Download - Implementation-Reference ASTM Records V3-x
Document: IMPREF3x.DOC
Boehringer Mannheim GmbH
Elecsys® 2010 Analyzer
Host Interface
Elecsys® 2010 - Host Interface
Implementation Reference ASTM Records
V 3.x - February 98
Revision Comments Author DateThis document is based on Mr.Mashiko´s (HID) implementationdocumentation
1.4 Changes from 1.3x to 1.4xindicated in blue color
Centner May 5, 97
2.x Changes from 1.4 to 2.xNone due to version was renamedNo extra document
Centner
3.x Changes from 2.x to 3.xindicated in dark green color
Centner Feb. 3, 98
3.x Correctionsindicated in dark green color
Centner Feb. 13, 98
Implementation Reference - ASTM Records
Elecsys® 2010 - Host Interface
Contents 1
Elecsys® 2010 - Host Interface Contents
1. Contents1. Contents....................................................................................................................................... 1
2. Test Order Message - Q/P/O ...................................................................................................... 4Purpose to use this message: ............................................................................................................... 4
Instrument Query — Host Answer ...................................................................................................... 4Host Query — Instrument Answer ..................................................................................................... 4Host Download.................................................................................................................................... 4Instrument Cancel Query .................................................................................................................... 5
Construction of Message:...................................................................................................................... 5Trace Example: ..................................................................................................................................... 5Details of Records: ................................................................................................................................ 6
3. Result Message - Q/P/O/R/C/RR/RC ........................................................................................ 18Purpose to use this message: ............................................................................................................. 18
Host Query — Instrument Answer .................................................................................................... 18Instrument à Host Upload................................................................................................................ 18
Construction of Message:.................................................................................................................... 19Trace Example: ................................................................................................................................... 19Trace Example: ................................................................................................................................... 20Details of Records: .............................................................................................................................. 21
4. Substance Data Message - XT/DP ........................................................................................... 30Purpose to use this message: ............................................................................................................. 30
Instrument à Host Upload................................................................................................................ 30Host Query — Instrument Answer .................................................................................................... 31Host Download.................................................................................................................................. 32
Construction of Message:.................................................................................................................... 32Trace Example: ................................................................................................................................... 32Trace Example: ................................................................................................................................... 35Details of Records: .............................................................................................................................. 38
5. Calibrator Parameters Message - XT/CP................................................................................. 43Purpose to use this message: ............................................................................................................. 43
Upload Instrument à Host................................................................................................................ 43Host Query — Instrument Answer .................................................................................................... 43Download Host à Instrument ........................................................................................................... 43
Construction of Message:.................................................................................................................... 44Trace Example: ................................................................................................................................... 44Details of Records: .............................................................................................................................. 45
6. Control Parameters Message - XT/QP .................................................................................... 48Purpose to use this message: ............................................................................................................. 48
Instrument à Host Upload................................................................................................................ 48Host Query — Instrument Answer .................................................................................................... 48Host Download.................................................................................................................................. 48
Construction of Message:.................................................................................................................... 49Trace Example: ................................................................................................................................... 49Details of Records: .............................................................................................................................. 50
7. Calibration Data Message - CR................................................................................................ 54Purpose to use this message: ............................................................................................................. 54
Instrument à Host Upload................................................................................................................ 54Construction of Message:.................................................................................................................... 54
2 Contents
Contents Elecsys® 2010 - Host Interface
Trace Example:....................................................................................................................................54Details of Records: ..............................................................................................................................55
8. Sample Status Message - SS ...................................................................................................58Purpose to use this message: .............................................................................................................58
Host Query — Instrument Answer ....................................................................................................58Construction of Message: ....................................................................................................................58Details of Records: ..............................................................................................................................58
Instrument Sending Record Pattern..................................................................................................58
9. Instrument Status Message - IS ...............................................................................................61Purpose to use this message: .............................................................................................................61
Host Query — Instrument Answer ...................................................................................................61Instrument à Host Upload................................................................................................................61
Construction of Message: ....................................................................................................................61Trace Example:....................................................................................................................................61Details of Records: ..............................................................................................................................62
10. Log File Message - LF...............................................................................................................65Purpose to use this message: .............................................................................................................65
Host Query — Instrument Answer ....................................................................................................65Construction of Message: ....................................................................................................................65Details of Records: ..............................................................................................................................65
11. Test Conditions Message - TC.................................................................................................68Purpose to use this message: .............................................................................................................68
Instrument à Host Upload................................................................................................................68Host Query — Instrument Answer ....................................................................................................68Host Download..................................................................................................................................68
Construction of Message: ....................................................................................................................68Trace Example:....................................................................................................................................69Details of Records: ..............................................................................................................................70
12. Test Application Message - TA ................................................................................................73Purpose to use this message: .............................................................................................................73
Instrument à Host Upload................................................................................................................73Trace Example:....................................................................................................................................73Details of Records: ..............................................................................................................................74
13. Instrument Configuration Message - IC ..................................................................................75Purpose to use this message: .............................................................................................................75
Host Query — Instrument Answer ...................................................................................................75Host Download..................................................................................................................................75Instrument à Host Upload................................................................................................................75
Construction of Message: ....................................................................................................................75Trace Example.....................................................................................................................................75Details of Records: ..............................................................................................................................76
14. Service Data Message - SD ......................................................................................................80Purpose to use this message: .............................................................................................................80
Instrument à Host Upload................................................................................................................80Host Query — Instrument Answer ....................................................................................................80
Construction of Message: ....................................................................................................................80Trace Example:....................................................................................................................................80Details of Records: ..............................................................................................................................82
15. Processing Message - PM ........................................................................................................84
Contents 3
Elecsys® 2010 - Host Interface Contents
Purpose to use this message: ............................................................................................................. 84Instrument à Host Upload................................................................................................................ 84
Construction of Message:.................................................................................................................... 84Trace Example: ................................................................................................................................... 84Details of Records: .............................................................................................................................. 84
16. Action Request / Action Confirmation Message - AR/AC...................................................... 86Purpose to use this message: ............................................................................................................. 86
Host Request — Instrument Confirmation ........................................................................................ 86Construction of Message:.................................................................................................................... 86Details of Records: .............................................................................................................................. 86
17. Header - H .................................................................................................................................. 88Details of Records: .............................................................................................................................. 88
18. Termination - L .......................................................................................................................... 89Details of Records: .............................................................................................................................. 89
19. Receive Acceptance Status...................................................................................................... 90When does Elecsys accept messages from host? (Disk Type) .......................................................... 90When does Elecsys accept messages from host? (Rack Type) ......................................................... 90
20. Data Alarm List .......................................................................................................................... 91
21. Index........................................................................................................................................... 94
4 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
2. Test Order Message - Q/P/O
Purpose to use this message:
Instrument Query — Host Answer
Purpose Host InstrumentQuery after reading sample bar-code(in case that Host answers positively)(Query Test order for specimens that have no Test order.)
L-11, L-21(M-2)
I-1(M-1)
Query after reading sample bar-code(in case that Host answers negatively)(Query Test order for specimens that have no Test order.)
L-11, L-22(M-2)
I-1(M-1)
Event: Sample rotor scanningà The instrument queries for test requisitions for each sample that has no testrequisitions using 1 message/ 1 sample.Host must answer these queries for each sample by a certain time while systembegins to make schedule for the sample.Even if the Host answers these queries after the certain time out limit, the instrumentaccepts these responses. In this case, the instrument regards these responses asbatch download. (ignores these responses as answers to previous queries)
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Test order(in case that Instrument answers positively)(Answer Information of Test Order that were requested fromHost.)
L-1(M-1)
I-11, I-21(M-2)
Query for Information of Test order(in case that Instrument answers negatively)(Answer negatively when there is no Information of TestOrder that were requested from Host.)
L-1(M-1)
I-11, I-22(M-2)
Event: The instrument receives a Test Order Message Query.à The instrument answers the query. (positively or negatively).Host queries for test selection information for a sample.(or for some samples).The instrument always answers 1 message/1 sample even if Host queries for testselection information for some samples using one message.
Host Download
Purpose Host InstrumentOrdering of tests and canceling of tests(Order tests for specimens and cancel the tests.)
L-11, L-23(M-2)
Event: The instrument receives a Test Order Message.à The instrument acts as following.
- new registration for unknown sample.- overwriting known sample.
Host downloads a Test Order Message for one sample.(or for some samples)
Test Order Message - Q/P/O 5
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
Instrument Cancel Query
Purpose Host InstrumentCanceling of the previous Query(Cancel the previous Query because of time out.)
I-2(M-1)
L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 Q0 L
Pattern 2 (M-2)Level Record
0 H1 P2 O0 L
Trace Example:
Instrument queries test selections for sample ‘000004’: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2Q|1|^000004^40^0^4^^SAMPLE^NORMAL||ALL
||||||||O<CR><ETX>C5<CR><LF> <ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
Host sends back test selections for sample ‘000004’: <ENQ> <ACK> <STX>1H|\^&||||||||||P<CR><ETX>0D<CR><LF> <ACK> <STX>2P|1<CR><ETX>3F<CR><LF> <ACK> <STX>3O|1|000004|40^0^4^^SAMPLE^NORMAL|^^^10\^^^50\^^^20\
^^^30|R|19970425085218|||||N||||||||||||||Q<CR><ETX>01<CR><LF>
<ACK> <STX>4L|1|F<CR><ETX>FF<CR><LF> <ACK> <EOT>
Instruments queries for test selections for control ‘PC U1’: <ENQ>
6 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
<ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2Q|1|^PC U1^37^0^1^^CONTROL^NORMAL||ALL
|||||||190599|O<CR><ETX>7D<CR><LF> <ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
Host sends back test selections and reduces volume: <ENQ> <ACK> <STX>1H|\^&||||||||||P<CR><ETX>0D<CR><LF> <ACK> <STX>2P|1<CR><ETX>3F<CR><LF> <ACK> <STX>3O|1|PC U1|37^0^1^^CONTROL^REDUCED|^^^10\^^^50\^^^20\^^^30|
R|19970403110957|||||N\Q||||||||||||||Q<CR><ETX>53<CR><LF>
<ACK> <STX>4L|1|F<CR><ETX>FF<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Request Information Record [Q Record] Pattern Pattern
Field Data Type Content I-1 I-2
1 Record type ID (Q) char[1] Q Q Q
2 Sequence number pos_int
3 Starting range ID number(patients/specimen Id/instrument specimen ID)
text^text^text
^text{^text}{^text}{^text}
^Sample ID^Sequence No.^S.Rotor No. or
RackID^Position No.
^^data carrier
type *5)^container
type *6)
^Sample ID^Sequence No.^S.Rotor No. or
RackID^Position No.
^^data carrier
type *5)^container
type *6)
4 ending range ID number [NONE] [NONE]
5 universal test ID text ALL ALL
6 nature of request timelimits
char[1] [NONE] [NONE]
7 beginning request resultsdate and time
d_t [NONE] [NONE]
8 ending request resultsdate and time
d_t [NONE] [NONE]
9 requesting physicianname
text [NONE] [NONE]
10 requesting physician text [NONE] [NONE]
Test Order Message - Q/P/O 7
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
Request Information Record [Q Record] Pattern Pattern
Field Data Type Content I-1 I-2telephone number
11 user field NO.1 text [NONE] [NONE]
12 user field NO.2 text Control Lot No Control Lot No
13 request informationstatus codes
char[1 O O A
Patient Information Record [P Record] Pattern
Field Data Type Content I-11
1 record type ID (P) char[1] P P
2 sequence number pos_int
3 practice assigned patientID
text [NONE]
4 laboratory assignedpatient ID
text patient ID
5 patient ID No.3 text [NONE]
6 patient name text^text^text^text
[NONE]
7 mothers maiden name text [NONE]
8 date of birth d [NONE]
9 patient sex char[1] [NONE]
10 patient race-ethnic origin text [NONE]
11 patient address text [NONE]
12 reserved field [NONE]
13 patient telephone number text [NONE]
14 attending physician ID text [NONE]
15 special field 1 text [NONE]
16 special field 2 text [NONE]
17 patient height (in cm) int{^unit}
[NONE]
18 patient weight in (kg) int{^unit}
[NONE]
19 patient's known orsuspected diagnosis
text [NONE]
20 patient activemedications
text [NONE]
21 patient's diet text [NONE]
22 practice field No.1 text [NONE]
23 practice field No.2 text [NONE]
8 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
Patient Information Record [P Record] Pattern
Field Data Type Content I-11
24 admission and dischargedates
d{\d}
[NONE]
25 admission status char[2] [NONE]
26 location text [NONE]
27 nature of alternativediagnostic code andclassifiers
text [NONE]
28 alternative diagnosticcode and classification
text [NONE]
29 patient religion char[1] ortext
[NONE]
30 marital status char[1] [NONE]
31 isolation status char[3] ortext
[NONE
32 language status text [NONE]
33 hospital service text [NONE]
34 hospital institution text [NONE]
35 dosage category text [NONE]
Test Order Record [O Record] Pattern Pattern
Field Data Type Content I-21 I-22
1 record type ID (O) char[1] O O O
2 sequence number pos_int
3 specimen ID text Sample ID Sample ID
4 instrument specimen ID text
{^text}{^text}{^text}{^text}{^text}
Sequence No.^S. Rotor No.
or RackID^Position No
^^data carrier
type *5)^container
type *6)
Sequence No.^S. Rotor No.
or RackID^Position No
^^
^
5 universal test ID text^text^text^text
{^text}
{^text}
{\ }
^^
^Test No.^ Auto Dil.
Reference *7)^Predilutionfactor *10)
{\ }
[NONE]
6 priority Char[1] S, R S: STATR: routine
[NONE]
Test Order Message - Q/P/O 9
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
Test Order Record [O Record] Pattern Pattern
Field Data Type Content I-21 I-22
7 request/ordered date andtime
d_t ordered date &time
[NONE]
8 specimen collection dateand time
d_t [NONE] [NONE]
9 collection end time d_t [NONE] [NONE]
10 collection volume (in ml) int{^unit}
[NONE] [NONE]
11 collector ID text [NONE] [NONE]
12 action code char[1]{\...}
C, N, Q N *4) [NONE]
13 danger code text [NONE] [NONE]
14 relevant clinicalinformation
text [NONE] [NONE]
15 date/time specimenreceived
d_t [NONE] [NONE]
16 specimen descriptor(type & source)
text^text
[NONE] [NONE]
17 ordering physician text [NONE] [NONE]
18 physician’s telephonenumber
text [NONE] [NONE]
19 user field No.1 text [NONE] [NONE]
20 user field No.2 text [NONE] [NONE]
21 laboratory field No.1 text [NONE] [NONE]
22 laboratory field No.1 text [NONE] [NONE]
23 date/time results reportedor last modified
d_t [NONE] [NONE]
24 instrument change tocomputer system
text [NONE] [NONE]
25 instrument section ID text [NONE] [NONE]
26 report types char[1]{\...}
Q, Z, O Q Z
27 reserved field [NONE] [NONE]
28 location or ward ofspecimen collection
text [NONE] [NONE]
29 nosocomial infection flag text [NONE] [NONE]
30 specimen service text [NONE] [NONE]
31 specimen institution text [NONE] [NONE]
10 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
Host Sending Record Pattern
Request Information Record [Q Record] Pattern
Field Data Type Content L-1
1 record type ID (Q) char[1] Q Q
2 sequence number pos_int
3 starting range ID number(patients/specimen Id/instrument specimen ID)
text^text^text
{^text}{^text}
^Sample ID^Sequence No.^S.Rotor No. or
RackID^Position No.
4 ending range ID number [NONE]
5 universal test ID text ALL
6 nature of request timelimits
char[1] [NONE]
7 beginning request resultsdate and time
d_t [NONE]
8 ending request resultsdate and time
d_t [NONE]
9 requesting physicianname
text [NONE]
10 requesting physiciantelephone number
text [NONE]
11 user field NO.1 text [NONE]
12 user field NO.2 text [NONE]
13 request informationstatus codes
char[1 O O
Patient Information Record [P Record] Pattern
Field Data Type Content L-11
1 record type ID (P) char[1] P P
2 sequence number pos_int
3 practice assigned patientID
text [NONE
4 laboratory assignedpatient ID
text patient ID
5 patient ID No.3 text [NONE]
6 patient name text^text^text^text
[NONE]
7 mothers maiden name text [NONE]
8 date of birth d [NONE]
9 patient sex char[1] [NONE]
Test Order Message - Q/P/O 11
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
Patient Information Record [P Record] Pattern
Field Data Type Content L-11
10 patient race-ethnic origin text [NONE]
11 patient address text [NONE]
12 reserved field [NONE]
13 patient telephone number text [NONE]
14 attending physician ID text [NONE]
15 special field 1 text [NONE]
16 special field 2 text [NONE]
17 patient height (in cm) int{^unit}
[NONE]
18 patient weight in (kg) int{^unit}
[NONE]
19 patient’s known orsuspected diagnosis
text [NONE]
20 patient activemedications
text [NONE]
21 patient’s diet text [NONE]
22 practice field No.1 text [NONE]
23 practice field No.2 text [NONE]
24 admission and dischargedates
d{\d}
[NONE]
25 admission status char[2] [NONE]
26 location text [NONE]
27 nature of alternativediagnostic code andclassifiers
text [NONE]
28 alternative diagnosticcode and classification
text [NONE]
29 patient religion char[1] ortext
[NONE]
30 marital status char[1] [NONE]
31 isolation status char[3] ortext
[NONE
32 language status text [NONE]
33 hospital service text [NONE]
34 hospital institution text [NONE]
35 dosage category text [NONE]
12 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
Test Order Record [O Record] Pattern Pattern Pattern
Field Data Type Content L-21 L-22 L-23
1 record type ID (O) char[1] O O O O
2 sequence number pos_int
3 specimen ID text Sample ID *3) Sample ID *3) Sample ID *3)
4 instrument specimen ID text
{^text}{^text}{^text}{^text}{^text}
Sequence No.^S. Rotor No.
or RackID^Position No
^^
^containertype *6)
Sequence No.^S. Rotor No.
or RackID^Position No
^^
^containertype *6)
Sequence No.^S. Rotor No.
or RackID^Position No
^^
^containertype *6)
5 universal test ID Text^text^text^text
{\ }
^^
^Test No.^ Auto Dil.
