01 - Cover Sheet 02 - Equipment Layout 03 - Equipment Layout Sections 04 - Floor - Electrical Layout 05 - Floor Structural Details 06 - Detailed Ceiling Layout 07 - Ceiling Structural Details 08 - Ceiling Support Details 09 - Power Requirements 10 - HVAC - Environment 11 - Equipment Details - Delivery 12 - Interconnections 13 - Disclaimer - Site Readiness New Orleans, LA Brian Ferrell DISCOVERY XR 656 PLUS (G2) REK GE Healthcare Date Drawn by Verified by S.O. (GON) PIM Ref & Rev Rev GE does not take responsibility for any damages resulting from changes on drawings made by others. Errors may occur by not referring to the complete set of final issue drawing. GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings, however caused. All dimensions are in millimeters unless otherwise specified. Do not scale from printed pdf files. GE accepts no responsibility or liability for defective work due to scaling from these drawings. /13 RAD-M076452-01B.DWG +1 337 371 7202 [email protected] REK 5643834-1EN Rev 5 B USA 1:48 4599253 12/FEB/2018 Childrens Hospital of New Orleans Scale REV DATE MODIFICATIONS A 21.Nov.17 Final version 01 based on DC-36034 B 12/FEB/2018 REVISION BASED ON DC-44442; CHANGE TO GON, POWER REQUIREMENTS, POWER DISTRIBUTION, CONTROL WALL HEIGHT, PIM REVISION A mandatory component of this drawing set is the GE Healthcare Pre Installation manual. Failure to reference the Pre Installation manual will result in incomplete documentation required for site design and preparation. Pre Installation documents for GE Healthcare products can be accessed on the web at: www.gehealthcare.com/siteplanning Sheet Format A3 Concession #### RAD ROOM 1 FINAL 01

Upload: others

Post on 26-Feb-2022

2 views

Category:

Documents


0 download

TRANSCRIPT

01 - Cover Sheet02 - Equipment Layout03 - Equipment Layout Sections04 - Floor - Electrical Layout05 - Floor Structural Details06 - Detailed Ceiling Layout07 - Ceiling Structural Details08 - Ceiling Support Details09 - Power Requirements10 - HVAC - Environment

11 - Equipment Details - Delivery12 - Interconnections13 - Disclaimer - Site Readiness

New Orleans, LA

Brian Ferrell

DISCOVERY XR 656 PLUS (G2)

REK

GE Healthcare

DateDrawn by Verified by S.O. (GON) PIM Ref & Rev Rev

GE does not take responsibility for any damages resulting from changes on drawings made by others. Errors may occur by not referring to the completeset of final issue drawing. GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings, however caused. Alldimensions are in millimeters unless otherwise specified. Do not scale from printed pdf files. GE accepts no responsibility or liability for defective work

due to scaling from these drawings. /13RAD-M076452-01B.DWG

+1 337 371 7202

[email protected]

REK 5643834-1EN Rev 5 B

USA

1:48

4599253 12/FEB/2018

Childrens Hospital of New Orleans

Scale

REV DATE MODIFICATIONSA 21.Nov.17 Final version 01 based on DC-36034B 12/FEB/2018 REVISION BASED ON DC-44442; CHANGE TO GON, POWER REQUIREMENTS,

POWER DISTRIBUTION, CONTROL WALL HEIGHT, PIM REVISION

A mandatory component of this drawing set is the GE Healthcare Pre Installation manual. Failure to reference the Pre Installation manual will result inincomplete documentation required for site design and preparation.

Pre Installation documents for GE Healthcare products can be accessed on the web at: www.gehealthcare.com/siteplanning

SheetFormat A3 Concession

####

RAD ROOM 1 FINAL

01

AA'

EQUIPMENT LAYOUT

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Equipment Layout 021:48

LEGEND

ARCHITECTURAL ELEMENTS

GE SUPPLIEDEQUIPMENT MOVEMENT

SERVICE CLEARANCE

MAIN AXIS

DIMENSIONS

CUSTOMER SUPPLIED ITEMS

ITEM DESCRIPTION DIMENSIONSLxWxH (in)

WEIGHT(lb)

1 ELEVATING TABLE G2 94.5x36.7x33.5 970

2 STANDARD WALL STAND G2 27x52.9x103.5 595

3 15'-2" RAILS FOR 3 m BRIDGE 182x131.5x3.74 319

4 OTS WITH 3 m BRIDGE 121.1x25.8x102.6 772

5 SYSTEM CABINET/GENERATOR 27.8x26.1x58.1 705

6 TETHER INTERFACE BOX (TIB) 10x5.75x12.2 15.40

7 MAIN DISCONNECT PANEL (MDP) 16x6x24 80

8 GRID HOLDER 20.15x8x19.9 30.40

9 IMAGE PASTING BARRIER 42.7x29x81.1 120

10 DONGLE ASSEMBLY 3.29x6.1x3.27 1.80

11 OPERATOR CONSOLE 47.2x29.5x47.4 79

12 PARTIAL UPS 8.4x16.1x12.8 76

13 WALL BOX 14.6x6.7x4.4 11

EXAM ROOM HEIGHT

CONTROL WALL HEIGHT

EXAM ROOM CEILING HEIGHT

7'-0"

9'-6"

Auto Image Pasting at Wall Stand

Volu

meR

AD

Vertical at Wall stand

Horizontal at Table

Horizontal at Extended Arm Wall stand

Cross-Table at Wall stand

The chart shows the applications possible to perform with the present equipment positioning,however the sales contract may not include all of them.

APPLICATIONS

Auto Image Pasting at Table YES

YES

NO

YES

YES

YES

REAR TO FRONT CROSS TABLE SHOT TO TABLE CENTER LINE.

THE FOLLOWING SHOTS ARE NOT AVAILABLE IN THIS LAYOUT:

1

3

4

2

5

6

7

8

9

11

10

1213

RAD EXAMROOM 1

CONTROLAREA

DRESSINGROOM

12'-8

"

22'-9

"

21'-1

0"

6'-2"

28'-9

"

6'-6

"

16'-2"

9'-3"

5'-7

"

0'-4

"4'

-3" 3'-0"

11'-0"

4'-2"

7'-2"

SIDE VIEW A-A'

FRONT VIEW B-B'

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Equipment Layout Sections 031:48

EXAM ROOM CEILING HEIGHTS

RECOMMENDED AND MINIMUM ROOM HEIGHTSCONFIGURATION SPECIFICATION CEILING HEIGHT

2m, 3m or 4-meter Bridge Recommended 2900 mm (114 in)

2m, 3m Bridge with Advanced Applications Minimum 2700 mm (106 in)

2m, 3m Bridge without Advanced Applications Minimum 2620 mm (103 in)

4m Bridge with Advanced Applications Minimum 2700 mm (106.25 in)

4m Bridge without Advanced Applications Minimum 2630 mm (103.25 in)

3m Bridge with Wallstand at Front Position Minimum 2870 mm (113 in)

4m Bridge with Wallstand at Front Position Minimum 2880 mm (113.25 in)

+ 9'-6"

± 0'-0"

+ 9'-6"

± 0'-0"1

5

67

8

10

1

2

2

33

34

4

6

7

5

AA'

FLOOR - ELECTRICAL LAYOUT

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Floor - Electrical Layout 041:48

RAD EXAMROOM 1

CONTROL AREA

NOTE: IF DOORWAY LEADS TO ANUNCONTROLLED AREA, A DOORINTERLOCK SWITCH ANDWARNING LIGHT WILL BEREQUIRED.

