AAssessment of an Earlobe Arterialized ssessment of an Earlobe Arterialized Blood Collector for use in MicrogravityBlood Collector for use in Microgravity
Thais Russomano* Thais Russomano* Marlise A dos Santos* Marlise A dos Santos*
João Castro* João Castro* John Whittle** John Whittle** Simon Evetts** Simon Evetts** John Ernsting** John Ernsting**
*Microgravity Laboratory/IPCT-PUCRS, Brazil
**Aerospace Medicine Group, King´s College London, UK
![Page 2: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/2.jpg)
There is currently no method of directly measuring arterial blood gas tensions in space.
Background
![Page 3: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/3.jpg)
There is currently no method of directly measuring arterial blood gas tensions in space.
An alternative to direct arterial measurement is earlobe arterialized blood sampling, an accurate technique for measuring blood gas tensions, which has been in use in clinical medicine and physiology for more than 30 years (Lilienthal JL & Riley RL, 1944).
Background
![Page 4: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/4.jpg)
There is currently no method of directly measuring arterial blood gas tensions in space.
An alternative to direct arterial measurement is earlobe arterialized blood sampling, an accurate technique for measuring blood gas tensions, which has been in use in clinical medicine and physiology for more than 30 years (Lilienthal JL & Riley RL, 1944).
This technique has not yet been examined in the microgravity environment due to the risk of environmental contamination with blood products.
Background
![Page 5: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/5.jpg)
The aim of the project is to develop a device and associated procedures that enable an acceptable estimation of arterial blood gas tensions and pH to be conducted accurately and safely in microgravity.
Aim
![Page 6: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/6.jpg)
Introduction
Earlobe Blood Sampling – The Technique
![Page 7: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/7.jpg)
Introduction
Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.
![Page 8: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/8.jpg)
Introduction
Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.
The skin is then cleaned with an alcohol swab and a small incision is made in the earlobe.
![Page 9: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/9.jpg)
Introduction
Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.
The skin is then cleaned with an alcohol swab and a small incision is made in the earlobe.
Blood is collected anaerobically via capillary tubes.
![Page 10: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/10.jpg)
Introduction
Earlobe Blood Sampling – The Technique The earlobe is rendered hyperemic by the application of a rubefacient cream containing 1% methyl nicotinate.
The skin is then cleaned with an alcohol swab and a small incision is made in the earlobe.
Blood is collected anaerobically via capillary tubes.
Blood is then analyzed using a blood gas analyzer.
![Page 11: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/11.jpg)
Introduction
Earlobe massage
Earlobe arterialized blood collection
![Page 12: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/12.jpg)
Characteristic Radial Artery Sample
Hyperemic Earlobe Sample
Discomfort Painful Pain Free
Potential Complications
Hematoma Hemorrhage
Infection (systemic) Wrist pain
Arterial Spasm
Hemorrhage Infection (cutaneous)
Ease of Use Requires trained medical personnel
Performed by non-medical personnel.
Potential Usage Hospital based research
Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.
Radial Artery / Arterialised Earlobe Collection
![Page 13: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/13.jpg)
Characteristic Radial Artery Sample
Hyperemic Earlobe Sample
Discomfort Painful Pain Free
Potential Complications
Hematoma Hemorrhage
Infection (systemic) Wrist pain
Arterial Spasm
Hemorrhage Infection (cutaneous)
Ease of Use Requires trained medical personnel
Performed by non-medical personnel.
Potential Usage Hospital based research
Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.
![Page 14: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/14.jpg)
Characteristic Radial Artery Sample
Hyperemic Earlobe Sample
Discomfort Painful Pain Free
Potential Complications
Hematoma Hemorrhage
Infection (systemic) Wrist pain
Arterial Spasm
Hemorrhage Infection (cutaneous)
Ease of Use Requires trained medical personnel
Performed by non-medical personnel.
Potential Usage Hospital based research
Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.
![Page 15: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/15.jpg)
Characteristic Radial Artery Sample
Hyperemic Earlobe Sample
Discomfort Painful Pain Free
Potential Complications
Hematoma Hemorrhage
Infection (systemic) Wrist pain
Arterial Spasm
Hemorrhage Infection (cutaneous)
Ease of Use Requires trained medical personnel
Performed by non-medical personnel.
Potential Usage Hospital based research
Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.
![Page 16: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/16.jpg)
Characteristic Radial Artery Sample
Hyperemic Earlobe Sample
Discomfort Painful Pain Free
Potential Complications
Hematoma Hemorrhage
Infection (systemic) Wrist pain
Arterial Spasm
Hemorrhage Infection (cutaneous)
Ease of Use Requires trained medical personnel
Performed by non-medical personnel.
Potential Usage Hospital based research
Hospital, clinic, rural center and university research.Aero medical and transport use. ISS and other space missions.
