Transcript
Page 1: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

23

Chapter 3

SFRA Basics

3.1 Introduction

Sweep Frequency Response Analysis (SFRA) testing has become a valuable tool for

verifying the geometric integrity of transformers. SFRA provides internal diagnostic

information using nonintrusive procedures. The SFRA test method has been proven to

provide accurate and repeatable measurements.

Power Transformers are specified to withstand the mechanical forces arising from

both shipping and subsequent in-service events, such as faults and lightning.

Transportation damage can occur if the clamping and restraints are inadequate; such

damage may lead to core and winding movement. The most severe in-service forces

arise from system faults and are axial and radial in nature. If the forces are excessive,

radial buckling or axial deformation can occur. With a core form design, the principal

forces are radially directed, whereas in a shell-form unit, they are axially directed.

This difference is likely to influence the types of damage found.

Once a transformer is damaged, even if only slightly, its ability to withstand further

short circuits is reduced. Utility personnel need to effectively identify such damage. A

visual inspection is costly and does not always produce the desired results or the

correct conclusion. Since so little of the winding is visible, little damage can be seen,

other than displaced support blocks. Often a complete teardown is required to

identify the problem. An alternative method is to implement field-diagnostic

techniques capable of detecting damage.

Sweep Frequency Response Analysis (SFRA) is a tool that can give an indication of

core or winding movement in transformers. The transformer is considered to be a

complex network of RLC components as shown in Fig. 3.1

Page 2: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

24

Fig. 3.1 Equivalent Geometry of Transformer

Any form of physical damage to the transformer results in the changes of this RLC

network. These changes are what we are looking for and employ frequency response

to highlight these small changes in the RLC network within the transformer. For

different frequencies the RLC network offers different impedance paths. Hence, the

transfer function at each frequency is a measure of the effective impedance of the

RLC network of the transformer. Any geometrical deformation changes the RLC

network, which in turn changes the transfer function at different frequencies and

hence highlights the area of concern [46]. The frequency response of such a network

is unique and, therefore, it can be considered as a fingerprint as shown in Fig. 3.2

Fig. 3.2 Principles of SFRA

Page 3: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

25

3.2 SFRA Theory

The primary objective of SFRA is to determine how the impedance of a test specimen

behaves over a specified range of frequencies. The impedance is a distributive

network of real and reactive electrical components. The components are passive and

can be modeled by resistors, inductors and capacitors. The reactive properties of a

given test specimen depend on, and are sensitive to, changes in frequency. The change

in impedance versus frequency can be dramatic in many cases. This behavior

becomes apparent when we model impedance as a function of frequency. The result is

a transfer function representation of the RLC network in the frequency domain.

When a transformer is subject to SFRA testing, the leads are configured to use four

terminals. These four terminals can be divided into two unique pairs – one pair each

for the input and output. These terminals can be modeled in a two terminal pair or a

two port network configuration as shown in Fig. 3.3.

Fig. 3.3 Two Port Network

Solving for the open-circuit impedance for each lumped element forms the

impedances Z11, Z22, Z12 and Z21. It should be noted that the negative terminals are

short circuited when transformers are tested. The transformer tank is common for both

negative and lower terminals. The transformer tank and lead ground shields must be

connected together to achieve a common mode measurement. This assures that no

external impedance is measured. Applying the connection in this manner helps reduce

the effects of noise. It is important to obtain a zero impedance between the lower or

negative terminals to assure a repeatable measurement.

Page 4: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

26

The transfer function of an RLC network is the ratio of the output and input frequency

responses when the initial conditions of the network are zero. Both magnitude and

phase relationship can be extracted from the transfer function.

The transfer function is represented in the frequency domain and is denoted by

Fourier variable H(jω) where (jω) denotes the presence of a frequency dependent

function, and w=2лf. The Fourier relationship for the input/output transfer function is

given by :

H(jω) = Voutput (jω)

Vinput (jω)

When a transfer function is reduced to its simplest form, it generates a ratio of two

polynomials. The main characteristics, such as half-power and resonance, of a transfer

function occur at the roots of the polynomials.

