PROTOCOL Europe - Diagenode s.a. / orders@diagenode.com / info@diagenode.com // North America - Diagenode Inc. / orders.na@diagenode.com / info.na@diagenode.com www.diagenode.com DNA Shearing for Bioruptor ® NGS Standard operating conditions Sample volume: 100 µl Tubes: Bioruptor ® NGS 0.65 ml Microtubes for DNA Shearing (Cat. No. WA-005-0500) Tube holder: 0.5/0.65 ml tube holder (Cat. No. UCD-pack 0.5) for 12 x 0.65 ml tubes Sonication buffer: TE (10 mM Tris, 1mM EDTA, pH 7.5 - 8.0) DNA concentration: 1-20 ng/µl (10 ng/µl recommended) Samples are vortexed (10-15 sec) and centrifuged (10 sec) before shearing. For optimal results samples should be stored on ice during 10-15 minutes prior to sonication. Temperature: 4°C – Bioruptor ® Water Cooler (Cat. No. BioAcc-Cool) & Single cycle valve for Water Cooler (Cat. No. VB-100-0001) Power setting: H position (High) Sonication cycle & total sonication time: varies depending on desired DNA size (see table) Note: Recommended protocols are subject to change without notice. Additional protocols are available on demand. 125 bp 135 bp 183 bp 200 bp 234 bp 319 bp 476 bp 696 bp 848 bp 1036 bp 1230 bp Programmable DNA size distributions, excellent reproducibility, and high dsDNA yields with Bioruptor ® NGS Figure shows different DNA size distributions of sheared genomic DNA produced by varying the duration of sonication using power setting high (H). The different curves depict a specific Bioruptor ® NGS run, optimized to produce specific mean sizes and size ranges for Next- Generation sequencing. All samples were analyzed on Bioanalyzer 2100 using DNA High Sensitivity chip.

Upload: others

Post on 16-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DNA Shearing for Bioruptor NGS · DNase free enzymatic digestion since they might generate a biased fragment distribution profile on microfluidics-based platform (e.g. Agilent Bioanalyzer)

PR

OT

OC

OL

Europe - Diagenode s.a. / [email protected] / [email protected] // North America - Diagenode Inc. / [email protected] / [email protected]

www.diagenode.com

DNA Shearing for Bioruptor® NGS

Standard operating conditions

Sample volume: 100µl

Tubes: Bioruptor®NGS0.65mlMicrotubesforDNAShearing(Cat.No.WA-005-0500)

Tube holder: 0.5/0.65mltubeholder(Cat.No.UCD-pack0.5)for12x0.65mltubes

Sonication buffer: TE(10mMTris,1mMEDTA,pH7.5-8.0)

DNA concentration:1-20ng/µl(10ng/µlrecommended)

Samplesarevortexed(10-15sec)andcentrifuged(10sec)beforeshearing.For optimal results samples should be stored on ice during 10-15 minutes prior to sonication.

Temperature: 4°C–Bioruptor®WaterCooler(Cat.No.BioAcc-Cool)&SinglecyclevalveforWaterCooler(Cat.No.VB-100-0001)

Power setting: Hposition(High)

Sonication cycle & total sonication time:variesdependingondesiredDNAsize(seetable)

Note: Recommended protocols are subject to change without notice. Additional protocols areavailableondemand.

125 bp

135 bp

183 bp

200 bp

234 bp

319 bp

476 bp

696 bp

848 bp

1036 bp

1230 bp

Programmable DNA size distributions, excellent reproducibility, and high dsDNA yields with Bioruptor® NGS

Figure shows different DNA size distributionsofshearedgenomicDNAproducedbyvaryingthedurationofsonicationusingpowersettinghigh(H).ThedifferentcurvesdepictaspecificBioruptor® NGS run, optimized to producespecificmeansizesandsizerangesforNext-Generationsequencing.AllsampleswereanalyzedonBioanalyzer2100usingDNAHighSensitivitychip.

