injuries and/or serious material damage are possible.

CAUTION warns of a poten-tially dangerous situation

Minor injuries or material dam-age are possible.

Tab. 1

Other information and symbols This indicates an activity (step) that needs to be carried out.

This indicates the first step of an activity to be carried out. Any 1. additional steps required are consecutively numbered.

This symbol refers to important information.

Project pumps

Page 3: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Use

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 3

2. Use 2.1. Proper use The pumps are intended for transferring and metering liquids and gases.

Owner's responsibility Only install and operate the pumps under the operating parameters and conditions described in Chapter 4, Technical data.

The pumps may be operated only when fully assembled.

Before transferring or metering a medium, check that the medium can be transferred without risk in the specific application case.

Before using a medium, check the compatibility of the materials of the pump head, pump housing, diaphragm and valves with the medium.

The temperature of the medium must lie within the permissible temperature range (see Chapter 4).

The media should not contain particles as these can prevent the pump from working correctly. If this cannot be guaranteed, a filter < 100 µm with a sufficiently large filter area must be used upstream of the pump. 2.2. Improper use The pumps may not be operated in an explosive atmosphere. For special modifications outside the standard technical specifica-tions, please contact a KNF pump specialist.

Operating parameters and conditions

Requirements for transferred medium

Page 4: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Safety

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 4

3. Safety

Note the safety precautions in Chapters 6. Installation and connection, and 7. Operation.

The pumps are built according to the generally recognized rules of technology and in accordance with the pertinent occupational safety and accident prevention regulations. Nevertheless, dangers can result during their use which lead to injuries to the user or others, or to damage to the pump or other property.

Only use the pumps in perfect working order and in accordance with their intended use. Always ensure adherence to the operating and installation instructions and work in a safety-conscious man-ner.

Make sure that only trained and instructed personnel or specially trained personnel work on the pumps. This especially applies to assembly, connection and servicing work.

Make sure that all personnel have read and understood the operat-ing and installation instructions, and in particular the "Safety" chapter.

Always ensure adherence to all pertinent accident prevention and safety regulations when working on and operating the pump.

Always observe the safety regulations when handling dangerous media.

Always ensure adherence to all information stickers on the pumps, such as flow direction arrows and type plates, and keep stickers in legible condition.

All replacement parts should be properly stored and disposed of in accordance with the applicable environmental protection regula-tions. Ensure adherence to the pertinent national and international regulations. This especially applies to parts contaminated with toxic substances.

Personnel

Working in a safety-conscious manner

Handling dangerous media

Notes

Environmental protection

Page 5: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Safety

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 5

The pumps are in accordance with the requirements of the guide-lines 2011/65/EU (ROHS2)

The pumps conform to EU safety requirements and guidelines for Electromagnetic interference 2004/108/EC.

As defined in the Machinery Directive 2006/42/EC, pumps are partly completed machines and not ready-for-use, the overall equipment must be made to fully conform with the requirements of the Directive before being brought into service. Always ensure implementation and enforcement of the basic requirements of the Machinery Directive 2006/42/EC according to Appendix I (General Principles).

The following harmonised standards are met:

NF 600 AA EN 55014-1

NF 600 DC / NF 1.600 DC EN 55022

EN 55011

NF 600 DCB EN 55014-1

EN 61000-6-1

All repairs to the pump(s) must be carried out by the relevant KNF Customer Service team.

Only use KNF original parts for all maintenance work.

EU directives/standards

Customer services and repairs

Page 6: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Technical data

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 6

4. Technical data Pump materials The pump type KP stands for:

Assembly Material1) Pump head * PP Valve plate / seals EPDM Diaphragm PTFE-coated

Tab. 2 1) according to DIN ISO 1629 and 1043.1

The pump type KT stands for:

Assembly Material1) Pump head * PP Valve plate / seals FFKM Diaphragm PTFE-coated

Tab. 3 1) according to DIN ISO 1629 and 1043.1

The pump type TT stands for:

Assembly Material1) Pump head * PVDF Valve plate / seals FFKM Diaphragm PTFE-coated

Tab. 4 1) according to DIN ISO 1629 and 1043.1

* Pump head 1 (Fig. 11): comprises a connecting plate and an intermedi-ate plate * Pump heads 2 – 4 (Fig. 11): comprise a head connection and an inter-mediate plate Hydraulic ratings