The goal of SFRA is to measure the impedance model of the test specimen. When we

measure the transfer function H(jω), it does not isolate the true specimen impedance

Z(jω). The true specimen impedance Z(jω) is the RLC network, which is positioned

between the instrument leads, and it does not include any impedance supplied by the

test instrument.

It must be noted that when using the voltage relationship, H(jω) is not always directly

related to Z(jω). For Z(jω) to be directly related to H(jω), a current must be

substituted for the output voltage and then Ohm’s law can be realized. However,

SFRA uses the voltage ratio relationship to determine H(jω). Since SFRA uses a 50

ohm impedance match measuring system, the 50 ohm impedance must be

incorporated into H(jω). The next equation shows the relationship of Z(jω) to H(jω) :

H(jw) = 50___

Z(jw) + 50

It is often useful to plot the magnitude and phase relationship of the transfer function

in logarithmic format. The units of magnitude and phase are in decibles (dB) and

degrees, respectively. Magnitude and phase are represented as follows :

Page 5: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

27

A(dB)=20 log10 (H(jw))

A(θ)= tan-1 (H(jw))

This format takes advantage of the asymptotic symmetry by using a logarithmic scale

for frequency. Plotting the phase relationship with the magnitude data helps determine

whether the system is resistive, inductive or capacitive. It is often useful to compare

resonance in the magnitude plots with the zero crossings in the phase relationship.

[46]

3.3 SFRA History

1960 Low Voltage Impulse Method (LVI) First proposed by W. Lech & L.

Tyminski in Poland for detecting transformer winding deformation.

1966 Result Published: W. Lech & L. Tyminski “Detecting transformer winding

damage – the low voltage impulse method” , Electric Review, no. 18, ERA, UK.

The method was used by Dr. Alexandr Dobishevsky in former USSR and within

Bonneville Power Administration, United States (Eldon Rogers)

1976 “Frequency Domain Analysis of Responses from L.V.I. Testing of Power

Transformers” Presented by A.G. Richenbacher at the 43rd Doble Int’l Client

Conference.

1980 In the 1980’s the Central Electricity Generating Board (CEGB) in the UK took

up the measurement technique and applied it transmission transformers. The French

also began to pursue measurements at the same time.

1990 On the breakup of the CEGB in the early 1990’s work in FRA was taken by

National Grid in the UK and resulted in several papers at Doble Client Conferences.

The technique has been spread further through Euro Doble Conferences and client

meetings and several utilities took up the technique.[16]

Many early practitioners tried impulse systems, and have continued to try them up to

the 2003. Though appealing in terms of speed, they have never been able to match the

range, resolution or repeatability of sweep methods and continue to reject such

methods.

As the basic technique developed by early users required laboratory based equipment

such as HP network analyzers, which were robust, but not field hardened and required

specialist operators. Upon a successful program of product development and field

Page 6: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

28

trials, Doble stepped in to provide field engineers and staff with a reliable and robust

tool for transformer analysis – the M5100. This outperforms the HP in terms of

measurement characteristics and field usability. [9]

3.4 SFRA Today

SFRA is today an established technology to investigate the mechanical and electrical

integrity of a transformer.

Before the 1990’s no designated SFRA equipment for transformer analysis existed.

Since then several specialized products have been introduced.

The value that SFRA provides to its users is quite easily established, by research of

published material and customer experience.

The international community, such as CIGRE, IEC and the IEEE has widely accepted

its use and is now providing guide lines.

CIGRE : CIGRE working group A2-26 – Mechanical Condition Assessment of

Transformer Windings created in 2004. The CIGRE WG A2-26 main objective is

to develop a guide on the Mechanical Condition Assessment of Transformer

Windings using the Frequency Response Analysis (FRA) method. The WG has

delivered a final report end 2007 for a publication before the group session in

August 2008. The WG is disbanded. UK will check with IEC TC 14 for a

possibility to have a part of this document publish as an IEC document. [11]

A new work item proposal 14/597/NP has been issued by IEC TC 14 on December

2008 based on the work done within WG A2-26 “Mechanical Condition Assessment

of Transformer Windings” and finalized with the publication of the TB 342 in April

2008. [12]

IEC : IEC published the FRA Standard, 60076-18, “ Power Transformers –

Part 18 : Measurement of Frequency Response” in July, 2012.