Reference *8){\ }
[NONE] ^^
^Test No.^ Auto Dil.
Reference *8){\ }
6 priority char[1 S, R S: STATR: routine *2)
[NONE] S: STATR: routine *2)
7 request/ordered dateand time
d_t [NONE] [NONE] [NONE]
8 specimen collectiondate and time
d_t [NONE] [NONE] [NONE]
9 collection end time d_t [NONE] [NONE] [NONE]
10 collection volume (inml)
int{^unit}
[NONE] [NONE] [NONE]
11 collector ID text [NONE] [NONE] [NONE]
12 action code char[1]{\ }
C, N, Q, A *1) *4) *9) [NONE] *1) *4) *9)
13 danger code text [NONE] [NONE] [NONE]
14 relevant clinicalinformation
text [NONE] [NONE] [NONE]
15 date/time specimenreceived
d_t [NONE] [NONE] [NONE]
16 specimen descriptor(type & source)
text^text
[NONE] [NONE] [NONE]
17 ordering physician text [NONE] [NONE] [NONE]
18 physician's telephonenumber
text [NONE] [NONE] [NONE]
19 user field No.1 text [NONE] [NONE] [NONE]
20 user field No.2 text [NONE] [NONE] [NONE]
21 laboratory field No.1 text [NONE] [NONE] [NONE]
22 laboratory field No.1 text [NONE] [NONE] [NONE]
Test Order Message - Q/P/O 13
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
Test Order Record [O Record] Pattern Pattern Pattern
Field Data Type Content L-21 L-22 L-23
23 date/time resultsreported or lastmodified
d_t [NONE] [NONE] [NONE]
24 instrument change tocomputer system
text [NONE] [NONE] [NONE]
25 instrument section ID text [NONE] [NONE] [NONE]
26 report types char[1]{\...}
Q, Z, O Q Z *11) O
27 reserved field [NONE] [NONE] [NONE]
28 location or ward ofspecimen collection
text [NONE] [NONE] [NONE]
29 nosocomial infectionflag
text [NONE] [NONE] [NONE]
30 specimen service text [NONE] [NONE] [NONE]
31 specimen institution text [NONE] [NONE] [NONE]
*1) N: Instrument replaces the existing test order information to a new test order informationbelonging to this specimen.
A: Instrument adds this new test order information to the existing test order informationbelonging to this specimen.Note: The behavior here now is adopted to the original ASTM standard specifications.
C: Instrument deletes this specimen from data base.
*2) In case of a registration update (no new registration, sample is already in the data base) it isimpossible to change the priority.(Example: Host is not allowed to change a sample from Routine to STAT or vice versa.)
Registration already assigned
priority
priority
(sent from Host)
priority
R „Normal“
„Normal“ S IMPOSSIBLE !
Update [NONE] „Normal“
R IMPOSSIBLE !
„STAT“ S „STAT“
[NONE] „STAT“
R „Normal“
New S „STAT“
[NONE] „Normal“
*** Rack Sampler/Rack Conveyer System (Query Mode)----- Can not assign „STAT“ priority. ( = S)
*3) Maximum length of Sample ID is 22.It is impossible to set below strings in this field as Sample ID. The instrument treats thesestrings as „Stop Bar-code“.
14 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
If there is any additional character (unless space) the number will be treated as a sample ID.„000“„000 „ [ „000“ + some spaces ]„ 000“ [ some spaces + „000“ ]„ 000 „ [ some spaces + „000“ + some spaces ]
*4) In case of control samples following two characters are added „\Q“
*5) data carrier type: „CONTROL“, „SAMPLE“
*6) container type: „NORMAL“, „REDUCED“
*7) Auto Dilution FactorReference no. in reference table corresponding to auto-dilution ratio is set in this field.Auto dilution factor is set as follows.
sample’s type Protocol Tests Auto Dil. factor
Diluent Protocol or Quantitative Tests [NONE]**
PretreatmentProtocol
Qualitative Tests [NONE]
„Sample“ Other Protocol
auto-dilution ratiois not assigned
[NONE]
(NOT DiluentProtocol)
(NOT PretreatmentProtocol)
Quantitative Tests auto-dilution ratiois assigned
Reference No. inReference Table iscorresponding toassigned auto-
dilution ratio
Qualitative Tests [NONE]
„Control“ [NONE]
**That means that the auto dilution factor (=component 5 of the Universal test ID) isnow empty instead of 0.
*8) Auto Dilution FactorReference No. in reference table corresponding to auto-dilution ratio is set in this field.Instrument treats this auto dilution factor as follows (available or NOT available).
Sample’s type Protocol Tests Auto Dil. factor
Diluent Protocolor
Quantitative Tests NOT Available(Use Default auto-dilution ratio)
PretreatmentProtocol
Qualitative Tests NOT Available(Use Default auto-dilution ratio)
„Sample“ Other Protocol(NOT Diluent
Protocol)
Quantitative Tests Available(assigning auto-dilution ratio is possible)
(NOT PretreatmentProtocol)
Qualitative Tests NOT Available(assigning auto-dilution ratio is
impossible)
„Control“ NOT Available(Use Default auto-dilution ratio)
Test Order Message - Q/P/O 15
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
*9) In case of a registration update (no new registration, sample is already in the data base), it isimpossible to change the sample’s type. .(Example: Host is not allowed to change a sample from Sample to Control or vice versa.)
Registration already assigned
sample’s type
action code
(sent from Host)
sample’s type
N or A „Sample“
„Sample“ N\Q or A\Q IMPOSSIBLE !
Update [NONE] „Sample“
N or A IMPOSSIBLE !
„Control“ N\Q or A\Q „Control“
[NONE] „Control“
N or A\ „Sample“
New N\Q or A\Q IMPOSSIBLE !
[NONE] „Sample“
*10) Pre dilution factor0: This sample has NOT been pre diluted.1: This sample has been pre diluted.
*11) In this case, the instrument deletes this specimen from Data Base. ( V03-06 ~ )
[NOTES]
1. It is impossible to download test selections for a new „CONTROL“ sample (For aCONTROL sample it is only possible to download test selections when it is alreadyregistered.)
2. In case that the Host answers with a test selection for an outstanding query, it is necessaryto set the same Sequence No and Sample ID as the instrument asked for in the QueryMessage. If these data (Sequence No and Sample ID) are not same as asked for from theinstrument, the instrument ignores this message as the answer to the Query and continueswaiting for the correct answer from Host.
In case no Sample ID was sent from the instrument („specimen ID“ field is empty), the Hostcan set the correct Sample ID in the answering Test Order Message. Also in this case, it isnecessary to answer with the same Sequence No as the instrument queried for.
3. The instrument registers test selections from Host as follows.
Contents in „specimen ID“field and „Instrument
specimen ID“ field
(sent from Host)Search
Registration
Comment
Patt. SeqNo
SampleID
RotorNo/RackID &
Position No
Key Update New
1 O Seq No O
2 O Sample ID O O If the same sample ID is assigned for
16 Test Order Message - Q/P/O
Test Order Message - Q/P/O Elecsys® 2010 - Host Interface
Contents in „specimen ID“field and „Instrument
specimen ID“ field
(sent from Host)Search
Registration
Comment
Patt. SeqNo
SampleID
RotorNo/RackID &
Position No
Key Update New
different samples, the instrument onlyupdates the sample which was found firstaccording to the search algorithm.
3 O RotorNo/RackID &
Position No
O O In case of a new registration, the instrumentassigns „@SeqNo“ as Sample IDautomatically.
4 O O Seq No O If the Host does not answer to a query withthe same Sample ID, the instrument doesnot update this sample.However if „@SeqNo“ was assigned assample ID (sample without bar-code), theinstrument overwrites this sample ID withthe one which was sent from Host.
5 O O Sample IDor
RotorNo/RackID &
Position No
O O If the instrument can not find acorresponding sample in database bysearching with „Sample ID“ key, then theinstrument searches the sample again byusing „RotorNo/ RackID & Position No“ key.If „@SeqNo“ was assigned as sample ID(sample without bar-code), the instrumentoverwrites this sample ID with the one whichwas sent from Host.In case of a registration update (button onthe screen), the instrument overwritesRotorNo/ RackID & Position No to the onethat were sent earlier from Host.
6 O O Seq No O Instrument does not overwrite RotorNo/RackID & Position No to the one that weresent from Host.
7 O O O Seq No O If the host does not answer to a query withthe corresponding Sample ID, theinstrument does not update the sample.If „@SeqNo“ was assigned as sample ID(sample without bar-code), the instrumentoverwrites this sample ID with the one whichwas sent from Host.The instrument does not overwrite RotorNo/RackID & Position No to the one that weresent from Host.
O: Set or Available,[space]: Not Set or NOT Available
Test Order Message - Q/P/O 17
Elecsys® 2010 - Host Interface Test Order Message - Q/P/O
4. When the Host queries for test selections of a sample, the instrument searches the testselections of the sample by using „Search Key“. „Search Key“ is as follows.
Contents in „starting rangeID number“ field
(sent from Host) Search Comment
Patt. SeqNo
SampleID
RotorNo/RackID &
Position No
Key
1 O Seq No
2 O Sample ID
3 O RotorNo/RackID &
Position No
4 O O Seq No „Sample ID“ is ignored.
5 O O Sample ID „RotorNo/RackID & Position No“ are ignored.
6 O O Seq No „RotorNo/RackID & Position No“ are ignored.
7 O O O Seq No „Sample ID“ and „RotorNo/RackID & Position No“ areignored.
18 Result Message - Q/P/O/R/C/RR/RC
Result Message - Q/P/O/R/C/RR/RC Elecsys® 2010 - Host Interface
3. Result Message - Q/P/O/R/C/RR/RC
Purpose to use this message:
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Result (Final Result)(in case that Instrument answers positively)(Answer Information of Result that were requested fromHost.)
L-1(M-1)
I-1, I-11I-21, I-31
(M-2)
Query for Information of Result (Edited or Normal Result)(in case that Instrument answers positively)(Answer Information of Result that were requested fromHost.)
L-2(M-1)
I-1, I-11I-21, I-31
(M-2)
Query for Information of Result(in case that Instrument answers negatively)(Answer negatively in case that all results of this specimenhave not been calculated yet.)
L-1(M-1)
I-1, I-12(M-3)
Event: The instrument receives a Result Message Query.à The instrument answers to the query.Host queries for result information for one sample.(or for some samples)The instrument always answers with 1 message/1 sample even if Host queries forresult information for some samples using one message.
Instrument à Host Upload
Purpose Host InstrumentUpload after all results of this specimen are calculatedUpload after pressing <DOC> key on Results screen.(Upload Information of Result of this specimen.)
I-1, I-13I-22, I-31I-41, I-51
(M-4)Upload after Instrument Power on.(Upload Information of Result of this specimen.)
I-1, I-13I-22, I-31
(M-2)L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: a) Automatic output of all the test results is generated (including ‘No Value”results)à The instrument uploads a Result Message for this sample.Normal samples and Control samples are applied.This is done in automatic upload mode.
b) Pressing <DOC> key on the Results Screen.à The instrument uploads a Result Message for the sample that is displayed on thecurrent screen.This is done even if all the test results are not yet generated.
c) Turn on the power switch of the instrument.à The instrument uploads messages for undocumented results of samples.The instrument uploads 1 message/1 sample.
Result Message - Q/P/O/R/C/RR/RC 19
Elecsys® 2010 - Host Interface Result Message - Q/P/O/R/C/RR/RC
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 Q0 L
<< more than one query may occur >>
Pattern 2 (M-2)Level Record
0 H1 P2 O3 R* C3 R } max. 60 results* C
: :: :
0 L<< 1 Message / 1 Sample >>
Trace Example:
Instrument sends results for control sample ‘000004’: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2P|1<CR><ETX>3F<CR><LF> <ACK> <STX>3O|1|000004|40^0^4^^SAMPLE^REDUCED|ALL|R|
19970425120322|||||X||||||||||||||O|||||<CR><ETX>6C<CR><LF> <ACK> <STX>4R|1|^^^10^^0|0.016|uIU/ml|0.230^3.80|L||F|||
19970425120351|19970425122213|<CR><ETX>D7<CR><LF> <ACK> <STX>5C|1|I|48^Below normal(expected)range|I<CR><ETX>57<CR><LF> <ACK> <STX>6R|2|^^^50^^0|1.17|ng/ml|0.846^2.02|||F|||
19970425120557|19970425122419|<CR><ETX>3CDE<CR><LF> <ACK> <STX>7R|3|^^^20^^0|4.41|ug/dl|5.13^13.52|L||F|||
19970425120433|19970425122255|<CR><ETX>7A<CR><LF> <ACK> <STX>0C|1|I|48^Below normal(expected)range|I<CR><ETX>52<CR><LF> <ACK> <STX>1R|4|^^^30^^0|0.091|ng/dl|1.01^1.79|L||F|||
19970425120515|19970425122337|<CR><ETX>71<CR><LF> <ACK> <STX>3C|1|I|48^Below normal(expected)range|I<CR><ETX>55<CR><LF> <ACK> <STX>2R|7|^^^400^^0|-1^0.001|COI|^|N||F|||19971215150922|
19971215152744|<CR><ETX>D9<CR><LF> <ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
20 Result Message - Q/P/O/R/C/RR/RC
Result Message - Q/P/O/R/C/RR/RC Elecsys® 2010 - Host Interface
Pattern 3 (M-3)Level Record
0 H1 P2 O0 L
<< 1 Message / 1 Sample >>
Pattern 4 (M-4)Level Record
0 H1 P2 O3 R* C4 M-RR4 M-RC3 R } max. 60 results* C4 M-RR4 M-RC
: :: :
0 L<< 1 Message / 1 Sample >>
Trace Example:
Instrument sends results for control ‘PC U1’: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2P|1<CR><ETX>3F<CR><LF> <ACK> <STX>3O|1|PC U1|37^0^1^^CONTROL^REDUCED|ALL|R|
19970425115705|||||X\Q||||||||||||||O|||||<CR><ETX>99<CR><LF> <ACK> <STX>4R|1|^^^10^^0|1.99|uIU/ml|1.69^2.43|||F|||
19970425115733|19970425121555|<CR><ETX>50<CR><LF> <ACK> <STX>5M|1|RR^BM^E1^1||14242\13924\14154\13391\12613\12836\
13062\13757\13053\13469\13147\12980\13312\13396\13501\13563\13639\13351\13155\12979\12857\13264\13391\13492\13343\13791\13345\13346\13752\12821\12993\13695\13696\14166\13736\13114\12690\12<ETB>9C<CR><LF>
<ACK> <STX>6517\13395\13264\12968\13375\13029\13117\13465\13267\
13507\12716\13380\13316\13145\12736\12679\12555\13334\13347\13635\13291\13372\13694\13063\13278\13166\13523\12749\12977\13293\12872\12550\12815\12657\12718\13367\13141\13495\13466\13869\13<ETB>06<CR><LF>
<ACK> <STX>7520\13464\13565\22636\29929\33176\34453\34874\34951\
34859\34748\34606\34414\34273\34127\33971\33799\33657\33534\33408\33303\33197\33114\32970\32860\32637\32585\32472\32434\32
Result Message - Q/P/O/R/C/RR/RC 21
Elecsys® 2010 - Host Interface Result Message - Q/P/O/R/C/RR/RC
285\32210\32150\32101\31976\31815\31758\31693\31590\31552\31396\31<ETB>3D<CR><LF>
<ACK> <STX>0296\31256\31194\31103\31156\31026\30899\30894\30786\
30742\30696\30591\30502\30582\30470\30363\30369\30350\30191\30183\30096\30065\30026\29933\29920\29881\29832\29783\29822\29769\29712\29527\29492\29499\29531\29487\29444\29403\29317\29248\29<ETB>5E<CR><LF>
<ACK> <STX>1190\29177\29196\29128\29081\29051\28995\29032\29019\
29002\29034\28952\28864\28814\28819\28791\28825\28762\28709\28655\28563\28528\28514\28488\28344\28360\28345\28352\28260\28343\28275\28286\28276\28344\28308\28255\28157\28067\28031\28130\28<ETB>AB<CR><LF>
<ACK> <STX>2156\28122\28175|28083<CR><ETX>1D<CR><LF> <ACK> <STX>3M|1|RC^BM^E1^1|10|TSH|190524|487|1|190599||<CR><ETX>18<CR><LF> <ACK> <STX>4R|2|^^^50^^0|1.45|ng/ml|1.11^1.70|||F|||
19970425115939|19970425121801|<CR><ETX>02<CR><LF> <ACK> <STX>5M|1|RR^BM^E1^1||13104\13200\13457\13463\13195\13161\..... .. .. <STX>2321\35302\35259|196679<CR><ETX>57<CR><LF> <ACK> <STX>3M|1|RC^BM^E1^1|50|T3|192955|240|1|190599||<CR><ETX>B1<CR><LF> <ACK> <STX>4R|3|^^^20^^0|10.49|ug/dl|8.10^12.41|||F|||
19970425115815|19970425121637|<CR><ETX>68<CR><LF> <ACK> <STX>5M|1|RR^BM^E1^1||14010\14245\13863\13399\13671\13047\..... .. ..<STX>2592\38558\38544|457109<CR><ETX>65<CR><LF> <ACK> <STX>3M|1|RC^BM^E1^1|20|T4|190764|257|1|190599||<CR><ETX>B3<CR><LF> <ACK> <STX>4R|4|^^^30^^0|1.52|ng/dl|1.36^1.84|||F|||
19970425115857|19970425121719|<CR><ETX>0A<CR><LF> <ACK> <STX>5M|1|RR^BM^E1^1||13277\13330\13316\13684\13247\13300\..... .. .. <STX>2033\28004\28064|28403<CR><ETX>0F<CR><LF> <ACK> <STX>3M|1|RC^BM^E1^1|30|FT4|192882|1228|1|190599||<CR><ETX>2C<CR><LF> <ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Patient Information Record [P Record] Pattern
Field Data Type Content I-1
1 record type ID (P) char[1] P P
2 sequence number pos_int
22 Result Message - Q/P/O/R/C/RR/RC
Result Message - Q/P/O/R/C/RR/RC Elecsys® 2010 - Host Interface
Patient Information Record [P Record] Pattern
Field Data Type Content I-1
3 practice assigned patientID
text [NONE]
4 laboratory assignedpatient ID
text patient ID
5 patient ID No.3 text [NONE]
6 patient name text^text^text^text
[NONE]
7 mothers maiden name text [NONE]
8 date of birth d [NONE]
9 patient sex char[1] [NONE]
10 patient race-ethnic origin text [NONE]
11 patient address text [NONE]
12 reserved field [NONE]
13 patient telephone number text [NONE]
14 attending physician ID text [NONE]
15 special field 1 text [NONE]
16 special field 2 text [NONE]
17 patient height (in cm) int{^unit}
[NONE]
18 patient weight in (kg) int{^unit}
[NONE]
19 patient's known orsuspected diagnosis
text [NONE]
20 patient activemedications
text [NONE]
21 patient's diet text [NONE]
22 practice field No.1 text [NONE]
23 practice field No.2 text [NONE]
24 admission and dischargedates
d{\d}
[NONE]
25 admission status char[2] [NONE]
26 location text [NONE]
27 nature of alternativediagnostic code andclassifiers
text [NONE]
28 alternative diagnosticcode and classification
text [NONE]
29 patient religion char[1] ortext
[NONE]
Result Message - Q/P/O/R/C/RR/RC 23
Elecsys® 2010 - Host Interface Result Message - Q/P/O/R/C/RR/RC
Patient Information Record [P Record] Pattern
Field Data Type Content I-1
30 marital status char[1] [NONE]
31 isolation status char[3] ortext
[NONE
32 language status text [NONE]
33 hospital service text [NONE]
34 hospital institution text [NONE]
35 dosage category text [NONE]
Test Order Record [O Record] Pattern Pattern Pattern
Field Data Type Content I-11 I-12 I-13
1 record type ID (O) char[1] O O O O
2 sequence number pos_int
3 specimen ID text Sample ID Sample ID Sample ID
4 instrument specimen ID text
{^text}{^text}{^text}{^text}
Sequence No.^S. Rotor No.
