Page 1: Haemodynamic Monitoring

Haemodynamic Monitoring

Theory and Practice

Page 2: Haemodynamic Monitoring

PiCCO Technology is a combination of transpulmonary thermodilution and pulse contour analysis

Principles of Measurement

Left HeartRight Heart

Pulmonary Circulation

Lungs

Body CirculationPULSIOCATHPULSIOCATH

CVC

PULSIOCATH arterial thermodilution catheter

central venous bolus injection

Introduction to the PiCCO-Technology – Function

Page 3: Haemodynamic Monitoring

Bolus injection

concentration changes over time(Thermodilution curve)

After central venous injection the cold bolus sequentially passes through the various intrathoracic compartments

The temperature change over time is registered by a sensor at the tip of the arterial catheter

Introduction to the PiCCO-Technology – Function

Left heartRight heart Lungs

RA RV LA LVPBV

EVLW

EVLW

Principles of Measurement

Page 4: Haemodynamic Monitoring

Intrathoracic Compartments (mixing chambers)

Introduction to the PiCCO-Technology – Function

Pulmonary Thermal Volume (PTV)

Intrathoracic Thermal Volume (ITTV)

Total of mixing chambers

RA RV LA LVPBV

EVLW

EVLW

Largest single mixing chamber

Page 5: Haemodynamic Monitoring

Haemodynamic Monitoring

E. Introduction to PiCCO Technology

1. Principles of function

2. Thermodilution

3. Pulse contour analysis

4. Contractility parameters

5. Afterload parameters

6. Extravascular Lung Water

7. Pulmonary Permeability

Page 6: Haemodynamic Monitoring

Tb x dt

(Tb - Ti) x Vi x

K

Tb

Injection

t

∫ =COTD a

Tb = Blood temperatureTi = Injectate temperatureVi = Injectate volume∫ ∆ Tb

. dt = Area under the thermodilution curve

K = Correction constant, made up of specific weight and specific heat of blood and injectate

The CO is calculated by analysis of the thermodilution curve using the modified Stewart-Hamilton algorithm

Calculation of the Cardiac Output

Introduction to the PiCCO-Technology – Thermodilution

Page 7: Haemodynamic Monitoring

The area under the thermodilution curve is inversely proportional to the CO.

36,5

37

5 10

Thermodilution curves

Normal CO: 5.5l/min

Introduction to the PiCCO-Technology – Thermodilution

36,5

37

36,5

37

Time

low CO: 1.9l/min

High CO: 19l/min

Time

Time

Temperature

Temperature

Temperature

Page 8: Haemodynamic Monitoring

Transpulmonary vs. Pulmonary Artery Thermodilution

Left heartRight Heart

Pulmonary Circulation

Lungs

Body Circulation

PULSIOCATH arterial thermo-dilution catheter

central venous bolus injection RA

RV

PA

LA

LV

Aorta

Transpulmonary TD (PiCCO) Pulmonary Artery TD (PAC)

In both procedures only part of the injected indicator passes the thermistor.

Nonetheless the determination of CO is correct, as it is not the amount of the detected indicator but the difference in temperature over time that is relevant!

Introduction to the PiCCO –Technology – Thermodilution

Page 9: Haemodynamic Monitoring

MTt: Mean Transit time the mean time required for the indicator to reach the detection point

DSt: Down Slope time the exponential downslope time of the thermodilution curve

Recirculation

t

e-1

Tb

From the characteristics of the thermodilution curve it is possible to determine certain time parameters

Extended analysis of the thermodilution curve

Introduction to the PiCCO-Technology – Thermodilution

Injection

In Tb

MTt DSt

Tb = blood temperature; lnTb = logarithmic blood temperature; t = time

Page 10: Haemodynamic Monitoring

Pulmonary Thermal Volume

PTV = Dst x CO

By using the time parameters from the thermodilution curve and the CO ITTV and PTV can be calculated

Calculation of ITTV and PTV

Introduction to the PiCCO-Technology – Thermodilution

Recirculation

t

e-1

Tb

Injection

In Tb

Intrathoracic Thermal Volume

ITTV = MTt x CO

MTt DSt

Page 11: Haemodynamic Monitoring

Pulmonary Thermal Volume (PTV)

Intrathoracic Thermal Volume (ITTV)

Calculation of ITTV and PTV

Einführung in die PiCCO-Technologie – Thermodilution

ITTV = MTt x CO

PTV = Dst x CO

RA RV LA LVPBV

EVLW

EVLW

Page 12: Haemodynamic Monitoring

GEDV is the difference between intrathoracic and pulmonary thermal volumes

Global End-diastolic Volume (GEDV)

Volumetric preload parameters – GEDV

RA RV LA LVPBV

EVLW

EVLW

ITTV

GEDV

PTV

Introduction to the PiCCO –Technology – Thermodilution

Page 13: Haemodynamic Monitoring

Volumetric preload parameters – ITBV

Intrathoracic Blood Volume (ITBV)

