Contrast Echocardiography
• DR PRASANTH S
![Page 2: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/2.jpg)
Introduction
• US contrast agents first used- mid 1970
• Gas containing microbubbles.
• First generation Contrast Agents:
Agitated saline with or without Indocyanine green.
![Page 3: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/3.jpg)
Agitated Saline
• Agitating a solution of saline
between two 10-mL syringes
• Each of which contains 5 mL of
saline and 0.1 to 0.5 mL of
room air
• Forceful agitation through a
three-way stopcock creates a
population of microbubbles
• ‘Dose’- 1- 5 ml
![Page 4: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/4.jpg)
Ideal contrast agent
• Non-toxic• Intravenously injectable• Has to behave similarly to blood• Crosses pulmonary filter• Resistant to intravascular and intra-cardiac
pressures• Stable throughout during the exam• Improve the Doppler signal-to-noise ratio
![Page 5: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/5.jpg)
![Page 6: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/6.jpg)
Recent microbubble formulations
Name Size (µm) Shell composition Gas content indicationAI-700 2.9 SYNTHETIC POLYMER PERFLUORO
CARBONMyocardial perfusion
CARDIOspere 4.0 POLYMER BILAYER NITROGEN Myocardial perfusion
DEFINITY(USA)
1.1-3.3 Lipid encapsulated PERFLUOROPROPANE
LV opacification
OPTISON7(USA)
2.0-4.5 DENATURED ALBUMIN
PERFLUOROPROPANE
LV opacification
SONOVUE(EUROPE, ASIA)
2.5 PHOSPHOLIPIDS SULPHUR HEXAFLUORIDE
Myocardial perfusion, LV opacification
![Page 7: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/7.jpg)
• Low surface tension.
• Resistant to ultrasound destruction.
• Slowly diffusing, insoluble, high molecular weight gases.
• 1.1 – 8 µm size, 5х10⁸ to 1.2x 10¹⁰ microbubbles per millilitre
• Single injection provide contrast effect for 3- 10 min.
• Safe – 4 deaths after 2 million use
• Contra indications
– Known Rt to Lt shunts
– Known hypersensitivity
![Page 8: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/8.jpg)
Ultrasound Interaction with Contrast Agent
<0.3 MI
>0.3 MI
![Page 9: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/9.jpg)
![Page 10: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/10.jpg)
Machine settings
• Dedicated contrast specific presets
• Mechanical Index; Power of US beam
Peak Negative acoustic pressure Transmitted Frequency
Routine B mode uses – High MI - 0.9 to 1.4
Low MI < 0.3
![Page 11: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/11.jpg)
Contrast Destruction
• High Mechanical Index
• High Frame rate
• Focal zone
• Near field
![Page 12: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/12.jpg)
![Page 13: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/13.jpg)
Fundamental Harmonic
![Page 14: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/14.jpg)
Continuous Imaging
Low Mechanical Index High Mechanical Index
![Page 15: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/15.jpg)
Intermittent imaging
• Triggered to ECG• In between imaging, no
ultrasound energy is delivered.
• Allows time for restitution of contrast effect.
• Analysis of wall motion-not possible.
• Evaluation of myocardial perfusion.
Continuous low MI imaging.
• Wall motion analysis in real time.
• Used for cavity opacification.
• Detection of very low concentration of myocardial contrast.