or RackID^Position No
^^data carrier
type *6^container
type *7)
[NONE] Sequence No.^S. Rotor No.
or RackID^Position No
^^data carrier
type *6^container
type *7)
5 universal test ID text^text^text^text
{^text}{^text}{^text}{\ }
ALL ALL ALL
6 priority char[1 S, R S: STATR: routine
[NONE] S: STATR: routine
7 request/ordered dateand time
d_t ordered date& time
[NONE] ordered date& time
8 specimen collectiondate and time
d_t [NONE] [NONE] [NONE]
9 collection end time d_t [NONE] [NONE] [NONE]
10 collection volume (inml)
int{^unit}
[NONE] [NONE] [NONE]
11 collector ID text [NONE] [NONE] [NONE]
12 action code char[1] X, Q X *3) X *3) X *3)
13 danger code text [NONE] [NONE] [NONE]
14 relevant clinicalinformation
text [NONE] [NONE] [NONE]
24 Result Message - Q/P/O/R/C/RR/RC
Result Message - Q/P/O/R/C/RR/RC Elecsys® 2010 - Host Interface
Test Order Record [O Record] Pattern Pattern Pattern
Field Data Type Content I-11 I-12 I-13
15 date/time specimenreceived
d_t [NONE] [NONE] [NONE]
16 specimen descriptor(type & source)
text^text
[NONE] [NONE] [NONE]
17 ordering physician text [NONE] [NONE] [NONE]
18 physician's telephonenumber
text [NONE] [NONE] [NONE]
19 user field No.1 text [NONE] [NONE] [NONE]
20 user field No.2 text [NONE] [NONE] [NONE]
21 laboratory field No.1 text [NONE] [NONE] [NONE]
22 laboratory field No.1 text [NONE] [NONE] [NONE]
23 date/time resultsreported or lastmodified
d_t [NONE] [NONE] [NONE]
24 instrument change tocomputer system
text [NONE] [NONE] [NONE]
25 instrument section ID text [NONE] [NONE] [NONE]
26 report types char[1]{\...}
Q, I, O Q I O
27 reserved field [NONE] [NONE] [NONE]
28 location or ward ofspecimen collection
text [NONE] [NONE] [NONE]
29 nosocomial infectionflag
text [NONE] [NONE] [NONE]
30 specimen service text [NONE] [NONE] [NONE]
31 specimen institution text [NONE] [NONE] [NONE]
Result Record [R Record] Pattern Pattern
Field Data Type Content I-21 I-22
1 record type ID (R) char[1] R R R
2 sequence number pos_int
3 universal test ID text^text^text^text
{^text}{^text}
^^^Test No.^Auto Dil.
Reference *9)^PredilutionFactor *11)
^^^Test No.^Auto Dil.
Reference *9)^PredilutionFactor *11)
4 data or measurementvalue
text{^...}
results *5) results *5)
5 units text units units
Result Message - Q/P/O/R/C/RR/RC 25
Elecsys® 2010 - Host Interface Result Message - Q/P/O/R/C/RR/RC
Result Record [R Record] Pattern Pattern
Field Data Type Content I-21 I-22
6 reference ranges text^text{\...}
Lower Limit^Upper Limit
*10)
Lower Limit^Upper Limit
*10)
7 result abnormal flags char[1] orchar[2]
*8) *8)
8 nature of abnormalitytesting
char[1]{\...}
[NONE] [NONE]
9 result status char[1]{\...}
F, X, Q, V,Y, +
*1) *2)
10 date of change ininstrument normativevalues or units
d [NONE] [NONE]
11 operator identification text [NONE] [NONE]
12 date/time test started d_t sampling time sampling time
13 date/time test completed d_t ready time ready time
14 instrument identification text [NONE] [NONE]
Comment Record [C Record] Pattern
Field Data Type Content I-31
1 record type ID (C) char[1] C C
2 sequence number pos_int
3 comment source char[1] I I
4 comment text text{^...}
Instrumentresult flag
(integer 1...50)^Comment forabnormal flags
5 comment type char[1] G I
Raw Result Record [M-RR Record] Pattern
Field Data Type Content I-41
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(RR)
char[2]^text^text^text
RR^BM^E1^1 RR^BM^E1^1
4 starting value int [NONE]
5 value list int{\...}
value list
6 effective signal text effective signal
26 Result Message - Q/P/O/R/C/RR/RC
Result Message - Q/P/O/R/C/RR/RC Elecsys® 2010 - Host Interface
Result Context Record [M-RC Record] Pattern
Field Data Type Content I-51
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(RC)
char[2]^text^text^text
RC^BM^E1^1 RC^BM^E1^1
4 BM test number text Test No.
5 reagent identifier text Test name
6 reagent lot number int lot number
7 reagent pack number int Rack Pack No.
8 control identifier text control No.
9 control lot number int control lotnumber
10 control pack number int [NONE]
11 Calibration status char[1] [NONE]
Host Sending Record Pattern
Request Information Record [Q Record] Pattern Pattern
Field Data Type Content L-1 L-1
1 record type ID (Q) char[1] Q Q Q
2 sequence number pos_int
3 starting range ID number(patients/specimen Id/instrument specimen ID)
text^text^text
^Sample ID^Sequence #
^Sample ID^Sequence #
4 ending range ID number text^text^text
[NONE] [NONE]
5 universal test ID text ALL ALL
6 nature of request timelimits
char[1] R [NONE] [NONE]
7 beginning request resultsdate and time
d_t [NONE] [NONE]
8 ending request resultsdate and time
d_t [NONE] [NONE]
9 requesting physicianname
text [NONE] [NONE]
10 requesting physiciantelephone number
text [NONE] [NONE]
11 Specimen Type char[1] N, Q *4) *4)
12 user field NO.2 text [NONE] [NONE]
Result Message - Q/P/O/R/C/RR/RC 27
Elecsys® 2010 - Host Interface Result Message - Q/P/O/R/C/RR/RC
Request Information Record [Q Record] Pattern Pattern
Field Data Type Content L-1 L-1
13 request informationstatus codes
char[1 F, N F N
*1) result status
No Value SystemBlocked
UserBlocked
UserReleased
Others
'F'request
set 'Q\X' set 'Q\Y' set 'Q\+' set 'Q\V' set 'Q\F'
'N'request
set 'Q\Y' set 'Q\+' set 'Q\V' set 'Q\F'
*2) result status:X : No ValueY : System Blocked+ : User BlockedV : User ReleasedF : Others
*3) In case of control samples the following two characters are added: ‘\Q’
*4) In case of normal samples set 'N'.In case of control samples set 'Q'.
*5) In case that a result is above upper limit (measuring range) the followingcharacter is added in the front of the result: ‘>’In case that a result is under lower limit (measuring range) the followingcharacter is added in the front of the result: ‘<’
For quantitative tests : a value corresponds to the analyte concentrationFor qualitative tests: data1 ^data2
data1 = ‘1’ : corresponds to „positive“ ‘0’ : corresponds to „border line“‘-1’ : corresponds to „negative“
data2 = Cut-off indexWhen „result status“ = ‘X’ (No Value), no data is set in this field ([NONE]).
*6) data carrier type : „CONTROL“, „SAMPLE“
*7) container type : „NORMAL“,“REDUCED“
28 Result Message - Q/P/O/R/C/RR/RC
Result Message - Q/P/O/R/C/RR/RC Elecsys® 2010 - Host Interface
*8) result abnormal flagsL : Below normal range (data alarm No = 48)H : Above normal range (data alarm No = 49)< : Below measuring range (data alarm No = 50)> : Above measuring range (data alarm No = 51)A : System Blocked (data alarm No = 1-11,17-22,24,30-37,45,47)N : All other flags
*9) Auto Dilution FactorReference no. in reference table corresponding to auto-dilution ratiois set in this field.Auto dilution factor is set as follows.
Sample’s type Protocol Tests Auto Dil. factor
Diluent Protocol or Quantitative Tests [NONE]**
PretreatmentProtocol
Qualitative Tests [NONE]
„Sample“ Other Protocol
auto-dilution ratiois not assigned
[NONE]
(NOT DiluentProtocol)
(NOT PretreatmentProtocol)
Quantitative Tests auto-dilution ratiois assigned
Reference No. inReference Table iscorresponding toassigned auto-
dilution ratio
Qualitative Tests [NONE]
„Control“ [NONE]
**That means that the auto dilution factor (=component 5 of the Universal test ID) isnow empty instead of 0.
*10) reference rangesreference ranges are set as follows.
Normal Range Check
Sample’s Type Tests Not Available Available
„Sample“
Quantitative Tests [NONE] Lower Limit
^Upper Limit
Qualitative Tests [NONE] [NONE]
„BM Control“ Lower Limit
^Upper Limit
Lower Limit
^Upper Limit
„Foreign Control“ Lower Limit
^Upper Limit
Lower Limit
^Upper Limit
*11) Pre dilution factor0: This sample has NOT been pre diluted.1: This sample has been pre diluted.
Result Message - Q/P/O/R/C/RR/RC 29
Elecsys® 2010 - Host Interface Result Message - Q/P/O/R/C/RR/RC
[NOTES]
1. When Host queries for the results of a sample the instrument searches the results of thesample by using „Search Key“.„Search Key“ is as follows.
Contents in „starting rangeID number“ field
(sent from Host) Search Comment
Patt. SeqNo
SampleID
RotorNo/RackID &
Position No
Key
1 O Seq No
2 O Sample ID
3 O - IMPOSSIBLE ! (can not search by this pattern)
4 O O Seq No „Sample ID“ is ignored.
5 O O Sample ID „RotorNo/RackID & Position No“ are ignored.
6 O O Seq No „RotorNo/RackID & Position No“ are ignored.
7 O O O Seq No „Sample ID“ and „RotorNo/RackID & PositionNo“ are ignored.
30 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
4. Substance Data Message - XT/DP
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload after completion of scanning of the reagent rotor(Test Reagent, Pretreatment, Blank Set)(Upload Information of rack packs that are on the reagentrotor. (18 positions))
I-1(M-4)
Upload after completion of scanning of the reagent rotor(There is no rack pack on this position.)(Upload Information of rack packs that are on the reagentrotor. (18 positions))
I-2(M-4)
Upload after completion of scanning of the reagent rotor(Diluent)(Upload Information of rack packs that are on the reagentrotor. (18 positions))
I-1, I-11(M-5)
Upload after completion of checking of liquid level of SystemReagent(Upload Information of System Reagent rack packs. (4positions))
I-3(M-1)
Upload after completion of scanning of Tip and Vesselmagazines(Upload Information of Tips and Vessels on the magazines.)
I-5(M-1)
Upload constantly (Instrument à Host)Upload when actual level becomes zero (Instrument à Host)(Upload Information of rack packs)
I-1I-3I-4I-5
(M-1)Upload when going to Stand-by status(Upload inventory Information with actual numbers.)
I-1I-3I-4I-5
I-11(M-3)
Upload after manual input of rack packs Information(Upload Information of rack packs that are on the reagentrotor.)
I-1(M-1)
Upload after occurring ‘Solid Waste Full’ Alarm(Upload Information of solid waste.)
I-4(M-1)
Upload after reset of Solid Waste Counter(Upload Information of solid waste.)
I-4(M-1)
Event: a) Scan of reagent rotorà The instrument uploads Substance Data Messages for the rack packs on thereagent rotor.The following rack packs are applied.
- Test Reagent, Pretreatment, Diluent, Blank SetThe instrument uploads 1 message/18 positions.
b) Liquid level check of the system reagents (AB/CC).à The instrument uploads Substance Data Messages for system reagent rack packs.
Substance Data Message - XT/DP 31
Elecsys® 2010 - Host Interface Substance Data Message - XT/DP
The instrument uploads 4 messages for AB1, CC1, AB2 and CC2.
c) Scan tip and vessel magazinesà The instrument uploads Substance Data Messages for tips and vessels.The instrument uploads 1 messages for tips and vessels.
d) Upload constantly during an operationà The instrument uploads Substance Data Messages constantly with a certain stepdistance.Following substances are applied.
- Test Reagent, Pretreatment, Diluent, Blank Set, System Reagent (AB/CC)- Tips & Vessel, Distilled Water, Liquid Waste
The instrument uploads 1 message/1 substance.
e) Upload if liquid level is zero during an operationà The instrument uploads Substance Data Messages of rack packs for which theliquid level became zero.Following Rack Packs are applied.
- Test Reagent, Pretreatment, Diluent, Blank Set, System Reagent (AB/CC)The instrument uploads 1 message/1 Rack Pack.
f) Going to Stand-by status after executing something functions.à The instrument uploads Substance Data MessagesFollowing substances are applied.
- Test Reagent, Pretreatment, Diluent, Blank Set, System Reagent (AB/CC)- Tip & Vessel, Distilled Water, Liquid Waste
g) Manual input of a rack pack no. on inventory screenà The instrument uploads information for the rack pack.
h) Occurrence of “Solid Waste Full” alarmà The instrument uploads information for solid waste.
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of rack packs that are on the reagentrotor(Test Reagent, Pretreatment, Blank Set)(Answer Information of rack pack on the reagent rotor positionthat was requested from Host. Answer parameter Informationtoo for Diluent rack pack.)
L-2(M-1)
I-1(M-1)
Query for Information of rack packs that are on the reagentrotor (Diluent)(Answer Information of rack pack on the reagent rotor positionthat was requested from Host. Answer parameter Informationtoo for Diluent rack pack.)
L-2(M-1)
I-1, I-11(M-2)
Query for Information of rack packs that are on the reagentrotor(There is no rack pack on this position.)(Answer Information of rack pack on the reagent rotor positionthat was requested from Host. Answer parameter Informationtoo for Diluent rack pack.)
L-2(M-1)
I-2(M-1)
Query for Information of System reagent rack packs(Answer actual level of each substances at this time.)
L-3(M-1)
I-3(M-1)
Query for Information of Distilled Water, Liquid and SolidWaste
L-4(M-1)
I-4(M-1)
32 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
(Answer actual level of each substances at this time.)Query for Information of Tips and Vessels(Answer actual level of each substances at this time.)
L-5(M-1)
I-5(M-1)
Event: The instrument receives a Substance Message Queryà The instrument answers to the HostHost queries for Substance Data information for a specific substance (or for somesubstances).The instrument always answers with 1 message/1 substance.