GEDV

ITBV

PBVRA RV LA LVPBV

EVLW

EVLW

Introduction to the PiCCO –Technology – Thermodilution

ITBV is the total of the Global End-Diastolic Volume and the blood volume in the pulmonary vessels (PBV)

Page 14: Haemodynamic Monitoring

ITBVTD (ml)

ITBV = 1.25 * GEDV – 28.4 [ml]

GEDV vs. ITBV in 57 Intensive Care Patients

IIntrantratthoracic horacic BBlood lood VVolume (olume (ITBVITBV))

Volumetric preload parameters – ITBV

Introduction to the PiCCO-Technology – Thermodilution

ITBV is calculated from the GEDV by the PiCCO Technology

0

1000

2000

3000

0 1000 2000 3000 GEDV (ml)

Sakka et al, Intensive Care Med 26: 180-187, 2000

Page 15: Haemodynamic Monitoring

Intrathoracic Compartments (mixing chambers)

Introduction to the PiCCO-Technology – Function

Pulmonary Thermal Volume (PTV)

Intrathoracic Thermal Volume (ITTV)

Total of mixing chambers

RA RV LA LVPBV

EVLW

EVLW

Largest single mixing chamber

Page 16: Haemodynamic Monitoring

ITTV

– ITBV

= EVLW

The Extravascular Lung Water is the difference between the intrathoracic thermal volume and the intrathoracic blood volume. It represents the amount of water in the lungs outside the blood vessels.

Calculation of Extravascular Lung Water (EVLW)

Introduction to the PiCCO –Technology – Extravascular Lung Water

Page 17: Haemodynamic Monitoring

Katzenelson et al,Crit Care Med 32 (7), 2004 Sakka et al, Intensive Care Med 26: 180-187, 2000

Gravimetry Dye dilution

EVLW from the PiCCO technology has been shown to have a good correlation with the measurement of extravascular lung water via the gravimetry and dye dilution reference methods

Validation of Extravascular Lung Water

n = 209r = 0.96

ELWI by gravimetry

ELWI by PiCCO

R = 0,97P < 0,001

Y = 1.03x + 2.49

0

10

20

30

20 30

40

10

ELWITD (ml/kg)

0

5

10

20

15 25

25

50 100 20

15

ELWIST (ml/kg)

Introduction to the PiCCO –Technology – Extravascular Lung Water

Page 18: Haemodynamic Monitoring

Haemodynamic Monitoring

E. Introduction to PiCCO Technology

1. Principles of function

2. Thermodilution

3. Pulse contour analysis

4. Contractility parameters

5. Afterload parameters

6. Extravascular Lung Water

7. Pulmonary Permeability

Page 19: Haemodynamic Monitoring

Transpulmonary Thermodilution

The pulse contour analysis is calibrated through the transpulmonary thermodilution and is a beat to beat real time analysis of the arterial pressure curve

Calibration of the Pulse Contour Analysis

Introduction to the PiCCO-Technology – Pulse contour analysis

Injection

Pulse Contour Analysis

T = blood temperature t = timeP = blood pressure

COCOTPDTPD= SV= SVTDTD

HRHR

Page 20: Haemodynamic Monitoring

PCCO = cal • HR •P(t)SVR

+ C(p) •dPdt

( ) dt

Cardiac Output

Patient- specific calibration factor (determined by thermodilution)

Heart rate Area under the pressure curve

Shape of the pressure curve

Aortic compliance

Systole

Introduction to the PiCCO-Technology – Pulse contour analysis

Parameters of Pulse Contour Analysis

Page 21: Haemodynamic Monitoring

SVSVmaxmax – SV – SVminminSVV =SVV =

SVSVmeanmean

SVSVmaxmax

SVSVminmin

SVSVmeanmean

The Stroke Volume Variation is the variation in stroke volume over the ventilatory cycle, measured over the previous 30 second period.

Parameters of Pulse Contour Analysis

Introduction to the PiCCO-Technology – Pulse Contour Analysis

Dynamic parameters of volume responsiveness – Stroke Volume Variation

Page 22: Haemodynamic Monitoring

PPPPmaxmax – PP – PPminminPPV =PPV =

PPPPmeanmean

The pulse pressure variation is the variation in pulse pressure over the ventilatory cycle, measured over the previous 30 second period.

Parameters of Pulse Contour Analysis

Introduction to the PiCCO-Technology – Pulse Contour Analysis

Dynamic parameters of volume responsiveness – Pulse Pressure Variation

PPPPmaxmax

PPPPmeanmean

PPPPminmin

Page 23: Haemodynamic Monitoring

Summary pulse contour analysis - CO and volume responsiveness

• The PiCCO technology pulse contour analysis is calibrated by transpulmonary thermodilution

• PiCCO technology analyses the arterial pressure curve beat by beat thereby providing real time parameters

• Besides cardiac output, the dynamic parameters of volume responsiveness SVV (stroke volume variation) and PPV (pulse pressure variation) are determined continuously

Introduction to the PiCCO-Technology – Pulse contour analysis