12'-8

"

1'-6

"

7'-1

0"

7'-2"

4'-7

"

4'-2"

ITEM QTY DESCRIPTION

1 Table anchoring (see Floor Structural Details)

2 Wall Stand anchoring (see Floor Structural Details)

3 Main disconnect panel @ 5'-0" A.F.F.

4 8" x 8" x 6" Flush floor junction box

5 1" Conduit running thru-ceiling

6 1.5" Conduit running thru-ceiling

7 2.5" Conduit running thru-ceiling

8 18" x 18" x 4" Flush junction box @ 0'-9" A.F.F. (TYP)[SYSTEM CABINET]

9 18" x 3.5" Vertical wall duct with minimum 2 dividers

10 6" deep Box above ceiling, size as required

11 2" Conduit running thru-floor

12 2.5" Conduit running thru-floor

13 6" x 6" x 4" Flush junction box @ 4'-0" A.F.F. (TYP)[TIB]

14 6" x 6" x 4" Flush junction box @ 1'-0" B.F.C. (TYP)[DONGLE]

15 8" x 8" x 4" Flush junction box @ 1'-6" A.F.F. (TYP)[WALL STAND]

16 12" x 6" x 4" Flush junction box @ 1'-6" A.F.F. (TYP)[CONSOLE]

Basic system

RJ 45 network socketSystem emergency off (SEO), (recommended height 4'-0" above floor)X-Ray room warning light control panel - Available from GEHC, Call: 800-279-7925 orlocal GE Installation Project ManagerX-Ray ON lamp (L1) - 24VDoor interlock switch (needed only if required by state/local codes)120V Duplex hospital grade outlet

CONDUITS REQUIRED FOR BASE SYSTEM

WARNING LIGHT CONTROL TO WARNING LIGHT ONE 16mm [1/2 in] CONDUIT

WARNING LIGHT CONTROL TO GENERATOR ONE 16mm [1/2 in] CONDUIT

WARNING LIGHT CONTROL TO 120-V 1Ø POWER CONDUIT AS REQUIRED

MAIN DISCONNECT PANEL TO GENERATOR ONE CONDUIT AS REQUIRED

MAIN DISCONNECT PANEL TO EMERGENCY OFF(CONTROL ROOM) ONE 16mm [1/2 in] CONDUIT

MAIN DISCONNECT PANEL TO FEEDER ONE CONDUIT AS REQUIRED

DOOR INTERLOCK SWITCH TO GENERATOR ONE 16mm [1/2 in] CONDUIT

GENERATOR TO TETHER INTERFACE BOX ONE 50mm [2 in] CONDUIT

CONTROL ROOM WALLBOX TO TETHER INTERFACE BOX ONE 25mm [1 in] CONDUIT

GENERATOR TO DONGLE ASSEMBLY ONE 25mm [1 in] CONDUIT

CONTROL ROOM WALLBOX TO DONGLE ASSEMBLY ONE 50mm [2 in] CONDUIT

1

4

2

15

3

13

14

16

6

7

5

11

12

8

9

10

0'-8

"

1'-8"7

4'-0

"

CABLE MANAGEMENT

CONDUIT IN THE FLOOR

NOT TO SCALE

Removable coverplate

VERTICAL DUCT ON WALL

TABLE ANCHORING

Anchors M8x190(7.5 in) suppliedby GE

TABLE BASE

The floor bearing the system is recommended to be concrete and the thickness to be determined by a StructuralEngineer to properly support the equipment loads. The supplied anchors require a minimum embedment of 90mm (3.54 in) into the concrete. If the floor thickness is less than 95 mm (3.74 in), it is recommended that the unitbe secured using a through-bolt method with a reinforcement plate on the back side.

SCALE 1:20

150mm[5.9in]

123mm[4.8in]

93mm [3.7in]

860mm[33.9in]920mm[36.2in]

460mm[18.1in]

460mm[18.1in] Cable inlet area

TABLE AXIS

WALLSTAND AXIS

Anchors M10x150 mm(5.9 in) supplied by GE

SCALE 1:10

Cable inlet area

WALLSTAND BASE

WALLSTAND ANCHORING

Concrete area for wall stand installation should be 1 m² (10.76 ft²).

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Floor Structural Details 05

AA'

CEILING LAYOUT

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Ceiling Layout 061:48

RAD EXAMROOM 1

CONTROL AREA

ITEM DESCRIPTION

STANDARD 3 RAILS SYSTEM

1

UNISTRUT OR EQUIVALENT SUPPORT IN CEILING FOR FASTENING CABLE DRAPE RAIL.SUPPORTS TO RUN CONTINUOUS WITH NO FITTINGS EXTENDING BELOW FACE OF UNISTRUTCHANNEL, RUN WALL TO WALL, BE PARALLEL, SQUARE, AND IN THE SAME HORIZONTALPLANE, FLUSH WITH THE FINISHED CEILING.RAILS ARE MOUNTED TO THESE SUPPORTS EVERY2'-2" AND REQUIRE 50 LBS. PR BOLT LOAD. METHODS OF SUPPORT THAT PERMITATTACHMENT TO STRUCTURAL STEEL OR THROUGH BOLTS IN THE CONCRETE SHOULD BEFAVORED. DO NOT USE SCREW ANCHORS IN DIRECT TENSION

2

UNISTRUT OR EQUIVALENT SUPPORT FOR FASTENING CEILING SUPPORTED EQUIPMENT.SUPPORTS TO RUN CONTINUOUS WITH NO FITTINGS EXTENDING BELOW FACE OF UNISTRUTCHANNEL, RUN WALL TO WALL, BE PARALLEL, SQUARE, AND IN THE SAME HORIZONTAL PLANE,FLUSH WITH THE FINISHED CEILING. RAILS ARE MOUNTED TO THESE SUPPORTS EVERY 2'-2"AND REQUIRE 350 LBS. (597 LBS. IN SEISMIC REGIONS) PER BOLT LOAD. METHODS OF SUPPORTTHAT PERMIT ATTACHMENT TO STRUCTURAL STEEL OR THROUGH BOLTS IN CONCRETE SHOULDBE FAVORED. DO NOT USE SCREW ANCHORS IN DIRECT TENSION.

1 2

0'-8" 3'-1" 4'-8"

FOCAL SPOT TRAVEL

SCALE 1:50

Focal spottravel area

3 m BRIDGE

4412mm[173.7in]

5042mm[198.5in]

Suspension columntravel area

645mm[25.4in]

480mm[18.9in]

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Ceiling Structural Details 07

CEILING REQUIREMENTS

To allow installation of the stationary rail cross-members, clearance is required between the ends of the stationaryrails and the walls.

It is recommended that sprinkler heads not be placed between the stationary rails. All sprinkler heads should bemounted so they do not extend downward more than 6.35 mm (0.25 in) from the ceiling while in the 'resting'position.

In addition, there should not be anything mounted in the ceiling (i.e. lights, A/C returns, etc) between thestationary rails. This is because the OTS longitudinal drive belt assembly is located on the movable bridge,approximately centered between the two stationary rails, and may come into contact with those ceiling-mounteditems during normal use.

Stationary rails are designed for top (ceiling) mounting. Rails can be ordered and are supplied in 10.2 cm (4 in)increments between 3.4 m (11.2 ft) and 5.64 m (18.5 ft), plus a 5.79 m (19 ft) length totaling 24 different sizes. Thechoice of length depends on room size, configuration and the possible presence of obstructions.

660.4 ±1.5mm(26 ±0.06in)

GE supplied 13mm (0.5") bolts for mounting stationary rails on Unistrut or equivalent. Distance between center ofholes is 660.4 mm (26in). Maximum load per bolt is 156kg (350 lbs), however each mounting screw must not "PULLOUT" or otherwise fail under a vertically downward dead load of 636kg (1400 lbs).

Stationary rail mountingpoints must be parallelwithin ±3mm (±0.125in)

When a 136kg (300 lb) force is appliedvertically downward, or horizontally atany stationary rail mounting point, theattachment interface must not deflectmore than 1.5mm (0.06in)

When a 22.7kg (50 lb) force is applied verticallyupward, downward or horizontally at any stationaryrail mounting point, the attachment interface mustnot deflect more than 1.5mm (0.06in)

When a 45.4kg (100 lb) force is applied verticallyupward at any stationary rail mounting point, theattachment interface must not deflect more than

1.5mm (0.06in)

Diagonals must be equalin length to within±6.5mm (±0.25in)

All mounting pointmust be located on

a common centerlinewithin ±1.5mm (±0.06in)

All mounting points must lie inthe same horizontal planewithin ±2.4mm (±0.09in)

OTS SUSPENSION RAILS MOUNTING SPECIFICATIONS

883 ±3mm(34.8 ±0.125in)

Auxilliary support railmounting points

3 m BRIDGE

1422 ±3mm(56 ±0.125in)

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Ceiling Support Details 081:20

DETAIL 2

NOT TO SCALE

DETAIL 1

NOT TO SCALE

CONTRACTOR SUPPLIED ANDINSTALLED "UNISTRUT" P1001OR EQUAL FLUSH WITHFINISHED CEILING

GE SUPPLIED UNISTRUT SPRINGNUTS WITH BOLTSGE SUPPLIED SELF-TAPPINGSCREWSGE SUPPLIED CABLE DRAPE RAIL

GE SUPPLIEDUNISTRUT SPRINGNUTS WITH BOLTS

XT RADIOGRAPHIC SUSPENSION, INBOARD MOUNTING

P1001

STATIONARY RAILUNISTRUT SYSTEM IS NOT SUPPLIEDOR INSTALLED BY GE HEALTHCARE

REFERENCE CEILING LAYOUTFOR THESE DIMENSIONS.