Host Download
Purpose Host InstrumentDownload Information of rack packs (actual level)(Test Reagent, Pretreatment, Blank Set, Diluent) *1)
L-1(M-1)
L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
*1) Download only actual level of each substance.Downloading actual level of following substances not available.system reagent ...
(actual level is set by checking liquid level)Distilled Water, Liquid Waste ...
(actual level is set by detector)Tips, Vessels ...
(actual level is set by mechanical scanning)
Event: The instrument receives a Substance Data Message.à The instrument overwrites information of the specific substance data.Host downloads 1 message/1 substance (or 1 message/some substances).If the Substance Data that was sent from Host is for a substance that the instrumentdo not know, then this data is not valid (will be ignored).
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-XT0 L
Trace Example:
Host queries for Substance Data at container position 12: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|XTQ^BM^E1^1|||||||||||||||||||12|<CR><ETX>18<CR><LF> <ACK> <STX>3L|1|I<CR><ETX>01<CR><LF> <ACK> <EOT>
Substance Data Message - XT/DP 33
Elecsys® 2010 - Host Interface Substance Data Message - XT/DP
Instrument sends back Substance Data for position 12: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|TR|30|FT4|||RP|SCE1|200|192882|
1228|19970423094333|||19970800|196||RD||12|A<CR><ETX>DB<CR><LF>
<ACK> <STX>3L|1|F<CR><ETX>FE<CR><LF> <ACK> <EOT>
Host queries for Substance Data at container position 2: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|XTQ^BM^E1^1|||||||||||||||||||2|<CR><ETX>E7<CR><LF> <ACK> <STX>3L|1|I<CR><ETX>01<CR><LF> <ACK> <EOT>
Instrument sends back Substance Data for position 2: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|TR|170|HCGSTAT|||RP|SCE1|100|170459|
1295|19971211162313|||19961200|96||RD||2|A<CR><ETX>DD<CR><LF> <ACK> <STX>3L|1|F<CR><ETX>FE<CR><LF> <ACK> <EOT>
Instruments actualizes Substance Data: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|AB|ABCC1|||||||0||||||100|||||A
<CR><ETX>23<CR><LF> <ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|CC|ABCC1|||||||0||||||100|||||A
<CR><ETX>26<CR><LF> <ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF>
34 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
<ACK> <STX>2M|1|XT^BM^E1^1|TP||||||||||||||303||TR|||A
<CR><ETX>85<CR><LF> <ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|VS||||||||||||||150||VR|||A
<CR><ETX>8C<CR><LF> <ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
Pattern 2 (M-2)Level Record
0 H1 M-XT2 M-DP
: } max. 40 records :
0 L
Pattern 3 (M-3)Level Record
0 H1 M-XT Rack pack Information of rack packs on a reagent rotor
...: *) in case of a diluent rack pack the parameter informationrecord (M-DP) is added.
...:0 L0 H1 M-XT à Tip information1 M-XT à Vessel information0 L0 H1 M-XT à system reagent information (Pos 1)1 M-XT1 M-XT à system reagent information (Pos 2)1 M-XT0 L0 H1 M-XT à Distilled Water information1 M-XT à Liquid Waste information1 M-XT à Solid Waste information0 L
Substance Data Message - XT/DP 35
Elecsys® 2010 - Host Interface Substance Data Message - XT/DP
Trace Example:
Instrument sends update for Substance Data <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|||||||||||||||||RD||1|R<CR><ETX>7D<CR><LF> <ACK> <STX>3M|2|XT^BM^E1^1|TR|40|T-UP|||RP|SCE1|200|190543|
17|19970424161410|||19970400|169||RD||2|A<CR><ETX>8C<CR><LF> <ACK> <STX>4M|3|XT^BM^E1^1|TR|310|AFP|||RP|SCE1|100|189872|
178|19970424161410|||19970400|96||RD||3|A<CR><ETX>83<CR><LF> <ACK> <STX>5M|4|XT^BM^E1^1|TR|10|TSH|||RP|SCE1|200|190524|
487|19970424161411|||19970800|182||RD||4|A<CR><ETX>92<CR><LF> <ACK> <STX>6M|5|XT^BM^E1^1|||||||||||||||||RD||5|R<CR><ETX>89<CR><LF> <ACK> <STX>7M|6|XT^BM^E1^1|TR|320|PSA|||RP|SCE1|100|192831|
446|19970424161411|||19970500|89||RD||6|A<CR><ETX>91<CR><LF> <ACK> <STX>0M|7|XT^BM^E1^1|TR|170|HCGSTAT|||RP|SCE1|100|190554|
100|19970424161411|||19970500|75||RD||7|A<CR><ETX>A7<CR><LF> <ACK> <STX>1M|8|XT^BM^E1^1|TR|210|CKMBSTAT|||RP|SCE1|100|190557|
75|19970424161411|||19970500|54||RD||8|A<CR><ETX>CB<CR><LF> <ACK> <STX>2M|9|XT^BM^E1^1|TR|60|FT3|||RP|SCE1|200|190884|
158|19970424161411|||19970400|182||RD||9|A<CR><ETX>7C<CR><LF> <ACK> <STX>3M|10|XT^BM^E1^1|||||||||||||||||RD||10|R<CR><ETX>DE<CR><LF> <ACK> <STX>4M|11|XT^BM^E1^1|||||||||||||||||RD||11|R<CR><ETX>E1<CR><LF> <ACK> <STX>5M|12|XT^BM^E1^1|TR|50|T3|||RP|SCE1|200|192955|
240|19970424161411|||19970700|184||RD||12|A<CR><ETX>8A<CR><LF>
<ACK> <STX>6M|13|XT^BM^E1^1|TR|20|T4|||RP|SCE1|200|190764|
257|19970424161411|||19971100|184||RD||13|A<CR><ETX>8A<CR><LF>
<ACK> <STX>7M|14|XT^BM^E1^1|TR|30|FT4|||RP|SCE1|200|192882|
1228|19970424161411|||19970800|184||RD||14|A<CR><ETX>0C<CR><LF>
<ACK> <STX>0M|15|XT^BM^E1^1|TR|200|TNTSTAT|||RP|SCE1|100|190560|
78|19970424161411|||19970500|18||RD||15|A<CR><ETX>FB<CR><LF> <ACK> <STX>1M|16|XT^BM^E1^1|TR|300|CEA|||RP|SCE1|100|192838|
577|19970424161411|||19970500|96||RD||16|A<CR><ETX>DA<CR><LF> <ACK> <STX>2M|17|XT^BM^E1^1|||||||||||||||||RD||17|R<CR><ETX>EB<CR><LF> <ACK> <STX>3M|18|XT^BM^E1^1|DR|1|DilUni2|||RP|SCE1|40000|7878787|
777|19970424161411|||19991200|40000||RD||18|A<CR><ETX>6B<CR><LF>
<ACK> <STX>4M|1|DP^BM^E1^1|10||<CR><ETX>73<CR><LF> <ACK> <STX>5M|2|DP^BM^E1^1|20||<CR><ETX>76<CR><LF> <ACK> <STX>6M|3|DP^BM^E1^1|30||<CR><ETX>79<CR><LF> <ACK>
36 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
<STX>7M|4|DP^BM^E1^1|40||<CR><ETX>7C<CR><LF> <ACK> <STX>0M|5|DP^BM^E1^1|50||<CR><ETX>77<CR><LF> <ACK> <STX>1M|6|DP^BM^E1^1|60||<CR><ETX>7A<CR><LF> <ACK> <STX>2M|7|DP^BM^E1^1|100||<CR><ETX>A7<CR><LF> <ACK> <STX>3M|8|DP^BM^E1^1|110||<CR><ETX>AA<CR><LF> <ACK> <STX>4M|9|DP^BM^E1^1|120||<CR><ETX>AD<CR><LF> <ACK> <STX>5M|10|DP^BM^E1^1|130||<CR><ETX>D7<CR><LF> <ACK> <STX>6M|11|DP^BM^E1^1|140||<CR><ETX>DA<CR><LF> <ACK> <STX>7M|12|DP^BM^E1^1|150||<CR><ETX>DD<CR><LF> <ACK> <STX>0M|13|DP^BM^E1^1|160||<CR><ETX>D8<CR><LF> <ACK> <STX>1M|14|DP^BM^E1^1|170||<CR><ETX>DB<CR><LF> <ACK> <STX>2M|15|DP^BM^E1^1|180||<CR><ETX>DE<CR><LF> <ACK> <STX>3M|16|DP^BM^E1^1|200||<CR><ETX>D9<CR><LF> <ACK> <STX>4M|17|DP^BM^E1^1|210||<CR><ETX>DC<CR><LF> <ACK> <STX>5M|18|DP^BM^E1^1|220||<CR><ETX>DF<CR><LF> <ACK> <STX>6M|19|DP^BM^E1^1|230||<CR><ETX>E2<CR><LF> <ACK> <STX>7M|20|DP^BM^E1^1|300||<CR><ETX>D9<CR><LF> <ACK> <STX>0M|21|DP^BM^E1^1|310||<CR><ETX>D4<CR><LF> <ACK> <STX>1M|22|DP^BM^E1^1|320||<CR><ETX>D7<CR><LF> <ACK> <STX>2M|23|DP^BM^E1^1|330||<CR><ETX>DA<CR><LF> <ACK> <STX>3M|24|DP^BM^E1^1|340||<CR><ETX>DD<CR><LF> <ACK> <STX>4M|25|DP^BM^E1^1|350||<CR><ETX>E0<CR><LF> <ACK> <STX>5M|26|DP^BM^E1^1|360||<CR><ETX>E3<CR><LF> <ACK> <STX>6M|27|DP^BM^E1^1|370||<CR><ETX>E6<CR><LF> <ACK> <STX>7M|28|DP^BM^E1^1|380||<CR><ETX>E9<CR><LF> <ACK> <STX>0M|29|DP^BM^E1^1|390||<CR><ETX>E4<CR><LF> <ACK> <STX>1M|30|DP^BM^E1^1|400||<CR><ETX>D5<CR><LF> <ACK> <STX>2M|31|DP^BM^E1^1|410||<CR><ETX>D8<CR><LF> <ACK> <STX>3M|32|DP^BM^E1^1|420||<CR><ETX>DB<CR><LF> <ACK> <STX>4M|33|DP^BM^E1^1|430||<CR><ETX>DE<CR><LF> <ACK> <STX>5M|34|DP^BM^E1^1|440||<CR><ETX>E1<CR><LF> <ACK> <STX>6M|35|DP^BM^E1^1|450||<CR><ETX>E4<CR><LF> <ACK> <STX>7M|36|DP^BM^E1^1|460||<CR><ETX>E7<CR><LF> <ACK>
Substance Data Message - XT/DP 37
Elecsys® 2010 - Host Interface Substance Data Message - XT/DP
<STX>0M|37|DP^BM^E1^1|470||<CR><ETX>E2<CR><LF> <ACK> <STX>1M|38|DP^BM^E1^1|480||<CR><ETX>E5<CR><LF> <ACK> <STX>2M|39|DP^BM^E1^1|490||<CR><ETX>E8<CR><LF> <ACK> <STX>3M|40|DP^BM^E1^1|500||<CR><ETX>D9<CR><LF> <ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|TP||||||||||||||254||TR|||A<CR><ETX>8A<CR><LF> <ACK> <STX>3M|2|XT^BM^E1^1|VS||||||||||||||126||VR|||A<CR><ETX>91<CR><LF> <ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|AB|ABCC1|||||||||||||110|||||A
<CR><ETX>F4<CR><LF> <ACK> <STX>3M|2|XT^BM^E1^1|CC|ABCC1|||||||||||||107|||||A
<CR><ETX>FF<CR><LF> <ACK> <STX>4M|3|XT^BM^E1^1|AB|ABCC2|||||||||||||19|||||A<CR><ETX>D1<CR><LF> <ACK> <STX>5M|4|XT^BM^E1^1|CC|ABCC2|||||||||||||20|||||A<CR><ETX>CE<CR><LF> <ACK> <STX>6L|1<CR><ETX>3F<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|DW||||||||||||||0|||||A<CR><ETX>70<CR><LF> <ACK> <STX>3M|2|XT^BM^E1^1|LW||||||||||||||0|||||A<CR><ETX>7A<CR><LF> <ACK> <STX>4M|3|XT^BM^E1^1|SW||||||||||||||83|||||A<CR><ETX>BE<CR><LF> <ACK> <STX>5L|1<CR><ETX>3E<CR><LF> <ACK> <EOT>
38 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
Pattern 4 (M-4)Level Record
0 H1 M-XT
...: Rack pack Information
...:0 L
Pattern 5 (M-5)Level Record
0 H1 M-XT2 M-DP
: } max. 40 records } Rack pack Information :
...:
...:0 L
Details of Records:
Instrument Sending Record Pattern
Substance Data Record [M-XT Record] Pattern Pattern Pattern Pattern Pattern
Field DataType
Content I-1 *1) I-2 *1) I-3 I-4 I-5
1 record type ID (M) char[1] M M ß ß ß ß
2 sequence number pos_int
3 record type sub-ID(XT)
char[2]^text^text^text
XT^BM^E1^1
XT^BM^E1^1
ß ß ß ß
4 substance type char[2] TR, DR,PR, BS,AB, CC,TP, VS,
DW, LW,SW
TR:Test Reagent
DR:Diluent
PR:Pretreatment
BS:Blank Set
[NONE] AB:Assay Buffer
CC:Cell Cleaner
DW:Distill. Water
LW:Liquid Waste
SW:Solid Waste
TP:TipVS:
Vessel
5 substanceidentifier
text Test ReagentPretreatment
Test No.Diluent
Diluent No.Blank Set
Blank Set No.
[NONE] ABCC1,ABCC2
[NONE] [NONE]
6 substance name text Test ReagentPretreatment
Test CodeDiluent
Blank SetName
[NONE] [NONE] [NONE] [NONE]
Substance Data Message - XT/DP 39
Elecsys® 2010 - Host Interface Substance Data Message - XT/DP
Substance Data Record [M-XT Record] Pattern Pattern Pattern Pattern Pattern
Field DataType
Content I-1 *1) I-2 *1) I-3 I-4 I-5
7 test number(s) text{\...}
[NONE] [NONE] [NONE] [NONE] [NONE]
8 data carrier type text BC,MBC,MBCC
[NONE] [NONE] [NONE] [NONE] [NONE]
9 container type text RP RP [NONE] [NONE] [NONE] [NONE]
10 container cap type text SCE1 SCE1 [NONE] [NONE] [NONE] [NONE]
11 initial volume int{^unit}
Test ReagentPretreatment
Blank SetNo. of tests
DiluentVolume
[NONE] [NONE] [NONE] [NONE]
12 substance lotnumber
int Lot No. [NONE] Lot No. [NONE] [NONE]
13 container/packagenumber
int{^int}
Rack PackNo.
[NONE] [NONE] [NONE] [NONE]
14 containerregistrationdate & time
d_t Rack Packfirst
registrationdate & time
[NONE] [NONE] [NONE] [NONE]
15 container firstopening/usagedate & time
d_t [NONE] [NONE] [NONE] [NONE] [NONE]
16 stability on board int [NONE] [NONE] [NONE] [NONE] [NONE]
17 expiring date d expiringdate of lot
*6)
[NONE] [NONE] [NONE] [NONE]
18 actual level int{^unit}
Test ReagentPretreatment
Blank Setreal # of tests
*4)Diluent real
volume
[NONE] percent.number
DistilledWater: 0: OK 1: EmptyLiquid Waste: 0: OK 1: FullSolid Waste: percentagenumber *5)
real numberof tips orvessels
19 available level int{^unit}
[NONE] [NONE] [NONE] [NONE] [NONE]
20 container carriertype
text RD, VR,TR
RD RD [NONE] [NONE] *3)
21 container carrier ID text [NONE] [NONE] [NONE] [NONE] [NONE]
22 container positionin carrier
int Position inthe reagent
rotor
Position inthe reagent
rotor
[NONE] [NONE] [NONE]
23 substance status char[1]{\...}
A, R, C A *2) R A A A
40 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
*1) I-1: in case that there is a Rack Pack in the position on thereagent rotor.I-2: in case that there is not a Rack Pack in the position onthe reagent rotor.
*2) If calibration request has been generated the following twocharacters are added. '\C'
*3) TR: Tip RackVR: Vessel Rack
*4) In case of pre treatment RackPack, if there is no AssayRackPack which corresponds to this pre treatment RackPack,no character is set in this field. ([NONE])
*5) In case of „Solid Waste Counter No Check Mode“, ‘0’ is set inthis field as a number of items of Solid Waste.
*6) In case that it is no need to check expiring date for thisRackPack, ‘000000’ is set in this field.
Diluent Parameters Record [M-DP Record] Pattern
Field Data Type Content I-11
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(DP)
char[2]^text^text^text
DP^BM^E1^1 DP^BM^E1^1
4 test number text Test No.
5 test lot number int [NONE]
6 Endogenous content /unit
text{^unit}
[NONE]
Host Sending Record Pattern
Substance Data Record [M-XT Record] Pattern Pattern Pattern Pattern Pattern
Field DataType
Content L-1 L-2 L-3 L-4 L-5
1 record type ID (M) char[1] M M ß ß ß ß
2 sequence number pos_int
3 record type sub-ID(XT)
(XTQ)
char[3]^text^text^text
XT^BM^E1^1
XTQ^BM^E1^1
XT^BM^E1^1
XTQ^BM^E1^1
ß ß ß
Substance Data Message - XT/DP 41
Elecsys® 2010 - Host Interface Substance Data Message - XT/DP
Substance Data Record [M-XT Record] Pattern Pattern Pattern Pattern Pattern
Field DataType
Content L-1 L-2 L-3 L-4 L-5
4 substance type char[2] TR, DR,PR, BS,AB, CC,TP, VS,
DW, LW,SW
TR:Test Reagent
DR:Diluent
PR:Pretreatment
BS:Blank Set
[NONE] AB:Assay Buffer
CC:Cell Cleaner
DW:Distill. Water
LW:Liquid Waste
SW:Solid Waste
TP:TipVS:
Vessel
5 substanceidentifier
text Test ReagentPretreatment
Test No.Diluent
Diluent No.Blank Set
Blank Set No.