Parameter Value Flow rate [l/min] 1), 2), 3) ≥ 6.0 Max. permissible pressure NF 600 [bar g] 1 Max. permissible pressure NF 1.600 [bar g]

6

Suction head [mWG] 3 Tab. 5 1) Measured with water at 20°C / at atmospheric pressure

2) Flow rates may vary from the values shown, depending on fluid viscosi-ty, pump head material and the hoses / hose connectors used. 1) Flowrate DC: ≥ 5.5 Hydraulic connections

Parameter Value NF 600 / NF 1.600 G 3/8" UNF 600 / UNF 1.600 NPT 3/8" Recommended hose size ID [mm] 10

Tab. 6

Page 7: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Technical data

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 7

Specifications NF 600 AA

Motor voltage 230V/ 50Hz

100V/50-60Hz

115V/60Hz

Power consumption [W] 99 115/85 98 Max. current at max. load [A] 0.43 1.15/0.85 0.85 Max. motor current [A] 0.55 1.7 1.1 Protection class [-] IP 54 IP 54 IP 54 Weight1) [g] 3000 3000 3000

Tab. 7

Specifications NF 600 DC

Motor voltage 24V Power consumption [W] 34 Max. current at max. load [A] 1.4 Max. motor current [A] 1.5 Protection class [-] IP 50 Weight1) [g] 1500

Tab. 8

Specifications NF 600 DCB

Motor voltage 24V Power consumption [W] 44 Max. current at max. load [A] 1.8 Max. motor current [A] 2 Protection class [-] IP 54 Weight1) [g] 1500

Tab. 9

Specifications NF 1.600 DC

Motor voltage 24V Power consumption [W] 75 Max. current at max. load [A] 3.1 Max. motor current [A] 4.9 Protection class [-] IP 50 Weight1) [g] 2500

Tab. 10 1) The weight may differ slightly from the stated value, depending on the version.

Electrostatic sensitive components (ESD)

Various parameters

Parameter Value Permissible ambient temperature range [°C]

+ 5 to + 40

Permissible media temperature [°C] + 5 to + 80 Permissible kinematic viscosity of the media [cSt]

≤ 150

Tab. 11

Page 8: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Assembly and function

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 8

5. Assembly and function Assembly

1 Inlet 2 Outlet 3 Connecting plate 4 Motor 5 Leads 6 Head connection 7 Mounting bracket * * AA version without mount-

ing bracket

Fig. 1: Diaphragm liquid pump NF 600

1 Ring head connection – inlet

2 Ring head connection – outlet

3 Diaphragm chamber 4 Diaphragm 5 Eccentric 6 Connecting rod 7 Drive

Fig. 2: Operating principle

Diaphragm liquid pumps are based on reciprocating displacement pump technology. The eccentric (5) transforms the circular move-ment of the motor shaft (7) into an oscillating movement of the diaphragms (4) via a connecting rod (6). Working in conjunction with the inlet (1) and outlet (2) valves, the movement of the dia-phragms generates the actual pump operation.

The use of 4 diaphragms enables 4 working strokes per revolution. The medium enters the pump via the ring head connection inlet (1) and is delivered to the diaphragm. The medium is then delivered to the pressure side connection via the ring head connection outlet (2).

Page 9: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Installation and connection

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 9

6. Installation and connection Only install and operate the pumps under the operating parameters and conditions described in Chapter 4, Technical data.

Observe the safety notes (see chapter 3).

6.1. Installation Before installation, store the pump at the installation location to

bring it up to ambient temperature.

Mounting dimensions (see Fig. 3 to 6)

Fig. 3: Mounting dimensions NF 600 DC

Fig. 4: Mounting dimensions NF 600 DCB

Mounting dimensions

Page 10: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Installation and connection

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 10

Fig. 5: Mounting dimensions NF 600 AA

Fig. 6: Mounting dimensions NF 1.600 DC

For pumps with fan: mount the pump so that the fan of the motor can suck in sufficient cooling air.

Make sure that the installation location is dry and the pump is protected against rain, splashes, hose and drip water.

Protect the pump against dust.