GE SUPPLIED CABLEDRAPE SUPPORT

CEILING HEIGHTAS SPECIFIED ONINSTALLATIONDRAWINGS

SUPPORTS TO SLAB AS REQUIRED

GE SUPPLIED CABLEDRAPE RAIL

CONTRACTOR SUPPLIED ANDINSTALLED FINISHED CEILING(CEILING & SUPPORTS MUSTNOT EXTEND BELOW FACE OFUNISTRUT)

GE SUPPLIEDCABLE DRAPESUPPORT

CONTRACTOR SUPPLIED ANDINSTALLED "UNISTRUT" P1001OR EQUAL FLUSH WITHFINISHED CEILING

GE SUPPLIED LONGITUDINALSTATIONARY RAIL

CONTRACTOR SUPPLIED ANDINSTALLED FINISHED CEILING(CEILING & SUPPORTS MUSTNOT EXTEND BELOW FACE OFUNISTRUT)

P1001

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Power Requirements 09

POWER REQUIREMENTS

POWER SUPPLY 380/400/415/440/460/480V ±10%, THREE-PHASE + GFREQUENCIES 50/60Hz ± 3HzPOWER DEMAND 97kVAMAXIMUM LINE RESISTANCEPER 2 PHASES (Ohm)

380V : 0.118 / 400V : 0.131 / 415V : 0.138440V : 0.154 / 480V : 0.185

Power supply should come into a main disconnect panel (MDP) containing the protective units and controls. The section of the supply cable should be calculated in accordance with its length and the maximum

permissible voltage drops. There must be discrimination between supply cable protective material at the beginning of the installation

(main low-voltage transformer side) and the protective devices in the MDP.

SUPPLY CHARACTERISTICS Power input must be separated from any others which may generate transients (elevators, air conditioning,

radiology rooms equipped with high speed film changers...) All equipment (lighting, power outlets, etc...) installed with GE system components must be powered

separately.

GROUND SYSTEM Equipotential : the equipotential link will be by means of an equipotential bar. This equipotential bar should

be connected to the protective earth conductors in the ducts of the non GE cableways and to additionalequipotential connections linking up all the conducting units in the rooms where GE units are located.

CABLES Power and cable installation must comply with the distribution diagram below. All cables must be isolated and flexible. Cable color codes must comply with standards for electrical installation.Case MDP furnished by GE : The cables for signals and remote control (SEO, XRL1...) will go to MDP with a pigtaillength of 1.5 m (4.9 ft), and will be connected during installation. Each conductor will be identified and isolated(screw connector).

CABLEWAYSThe general rules for laying cableways should meet the conditions laid down in current standards and regulations,with regard to: Protecting cables against water (cableways should be waterproof) Protecting cables against abnormal temperatures (proximity to heating pipes or ducts) Protecting cables against temperature shocks Replacing cables (cableways should be large enough for cables to be replaced) metal cableways should be

grounded.

POWER DISTRIBUTION

MDP Main Disconnect PanelSKL Generator (System Cabinet)SEO Emergency OFF button (Control Room), located 1.50m (4.9') above floorXRLC Warning Light ControlXRL1 Warning LightDLK1 Door Interlock Switch (needed only if required by state/local codes)

SEO

XRLC

DLK1

MDP

Main 380-480 VThree Phase Supply Ground cable (PE)

SKL

Cable SUPPLIED BYCUSTOMER

Equipment SUPPLIED BY GE

Equipment SUPPLIED BYCUSTOMER

Cable SUPPLIED BY GE

XRL1

14 Black14 White14 Green

*Black (3)*Green (1)

14 Black14 White14 Green

14 Black14 Green14 White

2 - NO. 14 Black

14 Black14 White14 Green

1 phasepower

14 White14 Red

*Black (3) *Green (1)

* Wire sized per the feedertable below

65kW JEDI FEEDER TABLEMINIMUM FEEDER WIRE SIZE IN mm² AND (AWG)

INPUTVOLTAGE

MINIMUM FEEDER WIRE LENGTH

15m (50') 30m (100') 46m (150') 61m (200') 77m (250') 92m (300') 107m(350')

122m(400')

138m(450')

380 VAC *21 (4) 27 (3) 34 (2) 53 (1/0) 34 (2) 53 (1/0) 107 (4/0) 127(250M)

152(300M)

400 VAC *21 (4) *21 (4) 34 (2) 42 (1) 34 (2) 67 (2/0) 85 (3/0) 107 (4/0) 127(250M)

420 VAC *21 (4) *21 (4) 34 (2) 42 (1) 53 (1/0) 67 (2/0) 85 (3/0) 107 (4/0) 107 (4/0)

440 VAC *21 (4) *21 (4) 27 (3) 34 (2) 42 (1) 53 (1/0) 67 (2/0) 85 (3/0) 107 (4/0)

460 VAC *21 (4) *21 (4) 27 (3) 34 (2) 42 (1) 53 (1/0) 67 (2/0) 85 (3/0) 85 (3/0)

480 VAC *21 (4) *21 (4) *21 (4) 34 (2) 42 (1) 53 (1/0) 53 (1/0) 67 (2/0) 85 (3/0)

* MINIMUM WIRE SIZE FOR CIRCUIT BREAKER, BASED ON RECOMMENDED OVERCURRENT PROTECTION

GENERAL NOTESIN ALL CASES QUALIFIED PERSONNEL MUST VERIFY THAT THE FEEDER (AT THE POINT OF TAKE-OFF) AND THE RUN TO THE RADIOLOGY

SYSTEM MEET ALL THE REQUIREMENTS STATED IN THE PIM.FOR A SINGLE UNIT INSTALLATION, THE MINIMUM TRANSFORMER SIZE IS 112.5KVa, SYNTHESIZED POWER FEED IS NOT ACCEPTABLE.

MAXIMUM ALLOWABLE TRANSIENT VOLTAGE EXCURSIONS ARE 2.5% OF RATED LINE VOLTAGE AT A MAXIMUM DURATION OF 5 CYCLESAND FREQUENCY OF 10 TIMES PER HOUR.

GROUND WIRE WILL BE SAME SIZE AS POWER CABLE. GROUND WILL RUN FROM THE EQUIPMENT BACK TO THE POWER SOURCE/MAINGROUNDING POINT AND ALWAYS TRAVEL IN THE SAME CONDUIT WITH THE FEEDERS AND NEUTRAL. NEUTRAL MUST BE TERMINATED

INSIDE THE MAIN DISCONNECT PANEL AND NOT AT ANY GE CABINET.

TEMPERATURE AND HUMIDITY SPECIFICATIONS

IN-USE CONDITIONS

STORAGE CONDITIONS

Storage longer than 90 days is not recommended.(1) Non-condensing

AIR RENEWALAccording to local standards.

NOTEIn case of using air conditioning systems that have a risk of water leakage it is recommended not to install it above electric equipment or totake measures to protect the equipment from dropping water.

EXAM ROOM CONTROL ROOM

TemperatureMin Max Min Max

15°C (59°F) 35°C (95°F) 15°C (59°F) 35°C (95°F)Temperature gradient < 10°C/h (18°F/h) < 10°C/h (18°F/h)Relative humidity (1) 30% to 60% 30% to 60%Humidity gradient < 30%/h < 30%/h

Temperature -5°C (-23°F) to +50°C (+122°F)Temperature gradient < 20°C/h (36°F/h)Relative humidity (1) 10% to 85%Humidity gradient < 30%/h

CONNECTIVITY REQUIREMENTS

Broadband Connections are necessary during the installation process and going forward to ensure full supportfrom the Engineering Teams for the customers system. Maximum performance and availability for the customerssystem is maintained and closely monitored during the lifetime of the system. Proactive and reactive maintenanceis available utilising the wide range of digital tools using the connectivity solutions listed below:

Site-to-Site VPN/GE Solution Site-to-Site VPN/Customer Solution Connection through Dedicated Service Network Internet Access - connectivity for InSite 2.0

The requirements for these connectivity solutions are explained in the broadband solutions catalogue (separatedocument).

