12706A
Test Fixture
OPERATION MANUAL
Manual Number FOE-00000048A00
C-1*
12706A Test Fixture Operation Manual
Table of Contents
1. Introduction .................................................................................................. 1-1
1.1 Product Overview ......................................................................................................................... 1-1
1.2 Before Using the Product ............................................................................................................. 1-2
1.2.1 Appearance Check .............................................................................................................. 1-2
1.2.2 Operating Environment ...................................................................................................... 1-2
1.2.3 Parts with Limited Service Life .......................................................................................... 1-3
2. Part Descriptions .......................................................................................... 2-1
3. Methods of Use ............................................................................................ 3-1
3.1 Attaching a Sample ....................................................................................................................... 3-1
3.2 Connecting to the Measuring Instruments .................................................................................... 3-2
3.2.1 Preparation before Connection ........................................................................................... 3-2
3.2.2 Connecting with the 8340A ................................................................................................ 3-3
3.2.3 Connecting with the 8240 ................................................................................................... 3-5
3.2.4 Connecting with the 8252 ................................................................................................... 3-7
3.3 Measurement ................................................................................................................................ 3-8 3.4 Sample Program ........................................................................................................................... 3-9 3.5 Replacing the Test Socket ........................................................................................................... 3-12
4. Specifications ............................................................................................... 4-1
Appendix ......................................................................................................... A1-1
A.1 Internal Connection ................................................................................................................... A1-1
A.2 12706A External View .............................................................................................................. A1-2
F-1*
12706A Test Fixture Operation Manual
List of Illustrations
No. Title Page
2-1 12706A Part Descriptions ............................................................................ 2-2 3-1 Attaching a Sample ..................................................................................... 3-1 3-2 Connecting with the 8340A ......................................................................... 3-3 3-3 Connecting with the 8240 and the 6161 ...................................................... 3-4 3-4 Replacing the Test Socket ............................................................................ 3-11
T-1*
12706A Test Fixture Operation Manual
List of Tables
No. Title Page
1-1 Parts with Limited Service Life ...................................................................... 1-3 3-1 Connectable Measuring Instruments and Necessary Cables/Accessories........ 3-2 3-2 Test Socket Standard ....................................................................................... 3-10
1-1
12706A Test Fixture Operation Manual
1.1 Product Overview
1. Introduction
1.1 Product Overview The 12706A test fixture is a sample box that connects to a measuring instrument such as 8340A, 8240 or 8252 to perform insulation resistance measurement and VSIM (voltage source current measurement) of capacitors and other electrical components.
Features:
(1) A shielded and insulated structure is adopted, allowing stable measurement of low cur-rent and high resistance.
(2) Voltage is applied by closing the cover and is shut off by opening the cover. (3) A contact signal according to opening and closing of the cover controls the measuring
instrument.
1-2
12706A Text Fixture Operation Manual
1.2 Before Using the Product
1.2 Before Using the Product 1.2.1 Appearance Check
When receiving the 12706A, inspect it for any damage during transportation. If there is any damage in appearance, please contact ADC Corporation or your sales represent-ative. Our address and phone number are listed at the end of this manual.
1.2.2 Operating Environment
(1) Use the 12706A in an area free from dust, corrosive gas and vibrations and away from di-rect sunlight.
(2) The ambient temperature should be -20°C to + 80°C, the relative humidity should be 80% or less.
(3) Storage
The storage temperature range is from -25°C to +80°C. When not in use for a long time, cover the 12706A with a plastic sheet or put in a card-board box, and store it in a low humidity area away from direct sunlight.
(4) Transportation
When transporting the 12706A, pack with the packaging materials used at delivery. If the packaging materials have been lost, pack the 12706A by using the following procedure:
a) Wrap the 12706A with a plastic sheet.
b) Put it in a cardboard box (with thickness of 5 mm or more) and protect it with cu-shioning materials.
c) Fix the outside with packing tape.