Page 2: DNA Shearing for Bioruptor NGS · DNase free enzymatic digestion since they might generate a biased fragment distribution profile on microfluidics-based platform (e.g. Agilent Bioanalyzer)

PR

OT

OC

OL

Europe - Diagenode s.a. / [email protected] / [email protected] // North America - Diagenode Inc. / [email protected] / [email protected]

www.diagenode.com

Target sizeCycle condition

(On/Off cycle time)Cycle number

150bp 30’’/30’’ 30

200bp 30’’/30’’ 13

400bp* 15’’/90’’ 7-8

1000bp* 5’’/90’’ 7-8

*Forlongerfragments(400upto1000bp),ashortcentrifugationstepafterhalfofthecyclenumberscansignificantlyimprovetheresults.Protocolsforothersizeranges(incl.longerfragmentsupto1300bp)areavailableonrequest.

Programmable DNA size distribution and excellent reproducibility with Bioruptor® NGSPanelAandBshowDNAsizedistributionsof200bpofshearedhumangenomicDNAafter13cycles(30secON/OFF)ofsonication.Bioruptor®NGS0.65mlmicrotubesforDNAShearing(WA-005-0500)wereused.AllsampleswereanalyzedonBioanalyzer2100usingDNAHighSensitivitychip.

PanelA:peakelectropherogramview

PanelB:gelvirtualview

A.

B.

The protocol settings listed above are recommended guidelines and actual results may vary depending on the type and amount of starting material, purity level, concentration and/or sample viscosity. It is highly recommended that a time course response experiment be carried out (e.g. varying the time of “on” and “off” durations as well as the number of cycles) to determine the appropriate treatment for your specific sample. Starting material with a smaller sample volume and a greater concentration than the recommended range may require a different time course to ensure homogenous shearing results.

Page 3: DNA Shearing for Bioruptor NGS · DNase free enzymatic digestion since they might generate a biased fragment distribution profile on microfluidics-based platform (e.g. Agilent Bioanalyzer)

PR

OT

OC

OL

Europe - Diagenode s.a. / [email protected] / [email protected] // North America - Diagenode Inc. / [email protected] / [email protected]

www.diagenode.com

Important comments about DNA shearing

The Diagenode ACT (Adaptative Cavitation Transfer technology) process is highly reproducible.However,attentionmustbepaidtothefollowingtreatmentattributestoensurebestresults:

• Tubes:Atpresent,therecommendedtubevesselsaretheDiagenode’sBioruptor®NGS0.65mlMicrotubesforDNAShearing(CatNo.WA-005-0500).Payattentionnottodamagethecapwhenclosingthetubessincethiscouldaltersonicationresults.

• Sample volume: The recommended volume of the Diagenode’s Bioruptor® NGS 0.65 mlMicrotubesforDNAShearing(CatNo.WA-005-0500)is100µl.Whenusinglowervolumes(e.g.≤50µl),lessreproducibleresultsmaybeobservedduetoanalterationoftheultrasonicwavesdistributioninthesamplefluid;thus,reducingtheefficiencyofsonicationwhichmayresultinbroadersizedistributionorlargerpeaks.

• Sample concentration:DiagenoderecommendsusingDNAconcentrationrangingbetween1and20ng/µl(10ng/µlrecommended).Usinglargerconcentration(e.g.50-100ng/µl)mayresultinbroaderpeaksorvariablepeakdistribution.

• Sample preparation:Sampleviscositymayhaveamajorimpactonsonicationresults.CarefulresuspensionofDNAsampleisstronglyrecommendedbeforesonicationprocessing.Multiplepipettingandgentlevortexingfollowedbyashortcentrifugationtorecoversamplevolumeatthebottomofthetubeisthereforestronglyrecommended.StoringDNAsamplesoniceduring10-15minutesbeforesonicationhasalsobeenshowntoimprovereproducibility.

• DNA quality: DNA quality and quantity must be considered carefully since bad quality andquantityDNAmayhaveseveral impactsonsonicationandnext-gensequencingdownstreamapplications.First,DNAcontamination(e.g.fromsuperfluousnucleicacidssuchasRNA,smallnucleicacidfragments,excessproteins,orothercontaminatingmaterials)mayinterferewithDNA measurement method leading to incorrect DNA quantitation thus. Also contaminatingRNA in genomic DNA preparation might generate a biased fragment distribution profile onmicrofluidics-basedplatform(e.g.AgilentBioanalyzer)oraltersonicationeffiency.