Page 7: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

29

The Scope of the IEC 60076 series covers the measurement technique and measuring

equipment to be used when a frequency response measurement is required either on-

site or in the factory either when the test object is new or at a later stage.

Interpretation of the result is not part of the normative text but some guidance is given

in Annex B. This standard is applicable to power transformers, reactors, phase

shifting transformers and similar equipment. [28]

IEEE : IEEE C57.149 Working Group published the “Guide for the Application

and Interpretation of Frequency Response Analysis for Oil immersed

Transformers” in March 2013.

IEEE C57.149 Working Group focused on the 3 failure modes, namely

• Radial Deformation

• Axial Deformation and

• Bulk Movement from Transportation.

3.5 Purpose of FRA Measurements

The main interest of FRA measurements on transformers is to detect winding

deformations that may result from the very large electromagnetic forces arising from

over-currents during through faults, tap-changer faults, faulty synchronization, etc.

Winding deformation eventually results in a transformer failure by damaging the

inter-turn insulation, resulting eventually in shorted turns, which means the immediate

end of service life. Transformers are expected to survive a number of short circuits

without failure but, once any significant winding deformation is produced, the

likelihood of surviving further short circuits is greatly reduced because of locally

increased electromagnetic stresses. Furthermore, any reduction in winding clamping

due to insulation shrinkage caused by ageing will also increase the likelihood of

failure by reducing the mechanical strength of the winding assemblies.

In addition to diagnosing failures after a short-circuit event, there is increasing interest

in detecting winding deformation damage prior to eventual failure during planned

outages, i.e. mechanical-condition assessments to assess the expected reliability of

Page 8: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

30

transformers in terms of any suspected increased susceptibility to failure under further

short circuits.

There is also an interest in using FRA measurements to detect any other problems that

result in changes to the inductance or capacitance distribution in transformers, e.g.

core faults or faulty grounding of cores or screens.

Another application for FRA measurements is to check the mechanical integrity of a

transformer after transportation, which usually means providing a reliable means of

confirming that the core and winding assembly have not suffered any mechanical

damage despite sustaining jolts during transportation. Note that for this application it

may be necessary to have reference results without oil and bushings, if that is how the

transformer is transported. Since transportation shocks are more likely to cause

damage to the core structure than to the windings, there is a slightly different focus for

this application. Because FRA measurements can provide information about the

consistency of geometric structures of windings and core, such tests are increasingly

being used as quality assurance checks. [15]

3.6 Definitions

Frequency Response Analysis (FRA)

Any measurements of the frequency dependency (to high frequencies, e.g. MHz) of

the electrical responses (transfer functions) of transformer windings to applied signals

which are made with the primary intention of detecting winding deformation through

the effects of resulting changes to capacitance or inductance distributions.

Sweep Frequency Method

A frequency response measured directly by injecting a signal of a variable frequency

at one terminal and measuring the response at another.

Impulse Voltage Method

A frequency response measured indirectly by injecting an impulse signal of a

particular shape at one terminal and measuring the response at another, and then

transforming the time domain measurements into frequency domain results.

Page 9: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

31

FRA Amplitude

The magnitude of the response relative to that of the injected signal, usually expressed

in dB calculated as 20*log 10(Vresponse/Vinjection).

FRA Phase Angle

The phase angle shift of the response relative to that of the injected signal.

Resonance Frequency

The frequencies corresponding to any local maxima or minima in the measured

amplitude response. [15]

3.7 Development and Variations in FRA Practices

It is important to realize that a great variety of FRA measurement techniques are

currently being used, not all of which have produced good results. Most of the

variations in the FRA technique can be traced to how the technique developed from

LVI. Differences in FRA practices arise from two main aspects:

How the measurement is made

Which measurement is made

The main variation in how FRA measurements are made concerns whether a sweep

frequency method (referred to as ‘SFRA’) or an impulse method (referred to as

‘IFRA’) is used. The first IFRA techniques used an impulse method with the same

double exponential type of impulse signal as used by LVI, with appropriate rise and

fall times to include components of the range of frequencies of interest. The impulse

is applied to one terminal and the form of the applied and the transmitted signal at

another terminal are recorded by a dual channel digital data acquisition system. One

key development from time domain LVI is that the two measured impulses are then

transformed into the frequency domain using the Fast Fourier Transform algorithm,

and then the calculated amplitudes of frequency components of the transmitted signal

are divided by the corresponding amplitudes of the applied signal to derive the

frequency response indirectly. This frequency response has the advantage over the