Protect the pump against vibrations and jolts.

The pump can be mounted in any orientation. The priming and accuracy of the pump is optimal if installed as in the illustration (Fig. 7).

Cooling air supply

Installation location

Mounting orientation

Fig. 7: Optimal mounting orienta-

tion / horizontal

Page 11: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Installation and connection

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 11

KNF recommends mechanically decoupling the pump from the

pipework system, e.g. by using flexible hoses or pipes. This prevents any oscillations of the pump being transferred to the system.

For pumps with alternating current motors:

WARNING

Danger of injury during operation Take protective measures against touching parts

which are energised (live), such as electrical connections or windings.

Take protective measures against touching moving parts (e.g. fan).

Danger of damaging the pump during operation

Take protective measures to ensure that no foreign objects can get into the pump or pene-trate the motor (see protection class).

Decoupling

Touch and foreign object pro-tection

Page 12: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Installation and connection

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 12

6.2. Electrical connection For pumps with alternating current motors:

DANGER

Risk of electric shock, danger of death Pump should only be connected by a specialist.

Only connect the pump when the power supply is turned off.

For pumps with direct current motors:

Pump should only be connected by a specialist.

Only connect the pump when the power supply is turned off.

All electrical connection work must adhere to the pertinent guidelines, regulations and technical standards.

Connecting the pump

Make sure that the power supply data match the data on the 1. motor type plate. The current consumption can be found on the type plate.

Connect the motor cables. Electrical data Chapter 4. 2.

Always check polarity of cables.

For DC motors: red motor cable: + black motor cable: -

In the case of brushless DC motors: incorrect polarity may damage the electronics!

If using AC motors, the power supply must not deviate more than a maximum of +/- 10 % from the specifications on the type plate.

In order to ensure that emissions comply with the EN 55014 -1 emission standard, this pump type with a brushless DC motor DCB must be fitted with an additional circuit. The additional circuit must be installed as closely as possible to the motor.

The additional circuit must be installed according to the following wiring diagram and its defined components in order to achieve the required level of attenuation.

Fig. 8: Additional circuit

Additional circuit for brushless DCB motors

C1: Condenser

≥ 2200 µF / UN = 12 V

≥ 1000 µF / UN = 24 V

Page 13: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Installation and connection

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 13

6.3. Hydraulic connection Only connect components to the pump that are designed to

handle the hydraulic data of the pump (see Chapter 4, Tech-nical data).

Only use hoses that are suitable for the maximum operating pressure of the pump (see Chapter 4).

Only use hoses that are chemically resistant to the liquids being pumped.

6.3.1. Connecting the pump

Arrows on the pump head indicate the flow direction.

Remove the protective caps from the connections. 1.

Connect the inlet and outlet lines. 2.

Keep the inlet line as short as possible in order to keep the priming process as brief as possible.

If the pump is used to build up pressure, make sure that all 3.

transition joints between hose and pump are secure in order to ensure that the hoses cannot come off.

Check that the hoses and transition joints are fitted correctly 4. and securely.

Check that the system is leak-tight. 5.

Connected components

Hoses

Page 14: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Operation

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 14

7. Operation The pumps should only be used under the operating parame-

ters/conditions described in Chapter 4, Technical data.

Ensure that the pumps are being used correctly (see Section 2.1).

Improper use of the pumps must be prevented (see Section 2.2).

Observe the safety notes (see chapter 3).

Pumps are components intended to be incorporated into another machine. The machine/equipment in which the pumps are installed must be made to fully comply with the pertinent regulations before being put into operation.

CAUTION

Risk of burning The drive heats up. Avoid contact with the drive.

Avoid contact with flammable materials.

Excessive pressures and its inherent dangers can be prevent-ed by using a bypass system with a pressure relief valve between the pressure and suction side of the pump. Further information is available from your KNF adviser (Telephone number: see first page).

If the pump stops running, reduce the pressure in the system

until it is at normal atmospheric pressure.

For pumps with thermal switch or electronic overload protection:

WARNING

Risk of physical injury and damage to the pump due to automatic start If the pump overheats and the thermal switch / elec-tronics stops pump operation, the pumps will restart automatically as soon as they have had time to cool down. Take steps to ensure that this cannot produce a

hazardous situation.