ENVIRONMENTAL SPECIFICATIONS

LIGHT REQUIREMENTS

For the electronic ballast of fluorescent lamp in exam room, the operating frequency should be above 42KHz.

MAGNETIC INTERFERENCE

To guarantee specified imaging performance :X-ray tubes and control console equipment must be located in ambient static field of less than 10 gauss.

ACOUSTIC OUTPUT

Measured 1 m from any point in system.In-use: less than 55 dBAStand-by: less than 55 dBA

HEAT DISSIPATION DETAILS

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans HVAC - Environment 10

ORT DESCRIPTIONHEAT DISSIPATION

(kW)HEAT DISSIPATION

(BTU)STANDBY IN-USE STANDBY IN-USE

Exam Room

Elevating Table G2 0.092 0.666 312 2272Table Detector power 0.017 0.017 58 58Wall Stand G2 (Standard/Extended) 0.023 0.094 78 321WS Detector power 0.017 0.017 58 58System Cabinet 0.714 1.427 2436 4869

OTS & Collimator 0.031 0.031 106 106

Tube Rotor 0 0.160 0 546

TIB 0.002 0.020 7 68

TOTAL 0.896 2.432 3055.000 8298.000

Control Room

Z420 PC and Monitor 0.176 0.253 601 863

UPS 0.009 0.013 31 44

TOTAL 0.185 0.266 632.000 907.000

THE CUSTOMER/CONTRACTOR SHOULD:

Provide an area adjacent to the installation site for delivery and unloading of the GE equipment. Ensure that the dimensions of all doors, corridors, ceiling heights are sufficient to accommodate the

movement of GE equipment from the delivery area into the definitive installation room. Ensure that access routes for equipment will accommodate the weights of the equipment and any

transportation, lifting and rigging equipment. Ensure that all necessary arrangements for stopping and unloading on public or private property not

belonging to the customer have been made.

DELIVERY

DIMENSIONS OF DELIVERY WITH DOLLY TRANSPORT EQUIPMENTEQUIPMENT DIMENSIONS WEIGHT

STANDARD WALLSTAND

LENGTH 2111 mm (83.1 in)

270 kg (595 lb) + fixtureWIDTH 911 mm (35.9 in)

HEIGHT 1860 mm (73.2 in)

G2 TABLE

LENGTH 2400 mm (94.5 in)

440 kg (970 lb) + dollyWIDTH 940 mm (37 in)

HEIGHT 800 mm (31.5")

Pay attention to the lengths of the rails! They can be also 5.79m (19 ft) have a shipping dimension of 5.92m x178mm x 76mm (233" x 7" x 3").

G2 TABLE WITH STANDARD WALLSTAND

SCALE 1:50

Service clearance,right hand load

Service clearance,left hand load

Column serviceclearance, either

side required

Column serviceclearance, either

side required

CLEARANCE AREAS

203mm [8in]

2400mm[94.5in]

3080mm[121.3in]

*Recommended service access clearance is 915 mm.

DONGLE POSITIONING

C

B

A

Pos 2

Pos 1 Pos 3

Pos 4

Detector insertion direction

DONGLE DEFAULT LOCATION AND ADJUSTING RANGE:

Dongle shall be positioned at the wall of detector insertion direction. B is the best position which is in the middle of the wall. The height requirement of dongle is 30 cm (11.8 in) lower than the ceiling. Position "A" to "C" (around ±1 m (3.2 ft)) are acceptable locations for dongle. There shall be no obstructions in the path between dongle and detector applications.

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Equipment Details - Delivery 11

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Interconnections 12

SKL

BRIDGE

WS

WBC1

OPERATORCONSOLE

EXAM ROOM

CONTROL ROOM

INTERCONNECTIONS

TIB

DONGLE (D)

TABLE

Measured from thefront of the bridge

15.5m (50.8')20.5m (67.2')

2M Bridge:14.5m (47.6')

3M/4M Bridge:19.5m (64')

Ethernet 49.5m (162.4')

17.5m (57.4')

2.8m (9.2') 19.5m (64')

19.5m (64')23.5m (77.1')

24.5m (80.3')

19m (62')14.5m (47.6')19.5m (64')

Ethernet 49.5m (162.4')

GENERAL SPECIFICATIONS

GE is not responsible for the installation of developers and associated equipment, lighting, cassette trays andprotective screens or derivatives not mentioned in the order.

The final study contains recommendations for the location of GE equipment and associated devices, electricalwiring and room arrangements. When preparing the study, every effort has been made to consider everyaspect of the actual equipment expected to be installed.

The layout of the equipment offered by GE, the dimensions given for the premises, the details provided forthe pre-installation work and electrical power supply are given according to the information noted duringon-site study and the wishes expressed by the customer.

The room dimensions used to create the equipment layout may originate from a previous layout and may notbe accurate as they may not have been verified on site. GE cannot take any responsibility for errors due tolack of information.

Dimensions apply to finished surfaces of the room. Actual configuration may differ from options presented in some typical views or tables. If this set of final drawings has been approved by the customer, any subsequent modification of the site must

be subject to further investigation by GE about the feasibility of installing the equipment. Any reservationsmust be noted.

The equipment layout indicates the placement and interconnection of the indicated equipment components. There may be local requirements that could impact the placement of these components. It remains thecustomer's responsibility to ensure that the site and final equipment placement complies with all applicablelocal requirements.

All work required to install GE equipment must be carried out in compliance with the building regulations andthe safety standards of legal force in the country concerned.

These drawings are not to be used for actual construction purposes. The company cannot take responsibilityfor any damage resulting therefrom.

CUSTOMER RESPONSIBILITIES

It is the responsibility of the customer to prepare the site in accordance with the specifications stated in thefinal study. A detailed site readiness checklist is provided by GE. It is the responsibility of the customer toensure all requirements are fulfilled and that the site conforms to all specifications defined in the checklist and final study. The GE Project Manager of Installation (PMI) will work in cooperation with the customer to followup and ensure that actions in the checklist are complete, and if necessary, will aid in the rescheduling of thedelivery and installation date.

Prior to installation, a structrual engineer of record must ensure that the floor and ceiling is designed in such away that the loads of the installed system can be securely borne and transferred. The layout of additionalstructural elements, dimensioning and the selection of appropriate installation methods are the soleresponsibility of the structural engineer. Execution of load bearing structures supporting equipment on theceiling, floor or walls are the customer's responsibility.

RADIO-PROTECTION

Suitable radiological protection must be determined by a qualified radiological physicist in conformation withlocal regulations. GE does not take responsibility for the specification or provision of radio-protection.

DISCLAIMER

THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND APPROVED THE PLANS IN THIS DOCUMENT.

DATE NAME SIGNATURE

Rev Date:Rev: /13RAD-M076452-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Disclaimer - Site Readiness 13

Customer Name: PMI Name:

GON/SO Number: Field Service Name:Equipment: Country/City or City/State:Required site assessment milestones Date of completion (dd/mm/yyyy)1) Check site before Equipment Delivery to Storage

2) Check site before installation start

Place an "X" in either Y or N column

Site Ready Checks at Installation Y NGeneral Site Planning

Room dimensions, including ceiling height, for all Exam, Equipment/Technical & Control rooms meets GE specifications.

Ceiling support structure, if indicated on the GE drawing, is in the correct location and at the correct height according to theOriginal Equipment Manufacturer specifications. Levelness and spacing has been measured, and is ready for the installationof any GE supplied components. Overhead support Structure has been confirmed with customer/contractor to meet requiredGE provided criteria.

Rooms that will contain equipment, including staging areas if applicable, are construction debris free. Precautions must betaken to prevent debris from entering rooms containing equipment.

Finished ceiling is installed. If applicable ceiling tiles installed per PMI discretion.

Adequate delivery route from truck to final place of installation has been reviewed with all stakeholders, allcommunications/notifications have occurred, arrangements have been made for special handling (rigging, elevator, fork lift,etc.). All floors along delivery route will support weight of the equipment, temporary reinforcements arranged if needed.

System power & grounding (PDB/MDP) is available as per GE specifications, installed at point of final connection and ready touse. Lock Out Tag Out is available.

System power and grounded audit has been scheduled to be completed during installation of equipment. (If Required) GEHCPM to confirmed if needed.

Adequate room illumination installed and working.

Cable ways (floor/wall/ceiling/Access Flooring) are available for installation of GE cables and are of correct length anddiameter. Cable ways routes per GE Final drawings and cable access openings areas installed at a time determined by GEHCPM. Surface floor duct can be installed at time of system installation.