LVI time response that it is independent of the shape of the applied impulse, so that

the result is more closely related to the test object and less to the test set-up, thereby

simplifying interpretation and improving repeatability. The impulse method of

Page 10: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

32

performing FRA measurements being a development of traditional high-voltage

impulse testing, some transformer manufacturers use their modern digital impulse

testing recorders to perform FRA measurements but recently purpose-built

transformer test instruments have become available to perform IFRA measurements.

Since the objective of FRA measurements is to obtain the frequency response of

windings, an alternative technique was proposed [28] to measure this directly using a

sweep frequency technique. A sine wave signal is applied to one terminal and the

amplitude and phase of the transmitted signal at another terminal are measured

relative to the applied signal for various frequencies over the frequency range of

interest. Early practitioners had to use general-purpose laboratory Network/Spectrum

Analyzer instruments, but more recently purpose-built transformer test instruments

have become available to perform SFRA measurements.

In principle, everything else being equal, the sweep frequency and impulse techniques

should be capable of producing the same result, and this has been demonstrated on

several occasions. For sweep frequency measurements, the accuracy depends on the

ability of the equipment to perform over the frequency range of interest, and to reject

noise at frequencies away from the measurement frequency. In order to obtain an

accurate derivation of the frequency response using the impulse technique, the

sampling frequency and record length of the digitizing equipment must be adequate to

faithfully record all frequency components of interest in both input and output

impulses, which must both return to zero at the end of the sampling period for the

FFT algorithm to be valid (in case no window function is applied), and the applied

impulse amplitude must be large enough to ensure that all noise components in the

output frequency distribution are insignificant. The introduction of Spectral Density

Estimates to the impulse technique helped to overcome the influence of noise in the

output signal and the result of the FRA measurement.

The other main way that variations in FRA results can be introduced by how the

measurement is made concerns practices involving test leads. A three-lead system

(separate leads for applying and measuring the signal at the input terminal) as shown

in Fig. 3.4 is recommended to avoid including the input lead in the measurement.

Page 11: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

33

Fig. 3.4 Color codes of leads at test set

When making high-frequency measurements, it is good practice to use coaxial test

leads with a good high-frequency bandwidth, and to ensure that the test leads are

terminated in their characteristic impedance, usually 50 ohms, to avoid reflections.

Good practice for grounding the shields of the coaxial cables is of primary importance

to achieve good repeatability.

The most basic and important variation in FRA results is introduced by which type of

measurement is made. Most SFRA users perform an end-to-end measurement in

which the input signal is applied to one end of every winding and the transmitted

signal at the other end is measured, as for a simple resistance measurement. For some

impulse users, following traditional impulse measurement practice, it is more usual to

inject a voltage at one terminal (usually an HV terminal) and measure the transferred

voltages to other windings, or currents in the injected winding (usually at the HV

neutral) to derive self or transferred impedances (or admittances). Variation can also

be introduced by different values of measuring impedances (50Ω/10Ω/1MΩ etc.)

and/or by the way other untested terminals are terminated. Some users prefer a

practice of grounding untested windings while others prefer to leave all other

terminals floating. Not surprisingly, these different measurements are not necessarily

equally effective in detecting mechanical displacement. Some recent work has been

done to compare the relative sensitivities of different connection techniques.

Page 12: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

34

Lately, a technique has been demonstrated where a complete transformer fingerprint

is measured such that subsequently any type of FRA curve can be calculated on

demand. A device connects to all the transformer terminals at the same time and

automatically measures all the linear properties of the transformer, i.e. the full

admittance matrix, without requiring any reconnection. In addition, this technique

allows the automated generation of high frequency terminal models of transformers

for network simulation purposes.