[NONE] ABCC1,ABCC2
[NONE] [NONE]
6 substance name text [NONE] [NONE] [NONE] [NONE] [NONE]
7 test number(s) text{\...}
[NONE] [NONE] [NONE] [NONE] [NONE]
8 data carrier type text BC,MBC,MBCC
[NONE] [NONE] [NONE] [NONE] [NONE]
9 container type text RP [NONE] [NONE] [NONE] [NONE] [NONE]
10 container cap type text SCE1 [NONE] [NONE] [NONE] [NONE] [NONE]
11 initial volume int{^unit}
[NONE] [NONE] [NONE] [NONE] [NONE]
12 substance lotnumber
int Lot No. [NONE] [NONE] [NONE] [NONE]
13 container/packagenumber
int{^unit}
Rack PackNo.
[NONE] [NONE] [NONE] [NONE]
14 containerregistrationdate & time
d_t [NONE] [NONE] [NONE] [NONE] [NONE]
15 container firstopening/usagedate & time
d_t [NONE] [NONE] [NONE] [NONE] [NONE]
16 stability on board int [NONE] [NONE] [NONE] [NONE] [NONE]
17 expiring date d [NONE] [NONE] [NONE] [NONE] [NONE]
18 actual level int{^unit}
Test ReagentPretreatment
Blank Setreal # of testsDiluent real
volume
[NONE] [NONE] [NONE] [NONE]
19 available level int{^unit}
[NONE] [NONE] [NONE] [NONE] [NONE]
20 container carriertype
text RD [NONE] [NONE] [NONE] [NONE] [NONE]
21 container carrier ID text [NONE] [NONE] [NONE] [NONE] [NONE]
22 container positionin carrier
int [NONE] Position inthe reagent
[NONE] [NONE] [NONE]
42 Substance Data Message - XT/DP
Substance Data Message - XT/DP Elecsys® 2010 - Host Interface
Substance Data Record [M-XT Record] Pattern Pattern Pattern Pattern Pattern
Field DataType
Content L-1 L-2 L-3 L-4 L-5
rotor
23 substance status char[1]{\...}
A, R, C A [NONE] [NONE] [NONE] [NONE]
Calibrator Parameters Message - XT/CP 43
Elecsys® 2010 - Host Interface Calibrator Parameters Message - XT/CP
5. Calibrator Parameters Message - XT/CP
Purpose to use this message:
Upload Instrument à Host
Purpose Host InstrumentUpload after completion of scanning of the Bar-code Card.(Upload information of Bar-code Card.)
I-1, I-11(M-1)
Event: Scan bar-code card of a calibrator.à The instrument uploads a Calibrator Parameters Message for the bar-code card.The instrument uploads 1 message/1 calibrator bar-code card.
Host Query — Instrument Answer
Purpose Host InstrumentQuery for information of Calibrator Parameters(Answer information of Calibrator Parameters that wererequested from Host.)
L-1, L-11(M-2)
I-1, I-11(M-1)
Event: The instrument receives a Calibrator Parameters Message Query.-----> The instrument answers to the query.The Host queries for Calibrator Parameters information for one test (or some test).The instrument always answers with 1 message/1 test.
Download Host à Instrument
Purpose Host InstrumentDownload Information of Calibrator Parameters(Download only target value of each Calibrator Parameters.)(If an instrument has never seen the Calibrator Parametersbefore, the downloading becomes invalid.)
L-2, L-12(M-2)
L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: The instrument receives a Calibrator Parameters Message.à The instrument overwrites information of the calibrator parameter.The Host downloads 1 message/1 test (or 1 message/ some tests).If the calibrator parameters that were sent from Host are data that the instrument doesnot know, then this data is not valid (will be ignored).
44 Calibrator Parameters Message - XT/CP
Calibrator Parameters Message - XT/CP Elecsys® 2010 - Host Interface
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-XT2 M-CP
: }max. 50 records :
0 L
Trace Example:
Instrument read Calibrator Bar-code card: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|RC||||||||190763|||||19971100||||||
<CR><ETX>CF<CR><LF> <ACK> <STX>3M|1|CP^BM^E1^1|20|A|47.50|99999999|<CR><ETX>71<CR><LF> <ACK> <STX>4M|2|CP^BM^E1^1|20|B|74.70|99999999|<CR><ETX>76<CR><LF> <ACK> <STX>5M|3|CP^BM^E1^1|20|C|0.000|99999999|<CR><ETX>67<CR><LF> <ACK> <STX>6M|4|CP^BM^E1^1|20|D|0.000|99999999|<CR><ETX>6A<CR><LF> <ACK> <STX>7M|5|CP^BM^E1^1|20|E|0.000|99999999|<CR><ETX>6D<CR><LF> <ACK> <STX>0M|6|CP^BM^E1^1|20|A|46.80|190619|<CR><ETX>E7<CR><LF> <ACK> <STX>1M|7|CP^BM^E1^1|20|B|139.0|190619|<CR><ETX>E5<CR><LF> <ACK> <STX>2M|8|CP^BM^E1^1|20|C|0.000|190619|<CR><ETX>DB<CR><LF> <ACK> <STX>3M|9|CP^BM^E1^1|20|D|0.000|190619|<CR><ETX>DE<CR><LF> <ACK> <STX>4M|10|CP^BM^E1^1|20|E|0.000|190619|<CR><ETX>08<CR><LF> <ACK> <STX>5M|11|CP^BM^E1^1|20|A|50.40|190764|<CR><ETX>10<CR><LF> <ACK> <STX>6M|12|CP^BM^E1^1|20|B|147.0|190764|<CR><ETX>16<CR><LF> <ACK> <STX>7M|13|CP^BM^E1^1|20|C|0.000|190764|<CR><ETX>0D<CR><LF> <ACK> <STX>0M|14|CP^BM^E1^1|20|D|0.000|190764|<CR><ETX>08<CR><LF> <ACK> <STX>1M|15|CP^BM^E1^1|20|E|0.000|190764|<CR><ETX>0B<CR><LF> <ACK> <STX>2L|1<CR><ETX>3B<CR><LF> <ACK> <EOT>
Calibrator Parameters Message - XT/CP 45
Elecsys® 2010 - Host Interface Calibrator Parameters Message - XT/CP
Pattern 2 (M-2)Level Record
0 H1 M-XT2 M-CP0 L
Details of Records:
Instrument Sending Record Pattern
Substance Data Record [M-XT Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(XT)
char[2]^text^text^text
XT^BM^E1^1 XT^BM^E1^1
4 substance type char[2] RC RC: Calibrator
5 substance identifier text [NONE]
6 substance name text [NONE]
7 test number(s) text{\...}
[NONE]
8 data carrier type text [NONE]
9 container type text [NONE]
10 container cap type text [NONE]
11 initial volume int{^unit}
[NONE]
12 substance lot number int Calibrator LotID
13 container/packagenumber
int{^int}
[NONE]
14 container registrationdate & time
d_t [NONE]
15 container firstopening/usage date &time
d_t [NONE]
16 stability on board int [NONE]
17 expiring date d expiring dateof lot
18 actual level int{^unit}
[NONE]
19 available level int{^unit}
[NONE]
46 Calibrator Parameters Message - XT/CP
Calibrator Parameters Message - XT/CP Elecsys® 2010 - Host Interface
Substance Data Record [M-XT Record] Pattern
Field Data Type Content I-1
20 container carrier type text [NONE]
21 container carrier ID text [NONE]
22 container position incarrier
int [NONE]
23 substance status char[1]{\...}
[NONE]
Calibrator Parameters Record [M-CP Record] Pattern
Field Data Type Content I-11
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(CP)
char[2]^text^text^text
CP^BM^E1^1 CP^BM^E1^1
4 test number text Test No.
5 calibrator level char[1] A, B, C, D, E A, B, C, D, E
6 target value / unit text{^unit}
reference value^unit
7 reagent lot number int Reagent LotNo.
8 reagent package number int{^int}
[NONE]
Host Sending Record Pattern
Substance Data Record [M-XT Record] Pattern Pattern
Field Data Type Content L-1 L-2
1 record type ID (M) char[1] M M M
2 sequence number pos_int
3 record type sub-ID(XT)
(XTQ)
char[3]^text^text^text
XT^BM^E1^1
XTQ^BM^E1^1 XTQ^BM^E1^1
XT^BM^E1^1
4 substance type char[2] RC RC: Calibrator RC: Calibrator
5 substance identifier text [NONE] [NONE]
6 substance name text [NONE] [NONE]
7 test number(s) text{\...}
[NONE] [NONE]
8 data carrier type text [NONE] [NONE]
9 container type text [NONE] [NONE]
Calibrator Parameters Message - XT/CP 47
Elecsys® 2010 - Host Interface Calibrator Parameters Message - XT/CP
Substance Data Record [M-XT Record] Pattern Pattern
Field Data Type Content L-1 L-2
10 container cap type text [NONE] [NONE]
11 initial volume int{^unit}
[NONE] [NONE]
12 substance lot number int [NONE] [NONE]
13 container/packagenumber
int{^int}
[NONE] [NONE]
14 container registrationdate & time
d_t [NONE] [NONE]
15 container firstopening/usage date &time
d_t [NONE] [NONE]
16 stability on board int [NONE] [NONE]
17 expiring date d [NONE] [NONE]
18 actual level int{^unit}
[NONE] [NONE]
19 available level int{^unit}
[NONE] [NONE]
20 container carrier type text [NONE] [NONE]
21 container carrier ID text [NONE] [NONE]
22 container position incarrier
int [NONE] [NONE]
23 substance status char[1]{\...}
[NONE] [NONE]
Calibrator Parameters Record [M-CP Record] Pattern Pattern
Field Data Type Content L-11 L-12
1 record type ID (M) char[1] M M M
2 sequence number pos_int
3 record type sub-ID(CP)
(CPQ)
char[3]^text^text^text
CP^BM^E1^1
CPQ^BM^E1^1 CPQ^BM^E1^1
CP^BM^E1^1
4 test number text Test No. Test No.
5 calibrator level char[1] A, B, C, D, E [NONE] A, B, C, D, E
6 target value / unit text{^unit}
[NONE] reference value
7 reagent lot number int [NONE] Reagent LotNo.
8 reagent package number int{^int}
[NONE] [NONE]
48 Control Parameters Message - XT/QP
Control Parameters Message - XT/QP Elecsys® 2010 - Host Interface
6. Control Parameters Message - XT/QP
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload after completion of scanning of the Bar-code Card(Upload Information of Bar-code Card.)
I-1, I-11(M-1)
Event: Scan bar-code card for a control.à The instrument uploads a Control Parameters Message of the bar-code card.The instrument uploads 1 message/1 control bar-code card.
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Control Parameters(Answer Information of Control Parameters that wererequested from Host.)
L-1, L-11(M-2)
I-1, I-11(M-3)
Event: The instrument receives a Control Parameters Message Query.à The instrument answers to the query.The Host queries for Control Parameters information for one (or some) control No.The instrument always answers 1 message/1 control No..
Host Download
Purpose Host InstrumentDownload Information of Control Parameters(Download only target value/unit and target range of eachControl Parameters.)(If an Instrument has never seen the Control Parametersbefore, the downloading becomes invalid.)
L-2, L-12(M-2)
L-*, I-*: refer to „Details of Records“M-*: refer to „Construction of Message“
Event: The instrument receives a Control Parameters Message.à The instrument overwrites the information of Control Parameters.The Host downloads 1 message/1 control (or 1 message/ some controls).A control is identified by control no. and control lot ID.If the Control Parameters that were sent from the Host are data that the instrumentdoes not now, then this data is not valid (will be ignored).
Control Parameters Message - XT/QP 49
Elecsys® 2010 - Host Interface Control Parameters Message - XT/QP
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-XT2 M-QP
: } max. 28 records :
0 L
Trace Example:
Instrument read Control Bar-code card: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|XT^BM^E1^1|CS|1|PCU1||||||1997402|||||19991200||||||<CR>
<ETX>53<CR><LF> <ACK> <STX>3M|1|QP^BM^E1^1|10||0.010^uIU/ml|0.008|0.012|N<CR>
<ETX>90<CR><LF> <ACK> <STX>4M|2|QP^BM^E1^1|20||114.0^nmol/l|90.06|137.9|N<CR>
<ETX>E6<CR><LF> <ACK> <STX>5M|3|QP^BM^E1^1|30||44.00^pmol/l|37.40|50.60|N<CR>
<ETX>E3<CR><LF> <ACK> <STX>6M|4|QP^BM^E1^1|40||1.13^TBI|0.893|1.37|N<CR>
<ETX>15<CR><LF> <ACK> <STX>7M|5|QP^BM^E1^1|310||12.60^IU/ml|8.82|16.38|N<CR>
<ETX>47<CR><LF> <ACK> <STX>0L|1<CR><ETX>39<CR><LF> <ACK> <EOT>
Pattern 2 (M-2)Level Record
0 H1 M-XT2 M-QP0 L
50 Control Parameters Message - XT/QP
Control Parameters Message - XT/QP Elecsys® 2010 - Host Interface
Pattern 3 (M-3)Level Record
0 H1 M-XT2 M-QP
: } max. 28 records :
1 M-XT } max. 5 pairs : :
0 L
Details of Records:
Instrument Sending Record Pattern
Substance Data Record [M-XT Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(XT)
char[2]^text^text^text
XT^BM^E1^1 XT^BM^E1^1
4 substance type char[2] CS CS: Control
5 substance identifier text Control No.
6 substance name text Control Name
7 test number(s) text{\...}
[NONE]
8 data carrier type text [NONE]
9 container type text [NONE]
10 container cap type text [NONE]
11 initial volume int{^unit}
[NONE]
12 substance lot number int Control Lot No
13 container/packagenumber
int{^int}
[NONE]
14 container registrationdate & time
d_t [NONE]
15 container firstopening/usage date &time
d_t [NONE]
16 stability on board int [NONE]
17 expiring date d expiring date oflot
18 actual level int [NONE]
Control Parameters Message - XT/QP 51
Elecsys® 2010 - Host Interface Control Parameters Message - XT/QP
Substance Data Record [M-XT Record] Pattern
Field Data Type Content I-1{^unit}
19 available level int{^unit}
[NONE]
20 container carrier type text [NONE]
21 container carrier ID text [NONE]
22 container position incarrier
int [NONE]
23 substance status char[1]{\...}
[NONE]
Control Parameters Record [M-QP Record] Pattern
Field Data Type Content I-11
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(QP)
text^text^text^text
QP^BM^E1^1 QP^BM^E1^1
4 test number text Test No.
5 reagent lot number int [NONE]
6 target value / unit text{^unit}
target value^unit
7 lower limit text Lower limit
8 upper limit text Upper limit
9 flags char[1] A, N *1)
*1) A: ActiveN: Not active
Host Sending Record Pattern
Substance Data Record [M-XT Record] Pattern Pattern
Field Data Type Content L-1 L-2
1 record type ID (M) char[1] M M M
2 sequence number pos_int
3 record type sub-ID(XT)
(XTQ)
text^text^text^text
XT^BM^E1^1
XTQ^BM^E1^1
XTQ^BM^E1^1 XT^BM^E1^1
4 substance type char[2] CS CS: Control CS: Control
5 substance identifier text Control No. Control No.
52 Control Parameters Message - XT/QP
Control Parameters Message - XT/QP Elecsys® 2010 - Host Interface
Substance Data Record [M-XT Record] Pattern Pattern
Field Data Type Content L-1 L-2
6 substance name text [NONE] [NONE]
7 test number(s) text{\...}
[NONE] [NONE]
8 data carrier type text [NONE] [NONE]
9 container type text [NONE] [NONE]
10 container cap type text [NONE] [NONE]
11 initial volume int{^unit}
[NONE] [NONE]
12 substance lot number int [NONE] Control lot No.
13 container/packagenumber
int{^int}
[NONE] [NONE]
14 container registrationdate & time
d_t [NONE] [NONE]
15 container firstopening/usage date &time
d_t [NONE] [NONE]
16 stability on board int [NONE] [NONE]
17 expiring date d [NONE] [NONE]
18 actual level int{^unit}
[NONE] [NONE]
19 available level int{^unit}
[NONE] [NONE]
20 container carrier type text [NONE] [NONE]
21 container carrier ID text [NONE] [NONE]
22 container position incarrier
int [NONE] [NONE]
23 substance status char[1]{\...}
[NONE] [NONE]
Control Parameters Record [M-QP Record] Pattern Pattern
Field Data Type Content L-11 L-12
1 record type ID (M) char[1] M M M
2 sequence number pos_int
3 record type sub-ID(QP)
(QPQ)
text^text^text^text
QP^BM^E1^1
QPQ^BM^E1^1 QPQ^BM^E1^1
QP^BM^E1^1
4 test number text [NONE] Test No.