Switching the pumps on and off The motor speed of the pump, and thus the flow rate, is adjustable and can also be regulated to some extent.

For more details, see Chapter 4, Technical data.

Duty cycle / impulse operation KNF pumps are designed for continuous operation.

Short start and stop cycles may adversely affect the service life of the brushed motors.

Pump standstill

Adjusting and controlling the motor speed

Impulse operation

Page 15: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Operation

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 15

If the pump is operated with short cycles in your application, please contact a KNF pump specialist for further information (Telephone number: see first page).

Turning the pump on In order to guarantee that the pump can start every time it is

advisable to reduce the back pressure to an acceptable level. This is also the case if there is a short power cut.

For more specific information contact the KNF specialist (Tele-phone number: see first page).

Turning the pump off KNF recommends: if pumping aggressive liquids, the pump

should be rinsed thoroughly prior to switch off (see Section 8.2.1), as this will help to lengthen the service life of the dia-phragm.

Ensure that the system is subject to normal atmospheric pres-sure (release the hydraulic pressure).

Flow rate NF 600

Fig. 9: Flow rate of the pump NF 600 Flow rate NF 1.600

Fig. 10: Flow rate of the pump NF 1.600

Page 16: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Servicing

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 16

8. Servicing 8.1. Servicing schedule Component Servicing interval Pump - Regular inspection for external

damage or leaks Pump head - Clean if the flow rate decreases, the

pump does not work or no vacuum is created (Section 8.2)

Diaphragm, valve plates and seals

- Change as soon as pumping capac-ity decreases, preferably sooner

Tab. 12

8.2. Cleaning

WARNING

Health hazard due to dangerous substances in the pump! Depending on the substance transferred, risk of caustic burns or poisoning. Wear protective clothing if necessary, e.g. pro-

tective gloves.

Rinse the pump with a neutral liquid and pump empty.

8.2.1. Flushing the pump If pumping aggressive media, KNF recommends flushing the

pump with air under atmospheric conditions for several minutes prior to switch off (if necessary for safety reasons: use an inert gas). This will extend the service life of the diaphragm.

8.2.2 Cleaning the pump 1. Rinse the pump with a neutral liquid and ensure that the

pump is free of any hazardous substances.

2. Pump empty

3. Switch off pump and remove mains plug from the socket.

4. Disconnect hoses from the pump head.

5. Where possible, wipe the components with a soft dry cloth. Do not use cleaning solvents as these may corrode plastic parts.

6. If there is compressed air available, blow off the separate parts.

7. We recommend replacing the diaphragm when the head parts are removed.

Information on procedure

Page 17: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Servicing

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 17

Preparation of dismantling

For the part numbering system in the subsequent steps refer to Fig.13.

1. Perform steps defined in Chapter 8.2.1 and 8.2.2.

2. Memorize the position of connecting plate (4) and the orien-tation of the hose connections. This defines head I.

Qty. Tools 1 Wrench for diaphragm, special tool *

1 T20 Torx screwdriver 1 Slotted Screwdriver No.2

Tab. 13 * Available with spare parts kit or as accessory, Chapter 10

Pump with mounting bracket (DC and DCB) Remove the 4 screws (1) from the mounting bracket (Fig 11). 1.

Dismantling the pump heads I II III IV

Loosen all 16 head screws (1) and 4 connecting screws (2) by 2. 2 revolutions in order to prevent any damage to the outer seal-ing lips.

Now dismantle the head screws (1) and connecting screws (2) 3. from the connecting plate I and remove entire head.

Follow the same procedure to dismantle pump heads II, III and 4. IV.

Tools

I

II III

IV

Fig. 12

1 1

1 1

Fig. 11

mounting bracket

Page 18: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Servicing

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 18

Removing the valves and seals Remove the anchor valves (5) and anchor valve seal (7) from 5.

the intermediate plate (6), the connecting plate (4) and head connection (3).

Remove the sealing rings (8) from the connecting plate and 6. head connection.

Removing the diaphragm

Starting with diaphragm head I.

Rotate motor shaft with screw driver until the diaphragm is at 7. its topmost position (upper dead center). Alignment of slot in the motor shaft has to be in direction of diaphragm. See Fig. 14.