In view of the wide variety of FRA practices in use, there would obviously be benefit

in carefully examining these with a view to standardizing those that have been shown

to be most effective, while allowing variety where this does not impact on

performance. [15]

3.8 Measurements on Different Winding Types

Most people immediately think of winding measurements as being only associated

with the high-voltage and the low-voltage windings. When considering SFRA

measurements, winding measurements realistically consist of five categories and not

just two. The winding categories are high-voltage, low-voltage, inter, series, and

common.

Short circuit measurements made on one winding while short circuiting another

winding are a variation on inter winding measurements.

It should be noted that inter-winding measurement is not a true winding measurement,

but rather the transfer impedance between two windings. The series and common

winding measurements describe the SFRA application as it is applied to auto

transformers. Regardless, certain expectations can be made for each.

These measurement types produce some predictable characteristics and properties.

Understanding these properties will minimize testing error and may help identify

problems. The following expectations exist for each of the following categories. [39]

Page 13: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

35

3.8.1 High-Voltage Winding

High-voltage winding measurements have greatest attenuation as compared to low

voltage and tertiary windings. Most traces start between –30 dB and –50 dB and are

initially inductive. High-voltage windings are much larger in overall size, which

contributes to greater complexity in its distributive network. High-voltage winding

measurements generally produce steeper resonances and more of them as compared to

its low-voltage counterpart. Fig. 3.5 illustrates these features.

Fig. 3.5 High Voltage winding

The traces shown in Fig. 3.5 are from different test specimens. Both traces are from

230 kV core-forms transformers, however one trace is from a delta connected

configuration and the other is from a wye connected configuration. [39]

3.8.2 Low-Voltage Winding

Low-voltage winding measurements have least attenuation as compared to the other

categories. Most traces start between –5 dB and –15 dB and are also initially

inductive This characteristic is due to the low impedance property of the high current

side of the transformer. The first peak after the core resonance generally approaches –

5 dB to 0 dB and is concave and smooth. As compared to the high-voltage winding

response, the low voltage winding fewer fluctuations and is slight smoother. Fig. 3.6

illustrates these features. Again, both traces in this figure are from different

transformers. [39]

Page 14: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

36

Fig. 3.6 Low-Voltage Winding

3.8.3 Inter-Winding

Inter-winding measurements always start with high attenuation, between –60 dB and

–90 dB, and are capacitive. If electrostatic interference is present, it will show up at

60 Hz and at the associated harmonics of 60 Hz during this measurement. Fig. 3.7

illustrates these features. These traces are very common; most inter-winding traces

adhere to one of the basic shapes shown below.

Fig. 3.7 Inter-Winding

Fig. 3.8 presents a high-voltage winding trace, a low-voltage winding trace, and an

inter-winding trace together from a common test specimen. This illustrates their

general relationship. It can be seen that the low-voltage winding has consistently

lower attenuation then the high-voltage winding. Also, low-voltage winding is much

Page 15: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

37

smoother at higher frequencies. This example was taken from a 10 MVA auxiliary

transformer. [39]

Fig. 3.8 Trace Relationship

3.8.4 Series and Common Winding

The series and common winding measurements are grouped together because of their

similarities. These measurements are associated with auto transformer. The naturally

low turns ratio of an auto transformer causes the series and common measurements to

be similar. However, if an LTC is present on either winding, the similarities will be

somewhat affected by the tap windings. Fig. 3.9 illustrate these features, and were

obtained from a General Electric 440MVA 345 kV auto-transformer. Electrostatic

interference was present during testing and is seen at 60 Hz. [39]

Fig. 3.9 Series and Common Winding

Page 16: 11 Chapter 3 SFRA Basics - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/50756/12/12_chapter3.pdf · Chapter 3 SFRA Basics ... Sweep Frequency Response Analysis (SFRA)

Chapter 3 SFRA Basics

38

3.9 Conclusion

Sweep Frequency Response Analysis (SFRA) is a tool that can give an indication of

core or winding movement in transformers. This is done by performing a

measurement, a simple one, looking at how well a transformer winding transmits a

low voltage signal that varies in frequency. Just how well a transformer does this is

related to its impedance, the capacitive and inductive elements of which are intimately

related to the physical construction of the transformer. Changes in frequency response

as measured by SFRA techniques may indicate a physical change inside the

transformer, the cause of which then needs to be identified and investigated.


Top Related