HVAC systems Installed, and the site meets minimum environmental operational system requirements.

Network outlets installed and computer network available and working.

Hospital IT/connectivity contacts have been engaged and information has been added to Project management tool. (IfRequired)

Floor levelness/flatness is measured and within tolerance, and there are no visible defects per GEHC specifications. FloorStrength and thickness have been discussed with customer/contractor and they have confirmed GE requirements are met.

Customer supplied countertops where GE equipment will be installed are in place.

Specific for CT & X-ray

Doors and windows complete or scheduled to be installed. If applicable, radiation protection (shielding) finished &radioprotection regulatory approval for installation obtained.

Status of work

General comments

System can be delivered PMI signature

Site ready for installation FS signature: optional

GLOBAL SITE READINESS CHECKLIST (DI)DOC1809666 Rev. 5

01 - Cover Sheet02 - Equipment Layout03 - Equipment Layout Sections04 - Floor - Electrical Layout05 - Floor Structural Details06 - Detailed Ceiling Layout07 - Ceiling Structural Details08 - Ceiling Support Details09 - Power Requirements10 - HVAC - Environment

11 - Equipment Details - Delivery12 - Interconnections13 - Disclaimer - Site Readiness

New Orleans, LA

Brian Ferrell

DISCOVERY XR 656 PLUS (G2)

REK

GE Healthcare

DateDrawn by Verified by S.O. (GON) PIM Ref & Rev Rev

GE does not take responsibility for any damages resulting from changes on drawings made by others. Errors may occur by not referring to the completeset of final issue drawing. GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings, however caused. Alldimensions are in millimeters unless otherwise specified. Do not scale from printed pdf files. GE accepts no responsibility or liability for defective work

due to scaling from these drawings. /13RAD-M076455-01B.DWG

+1 337 371 7202

[email protected]

REK 5643834-1EN Rev 5 B

USA

1:48

4599254 12/FEB/2018

Childrens Hospital of New Orleans

Scale

REV DATE MODIFICATIONSA 21.Nov.17 Final version 01 based on DC-36036B 12/FEB/2018 REVISION BASED ON DC-44443; CHANGE TO GON, POWER REQUIREMENTS,

POWER DISTRIBUTION, CONTROL WALL HEIGHT, PIM REVISION

A mandatory component of this drawing set is the GE Healthcare Pre Installation manual. Failure to reference the Pre Installation manual will result inincomplete documentation required for site design and preparation.

Pre Installation documents for GE Healthcare products can be accessed on the web at: www.gehealthcare.com/siteplanning

SheetFormat A3 Concession

####

RAD ROOM 2 FINAL

01

EQUIPMENT LAYOUT

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Equipment Layout 021:48

22'-9

"

11'-0"

16'-2"

6'-6

"

28'-9

"

5'-7

"

9'-3"6'-2"

12'-5

"

0'-5

"4'

-3"

4'-2"

7'-2"

2'-11"

LEGEND

ARCHITECTURAL ELEMENTS

GE SUPPLIEDEQUIPMENT MOVEMENT

SERVICE CLEARANCE

MAIN AXIS

DIMENSIONS

CUSTOMER SUPPLIED ITEMS

ITEM DESCRIPTION DIMENSIONSLxWxH (in)

WEIGHT(lb)

1 ELEVATING TABLE G2 94.5x36.7x33.5 970

2 STANDARD WALL STAND G2 27x52.9x103.5 595

3 18'-2" RAILS FOR 3 m BRIDGE 218x131.5x3.74 319

4 OTS WITH 3 m BRIDGE 121.1x25.8x102.6 772

5 SYSTEM CABINET/GENERATOR 27.8x26.1x58.1 705

6 TETHER INTERFACE BOX (TIB) 10x5.75x12.2 15.40

7 MAIN DISCONNECT PANEL (MDP) 16x6x24 80

8 GRID HOLDER 20.15x8x19.9 30.40

9 IMAGE PASTING BARRIER 42.7x29x81.1 120

10 DONGLE ASSEMBLY 3.29x6.1x3.27 1.80

11 OPERATOR CONSOLE 47.2x29.5x47.4 79

12 PARTIAL UPS 8.4x16.1x12.8 76

13 WALL BOX 14.6x6.7x4.4 11

EXAM ROOM HEIGHT

CONTROL WALL HEIGHT

EXAM ROOM CEILING HEIGHT

7'-0"

9'-6"

Auto Image Pasting at Wall Stand

Volu

meR

AD

Vertical at Wall stand

Horizontal at Table

Horizontal at Extended Arm Wall stand

Cross-Table at Wall stand

The chart shows the applications possible to perform with the present equipment positioning,however the sales contract may not include all of them.

APPLICATIONS

Auto Image Pasting at Table YES

YES

NO

YES

YES

YES

REAR TO FRONT CROSS TABLE SHOT TO TABLE CENTER LINE.

THE FOLLOWING SHOTS ARE NOT AVAILABLE IN THIS LAYOUT:

1

3

4

2

5

6

7

8

9

11

10

1213

RAD EXAMROOM 2

CONTROLAREA

DRESSINGROOM

15'-10"

SIDE VIEW A-A'

FRONT VIEW B-B'

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Equipment Layout Sections 031:48

EXAM ROOM CEILING HEIGHTS

RECOMMENDED AND MINIMUM ROOM HEIGHTSCONFIGURATION SPECIFICATION CEILING HEIGHT

2m, 3m or 4-meter Bridge Recommended 2900 mm (114 in)

2m, 3m Bridge with Advanced Applications Minimum 2700 mm (106 in)

2m, 3m Bridge without Advanced Applications Minimum 2620 mm (103 in)

4m Bridge with Advanced Applications Minimum 2700 mm (106.25 in)

4m Bridge without Advanced Applications Minimum 2630 mm (103.25 in)

3m Bridge with Wallstand at Front Position Minimum 2870 mm (113 in)

4m Bridge with Wallstand at Front Position Minimum 2880 mm (113.25 in)

+ 9'-6"

± 0'-0"

+ 9'-6"

± 0'-0"1

5

6

72

34

8

10

12

33

4

6

7

5

FLOOR - ELECTRICAL LAYOUT

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Floor - Electrical Layout 041:48

NOTE: IF DOORWAY LEADS TO ANUNCONTROLLED AREA, A DOORINTERLOCK SWITCH ANDWARNING LIGHT WILL BEREQUIRED.

1

4

2

15

13

14

16

6

7

5

7 12

11

8

9

10

12'-5

"

9'-2

"

7'-2"

4'-2"

4'-9

"

0'-8

"

RAD EXAMROOM 2

CONTROL AREA

1'-8"

ITEM QTY DESCRIPTION

1 Table anchoring (see Floor Structural Details)

2 Wall Stand anchoring (see Floor Structural Details)

3 Main disconnect panel @ 5'-0" A.F.F.

4 8" x 8" x 6" Flush floor junction box

5 1" Conduit running thru-ceiling

6 1.5" Conduit running thru-ceiling

7 2.5" Conduit running thru-ceiling

8 18" x 18" x 4" Flush junction box @ 0'-9" A.F.F. (TYP)[SYSTEM CABINET]

9 18" x 3.5" Vertical wall duct with minimum 2 dividers

10 6" deep Box above ceiling, size as required

11 2" Conduit running thru-floor

12 2.5" Conduit running thru-floor

13 6" x 6" x 4" Flush junction box @ 4'-0" A.F.F. (TYP)[TIB]

14 6" x 6" x 4" Flush junction box @ 1'-0" B.F.C. (TYP)[DONGLE]

15 8" x 8" x 4" Flush junction box @ 1'-6" A.F.F. (TYP)[WALL STAND]

16 12" x 6" x 4" Flush junction box @ 1'-6" A.F.F. (TYP)[CONSOLE]

Basic system

RJ 45 network socketSystem emergency off (SEO), (recommended height 4'-0" above floor)X-Ray room warning light control panel - Available from GEHC, Call: 800-279-7925 orlocal GE Installation Project ManagerX-Ray ON lamp (L1) - 24VDoor interlock switch (needed only if required by state/local codes)120V Duplex hospital grade outlet