1-3
12706A Test Fixture Operation Manual
1.2 Before Using the Product
1.2.3 Parts with Limited Service Life
The 12706A use the following parts with limited service life. Check these parts by reference to the durability in the table. If any abnormality such as a con-tact failure is found, the parts need replacing. Please contact ADC Corporation or your sales representative. For how to replace the test socket, refer to Chapter 3.5 "Test Socket Replacement."
Table 1-1 Parts with Limited Service Life
Part name Durability (times) ADC stock No. Manufacturer Standard name Test socket Mechanical:
100,000 or more Electrical: 50,000 or more
JCB-AB004EX01-1 Textool (3M) 201-0080-00-0605
Micro switch Mechanical: 100,000 or more Electrical: 50,000 or more
KSP-000698 OMRON SS-5GL2
VS contact switch Mechanical: 250,000 or more Electrical: 50,000 or more
MBN-68725A ADC -
2. P
1
2
Part Desc
1. INPUT: t
ConnectsTo be con
2. V SOURC
HI: ConnLO: Conn
criptions
terminal for m
to the MEAnnected to th
CE HI/LO: t
nects to the Vnects to the c
s
Figure 2-1
measuremen
ASURE of thehe input termi
terminal for v
V SOURCE ochassis.
12706A Pa
t
e test socket inal of a mea
voltage appli
of the test soc
12706A Te
art Descriptio
[4]. asuring instru
ication
cket [4].
est Fixture
2
ons
ument.
Operation M
2. Part Des
2-1
Manual
scriptions
2-2
12706A Text Fixture Operation Manual
2. Part Descriptions
3. LID SIGNAL: contact output by opening and closing the cover
When the cover is closed, the terminals are short-circuited. When it is opened, the termin-als are open-circuited. Used to turn the applied voltage ON/OFF and to start measurement.
4. Test socket: socket to mount a sample
Voltage is applied to the V SOURCE and measurement is performed on the MEASURE. Sample conditions: Axial lead part
Width: 23 mm or less, diameter: 14 mm or less Lead diameter: 1.14mm or less
5. LID SIGNAL switch: switch of the LID SIGNAL [3]
6. Applied voltage switch: switch that shuts off the voltage applied to the V SOURCE ter-minal of the test socket when the cover is opened.
The applied voltage switch has mechanical durability of 250,000 times operation. How-ever, after it reaches the end of its service life, it will be always OFF regardless of open-ing and closing of the cover. If a higher resistance or lower current value than an ex-pected value is obtained in kwon measurement, the end of service life of the switch may be reached.
a. Flash symbol Indicates high voltage danger. Printed for a terminal to which voltage higher than 1 kV is input.
b. Warning symbol Printed at a location where the operation manual needs to be referred to for protecting the 12706A from damage.
CAUTION
3. M
3.1 AT
P
1
2
Methods
Attaching This section d
The test soURE term
Procedure:
1. Check tha) Axib) Widc) Lea
2. Attach a
of Use
a Sampledescribes the
ocket has thinal for mea
hat a sample tial lead part dth: 23 mm oad diameter:
sample as sh
e e procedure o
e V SOURCasurement. B
to be measur
or less, diam1.14 mm or
hown below:
Figur
of attaching a
CE terminal fBe careful wh
red meets the
meter: 14 mm less
re 3-1 Atta
CAUTION
12706A Te
a sample to b
for voltage hen attachin
e following c
or less
aching a Sam
est Fixture
3.1
be measured
application ng a polar sa
conditions:
mple
Operation M
Attaching a
to the test so
and the MEample.
3-1
Manual
a Sample
ocket.
EAS-
3-2
12706A Text Fixture Operation Manual
3.2 Connecting to the Measuring Instruments
3.2 Connecting to the Measuring Instruments This section describes the procedure of connecting the 12706A to measuring instruments. For the details of each instrument, refer to its operation manual.
3.2.1 Preparation before Connection
The following table shows connectable instruments and necessary cables.
