Transcript
  • 5/11/2015 Calibrationadewpointmeasurementsystem

    http://control.com/thread/1277824746 1/2

    CalibrationadewpointmeasurementsystemHowtocalibrateadewpointmeasurementsystem?Howcouldwemakesurethatthereadingiscorrect?

    ByVahidon29June,201011:19am

    WehaveaGEdewpointmeasurementequipmentinstalledontheoutletgaspipelineofourplant,Iwouldneedtohavemoreinformationaboutthewaysthatwecouldmakesurethatthereadingiscorrect.Idoappreciateyourconcernandadvice.

    ByCSA on29June,20107:31pm

    Doyouhavetheinstructionmanualforthedevice?

    Arethereanyinstructionsinthemanualaboutverifyingcalibration/output?

    Theseareusuallyverydifficulttocalibratewithoutalabsetupwhichcanflowairofaknownhumidityoverthesensor.

    Ibelievethebestwayistosendthedevicetoapropercalibrationlab,unlessyoursitehasaverygoodinstrumentationshop/lab.

    BySam on30June,20108:21am

    Hithere,

    NotmuchyoucandotoverifytheiraccuracyinthefieldI'mafraid.Isuggestyoutaketheoriginaldatasheetsuppliedwiththeanalyzer/transmitterandseeifthereissomecalibrationconstantsthatyoucancheck.Consultthemanualsincetheymighthaveacalibrationvalidationprocedureinthere.ItwillbesomeconstantsvaluesyoucancheckonthedisplayorsomevoltagesyoucanmeasureonthePCB.

    SomethingyoucancheckisthemAoutputsaccordingtothefactorycalibratedzeroandspanvalues.Ifitisdewpointyouhaveontheoutputitshouldhaveafacilityonthematrixofthetransmittertosimulatethe0to150degreesCorwhateveryourfactorycalibrationshouldbeaccordingtothedatasheet.Youcanthenmakeadjustmentsonthetransmitter'soutputtomakesurethecorrectZ/Sdewpointvalues(asperdatasheet)giveaaccurate4to20mAoutput.

    Iknowthemanufacturercanalsosupplyyouwithvarioussaturatedsaltsolutionsthatyoucanuseinalaborw/shopenvironmenttoverifytheaccuracyofthetransmittersogetholdofthemandaskthemaboutthat.

    Anotherpossibilityistobuyahotfactorycalibratedstandbyidenticaltotheinstalledunitandjust

  • 5/11/2015 Calibrationadewpointmeasurementsystem

    http://control.com/thread/1277824746 2/2

    changeitoutwiththisspareonewhileyousendtheotheroneforthecalibrationverificationtothefactoryoneayearoronceeverytwoyears.Theremightbealimitedsensorshelfliveorspecialsensorstoragerequirementsinvolved,sospeaktothesupplierandseeifthismightbepossibleoption.

    YouruseofthissiteissubjecttothetermsandconditionssetforthunderLegalNoticesandthePrivacyPolicy.Pleasereadthosetermsandconditionscarefully.SubjecttotherightsexpresslyreservedtoothersunderLegalNotices,thecontentofthissiteandthecompilationthereofis19992015NerdsinControl,LLC.Allrightsreserved.


Top Related