Therefore it is highly recommended to use only high quality DNA when sonicating DNA forNext-Gensequencing librarypreparation.TheDNAsample tobeprocessedshouldbehighlypure,havinganOD260/280ratioofbetween1.8and2.0,andshouldbeas intactaspossible.DNAextractedusingstandardtechniques(e.g.ProteinaseKdigested,doublephenol/chloformextraction,ethanolprecipitated,treatmentwithRNase-DNasefreeenzymaticdigestiontoremovecontaminantRNA)orcommercialspin-columnbasedkitsarerecommended.

• Water temperature:PropagationofultrasoundinaliquidunavoidablyproducesheatthatcanultimatelyalterDNAsample(e.g.bythermaldenaturation).Toensurethebestpreservationofthesample,itisrecommendedtostartthesonicationprocesswithcoldwaterinthewaterbath.Duringsonication,especiallywhendoinglongsonicationruns,thetemperaturemustalsobecontrolled.Thisisobtainedbytheautomatictemperaturecontrol.

Note:ThepermanentinstallationoftheBioruptor®inacoldroomispossible,althoughnotsufficienttoavoidthetemperatureincreaseduetosonication.

• Automatic temperature control:Arecirculatingwatercoolerisusedtoguaranteetheautomatictemperaturecontrolofthewaterbathduringthewholesonicationprocess.Thiswatercooler(CatNo.BioAcc-cool)producesaregularwaterflowwithaconstantwaterlevelinthetank.Anadditionalregulatingvalve(SinglecyclevalveforWaterCooler,Cat.No.VB-100-0001)ensures

Page 4: DNA Shearing for Bioruptor NGS · DNase free enzymatic digestion since they might generate a biased fragment distribution profile on microfluidics-based platform (e.g. Agilent Bioanalyzer)

PR

OT

OC

OL

Europe - Diagenode s.a. / [email protected] / [email protected] // North America - Diagenode Inc. / [email protected] / [email protected]

www.diagenode.com

thatwaterwillonlybereplacedduringtheoffcycletoavoidanyinterferencebetweenthewaterflowandthesonicationprocess.

• Sonication time:Minoradjustmentsincyclenumbermaybemadetooptimizeresultsforvarioussampletypesandconcentrations.Cyclenumberlistedaboveisarecommendedguideline.Actualresultsmayvarydependingontheamountandtypeofstartingmaterial,concentration,viscosityand/orplastictubes.Diagenoderecommendssettingupatimedoseresponseexperimentfordeterminingappropriatecyclenumber.Largerlengthstartingmaterial(e.g.totalgenomicDNA)andhigherconcentrationmayrequirealongerdosetoensureahomogeneousshearingresult.

• Water bath: The sonication water bath is a critical component of the Bioruptor® sonicationsystem.

1. Water purity: Contaminants such as algae and particules may alter the ultrasonic wavespropagation,resultinginbroadersizedistributionorlargerpeaks.Bathwatershouldbepuredistilledwater,changedregularly.

2. Water bath maintenance: The water bath metal surface is fragile and requires a carefulmaintenance.Useonlysoftspongetoremovetraces.Neverusescratchscrubspongesincethiswouldaltertheultrasonicwaveemittersurface.

3. Water type:Distilledwater

Supplementary Data:

Please note that there are three main sources of variation in both peak base-pair size anddistribution:

1) ThephysicalprocessofDNA fragmentationmightnotbeentirely random inAT-orGC-richregions.

2) The analytical process to determine fragment size has inherent variances (for example, gelelectrophoresisandmicrofluidics-basedplatform).Therefore,fragmentdistributionsandpeakvalues,even fromtechnical replicates,maynotappear identical. If theshearedDNAsamplewillberesinorcolumnpurifiedorconcentratedpriortoanalysis,pleaseremembertotakeoutanaliquot foruseascontrolprior to thatstep.Columnpurificationandconcentrationof theshearedDNAwillgenerateabiasedfragmentdistributionprofileduetotheinherentgreaterlossofthesmallerDNAfragments.

3) RNA contamination in genomic DNA preparation should be carefully removed using RNase-DNasefreeenzymaticdigestionsincetheymightgenerateabiasedfragmentdistributionprofileonmicrofluidics-basedplatform(e.g.AgilentBioanalyzer)oraltersonicationeffiency.