Table 3-1 Connectable Measuring Instruments and Necessary Cables/Accessories
Model Terminal
Measuring instrument Voltage source
8340A 8240 8252 6161 6144
6243
For INPUT A01009
or A01239
A01009 A01009 - -
For V SOURCE A01240 - A01044*
+ A08531*
A01240 A01044*
+ A08531*
For LID SIGNAL A01036 - A01036 - -
A01009: TRIAX-TRIAX cable
A01239: High-voltage TRIAX-TRIAX cable
A01240: Banana-banana cable with cover
A01036: BNC-BNC cable
*: Supplied accessory
3-3
12706A Test Fixture Operation Manual
3.2 Connecting to the Measuring Instruments
3.2.2 Connecting with the 8340A
This section describes the procedure of connecting the 12706A with the 8340A. See Figure 3-2.
Procedure:
(1) Attach the short bar to the input section of the 8340A. (2) Connect the INPUT of the 8340A to the INPUT of the 12706A by using the TRI-
AX-TRIAX cable (A01009 or A01239). (3) Connect the V SOURCE HI of the 8340A to the V SOURCE HI of the 12706A by using
the banana-banana cable (A01240). The V SOURCE LO does not need connecting.
(4) Connect the LID SIGNAL of the 8340A to the LID SIGNAL of the 12706A by using the BNC-BNC cable (A01036-1500). (The V SOURCE of he 8340A is turned ON/OFF by closing/opening the cover of the 12706A.)
The 8340A outputs applied voltage of maximum 1000 V. Perform connection or measurement with great care. Fix the short bar of the 8340A between the GND, GUARD and LO terminals.
CAUTION
3-4
12706A
3.2 Co
A Text Fixtu
onnecting to t
ure Operatio
the Measurin
F
on Manual
ng Instrumen
Figure 3-2
nts
Connecting with the 83440A
3-5
12706A Test Fixture Operation Manual
3.2 Connecting to the Measuring Instruments
3.2.3 Connecting with the 8240
This section describes the procedure of connecting the 12706A with the 8240. See Figure 3-3.
Procedure:
(1) Connect the INPUT of the 8240 to the INPUT of the 12706A by using the TRIAX-TRIAX cable (A01009).
(2) Connect a voltage source such as the 6161 or 6144 to the V SOURCE (binding post) of the 12706A by using the banana-banana cable (A01240).
1. The output voltage of the voltage source must not exceed the maximum input voltage of the 8240. For the maximum input voltage of the 8240, refer to the 8240 Operation Ma-nual.
2. Be sure to turn ON the driving guard switch of the 8240 3. The maximum output voltage of the 6161 is 1100 V. Perform connection or
measurement with great care
CAUTION
3-6
12706A
3.2 Co
A Text Fixtu
onnecting to t
ure Operatio
the Measurin
Figure 3
on Manual
ng Instrumen
3-3 Connec
nts
cting with thhe 8240 and tthe 6161
3-7
12706A Test Fixture Operation Manual
3.2 Connecting to the Measuring Instruments
3.2.4 Connecting with the 8252
This section describes the procedure of connecting the 12706A with the 8252.
Procedure:
(1) Connect the INPUT of the 8252 to the INPUT of the 12706A by using the TRIAX-TRIAX cable (A01009).
(2) Connect the HI and LO of the 8252 (short circuit between LO and GND) to the V SOURCE of the 12706A by using the output cable (A01044) and the banana adapter (A08531).
(3) Connect the INTERLOCK of the 8252 to the LID SIGNAL of the 12706A by using the BNC-BNC cable (A01036-1500). (The voltage application of the 8252 can be controlled by closing/opening the cover of the 12706A.)
3-8
12706A Text Fixture Operation Manual
3.3 Measurement
3.3 Measurement Procedure:
(1) Attach a sample, referring to Section 3.1.
(2) Connect the 12706A with a measuring instrument, referring to Section 3.2.
(3) For the 8340A or 8252, set to voltage source current measurement (VSIM) or voltage source resistance measurement (VSRM). For the 8240, set to DC current measurement.
(4) Closing the cover of the 12706A will turn on the applied voltage switch, allowing mea-surement.
1. 12706A adopts a safety design. However, be careful when using high voltage as applied voltage.
2. Do not give vibrations to the 12706A or connected measuring instrument and cables during low current or high resistance measurement, otherwise measured values will vary.