Use the wrench (see accessories) to loosen the diaphragm (9). 8. In order not to damage the teeth of the diaphragm, press the wrench properly against the diaphragm and check for a correct fit. Make sure that no washers (10) fall into the pump housing. One washer per diaphragm!

Follow the same procedure to remove the three remaining 9. diaphragm of head II, III and IV, repeat steps 7 and 8.

We recommend replacing the 4 diaphragms (9). Cleaning the parts

Clean the diaphragm (9), anchor valves (5), anchor valve seals 10. (7), sealing rings (8), head connection (3) and connecting plate (4) with a cloth and then blow off with compressed air.

Blow off the intermediate plate (6), connecting plate (4) and 11. head connection (3) with compressed air.

Mounting the diaphragm Starting with diaphragm head I.

Rotate motor shaft with screw driver until the connecting rod is 12. at its topmost position (upper dead center). Alignment of slot in the motor shaft has to be in direction of diaphragm. See Fig. 14.

Screw the diaphragm (9) with one washer (10) on the connect-13. ing rod, making sure that no washers (10) fall into the pump housing.

Tighten the diaphragm using the special tool. In order not to 14. damage the teeth of the diaphragm, press the wrench properly against the diaphragm and check for a correct fit.

Torque 0.8 Nm **

Follow the same procedure to mount the three remaining 15. diaphragm of head II, III and IV, repeat steps 12 to 14.

Fig. 13

1 Head screws 2 Connecting screws 3 Head connection 4 Connecting plate 5 Anchor valves 6 Intermediate plate 7 Anchor valve seal 8 Head connection sealing ring 9 Diaphragm 10 Washer

Page 19: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Servicing

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 19

Mounting the valves and seals Insert the anchor valves (5) and anchor valve seals (7) in the 16. corresponding recesses of the intermediate plate (6), the head connection (3) and connecting plate (4).

Insert sealing rings (8) in the head connection (3) and connect-17. ing plate (4).

Mounting the pump heads I IV III II Starting with pump head I, mount anti-clockwise in order to prevent any damage to the outer sealing lips.

Rotate motor shaft with screw driver until the diaphragm is at 18. its topmost position (upper dead center). Alignment of slot in the motor shaft has to be in direction of diaphragm. See Fig. 14.

CAUTION: In order to ensure a leak-proof pump the dia-phragm has to be properly centered in the pump housing, Fig.15. If necessary adjust motor shaft slightly with screw driv-er until the diaphragm is centered.

Insert the intermediate plate (6). 19. CAUTION: Make sure that the orientation is correct so that the connections point in the right direction. A pin and hole system prevents incorrect assembly. See Fig. 16.

Mount the connecting plate (4) on the intermediate plate (6), 20. alternately tighten the four head screws (1) with torque 2.5 Nm**.

correctly aligned X incorrectly aligned

Fig. 16

Fig. 15

Fig. 14

Page 20: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Servicing

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 20

Follow the same procedure to mount pump heads IV, III, II, 21. repeat steps 18 to 20. CAUTION: When mounting head connection II, the two marked head screws (1) of pump head I (see Fig.17) must be loosened by 2 revolutions to prevent any damage to the seal-ing lips.

Tighten the two loosened head screws (1) of head I with torque 22. 2.5 Nm**.

Mount the connecting screws (2) with torque 1.2 Nm **. 23.

Mounting the mounting bracket (DC and DCB)

Tighten the mounting bracket with torque 2.5 Nm**. 24. ** Caution: Settling, fluctuations in temperature, transport, etc. may all cause variations in the breakaway torque.

8.3. Replacing diaphragm, valves and seals Same procedure as in section 8.2, "Cleaning" – with the exception of step 8: instead of cleaning the diaphragm (9), anchor valves (5), anchor valve seals (7) and sealing rings (8), these components must be replaced.

1

Fig. 17

Page 21: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Troubleshooting

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 21

9. Troubleshooting For pumps with alternating current motors:

DANGER

Risk of electric shock, danger of death Before working on the pump disconnect the pump

from the power supply. Ensure that the pump is de-energised.

Pump does not work

Cause Fault remedy Pump not connected to mains power supply.