![Page 17: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/17.jpg)
Development and Evaluation of a Earlobe Arterialized Blood Collector
![Page 18: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/18.jpg)
Prototype of the Earlobe Arterialized Blood (EAB) Collector
![Page 19: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/19.jpg)
Current Prototype of the EAB Collector
Blade module
Capillary tube module
Capillary tube
Body (module housing)
Ophthalmic blade
![Page 20: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/20.jpg)
A validation study of the EAB Collector was conducted in which simultaneous samples of arterial and arterialized blood were taken from the radial artery and the earlobe.
Method
![Page 21: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/21.jpg)
A validation study of the EAB Collector was conducted in which simultaneous samples of arterial and arterialized blood were taken from the radial artery and the earlobe.
Six healthy subjects breathed a gas mixture of 12.8% O2 in N2 (equivalent to breathing air at 12,000 feet) during 15 min of head-down tilt.
Method
![Page 22: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/22.jpg)
A validation study of the EAB Collector was conducted in which simultaneous samples of arterial and arterialized blood were taken from the radial artery and the earlobe.
Six healthy subjects breathed a gas mixture of 12.8% O2 in N2 (equivalent to breathing air at 12,000 feet) during 15 min of head-down tilt.
The blood samples were analyzed immediately.
Method
![Page 23: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/23.jpg)
MethodHead Down Tilt to partially replicate the
effects of microgravity
![Page 24: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/24.jpg)
Results
Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples
Radial arteryMean SD (range)
Arterialized EarlobeMean SD (range)
pH(pH Unit)
7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)
PO2
(mmHg)
42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)
PCO2
(mmHg)
34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)
SaO2
(%)
79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)
![Page 25: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/25.jpg)
Results
Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples
Radial arteryMean SD (range)
Arterialized EarlobeMean SD (range)
pH(pH Unit)
7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)
PO2
(mmHg)
42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)
PCO2
(mmHg)
34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)
SaO2
(%)
79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)
![Page 26: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/26.jpg)
Results
Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples
Radial arteryMean SD (range)
Arterialized EarlobeMean SD (range)
pH(pH Unit)
7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)
PO2
(mmHg)
42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)
PCO2
(mmHg)
34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)
SaO2
(%)
79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)
![Page 27: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/27.jpg)
Results
Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples
Radial arteryMean SD (range)
Arterialized EarlobeMean SD (range)
pH(pH Unit)
7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)
PO2
(mmHg)
42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)
PCO2
(mmHg)
34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)
SaO2
(%)
79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)
![Page 28: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/28.jpg)
Results
Blood Gas Data for Simultaneous Radial Artery and Earlobe Arterialized Blood Samples
Radial arteryMean SD (range)
Arterialized EarlobeMean SD (range)
pH(pH Unit)
7.43 0.02 (7.4 – 7.46) 7.43 0.02 (7.4 – 7.46)
PO2
(mmHg)
42.1 3.66 (38 – 47) 42.9 3.88 (37 – 50)
PCO2
(mmHg)
34.1 1.88 (31 – 37) 33.12 2.38 (29 – 37)
SaO2
(%)
79.0 3.85 (75 – 84.5) 79.9 3.29 (74 –85.6)
![Page 29: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/29.jpg)
Results
The mean difference in PO2 between arterialized earlobe and arterial samples was 0.25 ( ± 1.25) mmHg.
![Page 30: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/30.jpg)
Results
The mean difference in PO2 between arterialized earlobe and arterial samples was 0.25 ( ± 1.25) mmHg.
Comparison of PCO2 of the arterialized earlobe and arterial samples showed a mean difference of -1 ( ± 0.75) mmHg.
![Page 31: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/31.jpg)
Results
The mean difference in PO2 between arterialized earlobe and arterial samples was 0.25 ( ± 1.25) mmHg.
Comparison of PCO2 of the arterialized earlobe and arterial samples showed a mean difference of -1 ( ± 0.75) mmHg.
There was no difference between the pH values of the arterialized earlobe and arterial samples.
