  • The absolute measurement of the Photo Detection Efficiency (PDE) of a SiPM over the whole visible range using continuous light is not a straightforward task as it requires the accurate determination of the detector gain. We have developed at LAL a process to achieve an absolute PDE with an accuracy of 5 %.

    Conclusion

    • 5% absolutePDE accuracy @ all Vbias and (400 – 700 nm)

    • the SiPM gain is determined from the “current” PDE & “counting” PDE comparison need of a very well calibrated test bench

    • the direct gain measurement in dark conditions gives errors of 10 to 20 % need of detailed analysis of secondary effects (after-

    pulses, crosstalk) and dark noise distribution (ongoing study in collaboration with Fermilab).

    MPPC absolute PDE measurement with continuous and pulsed light

    Vincent Chaumat, Cyril Bazin, Nicoleta Dinu, Véronique Puill, Jean-François Vagnucci

    Research & Study Detector Group

    PHOTODET 2012

    LAL, Orsay, June

    IPMT PMT photo generated current (A), RSPMT PMT anode sensibility (A/W), GPMT PMT gain, SPMT PMT illuminated surface (mm²)

    Current method • continuous monochromatic source • incident flux