5 reagent lot number int [NONE] [NONE]
6 target value / unit text{^unit}
[NONE] target value
Control Parameters Message - XT/QP 53
Elecsys® 2010 - Host Interface Control Parameters Message - XT/QP
Control Parameters Record [M-QP Record] Pattern Pattern
Field Data Type Content L-11 L-12
7 lower limit text [NONE] target range*
8 upper limit text [NONE] [NONE]*
9 flags char[1] A, N [NONE] [NONE]
* No Changes. (Same as V02-XX)
54 Calibration Data Message - CR
Calibration Data Message - CR Elecsys® 2010 - Host Interface
7. Calibration Data Message - CR
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload after completion of Lot Calibration(Upload Information of Result of Lot Calibration.)
I-1(M-1)
Upload after completion of Rack Pack Calibration(Upload Information of Result of Rack Pack Calibration.)
I-2(M-1)
Upload after completion of Blank Set(Upload Information of Result of Blank Set.)
I-3(M-1)
Upload after releasing Calibration data(Upload Information of Result of Rack Pack or LotCalibration.)
I-1 or I-2(M-1)
I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: Output the measured calibration results of the test.à The instrument uploads a Calibration Result Message of this test.The instrument uploads 1 message/1 test.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-CR0 L
Trace Example:
Instrument sends Calibration Data: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|CR^BM^E1^1|30|192882|ng/dl|RD^1^12|1228|N^M||
19970423103101|O|I^0^----------\M^0^-----\S^0^----------\R^1.00^0.000\D^0^-----|252858.000000^4.043450^1.316730^8366.690430^0.000000\-376.218567^1.254267\1.000000|SD^1^1^192881^0\SD^1^2^19288<ETB>9D<CR><LF>
<ACK> <STX>31^0\SD^1^^^\SD^1^^^\SD^1^^^|48008^49049\14824^14835\-0.000^-
0.000\-0.000^-0.000\-0.000^-0.000<CR><ETX>50<CR><LF> <ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK>
Calibration Data Message - CR 55
Elecsys® 2010 - Host Interface Calibration Data Message - CR
<STX>2M|1|CR^BM^E1^1|170|170459|mIU/ml|RD^1^2|1295|N^M||19971211164831|O|I^0^----------\M^0^-----\S^0^----------\R^1.00\D^0^-----|801.572021^42261.300781^1.012350^5922030.000000^0.000000\-60043.859375^3.530345\1.000000|SD^0^1^417333^0\SD^0^2^4173<ETB>64<CR><LF>
<ACK> <STX>333^0\SD^^^^\SD^^^^\SD^^^^|17635^17652\172243^171273\
^\^\^||1<CR><ETX>D6<CR><LF> <ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Calibration Result [M-CR Record] Pattern Pattern Pattern
Field Data Type Content I-1 I-2 I-3
1 record type ID (M) char[1] M M M M
2 sequence number pos_int
3 record type sub-ID(CR)
char[2]^text^text^text
CR^BM^E1^1
CR^BM^E1^1 CR^BM^E1^1 CR^BM^E1^1
4 test number text Test No. Test No. Blank Set No
5 test lot number int Reagent LotNo.
Reagent LotNo.
Blank Set lotNo.
6 unit unit unit unit [NONE]
7 test containerdescription
text^text^text
RD^1
^position
RD^1
^position
RD^1
^position
8 test package number int Rack PackNo.
Rack PackNo.
Rack PackNo.
9 calibration method char[1]^char[1]
N^M, N^R,I^
N^M
N^R
I^
10 replication number int [NONE] [NONE] [NONE]
11 date/time calibrationcompleted
d_t result readydate/time
result readydate/time
result readydate/time
12 result status char[1] O,Q,F O: OKQ: Question
F: Failed
O: OKQ: Question
F: Failed
O: OKF: Failed
13 result characteristics char[1]^char[1]
^text{\...}
I, M, S, R,D, L, B, F,
T, C
*2) *2) *2)
14 curve parameters text{\...}
*1 *1 [NONE]
15 calibrators containerdescription
text^text
SD or RK *3)
^S.RotorNo.
SD or RK *3
^S.RotorNo.
[NONE]
56 Calibration Data Message - CR
Calibration Data Message - CR Elecsys® 2010 - Host Interface
Calibration Result [M-CR Record] Pattern Pattern Pattern
Field Data Type Content I-1 I-2 I-3^text^text^text{\...}
or RackID^position^lot No
^bottle No.{\...}
or RackID^position^lot No
^bottle No.{\...}
16 calibrators result text{^...}{\...}
1st signal^2nd signal
{\...}
1st signal^2nd signal
{\...}
1st signal^2nd signal^3rd signal^4th signal^5th signal^6th signal
17 Cut-off parameters text{^...} *4) *4) [NONE]
18 Expired RackPack Flag int *5) *5) [NONE]
*1) curve parameter[0]^curve parameter[1]^curve parameter[2]^curve parameter[3]^curve parameter[4]\2point correction parameter[0]^2point correction parameter[1]\1point correction parameter[0]
*** This field is used for quantitative tests. ***
*2) result characteristicscomponent : characteristicID ^ text1 ^ text2
[For Quantitative tests]
Criteria characteristicID text1 text2
Missing values I 0 (OK),1 (Question),
2 (Fail)
ex. AA--CC----
Monotony of curve M 0 (OK),1 (Question),
2 (Fail)
ex. A---E
Minimum signal S 0 (OK),1 (Question),
2 (Fail)
ex. AA--CC----
Calibration factor R Value ofCalibration
factor
[NONE]
Deviation of dupl.Measurement
D 0 (OK),1 (Question),
2 (Fail)
ex. A---E
[For Qualitative tests]
Criteria characteristicID text1 text2
Missing values I 0 (OK),1 (Question),
2 (Fail)
ex. AA--
Monotony of curve M [NONE] [NONE]
Minimum signal S [NONE] [NONE]
Calibration Data Message - CR 57
Elecsys® 2010 - Host Interface Calibration Data Message - CR
Calibration factor R [NONE] [NONE]
Deviation of dupl.Measurement
D 0 (OK),1 (Question),
2 (Fail)
ex. -B
Slope L 0 (OK), 2 (Fail) OK or NG
Min/Max signal B 0 (OK),1 (Question),
2 (Fail)
ex. >-<-
Minimum acceptabledifference
F 0 (OK), 2 (Fail) OK or NG
[For Blank Set]
Criteria characteristicID text1 text2
Achievement of Targets T 0 (OK), 2 (Fail) OK or NG
Comparison to last validBlankSet
C 0 (OK), 2 (Fail) OK or NG
*3) SD :SampleDiskRK : Rack
*4) cut-off parameter[0]^cut-off parameter[1]^cut-off parameter[2]^cut-off parameter[3]^cut-off parameter[4]^cut-off parameter[5]
*** This field is used for Qualitative tests. ***
*5) 0: Calibration performed with „Not Expired RackPack“1: Calibration performed with „Expired RackPack“
58 Sample Status Message - SS
Sample Status Message - SS Elecsys® 2010 - Host Interface
8. Sample Status Message - SS
Purpose to use this message:
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Sample Status on the Sample Rotor(Answer Information of Sample Status on the Sample Rotorthat was requested from Host.)
L-1(M-1)
I-1(M-1)
L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: The instrument receives a Sample Status Message Query.à The instrument answers to the query.The Host queries for Sample Status information for one (or some) position on thesample rotor.The instrument always answers 1 message/1 position.
Construction of Message:
Pattern 4 (M-4)Level Record
0 H1 M-SS0 L
Details of Records:
Instrument Sending Record Pattern
Sample Status Record [M-SS Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(SS)
char[2]^text^text^text
SS^BM^E1^1 SS^BM^E1^1
4 specimen ID text{^text}{^text}{^text}
Sample ID
Sample Status Message - SS 59
Elecsys® 2010 - Host Interface Sample Status Message - SS
Sample Status Record [M-SS Record] Pattern
Field Data Type Content I-1
5 position description text
^text^text^text^text
Sequence No^S.Rotor No or
RackID^Position No.
^^data carrier
type *3)^container
type *4)
6 status char[1]{\...}
I, A, P, N, G *1)
7 identification date andtime
d_t [NONE]
8 availability time t [NONE]
9 result ready time text^t
{\...}
Test No.^Result ready
time *5)
*1) Case status
No Sample on this position N
Sample is set on this position
before pipetting I
now pipetting P
all tests completed A\C
some tests can not be done A\X
prepare for next start/restart
(“regist” status)
G
Host Sending Record Pattern
Sample Status Record [M-SS Record] Pattern
Field Data Type Content L-11
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(SS)
(SSQ)
char[3}^text^text^text
SS^BM^E1^1
SSQ^BM^E1^1 SSQ^BM^E1^1
60 Sample Status Message - SS
Sample Status Message - SS Elecsys® 2010 - Host Interface
Sample Status Record [M-SS Record] Pattern
Field Data Type Content L-11
4 specimen ID text{^text}{^text}{^text}
Sample ID
*2)
5 position description text
^text^text
^S. Rotor No &RackID
^Position No
*2)
6 status char[1]{\...}
I, A, C, P, N, X [NONE]
7 identification date andtime
d_t [NONE]
8 availability time t [NONE]
9 result ready time text^t
{\...}
[NONE]
*2) If both field data (‘Sample ID’ and ‘Position No.) are set then‘Sample ID’ is used as key data.
*3) data carrier type : „CALIBRATOR“, „CONTROL“, „SAMPLE“
*4) container type : „NORMAL“,“REDUCED“
*5) In case of calibrator no character is set in this field [NONE]
Instrument Status Message - IS 61
Elecsys® 2010 - Host Interface Instrument Status Message - IS
9. Instrument Status Message - IS
Purpose to use this message:
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of instrument status and alarm level(Answer Information of instrument status and alarm level.)
L-1(M-1)
I-1(M-1)
Event: The instrument receives a Instrument Status Message Query.-----> The instrument answers the query.
Instrument à Host Upload
Purpose Host InstrumentUpload when an instrument status changes(Upload Information of instrument status and alarm level.)
I-1(M-1)
L-*, I-*: refer to „Details of Records“M-*: refer to „Construction of Message“
Event: Change of instrument statusà The instrument uploads information for instrument status.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-IS0 L
Trace Example:
Query from the Host: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|ISQ^BM^E1^1||||||||||<CR><ETX>CD<CR><LF> <ACK> <STX>3L|1|N<CR><ETX>06<CR><LF> <ACK> <EOT>
62 Instrument Status Message - IS
Instrument Status Message - IS Elecsys® 2010 - Host Interface
Answer from the Instrument: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|IS^BM^E1^1|P|O||||19971210175024|||1<CR><ETX>A1<CR><LF> <ACK> <STX>3L|1|F<CR><ETX>FE<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Instrument Status Record [M-IS Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(IS)
char[2]^text^text^text
IS^BM^E1^1 IS^BM^E1^1
4 instrument status char[1] B, O, S, H, R,P, M, I, A, T, E,
N, C, F,V, W, X, Y, Z, L
Q, G, J, K, D
*1)
5 alarm level of processing char[1] O, W, E, F, R,H, L, X, Z, S
Q, G
*2)
6 alarm level of inventory char[1] O, E [NONE]
7 alarm level of QC char[1] O, W, E [NONE]
8 alarm level of calibration char[1] O, W, E [NONE]
9 last result ready date andtime
d_t last result readydate and time
10 next STAT ready dateand time
d_t [NONE]
11 walk away data and time d_t [NONE]
12 operator identification Operator ID
Host Sending Record Pattern
Instrument Status Record [M-IS Record] Pattern
Field Data Type Content L-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
Instrument Status Message - IS 63
Elecsys® 2010 - Host Interface Instrument Status Message - IS
Instrument Status Record [M-IS Record] Pattern
Field Data Type Content L-1
3 record type sub-ID(IS)
(ISQ)
char[3]^text^text^text
IS^BM^E1^1
ISQ^BM^E1^1 ISQ^BM^E1^1
4 instrument status char[1] B, O, S, H, R,P, M, I, A, T, E,
N, C, FV, W, X, Y, Z,
L, Q, G, J, K, D
[NONE]
5 alarm level of processing char[1] O, W, E, F, R,H, L, X, Z, S
Q, G
[NONE]
6 alarm level of inventory char[1] O, E [NONE]
7 alarm level of QC char[1] O, W, E [NONE]
8 alarm level of calibration char[1] O, W, E [NONE]
9 last result ready date andtime
d_t [NONE]
10 next STAT ready dateand time
d_t [NONE]
11 walk away data and time d_t [NONE]
12 operator identification text [NONE]
*1 Instrument StatusInstrument Status Disk Rack
SamplerRack
ConveyerCLAS
Initialize I I I IStand By B B B BPreparation A A A AOperation O O O OS. STOP H - - -P. STOP T - - -STOP S S S SE. STOP E E E ET/M M M M MS. Scan (incl. Restart) N - - -R. Scan R R R RSLEEP P P P PC. Scan C C C CFinalization F F F FR.Stop - V V VL.Stop - W W WA.Stop - - X -A.Stop & R.Stop - Y Y YA.Stop & L.Stop - Z Z ZLine Only - - L -
64 Instrument Status Message - IS
Instrument Status Message - IS Elecsys® 2010 - Host Interface
*1 Instrument StatusA.Line Stop - - - QI.Line Stop - - - GA.Line Stop\R.Stop - - - JI.Line Stop\R.Stop - - - KFD Access D D D D
S.Stop:In case of Sample Stop occurs no more new sample will be pipetted.
P.Stop:Partial Stop occurs in case if any hard ware module does not work any more. Depending onthe importance of the module the run will be finished or stopped.
Stop:After the Stop key was pressed the actual time step will be performed before the systemstops in a defined status.
E.Stop:In case of a hardware crash the Emergency Stop stops the system in an undefined status.
L.Stop:The Line Stop occurs when the rack sampler or conveyer line does not work (A-line = inputbuffer; B-line = bar-code reader line; C-line = output buffer).
A.Stop:???
R.Stop:Rack Supply Stop occurs when there are no more racks to be fed.
S.Scan:During Sample Scan samples are registered into the data base.
R.Scan:During Reagent Scan reagents are registered into the data base.
C.Scan:During Calibrator Scan calibrator bar-code sheets are registered into the data base.
*2) alarm level of processingU/I Headline Disk RS RC CLASNo Alarm 0 0 0 0Alarm (Warning) W W W WAlarm (P.STOP) E - - -Alarm (STOP) F F F FAlarm (E.STOP) R R R RAlarm (S.Stop) H - - -Alarm (L.Stop) - L L LAlarm (A.Stop) - X X XAlarm (L.Stop & A.Stop) - Z Z ZAlarm (R.Stop) - S S SAlarm (A.Line Stop) - - - QAlarm (I.Line Stop) - - - G
Log File Message - LF 65
Elecsys® 2010 - Host Interface Log File Message - LF
10. Log File Message - LF
Purpose to use this message:
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Log File(Answer information of Log File within time range that wasrequested from Host.)
L-1(M-1)
I-1(M-1)
L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: The instrument receives a Log File Message Query.à The instrument answers to the query.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-LF0 L
Details of Records:
Instrument Sending Record Pattern
Log File Record [M-LF Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(LF)
char[2]^text^text^text
LF^BM^E1^1 LF^BM^E1^1
4 file name text [NONE]
5 start date/time d_t start date/time
6 end date/time d_t end date/time
7 message category char[1] W, E, F, R, A,H, L, X, Z, S
Q, G
*1)
66 Log File Message - LF
Log File Message - LF Elecsys® 2010 - Host Interface
Log File Record [M-LF Record] Pattern
Field Data Type Content I-1
8 message(s) char[1]^d_t^int^int
^text{\...}
*2)
Host Sending Record Pattern
Log File Record [M-LF Record] Pattern
Field Data Type Content L-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(LF)
(LFQ)
char[3]^text^text^text
LF^BM^E1^1
LFQ^BM^E1^1 LFQ^BM^E1^1
4 file name text [NONE]
5 start date/time d_t start date/time
6 end date/time d_t end date/time
7 message category char[1] W, E, F, R, A,H, L, X, Z, S
Q, G
*1)
8 message(s) char[1]^d_t^int^int
^text{\...}
[NONE]
*1) Message categoryAlarm Level Disk RS RC CLASWarning W W W WP.STOP E - - -STOP F F F FE.STOP R R R RS.STOP H - - -L.STOP - L L LA.STOP - X X XL.STOP & A.STOP - Z Z ZR.STOP - S S SA.Line STOP - - - QI.Line STOP - - - GAll Level A A A A
Log File Message - LF 67
Elecsys® 2010 - Host Interface Log File Message - LF
*2) message category^alarm occurrence date and time^alarm code^alarm sub code^alarm string{\...}
68 Test Conditions Message - TC
Test Conditions Message - TC Elecsys® 2010 - Host Interface
11. Test Conditions Message - TC
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload when press <close> button after editing data at a TestConditions Screen(Upload Information of Test Conditions)
I-1(M-1)
Event: Edit data of test condition on the test condition screen and update the database by pressing “Close” button.à The instrument uploads a Test Conditions Message for the test that is displayed onthe current screen.
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Test Conditions(Answer Information of Test Conditions that were requestedfrom Host.)
L-1(M-1)
I-1(M-1)
Event: The instrument receives a Test Conditions Message Query.à The instrument answers to the query.The Host queries for Test Conditions information for one (or some) test.The instrument always answers with 1 message/1 test.
Host Download
Purpose Host InstrumentDownload Information of Test Conditions(Download only reference range lower/upper limit of eachTest Conditions Information. If an instrument has never seenthe Test Conditions information before, the downloadingbecomes invalid.)
L-2(M-1)
L-*, I-*: refer to „Details of Records“M-*: refer to „Construction of Message“
Event: The instrument receives a Test Conditions Messageà The instrument overwrites the information of the Test Conditions.The Host downloads a message of Test Conditions for one (or some) test.If the Test Conditions that were sent from the Host are data for a test that theinstrument does not know, then this data is not valid (will be ignored).