CONDUITS REQUIRED FOR BASE SYSTEM

WARNING LIGHT CONTROL TO WARNING LIGHT ONE 16mm [1/2 in] CONDUIT

WARNING LIGHT CONTROL TO GENERATOR ONE 16mm [1/2 in] CONDUIT

WARNING LIGHT CONTROL TO 120-V 1Ø POWER CONDUIT AS REQUIRED

MAIN DISCONNECT PANEL TO GENERATOR ONE CONDUIT AS REQUIRED

MAIN DISCONNECT PANEL TO EMERGENCY OFF(CONTROL ROOM) ONE 16mm [1/2 in] CONDUIT

MAIN DISCONNECT PANEL TO FEEDER ONE CONDUIT AS REQUIRED

DOOR INTERLOCK SWITCH TO GENERATOR ONE 16mm [1/2 in] CONDUIT

GENERATOR TO TETHER INTERFACE BOX ONE 50mm [2 in] CONDUIT

CONTROL ROOM WALLBOX TO TETHER INTERFACE BOX ONE 25mm [1 in] CONDUIT

GENERATOR TO DONGLE ASSEMBLY ONE 25mm [1 in] CONDUIT

CONTROL ROOM WALLBOX TO DONGLE ASSEMBLY ONE 50mm [2 in] CONDUIT

1'-6

" 4'-0

"

3

CABLE MANAGEMENT

CONDUIT IN THE FLOOR

NOT TO SCALE

Removable coverplate

VERTICAL DUCT ON WALL

TABLE ANCHORING

Anchors M8x190(7.5 in) suppliedby GE

TABLE BASE

The floor bearing the system is recommended to be concrete and the thickness to be determined by a StructuralEngineer to properly support the equipment loads. The supplied anchors require a minimum embedment of 90mm (3.54 in) into the concrete. If the floor thickness is less than 95 mm (3.74 in), it is recommended that the unitbe secured using a through-bolt method with a reinforcement plate on the back side.

SCALE 1:20

150mm[5.9in]

123mm[4.8in]

93mm [3.7in]

860mm[33.9in]920mm[36.2in]

460mm[18.1in]

460mm[18.1in] Cable inlet area

TABLE AXIS

WALLSTAND AXIS

Anchors M10x150 mm(5.9 in) supplied by GE

SCALE 1:10

Cable inlet area

WALLSTAND BASE

WALLSTAND ANCHORING

Concrete area for wall stand installation should be 1 m² (10.76 ft²).

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Floor Structural Details 05

CEILING LAYOUT

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Ceiling Layout 061:48

0'-8"3'-1"4'-8"

ITEM DESCRIPTION

STANDARD 3 RAILS SYSTEM

1

UNISTRUT OR EQUIVALENT SUPPORT IN CEILING FOR FASTENING CABLE DRAPE RAIL.SUPPORTS TO RUN CONTINUOUS WITH NO FITTINGS EXTENDING BELOW FACE OF UNISTRUTCHANNEL, RUN WALL TO WALL, BE PARALLEL, SQUARE, AND IN THE SAME HORIZONTALPLANE, FLUSH WITH THE FINISHED CEILING.RAILS ARE MOUNTED TO THESE SUPPORTS EVERY2'-2" AND REQUIRE 50 LBS. PR BOLT LOAD. METHODS OF SUPPORT THAT PERMITATTACHMENT TO STRUCTURAL STEEL OR THROUGH BOLTS IN THE CONCRETE SHOULD BEFAVORED. DO NOT USE SCREW ANCHORS IN DIRECT TENSION

2

UNISTRUT OR EQUIVALENT SUPPORT FOR FASTENING CEILING SUPPORTED EQUIPMENT.SUPPORTS TO RUN CONTINUOUS WITH NO FITTINGS EXTENDING BELOW FACE OF UNISTRUTCHANNEL, RUN WALL TO WALL, BE PARALLEL, SQUARE, AND IN THE SAME HORIZONTAL PLANE,FLUSH WITH THE FINISHED CEILING. RAILS ARE MOUNTED TO THESE SUPPORTS EVERY 2'-2"AND REQUIRE 350 LBS. (597 LBS. IN SEISMIC REGIONS) PER BOLT LOAD. METHODS OF SUPPORTTHAT PERMIT ATTACHMENT TO STRUCTURAL STEEL OR THROUGH BOLTS IN CONCRETE SHOULDBE FAVORED. DO NOT USE SCREW ANCHORS IN DIRECT TENSION.

12

RAD EXAMROOM 2

CONTROL AREA

FOCAL SPOT TRAVEL

SCALE 1:50

Focal spottravel area

3 m BRIDGE

4412mm[173.7in]

5042mm[198.5in]

Suspension columntravel area

645mm[25.4in]

480mm[18.9in]

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Ceiling Structural Details 07

CEILING REQUIREMENTS

To allow installation of the stationary rail cross-members, clearance is required between the ends of the stationaryrails and the walls.

It is recommended that sprinkler heads not be placed between the stationary rails. All sprinkler heads should bemounted so they do not extend downward more than 6.35 mm (0.25 in) from the ceiling while in the 'resting'position.

In addition, there should not be anything mounted in the ceiling (i.e. lights, A/C returns, etc) between thestationary rails. This is because the OTS longitudinal drive belt assembly is located on the movable bridge,approximately centered between the two stationary rails, and may come into contact with those ceiling-mounteditems during normal use.

Stationary rails are designed for top (ceiling) mounting. Rails can be ordered and are supplied in 10.2 cm (4 in)increments between 3.4 m (11.2 ft) and 5.64 m (18.5 ft), plus a 5.79 m (19 ft) length totaling 24 different sizes. Thechoice of length depends on room size, configuration and the possible presence of obstructions.

660.4 ±1.5mm(26 ±0.06in)

GE supplied 13mm (0.5") bolts for mounting stationary rails on Unistrut or equivalent. Distance between center ofholes is 660.4 mm (26in). Maximum load per bolt is 156kg (350 lbs), however each mounting screw must not "PULLOUT" or otherwise fail under a vertically downward dead load of 636kg (1400 lbs).

Stationary rail mountingpoints must be parallelwithin ±3mm (±0.125in)

When a 136kg (300 lb) force is appliedvertically downward, or horizontally atany stationary rail mounting point, theattachment interface must not deflectmore than 1.5mm (0.06in)

When a 22.7kg (50 lb) force is applied verticallyupward, downward or horizontally at any stationaryrail mounting point, the attachment interface mustnot deflect more than 1.5mm (0.06in)

When a 45.4kg (100 lb) force is applied verticallyupward at any stationary rail mounting point, theattachment interface must not deflect more than

1.5mm (0.06in)

Diagonals must be equalin length to within±6.5mm (±0.25in)

All mounting pointmust be located on

a common centerlinewithin ±1.5mm (±0.06in)

All mounting points must lie inthe same horizontal planewithin ±2.4mm (±0.09in)

OTS SUSPENSION RAILS MOUNTING SPECIFICATIONS

883 ±3mm(34.8 ±0.125in)

Auxilliary support railmounting points

3 m BRIDGE

1422 ±3mm(56 ±0.125in)

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Ceiling Support Details 081:20

DETAIL 2

NOT TO SCALE

DETAIL 1

NOT TO SCALE

CONTRACTOR SUPPLIED ANDINSTALLED "UNISTRUT" P1001OR EQUAL FLUSH WITHFINISHED CEILING

GE SUPPLIED UNISTRUT SPRINGNUTS WITH BOLTSGE SUPPLIED SELF-TAPPINGSCREWSGE SUPPLIED CABLE DRAPE RAIL

GE SUPPLIEDUNISTRUT SPRINGNUTS WITH BOLTS

XT RADIOGRAPHIC SUSPENSION, INBOARD MOUNTING

P1001

STATIONARY RAILUNISTRUT SYSTEM IS NOT SUPPLIEDOR INSTALLED BY GE HEALTHCARE

REFERENCE CEILING LAYOUTFOR THESE DIMENSIONS.