Connect pump to mains supply.

Power supply is not switched on.

Switch on power supply.

Thermal switch or pump elec-tronics have responded.

Disconnect pump from mains power supply. Allow pump to cool. Identify and eliminate cause of overheating/overload.

Connections or pipes are blocked.

Check pipes and connections. Remove blockage.

External valve is closed or filter is blocked.

Check external valves and filters.

Diaphragm, anchor valves or seals are worn out.

Replace diaphragm, anchor valves and seals (see Section 8.3).

Tab. 14

Pump is not priming

Cause Fault remedy Suction side of pump not con-nected.

Connect the suction side of the pump.

Liquid in the container is too low.

Fill container.

Hose connections are not leak-tight.

Secure transition joints between hose and connections with clamps or clamping elements.

System valve is closed or filter is blocked.

Open the valve. Clean filter.

Pump head is filled with gas. The system is unable to handle the pressure on the pressure side

Reduce pressure on pressure side.

Particles in the pump. Clean the pump head (see Section 8.2). The head parts are not media-resistant.

Replace the pump head with a compatible version.

Incorrect interchange of outlet and inlet line connections.

Remove outlet and inlet lines and re-connect correctly.

Tab. 15

Page 22: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Troubleshooting

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 22

Flow rate, suction head or pressure head is too low The pump does not achieve the technical performance data stated on the data sheet. Cause Fault remedy Components in the system connected to the suction and pressure sides, such as hoses, valves or filters, are causing too much resistance.

Modify installation, check the cross-section of components.

Hose connections are not leak-tight.

Secure transition joints between hose and hose connections with clamps or clamping elements

Particles in the pump. Clean the pump head; install suction-side filter if required. Viscosity of the liquid is too high.

Contact KNF.

Incorrect interchange of outlet and inlet line connections.

Remove outlet and inlet lines and re-connect correctly.

The head parts are not media-resistant.

Replace the pump head with a compatible version.

Tab.16

Fault cannot be rectified If you are unable to identify any of the above causes, please send the pump to KNF customer services (see address on last page).

Flush the pump to clear the pump head of any hazardous or 1. aggressive fluids (see Section 8.2.1).

Dismantle the pump. 2.

Clean the pump (see Section 8.2.2). 3.

Send the pump, with completed decontamination statement 4. (see Chapter 11), to KNF customer services stating the nature of the pumped medium.

Page 23: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Spare parts and accessories

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 23

10. Spare parts and accessories Spare parts kit

Spare parts kit Order No. Spare parts kit NF 600 / NF 1.600 KP 157803 Spare parts kit NF 600 KT / TT 157804 Spare parts kit NF 1.600 KT / TT 157805

Tab. 17

Accessories

Accessories Order No. Wrench for diaphragm* 157776

Tab. 18

* Wrench for diaphragm is included in spare parts kit

Page 24: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 Decontamination declaration

KNF Flodos AG BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 24

11. Decontamination declaration

KNF shall only undertake to repair the pump on condition that the customer provides certification of the transferred media and the cleaning of the pump (decontamination declaration).

Therefore please fill in the document “decontamination decla-

ration” which has been delivered or download it directly from the website www.knf.com/downloads.

Enter the pump model, the Serial No. and the transferred media in the form below and sent the signed form together with the flushed and cleaned pump to your local KNF sales partner (see address on last page).

Customer decontamination declaration for repair order

We confirm that the only media that the model specified has pumped are those named and that the pump has been flushed and cleaned.

Pump model

Serial No.

Pumped media

The pump does not contain aggressive, biological, radioactive, poisonous, or other dangerous media.

Company Date/Signature

Page 25: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 For your notes

KNF Flodos BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 25

Page 26: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 For your notes

KNF Flodos BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 26

Page 27: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

Diaphragm liquid pump NF 600 For your notes

KNF Flodos BA_NF600_EN_06_157511 Translated from the Original Operating and Installation Instructions 27

Page 28: Diaphragm liquid pump NF 600 - KNF | Diaphragm Pumps & Systems€¦ · Diaphragm liquid pump NF 600 Installation and connection KNF Flodos AG BA_NF600_EN_06_157511 Translated from

KNF worldwide

Please find your local KNF partners at: www.knf.com