Construction of Message:
Pattern 4 (M-4)Level Record
0 H1 M-TC0 L
Test Conditions Message - TC 69
Elecsys® 2010 - Host Interface Test Conditions Message - TC
Trace Example:
Host queries for Test Conditions: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|TCQ^BM^E1^1||||||||||||<CR><ETX>C0<CR><LF> <ACK> <STX>3L|1|N<CR><ETX>06<CR><LF> <ACK> <EOT>
Instrument sends Test Conditions <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|TC^BM^E1^1|10|TSH|uIU/ml||1^1^2|0|0.270^4.20|
0||||y<CR><ETX>1C<CR><LF> <ACK> <STX>3M|2|TC^BM^E1^1|20|T4|nmol/l||1^1^2|0|66.00^174.0|
0||||y<CR><ETX>27<CR><LF> <ACK> <STX>4M|3|TC^BM^E1^1|30|FT4|pmol/l||1^1^2|0|13.00^23.00|
0||||y<CR><ETX>63<CR><LF> <ACK> <STX>5M|4|TC^BM^E1^1|40|T-UP|TBI||1^1^2|0|0.800^1.30|
0||||y<CR><ETX>1D<CR><LF> <ACK> <STX>6M|5|TC^BM^E1^1|60|FT3|pmol/l||1^1^2|0|4.00^7.80|
0||||y<CR><ETX>13<CR><LF> <ACK> <STX>7M|6|TC^BM^E1^1|100|E2|pg/ml||1^1^2|0|10.10^3960|
0||||y<CR><ETX>A9<CR><LF> <ACK> <STX>0M|7|TC^BM^E1^1|150|FSH|mIU/ml||1^1^2|0|0.110^198.0|
0||||y<CR><ETX>74<CR><LF> <ACK> <STX>1M|8|TC^BM^E1^1|200|TNTSTAT|ng/ml||1^1^2|0|0.000^0.100|
0||||y<CR><ETX>7A<CR><LF> <ACK> <STX>2M|9|TC^BM^E1^1|210|CKMBSTAT|ng/ml||1^1^2|0|0.000^4.00|
0||||y<CR><ETX>77<CR><LF> <ACK> <STX>3M|10|TC^BM^E1^1|300|CEA|ng/ml||1^1^2|0|0.000^4.60|
0||||y<CR><ETX>16<CR><LF> <ACK> <STX>4M|11|TC^BM^E1^1|310|AFP|IU/ml||1^1^2|0|0.000^6.40|
0||||y<CR><ETX>F0<CR><LF> <ACK> <STX>5M|12|TC^BM^E1^1|340|CA125|U/ml||1^1^2|0|0.000^35.00|
0||||y<CR><ETX>1F<CR><LF> <ACK> <STX>6M|13|TC^BM^E1^1|400|HBSAG|COI||1^1^2|0|^|
0||||<CR><ETX>88<CR><LF> <ACK> <STX>7M|14|TC^BM^E1^1|410|A-HBS|IU/l||1^1^2|0|2.00^10.00|
0||||y<CR><ETX>F7<CR><LF> <ACK> <STX>0M|15|TC^BM^E1^1|450|A-HBC|COI||1^1^2|0|^|
0||||<CR><ETX>5F<CR><LF> <ACK> <STX>1M|16|TC^BM^E1^1|600|B12|pg/ml||1^1^2|0|0.270^4.20|
0||||y<CR><ETX>00<CR><LF>
70 Test Conditions Message - TC
Test Conditions Message - TC Elecsys® 2010 - Host Interface
<ACK> <STX>2M|17|TC^BM^E1^1|670|CROSSL|ng/ml||1^1^2|0|0.270^4.20|
0||||y<CR><ETX>38<CR><LF> <ACK> <STX>3L|1|F<CR><ETX>FE<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Test Condition Record [M-TC Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(TC)
char[2]^text^text^text
TC^BM^E1^1 TC^BM^E1^1
4 test number text Test No.
5 test code text Test Code
6 unit name unit unit name
7 conversion factor text conversionfactor
8 replication number forsamples / controls /calibrators
int^int^int
1^1^2
9 auto-dilution factor int auto-dilutionfactor
10 reference range lower /upper limit
int^int
lower limit^upper limit *3)
11 test specific inventorythresholds
text{\...}
test specificinventory
thresholds
12 automatic rerun char[1] A, X [NONE]
13 rerun dilution factor int [NONE]
14 batch calibration char[1] y, n [NONE]
15 range check char[1] y, n *2)
Host Sending Record Pattern
Test Condition Record [M-TC Record] Pattern Pattern
Field Data Type Content L-1 L-2
1 record type ID (M) char[1] M M M
2 sequence number pos_int
Test Conditions Message - TC 71
Elecsys® 2010 - Host Interface Test Conditions Message - TC
Test Condition Record [M-TC Record] Pattern Pattern
Field Data Type Content L-1 L-2
3 record type sub-ID(TC)
(TCQ)
text^text^text^text
TC^BM^E1^1
TCQ^BM^E1^1 TCQ^BM^E1^1
TC^BM^E1^1
4 test number text Test No. *1) Test No. *1)
5 test code text [NONE] [NONE]
6 unit name unit [NONE] [NONE]
7 conversion factor text [NONE] [NONE]
8 replication number forsamples / controls /calibrators
int^int^int
[NONE] [NONE]
9 auto-dilution factor int [NONE] [NONE]
10 reference range lower /upper limit
int^int
[NONE] lower limit^upper limit
*4) *5)
11 test specific inventorythresholds
text{\...}
[NONE] [NONE]
12 automatic rerun char[1] A, X [NONE] [NONE]
13 rerun dilution factor int [NONE] [NONE]
14 batch calibration char[1] y, n [NONE] [NONE]
15 range check char[1] y, n [NONE] *2)
*1) If there is no data (NONE) in this field, an instrument sends alltest condition records to Host.
*2) y : Normal Range Check Function is available.n : Normal Range Check Function is not available.
*3) reference range lower/upper limitreference range is set as follows.
Tests Normal Range Check reference range
Quantitative Tests Not Available [NONE]
Available lower limit^upper limit
Qualitative Tests Not Available [NONE]
Available [NONE]
72 Test Conditions Message - TC
Test Conditions Message - TC Elecsys® 2010 - Host Interface
*4) reference range lower/upper limitInstrument treats this reference range as follows. (Available orNOT Available)
Tests Normal Range Check reference range
Quantitative Tests Not Available Available
Available Available
Qualitative Tests Not Available Not Available
Available Not Available
*5) reference range lower/upper limitmust be „lower limit < upper limit“
Test Application Message - TA 73
Elecsys® 2010 - Host Interface Test Application Message - TA
12. Test Application Message - TA
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload after executing C. Scan(Upload MBCC Information (586digits).)
I-1(M-1)
Upload after executing R. Scan(Upload MBC Information (586digits).)
I-1(M-2)
I-*: refer to „Details of Records“M-*: refer to „Construction of Message“
Event: a) Reagent Rotor scanà The instrument uploads MBC information (586 digits) of the rack packs on thereagent rotor.The instrument uploads 1message/18 positions.
b) Calibrator/control bar-code card scan.à The instrument uploads MBCC information (586 digits) of this bar-code card.The instrument uploads 1 message/1 calibrator (or control) bar-code card.
Pattern 1 (M-1)Level Record
0 H1 M-TA0 L
Pattern 1 (M-2)Level Record
0 H1 M-TA
: :
0 L
Trace Example:
Instrument updates after Reagent Scan (18 positions): <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|TA^BM^E1^1|MBC^<CR><ETX>49<CR><LF> <ACK> <STX>3M|2|TA^BM^E1^1|MBC^00000730205414080734234082728689704
010242070881124257269955222784000856322317776692976084451884107236094338414945841240361996008673689794278600039392085856219117763315723249227720861959956379520410800716598208614910419697<ETB>6C<CR><LF>
<ACK> <STX>4281244402755585073992234858530726385149212575697955326
8125520102420708811242572699552227840008563223177766929760844
74 Test Application Message - TA
Test Application Message - TA Elecsys® 2010 - Host Interface
51884107236094338414945841240361996008673689794278600039392085856219117763315723249227720861959956379520410800716598208614910<ETB>6B<CR><LF>
<ACK> <STX>5419697281244402755585073992234858530726385149138788721
66048859935212345678901234567890123456789012345678901234567890123451438<CR><ETX>F6<CR><LF>
<ACK> .. .. <ACK> <STX>2M|17|TA^BM^E1^1|MBC^<CR><ETX>80<CR><LF> <ACK> <STX>3M|18|TA^BM^E1^1|MBC^0003337441710178000026414732728591
045325267028048148156811648469334886272364100044644335119969615949687503268926064220048276755682909531999213857776265762209519421793614664344749070091307666161922310535701565974500756030<ETB>65<CR><LF>
<ACK> <STX>4696229226883953448239873669669769270523386774728906135
693527228421717197678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123456789012345678901234567890123<ETB>B1<CR><LF>
<ACK> <STX>5456789012345678901234567890123456789012345678901234567
890123456789012345678901234567890123456789012345678901234567890123455924<CR><ETX>1E<CR><LF>
<ACK> <STX>6L|1<CR><ETX>3F<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Test Application Record [M-TA Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(TA)
char[2]^text^text^text
TA^BM^E1^1 TA^BM^E1^1
4 application data char[3]^text
MBC^MBC data *2)
*2) 586 digits
Instrument Configuration Message - IC 75
Elecsys® 2010 - Host Interface Instrument Configuration Message - IC
13. Instrument Configuration Message - IC
Purpose to use this message:
Host Query — Instrument Answer
Purpose Host InstrumentQuery for Information of Instrument Configuration(Answer Information of Instrument Configurations that wererequested from Host.)
L-1(M-1)
I-1(M-1)
Event: The instrument receives an Instrument Configuration Message Query.à The instrument answers to the query.
Host Download
Purpose Host InstrumentDownload Information of Instrument Configuration (printingoption, uploading option, sending additional data)(Download only printing option, uploading option, and sendingadditional data and system clock.)
L-2(M-1)
Event: The instrument receives an Instrument Configuration Message.à The instrument overwrites the information of the instrument configuration.
Instrument à Host Upload
Purpose Host InstrumentUpload Information of Instrument Configuration when they arechanged (printing option, uploading option, system clock)(Upload Information of Instrument Configurations.)
I-1(M-1)
L-*, I-*: refer to „Details of Records“M-*: refer to „Construction of Message“
Event: Change of instrument configurationà The instrument uploads data of the instrument configuration.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-IC0 L
Trace Example
Host queries for Instrument Configuration: <ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|ICQ^BM^E1^1|||||||||||||||<CR><ETX>29<CR><LF> <ACK>
76 Instrument Configuration Message - IC
Instrument Configuration Message - IC Elecsys® 2010 - Host Interface
<STX>3L|1|N<CR><ETX>06<CR><LF> <ACK> <EOT>
Instrument sends Instrument Configuration:
<ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|IC^BM^E1^1||S|S|B||^All^^^|122755|070110^19971209^03 -
04^185||^MMDDYYYY^24|^^11||||<CR><ETX>08<CR><LF> <ACK> <STX>3L|1|F<CR><ETX>FE<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&||||||||||P||<CR><ETX>05<CR><LF> <ACK> <STX>2M|1|IC^BM^E1^1|||||ELEC2|^All^SD^^|112415||||||||<CR>
<ETX>79<CR><LF> <ACK> <STX>3L|1|N<CR><ETX>06<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Instrument Configuration Record [M-IC Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(IC)
char[2]^text^text^text
IC^BM^E1^1 IC^BM^E1^1
4 predilution factor int [NONE]
5 printing option char[1] S, M *1)
6 uploading option char[1] S, M *1)
7 documentation request char[1] P, U, B *3)
8 user defined instrumentname
text InstrumentName
9 sending additional data text^text^text
No^^^^,^All^^^,
^All^SD^^,^All^^RC^,^All^^^RR,
^All^SD^RC^,^All^SD^^RR,
^All^SD^RC^RR,^All^^RC^RR
*2)
Instrument Configuration Message - IC 77
Elecsys® 2010 - Host Interface Instrument Configuration Message - IC
Instrument Configuration Record [M-IC Record] Pattern
Field Data Type Content I-1
10 system clock t system clock
11 instrument information text^d
^text^text
serial no.^install date
^version^test count
12 keep function text 0, 1 [NONE]
13 language information text^text^text
^date format *4)
14 printer setup char[1]^text^text
^^
external printerpaper size in
inches
15 secondary cup type text HC, MC, PT [NONE]
16 number of secondaryadapters
int [NONE]
17 bar-code readerconfiguration
char[1]^char[1]^char[1]^char[1]
^int^char[1]
{\...}
[NONE]
18 other configuration data text{^...}
[NONE]
Host Sending Record Pattern
Instrument Configuration Record [M-IC Record] Pattern Pattern
Field Data Type Content L-1 L-2
1 record type ID (M) char[1] M M M
2 sequence number pos_int
3 record type sub-ID(IC)
(ICQ)
char[3]^text^text^text
IC^BM^E1^1
ICQ^BM^E1^1 ICQ^BM^E1^1
IC^BM^E1^1
4 predilution factor int [NONE] [NONE]
5 printing option char[1] S, M [NONE] *1)
6 uploading option char[1] S, M [NONE] *1)
7 documentation request char[1] P, U, B [NONE] *3)
8 user defined instrumentname
text [NONE] InstrumentName
9 sending additional data text^text
No^^^^,^All^^^,
[NONE] *2)
78 Instrument Configuration Message - IC
Instrument Configuration Message - IC Elecsys® 2010 - Host Interface
Instrument Configuration Record [M-IC Record] Pattern Pattern
Field Data Type Content L-1 L-2^text ^All^SD^^,
^All^^RC^,^All^^^RR,
^All^SD^RC^,^All^SD^^RR,
^All^SD^RC^RR,^All^^RC^RR
10 system clock t [NONE] system clock
11 instrument information text^d
^text^text
[NONE] [NONE]
12 keep function text 0, 1 [NONE] [NONE]
13 language information text^text^text
[NONE] [NONE]
14 printer setup char[1]^text^int
[NONE] [NONE]
15 secondary cup type text HC, MC, PT [NONE] [NONE]
16 number of secondaryadapters
int [NONE] [NONE]
17 bar-code readerconfiguration
char[1]^char[1]^char[1]^char[1]
^int^char[1]
{\...}
[NONE] [NONE]
18 other configuration data text{^...}
[NONE] [NONE]
*1) only the two options below are availableprinting option = ‘S’ & uploading option = ‘S’printing option = ‘M’ & uploading option = ‘M’
Sampling order .......SManual.....................M
Instrument Configuration Message - IC 79
Elecsys® 2010 - Host Interface Instrument Configuration Message - IC
*2) sending additional dataPossible contents:
No^^^^,^All^^^,^All^SD^^,^All^^RC^,^All^^^RR,^All^SD^RC^,^All^SD^^RR,^All^SD^RC^RR,^All^^RC^RR
E.g., No^^^^ means send no MSR’s at all;^All^SD^^RR means send all MSR’sbut SD record and RR record.
[NOTES]
When Host queries information with a MSR, the instrument sends MSRmessage for answer, regardless of this setting.
*3) documentation requestPrintout ...................PUpload ....................UBoth ........................B
*4) Date formatDDMMYYYY orMMDDYYYY orYYYMMDD
80 Service Data Message - SD
Service Data Message - SD Elecsys® 2010 - Host Interface
14. Service Data Message - SD
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload when an instrument detects Clot Detection(Upload pressure sensor values.)
I-1(M-1)
Upload when ADC module works(Upload ADC Raw Data.)
I-2(M-1)
Event: The instrument receives a Service Data Message Query.à The instrument answers to the query.
Host Query — Instrument Answer
Purpose Host InstrumentQuery for information of Service Data(Answer information of Service Data that was requested fromHost.)