GE SUPPLIED CABLEDRAPE SUPPORT

CEILING HEIGHTAS SPECIFIED ONINSTALLATIONDRAWINGS

SUPPORTS TO SLAB AS REQUIRED

GE SUPPLIED CABLEDRAPE RAIL

CONTRACTOR SUPPLIED ANDINSTALLED FINISHED CEILING(CEILING & SUPPORTS MUSTNOT EXTEND BELOW FACE OFUNISTRUT)

GE SUPPLIEDCABLE DRAPESUPPORT

CONTRACTOR SUPPLIED ANDINSTALLED "UNISTRUT" P1001OR EQUAL FLUSH WITHFINISHED CEILING

GE SUPPLIED LONGITUDINALSTATIONARY RAIL

CONTRACTOR SUPPLIED ANDINSTALLED FINISHED CEILING(CEILING & SUPPORTS MUSTNOT EXTEND BELOW FACE OFUNISTRUT)

P1001

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Power Requirements 09

POWER REQUIREMENTS

POWER SUPPLY 380/400/415/440/460/480V ±10%, THREE-PHASE + GFREQUENCIES 50/60Hz ± 3HzPOWER DEMAND 97kVAMAXIMUM LINE RESISTANCEPER 2 PHASES (Ohm)

380V : 0.118 / 400V : 0.131 / 415V : 0.138440V : 0.154 / 480V : 0.185

Power supply should come into a main disconnect panel (MDP) containing the protective units and controls. The section of the supply cable should be calculated in accordance with its length and the maximum

permissible voltage drops. There must be discrimination between supply cable protective material at the beginning of the installation

(main low-voltage transformer side) and the protective devices in the MDP.

SUPPLY CHARACTERISTICS Power input must be separated from any others which may generate transients (elevators, air conditioning,

radiology rooms equipped with high speed film changers...) All equipment (lighting, power outlets, etc...) installed with GE system components must be powered

separately.

GROUND SYSTEM Equipotential : the equipotential link will be by means of an equipotential bar. This equipotential bar should

be connected to the protective earth conductors in the ducts of the non GE cableways and to additionalequipotential connections linking up all the conducting units in the rooms where GE units are located.

CABLES Power and cable installation must comply with the distribution diagram below. All cables must be isolated and flexible. Cable color codes must comply with standards for electrical installation.Case MDP furnished by GE : The cables for signals and remote control (SEO, XRL1...) will go to MDP with a pigtaillength of 1.5 m (4.9 ft), and will be connected during installation. Each conductor will be identified and isolated(screw connector).

CABLEWAYSThe general rules for laying cableways should meet the conditions laid down in current standards and regulations,with regard to: Protecting cables against water (cableways should be waterproof) Protecting cables against abnormal temperatures (proximity to heating pipes or ducts) Protecting cables against temperature shocks Replacing cables (cableways should be large enough for cables to be replaced) metal cableways should be

grounded.

POWER DISTRIBUTION

MDP Main Disconnect PanelSKL Generator (System Cabinet)SEO Emergency OFF button (Control Room), located 1.50m (4.9') above floorXRLC Warning Light ControlXRL1 Warning LightDLK1 Door Interlock Switch (needed only if required by state/local codes)

SEO

XRLC

DLK1

MDP

Main 380-480 VThree Phase Supply Ground cable (PE)

SKL

Cable SUPPLIED BYCUSTOMER

Equipment SUPPLIED BY GE

Equipment SUPPLIED BYCUSTOMER

Cable SUPPLIED BY GE

XRL1

14 Black14 White14 Green

*Black (3)*Green (1)

14 Black14 White14 Green

14 Black14 Green14 White

2 - NO. 14 Black

14 Black14 White14 Green

1 phasepower

14 White14 Red

*Black (3) *Green (1)

* Wire sized per the feedertable below

65kW JEDI FEEDER TABLEMINIMUM FEEDER WIRE SIZE IN mm² AND (AWG)

INPUTVOLTAGE

MINIMUM FEEDER WIRE LENGTH

15m (50') 30m (100') 46m (150') 61m (200') 77m (250') 92m (300') 107m(350')

122m(400')

138m(450')

380 VAC *21 (4) 27 (3) 34 (2) 53 (1/0) 34 (2) 53 (1/0) 107 (4/0) 127(250M)

152(300M)

400 VAC *21 (4) *21 (4) 34 (2) 42 (1) 34 (2) 67 (2/0) 85 (3/0) 107 (4/0) 127(250M)

420 VAC *21 (4) *21 (4) 34 (2) 42 (1) 53 (1/0) 67 (2/0) 85 (3/0) 107 (4/0) 107 (4/0)

440 VAC *21 (4) *21 (4) 27 (3) 34 (2) 42 (1) 53 (1/0) 67 (2/0) 85 (3/0) 107 (4/0)

460 VAC *21 (4) *21 (4) 27 (3) 34 (2) 42 (1) 53 (1/0) 67 (2/0) 85 (3/0) 85 (3/0)

480 VAC *21 (4) *21 (4) *21 (4) 34 (2) 42 (1) 53 (1/0) 53 (1/0) 67 (2/0) 85 (3/0)

* MINIMUM WIRE SIZE FOR CIRCUIT BREAKER, BASED ON RECOMMENDED OVERCURRENT PROTECTION

GENERAL NOTESIN ALL CASES QUALIFIED PERSONNEL MUST VERIFY THAT THE FEEDER (AT THE POINT OF TAKE-OFF) AND THE RUN TO THE RADIOLOGY

SYSTEM MEET ALL THE REQUIREMENTS STATED IN THE PIM.FOR A SINGLE UNIT INSTALLATION, THE MINIMUM TRANSFORMER SIZE IS 112.5KVa, SYNTHESIZED POWER FEED IS NOT ACCEPTABLE.

MAXIMUM ALLOWABLE TRANSIENT VOLTAGE EXCURSIONS ARE 2.5% OF RATED LINE VOLTAGE AT A MAXIMUM DURATION OF 5 CYCLESAND FREQUENCY OF 10 TIMES PER HOUR.

GROUND WIRE WILL BE SAME SIZE AS POWER CABLE. GROUND WILL RUN FROM THE EQUIPMENT BACK TO THE POWER SOURCE/MAINGROUNDING POINT AND ALWAYS TRAVEL IN THE SAME CONDUIT WITH THE FEEDERS AND NEUTRAL. NEUTRAL MUST BE TERMINATED

INSIDE THE MAIN DISCONNECT PANEL AND NOT AT ANY GE CABINET.

TEMPERATURE AND HUMIDITY SPECIFICATIONS

IN-USE CONDITIONS

STORAGE CONDITIONS

Storage longer than 90 days is not recommended.(1) Non-condensing

AIR RENEWALAccording to local standards.

NOTEIn case of using air conditioning systems that have a risk of water leakage it is recommended not to install it above electric equipment or totake measures to protect the equipment from dropping water.

EXAM ROOM CONTROL ROOM

TemperatureMin Max Min Max

15°C (59°F) 35°C (95°F) 15°C (59°F) 35°C (95°F)Temperature gradient < 10°C/h (18°F/h) < 10°C/h (18°F/h)Relative humidity (1) 30% to 60% 30% to 60%Humidity gradient < 30%/h < 30%/h

Temperature -5°C (-23°F) to +50°C (+122°F)Temperature gradient < 20°C/h (36°F/h)Relative humidity (1) 10% to 85%Humidity gradient < 30%/h

CONNECTIVITY REQUIREMENTS

Broadband Connections are necessary during the installation process and going forward to ensure full supportfrom the Engineering Teams for the customers system. Maximum performance and availability for the customerssystem is maintained and closely monitored during the lifetime of the system. Proactive and reactive maintenanceis available utilising the wide range of digital tools using the connectivity solutions listed below:

Site-to-Site VPN/GE Solution Site-to-Site VPN/Customer Solution Connection through Dedicated Service Network Internet Access - connectivity for InSite 2.0

The requirements for these connectivity solutions are explained in the broadband solutions catalogue (separatedocument).

ENVIRONMENTAL SPECIFICATIONS

LIGHT REQUIREMENTS

For the electronic ballast of fluorescent lamp in exam room, the operating frequency should be above 42KHz.

MAGNETIC INTERFERENCE

To guarantee specified imaging performance :X-ray tubes and control console equipment must be located in ambient static field of less than 10 gauss.

ACOUSTIC OUTPUT

Measured 1 m from any point in system.In-use: less than 55 dBAStand-by: less than 55 dBA

HEAT DISSIPATION DETAILS

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans HVAC - Environment 10

ORT DESCRIPTIONHEAT DISSIPATION

(kW)HEAT DISSIPATION

(BTU)STANDBY IN-USE STANDBY IN-USE

Exam Room

Elevating Table G2 0.092 0.666 312 2272Table Detector power 0.017 0.017 58 58Wall Stand G2 (Standard/Extended) 0.023 0.094 78 321WS Detector power 0.017 0.017 58 58System Cabinet 0.714 1.427 2436 4869

OTS & Collimator 0.031 0.031 106 106

Tube Rotor 0 0.160 0 546

TIB 0.002 0.020 7 68

TOTAL 0.896 2.432 3055.000 8298.000

Control Room

Z420 PC and Monitor 0.176 0.253 601 863

UPS 0.009 0.013 31 44

TOTAL 0.185 0.266 632.000 907.000

THE CUSTOMER/CONTRACTOR SHOULD:

Provide an area adjacent to the installation site for delivery and unloading of the GE equipment. Ensure that the dimensions of all doors, corridors, ceiling heights are sufficient to accommodate the

movement of GE equipment from the delivery area into the definitive installation room. Ensure that access routes for equipment will accommodate the weights of the equipment and any

transportation, lifting and rigging equipment. Ensure that all necessary arrangements for stopping and unloading on public or private property not

belonging to the customer have been made.