CAUTION
3.4 STP8
1
ample PrThe followinPC9801 serie8340A, and l
1. Sample p
rogram ng shows a pes as personaeak current i
program list
program exaal computer is measured a
ample for min Figure 3
as i-t charact
12706A T
easurement -2. Voltage iteristic.
Test Fixture
3
by GPIB cois applied to
Operation M
3.4 Sample
ontrol using o the capacito
3-9
Manual
Program
the NEC or by the
3-10
12706A Text Fixture Operation Manual
3.4 Sample Program
2. Sample program descriptions
Line No. Description 120
130 140 150 160
170 180 190 200 210 220 230 240 250 260 270 280 290 300 310
Address 1 of the 8340A. (A variable does not start with a number. So, "R8340A" is used here.) Interface clear, remote enabled. Initialize the 8340A. Initialize the 8340A parameters. Set the 8340A parameters. "MO1"……………………….Sampling HOLD "PVS25"……………………..Source voltage 25 V ' Set the 8340A to discharge status. Make discharge time. Set the 8340A to voltage source and charge status. Make charge time. Set the 8340A to measurement status. ' Measure 30 times.
Give a measurement trigger to the 8340A. Capture data and print. Make measurement interval time.
To line 230. ' Set the 8340A to discharge status. Program end
3
3. Sample pprogram resu
ult
12706A TTest Fixture
3
Operation M
3.4 Sample
3-11
Manual
Program
3-12
12706A Text Fixture Operation Manual
3.5 Replacing the Test Socket
3.5 Replacing the Test Socket The test socket contact has durability of connecting/disconnecting 50,000 or more times. If contact failure due to abrasion occurs, the socket needs replacing. Please contact ADC Corpo-ration or your sales representative.
Replace the test socket in accordance with the following procedure:
Table 3-2 Test Socket Standard
Part name Stock No. Standard name Test socket JCE-AB004EX01-1 201-0080-000605
Textool (3M) Note: Quote the stock number of the part when ordering.
Procedure:
(1) Remove the six screws, and remove the cover. See Figure 3-4 (a).
(2) Remove the cable soldered on the test socket terminal. See Figure 3-4 (b).
(3) Remove the screws and the nuts, and replace the socket. (The screws and the nuts will be used again.) See Figure 3-4 (c).
(4) Attach the socket by the reversing above procedure.
F
Pan
Figure 3-4
Cabl
Panel
nel
Flat he
Replacing th
S
le
Test s
ead screw M2
12706A T
he Test Socke
oldering iron
socket
Nut
Spring wa
Test socke
2 x 12
Test Fixture
3.5 Repl
et
n
asher
et
Operation M
lacing the Te
3-13
Manual
est Socket
4-1
12706A Test Fixture Operation Manual
4. Specifications
4. Specifications
Dielectric strength INPUT signal line – V SOURCE HI 1.5 kV DC for a minute INPUT signal line – Chassis 1.5 kV DC for a minute V SOURCE HI – Chassis 1.5 kV DC for a minute
Insulation resistance Temperature: 0 to 40°C, relative humidity: 70% or less INPUT signal line – V SOURCE HI 1 × 1015 Ω or higher INPUT signal line – Chassis 1 × 1015 Ω or higher V SOURCE HI – Chassis 1 × 109 Ω or higher
Connector INPUT: triaxial connector V SOURCE HI: biding post (red) V SOURCE LO: biding post (black) LID SIGNAL: BNC connector
Socket Socket for an axial lead type part Attachable sample: 23 mm wide, 14 mm dia. 1.14 mm lead dia.
Operating environment Temperature range: -20°C to +80°C Relative humidity: 80% or less
Storage temperature -25°C to +80°C Dimensions Approx. 150 (W) x 80 (H) x 150 (D) mm Mass 1.5 kg
A1.1
Internall Connect
tion
12706A TTest Fixture Operation M
A
A1-1
Manual
Appendix
A1-2*
12706A
Appendix
A1.2
A Text Fixtu
x
12706A E
ure Operatio
External
on Manual
View