![Page 16: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/16.jpg)
Intermittent Triggered Imaging
oIntermittent Imaging
![Page 17: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/17.jpg)
Power Spectrum
Motion of the bubbles & their resonance in a stationary field
![Page 18: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/18.jpg)
Clinical Applications
![Page 19: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/19.jpg)
Detection and Utilization of Intracavitary contrast
• Enhanced visualization of the LV endocardial borders
• Improve reproducibility for wall motion analysis and
volumetric measurements
• Detection or exclusion of
– Intracavitary thrombus
– Ventricular noncompaction
– Atypical forms of HCM (Apical)
– Abnormal communication to the ventricular chamber
![Page 20: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/20.jpg)
Exclusion of Thrombus
![Page 21: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/21.jpg)
LV Thrombus
![Page 22: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/22.jpg)
Ventricular noncompaction
![Page 23: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/23.jpg)
![Page 24: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/24.jpg)
Spectral Doppler Enhancement
• Low concentrations of contrast
agents
• Enhancing the tricuspid
regurgitation jet
• Pulmonary vein flow
• Increasing intensity of a relatively
weak aortic stenosis jet
![Page 25: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/25.jpg)
Shunt Detection
• Right-to-left shunts - agitated saline - agent of choice
– Atrial septal defects of all types
– Patent foramen ovale - Valsalva and cough
– Pulmonary arteriovenous malformations - 5 to 15 cycles
– Larger ventricular septal defects during diastole
– Left SVC
• Left-to-right shunt
– Negative contrast effect
![Page 26: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/26.jpg)
Right-to-left shunt Negative contrast effect
![Page 27: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/27.jpg)
Atrial septal aneurysm with PFO
![Page 28: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/28.jpg)
Persistent Left SVC
![Page 29: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/29.jpg)
Myocardial Perfusion Contrast
• First recognized in the 1980s
• Preserved contrast effect in the myocardium - evidence of
microvascular integrity and blood flow to the area
• Analysis of myocardial flow - Time of appearance curve
• Multiple time appearance curve analyses - necessary
• Time of appearance curve requires a bolus effect
– wait 10 minutes
– purposeful destruction of the contrast agent - burst of high intensity
(high mechanical index) ultrasound
• Targeted to different regions of interest
• Performed under basal conditions & after vasodilator stress
![Page 30: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/30.jpg)
![Page 31: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/31.jpg)
Time of appearance curve
![Page 32: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/32.jpg)
• α is directly related to myocardial blood volume
• β is related to flow rate
• The product of α and β- proportional to myocardial blood flow
• Vasodilator results in an increase in flow velocity in those
areas not perfused by a stenosed artery
• Appearance of the contrast curves - differ in the normal and
diseased beds
![Page 33: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/33.jpg)
![Page 34: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/34.jpg)
Transcatheter alcohol septal ablation
• Performed for the Rx of HOCM.• Catheter is placed in the 1st septal perforator of LAD.• Controlled myocardial infarction for reduction of
proximal septal mass.• Before alcohol injection, diluted US contrast agent is
injected to the selected artery.• To ensure- no contrast reflux.• To confirm the presence and size of vascular bed.
![Page 35: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/35.jpg)
Transcatheter alcohol septal ablation
![Page 36: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/36.jpg)
![Page 37: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/37.jpg)
Attenuation & Shadowing
![Page 38: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/38.jpg)
Papillary Muscle Shadow
![Page 39: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/39.jpg)
Colour Artifact
![Page 40: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/40.jpg)
Competitive Flow
May be confused with a true negative contrast effect due to an atrial septal defect
![Page 41: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/41.jpg)
Prominent eustachian valve and margination of contrast-enhanced blood flow
May be confused with a true negative contrast effect due to an atrial septal defect
![Page 42: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/42.jpg)
Strain Rate Imaging
![Page 43: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/43.jpg)
Introduction
• Evaluation of a myocardial region with reference to an adjacent myocardial segment.
• Deformation analysis- analysis of ventricular mechanics or shapes during cardiac cycle.
• Myocardial strain, strain rate, torsion.
• Strain- percentage thickening or deformation of the myocardium during the cardiac cycle.
• Change of strain per unit of time is referred to as strain rate
![Page 44: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/44.jpg)
Strain & Strain rate
![Page 45: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/45.jpg)
• Strain calculated in three orthogonal planes- representing longitudinal, radial, circumferential contraction.