DELIVERY

DIMENSIONS OF DELIVERY WITH DOLLY TRANSPORT EQUIPMENTEQUIPMENT DIMENSIONS WEIGHT

STANDARD WALLSTAND

LENGTH 2111 mm (83.1 in)

270 kg (595 lb) + fixtureWIDTH 911 mm (35.9 in)

HEIGHT 1860 mm (73.2 in)

G2 TABLE

LENGTH 2400 mm (94.5 in)

440 kg (970 lb) + dollyWIDTH 940 mm (37 in)

HEIGHT 800 mm (31.5")

Pay attention to the lengths of the rails! They can be also 5.79m (19 ft) have a shipping dimension of 5.92m x178mm x 76mm (233" x 7" x 3").

G2 TABLE WITH STANDARD WALLSTAND

SCALE 1:50

Service clearance,right hand load

Service clearance,left hand load

Column serviceclearance, either

side required

Column serviceclearance, either

side required

CLEARANCE AREAS

203mm [8in]

2400mm[94.5in]

3080mm[121.3in]

*Recommended service access clearance is 915 mm.

DONGLE POSITIONING

C

B

A

Pos 2

Pos 1 Pos 3

Pos 4

Detector insertion direction

DONGLE DEFAULT LOCATION AND ADJUSTING RANGE:

Dongle shall be positioned at the wall of detector insertion direction. B is the best position which is in the middle of the wall. The height requirement of dongle is 30 cm (11.8 in) lower than the ceiling. Position "A" to "C" (around ±1 m (3.2 ft)) are acceptable locations for dongle. There shall be no obstructions in the path between dongle and detector applications.

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Equipment Details - Delivery 11

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Interconnections 12

SKL

BRIDGE

WS

WBC1

OPERATORCONSOLE

EXAM ROOM

CONTROL ROOM

INTERCONNECTIONS

TIB

DONGLE (D)

TABLE

Measured from thefront of the bridge

15.5m (50.8')20.5m (67.2')

2M Bridge:14.5m (47.6')

3M/4M Bridge:19.5m (64')

Ethernet 49.5m (162.4')

17.5m (57.4')

2.8m (9.2') 19.5m (64')

19.5m (64')23.5m (77.1')

24.5m (80.3')

19m (62')14.5m (47.6')19.5m (64')

Ethernet 49.5m (162.4')

GENERAL SPECIFICATIONS

GE is not responsible for the installation of developers and associated equipment, lighting, cassette trays andprotective screens or derivatives not mentioned in the order.

The final study contains recommendations for the location of GE equipment and associated devices, electricalwiring and room arrangements. When preparing the study, every effort has been made to consider everyaspect of the actual equipment expected to be installed.

The layout of the equipment offered by GE, the dimensions given for the premises, the details provided forthe pre-installation work and electrical power supply are given according to the information noted duringon-site study and the wishes expressed by the customer.

The room dimensions used to create the equipment layout may originate from a previous layout and may notbe accurate as they may not have been verified on site. GE cannot take any responsibility for errors due tolack of information.

Dimensions apply to finished surfaces of the room. Actual configuration may differ from options presented in some typical views or tables. If this set of final drawings has been approved by the customer, any subsequent modification of the site must

be subject to further investigation by GE about the feasibility of installing the equipment. Any reservationsmust be noted.

The equipment layout indicates the placement and interconnection of the indicated equipment components. There may be local requirements that could impact the placement of these components. It remains thecustomer's responsibility to ensure that the site and final equipment placement complies with all applicablelocal requirements.

All work required to install GE equipment must be carried out in compliance with the building regulations andthe safety standards of legal force in the country concerned.

These drawings are not to be used for actual construction purposes. The company cannot take responsibilityfor any damage resulting therefrom.

CUSTOMER RESPONSIBILITIES

It is the responsibility of the customer to prepare the site in accordance with the specifications stated in thefinal study. A detailed site readiness checklist is provided by GE. It is the responsibility of the customer toensure all requirements are fulfilled and that the site conforms to all specifications defined in the checklist and final study. The GE Project Manager of Installation (PMI) will work in cooperation with the customer to followup and ensure that actions in the checklist are complete, and if necessary, will aid in the rescheduling of thedelivery and installation date.

Prior to installation, a structrual engineer of record must ensure that the floor and ceiling is designed in such away that the loads of the installed system can be securely borne and transferred. The layout of additionalstructural elements, dimensioning and the selection of appropriate installation methods are the soleresponsibility of the structural engineer. Execution of load bearing structures supporting equipment on theceiling, floor or walls are the customer's responsibility.

RADIO-PROTECTION

Suitable radiological protection must be determined by a qualified radiological physicist in conformation withlocal regulations. GE does not take responsibility for the specification or provision of radio-protection.

DISCLAIMER

THE UNDERSIGNED, HEREBY CERTIFIES THAT I HAVE READ AND APPROVED THE PLANS IN THIS DOCUMENT.

DATE NAME SIGNATURE

Rev Date:Rev: /13RAD-M076455-01B.DWG B 12/FEB/2018DISCOVERY XR 656 PLUS (G2)Childrens Hospital of New Orleans Disclaimer - Site Readiness 13

Customer Name: PMI Name:

GON/SO Number: Field Service Name:Equipment: Country/City or City/State:Required site assessment milestones Date of completion (dd/mm/yyyy)1) Check site before Equipment Delivery to Storage

2) Check site before installation start

Place an "X" in either Y or N column

Site Ready Checks at Installation Y NGeneral Site Planning

Room dimensions, including ceiling height, for all Exam, Equipment/Technical & Control rooms meets GE specifications.

Ceiling support structure, if indicated on the GE drawing, is in the correct location and at the correct height according to theOriginal Equipment Manufacturer specifications. Levelness and spacing has been measured, and is ready for the installationof any GE supplied components. Overhead support Structure has been confirmed with customer/contractor to meet requiredGE provided criteria.

Rooms that will contain equipment, including staging areas if applicable, are construction debris free. Precautions must betaken to prevent debris from entering rooms containing equipment.

Finished ceiling is installed. If applicable ceiling tiles installed per PMI discretion.

Adequate delivery route from truck to final place of installation has been reviewed with all stakeholders, allcommunications/notifications have occurred, arrangements have been made for special handling (rigging, elevator, fork lift,etc.). All floors along delivery route will support weight of the equipment, temporary reinforcements arranged if needed.

System power & grounding (PDB/MDP) is available as per GE specifications, installed at point of final connection and ready touse. Lock Out Tag Out is available.

System power and grounded audit has been scheduled to be completed during installation of equipment. (If Required) GEHCPM to confirmed if needed.

Adequate room illumination installed and working.

Cable ways (floor/wall/ceiling/Access Flooring) are available for installation of GE cables and are of correct length anddiameter. Cable ways routes per GE Final drawings and cable access openings areas installed at a time determined by GEHCPM. Surface floor duct can be installed at time of system installation.

HVAC systems Installed, and the site meets minimum environmental operational system requirements.

Network outlets installed and computer network available and working.

Hospital IT/connectivity contacts have been engaged and information has been added to Project management tool. (IfRequired)

Floor levelness/flatness is measured and within tolerance, and there are no visible defects per GEHC specifications. FloorStrength and thickness have been discussed with customer/contractor and they have confirmed GE requirements are met.

Customer supplied countertops where GE equipment will be installed are in place.

Specific for CT & X-ray

Doors and windows complete or scheduled to be installed. If applicable, radiation protection (shielding) finished &radioprotection regulatory approval for installation obtained.

Status of work

General comments

System can be delivered PMI signature

Site ready for installation FS signature: optional

GLOBAL SITE READINESS CHECKLIST (DI)DOC1809666 Rev. 5