![Page 32: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/32.jpg)
Correlations between Arterial and Earlobe Blood Sample Measurements
r = 0.93 r = 0.94
r = 0.97 r = 0.92
![Page 33: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/33.jpg)
Discussion
Arterilaized Capillary- Arterial Mean ( SD)
PO2
mmHg
PCO2
mmHg Langlands & Wallace
(1965) (n=16)
0.62(4.1)
1.05(1.6)
Godfrey et al. (1971) (n=8)
2.09(2.48)
0.65(1.2)
Spiro & Dowdeswell (1971) (n=11)
-0.72(1.67)
1.0(1.91)
Dar et al. (1995) (n=55)
0.675(4.43)
0.75(2.25)
Present study (n=6)
0.25(1.25)
-1.0(0.75)
Study Author
![Page 34: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/34.jpg)
Discussion
Arterilaized Capillary- Arterial Mean ( SD)
PO2
mmHg
PCO2
mmHg Langlands & Wallace
(1965) (n=16)
0.62(4.1)
1.05(1.6)
Godfrey et al. (1971) (n=8)
2.09(2.48)
0.65(1.2)
Spiro & Dowdeswell (1971) (n=11)
-0.72(1.67)
1.0(1.91)
Dar et al. (1995) (n=55)
0.675(4.43)
0.75(2.25)
Present study (n=6)
0.25(1.25)
-1.0(0.75)
Study Author
![Page 35: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/35.jpg)
Discussion
Arterilaized Capillary- Arterial Mean ( SD)
PO2
mmHg
PCO2
mmHg Langlands & Wallace
(1965) (n=16)
0.62(4.1)
1.05(1.6)
Godfrey et al. (1971) (n=8)
2.09(2.48)
0.65(1.2)
Spiro & Dowdeswell (1971) (n=11)
-0.72(1.67)
1.0(1.91)
Dar et al. (1995) (n=55)
0.675(4.43)
0.75(2.25)
Present study (n=6)
0.25(1.25)
-1.0(0.75)
Study Author
![Page 36: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/36.jpg)
Discussion
Arterilaized Capillary- Arterial Mean ( SD)
PO2
mmHg
PCO2
mmHg Langlands & Wallace
(1965) (n=16)
0.62(4.1)
1.05(1.6)
Godfrey et al. (1971) (n=8)
2.09(2.48)
0.65(1.2)
Spiro & Dowdeswell (1971) (n=11)
-0.72(1.67)
1.0(1.91)
Dar et al. (1995) (n=55)
0.675(4.43)
0.75(2.25)
Present study (n=6)
0.25(1.25)
-1.0(0.75)
Study Author
![Page 37: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/37.jpg)
Discussion
Arterilaized Capillary- Arterial Mean ( SD)
PO2
mmHg
PCO2
mmHg Langlands & Wallace
(1965) (n=16)
0.62(4.1)
1.05(1.6)
Godfrey et al. (1971) (n=8)
2.09(2.48)
0.65(1.2)
Spiro & Dowdeswell (1971) (n=11)
-0.72(1.67)
1.0(1.91)
Dar et al. (1995) (n=55)
0.675(4.43)
0.75(2.25)
Present study (n=6)
0.25(1.25)
-1.0(0.75)
Study Author
![Page 38: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/38.jpg)
Discussion
Arterilaized Capillary- Arterial Mean ( SD)
PO2
mmHg
PCO2
mmHg Langlands & Wallace
(1965) (n=16)
0.62(4.1)
1.05(1.6)
Godfrey et al. (1971) (n=8)
2.09(2.48)
0.65(1.2)
Spiro & Dowdeswell (1971) (n=11)
-0.72(1.67)
1.0(1.91)
Dar et al. (1995) (n=55)
0.675(4.43)
0.75(2.25)
Present study (n=6)
0.25(1.25)
-1.0(0.75)
Study Author
![Page 39: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/39.jpg)
Discussion
Points of note.
![Page 40: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/40.jpg)
Discussion
Points of note.
• Arterialized blood collection from the earlobe is virtually pain free.
![Page 41: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/41.jpg)
Discussion
Points of note.
• Arterialized blood collection from the earlobe is virtually pain free.
• Anaesthetic is not required.
![Page 42: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/42.jpg)
Discussion
Points of note.
• Arterialized blood collection from the earlobe is virtually pain free.
• Anaesthetic is not required.
• Anaerobic blood collection is possible.
![Page 43: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/43.jpg)
Discussion
Points of note.
• Arterialized blood collection from the earlobe is virtually pain free.
• Anaesthetic is not required.
• Anaerobic blood collection is possible.
• EAB Collector prevents environmental contamination.
![Page 44: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/44.jpg)
Conclusion
This limited study suggests that:
![Page 45: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/45.jpg)
Conclusion
This limited study suggests that:
1. Arterialized blood sampled from the earlobe provides accurate estimations of arterial blood measurements of PO2, PCO2 and pH during microgravity simulation and in hypoxia.
![Page 46: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/46.jpg)
Conclusion
This limited study suggests that:
1. Arterialized blood sampled from the earlobe provides accurate estimations of arterial blood measurements of PO2, PCO2 and pH during microgravity simulation and in hypoxia.
2. Blood collection will be possible in microgravity without environmental contamination.
![Page 47: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/47.jpg)
Further Studies
1. To validate the EAB Collector during normoxia in other body positions (sitting and supine);
2. To test the efficiency of the EAB Collector in microgravity (parabolic flight).*
* Protocol under evaluation by ESA.
![Page 48: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/48.jpg)
Thank you for listening
Questions?
![Page 50: A ssessment of an Earlobe Arterialized Blood Collector for use in Microgravity](https://reader035.vdocuments.net/reader035/viewer/2022081603/5681431a550346895daf7456/html5/thumbnails/50.jpg)
Results
Individual Data