• Negative strain- shortening of segment.• Positive strain- lengthening of segment
![Page 46: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/46.jpg)
![Page 47: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/47.jpg)
Methods
Doppler tissue imaging
• Two discrete points are compared for change in velocity• Strain rate- primary parameter obtained• Strain –derived by integrating velocity over time.
Speckle tracking
• Actual location of discrete myocardial segments calculated. • Strain is the primary parameter.• Strain rate-derived by calculating change in distance over time.
![Page 48: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/48.jpg)
![Page 49: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/49.jpg)
SR- Doppler tissue imaging
![Page 50: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/50.jpg)
Speckle tracking
• ‘Speckles’ are small dots or groups of myocardial pixels that are created by the interaction of ultrasonic beams and the myocardium.
• Considered as acoustic fingerprint for that region.
• This enables to judge the direction of movement, the speed of such movement, and the distance of such movement of any points in the myocardium.
![Page 51: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/51.jpg)
Speckle
![Page 52: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/52.jpg)
Method
• Track the endocardial and epicardial borders of the left
ventricle
• Correctly define the region of interest (ROI) in the long or
short axis view
• Post-processing software automatically divides the
ventricle into six equally distributed segments
• 2D or 3D data set is produced
• Mathematical algorithms are applied to generate values
![Page 53: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/53.jpg)
• Strain is not uniform among all myocardial segments.
• Radial strain-Magnitude of basal parameters are higher than the apical values.
• Longitudinal strain- less variability fron apex to base.
• Circumferential strain- higher in anterior and lateral walls compared to posterior and septal.
• Normal longitudinal strain averages -20%• Normal radial strain about +40%
![Page 54: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/54.jpg)
![Page 55: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/55.jpg)
Normal Strain Displays Wave Forms ,Curved M-mode
![Page 56: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/56.jpg)
Normal Strain Displays- bulls eye presentation
![Page 57: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/57.jpg)
![Page 58: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/58.jpg)
Normal pattern Dilated cardiomyopathyDyssynchrony
![Page 59: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/59.jpg)
Velocity vector imaging
![Page 60: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/60.jpg)
VENTRICULAR TORSION
• Similar to the winding and Unwinding of a towel.
• Isovolumetric contraction the apex rotates clockwise
• Ejection phase apex rotates counterclockwise & base rotates
clockwise when viewed from the apex
• Diastole - relaxation of myocardial fibres - recoiling -
clockwise apical rotation.
![Page 61: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/61.jpg)
Myocardial mechanics
![Page 62: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/62.jpg)
• Rotation - Measure of the rotational movement of the myocardium in relation to an imaginary long axis line from apex to base drawn through the middle of LV cavity.
• Twist (degrees) is the net difference between apical and basal
rotation• Torsion - Twist divided by the vertical distance between the
apex and base and is expressed as degrees/cm.
![Page 63: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/63.jpg)
VENTRICULAR TORSION
![Page 64: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/64.jpg)
Applications
• CAD- Myocardial ischemia, Myocardial infarction, Myocardial
viability.
• Heart failure with normal LVEF
• Cardiac resynchronization therapy (CRT)
• DCM
• HCM.• Detection of subclinical diseases/early myocardial
involvement
![Page 65: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/65.jpg)
Applications
• Stress cardiomyopathy• Restrictive cardiomyopathy• Detection of rejection and coronary stenosis in heart
transplant patients.• Early detection of chemotherapy induced
cardiotoxicity.• Valvular heart disease
![Page 66: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/66.jpg)
![Page 67: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/67.jpg)
![Page 68: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/68.jpg)
![Page 69: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/69.jpg)
![Page 70: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/70.jpg)
![Page 71: Contrast Echocardiography DR PRASANTH S. Introduction US contrast agents first used- mid 1970 Gas containing microbubbles. First generation Contrast Agents:](https://reader037.vdocuments.net/reader037/viewer/2022102808/56649db15503460f94a9fd80/html5/thumbnails/71.jpg)