L-1(M-1)
I-3(M-1)
L-*, I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: a) Clot detection in operationà The instrument uploads pressure sensor values.
b) ADC module works in operationà The instrument uploads the ADC raw data.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-SD0 L
Trace Example:
Instrument send Service Data: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|SD^BM^E1^1||19970409^^10119||Calibrator\165\30\
R1\119^123^142^159^162^158^155^155^156^155^150^145^141^139^137^134^131^129^127^126^124^123^122^121^120^119^118^118^118^118^118^118^119^119^119^119^118^102^88^88^91^90^84^80^79^80^78^77^76^7<ETB>86<CR><LF>
<ACK> <STX>36^75^75^75^75^75^76^76^76^76^77^77^79^79^79^80^81^81
^82^82^83^83^86^90^92^91^90^90^90^90^90^89^89^88^88^87^84^80^77^78^80^80^79^78^78^79^79^78^78^78^78^78^78^78^78^78^78^78^7
Service Data Message - SD 81
Elecsys® 2010 - Host Interface Service Data Message - SD
8^78^78^78^78^78^78^73^68^61^59^59^65^255^255^255^255^255^255^255^<ETB>35<CR><LF>
<ACK> <STX>4255|ClotDetection|<CR><ETX>09<CR><LF> <ACK> <STX>5L|1<CR><ETX>3E<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|SD^BM^E1^1||19970409^^10119||Calibrator\165\30\
Sample\62^67^84^96^96^92^93^97^94^73^59^60^62^58^51^47^47^47^45^44^43^43^44^44^44^45^46^47^48^49^50^52^53^54^55^56^57^59^60^61^62^63^64^65^66^67^68^69^70^71^71^72^73^74^74^75^76^76^77^77^78<ETB>32<CR><LF>
<ACK> <STX>3^79^79^79^80^80^81^81^82^82^82^85^88^90^90^88^88^88^88
^87^87^86^85^84^83^82^81^81^80^79^78^77^77^76^75^72^68^65^66^68^67^66^66^66^66^66^66^66^65^66^65^65^66^65^68^70^72^70^66^65^255^255^255^255^255^255^255^255|ClotDetection|<CR><ETX>CA<CR><LF>
<ACK> <STX>4L|1<CR><ETX>3D<CR><LF> <ACK> <EOT>
<ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|SD^BM^E1^1||19970409^^10136||ABLevelCheck
^13874\13216\13330\13198\13477\13489\13397\13340\12850\13044\13900\13657\13112\13096\13342\13117\13547\13373\13052\13283\13361\13452\13642\13846\13536\13004\13103\13410\12896\13427\13014\12747\13<ETB>90<CR><LF>
<ACK> <STX>3077\13132\13810\13370\12833\13306\13462\14055\13522\
13696\13337\13187\13752\13710\13873\13349\13300\12898\13026\12903\13050\13221\12766\13550\13289\13350\13104\13380\13230\13710\13455\13220\13456\13090\12989\13167\14146\13911\13733\13509\13<ETB>A0<CR><LF>
<ACK> <STX>4619\12977\12868\13001\13390\13478\13448\13057\13432\
14641\16415\17284\17317\17861\17667\17546\17782\17874\18048\18406\18186\18509\18104\17859\18156\18099\18454\18838\18377\18915\18419\18539\18268\18281\17801\18205\18456\18332\18396\18459\18<ETB>90<CR><LF>
<ACK> <STX>5458\18435\18481\18481\18284\18477\18127\18303\18382\
18477\18298\17933\17959\18205\18087\18108\17851\18074\17893\17866\17966\17888\17871\18377\17986\18562\18445\18485\18543\18542\18212\18224\17834\17829\18090\18158\18004\17924\17873\17944\17<ETB>C7<CR><LF>
<ACK> <STX>6418\17616\17704\18022\17713\17793\17757\17821\17830\
18144\17823\17912\17866\17890\17997\17888\17601\17297\17514\18263\17525\17514\17236\17821\17896\18001\17691\18097\18398\18421\17848\18088\18243\17849\18216\17606\17701\18284\18186\18424\18<ETB>96<CR><LF>
<ACK> <STX>7078\17751\18581\18231\18079\18148\17791\18174|ADCRawData|
<CR><ETX>F6<CR><LF>
82 Service Data Message - SD
Service Data Message - SD Elecsys® 2010 - Host Interface
<ACK> <STX>0L|1|F<CR><ETX>FB<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Service Data Record [M-SD Record] Pattern Pattern Pattern
Field Data Type Content I-1 I-2 I-3
1 record type ID (M) char[1] M M M M
2 sequence number pos_int
3 record type sub-ID(SD)
text^text^text^text
SD^BM^E1^1 SD^BM^E1^1 SD^BM^E1^1 SD^BM^E1^1
4 maintenanceschedule service data
text^int
^long{\^...}
[NONE] [NONE] [NONE]
5 operation log d^long^long
install date^
^operationtime
install date^
^operationtime
install date^
^operationtime
6 trace text{\...}
S,R,M [NONE] [NONE] [NONE]
7 service data text{\...}
*1) *2) [NONE]
8 kind of service data text Clot Detection ADC RawData
[NONE]
9 left space fordetermination in aninstrument’s database
int [NONE] [NONE] left space fordeterminationin aninstrument’sdata base
*1) Sample Type \ Seq No \ Test No \ Pipetting Pos \Pressure Sensor Value ^Pressure Sensor Value ^...Sample Type: Sample
ControlCalibratorBlank Set
Pipetting Pos: R1R2DiluentSampleBeadsPSRMDL1DL2
Service Data Message - SD 83
Elecsys® 2010 - Host Interface Service Data Message - SD
*2) Action Pattern ^Raw Data \ Raw Data...Action Pattern: AB Level Check
Measuring CellSystem VolumeCalibratorBlank Set
Service Data Record [M-SD Record] Pattern
Field Data Type Content L-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID(SDQ)
text^text^text^text
SDQ^BM^E1^1 SDQ^BM^E1^1
4 maintenance scheduleservice data
text^int
^long{\^...}
[NONE]
5 operation log d^long^long
[NONE]
6 trace text{\...}
S,R,M [NONE]
7 service data text{\...}
[NONE]
8 kind of service data text [NONE]
9 left space fordetermination in aninstrument’s data base
int [NONE]
84 Processing Message - PM
Processing Message - PM Elecsys® 2010 - Host Interface
15. Processing Message - PM
Purpose to use this message:
Instrument à Host Upload
Purpose Host InstrumentUpload after generating alarm message(Upload Information of alarms that were generated.)
I-1(M-1)
I-*: refer to "Details of Records"M-*: refer to "Construction of Message"
Event: An alarm or some alarms are generated in the instrument.à The instrument uploads one ore more Processing Messages.The instrument uploads 1 message/1 alarm.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-PM0 L
Trace Example:
Instrument sends Alarm Message: <ENQ> <ACK> <STX>1H|\^&<CR><ETX>E5<CR><LF> <ACK> <STX>2M|1|PM^BM^E1^1|19970525132822|858^0|
Communication Message Format Error : 44-01-08|W<CR><ETX>EA<CR><LF>
<ACK> <STX>3L|1<CR><ETX>3C<CR><LF> <ACK> <EOT>
Details of Records:
Instrument Sending Record Pattern
Processing Messages Record [M-PM Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
Processing Message - PM 85
Elecsys® 2010 - Host Interface Processing Message - PM
Processing Messages Record [M-PM Record] Pattern
Field Data Type Content I-1
3 record type sub-ID(PM)
char[2]^text^text^text
PM^BM^E1^1 PM^BM^E1^1
4 occurrence date and time d_t occurrencedate & time
5 message code int^int
alarm code^alarm sub
code
6 message name text alarm namelanguage
dependand
7 message category char[1] W, E, F, R, H,L, X, Z, S
Q, G
*1
*1) Alarm Level Disk RS RC CLASWarning W W W WP. STOP E - - -STOP F F F FE. STOP R R R RS. STOP H - - -L. STOP - L L LA. STOP - X X XL. STOP & A. STOP - Z Z ZR. STOP - S S SA.Line STOP - - - QI.Line STOP - - - G
86 Action Request / Action Confirmation Message - AR/AC
Action Request / Action Confirmation Message - AR/ACElecsys® 2010 - Host Interface
16. Action Request / Action Confirmation Message - AR/AC
Purpose to use this message:
Host Request — Instrument Confirmation
Purpose Host InstrumentRequest START/STOPUpload confirmation record response to an action request thatwas requested by Host.
L-1(M-1)
I-1(M-2)
L-*, I-* : refer to "Details of Records"M-* : refer to "Construction of Message"
Event: The instrument receives an Action Request Message (Start/Stop)à The instrument uploads Confirmation Message in response to the Action Requestmessage.
Construction of Message:
Pattern 1 (M-1)Level Record
0 H1 M-AR0 L
Pattern 2 (M-2)Level Record
0 H1 M-AC0 L
Details of Records:
Host Sending Record Pattern
Action Request Record [M-AR Record] Pattern
Field Data Type Content L-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID (AR) char[2]^text^text^text
AR^BM^E1^1 AR^BM^E1^1
Action Request / Action Confirmation Message - AR/AC 87
Elecsys® 2010 - Host InterfaceAction Request / Action Confirmation Message - AR/AC
Action Request Record [M-AR Record] Pattern
Field Data Type Content L-1
4 command text STOPSTART
STOP orSTART
5 time out int [NONE]
6 additional parameters int{^int...}
*2)
Instrument Sending Record Pattern:
Action Confirmation Record [M-AC Record] Pattern
Field Data Type Content I-1
1 record type ID (M) char[1] M M
2 sequence number pos_int
3 record type sub-ID (AC) char[2]^text^text^text
AC^BM^E1^1 AC^BM^E1^1
4 command text STOPSTART
STOP orSTART
5 processing status char[1] O, A *1)
6 additional action result text [NONE]
*1) O: action processedA: not allowed in current system state
*2) command additional parametersSTART Rotor No.
88 Header - H
Header - H Elecsys® 2010 - Host Interface
17. Header - H
Details of Records:
Host Sending Record Pattern
Message Header Record [H Record] Pattern
Field Data Type Content I-1/L-1
1 record type ID (H) char[1] H H
2 delimiter definition char[4] |\^& |\^&
3 message control ID text [NONE]
4 access password text [NONE]
5 sender name or ID text{^...}
*1)
6 sender street address text [NONE]
7 reserved field [NONE]
8 sender telephonenumber
text [NONE]
9 characteristics of sender text [NONE]
10 receiver ID text *2)
11 comment of specialinstructions
text [NONE]
12 processing ID char[1] P, D *3)
13 version No. int 1 [NONE]
14 date and time ofmessage
d_t [NONE]
*1) text1^BM^E1^text2text1: Host defined instrument nametext2: Version # of the software
*2) text3text3: Host name
*3) processing IDP: ProductionD: Debugging
Termination - L 89
Elecsys® 2010 - Host Interface Termination - L
18. Termination - L
Details of Records:
Instrument / Host Sending Record Pattern
Message Terminator Record [H Record] Pattern
Field Data Type Content I-1/L-1
1 record type ID (L) char[1] L L
2 sequence number pos_int 1 1
3 termination code char[1] I,F,E,Q *1)
*1) null: for normal end of messageI: normal end of an empty message(no data found) sent in
response to a remote query messageF: normal end of message sent in response to a remote
query messageE: unknown system errorQ: for the response to an invalid remote query
90 Receive Acceptance Status
Receive Acceptance Status Elecsys® 2010 - Host Interface
19. Receive Acceptance StatusWhen does Elecsys accept messages from host? (Disk Type)
Instrument StatusM
essa
ges
from
Hos
t
Initi
aliz
e
Sta
ndB
y
Pre
Ope
ratio
n
Ope
ratio
n
S.S
top-
S.S
can
Sam
ple
Sto
p
Par
tial S
top
Fina
lizat
ion
Sto
p
Em
erge
ncy
Sto
p
Sle
ep
Sam
ple
Sca
n
Rea
gent
Sca
n
Car
d S
can
T/M
Test Order ü ü ü ü ü ü ü ü ü ü ü üSubstance Data ü üCalibrator Parameters ü üControl Parameters ü üTest Conditions ü üInstrument Configuration ü üAction Request(START) ü üAction Request(STOP) ü ü ü ü ü ü ü ü ü
When does Elecsys accept messages from host? (Rack Type)
Instrument Status
Mes
sage
s fro
m H
ost
Initi
aliz
e
Sta
ndB
y
Pre
Ope
ratio
n
Ope
ratio
n
Fina
lizat
ion
Sto
p
Em
erge
ncy
Sto
p
Sle
ep
Rea
gent
Sca
n
Car
d S
can
T/M
Rac
k S
uppl
y S
top
Line
Sto
p
Ana
lyze
r Sto
p
A.S
top&
R.S
top
A.S
top&
L.S
top
Line
Onl
y
Test Order ü ü ü ü ü ü ü ü ü ü ü ü ü üSubstance Data ü ü üCalibrator Parameters ü ü üControl Parameters ü ü üTest Conditions ü ü üInstrument Configuration ü ü üAction Request(START) üAction Request(STOP) ü ü ü ü ü ü ü ü ü ü ü
Data Alarm List 91
Elecsys® 2010 - Host Interface Data Alarm List
20. Data Alarm List
Data Alarm Flag Level Result SystemBlock
Explanation
0 Yes No No flag is attached to theresult
Power failure-operation stopped 1 1 No Yes Determination has beencanceled due to power fail orpower off.
E. Stop-operation stopped 2 2 No Yes Determination has beencanceled due toEMERGENCY STOP.
Stop-operation stopped 3 3 No Yes Determination has beencanceled due to STOP.
P. Stop-operation stopped 4 4 No Yes Determination has beencanceled due to PARTIALSTOP.
S. Stop-operation stopped 5 4 No Yes Determination has beencanceled due to SAMPLESTOP.
Error handling-operation stopped 6 5 No Yes Determination has beencanceled due to recoveryprocess.
Sample Short 7 6 No Yes The volume of sample isinsufficient.
Assay Reagent Short 8 6 No Yes The volume of assay reagentis insufficient.
Diluent Short 9 6 No Yes The volume of diluent isinsufficient.
Pretreatment Reagent Short 10 6 No Yes The volume of pretreatmentreagent is insufficient.
Diluted Sample Short 11 6 No Yes The volume of diluted sampleis insufficient.
Reagent disk temperature out ofrange
12 7 Yes No Reagent disk temperature isout of range.
Incubator temperature out ofrange
13 7 Yes No Incubator temperature is outof range.
Measuring cell temperature out ofrange
14 7 Yes No Measuring Cell temperature isout of range.
PC/CC temperature out of range 15 7 Yes No PC/CC temperature is out ofrange.
PC/CC short 16 7 Yes No The volume of PC/CC isinsufficient.
ADC data abnormal 17 7 No Yes ADC data abnormal.Peak out of range 18 6 No Yes The ratio of the maximum
signal to the minimum signalis out of range.
Blank ratio out of range 19 6 No Yes The ratio of the maximumblank signal to the minimumblank signal is out of range.
Signal peak position out of range 20 6 No Yes The position of the maximumsignal is out of range.
92 Data Alarm List
Data Alarm List Elecsys® 2010 - Host Interface
Data Alarm Flag Level Result SystemBlock
Explanation
Signal peak width out of range 21 6 No Yes Peak width is out of range.Signal out of range 22 6 No Yes Signal is out of range.Concentration out of expectedvalue range
23 6 --- --- Concentration is out of range.
Calculation Error 24 6 No Yes Internal calculation Error hasoccurred
No Calibration Data 25 6 No Yes During calibration is in processand no calibration data isavailable, this flag is attachedto all data.
Previous calibration used 26 8 Yes No Previous calibration data isused for this result calculation.
System blocked 27 --- --- The result is blocked by thesystem.
User blocked 28 --- --- The result is blocked by theuser.
User released 29 --- --- The result is released by theuser.
Premature LLD signal-SampleDisk
30 6 No Yes Hovering is detected atsample pipetting
Premature LLD signal-ReagentDisk (Assay reagent)
31 6 No Yes Hovering is detected at assayreagent pipetting
Premature LLD signal-ReagentDisk (Diluent reagent)
32 6 No Yes Hovering is detected at diluentpipetting
Premature LLD signal-ReagentDisk (Pretreatment reagent)
33 6 No Yes Hovering is detected atpretreatment pipetting
<reserve> 34 --- ---Assay reagent bubble detected 35 6 No Yes Film is detected at assay
reagent pipettingDiluent bubble detected 36 6 No Yes Film is detected at diluent
pipettingPretreatment bubble detected 37 6 No Yes Film is detected at
pretreatment pipetting<reserve> 38 --- ---<reserve> 39 --- ---PC level out of range 40 6 Yes No AB level is out of rangePC level check failed 41 6 Yes No AB level check failedMeasuring cell current out ofrange
42 6 Yes No Current is out of range
Measuring cell current checkfailed
43 6 Yes No Current check failed
PC/CC temperature unstable 44 6 Yes No AB/CC temperature isunstable
Abnormal aspiration 45 6 No Yes Clot is detected at samplepipetting
Potential carry over 46 6 Yes No Signal is lower than acalculated limit respective tothe prior measurement (carryover is possible)
Bar code sample ID error 47 6 No Yes During sample scan beforepipetting the sample ID couldnot be recognized
Data Alarm List 93
Elecsys® 2010 - Host Interface Data Alarm List
Data Alarm Flag Level Result SystemBlock
Explanation
Below expected value range 48 6 Yes No Below normal (expected)range
Above expected value range 49 6 Yes No Above normal (expected)range
Below measuring range 50 6 Yes No Below measuring rangeAbove measuring range 51 6 Yes No Above measuring rangeExpired Rack Pack 52 8 Yes No Expired Rack Pack
94 Index
Index Elecsys® 2010 - Host Interface
21. Index
Action Confirmation Record [M-AC Record] .......................................................................................................87Action Request Record [M-AR Record]................................................................................................................86
Calibration Result [M-CR Record] ........................................................................................................................55Calibrator Parameters Record [M-CP Record] ................................................................................................46, 47Comment Record [C Record] ................................................................................................................................25Control Parameters Record [M-QP Record] ....................................................................................................51, 52
Diluent Parameters Record [M-DP Record] ..........................................................................................................40
Instrument Configuration Record [M-IC Record]............................................................................................76, 77Instrument Status Record [M-IS Record]...............................................................................................................62
Log File Record [M-LF Record]......................................................................................................................65, 66
Patient Information Record [P Record] .......................................................................................................7, 10, 21Processing Messages Record [M-PM Record] ......................................................................................................84
Raw Result Record [M-RR Record] ......................................................................................................................25Request Information Record [Q Record] .....................................................................................................6, 10, 26Result Context Record [M-RC Record] .................................................................................................................26Result Record [R Record]......................................................................................................................................24
Sample Status Record [M-SS Record].............................................................................................................58, 59Service Data Record [M-SD Record] ..............................................................................................................82, 83Substance Data Record [M-TC Record] ................................................................................................................70Substance Data Record [M-XT Record]..................................................................................38, 40, 45, 46, 50, 51
Test Application Record [M-TA Record]..............................................................................................................74Test Order Record [O Record].....................................................................................................................8, 12, 23